Cardiac Science Q-Stress: troubleshooting and error codes
The ServiceForge knowledge base holds 29 documented symptoms for the Cardiac Science Q-Stress, with 54 ranked causes, 9 diagnostic checks with test points and expected values, and 18 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power on
Blue and/or green LED fails to illuminate when unit is powered on
Ranked causes
- 1. Power cord not plugged into isolation transformer (common)
- 2. Power cord not connected to chart recorder (common)
- 3. Power switch in OFF position (common)
- 4. No power to outlet (occasional)
- 5. Bad fuse(s) (occasional)
- 6. Defective power cord (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
BP monitor data error
Blood pressure monitor fails to acquire or transmit data to system
Ranked causes
- 1. Incorrect device selected (occasional)
- 2. Incorrectly configured synchronization (occasional)
- 3. Noisy ECG waveforms (occasional)
- 4. Bad microphone or improper microphone position (occasional)
- 5. Problem with BP monitor setup or configuration (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No recorder output
Chart recorder does not produce output
Ranked causes
- 1. Power switch in OFF position (common)
- 2. Paper is out (common)
- 3. Interface cable disconnected (common)
- 4. Defective interface cable (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treadmill communication disconnected
Loss of communication link with treadmill due to power or connection issue
Ranked causes
- 1. Treadmill unplugged or low voltage condition (common)
- 2. Incorrect COM or USB port selection (occasional)
- 3. Treadmill power is off (common)
- 4. Treadmill interface cable not connected properly (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication error with BP monitor - monitor disconnected
System cannot communicate with BP monitor; device is disconnected or turned off
Ranked causes
- 1. Monitor turned off (common)
- 2. Monitor not connected to PC (common)
- 3. Monitor not connected to correct port on PC (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication error with BP monitor - no data available
BP monitor connected but not transmitting data to system
Ranked causes
- 1. Monitor turned off (common)
- 2. Monitor not connected to PC (common)
- 3. Monitor not connected to correct port on PC (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ergometer communication disconnected
System unable to communicate with ergometer device
Ranked causes
- 1. Ergometer interface cable disconnected or not properly connected (common)
- 2. Ergometer control mode not set up correctly (occasional)
- 3. Ergometer power is off (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preamp not responding - power or connection issue
Preamp unit is not responding to commands; power or cable connection problem
Ranked causes
- 1. No power to preamp (common)
- 2. Preamp not connected or bad connection (common)
- 3. Faulty cable (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance Test
Checks preamp gain, noise, and impedance levels for each ECG channel; requires impedance service appliance connection
Ranked causes
- 1. Out-of-specification gain or impedance (unknown)
- 2. Out-of-specification noise level (unknown)
Measure preamp gain and impedance per channel
- Test point
- Service dialog results display
- Expected
- Pass
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer connection or power issue
Printer not connected or powered; bad cable connection
Ranked causes
- 1. Printer power off or disconnected (common)
- 2. Faulty printer cable (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adobe Reader displays wrong language
Adobe Reader application shows interface text in incorrect language; user needs to change language settings
Ranked causes
- 1. Language preference not set at application startup (common)
Verify Application Language setting is correct
- Test point
- Adobe Reader Preferences > International > Application Language
- Expected
- Choose at application startup
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blanking of ECG data
Holes or missing data appear in ECG waveform display and printouts due to system overload
Ranked causes
- 1. PC system overload from multiple requests (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Leads-Off Test patient safety
Checks preamp leads-off detection capability for each electrode; requires impedance service appliance and manual lead disconnect/reconnect
Ranked causes
- 1. Leads-off detection malfunction (unknown)
Verify leads-off detection for each electrode
- Test point
- Service dialog results per selected lead
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Self Test
Internal preamp status check to verify basic hardware function; displays pass/fail results in dialog box
Ranked causes
- 1. Preamp internal error (unknown)
Verify preamp internal status via self-test
- Test point
- Service dialog Status box
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Treadmill speed and grade caution
Manual calibration procedure for TM and ST series treadmills to verify and adjust speed and grade accuracy
Ranked causes
- 1. Treadmill speed or grade requires calibration verification (unknown)
Start calibration procedure
- Test point
- Treadmill Calibration tab
- Expected
- Start button available
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Database connection lost
Setup screens (protocol, procedure, final report, system) have no items in lists; application disconnected from database
Ranked causes
- 1. Application lost connection to database (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC card self-test error
DC card reported self-test error with specified error code, slot, and firmware version
Ranked causes
- 1. Preamp failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dialog messages from network printer server
Print job dialog messages appear originating from a different system on network
Ranked causes
- 1. Network printer server sending remote print notifications (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG card not responding - power or connection issue
ECG card is not responding; power or connection to AcqPack compromised
Ranked causes
- 1. No power or bad connection to AcqPack (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG card self-test error
ECG card reported self-test error with error code, slot, and firmware version
Ranked causes
- 1. Preamp failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrode disconnected patient safety
Electrode lead has disconnected from patient; leads off condition detected
Ranked causes
- 1. Electrode disconnected from patient (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preamp communication error - repeated self-test failure
Error detected communicating with preamp unit; self-test check failed
Ranked causes
- 1. Preamp failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preamp self-test error - communication failure
Preamp unit failed self-test; communication error with acquisition hardware detected
Ranked causes
- 1. Preamp failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preamp unit self-test error
Preamp unit is reporting a self-test error condition
Ranked causes
- 1. Preamp failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer out of paper
Printer paper tray is empty; cannot complete print job
Ranked causes
- 1. Out of paper in printer tray (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer unavailable or not found
System cannot locate printer or printer is not responding
Ranked causes
- 1. PC not configured with default printer (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treadmill cable disconnected
Treadmill interface cable is not connected; no signal received from treadmill
Ranked causes
- 1. Cable disconnected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treadmill grade control error caution
Treadmill unable to achieve or maintain desired grade setting
Ranked causes
- 1. Treadmill not calibrated (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treadmill motor overheated caution
Treadmill motor has overheated and stopped operating
Ranked causes
- 1. Motor thermal protection activated (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
An error occurred while printing | Printer is unavailable or not found by system | Printer connection or configuration issue | error |
An error was detected communicating with the preamp unit | Communication failure with preamp hardware | Preamp not responding to system commands | error |
A software self test error has been detected | Preamp unit failed internal self-test; reboot required | Self-test failure during system startup or operation | error |
BP monitor data error | Blood pressure monitor cannot communicate or acquire data | Device disconnection, incorrect configuration, or connection issue | warning |
Communication error with the BP monitor – BP Monitor is disc | BP monitor is physically disconnected or powered off | Power loss, cable disconnection, or incorrect port assignment | error |
Communication error with the BP monitor – no data is availab | BP monitor connected but not transmitting readings to system | Data acquisition failure after connection established | error |
DC card reported self test error code = xxx, slot = yyy, fir | DC card self-test failed with specified error code | DC card hardware or firmware issue | error |
ECG card reported self test error code = xxx, slot = yyy, fi | ECG card self-test failed with specified error code | ECG card hardware or firmware issue | error |
Electrode [name] Disconnected | Electrode lead has disconnected from patient; leads off detected | Physical disconnection of electrode from patient skin | error |
If the Setup screens have no item in their lists | Application has lost connection to the database | Database connectivity failure | error |
The ECG card is not responding | ECG card not communicating with AcqPack | Power or connection issue to AcqPack | error |
The ergometer communication is disconnected | System cannot communicate with ergometer device | Connection or configuration failure | error |
The preamp is not responding | Preamp unit not communicating with system | Power loss or connection failure | error |
The preamp unit is reporting a self test error | Preamp internal self-test failure detected | Preamp self-test execution failure | error |
The treadmill cable is disconnected | Treadmill interface cable is not connected; no signal received | Physical cable disconnection | error |
Treadmill communication is disconnected | Communication link between treadmill and system is lost | Power loss, voltage drop, or connection failure | error |
Treadmill grade control error | Treadmill unable to achieve or maintain desired grade setting | Grade adjustment command failure | warning |
Treadmill motor over heated | Treadmill motor has overheated and thermal protection has activated | Excessive motor temperature during operation | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.