Canon CXDI-701G/701C: troubleshooting and error codes
The ServiceForge knowledge base holds 15 documented symptoms for the Canon CXDI-701G/701C, with 48 ranked causes, 24 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Wireless communication signal strength level 0 or unstable
Signal strength display shows level 0 on GUI or signal strength level is not stable
Ranked causes
- 1. X-ray imaging room not tightly sealed (occasional)
- 2. Distance between access point and detector too great (common)
- 3. Structure blocking communication path (occasional)
- 4. Thermal therapy device or cordless phone interference (occasional)
- 5. Interference from other wireless LAN devices (occasional)
- 6. Multiple DR Wireless systems transferring images simultaneously (occasional)
- 7. PCA-WM-DB board failure (rare)
- 8. Access point malfunction (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector not recognized via infrared communication
Three LEDs on detector fail to light after infrared communication attempt
Ranked causes
- 1. Defective connection between image capture computer and ready indicator unit (unknown)
- 2. Problem in work environment (unknown)
- 3. LNK Controller does not start up (unknown)
- 4. AP/LNK configuration tool improperly set (unknown)
- 5. Ready indicator unit malfunction (unknown)
- 6. FPC-UIF board failure (unknown)
- 7. PCA-WM-DB board failure (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error displayed on GUI during CCS operation
Error message appears on the control software interface while Collection Control System is running
Ranked causes
- 1. LED status indicates detector malfunction (common)
- 2. LAN cable connection issue at any location (common)
- 3. Incorrect LAN cable type in use (occasional)
- 4. Detector cable disconnection or damage (common)
- 5. AC/DC box malfunction (occasional)
- 6. Detector hardware failure (occasional)
- 7. Image capture computer malfunction (rare)
Examine LED status on detector
- Test point
- All LEDs on detector unit
- Expected
- Normal operation state
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No X-ray exposure possible during imaging high voltage
X-ray generator does not trigger or produce exposure when commanded by detector
Ranked causes
- 1. Wireless communication loss or interference with RX_REQ signal (common)
- 2. Incorrect X-ray interface IP address configuration (occasional)
- 3. X-ray interface cable disconnection or poor connection (common)
- 4. PCA-XIF board failure in X-ray interface box (occasional)
- 5. Attempted exposure in non-generator-connection mode (rare)
Attempt X-ray exposure with wired connection
- Test point
- Detector to X-ray generator interface
- Expected
- X-ray exposure triggers successfully via wired link
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance test generates errors after calibration radiation
Performance test imaging fails or shows errors despite successful prior calibration; may indicate detector exposure or signal issues
Ranked causes
- 1. Uneven X-ray exposure on detector surface (common)
- 2. Grid attached during performance test (occasional)
- 3. Photo timer or mounting obstruction in field (occasional)
- 4. PCA-FE2 board malfunction in detector (rare)
Verify detector receives uniform X-ray exposure
- Test point
- entire detector surface
- Expected
- No shading anywhere on entire detector surface
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error during detector calibration
Error message appears while performing calibration procedure on the detector
Ranked causes
- 1. X-ray tube exposure field setting incorrect (common)
- 2. Exposure dosage out of range (common)
- 3. Detector calibration data corruption (occasional)
Verify X-ray tube exposure field setting
- Test point
- X-ray generator configuration
- Expected
- Correct exposure field geometry as required
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal image quality or defects in captured images patient safety
Images contain defects, lines, pixels, noise, or processing errors in captured X-ray images
Ranked causes
- 1. Detector imaging sensor malfunction (occasional)
- 2. Line or pixel defects in detector sensor (occasional)
Execute performance test and self-diagnosis imaging
- Test point
- Detector diagnostic imaging sequence
- Expected
- Test images acquired without defects
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue-green LED fails to light after infrared recognition
Three LEDs light during infrared communication and detector recognized successfully, but blue-green LED remains off
Ranked causes
- 1. Access point power not turned on (common)
- 2. Access point incorrectly configured (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware and FPGA update takes excessive time
Wireless update of firmware or FPGA runs significantly longer than expected
Ranked causes
- 1. Wireless communication interference or poor signal (common)
- 2. Insufficient battery charge on detector (common)
Verify wireless communication signal strength
- Test point
- Wireless link quality indicators
- Expected
- Strong signal with minimal interference
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-diagnosis offset correction test errors
Self-diagnosis Test E (offset correction test) detects abnormalities in detector panel or excessive noise in image data
Ranked causes
- 1. Detector panel trouble or extraneous noise (common)
- 2. Abnormal or defective pixels in detector (occasional)
Examine self-diagnosis images for abnormal pixels
- Test point
- detector output images
- Expected
- No abnormal or defective pixel patterns
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-diagnosis test pattern check errors
Self-diagnosis Test B (test pattern check) fails; indicates internal detector test image generation or communication issue
Ranked causes
- 1. PCA-FE2 board not properly generating test image (common)
- 2. Fault in image data communication channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensor adjustment and correction data generation radiation
Calibration procedure to perform dose checking and generate correction data for sensor characteristics maintenance
Ranked causes
- 1. Routine calibration after installation or photo timer adjustment (common)
Verify system operating normally after installation
- Test point
- overall system function
- Expected
- All system functions operational without errors
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector LEDs blinking
Three or two LEDs on detector blink
Ranked causes
- 1. Wireless communication interference (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image data transfer delay
Delayed transfer of image data from detector after image is taken
Ranked causes
- 1. Wireless communication interference or obstruction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preview image display delay
Delayed display of preview image on GUI
Ranked causes
- 1. Wireless communication interference or obstruction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.