Canon CXDI Control Software NE: troubleshooting and error codes

The ServiceForge knowledge base holds 44 documented symptoms for the Canon CXDI Control Software NE, with 55 ranked causes, 103 diagnostic checks with test points and expected values, and 214 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Second monitor image display magnification and cropping

Configuration of 3-phase fixed zooming ratios, threshold levels, and image cropping for second monitor display based on detector pixel pitch (0.160 mm or 0.125 mm) and monitor resolution

Ranked causes

  1. 1. Service tool registration of second monitor resolution (common)
  2. 2. Image cropping to first phase threshold levels triggers 100% magnification (common)
  3. 3. Image cropping between first and second phase thresholds triggers second phase magnification (common)
  4. 4. Image cropping to full screen or large size triggers third phase magnification (common)
First check

Register second monitor resolution in service tool

Test point
Service tool 'Set connected second monitor size' button
Expected
Orientation and resolution registered; [Apply] button clicked

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC Calibration protocol grid parameter error on Excel import

Error occurs when loading Protocol Sheet after importing Excel file exported from Protocol Database. Inappropriate grid parameter combination (GridType Off, Grid Suppression On) causes import failure.

Ranked causes

  1. 1. Inappropriate default AEC Calibration grid parameters installed (common)
  2. 2. Previously exported Excel file contains inconsistent grid parameters (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC Calibration Protocol setup for new detector

Need to create or modify AEC Calibration Protocol when adding a new detector to the system

Ranked causes

  1. 1. AEC Calibration Protocol deleted from database (occasional)
  2. 2. New detector added without protocol update (occasional)
First check

Verify workspace exists for detector

Test point
Service Tool Protocol Editor
Expected
Workspace listed for detector

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error with multiple detectors connected

Fatal error occurs during first calibration after detector detection with message about missing dp data

Ranked causes

  1. 1. Multiple detectors connected during initial calibration (common)
  2. 2. Version earlier than V1.05 installed (common)
First check

Verify number of detectors connected

Test point
Hardware connection
Expected
Only one detector connected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Grid installation angle measurement radiation

Measure grid installation angle relative to imaging unit sensor glass plate using audit_grid_NE_RF.exe software tool to verify alignment is within specification.

Ranked causes

  1. 1. Grid installation angle must be verified and corrected (common)
  2. 2. Image file format may require parameter configuration (occasional)
First check

Prepare software files

Test point
Image Capture Computer
Expected
audit_grid_NE_RF.exe and SecureConverter.dll copied to computer

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator communication settings configuration

Configure X-ray generator communication parameters including protocol, IP address, and display settings in service tool

Ranked causes

  1. 1. Generator communication settings must be configured for software to control X-ray exposure (common)
  2. 2. Response timeout and network address configuration (occasional)
First check

Open Generator settings from service tool TreeView

Test point
Service tool TreeView Generator menu
Expected
Generator setting screen displays

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software installation component selection and uninstallation cleanup

Select CANONDRSYSTEM component during installation and remove installation folder after uninstall

Ranked causes

  1. 1. Installation requires selection of CANONDRSYSTEM component (common)
  2. 2. Installation folder remains after program deletion and requires manual removal (occasional)
First check

Select CANONDRSYSTEM from component list

Test point
Component Selection screen checkbox
Expected
CANONDRSYSTEM checkbox selected

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software upgrade loses configuration settings

Certain settings revert to default after upgrading CXDI Control Software NE to new version

Ranked causes

  1. 1. Software upgrade process replaces configuration database (common)
  2. 2. Overwrap Software installed in V1.0.0 before upgrade (occasional)
First check

Backup database before upgrade

Test point
Database management
Expected
Backup file created

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Workspace creation and detector association

Create predetermined or custom workspace configuration and associate detector with position type

Ranked causes

  1. 1. Workspace must be created and associated with detector before calibration (common)
  2. 2. Custom workspace with optional protocol association (occasional)
First check

Open Workspace property settings in Service Tool

Test point
Property Editor Workspace menu
Expected
Workspace configuration screen displays

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alert window appears during SQL Server 2005 setup when upgrading from Ver.1.30 to Ver.1.40

Version upgrade processing is suspended when upgrading from Ver.1.30 or earlier to Ver.1.40; alert window is displayed during SQL Server 2005 setup

Ranked causes

  1. 1. SQL Server 2005 setup alert during version upgrade from Ver.1.30 to Ver.1.40 (common)
First check

Observe whether alert window appears during SQL Server 2005 setup phase of upgrade

Test point
SQL Server 2005 setup
Expected
Alert window may appear during upgrade from Ver.1.30 to Ver.1.40

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC calibration protocol selection radiation

Select and activate AEC calibration protocol for exposure testing and adjustment

Ranked causes

  1. 1. AEC calibration protocol must be selected for testing (common)
First check

Confirm AEC calibration protocols are available in tray

Test point
Protocol tray
Expected
AEC calibration protocols visible from section 4.8.2 setup

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Detector correction data creation radiation

Initial detector calibration and annual maintenance calibration to establish correction data for accurate imaging

Ranked causes

  1. 1. Detector calibration required during installation or annual maintenance (common)
First check

Establish calibration mode and select detector

Test point
Detector list in calibration interface
Expected
Detector selectable from list

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: EI display configuration

Configure EI (Exposure Index) display on screen at bottom right corner for proper visibility during examinations

Ranked causes

  1. 1. EI display not visible or requires initial setup (common)
First check

Verify EI is displayed on screen

Test point
Bottom right corner of display
Expected
EI value visible in bottom right corner

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Grid installation angle radiation

Verify that anti-scatter grid is installed at correct angle relative to X-ray focal point; angle must be within ±1.0° tolerance across three test images

Ranked causes

  1. 1. Grid installation angle out of tolerance (unknown)
First check

Measure grid installation angle using audit_grid_NE_RF.exe on first image

Test point
Grid relative to X-ray focal point
Expected
±1.0 or less, ±1.0°

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Phototimer adjustment radiation

Adjust photo-timer using Exposure Index (EI) to optimize patient dose and image quality. Requires exposing acrylics sheets at specified voltages and adjusting to target EI values per imaging unit model.

Ranked causes

  1. 1. Phototimer requires adjustment during installation (common)
First check

Confirm EI is displayed on Examination Screen

Test point
System Settings > Annotation tab
Expected
EI displayed at one of four corners

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Phototimer adjustment for AEC radiation

Adjust photo-timer to achieve target EI values at different kV settings (70, 100, 120 kV) using acrylic sheets; spectral sensitivity adjustment required

Ranked causes

  1. 1. Phototimer requires calibration for correct EI output across voltage settings (common)
First check

Expose X-ray through 5 cm acrylic at 70 kV and measure EI

Test point
Detector readout
Expected
EI 480 for CXDI-40EG/40G/401G/401G Compact/50G/55G/60G/501G/70G/701G/801G Wireless; EI 280 for CXDI-40EC/401C/401C Compact/50C/55C/501C/60C/70C/80C/701C/801C/50RF

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ROI adjustment after CXDI-70C wireless exposure

Workaround procedure to enable normal ROI adjustment on preview image after exposure with CXDI-70C wireless detector until bug-fixed software version is available.

Ranked causes

  1. 1. Software bug in Ver.1.30.0.17 requires intermediate image display before ROI adjustment (common)
First check

Select image B from available protocols or past image list

Test point
CXDI Control Software NE image selection interface
Expected
Image B displayed (must be different from target image A and previously exposed or in protocol)

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ROI adjustment workaround for CXDI-70C wireless

Procedure to perform successful ROI adjustment on preview images when using CXDI-70C wireless detector with affected software version

Ranked causes

  1. 1. Display sequence requirement for ROI adjustment with CXDI-70C wireless Ver.1.30.0.17 (common)
First check

Select image A for ROI adjustment

Test point
preview image
Expected
Image A displayed

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Second monitor display settings configuration

Configuration of preview annotations, laterality markers, and peripheral mask display on second monitor

Ranked causes

  1. 1. Second monitor display setting configuration via service tool (common)
First check

Confirm second monitor is connected and enabled

Test point
Service tool configuration
Expected
'Use second monitor' is ON and 2 monitors connected

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Second monitor gamma adjustment

Adjust gamma levels for secondary monitor in dual-monitor configuration using Gamma Adjustment Tool

Ranked causes

  1. 1. Second monitor gamma requires calibration (common)
First check

Verify second monitor is physically connected to video board DVI output

Test point
Video board DVI connector for second monitor
Expected
Second monitor connected and recognized by system

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Second monitor setup and configuration

Configure second monitor for dual-monitor operation including resolution, display options, and grid suppression algorithm

Ranked causes

  1. 1. Second monitor requires initial setup configuration (common)
First check

Verify video board has required DVI connectivity for dual monitors

Test point
Video board output connectors
Expected
Two DVI outputs present or DisplayPort-to-DVI converter available (for NVIDIA Quadro600)

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector registration with IP address configuration

Register X-ray detector with predetermined IP address and enroll detector information in service tool system

Ranked causes

  1. 1. Detector must be registered in service tool before operation (common)
First check

Connect detector to network

Test point
Network connection to Power Box
Expected
Detector connected and communicable

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector temperature dangerous level caution

Detector temperature has reached dangerous level; system displays warning button; imaging can continue

Ranked causes

  1. 1. Detector temperature reached dangerous threshold (common)
First check

Monitor detector temperature

Test point
Ferry detector temperature sensor
Expected
Temperature below dangerous threshold

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector temperature warning caution

Detector temperature has reached warning level; system displays warning button in GUI header; imaging can continue

Ranked causes

  1. 1. Detector temperature reached warning threshold (common)
First check

Monitor detector temperature via system notifications

Test point
Ferry detector temperature sensor
Expected
Temperature below warning threshold

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Detector unavailable during irradiation patient safety

Detector becomes invalid during X-ray exposure; irradiation stops automatically; error does not occur without user intervention

Ranked causes

  1. 1. Detector touched or disturbed during irradiation (occasional)
First check

Verify detector status before irradiation

Test point
Detector availability status in system
Expected
Detector available and not touched during irradiation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Double enlargement ratio compensation in distance measurement

DR Image Viewer applies enlargement ratio compensation twice when measuring distance on images that already have compensated enlargement ratio, producing incorrect results.

Ranked causes

  1. 1. Enlargement ratio compensation enabled in both NE/RF and DR Image Viewer (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error E010156203 displayed persistently after entering special character in SSID field

User inputs non-alphanumeric character like '(' in SSID field and presses Save button. Error E010156203 appears. Even after replacing the illegal character with valid input, the error persists permanently when Save is pressed again.

Ranked causes

  1. 1. Non-alphanumeric character entered in SSID field (common)
First check

Verify SSID field contains only alphanumeric characters

Test point
Configuration tab SSID field
Expected
Alphanumeric characters only

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

File transfer operation incomplete after database restoration

When DB Backup is executed during file transfer using Ver.1.30 or earlier and the backed-up database is restored, file transfer job remains incomplete in Process Viewer and transfer lamp continues flashing.

Ranked causes

  1. 1. Database backup executed during file transfer in Ver.1.30 then restored in Ver.1.40 (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Grid installation angle misalignment radiation

Relative angular misalignment between imaging unit sensor glass plate and grid exceeds acceptable tolerance

Ranked causes

  1. 1. Grid installed at incorrect angle relative to sensor (unknown)
First check

Verify detector type and image file extension

Test point
Image file extension
Expected
Extension .chi for CXDI-70C Wireless and subsequent products, or .crw for previous detectors (CXDI-40EG/EC, 50G/C, 55G/C, 60G/C)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Log database growth prevention

Configure logging settings to prevent database size overflow (4GB limit) that prevents system startup. Applicable to V1.01 and earlier; fixed in V1.05+.

Ranked causes

  1. 1. Operation logs cause database overflow in V1.01 or earlier (occasional)
First check

Verify software version

Test point
System information or About dialog
Expected
Version V1.01 or earlier; if V1.05 or later, no action required

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No calibration data present

Workspace calibration has not been completed; system cannot proceed with imaging without calibration

Ranked causes

  1. 1. Workspace calibration not completed (common)
First check

Verify calibration data exists for active workspace

Test point
Calibration data in system database
Expected
Valid calibration data present for target workspace

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No response during access point transmission when space appended to SSID

User enters space character at end of SSID text (example: 'abcd ') and sends information to access point using send button. System hangs with no response during encrypted transmission phase.

Ranked causes

  1. 1. Trailing space character in SSID field (common)
First check

Inspect SSID field for trailing spaces

Test point
SSID input field in AP/LNK Configuration tab
Expected
No space characters at end of SSID

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

OS network settings configuration for detector connection

Configure Windows network interface IP address and subnet mask for Power Box and detector connectivity

Ranked causes

  1. 1. Network interface requires predetermined IP configuration before detector registration (common)
First check

Open network status in Control Panel

Test point
Control Panel Network and Internet section
Expected
View network status and tasks option available

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Performance test: Detector maintenance and control checks radiation

Annual performance testing to verify detector maintenance and control based on previous calibration exposure settings

Ranked causes

  1. 1. Performance test required for detector maintenance and control (common)
First check

Adjust X-ray generator settings per previous calibration Exposure Memo

Test point
X-ray generator control panel
Expected
Settings match Exposure Memo from previous calibration

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Protocol Sheet import fails with duplicated Button layout lines

Error occurs when importing upgraded Protocol Sheet (from Ver.1.30 format) containing duplicated lines in Button layout sheet, even after successful upgrade to Ver.1.40 format.

Ranked causes

  1. 1. Duplicated lines present in Button layout sheet of Protocol Sheet (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Reject button not displayed in toolbar patient safety

After configuring toolbar settings to disable the Reject button in Group details Settings, the Reject button fails to appear when re-recording images, preventing cancellation of faulty recordings.

Ranked causes

  1. 1. Incorrect toolbar configuration: Retake button enabled and Reject button disabled simultaneously (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Required COM+ servers not installed

Necessary COM+ application servers are missing or not installed; fatal error prevents system operation

Ranked causes

  1. 1. COM+ servers not installed (rare)
First check

Verify all required COM+ servers are registered

Test point
COM+ service registry
Expected
MppsServer, PrintServer, StoreServer, ImageProcessingServer all registered

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

ROI adjustment fails after image darkening

ROI adjustment button does not function normally on preview image after exposure with CXDI-70C wireless; adjustment fails to apply

Ranked causes

  1. 1. CXDI Control Software NE/CXDI Controller RF Ver.1.30.0.17 with CXDI-70C wireless (common)
First check

Display two different images B and C before performing ROI adjustment

Test point
Preview image display
Expected
Images B and C displayed successfully

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ROI adjustment workaround procedure

Temporary workaround to perform ROI adjustment successfully on CXDI-70C wireless until bug-fixed software version is installed

Ranked causes

  1. 1. Software bug requires sequential image display before ROI adjustment (common)
First check

Select image A for ROI adjustment

Test point
Preview image display
Expected
Image A displayed on screen ready for adjustment

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self-diagnosis test execution

Monthly inspection of detectors and power box using test patterns to verify proper system operation.

Ranked causes

  1. 1. Routine system health check required (common)
First check

Select workspace for self-diagnosis

Test point
Detector/workspace list interface
Expected
Target workspace highlighted and selected

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Shell mode startup configuration

System startup configured for automatic login and application launch without desktop display

Ranked causes

  1. 1. Manual shell mode startup configuration needed (common)

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SSID configuration error with non-alphanumeric characters

Error displayed continuously and other symptoms occur when non-alphanumeric characters are input in SSID field of AP/LNK Tool Configuration tab in CCS V.1.30 or earlier

Ranked causes

  1. 1. Software limitation in CCS V.1.30 or earlier (common)
First check

Verify CCS software version installed

Test point
System configuration or software info screen
Expected
CCS V.1.40 or later

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System error E040501010 on database capacity overload

System error E040501010 occurs when database capacity exceeds 4GB SQL Server limit during continuous operation with Diskstore set to large media after 1000+ patient examinations

Ranked causes

  1. 1. Database capacity exceeds 4GB SQL Server upper limit (common)
First check

Verify current Diskstore save location and storage media capacity

Test point
Diskstore configuration settings
Expected
Save location set to DVD, USB memory ≤8GB, or storage media ≤8GB

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Workaround for ROI adjustment problem until upgrade

Temporary procedure to avoid ROI adjustment failure when using CXDI-70C wireless with affected software version

Ranked causes

  1. 1. Workaround required until Ver.1.30 bug-fix is installed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
020201003Detector remaining battery charge is lowActive detector battery drops below 4% of full chargeerror
020201004Gamma adjustment unavailable due to pattern file loading failureAll pattern files for simplified gamma adjustments fail to loaderror
020201005Failed to connect to barcode readerException during barcode reader connection initializationerror
020201006Cannot start specified examinationExamination specified during Studix operation or patient during magnetic card operation in ignored statuserror
020201007Detector communication error occurredReception timeout from detector or power-off during exposureerror
020201008Detector temperature exceeded allowable operating rangeDetector temperature reached dangerously high level, Sleep status activatederror
020201009Failed to start OverwrapSoftwareStartup failed due to invalid path or unexecutable file in OverwrapSoftware patherror
020201010Detector remaining battery charge is lowInactive detector battery drops below 4% of full chargeerror
020201011Imaging cannot be performed, no battery mounted in detectorWired connection attempted to detector with no battery mountederror
020201012Imaging cannot be performed due to low detector battery chargeActive detector battery drops below 4% or Ready selected for protocol with low battery detectorerror
020201013Imaging cannot be performed due to low detector battery chargeInactive detector battery drops below 4% or low-battery detector connectederror
020202001The detector temperature has reached warning levelsDetector temperature exceeds warning threshold as reported by Ferry subsystemwarning
020202002The detector temperature has reached dangerous levelsDetector temperature exceeds dangerous threshold as reported by Ferry subsystemwarning
020202003The detector is not available. Irradiation has finished.Detector becomes invalid during active X-ray irradiationerror
020202004Detector battery is weakeningDetector battery drops below 8% of full chargewarning
020202005Specified disk space could not be cleaned upSpace greater than examination to be deleted was specifiedwarning
020202006Could not acquire parameters required for reconfigurationReconfiguration parameters insufficient for Tomosynthesis protocolwarning
030100001No calibration dataSystem detects missing calibration data for active workspaceerror
030100005Required COM+ servers are not installed. contact a service engineer.System startup detects missing required COM+ serverscritical
030201001Detector communication errorAcquisition of detector information failed when inspection startederror
030201002Periodic noise detected in acquisition dataCalibration images taken without removing grid on grid-embedded detectorerror
030201003Cooling unit attached or detached during studyCooling unit attachment or detachment during active studyerror
030201004Calibration data check failedSPCC data check error on QCT completionerror
030201005Generator is disconnectedA generator disconnection event has occurrederror
030201006Generator communication failedGenerator communication failed when calibration, performance test or self-diagnosis startederror
030201007Generator disconnectedGenerator condition settings failederror
030201007Preparing for imaging failedSetting of conditions in generator or detector failederror
030201008Detector battery dischargedDetector battery charge level dropped below acceptable thresholderror
030201009Detector temperature exceeds operating rangeDetector temperature exceeded allowable operating rangeerror
030201010Detector battery not connectedBattery not installed with wired connectionerror
03020200EMemory allocation failedInsufficient memory when QC Tool started or examination startedcritical
03020201BCalibration data upload failedError occurred while SPCC data was being uploadedcritical
03020201CImage data transfer failedError during automatic image transfer without user switch operationcritical
03020201FDetector error occurredDetector error indication receivedcritical
030202021Control software/controller incompatible with detector firmwareFirmware compatibility check failedcritical
030202022Jumbo packet setting incorrect or incompatible network cardJumbo packet setting less than 4500 or network card not compatiblecritical
030203003Detector temperature exceeds operating rangeDetector temperature exceeded allowable operating rangewarning
03020FFFFInternal error occurredX-ray exposure and exposure cancel occur simultaneously or software bugcritical
040100001Invalid workspace specifiedUnregistered workspace seterror
040100008Invalid detector specifiedUnregistered detector specifiederror
040100023Generator communication failedInitialization processing of generator connection failederror
040100041Internal error occurredIllegal detector characteristics compensation dataerror
040100042Internal error occurredIllegal detector characteristics compensation dataerror
040100045Calibration data is invalidCalibration data not correctwarning
040100052Detector issued a warningWarning notification from detectorwarning
040100053Detector issued errorError notification received from detectorerror
040100054Detector issued fatal errorFatal error notification received from detectorcritical
040100056Generator disconnected during imaging setupRetry button canceled when setting imaging conditions in X-ray generator failederror
040100066IP addresses for wireless and wired reversedWired and wireless IP address settings reversed in Service Toolcritical
040101014Detector entered standby modeDetector not used for specified period while in Ready statewarning
040101021Connection to X-ray interface failedX-ray interface connection errorerror
040101029Battery charge-related error on detectorBattery charge-related error notification from detectorwarning
040300000The detector has not been started upA detector which has not been attached was selectederror
040300002An internal error occurredA detector which has not been attached was selected or hardware-related troubleerror
040300048Control software/controller incompatible with detector firmwareFirmware compatibility check failedcritical
040300050Jumbo packet setting incorrect or incompatible network cardJumbo packet setting less than 4500 or network card not compatiblecritical
040300051IP addresses for wireless and wired reversedWired and wireless IP address settings reversed in Service Toolcritical
040300053X-ray auto-detection enabled but detector does not support itX-ray auto-detection mode enabled but workspace with unsupported detector existscritical
040300054Detector firmware incompatibleDetector firmware compatibility check failedcritical
040400003Failed to acquire the system settingsDatabase data is defective or insufficient; table records insufficientcritical
040500003Exceeded the maximum number of the study order acquisitionsC-FIND-RSP study exceeding the maximum number of study order acquisitions set using the Service Tool have been receivederror
040500004Received some invalid studies. The invalid studies are not displayedData contravening DICOM rules contained in C-FIND-RSP receivederror
040500005The received tag contains an invalid value so the invalid value was replaced by nullData contravening DICOM rules contained in C-FIND-RSP receivederror
040500006Non-supported characters replaced with "?"Auxiliary kanjis received from RISerror
040500006Characters not included in the received character set were present. They may not be displayed correctlyCharacters which cannot be decoded properly were received from RIS or auxiliary kanjis received from RISerror
040500007Characters which do not match the current character code are used in this data. This character code will be properly changed automaticallyJapanese-language data received despite English (ASCII) set for Character Set of Service Toolwarning
040500008A date entry before 1753 is included in acquired ExamDate type (DA) data prior to January 1, 1753 received from RIS (PatientBirthDate or ScheduledProcedureStepStartDate)error
040500009Failed to copy a tag to C-STOREViolation of DICOM rules in data encountered when copying tags from C-FIND to C-STORE or tag contravenes DICOM ruleswarning
04050000AFailed to copy a tag to DicomFileViolation of DICOM rules in data when copying tags into DicomFile from C-FIND or tag contravenes DICOM ruleswarning
04050000CFailed to load DICOMDIRLoading of DICOMDIR failed because of damaged file or other reasonerror
04050000DAn output error occurredOutput information no longer available after system restart due to power discontinuity or other reasonerror
04050000FFailed to read DICOM fileLoading of DICOM file failed; file does not exist at specified location, file format non-compliant with DICOM, no access rights, or media corruptederror
040500010Failed to read DICOM file. Failed to create an image fileImage file creation failed due to insufficient memory, no access rights, or invalid image size tag valueerror
040501011Cannot continue the processing because a file already exists at the specified storage destinationFile with same name already exists at file save destinationerror
040501012Cannot continue processing; access to specified storage destination deniedSave destination folder cannot be accessederror
040501013Cannot continue processing; insufficient space at specified storage destinationNot enough space at specified storage destinationerror
040501014DICOM GSPS output was halted because the object image transfer was incomplete. Retry irradiation after object image transferC-STORE processing of GSPS executed while attempt to initiate C-STORE of target images (IHEEngine) failederror
040501016Network communication was cut off during file outputProblem occurred in network during media storage over network, making writing impossibleerror
040501017An unsupported DICOMDIR was loadedDICOMDIR file other than one created by CXDI NE/RF was loadederror
040502002Association negotiation failedAssociation negotiation failed with permanent rejection by service provider (ACSE-related function)error
040502003Association negotiation failedAssociation negotiation failed; abstract syntax sent not supported by SCPerror
040502004Association negotiation failedForward syntax sent is not supported by SCPerror
04050200DAssociation negotiation failedPermanent rejection by service provider (ACSE-related function), no reasonerror
04050200EAssociation negotiation failedPermanent rejection; protocol version not supportederror
04050200FAssociation negotiation failedTemporary rejection by service provider (ACSE-related), no reasonwarning
040502010Association negotiation failedTemporary rejection; protocol version not supportedwarning
040502011Association negotiation failedPermanent rejection by service provider (presentation-related); temporary congestionerror
040502012Association negotiation failedASSOCIATE-RJ received with RESULT=1, SOURCE=3, REASON=2; local restrictions exceedederror
040502013Association negotiation failedTemporary rejection by service provider (presentation-related); temporary congestionwarning
040502014Association negotiation failedTemporary rejection by service provider (presentation-related); local restrictions exceededwarning
04050201BAssociation was stoppedService provider aborted association; no reason givenwarning
04050201CAssociation was stoppedService provider aborted; PDU cannot be recognizederror
04050201DAssociation was stoppedService provider aborted; unexpected PDU receivederror
04050201EAssociation was stoppedService provider aborted; PDU parameters cannot be recognizederror
04050201FAssociation was stoppedService provider aborted; unexpected PDU parameterserror
040502020Association was stoppedService provider aborted; invalid PDU parameters valueerror
040502021Association negotiation failedNo response from communication partner within timeout periodwarning
040502023Association negotiation failedRELEASE-RQ received at time other than during association releaseerror
040502058Association negotiation failedNo communication partner; invalid IP address or porterror
040505001DICOM communication errorStatus tag (0000,0900) value error; non-existent attribute seterror
040505002DICOM communication errorStatus tag (0000,0900) value error; invalid attribute valueerror
040505003DICOM communication warning - attribute list errorStatus tag (0000,0900) value error with attribute list errorwarning
040505004DICOM communication errorStatus tag (0000,0900) value error; processing failederror
040505005DICOM communication errorStatus tag (0000,0900) value error; multiple SOP instanceserror
040505006DICOM communication errorStatus tag (0000,0900) value error; non-existent object instance seterror
040505007DICOM communication errorStatus tag (0000,0900) value error; non-existent event type seterror
040505008DICOM communication errorStatus tag (0000,0900) value error; non-existent argument seterror
040505009DICOM communication errorStatus tag (0000,0900) value error; invalid argument seterror
04050500ADICOM communication warning - attribute value out of rangeStatus tag (0000,0900) value error with out-of-range attribute valuewarning
04050500BDICOM communication errorStatus tag (0000,0900) value error; invalid object instance seterror
04050500CDICOM communication errorStatus tag (0000,0900) value error; non-existent SOP class seterror
04050500DDICOM communication errorStatus tag (0000,0900) value error; contradiction in class instanceerror
04050500EDICOM communication errorStatus tag (0000,0900) value error; missing attributeerror
04050500FDICOM communication errorStatus tag (0000,0900) value error; missing attribute value seterror
040505010DICOM communication errorStatus tag (0000,0900) value error; non-supported SOP class seterror
040505011DICOM communication errorStatus tag (0000,0900) value error; non-existent action type seterror
040505012DICOM communication errorStatus tag (0000,0900) value error; repeated startuperror
040505013DICOM communication error - status tag value errorStatus tag (0000,0900) value error received with unrecognized operationerror
040505014DICOM communication error - invalid argumentStatus tag (0000,0900) value error received with invalid argumenterror
040505015DICOM communication error - resource restrictionStatus tag (0000,0900) value error with resource restrictionerror
040505017DICOM communication error - incomprehensible settingStatus tag (0000,0900) value error with incomprehensible settingerror
040505018DICOM communication error - insufficient resourcesStatus tag (0000,0900) value error with insufficient resourceserror
04050501BDICOM communication error - C-FIND identifier mismatchC-FIND status tag (0000,0900) value error, identifier does not match SOP classerror
04050501CDICOM communication warning - data factor coercionC-STORE status tag (0000,0900) value error with data factor coercionwarning
04050501DDICOM communication warning - factor abandonedC-STORE status tag (0000,0900) value error with factor abandonedwarning
04050501EDICOM communication warning - dataset mismatchC-STORE status tag (0000,0900) value error, dataset does not match SOP classwarning
04050501FDICOM communication error - C-STORE dataset mismatchC-STORE status tag (0000,0900) value error, dataset does not match SOP classerror
040505020DICOM communication error - C-STORE no resourcesC-STORE status tag (0000,0900) value error with no resourceserror
040505027DICOM printer output error - film session structure invalidN-ACTION status tag (0000,0900) value error, BasicFilmSession SOP class, missing film boxerror
040505028DICOM printer output error - print queue fullN-ACTION status tag (0000,0900) value error, BasicFilmSession, print queue fullerror
040505029DICOM printer output error - image size exceeds box sizeN-ACTION status tag (0000,0900) value error, BasicFilmSession, image too largeerror
040505029Error occurred in DICOM printer outputStatus tag (0000,0900) value of N-ACTION when SOP class was BasicFilmSession received; image size larger than image box sizeerror
04050502ADICOM printer output error - linked image size exceeds boxN-ACTION status tag (0000,0900) value error, BasicFilmSession, linked image too largeerror
04050502AError in DICOM printer outputStatus tag (0000,0900) value of N-ACTION when SOP class was BasicFilmSession; linked print image size larger than image box sizeerror
04050502CDICOM printer output warning - print queue fullN-ACTION status tag (0000,0900) value error, BasicFilmSession, print queue fullerror
04050502EError in DICOM printer outputStatus tag (0000,0900) value of N-CREATE when SOP class was BasicFilmBox; existing unprinted film box exists; N-ACTION at film session level unsupportederror
04050502EDICOM printer output error - unprinted film box existsN-CREATE status tag (0000,0900) value error, BasicFilmSession, unprinted film box existserror
04050502FDICOM printer output error - printer memory insufficientN-SET status tag (0000,0900) value error, GrayScaleImageBox, insufficient printer memoryerror
040505030DICOM printer status warning receivedPrinter status tag (2110,0010) value of N-GET during Printer operation is WARNINGwarning
040505031DICOM communication response errorStatus tag (0000,0900) value not defined by DICOM received from communication partyerror
040506001DICOM printer failure detectedFAILURE status tag (2110,0010) value of N-GET from DICOM printererror
040507001DICOM StorageCommitment failed - reason 0x0110N-EVENT-REPORT failure of StorageCommitment with reason 0x0110error
040507002DICOM StorageCommitment failed - reason 0x0112N-EVENT-REPORT failure of StorageCommitment with reason 0x0112error
040507003DICOM StorageCommitment failed - reason 0x0119N-EVENT-REPORT failure of StorageCommitment with reason 0x0119error
040507004DICOM StorageCommitment failed - reason 0x0122N-EVENT-REPORT failure of StorageCommitment with reason 0x0122error
040507005DICOM StorageCommitment failed - reason 0x0131N-EVENT-REPORT failure of StorageCommitment with reason 0x0131error
040507006DICOM StorageCommitment failed - undefined reasonN-EVENT-REPORT failure of StorageCommitment with undefined reason codeerror
040507007DICOM StorageCommitment failed - undefined reasonN-EVENT-REPORT failure of StorageCommitment with undefined reason codeerror
040508001DICOM disc creation failedUnexpected error during DICOM disc creation processerror
040508002DICOM disc creation failed; disc recorder not foundSpecified disc recorder not detected during DICOM disc creation attempterror
040600008Failed to access target diskDisk drive inaccessible during free space check at media storage time (drive open, damaged, etc.)error
040600009Process interrupted; insufficient free disk spaceFree space check at media storage shows inadequate disk space availableerror
04060000BRetry not possible; error occurred during processPower discontinuity or software bug; key ID missing or image processing failederror
04060000CCannot execute old study information transferSystem attempting to transfer study created before data rebinding; same image file in useerror
04060000DOperation failed; write destination busySame media drive used for both media storage and DICOM disc creationerror
04060000ERetry not possible; DICOM disc creation error occurredDICOM disc creation failed; error displayed in process viewererror
040700006Failed to load image fileSystem attempts to read file that does not existerror
040700012Graphics card does not meet required specs; imaging may failInstalled graphics card performance below requirementscritical
040700100System memory settings invalidMemory size larger than installed memory specified by Service Toolcritical
040800100Examination cannot be started; non-existent protocol includedProtocol contained in pending examination has been deleted from databaseerror
040D0001FProviding free space on hard disk failedData equivalent to prescribed number of records cannot be deletederror
040D00201Network interface in monitoring settings not found; monitoring function disabledNetwork interface malfunction or improper connection during monitoring setuperror
040E01000Image analysis processing failedImage analysis processing failure during XdCom processingerror
040E01001Failed to load extra defect correction fileSystem unable to load extra defect correction fileerror
041000004Invalid studies are acquiredExam information containing illegal values for key items (ACC#, patient ID, patient name) acquired from Studixerror
041000005Entered information has invalid valueExam information containing illegal values for non-key items acquired from Studixerror
041000006Specified ID not found in examination fileExam information with specified key values did not exist in examination fileerror
041000008Error occurred during script-based input calibrationScript compilation error, script processing error, or abnormal script return valueerror
041000011This version of workflow library not supportedVersion mismatch between CCS and automatic examination packerror
041000012Error occurred in CCRHIS libraryStudix, Workflow, or magnetic card processing failederror
041000020Examination file loading halted - maximum reserved exams exceededMaximum number of reserved exams that can be registered to database exceedederror
041000021Invalid studies loaded and cannot be displayedInvalid value detected among values of required itemserror
041000022Entered information has invalid value in exam fileInvalid value detected among user-specified valueserror
041000023Failed to read examination fileExam file was not founderror
041000024Error occurred during automatic examination file updatingError during exam file monitoring when configuring automatic updateerror
041000025Failed to read examination fileError occurred during loading of exam fileerror
041100004CxdiInfo startup failedSpecified DLL does not exist or specified DLL lacks expected interfaceerror
050501002Flash ROM error detected in detectorFlash ROM ID error detected by detector AD boardwarning
050501003Flash ROM error detected in detectorFlash ROM Write error detected by detector AD boardwarning
050501004Flash ROM error detected in detectorFlash ROM Erase error detected by detector AD boardwarning
050501005There may be problem in images because of grid errorGrid warning detected by detector AD board during imagingwarning
050501005Grid error may have caused image abnormalityGrid warning detected but imaging continuedwarning
050501006Detector internal temperature exceeds upper limitTemperature inside detector exceeds upper limiterror
050501007Detector internal temperature approaching upper limitInternal detector temperature approaching upper limitwarning
050501008Drive circuit error occurredPLD error detected by detector AD boarderror
050501009Drive circuit error - PLD config errorPLD Config error detected by detector AD boarderror
05050100AAbnormality detected in detector structureIllegal value written for serial numbererror
05050100ADetector structure abnormality - invalid serial numberIllegal value written for serial numbererror
05050100BDetector structure abnormality - serial number mismatchDiscrepancy between detector serial number and DP numbererror
05050100BAbnormality detected in detector structureDiscrepancy between serial number of detector and DP numbererror
05050100CControl line connection error detected in detectorControl wire error detected by detector AD board, broken cable between detector and power boxerror
05050100DDetector control line connection error - cable length mismatchDiscrepancy between detector-power box cable length settings and XRAY board cableerror
05050100EUndefined error detected by detectorError detected in Hex format received when firmware or PLD was downloadedwarning
05050100FUndefined error detected by detectorError detected in Hex format received when firmware, PLD or DP file was downloadedwarning
050501010Failed to transfer image dataDMA timeout inside PCB-50Di when reduced-size images were readerror
050501011Transfer of image data was stoppedTransfer of Full image data was abortedwarning
050501012Abnormality detected in analog power supply for detectorAbnormality detected in analog power supply for detectorerror
050501013Grid ID loading error detectedGrid ID loading error was detectedwarning
050501014Grid error detectedGrid error detected at detector AD boarderror
050501015Grid error may have caused image abnormalityGrid warning detected by detector AD board but imaging continuedwarning
050501016Firmware version mismatchWrong detector firmware files written during firmware updateerror
050502009Detector communication error; cxdcap.ini missingcxdcap.ini file not installedcritical
050502026Detector communication error via EthernetCommunication error detected by Ethernet interfaceerror
E010156203SSID field contains non-alphanumeric characterPressing Save button after inputting special character (e.g. '(') in SSID fielderror
E040501010Database capacity exceeded SQL Server upper limitDiskstore save location set to large capacity media with 1000+ patient examinations performed continuously without system shutdownerror
ERROR_CHECK_LATERCheck later error allowing continued imagingNon-critical error such as DICOM communication fault occurswarning
ERROR_IMMEDIATEImmediate display error requiring user actionImaging failure such as exposure attempted but images not transferred or detector error occurserror
F040700012GPU does not meet required specifications or driver version insufficientCCS startup when GPU processing is enabled; GPU Compute capability below 2.0 or NVIDIA driver version too oldcritical
FATALFatal error condition requiring service callSystem experiences unrecoverable error such as detector no response or DB connection failurecritical
I0111SQL Server 2005 setup alert during system software upgradeUpgrading system software from Ver.1.30 or earlier to Ver.1.40warning
No second monitor is connectedSecond monitor detection failure during CCS startupWhen Use second monitor checkbox is enabled in service tool but system cannot detect connected secondary monitorerror
Parameter initialization requiredParameter file reset to default valuesaudit_grid_NE_RF_P.txt deleted after parameter configurationwarning
WARNINGWarning notification not requiring immediate actionNon-critical conditions such as low film, backup needed, detector temperature slightly elevatedwarning
Wrong image sizes errorParameter configuration error in audit_grid_NE_RF_P.txtCRW image files analyzed with incorrect image size parametererror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI