Canon CXDI-60G: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Canon CXDI-60G, with 55 ranked causes, 69 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
CXDI-60G/C errors during software startup radiation
System fails to reach ready state; log gathering tool execution and sequential connector/cable verification required
Ranked causes
- 1. Log data not gathered (common)
- 2. Power cable connections loose or disconnected (common)
- 3. Imaging unit, power box, or remote control lamp signal absent (occasional)
- 4. LAN cable disconnected at FC-24VE/E21A side (common)
- 5. LAN cable disconnected at power box (common)
- 6. Incorrect LAN cable type or category (occasional)
- 7. Sensor cable connector not properly positioned (occasional)
- 8. Sensor cable connection loose at power box (occasional)
- 9. Sensor cable damaged or faulty at power box side (occasional)
- 10. Sensor cable damaged or faulty at imaging unit side (occasional)
- 11. Power box hardware failure (occasional)
Execute log gathering tool
- Test point
- Diagnostic software
- Expected
- Log file generated without errors
9 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails at host startup
Imaging unit does not establish connection with control PC at system initialization; LAN or power communication failure.
Ranked causes
- 1. Power cable disconnected or loose (common)
- 2. Imaging unit, power box, or remote switch blue lamp not transmitting (common)
- 3. LAN cable disconnected or loose at FC-24VE (common)
- 4. LAN cable disconnected or loose at power box (common)
- 5. Incorrect LAN cable type used (occasional)
- 6. Sensor cable loose at imaging unit or power box (common)
- 7. Sensor cable damaged or faulty at power box (occasional)
- 8. Power box module failure (occasional)
- 9. Sensor cable damaged at imaging unit side (occasional)
Execute log gathering tool to collect diagnostic data
- Test point
- Host control PC
- Expected
- Log file generated without errors
10 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration fails during operation radiation
Error occurs when unit attempts to perform calibration sequence; exposure dosage or X-ray generator field issues prevent successful calibration.
Ranked causes
- 1. X-ray generator irradiation field incorrect (occasional)
- 2. Exposure dosage outside acceptable range (occasional)
- 3. Sensor self-diagnosis test failure (occasional)
- 4. Sensor cable loose or damaged at imaging unit or power box (common)
- 5. Sensor cable damaged at power box (occasional)
- 6. Power box module failure (occasional)
- 7. Sensor cable damaged at imaging unit side (occasional)
Verify X-ray generator irradiation field alignment
- Test point
- X-ray tube output aperture vs imaging unit sensor
- Expected
- Field fully covers sensor aperture without obstruction
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grid detection sensor not responding
Grid detection reed switch fails to activate when grid frame is positioned; magnet does not trigger sensor closure.
Ranked causes
- 1. Magnet surface distance exceeds 12 mm from sensor (common)
- 2. Magnetic material or obstruction between sensor and magnet (occasional)
- 3. Magnet magnetization direction incorrect (occasional)
- 4. Magnet insufficient flux density for detection distance (rare)
Measure distance from grid frame magnet surface to sensor unit
- Test point
- Grid detection sensor location (314.4 mm from left edge, 30 mm from right edge, 54.5 mm from effective area edge, 318 mm
- Expected
- 12 mm, 0 mm (exact)
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Secondary troubleshooting required after primary response radiation
Error persists after primary-level steps; secondary response performed at distributor site with loaner system swap and cable/module replacement
Ranked causes
- 1. Sensor cable connector position still incorrect (rare)
- 2. Imaging unit is faulty (occasional)
- 3. Error caused by imaging unit, not cables or power box (occasional)
- 4. Sensor cable connection still loose at imaging/power box sides (occasional)
Gather additional diagnostic data from primary response
- Test point
- Log files from primary troubleshooting
- Expected
- Complete data set captured for analysis
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal images produced during operation radiation
System generates images with visual defects, artifacts, or quality degradation
Ranked causes
- 1. Sensor cable degradation or connection fault (common)
- 2. Power box malfunction (occasional)
- 3. Imaging unit sensor or internal hardware failure (rare)
Execute log gathering tool
- Test point
- System logging function
- Expected
- Log files collected successfully
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error occurs after connector attachment or removal high voltage
Error recurs following connector operations on sensor cable
Ranked causes
- 1. Sensor cable connector misaligned or not in right position (common)
- 2. Power supply lines misaligned causing overcurrent (common)
- 3. Sensor unit or power box damaged (occasional)
Verify sensor cable connector position before power application
- Test point
- Sensor cable connector
- Expected
- Connector fully seated in correct position
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Errors occurring prior to ready transfer, during ready transfer, during X-ray exposure, or during exposure standby radiation
System error occurs at various stages during operation or standby mode
Ranked causes
- 1. Sensor cable connection fault or degraded connection (common)
- 2. Power box malfunction or failure (occasional)
- 3. Imaging unit hardware failure (rare)
Execute log gathering tool
- Test point
- System logging function
- Expected
- Log files collected without error
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bend relief repair on detector cable caution
Detector cable bend relief requires replacement and sealing with heat-shrinkable tubing to prevent damage and ensure proper connector protection.
Ranked causes
- 1. Bent or damaged bend relief on detector cable (unknown)
Wind sealing tape around detector cable at marked portion
- Test point
- Cable at red circle mark
- Expected
- Tape wound approximately 1.5 circumferences
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bend relief sealing tape wrapping caution
Sealing tape must be wound in specific pattern and tension around detector cable bend relief to ensure proper seal and prevent connector damage.
Ranked causes
- 1. Improper sealing tape application causing gaps or loose wraps (unknown)
Verify sealing tape starting point
- Test point
- Relay connector side of bend relief
- Expected
- Tape application begins at relay connector side
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector cable bend relief cracked or broken caution
Visible crack or breakage in the protective tube portion (bend relief) at the detector cable connector
Ranked causes
- 1. Mechanical stress or flexing of detector cable at connector (unknown)
Visually inspect bend relief for cracks or breakage
- Test point
- Detector cable connector protective tube portion
- Expected
- No visible cracks or breaks in protective tube
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grid unit lock base screw damage during assembly caution
Excessive tightening torque damages screw threads of the slide switch during grid lock assembly installation
Ranked causes
- 1. Excessive tightening torque applied to lock base screws (common)
Tighten lock base attachment screws
- Test point
- Screw XA4-1200-609 on grid lock assembly
- Expected
- 1kgf·cm (9.8N·cm), exact
- Tools
- torque wrench calibrated to 9.8 N·cm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grid unit lock base screw thread damage caution
Screw threads of slide switch (5) damaged or stripped during lock base installation due to excessive tightening torque
Ranked causes
- 1. Excessive torque applied when tightening lock base screw (XA4-1200-609) (common)
Tighten lock base screw (XA4-1200-609) to specification
- Test point
- Lock base (6) to slide switch (5) fastening screw
- Expected
- 1kgf·cm (9.8N·cm), do not exceed
- Tools
- Torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grid unit lock base screw threads damaged caution
Excessive torque applied when tightening screws to attach lock base causes damage to slide switch screw threads.
Ranked causes
- 1. Excessive torque applied to grid lock base screws (common)
Verify correct torque applied to grid lock base screw
- Test point
- Screw XA4-1200-609 at lock base
- Expected
- 1kgf·cm (9.8N·cm), exact
- Tools
- torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cable connector deterioration or wear
Connector gradually deteriorates or wears out due to material and structural characteristics, requiring replacement during maintenance visits
Ranked causes
- 1. Normal connector wear from repeated connect/disconnect cycles (common)
Inspect connector for deterioration or wear
- Test point
- Sensor cable connector (detector unit side and power box side)
- Expected
- Connector in good condition, no visible wear or deterioration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cable connector failure from excessive connect/disconnect cycles
Connector fails or becomes unreliable after reaching or exceeding maximum guaranteed operational cycles
Ranked causes
- 1. Connector use exceeds guaranteed connect/disconnect cycle limit of 4000 (occasional)
Verify connector has not exceeded 4000 connect/disconnect cycles
- Test point
- Sensor cable connector (detector unit and power box sides)
- Expected
- Connector cycle count less than 4000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cable connector misaligned or recessed
Terminal pins stick out 2–3 mm lower than normal, or terminal socket is flush with inner groove instead of projecting 3 mm upward.
Ranked causes
- 1. Connector terminal pins recessed or socket misaligned (unknown)
Inspect sensor cable connector on power box side
- Test point
- Connector terminal pins
- Expected
- Terminal pins stick out sufficiently
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cable connectors misaligned during removal or attachment
Loosening the back nut while removing and attaching detachable connectors may cause misalignment, compromising data transmission or physical fit.
Ranked causes
- 1. Back nut loosened during connector removal/attachment without adhesive locking (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cable detachable connector deterioration or wear
Connector shows signs of deterioration or wear during periodical maintenance or customer site visits
Ranked causes
- 1. Normal wear from connect/disconnect cycles exceeding service life (common)
Visually inspect detachable connector for signs of deterioration or wear
- Test point
- Sensor cable connector (detector unit side and power box side)
- Expected
- Connector appears undamaged and functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI