Canon CXDI-801G/801C: troubleshooting and error codes
The ServiceForge knowledge base holds 17 documented symptoms for the Canon CXDI-801G/801C, with 53 ranked causes, 31 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Signal strength level is 0 or unstable
Level 0 appears as signal strength display on GUI or signal strength level is not stable.
Ranked causes
- 1. X-ray imaging room not tightly sealed (occasional)
- 2. Distance between access point and detector too great (common)
- 3. Structure blocking communication between access point and detector (occasional)
- 4. Thermal therapy device, laser knife, or cordless phone near detector (occasional)
- 5. Interference from another wireless LAN device (occasional)
- 6. DR Wireless SYSTEM in adjacent rooms transferring images simultaneously (occasional)
- 7. Trouble with PCA-WM-DB board inside detector (rare)
- 8. Trouble with access point (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error displayed on GUI during CCS operation
Error message appears while Canon Control Software is running; may involve wireless communication, detector, or interface issues
Ranked causes
- 1. LED lighting status abnormal on detector (common)
- 2. LAN cable connection failure (common)
- 3. Incompatible or defective LAN cable type (occasional)
- 4. Detector cable connection failure (common)
- 5. Defective detector cable (occasional)
- 6. Defective AC/DC box (occasional)
- 7. Defective detector unit (rare)
- 8. Defective image capture computer (rare)
Inspect detector LED lighting status
- Test point
- Detector LEDs
- Expected
- Normal illumination pattern (as per operational specification)
6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infrared detector recognition fails, LEDs do not light
Detector was recognized by infrared communication but the three LEDs on the detector failed to light for about 1 second.
Ranked causes
- 1. Defective connection between image capture computer and ready indicator unit (unknown)
- 2. Problem in work environment (unit too close to monitor or in reflective signal environment) (unknown)
- 3. LNK Controller does not start up (unknown)
- 4. AP/LNK configuration tool improperly set (unknown)
- 5. Trouble with ready indicator unit or infrared data communication unit (unknown)
- 6. Trouble with FPC-UIF board inside detector (unknown)
- 7. Trouble with PCA-WM-DB board inside detector (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No X-ray exposure possible high voltage
X-ray generator does not respond to exposure commands; system unable to acquire radiographic images
Ranked causes
- 1. Wireless communication trouble prevents RX_REQ signal recognition (common)
- 2. Incorrect X-ray interface IP address setting (common)
- 3. Improper X-ray interface cable connection (common)
- 4. Defective PCA-XIF board in X-ray interface box (occasional)
- 5. X-ray exposure in non-generator-connection mode (rare)
Verify X-ray interface IP address setting
- Test point
- AP/LNK configuration tool IP settings
- Expected
- IP address correctly configured
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance test imaging fails radiation
Performance test mode 1 exposure shows errors or artifacts; entire detector surface not properly exposed to X-rays during calibration or test.
Ranked causes
- 1. Photo timer or mounting fixture captured in calibration image (occasional)
- 2. Grid attached during calibration or performance test (occasional)
- 3. Detector panel trouble or extraneous electrical noise (occasional)
- 4. PCA-FE2 board malfunction or detector cable fault (rare)
Verify X-ray generator geometric alignment with detector
- Test point
- Calibration image field
- Expected
- Entire detector surface uniformly exposed without photo timer or mounting fixture visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal image trouble occurred
Images display defects, artifacts, noise, or quality degradation; calibration and diagnostic imaging show abnormality
Ranked causes
- 1. Detector image acquisition defect (common)
- 2. Calibration data out of specification (common)
- 3. Defective detector (occasional)
Run performance test and self-diagnosis imaging
- Test point
- Detector performance diagnostics
- Expected
- Tests pass, no defects detected
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error during calibration
Error or failure occurs during detector calibration procedure; may affect image quality and system accuracy
Ranked causes
- 1. X-ray tube exposure field misconfiguration (common)
- 2. Incorrect exposure dosage (common)
- 3. Defective detector (rare)
Verify X-ray tube exposure field configuration
- Test point
- X-ray generator exposure field
- Expected
- Exposure field properly aligned and configured
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue-green LED fails to light after infrared recognition
The three LEDs of the detector lit for about 1 second with infrared communication and detector was recognized, but blue-green LED failed to light afterward.
Ranked causes
- 1. Access point power not turned on (common)
- 2. Incorrect access point settings using AP/LNK configuration tool (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware and FPGA updating takes excessive time with wireless
Wireless firmware or FPGA update process is significantly slower than expected or stalls; may indicate communication or power issue
Ranked causes
- 1. Wireless communication interference or dropout (common)
- 2. Insufficient battery charge (common)
Verify detector battery charge level
- Test point
- Battery charge indicator
- Expected
- 100% full charge
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image artifact or noise in processed output
Vertical or horizontal lines, rectangular granularity, or other artifacts obscure portions of detector field in processed images; may appear as phantom image on film output.
Ranked causes
- 1. Extraneous objects captured in calibration (photo timer, wires, fixtures) (occasional)
- 2. Detector panel defect or abnormal pixels (rare)
Process phantom image on high-resolution monitor and film
- Test point
- Processed image output
- Expected
- Artifacts visible consistently in both film and digital output; no artifacts visible if using conventional SF film baseline
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-diagnosis offset correction error (Test E)
Test E offset correction test detects errors during self-diagnostic evaluation; indicates detector panel fault or environmental noise.
Ranked causes
- 1. Detector panel trouble or sensor defect (occasional)
- 2. Extraneous electrical noise from environment (occasional)
Review self-diagnosis images for defective pixel distribution
- Test point
- Test E offset correction images
- Expected
- Uniform or randomly distributed defects consistent with normal sensor aging; no clustered anomalies
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-diagnosis test pattern error (Test B)
Test B test pattern inspection detects errors during self-diagnostic test; indicates image generation or data communication fault in detector.
Ranked causes
- 1. PCA-FE2 board not properly generating test image (common)
- 2. Image data communication channel fault (common)
Run self-diagnosis Test A device check
- Test point
- Detector internal circuits
- Expected
- No errors detected in device check
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: detector offset and correction radiation
Calibration procedure generates and updates sensor correction data to maintain normal characteristics and validate sensor performance against standards.
Ranked causes
- 1. Calibration required after system installation or photo timer adjustment (common)
Verify system is operating normally after installation
- Test point
- System initialization and detector communication
- Expected
- All detector and control software functions operational; no error messages
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector LEDs blink (two or three)
The two or three LEDs of the detector blink during operation.
Ranked causes
- 1. Weak wireless signal or communication interference (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image transfer data takes time
It takes time to transfer the data of the images taken from the detector.
Ranked causes
- 1. Weak wireless signal or interference (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line defect or pixel defect in images
Images show line artifacts, missing lines, or dead pixels; image quality degradation
Ranked causes
- 1. Detector requires DP file recalibration or replacement (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preview image takes time to display
It takes time to display a preview image on the GUI.
Ranked causes
- 1. Weak wireless signal or interference (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI