Canon CXDI-50G: troubleshooting and error codes
The ServiceForge knowledge base holds 54 documented symptoms for the Canon CXDI-50G, with 72 ranked causes, 112 diagnostic checks with test points and expected values, and 32 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Linearity check of transfer image density
Verify that density linearity of printed and displayed images matches CXDI transferred image density using SMPTE test pattern
Ranked causes
- 1. Inadequate printer and monitor calibration (common)
- 2. Incorrect CXDI printer parameter settings (common)
- 3. Faults in printer or monitor hardware (occasional)
Verify printer and monitor connections and output settings are configured
- Test point
- System settings and printer parameters
- Expected
- Printer and monitor output set to LINEAR; calibration completed
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM Transfer Error message displayed
Error message 'DICOM Transfer Error. Error occurred during the association. Retry?' appears when DICOM level communication fails despite TCP/IP connectivity.
Ranked causes
- 1. AE_TITLE not properly sent from CXDI to transmission destination (common)
- 2. Transmission destination IP address or port number setting incorrect (common)
- 3. AE_TITLE of transmission destination incorrect in CXDI settings (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grounding resistance out of specification after PCB-50LED attachment
After attaching PCB-50LED with aluminum spacer, resistance measurement between sensor unit cover and PCB-50LED through-hole exceeds 0.1 ohm limit, indicating inadequate grounding connection.
Ranked causes
- 1. Improper screw tightening or contact (common)
- 2. Aluminum spacer bent part touching rib (occasional)
- 3. Poor probe contact during resistance measurement (occasional)
Verify screw torque specification
- Test point
- Screw XB1-2260-606
- Expected
- 4 kgf·cm
- Tools
- Torque screwdriver
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network communication not established
CXDI-50G/50C imaging part cannot communicate with control PC via Ethernet for X-ray image transfer
Ranked causes
- 1. TCP/IP settings not configured on control PC network card (common)
- 2. Screwcap.ini not properly configured (common)
- 3. CXDI-50G/50C sensor network settings not configured (common)
Verify control PC TCP/IP configuration
- Test point
- Network card adapter settings
- Expected
- IP Address=192.168.100.10, Subnet Mask=255.255.255.0
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Non-linearity detected in transfer image density
Measurement values deviate too far from straight line or maximum density does not correspond to settings
Ranked causes
- 1. Inadequate calibration of printer and monitor (common)
- 2. Incorrect CXDI printer parameter settings (common)
- 3. Printer or monitor hardware fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alignment adjustment required between sensor and X-ray tube radiation
Sensor unit center must align with X-ray tube center with perpendicular crossing angle. Misalignment causes positioning errors in radiographic images.
Ranked causes
- 1. Sensor unit not centered horizontally or vertically relative to X-ray tube (common)
- 2. Stand or table base not level or perpendicular to X-ray beam (common)
Align cross line of X-ray tube with cross line of sensor unit
- Test point
- Cross line intersection at sensor center
- Expected
- Cross lines match exactly with [OK] alignment
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Firmware and PLD code verification
Verify firmware version, PLD code version, and hardware compatibility to ensure system operation and proper error handling
Ranked causes
- 1. Firmware or PLD code version mismatch or incompatibility (occasional)
- 2. Firmware and PLD code versions require verification via Telnet (occasional)
Verify firmware version in user mode
- Test point
- SYSTEM > SETUP MENU > VERSION INFORMATION
- Expected
- Firmware version, hardware version, and initialization code version display matches CXDI Software Combination List
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fixed ROI areas setup
Define and configure up to 5 fixed ROI areas (or 3 with AEC Display feature) on sensor with specific size, position, and pixel coordinates
Ranked causes
- 1. Fixed ROI areas not defined for sensor (common)
- 2. ROI pixel values not calculated (common)
Document required ROI area dimensions and position
- Test point
- Physical measurement of intended ROI
- Expected
- X, Y position (mm), W width (mm), H height (mm) recorded
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Photo timer adjustment radiation
Photo timer in imaging unit must be adjusted to set optimal X-ray exposure OFF time using specific Rex values
Ranked causes
- 1. Photo timer requires initial calibration (common)
- 2. Rex value must be set to specification during acrylic exposure (common)
Start system and enter calibration mode
- Test point
- CXDI application status
- Expected
- Normal radiography mode active, calibration completed
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM Connect Error message displayed
Error message 'DICOM Connect Error. Cannot connect to the target. Check network or port number setting. Retry?' appears when attempting to transmit images to printer or storage.
Ranked causes
- 1. Physical network connections not properly established (common)
- 2. Subnet mask settings incorrect (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image display timing delayed
Captured image display is delayed after X-ray exposure.
Ranked causes
- 1. RX_COM signal does not negate within timeout period (occasional)
- 2. RX_REQ not negated within specified time (occasional)
Measure time from RX_COM assertion to negation
- Test point
- RX_COM signal line
- Expected
- Should negate promptly after image read triggered
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect sensor identification or firmware version mismatch
System cannot identify sensor unit or firmware/PLD code versions do not match after sensor replacement
Ranked causes
- 1. Firmware or PLD code not installed or outdated (common)
- 2. Sensor unit not identified by system (occasional)
Check firmware and PLD code versions
- Test point
- System settings / firmware info
- Expected
- Firmware and PLD code versions match required versions
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System fails to start or operate after power unit replacement
System does not start or exposure/image capture/data transfer fails after power unit replacement
Ranked causes
- 1. Power unit jumper pins not set correctly for cable length (common)
- 2. PWB-40XRAY not properly installed or defective (occasional)
Identify power unit type and verify cable length
- Test point
- Power unit label
- Expected
- Confirm if 7m (white label, alphabet A) or 3m (pink label, alphabet B)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable brittleness or damage from bending
Imaging Unit Cable requires replacement due to insufficient bending strength in former design
Ranked causes
- 1. Original cable design has inadequate bending strength (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable clamp not properly secured caution
Sensor cable may become loose or disconnected from power box due to improper clamping.
Ranked causes
- 1. Cable clamp not installed or incorrectly positioned (common)
Verify cable clamp is installed on sensor cable
- Test point
- Sensor cable at power box connection
- Expected
- Cable clamp present and clamping shield portion
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable damage during bush tightening caution
Risk of cable damage when removing cables with excessive force or improper tightening technique during bush installation
Ranked causes
- 1. Improper tightening torque applied to bush (common)
Verify cable dimension and gap using gap gage before installation
- Test point
- Cable connection point at bush
- Expected
- Dimensions according to cable type: SENSOR cables 1.0 to 1.5mm, X-RAY I/F 0mm until stops
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable damage during tightening bush removal caution
Risk of cable damage when bushes are removed with excessive force during service operations
Ranked causes
- 1. Excessive removal force applied to tightening bush (common)
Verify cable dimensions with gap gage before installation
- Test point
- Cable connection point at tightening bush
- Expected
- Dimensions match specification per cable part number
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable disconnection during operation caution
Imaging unit cable comes off during use, requiring replacement with improved design featuring clamps
Ranked causes
- 1. Original cable design lacks disconnection prevention (common)
Verify cable is fully inserted into connector
- Test point
- Imaging unit cable connector
- Expected
- Cable seated with clamps engaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable needs replacement due to durability concerns
Imaging unit cable requires replacement to improve strength against bending and other mechanical stress
Ranked causes
- 1. Former cable lacks adequate bending strength (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable replacement required for imaging unit
Imaging unit cable needs replacement to improve bend strength and eliminate harmful substances
Ranked causes
- 1. Cable wear or age requiring replacement (unknown)
Verify cable type requirement
- Test point
- Imaging unit connector
- Expected
- 7m or 3m cable type matches unit from body number
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cables damaged during disconnection caution
Cables become damaged when removed with excessive force from power box connectors.
Ranked causes
- 1. Excessive disconnection force applied to cables (common)
Verify all cable lock nuts are loose before removal
- Test point
- Lock nuts on sensor, X-ray I/F, and remote cables
- Expected
- Lock nuts fully loosened, cable slides freely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Date and time setup
Set system date and time to local timezone; factory default is Japan standard time
Ranked causes
- 1. Date and time not set to local timezone (common)
Start CXDI application and open Administrator Setup
- Test point
- SYSTEM > SETUP MENU > ADMINISTRATOR SETUP menu path
- Expected
- ADMINISTRATOR SETUP MENU dialog appears
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Firmware installation
Write exposure code (firmware) into Flash ROM IC1 on PWB-50Di in imaging unit
Ranked causes
- 1. Firmware requires installation or update (occasional)
Verify CXDI system connection
- Test point
- Physical and network connection status
- Expected
- CXDI connected and powered on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Operation unit gamma correction
Procedure to adjust gamma of the operation unit display screen to match the appearance of the SMPTE test image on printer or external monitor.
Ranked causes
- 1. Operation unit display gamma not matched to printer or monitor output (common)
Verify linearity check procedure completed
- Test point
- System status and previous calibration records
- Expected
- Linearity check of transfer image density has been completed successfully
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PLD code installation
Write PLD code into PLD (IC11) on PWB-50Di in imaging unit
Ranked causes
- 1. PLD code requires installation or update (occasional)
Verify CXDI system connection
- Test point
- Physical and network connection status
- Expected
- CXDI connected and powered on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power unit jumper configuration
Jumper JP1 on PWB-40XRAY must be set correctly to match power unit cable length
Ranked causes
- 1. JP1 jumper not set for correct power unit cable length (common)
Identify power unit type from label
- Test point
- Power unit case label
- Expected
- White label with alphabet A (7m) OR pink label with alphabet B (3m)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PWB-50Di dipswitch configuration
PCB dipswitch settings must be configured after PWB-50Di replacement for proper operation
Ranked causes
- 1. Dipswitch settings not configured for installed PCB revision (common)
Verify PCB-50Di part number
- Test point
- PWB-50Di component marking
- Expected
- BG7-2999-040 or later revision
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SMPTE linearity and density verification
Procedure to verify that the SMPTE test image density response is linear across 32 grayscale steps from minimum to maximum density after installation or recalibration.
Ranked causes
- 1. Printer or monitor parameters not optimized for linear density output (common)
Set system to READY state in SMPTE exposure mode
- Test point
- Exposure screen mode selector
- Expected
- SMPTE mode selected and READY status displayed
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM communication error occurred
DICOM transmission failed with various possible error conditions
Ranked causes
- 1. Various DICOM transmission errors (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM communication warning from server
Server returns a warning response (0x18) after a successful transmission command
Ranked causes
- 1. Server warning response during successful DICOM transmission (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ethernet settings lost or normal code fails to start
LAN cannot connect with current Ethernet configuration; device requires factory reset to restore connectivity
Ranked causes
- 1. Normal code cannot execute or Ethernet configuration corrupted (unknown)
Verify initialization switch location and accessibility
- Test point
- Switch 1 on PWB-40 XRY module
- Expected
- Switch accessible and identifiable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exposed tabs on internal sensor unit damaged during disassembly patient safety
Tabs on the internal sensor unit are exposed when base cover is removed and may be damaged by contact
Ranked causes
- 1. Touching exposed tabs during disassembly or reassembly (common)
Verify clearance when removing or installing base cover
- Test point
- Tab area near grip
- Expected
- No contact with exposed tabs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware and PLD code upgrade for PWB-50Di
Specific firmware and PLD code versions required for PWB-50Di board; compatibility constraints with 50G A/D board
Ranked causes
- 1. Different firmware and PLD code requirements between PCB-50Di and PWB-50Di (common)
Verify PWB-50Di firmware version is 10200 or higher
- Test point
- Firmware file loaded
- Expected
- Firm50Di10200.hex or higher
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High noise in radiography detector output
Anti-noise stability improvement required; noise tolerance issue reported in prior service manual report CXDI-50G 05-013
Ranked causes
- 1. Insufficient grounding connection strength (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ICMP Echo Reply packet error detection
Tool detects malformed or unreachable ICMP Echo Reply packets during test sequence
Ranked causes
- 1. Destination unreachable or malformed reply packet (rare)
Monitor for premature sequence termination
- Test point
- ICMP Echo Reply packets
- Expected
- All ICMP Echo Reply packets valid until test count reached
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image quality degradation after component replacement
System produces images with degraded quality after replacing sensor unit, PWB, or LED
Ranked causes
- 1. Component installation incomplete or settings not configured (common)
Perform image quality test per Installation Manual section 5.2.6
- Test point
- System output images
- Expected
- Image quality within specification
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image quality verification for resolution and artifacts radiation
Final check of CXDI radiographic image quality to ensure no artifacts, shading, grid stripes, or other defects are present on output to monitor or printer.
Ranked causes
- 1. Calibration or self-test not performed (common)
Calibrate sensor system
- Test point
- SYSTEM menu CALIBRATION function
- Expected
- Calibration completes successfully
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Imaging unit not recognized by system
System fails to recognize imaging unit on startup, resulting in error. Communication with imaging unit fails when system powers on.
Ranked causes
- 1. Incorrect startup sequence not followed (common)
Verify power box main power is on before remote switch
- Test point
- Power box
- Expected
- Main power enabled first
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal sensor unit drops or shifts after base cover removal patient safety
Sensor unit becomes loose and drops or shifts down when base cover is removed or unit is tilted
Ranked causes
- 1. Base cover removed without proper support (common)
Verify unit position before removing base cover
- Test point
- Unit orientation
- Expected
- Unit in flat horizontal position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED board replacement compatibility
Technician must verify which LED board type is currently installed before replacement to avoid installing incorrect board type on CXDI-50G/50C units.
Ranked causes
- 1. Two different LED board types exist for CXDI-50G/50C (common)
Verify current LED board part number and serial number of unit
- Test point
- LED board PCB and unit serial number plate
- Expected
- BG7-2767-100 for older units or BM7-0076-000 for units 50G: 550001+ or 50C: 450001+
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED indicator not illuminating after PCB-50LED replacement
LED on PCB-50LED does not light up after module installation
Ranked causes
- 1. PCB-50LED not properly installed or defective (unknown)
Verify LED indicator illumination on PCB-50LED
- Test point
- PCB-50LED (BG7-2767)
- Expected
- LED lights
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line noise in reduced-size image transfer
Line noise observed when transferring reduced-size images, particularly in low dosage imaging of thick body areas
Ranked causes
- 1. Five chip jumpers present on PCB-50AD board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line noise in reduced-size image transfers
Line noise observed when transferring reduced-size images, particularly affecting low dosage diagnostic images of thick body areas at speed 400 or higher
Ranked causes
- 1. Five chip jumpers (R25, R27, R31, R32, R33) present on PCB-50AD (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line noise in reduced-size images
Line noise observed when transferring reduced-size images, particularly affecting low dosage areas in thick body imaging at speed 400 or higher
Ranked causes
- 1. Chip jumpers R25, R27, R31, R32, R33 installed on PCB-50AD (common)
Verify digital and analog ground separation
- Test point
- PCB-50AD digital GND and analog GND
- Expected
- Not short-circuited
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network communication reliability evaluation
Verify CXDI-50G network communication reliability using pingtest.exe tool to test packet transmission and reception
Ranked causes
- 1. Network communication failure or packet loss (unknown)
Perform ping test with default parameters
- Test point
- CXDI-50G IP address 192.168.100.11
- Expected
- Pingloss = 0/100,000 times packet loss count, 0 errors in 100,000 transmissions
- Tools
- pingtest.exe
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network communication reliability test
Evaluation of ping packet transmission and reception to CXDI-50G detector at IP 192.168.100.11 to verify zero packet loss
Ranked causes
- 1. Network communication failure or packet loss (unknown)
Run pingtest with default parameters
- Test point
- 192.168.100.11
- Expected
- Packets: Sent = 100000, Received = 100000, Lost = 0 (0% loss) packets, 0% loss
- Tools
- pingtest.exe
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network communication unreliable
CXDI-50G fails to maintain consistent network connectivity; packets are lost during transmission to the detector
Ranked causes
- 1. LAN card malfunction or configuration issue (unknown)
Run pingtest with default parameters
- Test point
- CXDI-50G IP address
- Expected
- Packets: Sent = 100000, Received = 100000, Lost = 0 (0% loss) packets, 0 packets lost
- Tools
- pingtest.exe
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Post-installation checklist verification patient safety
Comprehensive verification of all installation parameters including alignment, software versions, network settings, exposure conditions, image quality, and data integrity before clinical operation.
Ranked causes
- 1. Installation steps not completed (common)
Alignment with X-ray tube
- Test point
- Sensor center to X-ray tube center
- Expected
- Aligned horizontally and vertically with perpendicular crossing angle
12 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PWB-50Di board replacement procedure caution
Installation of replacement PWB-50Di board with spacer sheets and screws; firmware and PLD code compatibility requirements
Ranked causes
- 1. PWB-50Di board installation requiring proper screw torque and spacer placement (common)
Prepare required screws before starting replacement
- Test point
- Service technician inventory
- Expected
- XB1-2260406 screw prepared per technician
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recover corrupted system settings and firmware patient safety
System unable to start or operate normally due to corrupted settings or failed normal code boot
Ranked causes
- 1. Firmware or system settings corruption (rare)
Confirm CXDI system cannot start in normal mode
- Test point
- System boot sequence
- Expected
- System fails to reach normal operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reduced available imaging time
Available time window for X-ray imaging is reduced.
Ranked causes
- 1. RX_COM pre-delay timing exceeds specification (occasional)
Measure delay from RX_COM assertion to X-ray exposure start
- Test point
- RX_COM signal and X-ray output
- Expected
- Exposure starts within 0-100ms after RX_COM asserts
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RX_REQ signal retention too short
RX_REQ signal does not hold long enough for X-ray release signal (RX_COM) to be output.
Ranked causes
- 1. RX_REQ signal pulse width less than 300ms minimum (occasional)
Measure RX_REQ signal pulse width
- Test point
- RX_REQ signal line
- Expected
- Pulse width minimum 300ms
- Tools
- Oscilloscope
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Shading appears on captured images
Image shading or artifacts appear on radiographic images.
Ranked causes
- 1. RX_COM post-delay timing not met (occasional)
Measure delay from RX_COM negation to X-ray exposure stopping
- Test point
- RX_COM signal and X-ray output
- Expected
- Stop within 0-100ms after RX_COM negates
- Tools
- Oscilloscope, X-ray detector
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray I/F board damage from voltage mismatch patient safety
X-ray I/F board with 5V operating voltage may be damaged when multiple power boxes share same X-ray I/F channel and one board is replaced with 5V model while others operate at 12V
Ranked causes
- 1. Multiple power boxes sharing single X-ray I/F channel with mixed voltage boards (5V and 12V) (common)
Verify X-ray I/F channel assignment
- Test point
- Power box connection
- Expected
- Each power box connected to separate X-ray I/F channel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0x00000001 | Invalid parameter / DICOM connection error | Invalid parameter passed to print_stuff or send_image function | error |
0x00180012 | DICOM connection error - server not started or connection failed | Print server fails to start or cannot establish association with remote host | error |
0x10 | DICOM response error from server | Server responds with error after command request is sent | error |
0x18 | DICOM response warning (CCRTRANS_WRN_RESP) | Server returns warning response after command request | warning |
0x20 | DICOM printer status error | Printer returns error status during image transfer | error |
0x28 | DICOM printer status warning | Printer returns warning status during image transfer | warning |
Alert System Info Error (-6) | A/D board information has been updated | Occurs after entering Control PC serial number and restarting CXDI application | info |
Already initialized | Initialization function called after system already initialized | Software calls initialization function when system already in initialized state | error |
Association error - Other Messages | HIS or RIS server not responding to network association request | System attempts to connect to HIS or RIS during study data exchange | error |
Association error - TLS Cert Expiration of Validity | Network security certificate expired or invalid | System attempts secure TLS connection with RIS server during association | error |
Association error - TLS Doesn't Connect | RIS server not accepting network connection from CXDI system | System attempts to establish DICOM association with RIS via network | error |
Cannot generate Instance UID correctly | DICOM Instance Unique Identifier generation failed | System attempts to create new DICOM image with unique identifier | error |
CCRTRANS_ERR_DICOM_TRANSE | DICOM communication error | Various DICOM transmission failures | error |
CHMWM.DLL cannot load | DMW_PS2 V4 software component missing or corrupted | System attempts to initialize CHMWM.DLL library on startup | critical |
Destination unreachable | ICMP Echo Reply packet error received during ping test | Malformed or unreachable network response from CXDI-50G | error |
DICOM Connect Error | Cannot establish connection to target device at TCP/IP level | When attempting to transmit image to printer or storage device | error |
DICOM Lib initialize failure | MEDICOM DICOM library initialization failed | System attempts to initialize DICOM communications library during startup | critical |
DICOM Transfer Error | Communication failed at DICOM protocol level despite TCP/IP connectivity | When attempting image transfer after TCP/IP connection established | error |
Exam is not opened yet | Attempt to access study information before exam initialized | System attempts to read or close study data when exam not yet opened | warning |
Invalid parameter | Function called with null or invalid parameter | Software function receives null pointer or invalid parameter value | error |
Notification of study cannot be sent | MPPS server notification of study progress failed | System attempts to send examination progress information to MPPS server | warning |
Notification of study start cannot be sent | MPPS server notification of study initiation failed | System attempts to notify MPPS server of new study start | warning |
Not initialized yet | Function called before system initialization completed | Software calls non-initialization function before initialization function executed | error |
Other | DICOM communications error - various unspecified errors | Error return value other than documented codes | error |
Over 100 | Exposure count exceeds maximum within single flow | Number of X-ray exposures in current examination flow reaches or exceeds 101 | warning |
Patient Data is not found | Study matching search criteria could not be located in database | User executes find operation with specified search parameters | warning |
Pingloss > 0 | Network packet loss detected during communication test | When pingtest.exe receives fewer packets than sent to CXDI-50G at 192.168.100.11 | error |
Receive error | Message reception from RIS failed | System attempts to receive data or response from RIS server | error |
Send error | Message transmission to RIS failed | System attempts to send data or status message to RIS server | error |
Sensor Unit: Detect Error (-5100) | Sensor unit cannot be detected by control PC | Sensor serial number in PWB-50Di does not match sensor serial number stored in Control PC hard drive | error |
The number of received study information is over d | Study data count exceeds system configured maximum | Number of studies received from RIS exceeds parameter d | warning |
This function is not supported | Called function not available or supported in current system configuration | Software calls unsupported function | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI