The ServiceForge knowledge base holds 22 documented symptoms for the Canon GCOM, with 22 ranked causes, 0 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

BCC error may occur on send caution

Exposure conditions sent to X-ray generator may be corrupted due to external noise

Ranked causes

  1. 1. Electronic noise interference on communication cables (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot convert received data to numerical value

DLL cannot convert received X-ray generator data to numerical values

Ranked causes

  1. 1. DLL does not support this X-ray generator (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication circuit not established

Cannot establish communications circuit with X-ray generator

Ranked causes

  1. 1. X-ray generator powered off or cables disconnected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

DLL cannot communicate with X-ray generator after CXDI restart

Communication fails between DLL and X-ray generator when CXDI is restarted while X-ray generator power is ON

Ranked causes

  1. 1. Communication link not re-established after CXDI restart (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

EOT timeout error

X-ray generator did not send EOT (end of transmission) signal

Ranked causes

  1. 1. X-ray generator powered off or cables disconnected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exposure condition mismatch between operation unit and X-ray generator patient safety

Displayed kV, mA, msec values differ between operation unit and X-ray generator before irradiation

Ranked causes

  1. 1. Exposure parameters not synchronized after DLL sends settings (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exposure conditions not displayed on OPU patient safety

OPU is not displaying exposure conditions needed for X-ray generator communication

Ranked causes

  1. 1. OPU settings not configured to display exposure conditions; OPU display mismatched with X-ray generator (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fatal error within DLL

Error has occurred within the DLL software component

Ranked causes

  1. 1. Internal DLL error (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

File not found error

Required file has been deleted, moved, or renamed

Ranked causes

  1. 1. File missing from D:\CCR directory (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

INI file settings incorrect

Dialog box appears indicating configuration errors in cgbig3.ini file at startup

Ranked causes

  1. 1. Invalid INI file parameters (file name, section name, text length, exposure conditions, serial port settings) (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Insufficient memory

Not enough system memory available

Ranked causes

  1. 1. Insufficient system RAM or memory leak (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OPU not functioning normally patient safety

CXDI (OPU component) is not working properly

Ranked causes

  1. 1. OPU hardware or software malfunction (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Text data timeout error

X-ray generator does not send data after send request (TM3)

Ranked causes

  1. 1. X-ray generator powered off or cables disconnected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unknown data received from X-ray generator

Undefined data received that DLL does not support

Ranked causes

  1. 1. DLL does not support this X-ray generator model (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Win32 API error within DLL

Windows API error has occurred within DLL

Ranked causes

  1. 1. Windows system error affecting DLL (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator communication error - received data conversion failure

Actual irradiation information received cannot be converted to data reportable to OPU

Ranked causes

  1. 1. Incomplete description in [ExpCondFromGen] section of cgbig3.ini (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator communication error - received data length mismatch (long)

Received irradiation data from X-ray generator is longer than specified exposure conditions

Ranked causes

  1. 1. Conflict between actual irradiation information and [ExpCondFromGen] section in cgbig3.ini (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator communication error - received data length mismatch (short)

Received irradiation data from X-ray generator is shorter than expected

Ranked causes

  1. 1. Conflict between actual irradiation information and [ExpCondFromGen] section description in cgbig3.ini (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator communication error - text string mismatch

DLL cannot create text string to send to X-ray generator during communication

Ranked causes

  1. 1. Conflict between [InitOpu] and [ExpCondToGen] sections in cgbig3.ini (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator no response to send data

X-ray generator does not respond to data sent (RT2)

Ranked causes

  1. 1. X-ray generator powered off or cables disconnected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator no response to send request

X-ray generator does not respond to send requests at startup (RT1)

Ranked causes

  1. 1. X-ray generator powered off or cables disconnected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator not ready to receive

X-ray generator responds negatively to send requests

Ranked causes

  1. 1. X-ray generator not ready for receiving (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.