Canon CXDI NE: troubleshooting and error codes

The ServiceForge knowledge base holds 82 documented symptoms for the Canon CXDI NE, with 94 ranked causes, 102 diagnostic checks with test points and expected values, and 208 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

70C image loss

Images from 70C sensor are lost during capture or transmission to monitor

Ranked causes

  1. 1. Outdated 70C firmware (common)
  2. 2. Outdated Cisco AP link firmware (common)
  3. 3. Wireless channel congestion or interference (occasional)
First check

Verify CXDI NE software version

Test point
System software
Expected
v1.30.2.3 or higher

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC calibration protocol creation radiation

Create or modify AEC calibration protocol when detector is added or protocol deleted from database

Ranked causes

  1. 1. Detector added without AEC calibration protocol (common)
  2. 2. AEC calibration protocol deleted from database (occasional)
  3. 3. Existing AEC calibration protocol requires modification (occasional)
First check

Verify detector is registered and workspace created

Test point
Protocol Editor Service Tool
Expected
Detector and workspace appear in system

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: detector QC functions radiation

QC tool provides calibration, performance test and self-diagnosis functions for detector verification

Ranked causes

  1. 1. Detector requires periodic QC maintenance (common)
  2. 2. Performance verification needed (common)
  3. 3. Self-diagnosis verification required (common)
First check

Access QC Tool from CXDI Control Software NE

Test point
Tool screen
Expected
QC Tool available and responsive

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Firmware update: Ferry type FPD detector

Update firmware or FPGA code for Ferry protocol connected detectors using Ferry Type FPD Maintenance Tool

Ranked causes

  1. 1. Firmware update required for detector (occasional)
  2. 2. FPGA configuration update needed (occasional)
  3. 3. Sensor information file update required (rare)
First check

Enter local IP address of control computer

Test point
Local IP Address field
Expected
192.168.100.10

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor log collection and analysis

Collect sensor and diagnostic logs using Collection Tool for troubleshooting

Ranked causes

  1. 1. Diagnostic logs required for troubleshooting (common)
  2. 2. Database backup required (common)
  3. 3. QC result data collection needed (occasional)
First check

Start Collection Tool from Service Tool screen

Test point
Service Tool menu
Expected
Collection Tool option available and clickable

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector issued error radiation

Detector reports error condition; operator action required before continuing

Ranked causes

  1. 1. Detector error condition triggered (unknown)
  2. 2. Persistent detector error after operation (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator disconnected or communication failed radiation

System detects loss of generator connection; QC operations cannot proceed

Ranked causes

  1. 1. Generator communication failure during QC initiation (common)
  2. 2. Generator condition settings failed (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

70C communication failures

Communication errors between detector and system with 70C sensor

Ranked causes

  1. 1. Software incompatibility with 70C sensor (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

70C images lost radiation

X-ray images from 70C sensor are not captured or saved

Ranked causes

  1. 1. Software issue in CXDI NE with 70C sensor (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

70C sensor disconnection

Sensor connection is lost or becomes intermittent during operation

Ranked causes

  1. 1. Wireless communication failure due to firmware or channel issues (common)
First check

Collect system event logs for analysis

Test point
Windows System and Application logs
Expected
Log files collected without errors

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blank images displayed on CXDI NE system patient safety

Windows Aero themes enabled causing blank images in digital radiography systems

Ranked causes

  1. 1. Windows Aero themes enabled in system configuration (common)
First check

Verify Windows Aero themes status in system settings

Test point
Personalize menu, Theme selection
Expected
Theme set to Basic or High Contrast (not Aero)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration data invalid or out of date radiation

Stored calibration data is incorrect or expired; system rejects compensation

Ranked causes

  1. 1. Calibration data not correct or stale (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration data not available at startup radiation

Error displayed when control software starts up without prior calibration; system enters QC-tool-only mode, preventing normal operation.

Ranked causes

  1. 1. Detector calibration not yet performed (common)
First check

Verify calibration has been performed since system installation

Test point
Control software startup status
Expected
System enters normal operation mode, not QC-tool-only mode

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration data upload failed radiation

SPCC data upload error during QC Tool operation; system restart required

Ranked causes

  1. 1. Ferry communication or power connection error during upload (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC calibration protocol creation and modification radiation

Creating or modifying AEC calibration protocols after detector addition or protocol deletion; configuring image processing parameters for AEC calibration

Ranked causes

  1. 1. AEC Calibration Protocol needs creation or modification (common)
First check

For Case 1: Verify detector and workspace created

Test point
Service Tool - Protocol Editor
Expected
Detector registered, workspace visible in AEC Calibration Protocol

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC image processing parameters adjustment radiation

Adjust image processing parameters after AEC calibration protocol creation

Ranked causes

  1. 1. Image processing parameters require update for AEC calibration (common)
First check

Identify current image processing parameters

Test point
Protocol workspace image processing configuration
Expected
Before change state documented in section 5.1.3

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC phototimer adjustment radiation

Setting up AEC phototimer to achieve target EI (exposure indicator) values across different kV settings and acrylics thickness for various imaging units

Ranked causes

  1. 1. AEC calibration procedure required during installation/setup (common)
First check

Prepare acrylics sheets for exposure

Test point
calibration setup
Expected
5cm, 10cm, 20cm thickness sheets ready

7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC protocol deletion after adjustment patient safety

Removal and hiding of AEC calibration protocols from operator interface after service engineer completes phototimer adjustment, to prevent accidental use by general operators.

Ranked causes

  1. 1. AEC protocols must be hidden to prevent operator misuse after adjustment (common)
First check

Access Protocol Editor from System Settings

Test point
System button > Protocol Editor button
Expected
Protocol Editor screen displays with tray containing AEC protocols

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC protocol setup and activation radiation

Selecting and starting the AEC Calibration protocol through the system interface

Ranked causes

  1. 1. AEC Calibration protocol must be activated for phototimer adjustment (common)
First check

Open Utility Protocol category

Test point
system interface
Expected
Utility Protocol category visible and expandable

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Detector correction data creation radiation

Procedure to create detector correction data and establish system calibration baseline; required during installation and annually.

Ranked causes

  1. 1. Periodic calibration maintenance requirement (common)
First check

Establish calibration mode and select target detector

Test point
QC tool detector list
Expected
Detector appears in list and is selectable

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Input data transformation using regular expressions

Configure pattern matching and replacement rules for input calibration using regular expressions in the service tool

Ranked causes

  1. 1. Need to apply regex pattern replacement to input fields (common)
First check

Verify pattern field is set when input calibration function is ON

Test point
Service tool input calibration settings
Expected
Pattern must be defined; if not set, no processing occurs

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Input data transformation using Script with character limit overflow

Handle character limit exceeding during Script processing and determine appropriate operation

Ranked causes

  1. 1. Processed output exceeds character limit after Script execution (occasional)
First check

Monitor output character count during Script execution

Test point
Script processing output
Expected
Output within default character limit

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Input data transformation using VBScript

Configure input calibration using VBScript for magnetic card input and medical examination automation pack NE/Studix optional product

Ranked causes

  1. 1. Need to apply script-based transformation to input data (common)
First check

Verify Script is enabled and VBScript file is specified

Test point
Service tool Script setting and Script file name
Expected
Script setting ON; valid VBScript file path specified

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Input data using regular expressions

Configure input calibration for magnetic card input and medical examination automation pack NE to transform patient data (Accession Number, Patient Name variants, Patient ID, Patient Sex) using regular expression patterns.

Ranked causes

  1. 1. Setup of regular expression patterns for data transformation (common)
First check

Verify Input Correction setting is enabled

Test point
Input calibration setting screen
Expected
Input Correction = ON

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Input data using VBScript

Configure input calibration for magnetic card input and medical examination automation pack NE/Studix using VBScript to transform patient data with cross-item referencing capability.

Ranked causes

  1. 1. Setup of VBScript for data transformation with inter-item referencing (common)
First check

Verify Script setting is enabled

Test point
Service tool Studix setting screen
Expected
Script = ON

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Performance test execution radiation

Procedure to verify detector maintenance and control through performance testing; recommended annually.

Ranked causes

  1. 1. Periodic performance verification requirement (common)
First check

Adjust X-ray generator settings based on previous calibration

Test point
X-ray generator parameter configuration
Expected
Settings match values from Exposure Memo of previous calibration

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Phototimer (AEC) adjustment radiation

Automatic Exposure Control optimization to minimize patient dose while maintaining diagnostic image quality. Performed during installation using AEC calibration protocols with Exposure Index (EI) measurement.

Ranked causes

  1. 1. Phototimer requires spectral sensitivity adjustment across kV settings (common)
First check

Start CCS and perform calibration

Test point
CCS system
Expected
System ready for AEC adjustment

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remove AEC Calibration protocol after completion

Deleting AEC Calibration protocol and hiding Utility Protocol tab after all adjustments are completed

Ranked causes

  1. 1. AEC Calibration protocol must be removed after adjustment to prevent accidental activation by users (common)
First check

Enter system settings screen

Test point
system interface
Expected
System Settings Screen displayed

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCRHIS library error

Error occurred in CCRHIS library processing. May be due to incorrect settings, file I/O error or other problems.

Ranked causes

  1. 1. Studix processing failed due to incorrect settings or file I/O error (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collection Tool startup when Service Tool is active

Collection Tool cannot run while Service Tool is operating; unsaved settings prompt appears.

Ranked causes

  1. 1. Service Tool has unsaved data being edited (common)
First check

Observe dialog box when Collection Tool is invoked

Test point
Service Tool main window
Expected
Confirmation dialog appears asking to save settings

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collection Tool startup with unsaved data

Settings save confirmation dialog appears if Collection Tool is started while data is being edited in Service Tool

Ranked causes

  1. 1. Service Tool has unsaved edits (occasional)
First check

Check for settings save dialog on Collection Tool startup

Test point
Service Tool screen during Collection Tool initialization
Expected
Settings save confirmation dialog appears if edits pending

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication failures with 70C sensor radiation

Communication errors or dropped connections between detector and system

Ranked causes

  1. 1. Software incompatibility in CXDI NE with 70C sensor (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cooling unit attached or detached during study

Processing halted due to cooling unit operation during active acquisition

Ranked causes

  1. 1. Cooling unit operated during study (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector analog power supply abnormality high voltage

Abnormality detected in the analog power supply for the detector.

Ranked causes

  1. 1. Analog power supply abnormality in detector (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector battery depleted or insufficient charge

Detector battery charge is critically low; imaging operations halted

Ranked causes

  1. 1. Battery charge insufficient (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector calibration data check failed radiation

SPCC data validation failed; possible detector abnormality or power supply issue

Ranked causes

  1. 1. Detector power supply off or cable disconnected (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector characteristics compensation data invalid radiation

System encounters illegal detector calibration compensation data; internal processing error

Ranked causes

  1. 1. Corrupted or incorrect detector compensation data (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector communication error

Detector communication error occurred. Cable may be disconnected or power discontinuity detected.

Ranked causes

  1. 1. Ethernet interface communication error (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector communication error fatal

Fatal detector communication error. System restart required. cxdcap.ini file not installed.

Ranked causes

  1. 1. cxdcap.ini file not installed (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector control line connection error caution

Control line connection error detected in detector. Broken cable between detector and power box may be involved.

Ranked causes

  1. 1. Control wire error or broken cable between detector and power box (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector control wire cable length mismatch caution

Discrepancy between cable length settings for detector-to-power box connection and XRAY board cable.

Ranked causes

  1. 1. Cable length settings discrepancy between detector cable and XRAY board cable (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector hardware error occurred radiation

Internal detector malfunction; system restart required

Ranked causes

  1. 1. Detector internal error (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector internal temperature approaching upper limit caution

Internal temperature of detector is close to upper limit. Do not leave detector in Ready condition.

Ranked causes

  1. 1. Detector temperature approaching upper limit (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector internal temperature exceeds upper limit

Detector temperature warning; continues operation but at elevated risk

Ranked causes

  1. 1. High ambient temperature or insufficient detector cooling (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector issued fatal error radiation

Detector reports fatal error condition; system restart required

Ranked causes

  1. 1. Detector fatal error condition triggered (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector issued warning radiation

Detector reports non-critical warning condition; operation may continue

Ranked causes

  1. 1. Detector warning condition triggered (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector serial number discrepancy caution

Discrepancy between serial number of detector and DP number detected.

Ranked causes

  1. 1. Serial number discrepancy between detector and DP (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector structure abnormality detected caution

Abnormality detected in detector structure. Illegal value written for serial number.

Ranked causes

  1. 1. Illegal value written for detector serial number (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector temperature critically high

Detector internal temperature exceeds safe operating limit; processing halted

Ranked causes

  1. 1. Abnormally high detector temperature (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive circuit error caution

PLD error detected by detector AD board. System cannot proceed.

Ranked causes

  1. 1. PLD error detected by detector AD board (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive circuit PLD config error caution

PLD Config error detected by detector AD board. System cannot proceed.

Ranked causes

  1. 1. PLD Config error detected by detector AD board (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Entered information has invalid value patient safety

Exam information containing illegal values for items other than key items was acquired. Invalid values are replaced with blanks.

Ranked causes

  1. 1. Exam information with illegal values for non-key items from Studix (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code E040100054

System is receiving E040100054 error code indicating communication or image loss issue

Ranked causes

  1. 1. Combination of outdated firmware and wireless interference (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Extra defect correction file not loaded

System fails to load the extra defect correction file required for imaging. Error occurs during system operation.

Ranked causes

  1. 1. Extra defect correction file does not exist (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Firmware backup area management caution

Understand double firmware architecture for safe updates

Ranked causes

  1. 1. Understand backup firmware protection (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Firmware update: IP address configuration

Change IP addresses for Ethernet-connected detectors using Ferry Type FPD Maintenance Tool

Ranked causes

  1. 1. IP address change required for detector network configuration (occasional)
First check

Confirm detector supports Ethernet connection

Test point
Detector model type
Expected
CXDI-40EG, 40EC, 40G Compact, 50G, 50C, 55G, 55C, 60G, 60C, CXDI-401G/C or CXDI-401G/C Compact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Firmware version mismatch

Firmware version differs from expected. Wrong detector firmware may have been written.

Ranked causes

  1. 1. Wrong detector firmware written (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flash ROM erase error

Flash ROM error detected in detector. Flash ROM erase error occurred.

Ranked causes

  1. 1. Flash ROM erase error detected by detector AD board (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flash ROM ID error

Flash ROM error detected in detector. Flash ROM ID error occurred.

Ranked causes

  1. 1. Flash ROM ID error detected by detector AD board (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flash ROM write error

Flash ROM error detected in detector. Flash ROM write error occurred.

Ranked causes

  1. 1. Flash ROM write error detected by detector AD board (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Generator initialization communication failed
  • Gray pixel values displayed in image
  • Grid error at detector
  • Grid error may have caused image abnormality
  • Grid ID loading error
  • Image data transfer failed
  • Image data transfer stopped
  • Image rotation not applied to output
  • Invalid detector specified in system settings
  • Invalid studies acquired
  • Invalid workspace specified in system settings
  • Memory allocation failed
  • No battery attached to detector
  • Periodic noise detected in acquisition data
  • Phototimer adjustment: AEC calibration setup
  • Self-diagnosis: Detector and power box system test
  • Specified ID not found in examination file
  • System setting: Enable EI (Exposure Indicator) display annotation
  • Undefined detector error file download
  • Undefined detector error hex format
  • Verify AEC Calibration Protocol configuration
  • Workspace creation required before calibration

22 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
020201003Detector battery charge dropped below 8%Detector battery remaining charge falls below 8% of full chargeerror
020201004Gamma adjustment unavailable due to pattern file loading failureAll pattern files for simplified gamma adjustments failed to loaderror
020201004Unexpected error occurred. Gamma adjustment is not available.All loading of pattern files used for simplified gamma adjustments has failed.error
020201005Failed to connect to the barcode readerException occurred in barcode reader connection initialization processerror
020201005Barcode reader connection failedException during barcode reader connection initializationerror
020201006Cannot start specified examination.Examination specified during Studix operation when specification using ccrhis.dvr file is ignored, or patient specified during magnetic card operation when specification is ignored.error
020201007Detector communication error during exposureReception timeout from detector or detector power-off during exposureerror
020201008Detector temperature reached dangerous levelsDetector temperature reached dangerously high level triggering Sleep statuserror
020201008The detector temperature has reached dangerous levels. Images cannot be captured until temperature has dropped.Detector temperature reached dangerously high level, establishing Sleep status.error
020201009Failed to start up the OverwrapSoftwareStartup failed due to invalid path of overlapping software set by service toolerror
020201009OverwrapSoftware startup failedStartup failed due to invalid overlapping software path configurationerror
020202001Detector temperature reached warning levelDetector temperature notification from Ferry indicates warning level reachedwarning
020202002Detector temperature reached dangerous level (warning state)Ferry notification that detector temperature reached dangerous levelwarning
020202003The detector is not available. Irradiation has finished.Notification received that detector became invalid during irradiation.warning
020202004The detector battery is weakening. Charge the detector battery.Remaining charge of detector battery dropped below 4% of full charge.warning
020202005Specified disk space could not be cleaned upSpace greater than examination to be deleted was specifiedwarning
030100001No calibration data availableWorkspace calibration has not been completederror
030100005Required COM+ servers not installedOne or more necessary COM+ servers not installed during system startupcritical
030201001Detector communication error occurredAcquisition of detector information from capture manager failed when inspection startederror
030201002Periodic noise detected in acquisition dataGrid-embedded detector calibration performed with grid in placeerror
030201003Cooling unit operation during studyCooling unit attached or detached during active acquisitionerror
030201003The cooling unit has been attached or detached. Processing was halted.Cooling unit was operated (attached or detached) during the study.error
030201004Calibration data (SPCC) check failedQCT completion returns error on SPCC data validationerror
030201005Generator disconnectedGenerator disconnection event occurrederror
030201006Generator is disconnected (communication failure)Generator communication failure during calibration, performance test, or self-diagnosis initiationerror
030201007Generator is disconnected (condition setting failure)Generator condition setting fails during calibration, performance test, or self-diagnosiserror
030201007Imaging preparation failedSetting of conditions in generator or detector failederror
030201008Detector battery charge insufficientDetector battery remaining charge is deficienterror
030201008Detector cannot be used because the battery has lost its charge. Processing was halted.Remaining charge of detector battery dropped to deficient level.error
030201009The detector temperature has reached dangerous levels. Processing was halted.Notification received from detector of abnormally high temperature.error
030201009Detector temperature reached dangerous level during QCDetector notification of abnormally high temperatureerror
030201010No battery is attached to the detector. Attach the battery.Battery has not been installed with a wired connection.error
03020200EMemory allocation failedInsufficient system RAM when QC Tool starts or examination beginscritical
03020201BCalibration data upload failedError during SPCC data upload operation in QC Toolcritical
03020201CImage data transfer failedError during automatic image transfer from detector to systemcritical
03020201FDetector error occurredInternal detector error indication during operationcritical
030203003Detector internal temperature exceeds upper limitDetector temperature reaches warning thresholdwarning
03020FFFFAn internal error occurredBoth X-ray exposure and exposure cancel occur at the same time, or software bugcritical
040100001Invalid workspace has been specifiedAttempt to use unregistered workspace configurationerror
040100008Invalid detector has been specifiedAttempt to use unregistered detector configurationerror
040100023Generator communication failedGenerator initialization communication error at system setuperror
040100041Internal error - illegal detector compensation dataDetector characteristics compensation data is illegal during image processingerror
040100042Internal error - illegal detector compensation dataDetector characteristics compensation data is illegal during operationerror
040100045Calibration data is invalidCalibration data is not correct or corruptedwarning
040100052Detector issued a warningWarning notification from detector occurredwarning
040100053Detector issued an errorError condition reported by detector during operationerror
040100054Detector issued a fatal errorFatal error condition reported by detector during operationcritical
040100056Generator disconnectedRetry button canceled during X-ray generator imaging condition setup failureerror
040300000Detector has not been started upSelected detector not attached or detector power supply issueerror
040300002Internal error occurredSelected detector not attached or hardware malfunctionerror
040400003Failed to acquire system settingsDatabase data defective, insufficient table recordscritical
040500003Exceeded maximum number of study order acquisitionsC-FIND-RSP study exceeds maximum configured acquisitionserror
040500004Received some invalid studiesData contravening the DICOM rules are contained in C-FIND-RSP receivederror
040500005Received tag contains invalid value; value replaced by nullData contravening DICOM rules in C-FIND-RSP receivederror
040500006Unsupported character set receivedCharacters not in received character set present; auxiliary kanjis or decoding failureerror
040500006Characters not in received character set present, may not display correctlyCharacters cannot be decoded properly or auxiliary kanjis received from RISerror
040500007Characters do not match current character code; automatic correction appliedJapanese-language data received despite English (ASCII) character set configuredwarning
040500007Character code mismatchData received with character code different from service tool settingwarning
040500008Date entry before 1753 included in acquired ExamDate type (DA) data prior to January 1, 1753 received from RISerror
040500009Failed to copy tag to C-STOREDICOM rule violation in data during tag copy process from C-FIND to C-STORE; non-existent or invalid tag specifiedwarning
04050000AFailed to copy tag to DicomFileViolation of DICOM rules when copying tags from C-FIND to DicomFile, or non-existent tag specifiedwarning
04050000CFailed to load DICOMDIRDICOMDIR file loading failure due to damage or corruptionerror
04050000DOutput error occurredOutput information unavailable after system restart due to power discontinuityerror
040501011Cannot continue processing; file already exists at storage destinationFile with same name exists at configured save destinationerror
040501012Cannot continue processing; access to storage destination deniedSave destination folder cannot be accessed; insufficient user permissionserror
040501013Insufficient storage spaceNot enough space at specified storage destinationerror
040501013Cannot continue processing; insufficient space at storage destinationNot enough disk space available at configured save destinationerror
040501014DICOM GSPS output halted; object image transfer incompleteC-STORE processing of GSPS executed while C-STORE of target images failed or incompleteerror
040501016Network communication cut off during file outputNetwork problem or interruption during media storage over networkerror
040501017Unsupported DICOMDIR loadedDICOMDIR file not created by CXDI NE/RF loadederror
040501018Output error occurred during version upgradeVersion upgrade from pre-V1.30 performed while media storage queue existserror
040502002Association negotiation failedPermanent rejection by service provider (ACSE-related function) with no reasonerror
040502003Association negotiation failed - unsupported abstract syntaxService provider rejected unsupported abstract syntax senterror
040502004Association negotiation failed - unsupported forward syntaxService provider rejected unsupported forward syntaxerror
04050200DAssociation negotiation failed - ACSE rejection without reasonPermanent rejection by service provider (ACSE-related function)error
04050200EAssociation negotiation failed - unsupported protocol versionProtocol version not supported by service providererror
04050200FAssociation negotiation failed - temporary service rejectionTemporary rejection by service provider (ACSE-related function)warning
040502010Association negotiation failed - temporary protocol version rejectionTemporary rejection due to unsupported protocol versionwarning
040502011Association negotiation failed - temporary congestionPermanent rejection by service provider (presentation-related) due to temporary congestionwarning
040502012Association negotiation failed - local restrictions exceededASSOCIATE-RJ received with RESULT=1, SOURCE=3, REASON=2 during negotiationerror
040502013Association negotiation failed - presentation temporary congestionTemporary rejection by service provider (presentation-related) due to congestionwarning
040502014Association negotiation failed - presentation restrictions exceededTemporary rejection by service provider (presentation-related) - local restrictions exceededwarning
04050201BAssociation stopped - aborted by service providerAssociation aborted by service provider without reasonwarning
04050201CAssociation stopped - PDU not recognizedAssociation aborted because service provider cannot recognize PDUerror
04050201DAssociation stopped - unexpected PDU receivedAssociation aborted due to unexpected PDU from service providererror
04050201EAssociation stopped - PDU parameters not recognizedAssociation aborted because PDU parameters cannot be recognizederror
04050201FAssociation stopped - unexpected PDU parametersAssociation aborted due to unexpected PDU parameterserror
040502020Association stopped - invalid PDU parameter valueAssociation aborted due to invalid PDU parameter valueerror
040502021Association negotiation failed - timeout on readout standbyNo response from communication partner within readout standby timeout periodwarning
040502023Association negotiation failed - RELEASE-RQ received out of sequenceRELEASE-RQ received at time other than association release phaseerror
040502058Association negotiation failed - invalid IP or portAssociation initiated with no valid communication partner IP address and porterror
040505001DICOM communication error - unsupported attribute tagStatus tag (0000,0900) value error received; non-existent attribute seterror
040505002DICOM communication error - invalid attribute valueStatus tag (0000,0900) value error; invalid attribute value seterror
040505003DICOM communication warning due to attribute list errorStatus tag (0000,0900) value error with attribute list problemwarning
040505004DICOM communication error, status tag (0000,0900) value errorProcessing failederror
040505005DICOM communication error - multiple SOP instancesStatus tag (0000,0900) value error; multiple SOP instances presenterror
040505006DICOM communication error - non-existent object instanceStatus tag (0000,0900) value error; non-existent object instance seterror
040505007DICOM communication error - non-existent event typeStatus tag (0000,0900) value error with non-existent event type seterror
040505008DICOM communication error - non-existent argumentStatus tag (0000,0900) value error with non-existent argument seterror
040505009DICOM communication error due to invalid argument in status tag (0000,0900)Status tag (0000,0900) value error received with invalid argument seterror
04050500ADICOM communication warning - attribute value out of rangeStatus tag (0000,0900) value error with attribute value outside rangewarning
04050500BDICOM communication error due to invalid object instance in status tagStatus tag (0000,0900) value error with invalid object instance seterror
04050500CDICOM communication error - non-existent SOP classStatus tag (0000,0900) value error with non-existent SOP class seterror
04050500DDICOM communication error - class instance contradictionStatus tag (0000,0900) value error with contradiction in class instanceerror
04050500EDICOM communication error due to missing attribute in status tagStatus tag (0000,0900) value error with missing attributeerror
04050500FDICOM communication error due to missing attribute valueStatus tag (0000,0900) value error with missing attribute value seterror
040505010DICOM communication error - non-supported SOP classStatus tag (0000,0900) value error with non-supported SOP class seterror
040505011DICOM communication error due to non-existent action typeStatus tag (0000,0900) value error with non-existent action typeerror
040505012DICOM communication error due to repeated startupStatus tag (0000,0900) value error with repeated startup conditionerror
040505013DICOM communication error - unrecognized operationStatus tag (0000,0900) value error with unrecognized operationerror
040505014DICOM communication error due to erroneous input argumentStatus tag (0000,0900) value error with incorrectly input argumenterror
040505015DICOM communication error due to resource restrictionStatus tag (0000,0900) value error with resource limitationerror
040505017DICOM communication error - ununderstandable settingStatus tag (0000,0900) value error with unintelligible settingerror
040505018DICOM communication error due to insufficient resourcesStatus tag (0000,0900) value error with insufficient resourceserror
04050501BDICOM C-FIND error due to identifier/SOP class mismatchC-FIND status tag (0000,0900) value error with identifier mismatcherror
04050501CDICOM C-STORE warning - data factor coercionStatus tag (0000,0900) value error in C-STORE with data coercionwarning
04050501DDICOM C-STORE warning - factor abandonedStatus tag (0000,0900) value error in C-STORE where factor was droppedwarning
04050501EDICOM C-STORE warning - dataset/SOP class mismatchStatus tag (0000,0900) value error in C-STORE where dataset does not match SOPwarning
04050501FDICOM C-STORE error due to dataset/SOP class mismatchC-STORE status tag (0000,0900) value error with dataset mismatcherror
040505020DICOM C-STORE error due to no resources at destinationC-STORE status tag (0000,0900) value error with no resourceserror
040505027DICOM printer output error due to missing film box SOP instancesN-ACTION status tag (0000,0900) error with BasicFilmSession classerror
040505028DICOM printer output error due to full print queueN-ACTION status tag (0000,0900) error; print queue fullerror
040505029DICOM printer output error due to image size exceeding image box sizeN-ACTION status tag (0000,0900) error with oversized imageerror
04050502ADICOM printer output error due to linked print image exceeding box sizeN-ACTION status tag (0000,0900) error with linked image too largeerror
04050502CDICOM printer output N-ACTION error; print queue is full for BasicFilmBoxN-ACTION status tag (0000,0900) value error when SOP class was BasicFilmBoxerror
04050502EDICOM printer error due to unprinted film box and unsupported film session operationN-CREATE status tag (0000,0900) error with BasicFilmSession classerror
04050502FDICOM printer output N-SET error; insufficient printer memory for imagesN-SET status tag (0000,0900) value error when SOP class was GrayScaleImageBoxerror
040505030DICOM printer output warning; printer status is WARNINGPrinter status tag (2110,0010) value of N-GET during Printer operation is WARNINGwarning
040506001DICOM printer output halted due to printer failure statusPrinter status tag (2110,0010) N-GET received with FAILURE statuserror
040507001DICOM StorageCommitment failed with reason code 0x0110N-EVENT-REPORT failure with reason (0008,1197) = 0x0110error
040507002DICOM StorageCommitment failed with reason code 0x0112N-EVENT-REPORT failure with reason (0008,1197) = 0x0112error
040507003DICOM StorageCommitment failed with reason code 0x0119N-EVENT-REPORT failure with reason (0008,1197) = 0x0119error
040507004DICOM StorageCommitment failed with reason code 0x0122N-EVENT-REPORT failure with reason (0008,1197) = 0x0122error
040507005DICOM StorageCommitment failed with reason code 0x0131N-EVENT-REPORT failure with reason (0008,1197) = 0x0131error
040507006DICOM StorageCommitment failed with undefined reason codeN-EVENT-REPORT failure with non-DICOM-defined reason (0008,1197)error
040507007DICOM StorageCommitment failed with undefined reason codeN-EVENT-REPORT failure with non-DICOM-defined reason (0008,1197)error
040600008Disc access failedDisk could not be accessed during media storage free space checkerror
040600008Disk access failed during media storage free space checkDisk could not be accessed during free space check at time of media storageerror
040600009Insufficient free disk space during processFree space check at media storage shows insufficient spaceerror
04060000BProcess error prevents retryPower discontinuity, software bug, or image processing failureerror
04060000BRetry failed due to error during media output processingPower discontinuity or software bug; ID key did not exist or image processing failederror
04060000COld study info cannot execute; created before rebindingSystem attempting transfer of old study from pre-rebindingerror
040700006Failed to load image file; file does not existAttempt to load non-existent image fileerror
040700100System memory settings invalid; exceeds installed memoryService tool specifies memory size larger than installed RAMcritical
040700100System memory settings are invalid; exceeds installed capacityService tool specified memory size larger than installed memorycritical
040800100Examination cannot be started due to non-existent protocolProtocol contained in pending examination has been deleted from databaseerror
040D0001FProviding free space on the hard disk failedSystem unable to delete data equivalent to prescribed number of recordserror
040E01000Image analysis processing failedImage analysis processing routine failed during executionerror
040E01001Failed to load extra defect correction fileExtra defect correction file does not exist or is missingerror
041000004Invalid studies are acquiredExam information containing illegal values for key items (ACC#, patient ID, patient name) acquired from Studixerror
041000005Entered information has invalid valueExam information containing illegal values for non-key items acquired from Studixerror
041000006Specified ID is not found in the examination fileExam information with specified key values did not exist in examination fileerror
041000008An error has occurred during script-based input calibrationScript compilation error, script processing error, script standard error output, or script return value other than 0error
041000010An error has occurred in CCRHIS libraryStudix processing failed due to incorrect settings, file I/O error or other problemerror
041000011Workflow library version is not supportedVersion of CCS does not match automatic examination pack versionerror
041000012Error occurred in CCRHIS libraryStudix, Workflow, or magnetic card processing failed due to invalid setting or file I/O errorerror
041000020Examination file loading halted due to maximum reserved exams exceededMaximum number of reserved exams registered to database exceedederror
041000021Invalid studies loaded and cannot be displayedInvalid value detected in required items of exam informationerror
041000022Entered information has invalid value in user-specified itemsInvalid value detected in user-specified valueserror
041000023Failed to read examination fileExam file not found at specified locationerror
041000024Error occurred during automatic examination file updatingError during exam file monitoring with automatic update configurationerror
041000025Failed to read examination file during loadingError occurred during loading of exam fileerror
041100004CxdiInfo startup failedSpecified DLL does not exist or lacks expected interfaceerror
050501002A flash ROM error was detected in the detectorFlash ROM ID error detected by detector AD boardwarning
050501003A flash ROM error was detected in the detectorFlash ROM Write error detected by detector AD boardwarning
050501004A flash ROM error was detected in the detectorFlash ROM Erase error detected by detector AD boardwarning
050501005There may be problem in the images because of a grid errorGrid warning detected by detector AD board but imaging continuedwarning
050501005Grid error may have caused image abnormalityGrid warning detected during imagingwarning
050501006Detector internal temperature exceeds upper limitDetector internal temperature exceeds safety thresholderror
050501007The internal temperature of the detector is nearing its upper limitDetector internal temperature approaching upper limitwarning
050501008A drive circuit error occurredPLD error detected by detector AD boarderror
050501009A drive circuit error occurredPLD Config error detected by detector AD boarderror
05050100AAn abnormality was detected in the detector structureIllegal value written for detector serial numbererror
05050100BAn abnormality was detected in the detector structureDiscrepancy between detector serial number and DP numbererror
05050100BDetector serial number discrepancyMismatch between detector serial number and DP numbererror
05050100CA control line connection error was detected in the detectorControl wire error detected by detector AD board; broken cable between detector and power boxerror
05050100DA control line connection error was detected in the detectorControl wire error detected by detector AD board; cable length mismatch between detector-power box and XRAY boarderror
05050100DControl wire cable length mismatchCable length settings mismatch between detector and XRAY boarderror
05050100EUndefined detector error hex formatHex format error during firmware or PLD downloadwarning
05050100EAn error was detected by the detectorUndefined error detected by detector AD board; Hex format error during firmware or PLD downloadwarning
05050100FAn error was detected by the detectorUndefined error detected by detector AD board; Hex format error during firmware, PLD or DP file downloadwarning
05050100FUndefined detector error file downloadHex format error during firmware, PLD, or DP file downloadwarning
050501010Image data transfer failedDMA timeout in PCB-Di during reduced-size image readerror
050501010Failed to transfer the image dataDMA inside PCB-Di timed out when reduced-size images were readerror
050501011The transfer of the image data was stoppedTransfer of Full image data was abortedwarning
050501012An abnormality was detected in the analog power supply for the detectorAbnormality detected in detector analog power supplyerror
050501013A grid ID loading error was detectedGrid ID loading error detectedwarning
050501014A grid error was detectedGrid error detected at the detector AD boarderror
050501015A grid error may have caused an image abnormalityGrid warning detected by detector AD board but imaging continuedwarning
050501016The firmware is differentWrong detector files written during firmware writingerror
050502009A detector communication error has occurredcxdcap.ini file has not been installedcritical
050502026A detector communication error occurredCommunication error detected by Ethernet interface; cable disconnected; detector power discontinuityerror
E04010005470C sensor communication failure or image loss errorLoss of communication between 70C sensor and system or failure to complete image transmissionerror
E040500005MWL DICOM violation errorIllegal DICOM data received from MWLerror
E040500006MWL DICOM violation errorIllegal DICOM data received from MWLerror
E040500008MWL DICOM violation errorIllegal DICOM data received from MWLerror
E040501010System error from exceeded SQL Server database size limitDiskstore database exceeds 4GB capacity when more than 1000 studies stored continuously on large-capacity media without shutdowncritical
ERRORError condition requiring user action; imaging can continue once remediedSystem enters error stateerror
ERRORError level system error affecting operationError condition occurs during system operationerror
FATALFatal condition requiring service call; imaging cannot be continuedSystem enters fatal error statecritical
FATALFatal level system error that prevents normal operationOccurs during any operational state; error dialog appears automaticallycritical
Pattern Match ErrorError occurred while applying regular expression replacementRegular expression processing fails on one or more items during input calibrationwarning
Script compilation errorSyntax or compilation error detected in VBScript codeInvalid VBScript syntax when Script execution beginserror
Script ErrorError occurred during VBScript executionVBScript execution fails, compilation error, or error in return valueerror
Script return value errorScript return value is invalid or unexpected typeScript completes but return value does not match expected typeerror
Standard error output in ScriptStandard error message generated within Script executionScript encounters runtime error and outputs to stderrwarning
WARNINGWarning condition; no immediate user action requiredSystem enters warning statewarning
WARNINGWarning level non-critical system alertWarning condition detected during system check or operationwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.