Boston Scientific Galaxy: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Boston Scientific Galaxy, with 38 ranked causes, 10 diagnostic checks with test points and expected values, and 6 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
MD5 will not pullback
Display shows normal incremental pullback but MD5 fails to physically retract, or clutch disengaged or motor overload error messages appear
Ranked causes
- 1. MD5 not seated properly in sled (common)
- 2. Clutch lever physically stuck (occasional)
- 3. Sterile bag caught in sled gear (common)
- 4. Defective sled (occasional)
- 5. Catheter restriction from Touhy-Borst too tight (occasional)
- 6. Defective MD5 (occasional)
Verify MD5 seating in sled
- Test point
- MD5/sled interface both sides
- Expected
- MD5 fully seated and stable
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dark images
Ultrasound images appear darker than expected or with reduced brightness and clarity
Ranked causes
- 1. Catheter not fully seated in MD5 (common)
- 2. Occluded or defective catheter (common)
- 3. System software issue (occasional)
- 4. Defective MD5 (occasional)
- 5. MD5 started too soon after catheter insertion (common)
Verify catheter seating and ID stability
- Test point
- Catheter in MD5, catheter ID display
- Expected
- Catheter fully seated, catheter ID stable and displayed
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error messages following boot process or during application
System displays error messages such as Internal sub-system failure, DAQ, NI-DAQ, or VME Watch Dog/Dog Bark during startup or operation
Ranked causes
- 1. UPS battery not fully charged (common)
- 2. MD5 Lemo connector not fully plugged in (common)
- 3. UPS malfunction or VME circuit board connection issue (occasional)
- 4. Persistent hardware or firmware issue (rare)
Verify UPS fully charged and system rebooted
- Test point
- System power and UPS battery
- Expected
- UPS shows full charge, system boots without error messages
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No VCR playback video - blank or gray display
VCR playback shows no video on Galaxy display (appears blank or light gray) but audio is present
Ranked causes
- 1. VCR video input not set to S-VIDEO (common)
- 2. Video cable disconnected at Extron Scan Converter (occasional)
- 3. System software or hardware issue (occasional)
Check VCR Video Input switch setting
- Test point
- VCR Video Input switch behind hidden front door
- Expected
- Switch set to S-VIDEO
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video on display screen when powered up
Display screen is blank or shows no signal after system power-on while other peripherals power normally
Ranked causes
- 1. Display or video subsystem failure (rare)
- 2. System not receiving AC power (common)
- 3. Depleted UPS battery on brand new unit (occasional)
Verify AC power to system and check fuses
- Test point
- Fuses at back of system, AC outlet
- Expected
- Fuses intact, system has AC power
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printed images washed out and blurry
Printed ultrasound images lack contrast, clarity, or appear faded
Ranked causes
- 1. Paper installed incorrectly (common)
- 2. Old or damaged printer paper (common)
- 3. Wrong paper type (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message when selecting record: Internal sub-system not responding VCR
System displays error message Internal sub-system not responding VCR when user attempts to record
Ranked causes
- 1. Write-protect tab removed from videocassette (common)
- 2. Videocassette not installed (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
2D cut-plane jumps when selecting in LongView mode
When rotating 2D cut-plane line in LongView mode and clicking to select, 2D image jumps several frames and does not match selected location
Ranked causes
- 1. Normal behavior from different data display methods for LongView and 2D (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio and images out of sync during digital playback
Audio and video become progressively out of synchronization during digital playback, especially at end of record
Ranked causes
- 1. Inherent difference in how audio and image data are digitally recorded (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot read alphanumeric text in VCR playback
Small text and symbols on VCR playback are too blurry or degraded to read clearly
Ranked causes
- 1. Low quality of converted VCR signal (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CD/MO cannot be removed message
System displays message CD/MO cannot be removed until this run is active when attempting to eject media
Ranked causes
- 1. Current run is still active or message wording confusing (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DOS window appears with operation completed successfully message on boot
A DOS window appears during system boot with message The operation was completed successfully preventing normal startup
Ranked causes
- 1. File Manager folder exists without any files (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No image with correct catheter ID and no error messages
Display shows proper catheter identification and stability but no ultrasound image is produced
Ranked causes
- 1. Catheter or system malfunction (unknown)
Try another catheter or Simulator
- Test point
- Catheter or Simulator unit
- Expected
- Image appears with alternate catheter or Simulator
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Premature UPS battery failure
UPS battery does not hold charge or fails prematurely
Ranked causes
- 1. System unplugged while UPS battery not fully charged (common)
Verify UPS battery fully charged
- Test point
- UPS power backup test
- Expected
- UPS holds charge for at least 8 minutes (snooze function timeout)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print starts then alarm light comes on and stops or crooked output
Printer begins printing but stops when alarm light activates, or output emerges crooked
Ranked causes
- 1. Printer paper stuck on inside roller from jam (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not found error when printing
Error message Printer not found appears when user selects print function
Ranked causes
- 1. Mitsubishi printer alarm light on from snooze or power down during print (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer paper cuts automatically one sheet at a time
Printer cuts paper automatically instead of continuous feed, causing sheets to fall on floor
Ranked causes
- 1. Default printer configuration in Galaxy system (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VCR playback quality poor
Video recorded to VCR and played back shows reduced image quality compared to live display
Ranked causes
- 1. Inherent quality loss from computer to VCR to computer conversion (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VCR playback video not centered during playback
Video image displayed off-center or not properly framed during VCR playback
Ranked causes
- 1. Extron Scan Converter needs replacement or firmware update (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
CD/MO cannot be removed until this run is active | Media cannot be ejected because current run state prevents removal (message text is misleading) | User attempts to eject CD or MO media | info |
Internal sub-system failure, DAQ, NI-DAQ, VME Watch Dog / Do | Internal system subsystem failure detected by watchdog timer | During boot process or during application operation | critical |
Internal sub-system not responding - VCR | VCR subsystem is not responding to record commands | User selects record function | error |
MDU Clutch Disengaged - Pullback Stopped | MD5 clutch mechanism is stuck and will not allow pullback operation | Pullback initiated but clutch lever is physically stuck | error |
MDU Pullback Motor Overload - Pullback Stopped | Motor is overloaded and pullback operation has been halted | Pullback attempted but motor detected excessive load | error |
Printer not found | System cannot locate or communicate with connected printer | User selects print function while printer has alarm light on | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI