Compression Therapy Concepts VP 500: troubleshooting and error codes
The ServiceForge knowledge base holds 3 documented symptoms for the Compression Therapy Concepts VP 500, with 12 ranked causes, 12 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Low pressure light and/or audible alarm activates patient safety
Low pressure indicator LED illuminates and/or audible alarm sounds during operation; indicates insufficient pressure at garment or system failure
Ranked causes
- 1. Loose or disconnected air tubing connections (common)
- 2. Defective garment (occasional)
- 3. Bellows/pressure control failure (occasional)
- 4. Timing motor malfunction (occasional)
- 5. Microswitch failure (occasional)
Verify all tube connections secure
- Test point
- Air tubing from pump to garment
- Expected
- All connections seated firmly; no leaks or disconnections
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump functioning but no air output
Pump runs but produces no pressurized air at outlet ports despite motor and compressor operation
Ranked causes
- 1. Bellows/diaphragm failure in compressor (occasional)
- 2. Internal air tubing loose or disconnected (common)
- 3. Compressor coil failure (rare)
- 4. Timing motor malfunction (occasional)
Verify all tube connections
- Test point
- Internal air tubing and compressor outlet
- Expected
- All tubes seated firmly and not obstructed
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump not functioning
Pump fails to operate; no motor activation or compressor action
Ranked causes
- 1. Unit not plugged in or power cord defective (common)
- 2. PC board malfunction (occasional)
- 3. Fuses loose or burnt out (common)
Verify power connection
- Test point
- Power plug and outlet
- Expected
- Plug inserted firmly into working outlet
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI