Codman CMC-III: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the Codman CMC-III, with 18 ranked causes, 18 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Low power output
Generator produces less cutting or coagulation power than expected
Ranked causes
- 1. Low line voltage (common)
- 2. Incorrect test load (common)
- 3. Internal calibration change (rare)
Measure supply line voltage
- Test point
- Power supply input
- Expected
- 120V, 220V or 240V ±10%
- Tools
- DMM
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not operate high voltage
Power-on indicator light is off or non-functional
Ranked causes
- 1. Blown fuse (common)
- 2. Loose wire at power cord plug (common)
- 3. No power at wall outlet (occasional)
Verify fuse status
- Test point
- Main power fuse and internal fuses F400/F402
- Expected
- Fuses intact and rated correctly for supply voltage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Output power high voltage
Adjust target DAC calibration values for any of 40 power output settings to match desired output power into 50 or 400 ohm loads
Ranked causes
- 1. Factory DAC calibration values require adjustment (common)
- 2. EEPROM error or blank memory at first power-up (occasional)
Enter calibration mode and select power setting
- Test point
- P203 pin 4
- Expected
- Unit announces 'calibrate output power'; Malis Unit setting and DAC value display on numeric LED
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erratic power output
Power output fluctuates or is inconsistent during operation
Ranked causes
- 1. Loose or dirty connections between forceps cord and jacks (common)
- 2. Intermittent break in forceps cord (common)
Inspect bipolar cord plugs and jacks
- Test point
- Forceps cord connectors and generator output jacks
- Expected
- No visible oxidation, corrosion or looseness
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive leakage current high voltage
Unit shows excessive leakage current to ground or chassis
Ranked causes
- 1. Shorted output transformer (rare)
Measure leakage current to ground
- Test point
- Generator chassis to ground
- Expected
- Within safe limits per IEC or applicable standard
- Tools
- Leakage current meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive power output patient safety
Generator produces more power than set or expected
Ranked causes
- 1. Internal calibration change (rare)
Measure output power at known setting
- Test point
- Generator output with standard test load
- Expected
- Power output matches front panel setting within tolerance
- Tools
- Power meter or test load with measurement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power output
Generator produces no cutting or coagulation power
Ranked causes
- 1. Broken wire in forceps cord (common)
Test forceps cord continuity
- Test point
- Forceps cord pins and connectors
- Expected
- Continuity present on both conductors
- Tools
- DMM ohms function
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voice indicator announces Error
Voice alert announces error condition
Ranked causes
- 1. Two front panel setting controls pressed simultaneously (occasional)
Verify single control operation
- Test point
- Front panel controls
- Expected
- Only one control pressed at a time
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voice indicator announces Footpedal error
Voice alert announces footpedal error
Ranked causes
- 1. Two footpedals pressed simultaneously (occasional)
Verify single footpedal operation
- Test point
- Pneumatic and electric footpedals
- Expected
- Only one pedal activated at a time
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voice indicator announces Internal fuse high voltage
Voice alert announces internal fuse blown; power setting display flashes 00
Ranked causes
- 1. Blown internal fuse (common)
Identify internal fuse status and type
- Test point
- Fuses F400 and F402 on power supply board
- Expected
- Fuses intact and rated for current supply voltage (110/120V, 220V, or 240V)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voice indicator announces Internal power error high voltage
Voice alert announces internal power error condition
Ranked causes
- 1. Operating malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voice indicator announces Memory error
Voice alert announces memory error during operation
Ranked causes
- 1. Operating malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.