Canon CXDI Control Software NE: troubleshooting and error codes
The ServiceForge knowledge base holds 44 documented symptoms for the Canon CXDI Control Software NE, with 55 ranked causes, 103 diagnostic checks with test points and expected values, and 214 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Second monitor image display magnification and cropping
Configuration of 3-phase fixed zooming ratios, threshold levels, and image cropping for second monitor display based on detector pixel pitch (0.160 mm or 0.125 mm) and monitor resolution
Ranked causes
- 1. Service tool registration of second monitor resolution (common)
- 2. Image cropping to first phase threshold levels triggers 100% magnification (common)
- 3. Image cropping between first and second phase thresholds triggers second phase magnification (common)
- 4. Image cropping to full screen or large size triggers third phase magnification (common)
Register second monitor resolution in service tool
- Test point
- Service tool 'Set connected second monitor size' button
- Expected
- Orientation and resolution registered; [Apply] button clicked
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AEC Calibration protocol grid parameter error on Excel import
Error occurs when loading Protocol Sheet after importing Excel file exported from Protocol Database. Inappropriate grid parameter combination (GridType Off, Grid Suppression On) causes import failure.
Ranked causes
- 1. Inappropriate default AEC Calibration grid parameters installed (common)
- 2. Previously exported Excel file contains inconsistent grid parameters (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AEC Calibration Protocol setup for new detector
Need to create or modify AEC Calibration Protocol when adding a new detector to the system
Ranked causes
- 1. AEC Calibration Protocol deleted from database (occasional)
- 2. New detector added without protocol update (occasional)
Verify workspace exists for detector
- Test point
- Service Tool Protocol Editor
- Expected
- Workspace listed for detector
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error with multiple detectors connected
Fatal error occurs during first calibration after detector detection with message about missing dp data
Ranked causes
- 1. Multiple detectors connected during initial calibration (common)
- 2. Version earlier than V1.05 installed (common)
Verify number of detectors connected
- Test point
- Hardware connection
- Expected
- Only one detector connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Grid installation angle measurement radiation
Measure grid installation angle relative to imaging unit sensor glass plate using audit_grid_NE_RF.exe software tool to verify alignment is within specification.
Ranked causes
- 1. Grid installation angle must be verified and corrected (common)
- 2. Image file format may require parameter configuration (occasional)
Prepare software files
- Test point
- Image Capture Computer
- Expected
- audit_grid_NE_RF.exe and SecureConverter.dll copied to computer
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator communication settings configuration
Configure X-ray generator communication parameters including protocol, IP address, and display settings in service tool
Ranked causes
- 1. Generator communication settings must be configured for software to control X-ray exposure (common)
- 2. Response timeout and network address configuration (occasional)
Open Generator settings from service tool TreeView
- Test point
- Service tool TreeView Generator menu
- Expected
- Generator setting screen displays
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software installation component selection and uninstallation cleanup
Select CANONDRSYSTEM component during installation and remove installation folder after uninstall
Ranked causes
- 1. Installation requires selection of CANONDRSYSTEM component (common)
- 2. Installation folder remains after program deletion and requires manual removal (occasional)
Select CANONDRSYSTEM from component list
- Test point
- Component Selection screen checkbox
- Expected
- CANONDRSYSTEM checkbox selected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade loses configuration settings
Certain settings revert to default after upgrading CXDI Control Software NE to new version
Ranked causes
- 1. Software upgrade process replaces configuration database (common)
- 2. Overwrap Software installed in V1.0.0 before upgrade (occasional)
Backup database before upgrade
- Test point
- Database management
- Expected
- Backup file created
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Workspace creation and detector association
Create predetermined or custom workspace configuration and associate detector with position type
Ranked causes
- 1. Workspace must be created and associated with detector before calibration (common)
- 2. Custom workspace with optional protocol association (occasional)
Open Workspace property settings in Service Tool
- Test point
- Property Editor Workspace menu
- Expected
- Workspace configuration screen displays
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alert window appears during SQL Server 2005 setup when upgrading from Ver.1.30 to Ver.1.40
Version upgrade processing is suspended when upgrading from Ver.1.30 or earlier to Ver.1.40; alert window is displayed during SQL Server 2005 setup
Ranked causes
- 1. SQL Server 2005 setup alert during version upgrade from Ver.1.30 to Ver.1.40 (common)
Observe whether alert window appears during SQL Server 2005 setup phase of upgrade
- Test point
- SQL Server 2005 setup
- Expected
- Alert window may appear during upgrade from Ver.1.30 to Ver.1.40
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AEC calibration protocol selection radiation
Select and activate AEC calibration protocol for exposure testing and adjustment
Ranked causes
- 1. AEC calibration protocol must be selected for testing (common)
Confirm AEC calibration protocols are available in tray
- Test point
- Protocol tray
- Expected
- AEC calibration protocols visible from section 4.8.2 setup
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector correction data creation radiation
Initial detector calibration and annual maintenance calibration to establish correction data for accurate imaging
Ranked causes
- 1. Detector calibration required during installation or annual maintenance (common)
Establish calibration mode and select detector
- Test point
- Detector list in calibration interface
- Expected
- Detector selectable from list
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EI display configuration
Configure EI (Exposure Index) display on screen at bottom right corner for proper visibility during examinations
Ranked causes
- 1. EI display not visible or requires initial setup (common)
Verify EI is displayed on screen
- Test point
- Bottom right corner of display
- Expected
- EI value visible in bottom right corner
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Grid installation angle radiation
Verify that anti-scatter grid is installed at correct angle relative to X-ray focal point; angle must be within ±1.0° tolerance across three test images
Ranked causes
- 1. Grid installation angle out of tolerance (unknown)
Measure grid installation angle using audit_grid_NE_RF.exe on first image
- Test point
- Grid relative to X-ray focal point
- Expected
- ±1.0 or less, ±1.0°
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Phototimer adjustment radiation
Adjust photo-timer using Exposure Index (EI) to optimize patient dose and image quality. Requires exposing acrylics sheets at specified voltages and adjusting to target EI values per imaging unit model.
Ranked causes
- 1. Phototimer requires adjustment during installation (common)
Confirm EI is displayed on Examination Screen
- Test point
- System Settings > Annotation tab
- Expected
- EI displayed at one of four corners
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Phototimer adjustment for AEC radiation
Adjust photo-timer to achieve target EI values at different kV settings (70, 100, 120 kV) using acrylic sheets; spectral sensitivity adjustment required
Ranked causes
- 1. Phototimer requires calibration for correct EI output across voltage settings (common)
Expose X-ray through 5 cm acrylic at 70 kV and measure EI
- Test point
- Detector readout
- Expected
- EI 480 for CXDI-40EG/40G/401G/401G Compact/50G/55G/60G/501G/70G/701G/801G Wireless; EI 280 for CXDI-40EC/401C/401C Compact/50C/55C/501C/60C/70C/80C/701C/801C/50RF
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ROI adjustment after CXDI-70C wireless exposure
Workaround procedure to enable normal ROI adjustment on preview image after exposure with CXDI-70C wireless detector until bug-fixed software version is available.
Ranked causes
- 1. Software bug in Ver.1.30.0.17 requires intermediate image display before ROI adjustment (common)
Select image B from available protocols or past image list
- Test point
- CXDI Control Software NE image selection interface
- Expected
- Image B displayed (must be different from target image A and previously exposed or in protocol)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ROI adjustment workaround for CXDI-70C wireless
Procedure to perform successful ROI adjustment on preview images when using CXDI-70C wireless detector with affected software version
Ranked causes
- 1. Display sequence requirement for ROI adjustment with CXDI-70C wireless Ver.1.30.0.17 (common)
Select image A for ROI adjustment
- Test point
- preview image
- Expected
- Image A displayed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Second monitor display settings configuration
Configuration of preview annotations, laterality markers, and peripheral mask display on second monitor
Ranked causes
- 1. Second monitor display setting configuration via service tool (common)
Confirm second monitor is connected and enabled
- Test point
- Service tool configuration
- Expected
- 'Use second monitor' is ON and 2 monitors connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Second monitor gamma adjustment
Adjust gamma levels for secondary monitor in dual-monitor configuration using Gamma Adjustment Tool
Ranked causes
- 1. Second monitor gamma requires calibration (common)
Verify second monitor is physically connected to video board DVI output
- Test point
- Video board DVI connector for second monitor
- Expected
- Second monitor connected and recognized by system
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Second monitor setup and configuration
Configure second monitor for dual-monitor operation including resolution, display options, and grid suppression algorithm
Ranked causes
- 1. Second monitor requires initial setup configuration (common)
Verify video board has required DVI connectivity for dual monitors
- Test point
- Video board output connectors
- Expected
- Two DVI outputs present or DisplayPort-to-DVI converter available (for NVIDIA Quadro600)
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector registration with IP address configuration
Register X-ray detector with predetermined IP address and enroll detector information in service tool system
Ranked causes
- 1. Detector must be registered in service tool before operation (common)
Connect detector to network
- Test point
- Network connection to Power Box
- Expected
- Detector connected and communicable
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector temperature dangerous level caution
Detector temperature has reached dangerous level; system displays warning button; imaging can continue
Ranked causes
- 1. Detector temperature reached dangerous threshold (common)
Monitor detector temperature
- Test point
- Ferry detector temperature sensor
- Expected
- Temperature below dangerous threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector temperature warning caution
Detector temperature has reached warning level; system displays warning button in GUI header; imaging can continue
Ranked causes
- 1. Detector temperature reached warning threshold (common)
Monitor detector temperature via system notifications
- Test point
- Ferry detector temperature sensor
- Expected
- Temperature below warning threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector unavailable during irradiation patient safety
Detector becomes invalid during X-ray exposure; irradiation stops automatically; error does not occur without user intervention
Ranked causes
- 1. Detector touched or disturbed during irradiation (occasional)
Verify detector status before irradiation
- Test point
- Detector availability status in system
- Expected
- Detector available and not touched during irradiation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Double enlargement ratio compensation in distance measurement
DR Image Viewer applies enlargement ratio compensation twice when measuring distance on images that already have compensated enlargement ratio, producing incorrect results.
Ranked causes
- 1. Enlargement ratio compensation enabled in both NE/RF and DR Image Viewer (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error E010156203 displayed persistently after entering special character in SSID field
User inputs non-alphanumeric character like '(' in SSID field and presses Save button. Error E010156203 appears. Even after replacing the illegal character with valid input, the error persists permanently when Save is pressed again.
Ranked causes
- 1. Non-alphanumeric character entered in SSID field (common)
Verify SSID field contains only alphanumeric characters
- Test point
- Configuration tab SSID field
- Expected
- Alphanumeric characters only
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
File transfer operation incomplete after database restoration
When DB Backup is executed during file transfer using Ver.1.30 or earlier and the backed-up database is restored, file transfer job remains incomplete in Process Viewer and transfer lamp continues flashing.
Ranked causes
- 1. Database backup executed during file transfer in Ver.1.30 then restored in Ver.1.40 (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grid installation angle misalignment radiation
Relative angular misalignment between imaging unit sensor glass plate and grid exceeds acceptable tolerance
Ranked causes
- 1. Grid installed at incorrect angle relative to sensor (unknown)
Verify detector type and image file extension
- Test point
- Image file extension
- Expected
- Extension .chi for CXDI-70C Wireless and subsequent products, or .crw for previous detectors (CXDI-40EG/EC, 50G/C, 55G/C, 60G/C)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Log database growth prevention
Configure logging settings to prevent database size overflow (4GB limit) that prevents system startup. Applicable to V1.01 and earlier; fixed in V1.05+.
Ranked causes
- 1. Operation logs cause database overflow in V1.01 or earlier (occasional)
Verify software version
- Test point
- System information or About dialog
- Expected
- Version V1.01 or earlier; if V1.05 or later, no action required
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No calibration data present
Workspace calibration has not been completed; system cannot proceed with imaging without calibration
Ranked causes
- 1. Workspace calibration not completed (common)
Verify calibration data exists for active workspace
- Test point
- Calibration data in system database
- Expected
- Valid calibration data present for target workspace
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No response during access point transmission when space appended to SSID
User enters space character at end of SSID text (example: 'abcd ') and sends information to access point using send button. System hangs with no response during encrypted transmission phase.
Ranked causes
- 1. Trailing space character in SSID field (common)
Inspect SSID field for trailing spaces
- Test point
- SSID input field in AP/LNK Configuration tab
- Expected
- No space characters at end of SSID
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OS network settings configuration for detector connection
Configure Windows network interface IP address and subnet mask for Power Box and detector connectivity
Ranked causes
- 1. Network interface requires predetermined IP configuration before detector registration (common)
Open network status in Control Panel
- Test point
- Control Panel Network and Internet section
- Expected
- View network status and tasks option available
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance test: Detector maintenance and control checks radiation
Annual performance testing to verify detector maintenance and control based on previous calibration exposure settings
Ranked causes
- 1. Performance test required for detector maintenance and control (common)
Adjust X-ray generator settings per previous calibration Exposure Memo
- Test point
- X-ray generator control panel
- Expected
- Settings match Exposure Memo from previous calibration
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Protocol Sheet import fails with duplicated Button layout lines
Error occurs when importing upgraded Protocol Sheet (from Ver.1.30 format) containing duplicated lines in Button layout sheet, even after successful upgrade to Ver.1.40 format.
Ranked causes
- 1. Duplicated lines present in Button layout sheet of Protocol Sheet (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reject button not displayed in toolbar patient safety
After configuring toolbar settings to disable the Reject button in Group details Settings, the Reject button fails to appear when re-recording images, preventing cancellation of faulty recordings.
Ranked causes
- 1. Incorrect toolbar configuration: Retake button enabled and Reject button disabled simultaneously (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Required COM+ servers not installed
Necessary COM+ application servers are missing or not installed; fatal error prevents system operation
Ranked causes
- 1. COM+ servers not installed (rare)
Verify all required COM+ servers are registered
- Test point
- COM+ service registry
- Expected
- MppsServer, PrintServer, StoreServer, ImageProcessingServer all registered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROI adjustment fails after image darkening
ROI adjustment button does not function normally on preview image after exposure with CXDI-70C wireless; adjustment fails to apply
Ranked causes
- 1. CXDI Control Software NE/CXDI Controller RF Ver.1.30.0.17 with CXDI-70C wireless (common)
Display two different images B and C before performing ROI adjustment
- Test point
- Preview image display
- Expected
- Images B and C displayed successfully
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROI adjustment workaround procedure
Temporary workaround to perform ROI adjustment successfully on CXDI-70C wireless until bug-fixed software version is installed
Ranked causes
- 1. Software bug requires sequential image display before ROI adjustment (common)
Select image A for ROI adjustment
- Test point
- Preview image display
- Expected
- Image A displayed on screen ready for adjustment
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-diagnosis test execution
Monthly inspection of detectors and power box using test patterns to verify proper system operation.
Ranked causes
- 1. Routine system health check required (common)
Select workspace for self-diagnosis
- Test point
- Detector/workspace list interface
- Expected
- Target workspace highlighted and selected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Shell mode startup configuration
System startup configured for automatic login and application launch without desktop display
Ranked causes
- 1. Manual shell mode startup configuration needed (common)
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SSID configuration error with non-alphanumeric characters
Error displayed continuously and other symptoms occur when non-alphanumeric characters are input in SSID field of AP/LNK Tool Configuration tab in CCS V.1.30 or earlier
Ranked causes
- 1. Software limitation in CCS V.1.30 or earlier (common)
Verify CCS software version installed
- Test point
- System configuration or software info screen
- Expected
- CCS V.1.40 or later
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error E040501010 on database capacity overload
System error E040501010 occurs when database capacity exceeds 4GB SQL Server limit during continuous operation with Diskstore set to large media after 1000+ patient examinations
Ranked causes
- 1. Database capacity exceeds 4GB SQL Server upper limit (common)
Verify current Diskstore save location and storage media capacity
- Test point
- Diskstore configuration settings
- Expected
- Save location set to DVD, USB memory ≤8GB, or storage media ≤8GB
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Workaround for ROI adjustment problem until upgrade
Temporary procedure to avoid ROI adjustment failure when using CXDI-70C wireless with affected software version
Ranked causes
- 1. Workaround required until Ver.1.30 bug-fix is installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
020201003 | Detector remaining battery charge is low | Active detector battery drops below 4% of full charge | error |
020201004 | Gamma adjustment unavailable due to pattern file loading failure | All pattern files for simplified gamma adjustments fail to load | error |
020201005 | Failed to connect to barcode reader | Exception during barcode reader connection initialization | error |
020201006 | Cannot start specified examination | Examination specified during Studix operation or patient during magnetic card operation in ignored status | error |
020201007 | Detector communication error occurred | Reception timeout from detector or power-off during exposure | error |
020201008 | Detector temperature exceeded allowable operating range | Detector temperature reached dangerously high level, Sleep status activated | error |
020201009 | Failed to start OverwrapSoftware | Startup failed due to invalid path or unexecutable file in OverwrapSoftware path | error |
020201010 | Detector remaining battery charge is low | Inactive detector battery drops below 4% of full charge | error |
020201011 | Imaging cannot be performed, no battery mounted in detector | Wired connection attempted to detector with no battery mounted | error |
020201012 | Imaging cannot be performed due to low detector battery charge | Active detector battery drops below 4% or Ready selected for protocol with low battery detector | error |
020201013 | Imaging cannot be performed due to low detector battery charge | Inactive detector battery drops below 4% or low-battery detector connected | error |
020202001 | The detector temperature has reached warning levels | Detector temperature exceeds warning threshold as reported by Ferry subsystem | warning |
020202002 | The detector temperature has reached dangerous levels | Detector temperature exceeds dangerous threshold as reported by Ferry subsystem | warning |
020202003 | The detector is not available. Irradiation has finished. | Detector becomes invalid during active X-ray irradiation | error |
020202004 | Detector battery is weakening | Detector battery drops below 8% of full charge | warning |
020202005 | Specified disk space could not be cleaned up | Space greater than examination to be deleted was specified | warning |
020202006 | Could not acquire parameters required for reconfiguration | Reconfiguration parameters insufficient for Tomosynthesis protocol | warning |
030100001 | No calibration data | System detects missing calibration data for active workspace | error |
030100005 | Required COM+ servers are not installed. contact a service engineer. | System startup detects missing required COM+ servers | critical |
030201001 | Detector communication error | Acquisition of detector information failed when inspection started | error |
030201002 | Periodic noise detected in acquisition data | Calibration images taken without removing grid on grid-embedded detector | error |
030201003 | Cooling unit attached or detached during study | Cooling unit attachment or detachment during active study | error |
030201004 | Calibration data check failed | SPCC data check error on QCT completion | error |
030201005 | Generator is disconnected | A generator disconnection event has occurred | error |
030201006 | Generator communication failed | Generator communication failed when calibration, performance test or self-diagnosis started | error |
030201007 | Generator disconnected | Generator condition settings failed | error |
030201007 | Preparing for imaging failed | Setting of conditions in generator or detector failed | error |
030201008 | Detector battery discharged | Detector battery charge level dropped below acceptable threshold | error |
030201009 | Detector temperature exceeds operating range | Detector temperature exceeded allowable operating range | error |
030201010 | Detector battery not connected | Battery not installed with wired connection | error |
03020200E | Memory allocation failed | Insufficient memory when QC Tool started or examination started | critical |
03020201B | Calibration data upload failed | Error occurred while SPCC data was being uploaded | critical |
03020201C | Image data transfer failed | Error during automatic image transfer without user switch operation | critical |
03020201F | Detector error occurred | Detector error indication received | critical |
030202021 | Control software/controller incompatible with detector firmware | Firmware compatibility check failed | critical |
030202022 | Jumbo packet setting incorrect or incompatible network card | Jumbo packet setting less than 4500 or network card not compatible | critical |
030203003 | Detector temperature exceeds operating range | Detector temperature exceeded allowable operating range | warning |
03020FFFF | Internal error occurred | X-ray exposure and exposure cancel occur simultaneously or software bug | critical |
040100001 | Invalid workspace specified | Unregistered workspace set | error |
040100008 | Invalid detector specified | Unregistered detector specified | error |
040100023 | Generator communication failed | Initialization processing of generator connection failed | error |
040100041 | Internal error occurred | Illegal detector characteristics compensation data | error |
040100042 | Internal error occurred | Illegal detector characteristics compensation data | error |
040100045 | Calibration data is invalid | Calibration data not correct | warning |
040100052 | Detector issued a warning | Warning notification from detector | warning |
040100053 | Detector issued error | Error notification received from detector | error |
040100054 | Detector issued fatal error | Fatal error notification received from detector | critical |
040100056 | Generator disconnected during imaging setup | Retry button canceled when setting imaging conditions in X-ray generator failed | error |
040100066 | IP addresses for wireless and wired reversed | Wired and wireless IP address settings reversed in Service Tool | critical |
040101014 | Detector entered standby mode | Detector not used for specified period while in Ready state | warning |
040101021 | Connection to X-ray interface failed | X-ray interface connection error | error |
040101029 | Battery charge-related error on detector | Battery charge-related error notification from detector | warning |
040300000 | The detector has not been started up | A detector which has not been attached was selected | error |
040300002 | An internal error occurred | A detector which has not been attached was selected or hardware-related trouble | error |
040300048 | Control software/controller incompatible with detector firmware | Firmware compatibility check failed | critical |
040300050 | Jumbo packet setting incorrect or incompatible network card | Jumbo packet setting less than 4500 or network card not compatible | critical |
040300051 | IP addresses for wireless and wired reversed | Wired and wireless IP address settings reversed in Service Tool | critical |
040300053 | X-ray auto-detection enabled but detector does not support it | X-ray auto-detection mode enabled but workspace with unsupported detector exists | critical |
040300054 | Detector firmware incompatible | Detector firmware compatibility check failed | critical |
040400003 | Failed to acquire the system settings | Database data is defective or insufficient; table records insufficient | critical |
040500003 | Exceeded the maximum number of the study order acquisitions | C-FIND-RSP study exceeding the maximum number of study order acquisitions set using the Service Tool have been received | error |
040500004 | Received some invalid studies. The invalid studies are not displayed | Data contravening DICOM rules contained in C-FIND-RSP received | error |
040500005 | The received tag contains an invalid value so the invalid value was replaced by null | Data contravening DICOM rules contained in C-FIND-RSP received | error |
040500006 | Non-supported characters replaced with "?" | Auxiliary kanjis received from RIS | error |
040500006 | Characters not included in the received character set were present. They may not be displayed correctly | Characters which cannot be decoded properly were received from RIS or auxiliary kanjis received from RIS | error |
040500007 | Characters which do not match the current character code are used in this data. This character code will be properly changed automatically | Japanese-language data received despite English (ASCII) set for Character Set of Service Tool | warning |
040500008 | A date entry before 1753 is included in acquired Exam | Date type (DA) data prior to January 1, 1753 received from RIS (PatientBirthDate or ScheduledProcedureStepStartDate) | error |
040500009 | Failed to copy a tag to C-STORE | Violation of DICOM rules in data encountered when copying tags from C-FIND to C-STORE or tag contravenes DICOM rules | warning |
04050000A | Failed to copy a tag to DicomFile | Violation of DICOM rules in data when copying tags into DicomFile from C-FIND or tag contravenes DICOM rules | warning |
04050000C | Failed to load DICOMDIR | Loading of DICOMDIR failed because of damaged file or other reason | error |
04050000D | An output error occurred | Output information no longer available after system restart due to power discontinuity or other reason | error |
04050000F | Failed to read DICOM file | Loading of DICOM file failed; file does not exist at specified location, file format non-compliant with DICOM, no access rights, or media corrupted | error |
040500010 | Failed to read DICOM file. Failed to create an image file | Image file creation failed due to insufficient memory, no access rights, or invalid image size tag value | error |
040501011 | Cannot continue the processing because a file already exists at the specified storage destination | File with same name already exists at file save destination | error |
040501012 | Cannot continue processing; access to specified storage destination denied | Save destination folder cannot be accessed | error |
040501013 | Cannot continue processing; insufficient space at specified storage destination | Not enough space at specified storage destination | error |
040501014 | DICOM GSPS output was halted because the object image transfer was incomplete. Retry irradiation after object image transfer | C-STORE processing of GSPS executed while attempt to initiate C-STORE of target images (IHEEngine) failed | error |
040501016 | Network communication was cut off during file output | Problem occurred in network during media storage over network, making writing impossible | error |
040501017 | An unsupported DICOMDIR was loaded | DICOMDIR file other than one created by CXDI NE/RF was loaded | error |
040502002 | Association negotiation failed | Association negotiation failed with permanent rejection by service provider (ACSE-related function) | error |
040502003 | Association negotiation failed | Association negotiation failed; abstract syntax sent not supported by SCP | error |
040502004 | Association negotiation failed | Forward syntax sent is not supported by SCP | error |
04050200D | Association negotiation failed | Permanent rejection by service provider (ACSE-related function), no reason | error |
04050200E | Association negotiation failed | Permanent rejection; protocol version not supported | error |
04050200F | Association negotiation failed | Temporary rejection by service provider (ACSE-related), no reason | warning |
040502010 | Association negotiation failed | Temporary rejection; protocol version not supported | warning |
040502011 | Association negotiation failed | Permanent rejection by service provider (presentation-related); temporary congestion | error |
040502012 | Association negotiation failed | ASSOCIATE-RJ received with RESULT=1, SOURCE=3, REASON=2; local restrictions exceeded | error |
040502013 | Association negotiation failed | Temporary rejection by service provider (presentation-related); temporary congestion | warning |
040502014 | Association negotiation failed | Temporary rejection by service provider (presentation-related); local restrictions exceeded | warning |
04050201B | Association was stopped | Service provider aborted association; no reason given | warning |
04050201C | Association was stopped | Service provider aborted; PDU cannot be recognized | error |
04050201D | Association was stopped | Service provider aborted; unexpected PDU received | error |
04050201E | Association was stopped | Service provider aborted; PDU parameters cannot be recognized | error |
04050201F | Association was stopped | Service provider aborted; unexpected PDU parameters | error |
040502020 | Association was stopped | Service provider aborted; invalid PDU parameters value | error |
040502021 | Association negotiation failed | No response from communication partner within timeout period | warning |
040502023 | Association negotiation failed | RELEASE-RQ received at time other than during association release | error |
040502058 | Association negotiation failed | No communication partner; invalid IP address or port | error |
040505001 | DICOM communication error | Status tag (0000,0900) value error; non-existent attribute set | error |
040505002 | DICOM communication error | Status tag (0000,0900) value error; invalid attribute value | error |
040505003 | DICOM communication warning - attribute list error | Status tag (0000,0900) value error with attribute list error | warning |
040505004 | DICOM communication error | Status tag (0000,0900) value error; processing failed | error |
040505005 | DICOM communication error | Status tag (0000,0900) value error; multiple SOP instances | error |
040505006 | DICOM communication error | Status tag (0000,0900) value error; non-existent object instance set | error |
040505007 | DICOM communication error | Status tag (0000,0900) value error; non-existent event type set | error |
040505008 | DICOM communication error | Status tag (0000,0900) value error; non-existent argument set | error |
040505009 | DICOM communication error | Status tag (0000,0900) value error; invalid argument set | error |
04050500A | DICOM communication warning - attribute value out of range | Status tag (0000,0900) value error with out-of-range attribute value | warning |
04050500B | DICOM communication error | Status tag (0000,0900) value error; invalid object instance set | error |
04050500C | DICOM communication error | Status tag (0000,0900) value error; non-existent SOP class set | error |
04050500D | DICOM communication error | Status tag (0000,0900) value error; contradiction in class instance | error |
04050500E | DICOM communication error | Status tag (0000,0900) value error; missing attribute | error |
04050500F | DICOM communication error | Status tag (0000,0900) value error; missing attribute value set | error |
040505010 | DICOM communication error | Status tag (0000,0900) value error; non-supported SOP class set | error |
040505011 | DICOM communication error | Status tag (0000,0900) value error; non-existent action type set | error |
040505012 | DICOM communication error | Status tag (0000,0900) value error; repeated startup | error |
040505013 | DICOM communication error - status tag value error | Status tag (0000,0900) value error received with unrecognized operation | error |
040505014 | DICOM communication error - invalid argument | Status tag (0000,0900) value error received with invalid argument | error |
040505015 | DICOM communication error - resource restriction | Status tag (0000,0900) value error with resource restriction | error |
040505017 | DICOM communication error - incomprehensible setting | Status tag (0000,0900) value error with incomprehensible setting | error |
040505018 | DICOM communication error - insufficient resources | Status tag (0000,0900) value error with insufficient resources | error |
04050501B | DICOM communication error - C-FIND identifier mismatch | C-FIND status tag (0000,0900) value error, identifier does not match SOP class | error |
04050501C | DICOM communication warning - data factor coercion | C-STORE status tag (0000,0900) value error with data factor coercion | warning |
04050501D | DICOM communication warning - factor abandoned | C-STORE status tag (0000,0900) value error with factor abandoned | warning |
04050501E | DICOM communication warning - dataset mismatch | C-STORE status tag (0000,0900) value error, dataset does not match SOP class | warning |
04050501F | DICOM communication error - C-STORE dataset mismatch | C-STORE status tag (0000,0900) value error, dataset does not match SOP class | error |
040505020 | DICOM communication error - C-STORE no resources | C-STORE status tag (0000,0900) value error with no resources | error |
040505027 | DICOM printer output error - film session structure invalid | N-ACTION status tag (0000,0900) value error, BasicFilmSession SOP class, missing film box | error |
040505028 | DICOM printer output error - print queue full | N-ACTION status tag (0000,0900) value error, BasicFilmSession, print queue full | error |
040505029 | DICOM printer output error - image size exceeds box size | N-ACTION status tag (0000,0900) value error, BasicFilmSession, image too large | error |
040505029 | Error occurred in DICOM printer output | Status tag (0000,0900) value of N-ACTION when SOP class was BasicFilmSession received; image size larger than image box size | error |
04050502A | DICOM printer output error - linked image size exceeds box | N-ACTION status tag (0000,0900) value error, BasicFilmSession, linked image too large | error |
04050502A | Error in DICOM printer output | Status tag (0000,0900) value of N-ACTION when SOP class was BasicFilmSession; linked print image size larger than image box size | error |
04050502C | DICOM printer output warning - print queue full | N-ACTION status tag (0000,0900) value error, BasicFilmSession, print queue full | error |
04050502E | Error in DICOM printer output | Status tag (0000,0900) value of N-CREATE when SOP class was BasicFilmBox; existing unprinted film box exists; N-ACTION at film session level unsupported | error |
04050502E | DICOM printer output error - unprinted film box exists | N-CREATE status tag (0000,0900) value error, BasicFilmSession, unprinted film box exists | error |
04050502F | DICOM printer output error - printer memory insufficient | N-SET status tag (0000,0900) value error, GrayScaleImageBox, insufficient printer memory | error |
040505030 | DICOM printer status warning received | Printer status tag (2110,0010) value of N-GET during Printer operation is WARNING | warning |
040505031 | DICOM communication response error | Status tag (0000,0900) value not defined by DICOM received from communication party | error |
040506001 | DICOM printer failure detected | FAILURE status tag (2110,0010) value of N-GET from DICOM printer | error |
040507001 | DICOM StorageCommitment failed - reason 0x0110 | N-EVENT-REPORT failure of StorageCommitment with reason 0x0110 | error |
040507002 | DICOM StorageCommitment failed - reason 0x0112 | N-EVENT-REPORT failure of StorageCommitment with reason 0x0112 | error |
040507003 | DICOM StorageCommitment failed - reason 0x0119 | N-EVENT-REPORT failure of StorageCommitment with reason 0x0119 | error |
040507004 | DICOM StorageCommitment failed - reason 0x0122 | N-EVENT-REPORT failure of StorageCommitment with reason 0x0122 | error |
040507005 | DICOM StorageCommitment failed - reason 0x0131 | N-EVENT-REPORT failure of StorageCommitment with reason 0x0131 | error |
040507006 | DICOM StorageCommitment failed - undefined reason | N-EVENT-REPORT failure of StorageCommitment with undefined reason code | error |
040507007 | DICOM StorageCommitment failed - undefined reason | N-EVENT-REPORT failure of StorageCommitment with undefined reason code | error |
040508001 | DICOM disc creation failed | Unexpected error during DICOM disc creation process | error |
040508002 | DICOM disc creation failed; disc recorder not found | Specified disc recorder not detected during DICOM disc creation attempt | error |
040600008 | Failed to access target disk | Disk drive inaccessible during free space check at media storage time (drive open, damaged, etc.) | error |
040600009 | Process interrupted; insufficient free disk space | Free space check at media storage shows inadequate disk space available | error |
04060000B | Retry not possible; error occurred during process | Power discontinuity or software bug; key ID missing or image processing failed | error |
04060000C | Cannot execute old study information transfer | System attempting to transfer study created before data rebinding; same image file in use | error |
04060000D | Operation failed; write destination busy | Same media drive used for both media storage and DICOM disc creation | error |
04060000E | Retry not possible; DICOM disc creation error occurred | DICOM disc creation failed; error displayed in process viewer | error |
040700006 | Failed to load image file | System attempts to read file that does not exist | error |
040700012 | Graphics card does not meet required specs; imaging may fail | Installed graphics card performance below requirements | critical |
040700100 | System memory settings invalid | Memory size larger than installed memory specified by Service Tool | critical |
040800100 | Examination cannot be started; non-existent protocol included | Protocol contained in pending examination has been deleted from database | error |
040D0001F | Providing free space on hard disk failed | Data equivalent to prescribed number of records cannot be deleted | error |
040D00201 | Network interface in monitoring settings not found; monitoring function disabled | Network interface malfunction or improper connection during monitoring setup | error |
040E01000 | Image analysis processing failed | Image analysis processing failure during XdCom processing | error |
040E01001 | Failed to load extra defect correction file | System unable to load extra defect correction file | error |
041000004 | Invalid studies are acquired | Exam information containing illegal values for key items (ACC#, patient ID, patient name) acquired from Studix | error |
041000005 | Entered information has invalid value | Exam information containing illegal values for non-key items acquired from Studix | error |
041000006 | Specified ID not found in examination file | Exam information with specified key values did not exist in examination file | error |
041000008 | Error occurred during script-based input calibration | Script compilation error, script processing error, or abnormal script return value | error |
041000011 | This version of workflow library not supported | Version mismatch between CCS and automatic examination pack | error |
041000012 | Error occurred in CCRHIS library | Studix, Workflow, or magnetic card processing failed | error |
041000020 | Examination file loading halted - maximum reserved exams exceeded | Maximum number of reserved exams that can be registered to database exceeded | error |
041000021 | Invalid studies loaded and cannot be displayed | Invalid value detected among values of required items | error |
041000022 | Entered information has invalid value in exam file | Invalid value detected among user-specified values | error |
041000023 | Failed to read examination file | Exam file was not found | error |
041000024 | Error occurred during automatic examination file updating | Error during exam file monitoring when configuring automatic update | error |
041000025 | Failed to read examination file | Error occurred during loading of exam file | error |
041100004 | CxdiInfo startup failed | Specified DLL does not exist or specified DLL lacks expected interface | error |
050501002 | Flash ROM error detected in detector | Flash ROM ID error detected by detector AD board | warning |
050501003 | Flash ROM error detected in detector | Flash ROM Write error detected by detector AD board | warning |
050501004 | Flash ROM error detected in detector | Flash ROM Erase error detected by detector AD board | warning |
050501005 | There may be problem in images because of grid error | Grid warning detected by detector AD board during imaging | warning |
050501005 | Grid error may have caused image abnormality | Grid warning detected but imaging continued | warning |
050501006 | Detector internal temperature exceeds upper limit | Temperature inside detector exceeds upper limit | error |
050501007 | Detector internal temperature approaching upper limit | Internal detector temperature approaching upper limit | warning |
050501008 | Drive circuit error occurred | PLD error detected by detector AD board | error |
050501009 | Drive circuit error - PLD config error | PLD Config error detected by detector AD board | error |
05050100A | Abnormality detected in detector structure | Illegal value written for serial number | error |
05050100A | Detector structure abnormality - invalid serial number | Illegal value written for serial number | error |
05050100B | Detector structure abnormality - serial number mismatch | Discrepancy between detector serial number and DP number | error |
05050100B | Abnormality detected in detector structure | Discrepancy between serial number of detector and DP number | error |
05050100C | Control line connection error detected in detector | Control wire error detected by detector AD board, broken cable between detector and power box | error |
05050100D | Detector control line connection error - cable length mismatch | Discrepancy between detector-power box cable length settings and XRAY board cable | error |
05050100E | Undefined error detected by detector | Error detected in Hex format received when firmware or PLD was downloaded | warning |
05050100F | Undefined error detected by detector | Error detected in Hex format received when firmware, PLD or DP file was downloaded | warning |
050501010 | Failed to transfer image data | DMA timeout inside PCB-50Di when reduced-size images were read | error |
050501011 | Transfer of image data was stopped | Transfer of Full image data was aborted | warning |
050501012 | Abnormality detected in analog power supply for detector | Abnormality detected in analog power supply for detector | error |
050501013 | Grid ID loading error detected | Grid ID loading error was detected | warning |
050501014 | Grid error detected | Grid error detected at detector AD board | error |
050501015 | Grid error may have caused image abnormality | Grid warning detected by detector AD board but imaging continued | warning |
050501016 | Firmware version mismatch | Wrong detector firmware files written during firmware update | error |
050502009 | Detector communication error; cxdcap.ini missing | cxdcap.ini file not installed | critical |
050502026 | Detector communication error via Ethernet | Communication error detected by Ethernet interface | error |
E010156203 | SSID field contains non-alphanumeric character | Pressing Save button after inputting special character (e.g. '(') in SSID field | error |
E040501010 | Database capacity exceeded SQL Server upper limit | Diskstore save location set to large capacity media with 1000+ patient examinations performed continuously without system shutdown | error |
ERROR_CHECK_LATER | Check later error allowing continued imaging | Non-critical error such as DICOM communication fault occurs | warning |
ERROR_IMMEDIATE | Immediate display error requiring user action | Imaging failure such as exposure attempted but images not transferred or detector error occurs | error |
F040700012 | GPU does not meet required specifications or driver version insufficient | CCS startup when GPU processing is enabled; GPU Compute capability below 2.0 or NVIDIA driver version too old | critical |
FATAL | Fatal error condition requiring service call | System experiences unrecoverable error such as detector no response or DB connection failure | critical |
I0111 | SQL Server 2005 setup alert during system software upgrade | Upgrading system software from Ver.1.30 or earlier to Ver.1.40 | warning |
No second monitor is connected | Second monitor detection failure during CCS startup | When Use second monitor checkbox is enabled in service tool but system cannot detect connected secondary monitor | error |
Parameter initialization required | Parameter file reset to default values | audit_grid_NE_RF_P.txt deleted after parameter configuration | warning |
WARNING | Warning notification not requiring immediate action | Non-critical conditions such as low film, backup needed, detector temperature slightly elevated | warning |
Wrong image sizes error | Parameter configuration error in audit_grid_NE_RF_P.txt | CRW image files analyzed with incorrect image size parameter | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI