Ultrasound therapy systems · Intelect / Genisys family

Chattanooga 2776: troubleshooting and error codes

The ServiceForge knowledge base holds 17 documented symptoms for the Chattanooga 2776, with 51 ranked causes, 18 diagnostic checks with test points and expected values, and 14 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Ultrasound duty cycle output incorrect

Power meter readings fall outside specified ranges for 10%, 20%, 50%, or 100% duty cycle settings at 1 MHz or 3.3 MHz frequency

Ranked causes

  1. 1. Bad or contaminated degassed water in power meter (common)
  2. 2. Wrong power meter model used (occasional)
  3. 3. Applicator out of calibration or defective (occasional)
  4. 4. Ultrasound board defective (rare)
  5. 5. Control board defective (rare)
First check

Verify degassed water is in power meter reservoir

Test point
Power meter reservoir
Expected
Clear degassed water, no aeration bubbles visible

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 201: Applicator calibration save error

System fails to save calibration data to applicator EEPROM during calibration procedure.

Ranked causes

  1. 1. Bad Applicator (common)
  2. 2. Bad Applicator Cable (common)
  3. 3. Applicator PCB Connector Bad (occasional)
  4. 4. Bad Ultrasound Board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 202: General applicator calibration error

System encounters unspecified error while attempting to calibrate an applicator.

Ranked causes

  1. 1. Bad Applicator (common)
  2. 2. Bad Applicator Cable (common)
  3. 3. Applicator PCB Connector Bad (occasional)
  4. 4. Bad Ultrasound Board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 203: Applicator EEPROM save timeout

System timeout while attempting to save calibration data to applicator EEPROM.

Ranked causes

  1. 1. Bad Applicator (common)
  2. 2. Bad Applicator Cable (common)
  3. 3. Applicator PCB Connector Bad (occasional)
  4. 4. Bad Ultrasound Board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ultrasound applicator output out of range

Power meter readings fall outside specified ranges for applicator at given settings.

Ranked causes

  1. 1. Bad degassed water in power meter (common)
  2. 2. Wrong power meter used (occasional)
  3. 3. Bad or out of calibration applicator (occasional)
  4. 4. Bad Ultrasound Board (occasional)
First check

Set up power meter per manual

Test point
Power meter reservoir
Expected
Applicator submerged in degassed water, centered over SS cone

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Applicator not calibrated (NO CAL displayed)

Display shows 'NO CAL' icon after applicator is connected.

Ranked causes

  1. 1. Applicator not calibrated (common)
  2. 2. Bad Applicator (occasional)
  3. 3. Bad Ultrasound PCB (occasional)
First check

Connect applicator and check display

Test point
Display center
Expected
Applicator detected icon without NO CAL icon

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 300: Critical ultrasound board error

System detects critical ultrasound board fault of unknown origin.

Ranked causes

  1. 1. Ultrasound Board to Control Board header loose or disconnected (common)
  2. 2. Bad Ultrasound Board (occasional)
  3. 3. Bad Control Board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 301: Ultrasound board not detected

System cannot detect presence of ultrasound board during initialization.

Ranked causes

  1. 1. Ultrasound Board to Control Board header loose or disconnected (common)
  2. 2. Bad Ultrasound Board (occasional)
  3. 3. Bad Control Board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 302: Ultrasound board reset

Ultrasound board has unexpectedly reset for unknown reason.

Ranked causes

  1. 1. Ultrasound Board to Control Board header loose or disconnected (common)
  2. 2. Bad Ultrasound Board (occasional)
  3. 3. Bad Control Board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 303: Ultrasound board read error

System error reading data from ultrasound board.

Ranked causes

  1. 1. Ultrasound Board to Control Board header loose or disconnected (common)
  2. 2. Bad Ultrasound Board (occasional)
  3. 3. Bad Control Board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 304: Ultrasound board write error

System error writing data to ultrasound board.

Ranked causes

  1. 1. Ultrasound Board to Control Board header loose or disconnected (common)
  2. 2. Bad Ultrasound Board (occasional)
  3. 3. Bad Control Board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 305: Applicator calibration error

System error encountered during applicator calibration procedure.

Ranked causes

  1. 1. Ultrasound Board to Control Board header loose or disconnected (common)
  2. 2. Bad Ultrasound Board (occasional)
  3. 3. Bad Control Board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan not blowing outward caution

Fan either does not blow air or blows inward instead of outward from unit rear.

Ranked causes

  1. 1. Fan wired wrong (occasional)
  2. 2. Bad Fan (occasional)
  3. 3. Bad Power Supply (occasional)
First check

Verify fan direction at unit rear

Test point
Unit rear outlet
Expected
Air blowing outward

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Applicator not detected

Display shows 'Applicator not detected' icon after applicator is plugged in.

Ranked causes

  1. 1. Bad Applicator (common)
  2. 2. Bad Ultrasound Board (occasional)
First check

Power on unit without applicator

Test point
Display center
Expected
Applicator not detected icon appears

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main menu does not display

Power button LED present but main menu fails to appear on display after startup.

Ranked causes

  1. 1. Bad Control Board (common)
  2. 2. Bad Power Supply (occasional)
First check

Check power button LED status

Test point
Power button LED
Expected
LED flashing

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 200: EEPROM access error

System displays error accessing internal EEPROM for configuration and protocol storage.

Ranked causes

  1. 1. Bad Control Board (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not start

System does not power on or respond to power button activation.

Ranked causes

  1. 1. Bad Power Supply (common)
First check

Attempt system startup

Test point
Power button
Expected
Unit powers on with beep tones

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
100WARN_US_UNPLUGGED: Technician attempted to calibrate ultrasound applicator but none is plugged into systemCalibration initiated without applicator connectedwarning
101WARN_US_APPL_BECAME_UNPLUGGED: Applicator came unplugged from system during therapyApplicator disconnection detected while therapy is activewarning
200Error accessing internal EEPROMSystem configuration or protocol read/write failureerror
201Error saving calibration data to applicator EEPROMApplicator calibration data save failureerror
202General applicator calibration errorAny error during applicator calibration procedureerror
203Applicator EEPROM save timeoutCalibration data save exceeds time limiterror
300Critical ultrasound board errorUltrasound board malfunction of unknown origincritical
301Ultrasound board not detectedUltrasound board not responding during initializationcritical
302Ultrasound board resetUltrasound board unexpected reset eventcritical
303Ultrasound board read errorFailed to read data from ultrasound boardcritical
304Ultrasound board write errorFailed to write data to ultrasound boardcritical
305Applicator calibration errorError during calibration procedure executioncritical
Icon- No applicator plugged into systemNo applicator is connected to the systemApplicator connection issue detected during operation or calibrationwarning
Icon- Sound Head overtempSound Head has overheated to a temperature that could cause damage to the crystalSound Head temperature exceeds safe limitwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI