Canon CXDI-801W: troubleshooting and error codes
The ServiceForge knowledge base holds 2 documented symptoms for the Canon CXDI-801W, with 2 ranked causes, 1 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Cannot perform next radiographic examination without power cycle
After image transfer with radio environment data loss and timeout, detector requires power off/on before next exam
Ranked causes
- 1. Image transfer timeout during data retransmission after radio interference (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image transfer times out during data retransmission
During image transfer, abnormalities in radio environment cause frequent data loss. Before retransmission of missing data completes, transfer times out, preventing next radiographic examination until detector power is cycled.
Ranked causes
- 1. Radio environment abnormalities causing data loss during transfer (occasional)
Verify detector firmware version
- Test point
- Detector settings or system information
- Expected
- Compatible version with CXDI Controller RF or CXDI Control Software NE Ver.2.1 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI