Boston Scientific AUTOLITH Touch: troubleshooting and error codes
The ServiceForge knowledge base holds 15 documented symptoms for the Boston Scientific AUTOLITH Touch, with 19 ranked causes, 16 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No power to unit
Unit does not respond to commands; LCD touch screen dark; no LED indicator illuminated
Ranked causes
- 1. Power cord not connected or disconnected (common)
- 2. Power cord damaged or deteriorated (occasional)
- 3. Wall outlet not powered (occasional)
Verify power cord is connected
- Test point
- Rear panel power connector
- Expected
- Power cord fully seated and secured
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe automatically flagged for replacement despite continued use patient safety
Touch screen displays REPLACE PROBE indicator or message even though probe appears functional
Ranked causes
- 1. Probe has reached pulse limit for safe operation (common)
- 2. Unit powered off and back on with same probe (occasional)
Observe pulse count on display
- Test point
- LCD touch screen display field
- Expected
- Total pulses delivered shown; compare to probe specification limits
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not discharge pulse when footswitch activated high voltage
No high-voltage discharge occurs when operator steps on foot-pedal despite unit powered on and probe connected
Ranked causes
- 1. Extender cable or probe not properly connected (common)
- 2. Unit in standby mode (occasional)
Verify extender cable connection
- Test point
- Probe connector on front panel
- Expected
- Cable fully seated and engaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration recommended alert on startup
Unit displays 'Calibration Recommended' screen for 10 seconds at power-on after extended use
Ranked causes
- 1. Unit has delivered approximately 250,000 pulses or 5 years of service (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CALIBRATION RECOMMENDED screen appears after 250,000 pulses
Calibration recommended screen displays after 250,000 pulses have been delivered and remains visible for 10 seconds before auto-transitioning
Ranked causes
- 1. Routine maintenance interval reached (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CONNECT PROBE message displayed and unit will not start
After power-up, main control screen displays CONNECT PROBE message and will not allow operation
Ranked causes
- 1. Probe not connected or not designated for model (common)
Verify probe is connected to extender cable
- Test point
- Probe connector at extender cable
- Expected
- Probe firmly locked with audible click
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Endoscope damaged during procedure patient safety
Endoscope lens or distal end damaged; operator fired probe tip while inside endoscope
Ranked causes
- 1. Probe fired while tip inside endoscope instead of extended beyond distal lens (common)
Verify probe tip extension distance before firing
- Test point
- Distal tip of probe relative to endoscope distal lens
- Expected
- Probe tip at least 5mm from distal lens of endoscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Extender cable lost or missing
Extender cable not available; unit cannot be operated
Ranked causes
- 1. Extender cable misplaced, lost, or damaged (occasional)
Verify extender cable is present and accounted for
- Test point
- Equipment storage, surgical cart, sterilization area
- Expected
- Extender cable located and ready for use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect irrigating solution used during procedure patient safety
Non-saline irrigant introduced into endoscopic field; may affect shock wave efficacy or damage equipment
Ranked causes
- 1. Wrong irrigating solution selected or inadvertently used (common)
Verify irrigating solution identity before use
- Test point
- Irrigation fluid container and label
- Expected
- .9% normal physiological saline
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
INSPECT PROBE alert at 90% life patient safety
Alert symbol appears, audible beep (3 times), and INSPECT PROBE message displayed when probe reaches 90% of useful life
Ranked causes
- 1. Probe approaching end of useful life (common)
Observe probe for unusual or unexpected performance
- Test point
- EHL probe output and function
- Expected
- Normal pulse delivery with expected fragmentation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REPLACE PROBE message displayed, unit stops firing patient safety
Probe has reached 100% useful life, REPLACE PROBE message appears, audible alert sounds, and unit will not deliver EHL pulses
Ranked causes
- 1. Probe reached end of useful life (100%) (common)
Verify probe is no longer functional
- Test point
- EHL probe output
- Expected
- No pulses delivered when foot pedal pressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Report Fault Code screen appears on startup
Internal fault detected and displayed on screen with service contact information, unit will not proceed beyond fault code screen
Ranked causes
- 1. Internal circuit or system fault (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tissue damage or organ perforation during procedure patient safety
Patient experiences unexpected pain, bleeding, or perforation detected during or after lithotripsy
Ranked causes
- 1. Probe distal tip pressed against tissue during discharge (common)
Verify probe tip positioning
- Test point
- Distal tip position inside biliary duct or target location
- Expected
- Probe tip not in contact with tissue walls; suspended in fluid
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not deliver EHL pulses patient safety
Foot-pedal depression does not result in EHL pulses being discharged into saline or tissue
Ranked causes
- 1. Faulty EHL probe (common)
Substitute a new EHL probe and test discharge
- Test point
- EHL probe tip in 0.9% saline solution
- Expected
- 5 EHL pulses delivered as set
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
LED on front panel does not change from blue to green when power switch is turned on
Ranked causes
- 1. Fuse open or blown (common)
Verify power switch is in on position
- Test point
- Rear panel power switch (A)
- Expected
- Switch fully depressed to ON position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.