Canon CXDI Control NE: troubleshooting and error codes
The ServiceForge knowledge base holds 3 documented symptoms for the Canon CXDI Control NE, with 8 ranked causes, 13 diagnostic checks with test points and expected values, and 135 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Monitor grayscale not properly adjusted
Small squares in grayscale pattern test are not all visible at similar contrast; some squares barely visible or invisible due to color similarity with background.
Ranked causes
- 1. Monitor gamma value not set correctly to 2.2 (common)
- 2. Luminance or contrast ratio not properly configured (occasional)
- 3. Matte LCD display with anti-glare indentations causing light diffusion (occasional)
- 4. High reflected ambient light reducing contrast ratio (occasional)
Verify all eight small squares in P-value pattern are visible
- Test point
- Monitor display P-value pattern with 8 patches
- Expected
- All eight small squares y1 to y8 visible at similar contrast and clarity patches, All visible; difference in P-values 1.5% normal mode, 1.2% t
- Tools
- Visual inspection
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gamma adjustment for monitor DICOM conformance
Setting gamma value to achieve DICOM GSDF display characteristics on non-DICOM monitors using luminance measurements and visual adjustment
Ranked causes
- 1. Non-DICOM monitor installed; need to simulate DICOM grayscale curve (common)
- 2. Type A: Monitor gamma measurement with luminance meter (occasional)
- 3. Type B: Monitor gamma visual adjustment without luminance meter (occasional)
Launch Gamma Adjustment Tool
- Test point
- Service Tool menu
- Expected
- Gamma adjustment type selection dialog appears with three options: Off, With_Luminance_Meter, Without_Luminance_Meter
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor grayscale using Gamma Adjustment Tool (Type B: Without Luminance Meter)
Procedure to calibrate monitor grayscale to resemble DICOM standard using service tool without luminance meter, enabling general-purpose monitors to be used for diagnostic imaging.
Ranked causes
- 1. Initial gamma adjustment required for diagnostic imaging setup (common)
Start Gamma Adjustment Tool and select configuration mode
- Test point
- Service Tool menu, Tool tab
- Expected
- Gamma Adjustment Tool opens; select Without Luminance Meter option selection, Exact selection
- Tools
- Service Tool software, monitor
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
020201003 | Detector remaining battery charge is low | Active detector on external power supply battery drops below 4% full charge | error |
020201004 | Unexpected error occurred. Gamma adjustment is not available | All loading of pattern files for simplified gamma adjustments failed | error |
020201005 | Failed to connect to barcode reader | Exception during barcode reader connection initialization process | error |
020201006 | Cannot start specified examination | Examination specified in status where ccrhis.dvr file specification is ignored or patient specified during magnetic card ignore status | error |
020201007 | Detector communication error occurred | Reception timeout from detector or power-off during exposure | warning |
020201008 | Detector temperature exceeded allowable operating range | Detector temperature reached dangerously high level triggering Sleep status | error |
020201009 | Failed to start up OverwrapSoftware | Startup failed due to invalid path set in Service Tool | error |
020201010 | Detector remaining battery charge is low | Inactive detector on external power supply battery drops below 4% full charge | error |
020201011 | Imaging cannot be performed because no battery is mounted in detector | Wired connection made to wired/wireless cable type detector without battery | error |
020201012 | Imaging cannot be performed because battery charge is low | Active detector on battery power drops below 4% full charge | error |
020201013 | Imaging cannot be performed because battery charge is low | Inactive detector on battery power drops below 4% full charge | error |
020202001 | Detector temperature may exceed allowable operating range | Detector temperature reached warning level threshold | warning |
020202002 | Detector temperature exceeded allowable operating range | Detector temperature reached dangerous level | warning |
020202003 | Detector is not available. Irradiation has finished | Detector becomes invalid during irradiation | warning |
020202004 | Detector battery is weakening | Detector battery charge drops below 8% of full capacity | warning |
020202005 | Specified disk space could not be cleaned up | Specified space greater than examination to be deleted | warning |
020202006 | Could not acquire parameters required for reconfiguration | Reconfiguration parameters insufficient for Tomosynthesis protocol | warning |
030100001 | No calibration data | Workspace calibration not completed | error |
030100005 | Required COM+ servers are not installed | One or more necessary COM+ servers not installed | critical |
030201002 | Periodic noise detected in acquisition data | Calibration images taken without removing grid on grid-embedded detector | error |
030201003 | Cooling unit has been attached or detached | Cooling unit attached or detached during study | error |
030201005 | Generator is disconnected | A generator disconnection event has occurred | error |
030201006 | Generator is disconnected | Generator communication has failed when calibration, performance test or self-diagnosis is started | error |
030201007 | Preparing for imaging failed | The setting of conditions in the generator or detector failed | error |
030201008 | Detector battery has lost its charge | The remaining charge of the detector's battery has dropped and is now deficient | error |
030201009 | Detector temperature has exceeded the allowable operating range | Received notification from the detector of an abnormally high temperature | error |
030201010 | No battery is attached to the detector | The battery has not yet been installed with a wired connection | error |
03020200E | Memory allocation failed | When the QC Tool was started or examination was started, the attempt to secure the memory area failed due to insufficient memory | critical |
03020201B | The uploading of the calibration data has failed | An error occurred while the SPCC data was being uploaded | critical |
03020201C | Image data transfer has failed | The transfer of images caused an error for some reason | critical |
03020201F | A detector error occurred | A detector error was indicated | critical |
030202021 | The control software/controller is not compatible with the current detector firmware | There is no compatibility with the firmware | critical |
030202022 | The Jumbo packet setting is incorrect or a network card not compatible with Jumbo packet is used | Either a setting for more than 4500 Jumbo packets has not been established or a network card which is compatible with Jumbo packets is not being used | critical |
030203003 | The detector temperature has exceeded the allowable operating range | Temperature trouble was notified by the detector | warning |
03020FFFF | An internal error occurred | This error happens when both X-ray exposure and exposure cancel occur at the same time, or due to a software bug | critical |
040100008 | An invalid detector has been specified | A detector which has not been registered was specified | error |
040100023 | Generator communication failed | The initialization processing of the connection to the generator failed. Illegal IP setting of generator, etc. | error |
040100045 | Calibration data is invalid | Calibration data is not correct | warning |
040100052 | The detector issued a warning | This error occurs when warning from detector occurred | warning |
040100053 | The detector issued an error | This error occurs when error from detector occurred | error |
040100054 | The detector issued a fatal error | This error occurs when fatal error from detector occurred | critical |
040100056 | Generator is disconnected | This error occurs if the retry button is canceled when the setting of the imaging conditions in the X-ray generator failed | error |
040100066 | The IP addresses for the wireless and wired configurations have been reversed | The wired and wireless IP address settings of the Service Tool are reversed | critical |
040101014 | The detector has entered the standby state | The detector transitioned to auto sleep mode after a specified period of time elapsed while in the Ready state | warning |
040101021 | Connection to X-ray interface failed | X-ray interface connection error | error |
040101029 | A battery charge-related error has occurred on the detector | A battery charge-related error notification was issued by the detector | warning |
040300048 | The control software/controller is not compatible with the current detector firmware | There is no compatibility with the firmware | critical |
040300050 | The Jumbo packet setting is incorrect or a network card not compatible with Jumbo packet is used | Either a setting for more than 4500 Jumbo packets has not been established or a network card which is compatible with Jumbo packets is not being used | critical |
040300051 | The IP addresses for the wireless and wired configurations have been reversed | The wired and wireless IP address settings of the Service Tool are reversed | critical |
040300053 | The setting for imaging with X-ray auto-detection has been enabled, but activation is not possible | The X-ray auto-detection mode setting was enabled with the Service Tool, but a workspace with a detector that does not support X-ray auto-detection exists | critical |
040300054 | The firmware of the detector being used is not compatible | The firmware is not compatible | critical |
040400003 | Failed to acquire the system settings | The DB data is defective or deficient. The table records are insufficient. | critical |
040500003 | Exceeded the maximum number of study order acquisitions | C-FIND-RSP study exceeding the maximum number of study order acquisitions set using the Service Tool received | error |
040500004 | Received invalid studies violating DICOM rules | C-FIND-RSP contains data contravening DICOM rules | error |
040500005 | Received tag contains invalid value replaced by null | C-FIND-RSP contains data contravening DICOM rules | error |
040500006 | Characters not in received character set present | Characters which cannot be decoded properly received from RIS or auxiliary kanjis received | error |
040500007 | Characters do not match current character code | Japanese-language data received despite English (ASCII) set in Service Tool Character Set | warning |
040500008 | Date entry before 1753 included in acquired Exam | Date type (DA) data prior to January 1, 1753 received from RIS | error |
040500009 | Failed to copy a tag to C-STORE | DICOM rules violation in data encountered when copying tags from C-FIND to C-STORE | warning |
04050000A | Failed to copy a tag to DicomFile | DICOM rules violation in data when copying tags from C-FIND to DicomFile | warning |
04050000C | Failed to load DICOMDIR | Loading of DICOMDIR failed due to damaged file or other reason | error |
04050000D | An output error occurred | Output information no longer available after system restart due to power discontinuity | error |
04050000F | Failed to read DICOM file | Loading of DICOM file failed; file does not exist, format non-conformant, access denied, or media corrupted | error |
040500010 | Failed to read DICOM file (image creation) | Failed to create image file due to insufficient memory, access rights, or invalid tag values | error |
040501011 | Cannot continue processing; file already exists at storage destination | File with same name already exists at save destination | error |
040501012 | Cannot continue processing; access to storage destination denied | Save destination folder cannot be accessed | error |
040501013 | Cannot continue processing; insufficient space at storage destination | Not enough storage space at specified destination | error |
040501014 | DICOM GSPS output halted due to incomplete object image transfer | C-STORE processing of GSPS executed while C-STORE of target images (IHEEngine) failed | error |
040501016 | Network communication cut off during file output | Problem occurred in network during media storage over network, making writing impossible | error |
040501017 | An unsupported DICOMDIR was loaded | DICOMDIR file other than one created by CXDI NE/RF was loaded | error |
040505029 | Error in DICOM printer output; image size exceeds image box size | Image size larger than the image box size during BasicFilmSession N-ACTION | error |
04050502A | Error in DICOM printer output; linked print image size exceeds image box size | Image size larger than image box size during BasicFilmSession N-ACTION | error |
04050502C | DICOM printer output failed; print queue full | Print job SOP instances cannot be created; print queue full | error |
04050502E | DICOM printer output error; existing film box not printed or N-ACTION unsupported | Existing film box not yet printed; N-ACTION at film session not supported | error |
04050502F | DICOM printer output error; insufficient printer memory | Printer memory insufficient to save images during GrayScaleImageBox N-SET | error |
040505030 | DICOM printer warning status received despite successful output | Printer status tag (2110,0010) value of N-GET is WARNING | warning |
040505031 | DICOM communication error with undefined status tag value | Status tag (0000,0900) value not defined by DICOM standard received | error |
040506001 | Output halted; DICOM printer reported failure status | Printer status tag (2110,0010) value of N-GET is FAILURE | error |
040507001 | DICOM StorageCommitment failed; reason 0x0110 | N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0110 | error |
040507002 | DICOM StorageCommitment failed; reason 0x0112 | N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0112 | error |
040507003 | DICOM StorageCommitment failed; reason 0x0119 | N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0119 | error |
040507004 | DICOM StorageCommitment failed; reason 0x0122; save service not supported | N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0122 | error |
040507005 | DICOM StorageCommitment failed; reason 0x0131 | N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0131 | error |
040507006 | DICOM StorageCommitment failed; reason 0x0213; output destination capacity insufficient | N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0213 | error |
040507007 | DICOM StorageCommitment failed; non-DICOM reason code received | N-EVENT-REPORT of StorageCommitment received undefined reason (0008,1197) | error |
040508001 | DICOM disc creation failed | Unexpected error during DICOM disc creation process | error |
040508002 | DICOM disc creation failed; specified disc recorder not found | Specified disc recorder unavailable when attempting DICOM disc creation | error |
040600008 | Job Manager: Failed to access target disk | Disk could not be accessed during free space check or media storage | error |
040600009 | Job Manager: Insufficient free disk space; process interrupted | Not enough free space available at time of media storage | error |
04060000B | Job Manager: Retry not possible; error in process | Power discontinuity or software bug; key ID missing or image processing failed | error |
04060000C | Job Manager: Cannot transfer old study; rebinding data conflict | System attempting to transfer study created before data rebinding | error |
04060000D | Job Manager: Operation failed; write destination busy | Same media drive used for both media storage and DICOM disc creation | error |
04060000E | Job Manager: DICOM disc creation error; retry not possible | DICOM disc creation failed for unspecified reason | error |
040700006 | Memory Manager: Failed to load image file | Image file to be read does not exist | error |
040700012 | Memory Manager: Graphics card does not meet required specifications | Installed graphics card performance below requirements | critical |
040700100 | Memory Manager: Invalid system memory settings | Memory size larger than installed memory specified by Service Tool | critical |
040800100 | Protocol Manager: Cannot start examination; non-existent protocol included | Protocol contained in pending examination deleted from database | error |
040D0001F | Utility Manager: Failed to provide free disk space; data deletion unsuccessful | Cannot delete data equivalent to prescribed number of records | error |
040D00201 | Utility Manager: Network interface not found; monitoring disabled | Network interface specified in monitoring settings not found or disconnected | error |
040E01000 | XdCom Manager: Image analysis processing failed | Image analysis processing failed during automated image adjustment | error |
040E01001 | XdCom Manager: Failed to load extra defect correction file | Extra defect correction file required for images does not exist | error |
041000004 | Invalid studies acquired with illegal key item values (ACC#, patient ID, patient name) | Exam information containing illegal values for key items acquired from Studix | error |
041000005 | Entered information has invalid value; invalid values replaced with blanks | Exam information with illegal values in non-key items acquired from Studix | error |
041000006 | Specified ID not found in examination file | Exam information with specified key values did not exist in examination file | error |
041000008 | Error occurred during script-based input calibration | Script compilation error, script processing error, script standard error output, or non-zero return value | error |
041000011 | Workflow library version not supported | Version mismatch between CCS and automatic examination pack | error |
041000012 | Error occurred in CCRHIS library | Studix, Workflow, or magnetic card processing failed due to invalid startup setting or I/O error | error |
041000020 | Examination file loading halted; maximum reserved exams exceeded | Maximum number of reserved exams registered to database exceeded | error |
041000021 | Invalid studies loaded; cannot be displayed; check examination file | Invalid value detected in required items during file load | error |
041000022 | Entered information invalid; values replaced with blanks; check examination file | Invalid value detected in user-specified values during file load | error |
041000023 | Failed to read examination file; check setting information | Exam file not found at specified location | error |
041000024 | Error occurred during automatic examination file updating | Error during exam file monitoring when configuring automatic update | error |
041000025 | Failed to read examination file; check examination file | Error occurred during exam file loading | error |
041100004 | CxdiInfo startup failed; check settings involved | Specified DLL does not exist or DLL lacks expected interface | error |
050501002 | Flash ROM error detected in detector | Flash ROM warning detected by detector AD board with flash ROM ID error | warning |
050501003 | Flash ROM error detected in detector | Flash ROM warning detected by detector AD board with flash ROM write error | warning |
050501004 | Flash ROM error detected in detector | Flash ROM warning detected by detector AD board with flash ROM erase error | warning |
050501005 | Image quality may be affected due to grid error | Grid warning detected by detector AD board but imaging continued | warning |
050501008 | Drive circuit error occurred | PLD error detected by detector AD board | error |
050501009 | Drive circuit error occurred | PLD config error detected by detector AD board | error |
05050100A | Abnormality detected in detector structure | Illegal value written for detector serial number | error |
05050100B | Abnormality detected in detector structure | Discrepancy between detector serial number and DP number | error |
05050100C | Control line connection error detected in detector | Control wire error detected by detector AD board; broken cable likely | error |
05050100D | Control line connection error detected in detector | Control wire error with discrepancy between cable length settings | error |
05050100E | Error detected by detector | Undefined error with error detected in hex format during firmware or PLD download | warning |
05050100F | Error detected by detector | Undefined error with hex format error during firmware, PLD or DP file download | warning |
050501010 | Failed to transfer image data | DMA inside PCB-50Di timed out when reading reduced-size images | error |
050501011 | Transfer of image data stopped | Full image data transfer aborted | warning |
050501012 | Abnormality detected in analog power supply for detector | Abnormality detected in analog power supply for detector | error |
050501013 | Grid ID loading error detected | Grid ID failed to load | warning |
050501014 | Grid error detected | Grid error detected at detector AD board | error |
050501015 | Grid error may have caused image abnormality | Grid warning detected by detector AD board but imaging continued | warning |
050501016 | Firmware version mismatch | Wrong detector firmware written during firmware update | error |
050502009 | Detector communication error; restart system | cxdcap.ini file not installed | critical |
050502026 | Detector communication error; cable may be disconnected or power discontinuity | Communication error detected by Ethernet interface | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI