Canon CXDI Control NE: troubleshooting and error codes

The ServiceForge knowledge base holds 3 documented symptoms for the Canon CXDI Control NE, with 8 ranked causes, 13 diagnostic checks with test points and expected values, and 135 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Monitor grayscale not properly adjusted

Small squares in grayscale pattern test are not all visible at similar contrast; some squares barely visible or invisible due to color similarity with background.

Ranked causes

  1. 1. Monitor gamma value not set correctly to 2.2 (common)
  2. 2. Luminance or contrast ratio not properly configured (occasional)
  3. 3. Matte LCD display with anti-glare indentations causing light diffusion (occasional)
  4. 4. High reflected ambient light reducing contrast ratio (occasional)
First check

Verify all eight small squares in P-value pattern are visible

Test point
Monitor display P-value pattern with 8 patches
Expected
All eight small squares y1 to y8 visible at similar contrast and clarity patches, All visible; difference in P-values 1.5% normal mode, 1.2% t
Tools
Visual inspection

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gamma adjustment for monitor DICOM conformance

Setting gamma value to achieve DICOM GSDF display characteristics on non-DICOM monitors using luminance measurements and visual adjustment

Ranked causes

  1. 1. Non-DICOM monitor installed; need to simulate DICOM grayscale curve (common)
  2. 2. Type A: Monitor gamma measurement with luminance meter (occasional)
  3. 3. Type B: Monitor gamma visual adjustment without luminance meter (occasional)
First check

Launch Gamma Adjustment Tool

Test point
Service Tool menu
Expected
Gamma adjustment type selection dialog appears with three options: Off, With_Luminance_Meter, Without_Luminance_Meter

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monitor grayscale using Gamma Adjustment Tool (Type B: Without Luminance Meter)

Procedure to calibrate monitor grayscale to resemble DICOM standard using service tool without luminance meter, enabling general-purpose monitors to be used for diagnostic imaging.

Ranked causes

  1. 1. Initial gamma adjustment required for diagnostic imaging setup (common)
First check

Start Gamma Adjustment Tool and select configuration mode

Test point
Service Tool menu, Tool tab
Expected
Gamma Adjustment Tool opens; select Without Luminance Meter option selection, Exact selection
Tools
Service Tool software, monitor

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
020201003Detector remaining battery charge is lowActive detector on external power supply battery drops below 4% full chargeerror
020201004Unexpected error occurred. Gamma adjustment is not availableAll loading of pattern files for simplified gamma adjustments failederror
020201005Failed to connect to barcode readerException during barcode reader connection initialization processerror
020201006Cannot start specified examinationExamination specified in status where ccrhis.dvr file specification is ignored or patient specified during magnetic card ignore statuserror
020201007Detector communication error occurredReception timeout from detector or power-off during exposurewarning
020201008Detector temperature exceeded allowable operating rangeDetector temperature reached dangerously high level triggering Sleep statuserror
020201009Failed to start up OverwrapSoftwareStartup failed due to invalid path set in Service Toolerror
020201010Detector remaining battery charge is lowInactive detector on external power supply battery drops below 4% full chargeerror
020201011Imaging cannot be performed because no battery is mounted in detectorWired connection made to wired/wireless cable type detector without batteryerror
020201012Imaging cannot be performed because battery charge is lowActive detector on battery power drops below 4% full chargeerror
020201013Imaging cannot be performed because battery charge is lowInactive detector on battery power drops below 4% full chargeerror
020202001Detector temperature may exceed allowable operating rangeDetector temperature reached warning level thresholdwarning
020202002Detector temperature exceeded allowable operating rangeDetector temperature reached dangerous levelwarning
020202003Detector is not available. Irradiation has finishedDetector becomes invalid during irradiationwarning
020202004Detector battery is weakeningDetector battery charge drops below 8% of full capacitywarning
020202005Specified disk space could not be cleaned upSpecified space greater than examination to be deletedwarning
020202006Could not acquire parameters required for reconfigurationReconfiguration parameters insufficient for Tomosynthesis protocolwarning
030100001No calibration dataWorkspace calibration not completederror
030100005Required COM+ servers are not installedOne or more necessary COM+ servers not installedcritical
030201002Periodic noise detected in acquisition dataCalibration images taken without removing grid on grid-embedded detectorerror
030201003Cooling unit has been attached or detachedCooling unit attached or detached during studyerror
030201005Generator is disconnectedA generator disconnection event has occurrederror
030201006Generator is disconnectedGenerator communication has failed when calibration, performance test or self-diagnosis is startederror
030201007Preparing for imaging failedThe setting of conditions in the generator or detector failederror
030201008Detector battery has lost its chargeThe remaining charge of the detector's battery has dropped and is now deficienterror
030201009Detector temperature has exceeded the allowable operating rangeReceived notification from the detector of an abnormally high temperatureerror
030201010No battery is attached to the detectorThe battery has not yet been installed with a wired connectionerror
03020200EMemory allocation failedWhen the QC Tool was started or examination was started, the attempt to secure the memory area failed due to insufficient memorycritical
03020201BThe uploading of the calibration data has failedAn error occurred while the SPCC data was being uploadedcritical
03020201CImage data transfer has failedThe transfer of images caused an error for some reasoncritical
03020201FA detector error occurredA detector error was indicatedcritical
030202021The control software/controller is not compatible with the current detector firmwareThere is no compatibility with the firmwarecritical
030202022The Jumbo packet setting is incorrect or a network card not compatible with Jumbo packet is usedEither a setting for more than 4500 Jumbo packets has not been established or a network card which is compatible with Jumbo packets is not being usedcritical
030203003The detector temperature has exceeded the allowable operating rangeTemperature trouble was notified by the detectorwarning
03020FFFFAn internal error occurredThis error happens when both X-ray exposure and exposure cancel occur at the same time, or due to a software bugcritical
040100008An invalid detector has been specifiedA detector which has not been registered was specifiederror
040100023Generator communication failedThe initialization processing of the connection to the generator failed. Illegal IP setting of generator, etc.error
040100045Calibration data is invalidCalibration data is not correctwarning
040100052The detector issued a warningThis error occurs when warning from detector occurredwarning
040100053The detector issued an errorThis error occurs when error from detector occurrederror
040100054The detector issued a fatal errorThis error occurs when fatal error from detector occurredcritical
040100056Generator is disconnectedThis error occurs if the retry button is canceled when the setting of the imaging conditions in the X-ray generator failederror
040100066The IP addresses for the wireless and wired configurations have been reversedThe wired and wireless IP address settings of the Service Tool are reversedcritical
040101014The detector has entered the standby stateThe detector transitioned to auto sleep mode after a specified period of time elapsed while in the Ready statewarning
040101021Connection to X-ray interface failedX-ray interface connection errorerror
040101029A battery charge-related error has occurred on the detectorA battery charge-related error notification was issued by the detectorwarning
040300048The control software/controller is not compatible with the current detector firmwareThere is no compatibility with the firmwarecritical
040300050The Jumbo packet setting is incorrect or a network card not compatible with Jumbo packet is usedEither a setting for more than 4500 Jumbo packets has not been established or a network card which is compatible with Jumbo packets is not being usedcritical
040300051The IP addresses for the wireless and wired configurations have been reversedThe wired and wireless IP address settings of the Service Tool are reversedcritical
040300053The setting for imaging with X-ray auto-detection has been enabled, but activation is not possibleThe X-ray auto-detection mode setting was enabled with the Service Tool, but a workspace with a detector that does not support X-ray auto-detection existscritical
040300054The firmware of the detector being used is not compatibleThe firmware is not compatiblecritical
040400003Failed to acquire the system settingsThe DB data is defective or deficient. The table records are insufficient.critical
040500003Exceeded the maximum number of study order acquisitionsC-FIND-RSP study exceeding the maximum number of study order acquisitions set using the Service Tool receivederror
040500004Received invalid studies violating DICOM rulesC-FIND-RSP contains data contravening DICOM ruleserror
040500005Received tag contains invalid value replaced by nullC-FIND-RSP contains data contravening DICOM ruleserror
040500006Characters not in received character set presentCharacters which cannot be decoded properly received from RIS or auxiliary kanjis receivederror
040500007Characters do not match current character codeJapanese-language data received despite English (ASCII) set in Service Tool Character Setwarning
040500008Date entry before 1753 included in acquired ExamDate type (DA) data prior to January 1, 1753 received from RISerror
040500009Failed to copy a tag to C-STOREDICOM rules violation in data encountered when copying tags from C-FIND to C-STOREwarning
04050000AFailed to copy a tag to DicomFileDICOM rules violation in data when copying tags from C-FIND to DicomFilewarning
04050000CFailed to load DICOMDIRLoading of DICOMDIR failed due to damaged file or other reasonerror
04050000DAn output error occurredOutput information no longer available after system restart due to power discontinuityerror
04050000FFailed to read DICOM fileLoading of DICOM file failed; file does not exist, format non-conformant, access denied, or media corruptederror
040500010Failed to read DICOM file (image creation)Failed to create image file due to insufficient memory, access rights, or invalid tag valueserror
040501011Cannot continue processing; file already exists at storage destinationFile with same name already exists at save destinationerror
040501012Cannot continue processing; access to storage destination deniedSave destination folder cannot be accessederror
040501013Cannot continue processing; insufficient space at storage destinationNot enough storage space at specified destinationerror
040501014DICOM GSPS output halted due to incomplete object image transferC-STORE processing of GSPS executed while C-STORE of target images (IHEEngine) failederror
040501016Network communication cut off during file outputProblem occurred in network during media storage over network, making writing impossibleerror
040501017An unsupported DICOMDIR was loadedDICOMDIR file other than one created by CXDI NE/RF was loadederror
040505029Error in DICOM printer output; image size exceeds image box sizeImage size larger than the image box size during BasicFilmSession N-ACTIONerror
04050502AError in DICOM printer output; linked print image size exceeds image box sizeImage size larger than image box size during BasicFilmSession N-ACTIONerror
04050502CDICOM printer output failed; print queue fullPrint job SOP instances cannot be created; print queue fullerror
04050502EDICOM printer output error; existing film box not printed or N-ACTION unsupportedExisting film box not yet printed; N-ACTION at film session not supportederror
04050502FDICOM printer output error; insufficient printer memoryPrinter memory insufficient to save images during GrayScaleImageBox N-SETerror
040505030DICOM printer warning status received despite successful outputPrinter status tag (2110,0010) value of N-GET is WARNINGwarning
040505031DICOM communication error with undefined status tag valueStatus tag (0000,0900) value not defined by DICOM standard receivederror
040506001Output halted; DICOM printer reported failure statusPrinter status tag (2110,0010) value of N-GET is FAILUREerror
040507001DICOM StorageCommitment failed; reason 0x0110N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0110error
040507002DICOM StorageCommitment failed; reason 0x0112N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0112error
040507003DICOM StorageCommitment failed; reason 0x0119N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0119error
040507004DICOM StorageCommitment failed; reason 0x0122; save service not supportedN-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0122error
040507005DICOM StorageCommitment failed; reason 0x0131N-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0131error
040507006DICOM StorageCommitment failed; reason 0x0213; output destination capacity insufficientN-EVENT-REPORT of StorageCommitment received reason (0008,1197) 0x0213error
040507007DICOM StorageCommitment failed; non-DICOM reason code receivedN-EVENT-REPORT of StorageCommitment received undefined reason (0008,1197)error
040508001DICOM disc creation failedUnexpected error during DICOM disc creation processerror
040508002DICOM disc creation failed; specified disc recorder not foundSpecified disc recorder unavailable when attempting DICOM disc creationerror
040600008Job Manager: Failed to access target diskDisk could not be accessed during free space check or media storageerror
040600009Job Manager: Insufficient free disk space; process interruptedNot enough free space available at time of media storageerror
04060000BJob Manager: Retry not possible; error in processPower discontinuity or software bug; key ID missing or image processing failederror
04060000CJob Manager: Cannot transfer old study; rebinding data conflictSystem attempting to transfer study created before data rebindingerror
04060000DJob Manager: Operation failed; write destination busySame media drive used for both media storage and DICOM disc creationerror
04060000EJob Manager: DICOM disc creation error; retry not possibleDICOM disc creation failed for unspecified reasonerror
040700006Memory Manager: Failed to load image fileImage file to be read does not existerror
040700012Memory Manager: Graphics card does not meet required specificationsInstalled graphics card performance below requirementscritical
040700100Memory Manager: Invalid system memory settingsMemory size larger than installed memory specified by Service Toolcritical
040800100Protocol Manager: Cannot start examination; non-existent protocol includedProtocol contained in pending examination deleted from databaseerror
040D0001FUtility Manager: Failed to provide free disk space; data deletion unsuccessfulCannot delete data equivalent to prescribed number of recordserror
040D00201Utility Manager: Network interface not found; monitoring disabledNetwork interface specified in monitoring settings not found or disconnectederror
040E01000XdCom Manager: Image analysis processing failedImage analysis processing failed during automated image adjustmenterror
040E01001XdCom Manager: Failed to load extra defect correction fileExtra defect correction file required for images does not existerror
041000004Invalid studies acquired with illegal key item values (ACC#, patient ID, patient name)Exam information containing illegal values for key items acquired from Studixerror
041000005Entered information has invalid value; invalid values replaced with blanksExam information with illegal values in non-key items acquired from Studixerror
041000006Specified ID not found in examination fileExam information with specified key values did not exist in examination fileerror
041000008Error occurred during script-based input calibrationScript compilation error, script processing error, script standard error output, or non-zero return valueerror
041000011Workflow library version not supportedVersion mismatch between CCS and automatic examination packerror
041000012Error occurred in CCRHIS libraryStudix, Workflow, or magnetic card processing failed due to invalid startup setting or I/O errorerror
041000020Examination file loading halted; maximum reserved exams exceededMaximum number of reserved exams registered to database exceedederror
041000021Invalid studies loaded; cannot be displayed; check examination fileInvalid value detected in required items during file loaderror
041000022Entered information invalid; values replaced with blanks; check examination fileInvalid value detected in user-specified values during file loaderror
041000023Failed to read examination file; check setting informationExam file not found at specified locationerror
041000024Error occurred during automatic examination file updatingError during exam file monitoring when configuring automatic updateerror
041000025Failed to read examination file; check examination fileError occurred during exam file loadingerror
041100004CxdiInfo startup failed; check settings involvedSpecified DLL does not exist or DLL lacks expected interfaceerror
050501002Flash ROM error detected in detectorFlash ROM warning detected by detector AD board with flash ROM ID errorwarning
050501003Flash ROM error detected in detectorFlash ROM warning detected by detector AD board with flash ROM write errorwarning
050501004Flash ROM error detected in detectorFlash ROM warning detected by detector AD board with flash ROM erase errorwarning
050501005Image quality may be affected due to grid errorGrid warning detected by detector AD board but imaging continuedwarning
050501008Drive circuit error occurredPLD error detected by detector AD boarderror
050501009Drive circuit error occurredPLD config error detected by detector AD boarderror
05050100AAbnormality detected in detector structureIllegal value written for detector serial numbererror
05050100BAbnormality detected in detector structureDiscrepancy between detector serial number and DP numbererror
05050100CControl line connection error detected in detectorControl wire error detected by detector AD board; broken cable likelyerror
05050100DControl line connection error detected in detectorControl wire error with discrepancy between cable length settingserror
05050100EError detected by detectorUndefined error with error detected in hex format during firmware or PLD downloadwarning
05050100FError detected by detectorUndefined error with hex format error during firmware, PLD or DP file downloadwarning
050501010Failed to transfer image dataDMA inside PCB-50Di timed out when reading reduced-size imageserror
050501011Transfer of image data stoppedFull image data transfer abortedwarning
050501012Abnormality detected in analog power supply for detectorAbnormality detected in analog power supply for detectorerror
050501013Grid ID loading error detectedGrid ID failed to loadwarning
050501014Grid error detectedGrid error detected at detector AD boarderror
050501015Grid error may have caused image abnormalityGrid warning detected by detector AD board but imaging continuedwarning
050501016Firmware version mismatchWrong detector firmware written during firmware updateerror
050502009Detector communication error; restart systemcxdcap.ini file not installedcritical
050502026Detector communication error; cable may be disconnected or power discontinuityCommunication error detected by Ethernet interfaceerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI