Canon CXDI-70C: troubleshooting and error codes
The ServiceForge knowledge base holds 84 documented symptoms for the Canon CXDI-70C, with 169 ranked causes, 153 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Wireless communication signal strength Level 0 or unstable
Signal strength display shows Level 0 on GUI or signal strength level is not stable
Ranked causes
- 1. X-ray imaging room not tightly sealed or door open (common)
- 2. Distance between access point and detector too great (common)
- 3. Structure blocking communication between access point and detector (common)
- 4. Thermal therapy device, laser knife, cordless telephone near detector (occasional)
- 5. Interference from other wireless LAN device signals (common)
- 6. Multiple CXDI-70C systems in adjacent rooms transferring images simultaneously (occasional)
- 7. Trouble with PCA-11nWM board inside detector (rare)
- 8. Trouble with access point (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless signal strength level 0 or unstable
Signal strength display on GUI shows Level 0 or signal strength level is not stable
Ranked causes
- 1. X-ray imaging room not tightly sealed or door open (common)
- 2. Distance between access point and detector too great (common)
- 3. Structure blocking communication between access point and detector (occasional)
- 4. Thermal therapy device, laser knife, cordless telephone near detector (occasional)
- 5. Interference from another wireless LAN device (occasional)
- 6. Adjacent CXDI-70C or CXDI-80C transferring images simultaneously (occasional)
- 7. Trouble with PCA-WM-DB board (occasional)
- 8. Trouble with access point (occasional)
Verify X-ray imaging room is sealed
- Test point
- Door and room sealing condition
- Expected
- Door closed, room tightly sealed
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless signal strength shows Level 0 or unstable
Signal strength displays as Level 0 on GUI or fluctuates without stabilizing
Ranked causes
- 1. X-ray imaging room not tightly sealed or door open (common)
- 2. Distance between access point and detector too great (common)
- 3. Structure blocking communication between access point and detector (occasional)
- 4. Nearby thermal therapy device, laser knife, or cordless telephone (occasional)
- 5. Interference from another wireless LAN device (common)
- 6. Two CXDI-70C systems in adjacent rooms transferring images simultaneously (occasional)
- 7. Trouble with PCA-11nWM board inside detector (rare)
- 8. Trouble with access point (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infrared communication detection fails
Detector was recognized by infrared but three LEDs on detector failed to light for about 1 second
Ranked causes
- 1. Defective connection between image capture computer and infrared data communication unit (unknown)
- 2. Problem in work environment (common)
- 3. Failure of LNK Controller service tool to start up (unknown)
- 4. Incorrect AP/LNK configuration tool COM port setting (common)
- 5. Trouble with infrared data communication unit (occasional)
- 6. Trouble with FPC-UIF board inside detector (occasional)
- 7. Trouble with PCA-WM-DB board inside detector (occasional)
Verify infrared data communication unit connection
- Test point
- Connection between image capture computer and infrared unit
- Expected
- Secure connection
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infrared communication detector recognition failed
Detector recognized via infrared but three LEDs on detector failed to light for about 1 second
Ranked causes
- 1. Defective connection between image capture computer and infrared data communication unit (unknown)
- 2. Problem in work environment (unknown)
- 3. Failure of LNK Controller service tool to start up (unknown)
- 4. Incorrect COM port setting in AP/LNK configuration tool (unknown)
- 5. Trouble with infrared data communication unit (unknown)
- 6. Trouble with FPC-UIF board inside detector (unknown)
- 7. Trouble with PCA-11nWM board inside detector (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infrared communication with detector fails
Detector not recognized via infrared communication; three LEDs on detector do not light for about 1 second
Ranked causes
- 1. Defective connection between image capture computer and infrared data communication unit (unknown)
- 2. Problem in work environment (monitor proximity or signal reflection) (unknown)
- 3. Failure of LNK Controller service tool to start up (unknown)
- 4. Incorrect COM port setting in AP/LNK configuration tool (unknown)
- 5. Trouble with infrared data communication unit (unknown)
- 6. Trouble with FPC-UIF board (board inside detector) (unknown)
- 7. Trouble with PCA-11nWM board (board inside detector) (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error displayed in GUI during CCS operation
Error message appears on control software GUI while CCS is running
Ranked causes
- 1. Detector cable connection issue (common)
- 2. LAN cable connection problem (common)
- 3. Detector LED status abnormal (occasional)
- 4. Defective AC/DC box (rare)
- 5. Defective detector unit (rare)
- 6. Defective image capture computer (rare)
Verify LED lighting status on detector
- Test point
- Detector LEDs
- Expected
- All LEDs illuminated and normal state
6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No X-ray exposure possible high voltage
System unable to perform X-ray exposure
Ranked causes
- 1. Wireless communication trouble (occasional)
- 2. Incorrect X-ray interface IP address setting (common)
- 3. Improper connection of X-ray interface cable (common)
- 4. Defective PCA-XIF board in X-ray interface box (occasional)
Verify X-ray interface IP address setting
- Test point
- IP address in AP/LNK configuration tool
- Expected
- Correct IP address for X-ray interface
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No X-ray exposure possible radiation
System cannot trigger or execute X-ray exposures
Ranked causes
- 1. Wireless communication trouble affecting X-ray control signal (common)
- 2. Incorrect X-ray interface IP address configuration (occasional)
- 3. X-ray interface cable improperly connected (occasional)
- 4. Defective PCA-XIF board in X-ray interface box (rare)
Test X-ray control with wired connection
- Test point
- Wired X-ray interface cable connected to X-ray generator
- Expected
- X-ray exposure possible when triggered
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance test error persistence after recalibration radiation
If performance test error persists after recalibration, investigate full detector exposure, exposure proper without shading, grid removed, and defects corrected.
Ranked causes
- 1. Detector not fully exposed to X-rays (common)
- 2. X-ray exposure improper with shading on detector surface (common)
- 3. Grid attached during performance test (occasional)
- 4. Defects not properly rectified (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-diagnosis test errors
Self-diagnostic test function includes Test A device check, Test B test pattern check, and Test E offset correction test to verify sensor integrity without X-ray exposure.
Ranked causes
- 1. PCA-FE2 board not generating test image properly (occasional)
- 2. Fault in image data communication channel (occasional)
- 3. Detector cable connection problem or broken wire (occasional)
- 4. Detector panel trouble or extraneous noise (Test E) (occasional)
Review self-diagnosis images for abnormal or defective pixels
- Test point
- Self-diagnosis image
- Expected
- No abnormal or defective pixels
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error displayed on GUI during operation
Error message appears on control software interface while CCS is running
Ranked causes
- 1. Defective detector cable (unknown)
- 2. Faulty AC/DC box (unknown)
- 3. Defective detector (unknown)
- 4. Faulty image capture computer (unknown)
Check LED lighting status on detector
- Test point
- detector LEDs
- Expected
- normal operation as documented
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration imaging error
Error occurs during detector calibration process
Ranked causes
- 1. X-ray tube exposure field incorrect (common)
- 2. Exposure dosage incorrect (common)
- 3. Defective detector unit (occasional)
Verify X-ray tube exposure field alignment
- Test point
- X-ray generator
- Expected
- Exposure field properly aligned with detector coverage area
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error during calibration
Error message displayed when performing calibration procedure
Ranked causes
- 1. Incorrect X-ray tube exposure field (unknown)
- 2. Improper exposure dosage (unknown)
- 3. Defective detector (unknown)
Check X-ray tube exposure field
- Test point
- X-ray generator settings
- Expected
- correct exposure field for calibration
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Errors during performance test imaging radiation
Errors occur during performance test imaging possibly related to photo timer installation or detector panel issues.
Ranked causes
- 1. Photo timer artifact or geometric deviation during calibration (occasional)
- 2. Low signal-to-noise ratio in photo timer installed parts area (occasional)
- 3. Defective detector panel or pixels (rare)
Verify detector surface X-ray exposure during performance test
- Test point
- Entire detector panel surface
- Expected
- Full exposure under same calibration conditions with no subject, no shading anywhere
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Errors in test pattern inspection (Test B) radiation
Self-diagnosis test B detects errors during test pattern inspection indicating potential data path or communication problems.
Ranked causes
- 1. Malfunction of PCB (PCA-FE board) inside detector (occasional)
- 2. Detector cable connection problem or broken wire (occasional)
- 3. Wireless communication trouble (occasional)
Inspect detector cable connections
- Test point
- Detector cable connectors and wiring
- Expected
- Connectors fully seated, no visible damage or corrosion
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue-green LED fails to light after infrared recognition
Three LEDs light for about 1 second with infrared communication and detector recognized successfully but blue-green LED failed to light
Ranked causes
- 1. Access point power not turned on (common)
- 2. Incorrect access point settings in AP/LNK configuration tool (common)
Verify access point power status
- Test point
- Access point power indicator
- Expected
- Power on, indicator lit
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Firmware and FPGA update for CCS V1.4x/V2.0x environment
Step-by-step procedure to update firmware (01170200) and FPGA (02010002) code when user point and rewriting environment are both CCS V1.4x or V2.0x
Ranked causes
- 1. Required firmware update for CCS V1.4x/V2.0x environment (common)
- 2. Required FPGA update for CCS V1.4x/V2.0x environment (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Firmware and FPGA update for CCS V2.1x environment
Step-by-step procedure to update firmware (01180200) and FPGA (02030000) code when user point is CCS V2.1x
Ranked causes
- 1. Required firmware update for CCS V2.1x environment (common)
- 2. Required FPGA update for CCS V2.1x environment (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot access PCA-XIF board web interface
Web browser cannot connect to board management screen; IP address unknown or communication failed
Ranked causes
- 1. Unknown or changed IP address (occasional)
- 2. Incorrect network connection or cable type (occasional)
Verify board IP address is set to default
- Test point
- BOOTSEL/RESET switch operation
- Expected
- Status LED indicates successful boot
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector wireless communication trouble
Detector fails to communicate wirelessly with system; may recur intermittently
Ranked causes
- 1. Detector malfunction not resolved by primary response (unknown)
- 2. Wireless environment interference or misconfiguration (unknown)
Verify detector is recognized by system
- Test point
- Infrared communication link
- Expected
- Detector recognized and connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ensure channel frequencies do not overlap with existing wireless systems
Guidance on setting CXDI-70C wireless channels to avoid partial frequency overlap that degrades both CXDI-70C and existing wireless LAN performance
Ranked causes
- 1. Partial frequency overlap between CXDI-70C HT20 and existing HT40 systems (common)
- 2. Partial frequency overlap with variable transfer time (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error during calibration patient safety
Errors occur during calibration procedure
Ranked causes
- 1. X-ray tube exposure field incorrect (common)
- 2. Incorrect exposure dosage (common)
Verify X-ray tube exposure field
- Test point
- X-ray generator exposure field
- Expected
- Correct exposure field coverage
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error or abnormal image trouble
Detector displays errors or produces abnormal/defective images
Ranked causes
- 1. Detector hardware fault (unknown)
- 2. Line or pixel defects in detector (occasional)
Verify sufficient data was collected at primary response
- Test point
- collected data volume
- Expected
- adequate data collected
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware and FPGA update takes excessive time
Firmware and FPGA updating with wireless communication takes longer than expected
Ranked causes
- 1. Wireless communication trouble (common)
- 2. Insufficient battery charge (common)
Verify detector battery charge level
- Test point
- Battery charge indicator on detector or GUI
- Expected
- 100% battery charge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware or FPGA update over wireless takes excessive time
Firmware or FPGA update process takes much longer than expected when using wireless connection
Ranked causes
- 1. Wireless communication trouble or interference (common)
- 2. Insufficient battery charge (common)
Check wireless signal quality and interference
- Test point
- Wireless communication link
- Expected
- Strong signal without excessive interference from other wireless LAN devices
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware update fails or becomes corrupted
Normal area firmware erased or damaged during update; communication error or power loss during updating process
Ranked causes
- 1. Communication error during normal area firmware update (occasional)
- 2. Power loss during normal area firmware update (occasional)
Verify backup firmware can be booted
- Test point
- STATUS LED
- Expected
- LED blinks during backup booting
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Images have shading or other artifacts
Captured images display shading, grid stripes, or quality degradation after X-ray exposure
Ranked causes
- 1. RX_COM to X-ray exposure stop delay exceeds specification (occasional)
- 2. Detector cable connector area subjected to mechanical stress or strain (occasional)
Measure time from RX_COM negation to X-ray exposure stop
- Test point
- RX_COM signal and X-ray exposure signal at generator interface
- Expected
- max.100ms, ≤100ms
- Tools
- Oscilloscope
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infrared communication succeeds but blue-green LED does not light
Three LEDs light for about 1 second and detector is recognized via infrared, but blue-green LED fails to light during wireless setup
Ranked causes
- 1. Power of wireless access point not turned on (common)
- 2. Incorrect setting of access point using AP/LNK configuration tool (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infrared communication successful but blue-green LED failed to light
Three LEDs lit for about 1 second during infrared communication and detector recognized successfully, but blue-green LED did not light
Ranked causes
- 1. Power of access point not turned on (common)
- 2. Incorrect access point setting using AP/LNK configuration tool (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IP address change does not take effect
New IP address configured but system still uses old address after power cycle
Ranked causes
- 1. New IP address not written to flash memory before power cycle (common)
- 2. AP/LNK configuration not updated in Service Tool (common)
Verify new IP address written to flash memory
- Test point
- PCA-XIF board telnet connection
- Expected
- ipadr command shows new address in flash memory field
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance test error on photo timer use radiation
Using photo timer with stand mounting may cause performance test error depending on type and mounting configuration. Extraneous images captured due to geometrical misalignment or fixtures.
Ranked causes
- 1. Photo timer image captured due to X-ray generator misalignment (occasional)
- 2. Photo timer mounting fixture or wires captured in calibration image (occasional)
Review calibration, performance test and self-diagnosis images for extraneous content
- Test point
- Image content
- Expected
- No phantom images or photo timer artifacts
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slow firmware or FPGA update over wireless connection
Firmware and FPGA updating takes excessive time when using wireless communication
Ranked causes
- 1. Wireless communication trouble (common)
- 2. Insufficient battery charge (common)
Check battery charge level
- Test point
- battery indicator
- Expected
- 100% charge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not connect with existing CXDI detector
Cannot establish network communication between CXDI-70C Wireless and existing CXDI system (40EG/EC, 40G, 50G/C, 60G/C, 55G/C) after installation
Ranked causes
- 1. IP address conflict between 70C Wireless and existing detector (common)
- 2. 70C detector IP address conflicts with other 70C detector in same location (occasional)
Verify IP address assignment of existing detector before installation
- Test point
- Existing detector network configuration
- Expected
- IP address is 192.168.100.11 (factory default) or has been previously changed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal image artifacts or quality issues patient safety
Image shows artifacts, noise, or quality degradation during operation
Ranked causes
- 1. Detector imaging function degradation (common)
Perform performance test and self-diagnosis imaging
- Test point
- Detector and control software
- Expected
- Test and self-diagnosis images display normally without artifacts
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal image quality
Image data shows defects, artifacts, or quality degradation
Ranked causes
- 1. Defective detector (unknown)
Perform performance test and self-diagnosis imaging
- Test point
- detector self-test
- Expected
- normal diagnostic results
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal image trouble patient safety
Images show abnormal appearance or artifacts
Ranked causes
- 1. Detector malfunction (occasional)
Run performance test and self-diagnosis imaging
- Test point
- Detector and system performance
- Expected
- All tests pass; self-diagnosis images normal
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC/DC box, PCA-PB board, and power supply replacement commissioning high voltage
After power supply component replacement, system must verify boot and normal operation
Ranked causes
- 1. AC/DC box, PCA-PB board, or AC/DC power supply replaced (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access point wireless communication trouble
Access point fails to establish or maintain wireless communication; may occur at installation site
Ranked causes
- 1. Access point malfunction or misconfiguration (unknown)
Connect access point to loan CXDI-70C system
- Test point
- Access point connection and power
- Expected
- Access point powered and connected to system
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery connector pad torn or deteriorated patient safety
Connector pad on battery holder shows tears, wear, or material degradation from repeated battery installation/removal cycles. Risk of connector pin damage and foreign matter entry causing shorts.
Ranked causes
- 1. Connector pad material fatigue from repeated use (common)
Inspect connector pads for tears or deterioration
- Test point
- Battery holder connector pad mounting surface
- Expected
- Pads intact, no tears or worn areas
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery connector pads peeled off
Connector pads on battery holder have detached; risk of damaged pins or short circuit if battery reattached
Ranked causes
- 1. Connector pads adhesive failure (common)
Inspect battery connector pad condition and adhesion
- Test point
- Battery holder connector pads
- Expected
- Pads firmly adhered to mounting surface
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery connector pins damaged patient safety
Connector pins on battery or holder show physical damage, bending, or corrosion, preventing proper battery contact or insertion.
Ranked causes
- 1. Torn connector pad allowed pin contact damage (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery connector pins damaged or short circuit caution
Connector pins may be damaged or short-circuit caused by foreign substances entering through gap when battery loaded without connector pad
Ranked causes
- 1. Battery connector pad missing or deteriorated (common)
Verify connector pad is properly positioned
- Test point
- Battery connector pad location
- Expected
- Opening not covered by pad, pad does not exceed blue edge line by more than 0.1mm, <=0.1mm from edge line
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AC/DC box unit, PCA-PB board, and power supply replacement caution
Post-replacement verification after AC/DC box unit, PCA-PB board, or AC/DC power supply swap
Ranked causes
- 1. AC/DC box unit, PCA-PB board, or AC/DC power supply replaced (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: detector cable replacement caution
Post-replacement verification after detector cable swap
Ranked causes
- 1. Detector cable replaced (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: detector replacement
Post-replacement verification and calibration workflow after detector unit swap
Ranked causes
- 1. Detector unit replaced (common)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensor serial number setup after FE board exchange
Configure sensor serial number on PCA-FE2 board via telnet after replacing FE board on CXDI-70C Wireless detector
Ranked causes
- 1. FE board replaced, sensor serial number requires reconfiguration (common)
Verify PC TCP/IP network configuration
- Test point
- PC network adapter
- Expected
- IP address 192.168.100.10, subnet mask 255.255.255.0
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: X-ray interface box and PCA-XIF board replacement
Post-replacement verification after X-ray interface box or PCA-XIF board swap
Ranked causes
- 1. X-ray interface box or PCA-XIF board replaced (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel interference between wireless systems
Guidance on wireless channel settings when DR Wireless systems share environment with existing wireless LAN
Ranked causes
- 1. Suboptimal wireless channel selection for environment (common)
Identify existing wireless LAN system configuration
- Test point
- Existing wireless LAN band and mode (HT20 or HT40)
- Expected
- Configuration documented
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector cable replacement commissioning
After detector cable replacement, system must verify boot and normal wired operation
Ranked causes
- 1. Detector cable has been replaced (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector error or abnormal image trouble
Detector generates error codes or produces abnormal images; may occur intermittently
Ranked causes
- 1. Detector defect requires tertiary response analysis (unknown)
Verify sufficient troubleshooting data was collected
- Test point
- Collection Tool files from primary response
- Expected
- Adequate data collected; if insufficient, request additional data
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector LEDs blink abnormally
Three or two LEDs of the detector blink
Ranked causes
- 1. Wireless communication trouble (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector LEDs blink intermittently
Three or two LEDs on detector blink instead of steady operation
Ranked causes
- 1. Wireless communication interference or instability (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector LEDs blinking
Three or two LEDs of the detector are blinking
Ranked causes
- 1. Wireless communication interference or unstable signal (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector replacement commissioning
After detector unit replacement, system must be reconfigured and validated before return to service
Ranked causes
- 1. Detector unit has been replaced (common)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector warping exceeds standard patient safety
X-ray detector unit shows visible warping or deformation measured at multiple points on the surface.
Ranked causes
- 1. Detector warping due to handling or impact (occasional)
Measure detector warp at eight points
- Test point
- Detector unit flat surface
- Expected
- 2mm or less
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error or abnormal image trouble recurs
Detector produces errors or abnormal images after primary response troubleshooting
Ranked causes
- 1. Detector hardware defect (unknown)
Detector recognition via infrared communication
- Test point
- infrared communication to loan system
- Expected
- Detector is recognized
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Errors in offset image inspection (Test E) radiation
Self-diagnosis test E detects errors during offset image inspection indicating detector panel trouble or external noise issues.
Ranked causes
- 1. Detector panel trouble or extraneous noise (common)
Verify calibration before concluding detector panel defect
- Test point
- Calibration status and offset image results
- Expected
- Calibration successfully accomplished, offset image shows acceptable noise levels
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Establish telnet connection to PCA-XIF board
Connect to board via telnet for command-line management; requires telnet client or TeraTerm software
Ranked causes
- 1. Need command-line access to board (occasional)
Install/enable telnet client software
- Test point
- PC operating system telnet availability
- Expected
- TeraTerm installed or Windows telnet client enabled
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware and FPGA code update for CXDI-70C Wireless
Update firmware, FPGA code, and DP file on detector with PCA-FE2 board via wired connection using CCS service tool
Ranked causes
- 1. Routine firmware and FPGA maintenance update (common)
Verify PC is connected to detector with wiring unit WU-1A
- Test point
- WU-1A connector
- Expected
- Wiring unit WU-1A and sensor cable connected to detector
10 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Firmware and FPGA code update procedure
- Image data transfer takes excessive time
- Image display timing delayed
- Image preview display takes excessive time
- Imaging time reduced
- Incomplete PCA-XIF board log file download
- Line defect or pixel defect in images
- Line defect or pixel defect in images
- Many abnormal or defective pixels in self-diagnosis images
- PCA-FE board firmware or FPGA update failed or corrupted
- Reset IP address to default
- Return to normal mode after backup mode firmware/FPGA update
- Slow image data transfer
- Slow image data transfer
- Slow preview image display
- Slow preview image display
- System error displayed on GUI during operation
- Update backup area firmware
- Update normal area firmware via web browser
- Web browser prompts for credentials
- Wireless communication trouble
- Wireless communication trouble recurs
- Wireless signal strength degradation
- X-ray interface box and PCA-XIF board replacement commissioning
24 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI