CardioQuip MCH: troubleshooting and error codes
The ServiceForge knowledge base holds 70 documented symptoms for the CardioQuip MCH, with 84 ranked causes, 107 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No water flow during prime cycle
Prime button activated but pump does not turn on or valve does not switch, or water flow bursts not observed every ten seconds
Ranked causes
- 1. Kinks or closed valves blocking water flow in hose loop (common)
- 2. Insufficient water in ice tank (occasional)
- 3. Pump or valve malfunction (rare)
Inspect hose connections and water level in ice tank
- Test point
- ice tank, attached hoses
- Expected
- No kinks or closed valves; water level adequate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water leaks from unit caution
Water accumulation observed on or under the MCH unit, indicating internal or external water connection leaks.
Ranked causes
- 1. Loose internal water connections (occasional)
- 2. Damaged external water connectors (occasional)
- 3. Internal water system leak (rare)
Inspect external water connectors for leaks and physical damage
- Test point
- Water inlet, water outlet, drain connectors on exterior
- Expected
- No visible leaks, no physical damage, connectors intact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet or controls physically damaged caution
Dents or other visible physical damage to external cabinet and control surfaces
Ranked causes
- 1. Shipping or handling damage (occasional)
- 2. Impact during transport or storage (occasional)
Visually inspect external cabinet and all controls
- Test point
- Cabinet exterior and control panels
- Expected
- No dents or visible damage
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect power outlet or electrical connection high voltage
MCH connected to unsuitable mains supply without proper grounding or amperage
Ranked causes
- 1. Not using hospital grade grounded 20A outlet (common)
- 2. Mains supply lacks protective earth (occasional)
Verify outlet is dedicated 20A hospital grade
- Test point
- Mains outlet label and receptacle
- Expected
- 20 Ampere AC hospital grade designation visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal cabinet water leaks or loose connections caution
Water leaks, loose water connections, or visible damage found inside cabinet during service inspection
Ranked causes
- 1. Loose water connection fittings or tubing (occasional)
- 2. Damaged internal water system component (occasional)
Qualified service personnel to remove side panels and inspect interior
- Test point
- All internal water connections and cabinet interior
- Expected
- No water leaks, all connections tight, no visible damage
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal wiring damage or loose connections high voltage
Exposed wires, loose electrical connections, or visible wiring damage found during service inspection
Ranked causes
- 1. Loose wiring connections from vibration or improper handling (occasional)
- 2. Wire insulation damage from abrasion or impact (occasional)
Qualified service personnel to inspect all internal wiring
- Test point
- All internal wiring and connections
- Expected
- No exposed wires, all connections tight, no visible damage or abrasion
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output water temperature not within ±0.5°C of setpoint patient safety
Temperature displayed on system does not match actual water output temperature within ±0.5°C tolerance at various setpoints
Ranked causes
- 1. Temperature sensor calibration error (common)
- 2. Insufficient ice in cold water tank for below-nominal setpoints (occasional)
Set system to various temperatures and allow stabilization
- Test point
- water output temperature port
- Expected
- Output temperature within ±0.5°C of displayed setpoint, ±0.5°C
- Tools
- Reliable temperature standard
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output water temperature not within setpoint tolerance patient safety
Measured output water temperature deviates from display setpoint by more than ±0.5°C across temperature scale
Ranked causes
- 1. Temperature sensor calibration drift (occasional)
- 2. Thermostat or heating/cooling element malfunction (rare)
Measure water output temperature at various setpoints
- Test point
- Water output
- Expected
- Within ± 0.5°C of display setpoint, ± 0.5°C
- Tools
- Reliable temperature standard
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overtemp alarm does not activate at 42.5°C patient safety
During alarm test, system fails to activate Overtemp alarm when water temperature reaches or exceeds 42.5°C setpoint
Ranked causes
- 1. Overtemp alarm threshold not set or sensor malfunction (occasional)
- 2. Alarm output driver failure (rare)
Stabilize system at 42°C on RUN screen
- Test point
- Display, water output temperature
- Expected
- Temperature stabilized at 42°C (±0.5°C), ± 0.5°C
- Tools
- Reliable temperature standard
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water connector leaks or damage caution
External water connectors show signs of leaks, physical damage, or water accumulation under unit
Ranked causes
- 1. Loose or damaged water connection (occasional)
- 2. Internal leak (occasional)
Visually inspect external water connectors
- Test point
- All water inlet and outlet connectors
- Expected
- No visible leaks, cracks, or physical damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water connector leaks or damage detected caution
Leaks, physical damage to external water connectors, or water accumulation on or under unit
Ranked causes
- 1. Loose or damaged water connection fittings (occasional)
- 2. Internal water system leak (occasional)
Visually inspect all external water connectors
- Test point
- All external water connection points
- Expected
- No leaks, no physical damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water leak detected caution
Water accumulation on or under unit, or visible leaks at external water connectors
Ranked causes
- 1. Damaged external water connectors (occasional)
- 2. Internal water connection failure (occasional)
Inspect external water connectors and connections
- Test point
- All external water inlet and outlet connectors
- Expected
- No visible leaks or water accumulation
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet fan not operational caution
Cabinet fan located under system is not running during power-on or normal operation
Ranked causes
- 1. Cabinet fan not running or blocked (unknown)
Turn on mains power switch
- Test point
- Front panel power switch
- Expected
- Switch in ON position, system initializes
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet or control physical damage caution
External cabinet or controls show dents or other visible damage
Ranked causes
- 1. Damage during shipping or handling (common)
Visually inspect cabinet exterior and all controls
- Test point
- Entire external surface
- Expected
- No dents or visible physical damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot achieve setpoint below nominal water temperature
System unable to cool output water below nominal tank water temperature even with ice in cold water tank
Ranked causes
- 1. Insufficient ice in cold water tank (common)
Verify ice level in cold water tank
- Test point
- Cold water tank
- Expected
- Adequate ice present for desired setpoint
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control Module positioning and accessibility caution
MCH positioned in manner that restricts quick access to mains power plug or difficult to unmount Control Module
Ranked causes
- 1. MCH mains plug location obscured or blocked (occasional)
Verify mains power plug is easily accessible
- Test point
- MCH mains plug location
- Expected
- Plug within immediate reach and unobstructed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control Module positioning restricts power access caution
MCH positioned without easy access to mains power plug preventing quick disconnection in emergency
Ranked causes
- 1. Improper equipment positioning (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control Module shipping damage caution
Visually inspect for scratched or cracked screen, dents, broken or exposed wires, bent pins after unpacking
Ranked causes
- 1. Shipping damage during transport (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control panel or articulated arm mechanical damage caution
Sharp objects used on switches or excessive tightening of clamp/arm lock causing mechanical damage
Ranked causes
- 1. Improper tool use or overtightening (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dry operation without water circulation caution
System operated without water causing equipment damage to pump and internal components
Ranked causes
- 1. System drained while in operating mode (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dry system operation damage caution
MCH operated without sufficient water or in dry-run mode
Ranked causes
- 1. System completely drained during operating mode (occasional)
Verify water level before operating MCH
- Test point
- Tank water level indicator or sight glass
- Expected
- Water level above minimum mark
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment damage from dry operation caution
MCH operated with insufficient water in cold water tank, risking pump and cooling system damage
Ranked causes
- 1. Water level in cold water tank drops below 4 inches (common)
Inspect water level in cold water tank
- Test point
- Cold water tank
- Expected
- At least 4 inches of water
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment damage from improper lifting caution
MCH lifted by drain or water connections causes structural damage to connectors and internal components.
Ranked causes
- 1. Lifting MCH by drain or water connections instead of base (common)
Inspect drain and water connectors for damage or looseness
- Test point
- Drain outlet, water inlet, water outlet connectors
- Expected
- Connectors secure, no visible damage, no leaks
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive bacteria buildup in tank
Water tank contains excessive bacteria requiring quarterly disinfection to maintain proper operation
Ranked causes
- 1. Normal bacteria accumulation over time (common)
Prime the system
- Test point
- Front panel Prime button and water tank
- Expected
- Pump runs at full speed, valve switches from COOLING to RECIRC approximately every 10 seconds, observe 10-second water flow bursts
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive control panel or articulated arm force caution
Control panel clamp or articulated arm lock overtightened, causing mechanical damage
Ranked causes
- 1. Excessive tightening force applied to clamp or lock (occasional)
Verify control panel clamp tension
- Test point
- Control panel clamp handle
- Expected
- Snug but easily removable by hand
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External cabinet or control damage caution
Dents, cracks, or other visible physical damage to external cabinet or control panels.
Ranked causes
- 1. Shipping or handling damage (occasional)
Inspect external cabinet and control panels for dents and physical damage
- Test point
- Entire exterior cabinet surface, control panel, handle
- Expected
- No visible dents, cracks, or damage; smooth finish
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External cabinet or controls damaged caution
Dents or other physical damage visible on external cabinet and controls
Ranked causes
- 1. Shipping damage or rough handling (occasional)
Visually inspect external cabinet and all controls
- Test point
- Cabinet exterior and control panel
- Expected
- No dents or physical damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hose water flow blocked
Kinks, closed valves, or obstructions in attached hoses preventing water circulation
Ranked causes
- 1. Kinked or obstructed hose (common)
Inspect attached hoses for kinks or blockages
- Test point
- All hose connections
- Expected
- No kinks, closed valves, or obstructions; clear water flow path
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper mains power connection high voltage
MCH not connected to hospital grade grounded outlet or non-protective earth supply creating electrical shock hazard
Ranked causes
- 1. Connection to non-grounded or inadequate outlet (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incompatible heat exchanger pressure rating patient safety
Heat exchanger or blanket with max pressure rating below 83 kPa connected to MCH
Ranked causes
- 1. Connected equipment rated below 83 kPa maximum pressure (occasional)
Check heat exchanger/blanket pressure rating
- Test point
- Equipment manufacturer label or documentation
- Expected
- Maximum pressure rating 83 kPa (12 psi) or higher
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal water leaks or loose connections caution
Internal water connections are loose, leaking, or showing visible damage inside cabinet
Ranked causes
- 1. Loose or deteriorated water connection (occasional)
Visually inspect all internal water lines and connectors
- Test point
- All internal water connections, hoses, and fittings
- Expected
- No leaks, drips, or loose connections observed; all fittings secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal wiring damaged or loose high voltage
Exposed wires, loose connections, or visible damage detected inside cabinet during inspection
Ranked causes
- 1. Vibration or handling damage during shipping (occasional)
Visual inspection of all internal wiring
- Test point
- All internal wiring harnesses and connections
- Expected
- No exposed wires, all connections tight and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage current exceeds 100µA high voltage
Electrical safety analyzer measures leakage current above 100µA under any operating condition (power on/off, normal/reverse polarity, grounded/ungrounded)
Ranked causes
- 1. Unspecified electrical fault or component degradation (unknown)
Measure leakage current with calibrated electrical safety analyzer
- Test point
- MCH unit ground and line connections
- Expected
- less than 100µA, <100µA
- Tools
- Calibrated electrical safety analyzer
Leakage current exceeds limits high voltage
Electrical leakage current measurement indicates unsafe levels or significant increase from baseline
Ranked causes
- 1. Electrical insulation degradation or fault in MCH unit (unknown)
Measure leakage current with power on, normal polarity, grounded
- Test point
- MCH unit terminals to ground
- Expected
- Less than 100µA, < 100µA
- Tools
- Calibrated electrical safety analyzer
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing or improperly affixed warning labels patient safety
Warning labels are not present or are not properly attached to the unit.
Ranked causes
- 1. Labels damaged or lost during shipping (rare)
Verify all warning labels are present and properly affixed
- Test point
- All label locations on MCH exterior
- Expected
- All warning labels present and securely affixed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output temperature does not match setpoint
Measured water output temperature deviates from setpoint setting beyond tolerance; displayed temperature differs from actual temperature
Ranked causes
- 1. Temperature sensor or control circuit calibration error (occasional)
Allow system to stabilize at setpoint
- Test point
- Display screen and temperature indicator
- Expected
- Temperature stable at setpoint for minimum stabilization period
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output water temperature exceeds maximum setpoint patient safety
Actual water output temperature exceeds the 42°C maximum setpoint limit beyond tolerance
Ranked causes
- 1. Heating control circuit overshoot or thermostat malfunction (occasional)
Measure output water temperature at 42°C setpoint
- Test point
- Water output
- Expected
- 42°C ± 0.5°C, ± 0.5°C
- Tools
- Reliable temperature standard
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overtemp alarm activates patient safety
OVERTEMP screen displayed, audible alarm sounds when water temperature exceeds safe limits
Ranked causes
- 1. Cold water tank temperature too high (above 42°C) (common)
Silence alarm and observe screen
- Test point
- OVERTEMP screen display
- Expected
- OVERTEMP screen remains visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overtemp Alarm does not activate patient safety
System fails to trigger Overtemp Alarm when water temperature exceeds setpoint; OVERTEMP screen does not appear, audible alarm does not sound, or pump does not stop
Ranked causes
- 1. Water temperature exceeds setpoint without alarm triggering (common)
Verify system is in RUN mode and temperature stabilized
- Test point
- Display screen and temperature control status
- Expected
- RUN screen visible, temperature at setpoint 42°C
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord damage high voltage
Power cord shows cuts, exposed wiring, bent pins, or other physical damage that creates electrical hazard.
Ranked causes
- 1. Shipping damage to power cord (occasional)
Inspect power cord for cuts, exposed wire, bent or missing pins
- Test point
- Entire length of power cord and plug
- Expected
- Cord intact, no cuts or exposed wires, all pins present and straight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord damage suspected high voltage
Cuts, exposed wire, bent or missing pins on power cord observed during inspection
Ranked causes
- 1. Power cord damage from shipping or handling (occasional)
Visually inspect entire length of power cord
- Test point
- Power cord entire length
- Expected
- No cuts, exposed wire, bent or missing pins
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord damaged or defective high voltage
Power cord shows cuts, exposed wire, bent or missing pins that could create electrical hazard
Ranked causes
- 1. Physical damage during shipping or handling (common)
Visually inspect power cord for damage
- Test point
- Entire length of power cord
- Expected
- No cuts, exposed wire, bent or missing pins
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord damaged or unsafe high voltage
Cuts, exposed wire, bent or missing pins visible on power cord
Ranked causes
- 1. Damage during shipping or handling (occasional)
Visually inspect entire length of power cord
- Test point
- Power cord
- Expected
- No cuts, exposed wire, bent or missing pins
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not turn on after priming
Prime button pressed but pump fails to activate; no water flow in ice tank or valve does not cycle
Ranked causes
- 1. Pump or valve solenoid not responding to prime command (unknown)
Verify water in ice tank before priming
- Test point
- Ice tank
- Expected
- Approximately three gallons (12 liters) for MCH-1000(i) or one gallon (4 liters) for MCH-1000(m)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump or valve not responding to Prime button
After pressing Prime button, pump does not turn on or valve does not cycle between cool and recirculation modes
Ranked causes
- 1. Prime button not functional or electrical issue (unknown)
Press Prime button and observe pump and valve response
- Test point
- Prime button, pump outlet, recirculation valve
- Expected
- Pump turns on, valve switches from cool to recirculation mode approximately every ten seconds with ten-second bursts of water flow in ice tank
Sharp object damage to control panel switches caution
Sharp objects used to adjust or select control panel switches
Ranked causes
- 1. Sharp objects used on control panel switches (occasional)
Inspect control panel switches for damage
- Test point
- Switch surfaces and edges
- Expected
- No scratches, gouges, or cracks visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Significant increase in leakage current high voltage
Leakage current shows marked increase from baseline measurement, indicating potential component degradation or developing fault
Ranked causes
- 1. Component deterioration or insulation breakdown (unknown)
System does not enter cooling mode
When temperature is set to 0°C from STANDBY, system fails to enter cooling mode; water does not flow into ice tank or temperature control status does not show COOLING
Ranked causes
- 1. Temperature control system not switching to cooling mode at 0°C setpoint (unknown)
Verify system in STANDBY mode
- Test point
- Display screen
- Expected
- STANDBY screen visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not enter cooling mode at 0°C setpoint
When temperature set to 0°C and Run button pressed, system fails to enter cooling mode or water does not flow into ice tank
Ranked causes
- 1. Temperature control circuit malfunction (unknown)
Set temperature to 0°C and press Run button
- Test point
- Display screen, water flow to ice tank
- Expected
- System enters cooling mode; water flows into ice tank; temperature control status reads COOLING
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
System does not enter heating mode
When temperature is set to 42°C, system fails to enter heating mode; no water flow in ice tank or temperature control status does not show HEATING
Ranked causes
- 1. Temperature control system not switching to heating mode at 42°C setpoint (unknown)
Set temperature setpoint to 42°C
- Test point
- Temperature control interface
- Expected
- 42°C displayed as setpoint
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not enter heating mode at 42°C setpoint
When temperature set to 42°C, system fails to enter heating mode or water continues flowing into ice tank
Ranked causes
- 1. Heating circuit or valve malfunction (unknown)
Set temperature to 42°C and observe system response
- Test point
- Display screen, ice tank water flow
- Expected
- System enters heating mode; no water flow in ice tank; temperature control status reads HEATING
System does not run through successful power-on test
Unit powers on but system test does not complete successfully; display does not confirm test completion
Ranked causes
- 1. System test failure, hardware malfunction, or initialization error (unknown)
Turn on mains power switch and observe boot sequence
- Test point
- Display screen
- Expected
- System initializes; display shows boot sequence or welcome screen
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System drainage during clinical operation patient safety
Equipment connected to MCH is drained, causing entire heat exchanger network to drain
Ranked causes
- 1. Draining connected equipment during operation (common)
Verify clinical operation status before draining
- Test point
- MCH operational mode indicator
- Expected
- MCH not in active clinical use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System drained during clinical operation patient safety
Heat exchanger network drained by draining attached equipment causing patient injury or system damage
Ranked causes
- 1. Draining attached equipment during operation (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
Mains power switch is ON but unit does not initialize or system test does not run
Ranked causes
- 1. No power at hospital-grade outlet or incorrect voltage/frequency (common)
Verify mains power switch is OFF before connecting
- Test point
- Front panel power switch
- Expected
- Switch in OFF position
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not run system test on power-up
After plugging in and turning on mains power switch, unit fails to run through successful system test
Ranked causes
- 1. Power not supplied or outlet mismatch (common)
Verify mains power switch is OFF before connecting
- Test point
- Front panel mains power switch
- Expected
- OFF position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit operates with insufficient water in tank caution
MCH running with less than 4 inches of water in cold water tank
Ranked causes
- 1. Inadequate water level in cold water tank during operation (unknown)
Verify water level in cold water tank before operation
- Test point
- Cold water tank
- Expected
- Minimum 4 inches of water depth
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit operates with insufficient water level caution
MCH runs with less than 4 inches of water in cold water tank
Ranked causes
- 1. Dry operation without minimum water level (rare)
Verify water level in cold water tank
- Test point
- Cold water tank
- Expected
- Water level at or above 4 inches
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning labels missing or improperly affixed patient safety
Warning labels not properly installed on equipment exterior
Ranked causes
- 1. Labels damaged or dislodged during shipping (occasional)
Verify all required warning labels present and secure
- Test point
- All label locations on unit exterior
- Expected
- All warning labels properly affixed and legible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning labels missing or not properly affixed patient safety
Warning labels not present or improperly attached to unit
Ranked causes
- 1. Damage during shipping or installation (rare)
Verify all warning labels present and securely affixed
- Test point
- All external cabinet surfaces
- Expected
- All warning labels present and properly affixed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Water hose flow blocked or obstructed
- Water level drops below 4 inches
- Water level in cold water tank drops below four inches during operation
- Water level in ice tank drops below 4 inches
- Water strainer contamination
- Water strainer contamination or damage
- Water strainers clogged or dirty
- Water tank contains solids or debris
- Water tank contains solids, debris, or foul odor
- Water tank contains solids, debris, or foul smell
10 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI