Cardinal Health AirLife nCPAP: troubleshooting and error codes
The ServiceForge knowledge base holds 71 documented symptoms for the Cardinal Health AirLife nCPAP, with 90 ranked causes, 81 diagnostic checks with test points and expected values, and 24 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power-on self-test: Insufficient or missing gas source pressure
Air and/or oxygen gas sources not connected or below minimum 41 psig
Ranked causes
- 1. Gas source not connected or below minimum pressure (common)
- 2. Clogged inlet filter screens (occasional)
- 3. Faulty or defective inlet pressure sensor (rare)
Verify gas sources are connected
- Test point
- air and oxygen inlet connectors
- Expected
- both sources connected securely
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging
Battery icon does not show lightning bolt indicating charging status after AC Adapter connected
Ranked causes
- 1. AC Adapter not connected or loose connection (common)
- 2. Defective battery or charging circuit (rare)
Verify AC Adapter is properly connected
- Test point
- Power connector on back of driver
- Expected
- AC Adapter fully seated; lightning bolt visible in battery icon
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bleed flow outside specification caution
Measured bleed flow rate not within acceptable range with circuit and generator connected
Ranked causes
- 1. Loose internal plumbing connections (common)
- 2. Kinked tubing restricting flow (occasional)
Measure bleed flow rate
- Test point
- Bleed flow outlet port on bottom housing
- Expected
- 0.7 to 3.0 LPM (acceptable range at inlet pressure 41-66 PSI), 0.7-3.0 LPM
- Tools
- Flow meter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
FiO2 delivery outside tolerance patient safety
Oxygen concentration delivered does not meet ±3% accuracy specification at set points
Ranked causes
- 1. Incomplete warm-up calibration (occasional)
- 2. Independent O2 sensor not calibrated (occasional)
Measure FiO2 at 21% setpoint
- Test point
- Driver LCD screen and patient outlet via independent analyzer
- Expected
- Within ±3% of setpoint (18-24%), ±3%
- Tools
- O2 analyzer, DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas source pressure out of range patient safety
Driver alarms and displays source fault when one gas source falls outside 41-66 psig range
Ranked causes
- 1. Medical air or oxygen supply pressure outside specified range (41-66 psig) (occasional)
- 2. Gas supply line not connected or switched to new source (common)
Verify gas source pressures
- Test point
- Medical air and oxygen supply regulators
- Expected
- Both sources regulated at 41-66 psig (nominal 50 psig)
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery power remaining caution
Audible and visual alarm with dual thresholds; Medium Priority at <10% (approx. 24 minutes remaining); High Priority at <5% (approx. 12 minutes); 2-second delay
Ranked causes
- 1. Battery charge depletes below 10% capacity (common)
- 2. Battery charge depletes below 5% capacity (common)
Monitor battery capacity display
- Test point
- Battery indicator on screen
- Expected
- Above 10% capacity
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low flow alarm activates patient safety
Low Flow alarm activates during operation, measured flow rate less than 3 lpm
Ranked causes
- 1. Occluded expiratory limb or kinked tubing (common)
- 2. Defective patient circuit or generator (occasional)
Measure flow rate at patient circuit connection
- Test point
- Patient Circuit Connection
- Expected
- Greater than or equal to 3 lpm
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure in patient circuit patient safety
Pressure 2 cm H2O below set value; audible alarm, yellow flashing warning LED, message bar
Ranked causes
- 1. Leak in tubes, circuit, nasal prongs or mask (common)
- 2. Insufficient supply gases (common)
Inspect circuit for leaks or disconnections
- Test point
- tubes, circuit, nasal prongs, mask connections
- Expected
- all connections sealed, no visible leaks
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient treatment mode: Air pressure below minimum during operation patient safety
Air source pressure dropped below 41 psig; user may remove device from service or select Oxygen Only mode (FiO2 goes to 100%)
Ranked causes
- 1. Air source pressure below minimum 41 psig (common)
- 2. Faulty inlet pressure sensor (rare)
Verify air supply pressure is at least 41 psig
- Test point
- air inlet pressure gauge or regulator
- Expected
- minimum 41 psig
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient treatment mode: Air pressure exceeds maximum during operation patient safety
Air source pressure exceeded 66 psig; user may remove device from service or select Oxygen Only mode (FiO2 goes to 100%)
Ranked causes
- 1. Air source pressure above maximum 66 psig (common)
- 2. Faulty inlet pressure sensor (rare)
Verify air supply pressure does not exceed 66 psig
- Test point
- air inlet pressure gauge or regulator
- Expected
- maximum 66 psig
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient treatment mode: Air source disconnected or depleted patient safety
Air source has become disconnected, tank run out, or hose/fittings leaking excessively; FiO2 drops below specification
Ranked causes
- 1. Air source disconnected, tank empty, or excessive leak (common)
- 2. Clogged inlet filter screen (occasional)
Verify air tank is connected and contains gas
- Test point
- air tank connection and regulator
- Expected
- connected securely, tank not empty
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient treatment mode: Oxygen pressure below minimum during operation patient safety
Oxygen source pressure dropped below 41 psig; user may remove device from service or select Air Only mode (FiO2 goes to 21%)
Ranked causes
- 1. Oxygen source pressure below minimum 41 psig (common)
- 2. Faulty inlet pressure sensor (rare)
Verify oxygen supply pressure is at least 41 psig
- Test point
- oxygen inlet pressure gauge or regulator
- Expected
- minimum 41 psig
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient treatment mode: Oxygen pressure exceeds maximum during operation patient safety
Oxygen source pressure exceeded 66 psig; user may remove device from service or select Air Only mode (FiO2 goes to 21%)
Ranked causes
- 1. Oxygen source pressure above maximum 66 psig (common)
- 2. Faulty inlet pressure sensor (rare)
Verify oxygen supply pressure does not exceed 66 psig
- Test point
- oxygen inlet pressure gauge or regulator
- Expected
- maximum 66 psig
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient treatment mode: Oxygen source disconnected or depleted patient safety
Oxygen source has become disconnected, tank run out, or hose/fittings leaking excessively; FiO2 drops below specification
Ranked causes
- 1. Oxygen source disconnected, tank empty, or excessive leak (common)
- 2. Clogged inlet filter screen (occasional)
Verify oxygen tank is connected and contains gas
- Test point
- oxygen tank connection and regulator
- Expected
- connected securely, tank not empty
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test: Excessive gas source pressure
Air and/or oxygen gas sources exceed maximum 66 psig
Ranked causes
- 1. Gas source pressure above maximum specification (common)
- 2. Faulty or defective inlet pressure sensor (rare)
Measure gas source inlet pressures
- Test point
- air and oxygen inlet pressure
- Expected
- 41–66 psig (maximum 66 psig)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test: Insufficient oxygen sensor bleed flow
Oxygen sensor bleed flow below 0.9 lpm at 50 psig; 2-point O2 sensor calibration criteria not met
Ranked causes
- 1. Occluded oxygen sensor bleed port or inlet filter screens (common)
- 2. Faulty or defective O2 sensor (rare)
Inspect and clean inlet filter screens and O2 sensor filters
- Test point
- inlet filters and oxygen sensor area
- Expected
- clean, unobstructed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test: Patient circuit pressure not cleared patient safety
Pressure elevated in patient circuit during start-up; patient pressure sensor failed zero calibration
Ranked causes
- 1. Proximal pressure line connected or pressure trapped during startup (common)
- 2. Faulty or defective patient pressure sensor (occasional)
Disconnect proximal pressure line and cycle power
- Test point
- proximal pressure line connection to device
- Expected
- disconnected, then device powered off and on
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on or LEDs do not illuminate
After powering on unit, LEDs fail to illuminate during power-on self test
Ranked causes
- 1. AC Adapter not properly connected or defective (common)
- 2. Incorrect AC Adapter type (occasional)
Verify AC Adapter is connected to driver power connector
- Test point
- Power connector on back of driver
- Expected
- AC Adapter fully seated and connected to properly grounded AC outlet
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air filter dirty or wet
Inlet air filter and water trap on back of driver appear soiled or contain moisture
Ranked causes
- 1. Normal accumulation of dirt or moisture in air filter (common)
Visually inspect inlet air filter and water trap
- Test point
- Back of driver
- Expected
- Filter clean and dry
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air pressure high patient safety
Audible alarm, red flashing fault LED, message bar
Ranked causes
- 1. Air supply pressure exceeds specification (common)
Verify air supply pressure does not exceed maximum
- Test point
- air inlet pressure gauge or regulator
- Expected
- 41–66 psig (maximum 66 psig)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air pressure low patient safety
Audible alarm, red flashing fault LED, message bar
Ranked causes
- 1. Air supply pressure below specification (common)
Verify air supply pressure is within specification
- Test point
- air inlet pressure gauge or regulator
- Expected
- 41–66 psig
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm mute and system startup delay patient safety
Alarms delayed 120 seconds after entering Patient Treatment Mode for patient setup; can be muted an additional 120 seconds by pressing Alarm Mute/Reset button or entering Standby Mode
Ranked causes
- 1. User needs additional time for patient setup beyond initial 120-second delay (common)
Observe alarm behavior after Patient Treatment Mode entry
- Test point
- System timer and alarm status
- Expected
- Alarms inactive for 120 seconds; Alarm Mute/Reset button responsive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge fail caution
Audible and visual alarm indicating battery is not charging properly; verified at startup and during operation; 5-second delay
Ranked causes
- 1. Battery fails to charge during normal charging operation (rare)
Observe battery charge indicator
- Test point
- Battery level display during charging
- Expected
- Battery capacity increasing while connected to charger
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge fault
Audible alarm, red flashing fault LED, message bar, red power LED, LCD indicator active
Ranked causes
- 1. Battery charging circuit malfunction or defective battery (common)
Verify AC adaptor is always connected during operation
- Test point
- AC adaptor connection
- Expected
- connected to power at all times
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charging not indicated
Lightning bolt icon, green LED, plug icon, or 'Charge' text not illuminated/displayed during AC power connection
Ranked causes
- 1. Charging system malfunction (unknown)
Observe charging indicators with AC power connected
- Test point
- Battery icon, LED at plug decal, LCD display
- Expected
- Lightning bolt icon illuminated, green LED lit, plug icon displayed, 'Charge' text shown
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery insufficient run time
Battery operates less than 4 hours on full charge, or low battery alarms do not activate
Ranked causes
- 1. Battery aged beyond 2 year service life (common)
Monitor battery run time and alarm activation
- Test point
- Battery icon and audible alarms
- Expected
- 4 hours total operation including low battery alarm periods; low battery alarms activate before shutdown
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low (critical level)
Approximately 12 minutes remaining; audible alarm, red flashing fault LED, message bar, red power LED
Ranked causes
- 1. Battery critically depleted on DC power (common)
Verify AC adaptor is connected
- Test point
- AC adaptor connection to device
- Expected
- connected to power source and device
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low (medium level)
Approximately 24 minutes remaining; audible alarm, yellow flashing warning LED, yellow power LED
Ranked causes
- 1. Battery discharging on DC power without AC charging (common)
Verify AC adaptor is connected to charger port
- Test point
- AC adaptor connection
- Expected
- connected and charging initiated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bleed flow not dropping to zero
Bleed flow continues when main power switch is turned off
Ranked causes
- 1. Valve malfunction or stuck solenoid (unknown)
Verify bleed flow stops on power-off
- Test point
- Bleed flow outlet port
- Expected
- Zero flow when power switch off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit disconnect detected patient safety
Audible alarm, yellow flashing warning LED, message bar; flow rate adjusted to target ±2 lpm
Ranked causes
- 1. Leak or disconnection in circuit, nasal prongs or mask (common)
Inspect circuit for leaks or disconnections
- Test point
- all tubing and interface connections
- Expected
- all connections sealed and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPAP pressure outside tolerance patient safety
Delivered CPAP pressure does not meet ±1 cm H2O accuracy at set levels
Ranked causes
- 1. Prongs occluded or connections not secure (common)
Measure CPAP pressure at 5 cm H2O
- Test point
- Driver LED bar graph, LCD display, and independent manometer
- Expected
- 5 ± 1 cm H2O (4-6 cm H2O), ±1 cm H2O
- Tools
- Manometer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevated CPAP pressure patient safety
Audible and visual Medium Priority alarm; indicates pressure is above target set level
Ranked causes
- 1. Measured CPAP pressure is 2 cm H2O above set level with 5-second duration (unknown)
Measure CPAP pressure
- Test point
- Pressure sensor
- Expected
- Within 2 cm H2O of Set CPAP level, ±2 cm H2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevated pressure in patient circuit patient safety
Pressure 2 cm H2O above set value; audible alarm, yellow flashing warning LED, message bar displayed
Ranked causes
- 1. Exhaust tube occluded or restricted (common)
Inspect exhaust tube for occlusions or restrictions
- Test point
- exhaust tube
- Expected
- clear, unobstructed flow
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive circuit leak detected patient safety
Leak 3 lpm over target flow rate or flow rate >15 lpm; audible alarm, yellow flashing warning LED, message bar
Ranked causes
- 1. Leak or disconnection in circuit, nasal prongs or mask (common)
Inspect circuit and interface for leaks
- Test point
- all connections and patient interface
- Expected
- no visible leaks, secure connections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive flow leak detected patient safety
Audible and visual Medium Priority alarm; flow exceeds target by 3 lpm or exceeds 15 lpm; 5-second delay
Ranked causes
- 1. Flow rate exceeds Target Flow by 3 lpm or absolute flow exceeds 15 lpm for 5 seconds (occasional)
Measure delivered flow rate
- Test point
- Flow sensor
- Expected
- Not exceed Target Flow + 3 lpm and not exceed 15 lpm absolute
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive patient leak during CPAP delivery patient safety
Large leak exists at patient interface that cannot be corrected by adjusting mask or prongs; inadequate CPAP pressure maintenance
Ranked causes
- 1. Improper patient interface fit or positioning (common)
Verify adequate CPAP pressure at patient
- Test point
- Pressure line connection at patient circuit
- Expected
- CPAP pressure matches Set value on main screen
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External power disconnected
Single audible burst, indicator on display, green power LED turns off
Ranked causes
- 1. AC adaptor unplugged or disconnected (common)
Verify AC adaptor is plugged in
- Test point
- AC adaptor connector
- Expected
- connected to power outlet and device
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fatal software fault patient safety
Continuous audible alarm, red fault LED, warning screen; driver ceases operation, no patient flow
Ranked causes
- 1. Critical software fault that prevents safe operation (rare)
Immediately disconnect patient from device
- Test point
- patient interface
- Expected
- patient safely disconnected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FiO2 limit exceeded patient safety
Audible alarm, yellow flashing warning LED, message bar displays 'FiO2 Limit'
Ranked causes
- 1. Supply gases insufficient or out of specification (common)
Verify air and oxygen supply presence and pressure
- Test point
- gas inlet connections
- Expected
- within 41–66 psig range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FiO2 warning patient safety
Audible alarm, red flashing fault LED, message bar says 'FiO2 Warning'; device continues CPAP and FiO2 delivery
Ranked causes
- 1. Supply gases insufficient or sensor drift out of range (common)
Verify gas supplies connected and in specification
- Test point
- air and oxygen inlet
- Expected
- 41–66 psig
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow or pressure sensor calibration failure patient safety
Audible and visual Sensor Calibration alarm; occurs during Power-On Self Test; device must be removed from service
Ranked causes
- 1. Flow sensor or pressure sensor fails calibration test during startup (rare)
Complete Power-On Self Test
- Test point
- Startup sensor calibration sequence
- Expected
- Both flow and pressure sensors calibrate successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow rate outside tolerance patient safety
Delivered flow rate does not meet ±2 LPM accuracy compared to independent flow meter
Ranked causes
- 1. Circuit or generator connection loose (common)
Measure flow rate at CPAP 5 cm H2O
- Test point
- Driver LCD display and independent flow meter
- Expected
- Within ±2 LPM of independent meter reading, ±2 LPM
- Tools
- Flow meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow/pressure sensor failure
Audible alarm, yellow flashing warning LED, Sensor Calibration Failure warning screen
Ranked causes
- 1. Flow or pressure sensor failed zero calibration (common)
Disconnect patient from entire circuit
- Test point
- patient interface connection
- Expected
- completely disconnected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse blown caution
3.0 Amp 250 V fuse on rear of unit has failed
Ranked causes
- 1. Electrical overload or component failure (unknown)
Verify fuse is blown
- Test point
- Fuse on rear of unit
- Expected
- Fuse element intact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure alarm not activating patient safety
No audible alarm or visual indication when exhaust port is occluded
Ranked causes
- 1. High pressure sensor or alarm circuit malfunction (rare)
Occlude exhaust port and observe alarm
- Test point
- Audible alarm and LCD display
- Expected
- High pressure alarm activates immediately with CPAP and flow dropping to zero
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure in patient circuit patient safety
Audible and visual High Priority alarm; pressure relief solenoid activates and patient delivery solenoid closes, reducing circuit pressure to zero; system reinitializes flow every 3 seconds
Ranked causes
- 1. Patient circuit pressure exceeds 12 cm H2O for 2 seconds or reaches 18 cm H2O (unknown)
Monitor circuit pressure during operation
- Test point
- Pressure sensor reading
- Expected
- Below 12 cm H2O, Normal operating range
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low air inlet pressure alarm not activating patient safety
Audible alarm and 'Air Pressure Low' message do not appear when air source is disconnected
Ranked causes
- 1. Air inlet pressure sensor malfunction (rare)
Observe alarm on air source disconnect
- Test point
- Audible alarm and LCD message bar
- Expected
- 'Air Pressure Low' message with high priority red flashing LED and audible alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low CPAP pressure patient safety
Audible and visual Medium Priority alarm; indicates pressure is below target set level with 14-second delay
Ranked causes
- 1. Measured CPAP pressure is 2 cm H2O below set level for 14 seconds (unknown)
Measure CPAP pressure
- Test point
- Pressure sensor
- Expected
- Within 2 cm H2O of Set CPAP level, ±2 cm H2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low flow delivery patient safety
Audible and visual High Priority alarm; flow rate at or below 3 lpm; 10-second delay
Ranked causes
- 1. Delivered flow is ≤3 lpm for 10 seconds (common)
Measure gas supply inlet pressures
- Test point
- Supply pressure inlet
- Expected
- Within specified gas supply range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low flow detected patient safety
Flow rate ≤3 lpm; audible alarm, red flashing fault LED, message bar
Ranked causes
- 1. Exhaust tube occluded or restricted (common)
Inspect exhaust tube for blockages
- Test point
- exhaust tube
- Expected
- clear, unobstructed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen inlet pressure alarm not activating patient safety
Audible alarm and 'Oxygen Pressure Low' message do not appear when oxygen source is disconnected
Ranked causes
- 1. Oxygen inlet pressure sensor malfunction (rare)
Observe alarm on oxygen source disconnect
- Test point
- Audible alarm and LCD message bar
- Expected
- 'Oxygen Pressure Low' message with high priority red flashing LED and audible alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure/disconnect alarm not activating patient safety
No audible alarm or visual indication when prongs are removed or circuit disconnected
Ranked causes
- 1. Pressure sensor or alarm circuit malfunction (rare)
Remove prong occlusion and observe alarm response
- Test point
- Audible alarm and LCD message bar
- Expected
- Medium priority audible alarm with yellow LED and 'Disconnect' or 'Low Pressure' text within 15 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mechanical high pressure relief not within specification patient safety
Pop-off pressure measured at patient outlet not within 211 ± 22 cm H2O
Ranked causes
- 1. High pressure relief valve setting out of adjustment (unknown)
Measure dead-head pressure at patient outlet
- Test point
- Patient outlet port with high range manometer
- Expected
- 211 ± 22 cm H2O (189-233 cm H2O), ±22 cm H2O
- Tools
- High range manometer (≥250 cm H2O capability)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Non-fatal software fault
Audible alarm, red flashing fault LED, message bar
Ranked causes
- 1. Software fault detected by watchdog that does not affect operation (unknown)
Note and record the displayed fault code
- Test point
- message bar or LCD display
- Expected
- specific fault code visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen Bleed Flow out of specification
Oxygen Bleed Flow measured at the Oxygen Sensor Bleed Port does not fall within acceptable range, indicating potential FiO2 accuracy issues
Ranked causes
- 1. Loose connections or kinked tubing in driver plumbing (common)
Measure Oxygen Bleed Flow at Oxygen Sensor Bleed Port on underside left-rear of driver
- Test point
- Oxygen Sensor Bleed Port
- Expected
- 0.9 to 3.0 lpm at 50 ± 1 psig inlet pressure; acceptable range 0.7 – 3.0 lpm at 41 – 66 psig, 0.7 – 3.0 lpm
- Tools
- Flow meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen concentration out of range patient safety
Audible Medium or High Priority alarm; FiO2 outside ±3% tolerance with variance-duration product exceeding threshold
Ranked causes
- 1. Variance from set FiO2 exceeds ±3% for duration where variance×seconds >20 (Medium) or >30 (High) (occasional)
Measure actual FiO2 output
- Test point
- Oxygen sensor
- Expected
- Set %O2 ±3%, ±3%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen pressure high patient safety
Audible alarm, red flashing fault LED, message bar
Ranked causes
- 1. Oxygen supply pressure exceeds specification (common)
Verify oxygen supply pressure does not exceed maximum
- Test point
- oxygen inlet pressure gauge or regulator
- Expected
- 41–66 psig (maximum 66 psig)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen pressure low patient safety
Audible alarm, red flashing fault LED, message bar
Ranked causes
- 1. Oxygen supply pressure below specification (common)
Verify oxygen supply pressure is within specification
- Test point
- oxygen inlet pressure gauge or regulator
- Expected
- 41–66 psig
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor calibration error patient safety
Audible alarm, red flashing fault LED, Oxygen Calibration Error warning screen; device provides CPAP and FiO2 but FiO2 may not be accurate
Ranked causes
- 1. Oxygen sensor malfunction or calibration criteria not met (common)
Remove patient and power cycle device
- Test point
- power button
- Expected
- device powers off then restarts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient circuit disconnect or loose connection patient safety
Audible and visual Medium Priority alarm; detects pressure line or patient interface disconnect; 5-second delay
Ranked causes
- 1. Flow and pressure conditions meet alarm thresholds simultaneously (e.g., flow ≥6 lpm AND CPAP ≤2 cm H2O for 5 seconds) (common)
Verify patient interface connection
- Test point
- Patient interface attachment point
- Expected
- Securely seated with no visible gaps
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Patient treatment mode warning: Non-fatal software fault detected
- Patient treatment mode warning: Oxygen sensor drift calibration failure
- Patient treatment mode warning: Running on DC power beyond 4 hours
- Patient treatment mode: High FiO2 warning for prolonged delivery
- Power loss not indicated
- Power-on self-test: Flow sensor failure
- Power-on self-test: Oxygen sensor calibration failure
- Power-on self-test: Source valve or delivery valve failure
- Software fault detected
- Standby Mode operation
- Supply gas pressure out of range
11 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Air Pressure High | Air inlet pressure exceeded maximum 66 psig | Supply pressure rises above specification during operation | critical |
Air Pressure Low | Air inlet pressure dropped below minimum 41 psig | Supply pressure falls out of specification during operation | critical |
Alarm Mute/Reset button function | Resets alarm state after High Priority condition clears; provides additional 120-second mute window during patient setup | User presses Alarm Mute/Reset button after High Priority alarm or during patient setup initialization | info |
Battery Charge Fault | Battery charging circuit has failed; device cannot charge battery | Detected during power-on self-test or during operation | critical |
Battery Low Dual Stage Alarm (~12 minutes remaining) | Battery charge level critically depleted with minimal time before shutdown | Continued battery operation without AC input after initial warning | critical |
Battery Low Dual Stage Alarm (~24 minutes remaining) | Battery charge level critically low with medium remaining time | Device operating on battery power for extended period without AC input | warning |
Disconnect (Table Based) | Circuit disconnect detected by pressure or flow monitoring | Tubing or interface disconnection develops during operation | warning |
Elevated Pressure (2 cm H2O over Set Value) | Pressure in patient circuit 2 cm H2O above the set pressure | Exhaust tube occlusion or restriction develops during operation | warning |
External Power Disconnect | AC adaptor has been unplugged; device switched to battery | AC adaptor cable removed from power outlet or device connector | info |
Fatal Software Fault | Critical software error; device ceases operation immediately | Software failure detected by watchdog that prevents safe operation | critical |
FiO2 Limit | Oxygen concentration cannot be maintained within specified range | Oxygen supply fails or becomes insufficient during operation | warning |
FiO2 Warning | Oxygen concentration drifted outside specification; alarm active but therapy continues | Oxygen sensor calibration criteria not met; oxygen supply degradation | critical |
Flow/Pressure Sensor Failure | Flow or pressure sensor failed zero-point calibration during startup | Detected during power-on self-test; prevents normal operation | info |
High Pressure | Pressure in patient circuit exceeds safe limits | Patient circuit pressure rises above safe operating threshold during therapy | error |
High Pressure (≥12 cm H2O for 2 seconds or ≥18 cm H2O instan | Pressure in patient circuit critically elevated; safety release activated | Exhaust obstruction or system malfunction during patient therapy | critical |
High Priority alarm (flashing Red Fault LED) | High priority alarm condition detected; audible alarm active; Red Fault LED flashes | High priority alarm condition occurs (e.g., High Pressure, Low Flow, Software Fault, Oxygen Calibration, Source Pressure) | critical |
Leak | Excessive leak detected in patient circuit (>3 lpm above target) | Large leak or disconnection during patient therapy | warning |
Low Flow | Measured flow rate is less than 3 lpm | Flow rate detected at patient circuit drops below 3 lpm during operation | warning |
Low Pressure (2 cm H2O under Set Value) | Pressure in patient circuit 2 cm H2O below the set pressure | Leak in circuit or supply gas insufficient | warning |
Medium Priority alarm (Yellow Warning LED) | Medium priority alarm condition detected; audible alarm and Yellow Warning LED active | Medium priority alarm condition occurs (e.g., Elevated Pressure, Low Pressure, Disconnect, Leak, Low Battery) | warning |
Non-Fatal Software Fault | Software error detected by watchdog that does not prevent operation | Software anomaly detected during self-test or operation | error |
Oxygen Pressure High | Oxygen inlet pressure exceeded maximum 66 psig | Supply pressure rises above specification | critical |
Oxygen Pressure Low | Oxygen inlet pressure dropped below minimum 41 psig | Supply pressure falls out of specification during operation | critical |
Oxygen Sensor Calibration Error | Oxygen sensor failed 2-point calibration; device cannot verify FiO2 accuracy | Oxygen sensor malfunction detected during calibration or periodic check | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI