Bird VIP Gold: troubleshooting and error codes
The ServiceForge knowledge base holds 86 documented symptoms for the Bird VIP Gold, with 117 ranked causes, 221 diagnostic checks with test points and expected values, and 14 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
High Pressure alarm not accurate patient safety
High Pressure alarm does not appear to be accurate
Ranked causes
- 1. Pressure relief valve setting incorrect (common)
- 2. Exhalation valve body defective (occasional)
- 3. Patient circuit or valve body faulty (occasional)
- 4. Transducer not properly verified (occasional)
Check pressure relief valve setting
- Test point
- Pressure relief valve
- Expected
- Set per manual specifications
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low FIO2 (Fraction of Inspired Oxygen) patient safety
FIO2 reading is lower than expected
Ranked causes
- 1. Oxygen knob set incorrectly (common)
- 2. Oxygen analyzer accuracy questionable (occasional)
- 3. Blender requires overhaul (occasional)
- 4. Blender calibration incorrect (occasional)
Set oxygen knob
- Test point
- O2 percentage control knob
- Expected
- Set to desired FIO2 percentage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitored exhaled tidal volumes are high or low patient safety
Tidal volume reading does not match expected or patient delivered volume
Ranked causes
- 1. Flow sensor not properly installed or defective (common)
- 2. Water migration in sensor tubing (occasional)
- 3. Mucus or debris in sensor (occasional)
- 4. Patient circuit or endotracheal tube leak (common)
Verify flow sensor is properly installed
- Test point
- Flow sensor installation position
- Expected
- Flow sensor in correct patient circuit position
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressurized Circuit Test fails (UVT)
User Verification Test Pressurized Circuit Test fails
Ranked causes
- 1. Leaks in humidifier or patient breathing circuit (common)
- 2. Pressure Relief Valve not at full clockwise position (occasional)
- 3. Inspiratory Pressure Control or High Pressure Alarm not at maximum (occasional)
- 4. Transducers need calibration (occasional)
Inspect humidifier for leaks
- Test point
- Humidifier
- Expected
- No visible leaks
8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System will not hold PEEP patient safety
PEEP pressure does not remain stable at set level
Ranked causes
- 1. Patient breathing circuit connections leaking (common)
- 2. Exhalation valve diaphragm misaligned or damaged (common)
- 3. Internal tubing incorrectly connected (occasional)
- 4. Faulty exhalation valve body (occasional)
Check patient breathing circuit connections for leaks
- Test point
- All circuit connections
- Expected
- No leaks
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent hop alarm during power-on self-test (POST) patient safety
When power switch turned ON, ventilator sounds audible alarm, illuminates Vent hop indicator, and retains code number indicating failed POST test
Ranked causes
- 1. Mode select switch not set to mode (common)
- 2. Air and oxygen sources not turned on or not properly connected; gas pressures out of range (common)
- 3. AC/DC switch set incorrectly or AC line voltage set wrong (occasional)
- 4. DC power source not providing 12VDC or insufficient current capacity (occasional)
Verify mode select switch setting
- Test point
- Mode select switch
- Expected
- Set to desired operating mode
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender Alarm or Low Inlet Gas Pressure Alarm patient safety
Blender Alarm or Low Inlet Gas Pressure Alarm appears, with or without Vent hop alarm
Ranked causes
- 1. Inlet gas pressures out of range or pressure difference too high (common)
- 2. Blender calibration incorrect (occasional)
- 3. Blender requires overhaul (occasional)
Measure inlet gas pressure from air source
- Test point
- Air inlet pressure port
- Expected
- 35-70 PSlG (2.4-4.8 bar), 35-70 PSlG
- Tools
- Pressure gauge
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit Leak Test fails (UVT)
User Verification Test Circuit Leak Test fails
Ranked causes
- 1. Leaks in patient breathing circuit or humidifier (common)
- 2. Pressure Relief Valve needs adjustment or replacement (occasional)
- 3. Exhalation valve, Safety Valve Assembly, or Jet pump components defective (occasional)
Check patient breathing circuit for leaks
- Test point
- Patient circuit tubing and connections
- Expected
- No leaks; circuit intact
10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manometer needle not moving during breaths patient safety
Manometer fails to show needle movement when ventilator delivers breaths
Ranked causes
- 1. Occluded or disconnected proximal airway line (common)
- 2. Faulty transducer or disconnected manometer tubing (occasional)
- 3. Transducer verification failure (occasional)
Inspect proximal airway line for occlusion or disconnection
- Test point
- Proximal airway line
- Expected
- Line clear and connected
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto cycling during assisted ventilation patient safety
Ventilator cycling repeatedly without patient effort or command
Ranked causes
- 1. Patient breathing circuit leaks exceed assist sensitivity setting (common)
- 2. Assist sensitivity setting too high (common)
Check patient breathing circuit connections for leaks
- Test point
- All circuit connections
- Expected
- No leaks or leaks below assist sensitivity setting
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit Fault Alarm patient safety
Circuit Fault Alarm appears
Ranked causes
- 1. Passages in exhalation valve body occluded (common)
- 2. Inspiratory/Expiratory limb of patient breathing circuit occluded (common)
Inspect exhalation valve body passages
- Test point
- Exhalation valve body interior
- Expected
- Passages clear; no blockage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low peak pressure alarm patient safety
Alarm sounds indicating peak airway pressure is below set threshold
Ranked causes
- 1. Patient disconnected or circuit leak (common)
- 2. Water present in breathing circuit (occasional)
Confirm patient breathing circuit connection
- Test point
- Patient circuit connection
- Expected
- Patient circuit connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display will not scan monitored parameters patient safety
Monitor display does not scan or update monitored parameters
Ranked causes
- 1. Select button not activated for scanning (common)
- 2. SW504 switch 5 not in ON position (common)
Press and hold Select button for three seconds
- Test point
- Select button
- Expected
- Automatic scanning starts
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Peak Flow does not appear accurate
Peak Flow reading does not appear to be accurate
Ranked causes
- 1. Tidal Volume range incorrect for altitude; pressure setting wrong (common)
- 2. Transducers require calibration (occasional)
Run Performance Test for tidal volume verification
- Test point
- Ventilator controls
- Expected
- Tidal Volume range correct for altitude
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Power switch turned ON but unit does not power up; ON indicator does not illuminate
Ranked causes
- 1. No power at AC outlet, AC/DC select switch in wrong position, or AC voltage setting incorrect (common)
- 2. Power Supply PCB defective; incorrect DC voltages (occasional)
Verify AC outlet has power
- Test point
- AC outlet
- Expected
- Power present
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on (external DC source)
Ventilator attached to external DC power source but does not power up
Ranked causes
- 1. Battery not charged or AC/DC select switch in AC position (common)
- 2. Power Supply PCB defective (occasional)
Test with known good power source
- Test point
- AC outlet or battery
- Expected
- Unit powers up with good source
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inlet pressure alarm not triggering at correct threshold caution
Blender alarm does not activate at 30 ±2.0 PSIG when air inlet pressure reduced, or alarm does not reset at or below 44 PSIG when pressure restored
Ranked causes
- 1. Blender pressure alarm switch malfunction or miscalibration (common)
Blender alarm activation during air pressure reduction
- Test point
- air inlet pressure gauge, alarm indicator
- Expected
- Alarm activates at 30 ±2.0 PSIG, ±2.0
- Tools
- air supply regulator, pressure gauge, visual/audible alarm
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Silence button does not silence alarm for correct duration
Elapsed time between Alarm Silence button activation and alarm reinstatement not 60 ±5 seconds
Ranked causes
- 1. Alarm Silence duration timer circuit malfunction (common)
Alarm Silence duration
- Test point
- stopwatch, alarm indicators
- Expected
- 60 ±5 seconds between button activation and alarm reinstatement, ±5
- Tools
- stopwatch, Alarm Silence button, alarm indicators
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient Pressure Test fails (UVT)
User Verification Test Ambient Pressure Test fails
Ranked causes
- 1. Internal transducers reading flow through patient breathing circuit (common)
Disconnect patient breathing circuit
- Test point
- Patient circuit inlet
- Expected
- Circuit fully disconnected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apnea alarm does not activate within timeout patient safety
Audible and visual alarm do not activate within 20 ±2 seconds after activating Manual Breath button with Rate set to zero
Ranked causes
- 1. Apnea alarm timing circuit malfunction or disabled (common)
Apnea alarm activation time
- Test point
- stopwatch, alarm indicators (audible and visual)
- Expected
- 20 ±2 seconds from Manual Breath activation to alarm onset, ±2
- Tools
- stopwatch, manual breath button, alarm indicators
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm too loud, too soft, or absent (UVT) caution
Audible alarm is too loud, too low, or absent during Audible Alarm Test
Ranked causes
- 1. Alarm volume control setting incorrect; alarm silenced (common)
Check Silence indicator status
- Test point
- Silence indicator LED
- Expected
- Silence indicator NOT illuminated
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breath rate inaccurate
Breath interval does not measure 6.0 ±0.2 seconds or breath rate not 10.0 ±0.3 breaths per minute
Ranked causes
- 1. Breath rate timing circuit malfunction or calibration drift (common)
Measure breath interval at Rate setting of 10
- Test point
- manometer, stopwatch
- Expected
- 6.0 ±0.2 seconds per breath, ±0.2
- Tools
- stopwatch, manometer, test lung
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Circuit Leak Test
Verify patient circuit leak rate using data from pressurized circuit test
Ranked causes
- 1. Patient circuit leak excessive or leak test failure (unknown)
Initiate leak test after PRS PASS
- Test point
- Monitor display
- Expected
- Leak test begins; display shows LEAKPASS if passed or HIGHFLOW/LEAKFAIL if failed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation Pressure Transducer Span
Exhalation pressure reading out of specification; requires span adjustment via potentiometer R342
Ranked causes
- 1. Exhalation pressure transducer span drift (unknown)
Press Select button to display exhalation pressure at 50 cmH2O
- Test point
- Monitor display
- Expected
- X 50.00 ? 1 .OO cmH2O, ±1.00 cmH2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation pressure transducer zero
Exhalation pressure transducer must read 0.00 ± 0.25 cmH₂O at zero reference; adjusted via R349 potentiometer
Ranked causes
- 1. Exhalation transducer out of zero specification (common)
Check exhalation pressure transducer zero reading
- Test point
- Monitor display exhalation pressure (X)
- Expected
- 0.00 ± 1.00 cmH₂O, ±1.00
- Tools
- Ventilator monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation purge flow
Exhalation purge flow must be 0.03 to 0.08 LPM at exhalation pressure port
Ranked causes
- 1. Exhalation purge flow out of specification (common)
Measure exhalation purge flow rate
- Test point
- Exhalation pressure port (right hole)
- Expected
- 0.03 to 0.08 LPM, 0.03–0.08
- Tools
- Flow tube
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation Valve current
Verify exhalation valve test circuit current reading on DMM during XVLV TST mode
Ranked causes
- 1. Exhalation valve current out of specification (unknown)
Plug Exhalation Valve into current measuring test fixture and connect DMM
- Test point
- Exhalation Valve test fixture
- Expected
- DMM set to 200 mA range
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow Valve Reference Voltage
Verify flow valve reference voltage at test points TP 1003, TP 1004, and TP 1005 during FLOW CAL mode
Ranked causes
- 1. Flow valve reference voltage out of specification (unknown)
Initiate flow valve test mode
- Test point
- Monitor display
- Expected
- Display reads FLOW CAL after pressing Manual Breath twice
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Inspiratory Pressure Verification patient safety
Verify peak inspiratory pressure (PIP) delivery at 30 cmH2O during infant flow volume test
Ranked causes
- 1. Peak inspiratory pressure out of specification (unknown)
Position ventilator on stand at 7 degree angle
- Test point
- Ventilator placement
- Expected
- Ventilator on ventilator stand or tilted back 7 degrees
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Jet Pump Flow and Pressure caution
Verify jet pump exhalation flow rate and back-pressure at exhalation valve inlet
Ranked causes
- 1. Jet pump flow or back-pressure out of specification (unknown)
Set ventilator to standard settings with PEEP/CPAP zero
- Test point
- Control panel settings
- Expected
- Ventilator configured per standard settings except PEEP/CPAP = 0
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Machine Pressure Transducer Span
Machine pressure reading out of specification; requires span adjustment via potentiometer R330
Ranked causes
- 1. Machine pressure transducer span drift (unknown)
Press Select button to display machine pressure at 50 cmH2O
- Test point
- Monitor display
- Expected
- M 50.00 k2.00 cmH20 cmH2O, ±2.00 cmH2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Machine pressure transducer zero
Machine pressure transducer must read 0.00 ± 0.25 cmH₂O at zero reference; adjusted via R337 potentiometer
Ranked causes
- 1. Machine transducer out of zero specification (common)
Check machine pressure transducer zero reading
- Test point
- Monitor display machine pressure (M)
- Expected
- 0.00 ± 2.00 cmH₂O, ±2.00
- Tools
- Ventilator monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressurized Circuit Test caution
Verify ventilator circuit pressurization to 55-65 cmH2O and transducer integrity at pressure
Ranked causes
- 1. Pressurized circuit test failure or transducer fault at pressure (unknown)
Block patient wye and press Select
- Test point
- Monitor display
- Expected
- Display shows PRSR TST BLOCK PATIENT WYE PRESS SELECT
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Proximal Pressure Transducer Span
Proximal pressure reading out of specification; requires span adjustment via potentiometer R318
Ranked causes
- 1. Proximal pressure transducer span drift (unknown)
Press Select button twice to display proximal pressure zero
- Test point
- Monitor display
- Expected
- P 0.00 + 1 .OO cmH2O
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Proximal pressure transducer zero
Proximal pressure transducer must read 0.00 ± 0.25 cmH₂O at zero reference; adjusted via R325 potentiometer
Ranked causes
- 1. Proximal transducer out of zero specification (common)
Check proximal pressure transducer zero reading
- Test point
- Monitor display proximal pressure (P)
- Expected
- 0.00 ± 1.00 cmH₂O, ±1.00
- Tools
- Ventilator monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Proximal purge pressure caution
Proximal purge pressure must be set to 60 ± 3.0 cmH₂O and purge flow to 0.03–0.08 LPM
Ranked causes
- 1. Purge valve pressure out of specification (common)
Verify proximal purge pressure via master manometer
- Test point
- Tube #3 (system manometer tube)
- Expected
- 60 ± 3.0 cmH₂O, ±3.0
- Tools
- Master manometer, tee fitting, proximal port blocked
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System pressure regulator
System pressure must be set to 25.0 PSIG for proper ventilator operation; verified via master manometer and DMM voltage measurement
Ranked causes
- 1. System pressure regulator out of specification (common)
Verify system pressure via master manometer
- Test point
- Tube assembly #24
- Expected
- 25.0 ± 0.5 PSIG, ±0.5
- Tools
- Master manometer, tee fitting
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Transducer Zero Test
Verify all three pressure transducers (proximal, exhalation, machine) under no-flow condition
Ranked causes
- 1. Pressure transducer zero calibration failure (unknown)
Attach test circuit to ventilator
- Test point
- PIN 10383 test circuit connection
- Expected
- Test circuit properly connected per Figure 5.11 without test lung
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Volume Monitor Flow Transducer Zero
Volume monitor flow transducer zero calibration without connected flow sensor or test harness
Ranked causes
- 1. Volume monitor flow transducer zero point drift (unknown)
Disconnect flow sensor and test harness tubes 14 & 17
- Test point
- Volume Monitor PCB
- Expected
- Tubes disconnected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Volume Monitor High Pressure Flow Transducer
Volume monitor high pressure (HP) flow transducer 2 cmH2O span calibration verification
Ranked causes
- 1. High pressure flow transducer span out of specification (unknown)
Display high pressure flow calibration menu
- Test point
- Monitor display
- Expected
- HP F-CAL
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Volume Monitor Low Pressure Flow Transducer
Volume monitor low pressure (LP) flow transducer 2 cmH2O span calibration verification
Ranked causes
- 1. Low pressure flow transducer span out of specification (unknown)
Display low pressure flow calibration menu
- Test point
- Monitor display
- Expected
- LP F-CAL
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Watchdog Timer
Verify watchdog timer functionality and millisecond counter tolerance during WDOG TST mode
Ranked causes
- 1. Watchdog timer out of tolerance (unknown)
Initiate watchdog test
- Test point
- Monitor display
- Expected
- Display reads WDOG TST
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit Fault alarm (proximal pressure line disconnect)
Audible and visual Circuit Fault alarm activates; Circuit Fault indicator light illuminates
Ranked causes
- 1. Proximal pressure line disconnected from ventilator (common)
Disconnect proximal pressure line and verify alarms activate
- Test point
- Proximal pressure port on front panel
- Expected
- Audible and visual Circuit Fault alarm activate, Circuit Fault indicator light illuminates
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit fault alarm active
Circuit fault alarm with High Pressure display flashing
Ranked causes
- 1. Unknown circuit fault condition (unknown)
Press Reset button and verify Circuit Fault alarm cancels
- Test point
- Control panel Reset button
- Expected
- Circuit Fault alarm cancels, High Pressure display stops flashing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control variable, push button, or Mode Select switch fails (UVT)
Variable control, push button switch, or Mode Select switch fail to perform correctly during Controls Test
Ranked causes
- 1. Poor solder joints on main board; control alignment or length incorrect (occasional)
Remove Main Display boards from panel
- Test point
- Main Display boards
- Expected
- Boards removed and accessible
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes appear in monitor display window patient safety
Monitor displays error or alarm codes
Ranked causes
- 1. System fault condition detected (common)
Record displayed error codes
- Test point
- Monitor display window
- Expected
- Specific 4-digit or alphanumeric code
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation valve blockage alarm active patient safety
Audible alarm sounds indicating exhalation valve outlet is blocked
Ranked causes
- 1. Exhalation valve body outlet blocked (common)
Verify audible alarm cancels after removing blockage
- Test point
- Exhalation Valve Body outlet
- Expected
- Audible alarm cancels
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High Breath Rate alarm active caution
Audible and visual High Breath Rate alarm activates
Ranked causes
- 1. High Breath Rate control set below set rate (common)
Verify High Breath Rate alarm activates when control set below set rate
- Test point
- High Breath Rate control
- Expected
- Audible and visual High Breath Rate alarm activate
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High Pressure Alarm does not activate at correct threshold patient safety
High Pressure audible and visual alarms do not activate at 60.0 ±6.0 cmH₂O when High Pressure Alarm set to 60.0 cmH₂O and patient wye occluded
Ranked causes
- 1. High Pressure Alarm threshold comparator malfunction or miscalibration (common)
High Pressure Alarm activation threshold
- Test point
- system pressure gauge, alarm indicators
- Expected
- Audible and visual alarms activate at 60.0 ±6.0 cmH₂O, ±6.0
- Tools
- system pressure gauge, Manual Breath button, alarm indicators
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High Tidal Volume alarm active caution
Audible and visual High Tidal Volume alarm activates
Ranked causes
- 1. High Tidal Volume control set too low (common)
Verify High Tidal Volume alarm activates when control set to 300mL
- Test point
- High Tidal Volume control
- Expected
- Audible and visual High Tidal Volume alarm activate
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Indicator lamps or display segments fail to illuminate (UVT)
One or more indicator lamps or display segments fail to illuminate during Lamp Test; Vent hop or External DC visuals will not illuminate during test
Ranked causes
- 1. Ribbon cable not connected or loose; LEDs or seven-segment decoders not seated properly (common)
Verify ribbon cable PIN 15481 connection
- Test point
- Ribbon cable at Main Display
- Expected
- Cable connected and tight
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of gas supply alarm active patient safety
Audible and visual Low Inlet Gas alarms activate when air and oxygen supply removed
Ranked causes
- 1. Air and oxygen supply pressure removed (common)
Remove air and oxygen supply and verify alarms activate
- Test point
- Gas supply connections
- Expected
- Audible and visual Low Inlet Gas alarms activate
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Inlet Gas Pressure alarm active caution
Audible and visual Low Inlet Gas alarm activates when inlet pressure reduced
Ranked causes
- 1. Inlet gas pressure below threshold (30 PSIG) (common)
Set Pressure Support and Sensitivity parameters
- Test point
- Pressure Support control, Sensitivity control
- Expected
- Pressure Support set to 10 cmH20 or above, Sensitivity set to 1.0 cmH20
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Minute Volume alarm active caution
Audible and visual Low Minute Volume alarm activates
Ranked causes
- 1. Low Minute Volume control set too high (common)
Verify Low Minute Volume alarm activates when control set to 25.0L
- Test point
- Low Minute Volume control
- Expected
- Audible and visual Low Minute Volume alarm activate
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Peak Pressure Alarm does not activate correctly patient safety
Audible and visual alarms do not activate when Low Peak Pressure Limit set above observed peak pressure, or do not deactivate when set to OFF
Ranked causes
- 1. Low Peak Pressure Alarm comparator or control circuit malfunction (common)
Peak pressure observation during standard operation
- Test point
- system manometer
- Expected
- Peak pressure reading visible during inspiratory cycle
- Tools
- system manometer
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual Breath button does not cancel audible alarm
Audible alarm does not cancel after pressing Manual Breath button three times
Ranked causes
- 1. Audible alarm cancellation circuit or Manual Breath control logic malfunction (common)
Audible alarm cancellation with three Manual Breath activations
- Test point
- Manual Breath button, audible alarm
- Expected
- Audible alarm cancels after third Manual Breath button press
- Tools
- Manual Breath button, alarm indicator
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual breath does not deliver one breath patient safety
Pressing Manual breath button does not deliver a single breath to patient circuit
Ranked causes
- 1. Manual breath mechanism malfunction or control circuit failure (common)
Manual breath button response
- Test point
- Manual breath button, test lung pressure rise
- Expected
- One breath delivered upon button press
- Tools
- test lung, manometer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display breath rate (BPM) out of range
BPM LED lit, but monitor display does not read 10 ±2.0 breaths per minute
Ranked causes
- 1. Breath rate sensor or display calculation error (common)
Read breath rate (BPM) on monitor display
- Test point
- monitor display, BPM LED
- Expected
- 10 ±2.0 BPM, ±2.0
- Tools
- ventilator monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display I:E Ratio out of range
I:E Ratio LED lit, but monitor display does not read 1:6.5 to 1:9.0
Ranked causes
- 1. I:E ratio calculation circuit malfunction (common)
Read I:E Ratio on monitor display
- Test point
- monitor display, I:E Ratio LED
- Expected
- 1:6.5 to 1:9.0
- Tools
- ventilator monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display Inspiratory Time (Ti) out of range
Ti LED lit, but monitor display does not read 0.6 ±0.1 seconds
Ranked causes
- 1. Inspiratory time measurement or display circuit malfunction (common)
Read Inspiratory Time (Ti) on monitor display
- Test point
- monitor display, Ti LED
- Expected
- 0.6 ±0.1 seconds, ±0.1
- Tools
- ventilator monitor display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Monitor display Mean Airway Pressure (MAP) out of range
- Monitor display minute ventilation (VE) out of range
- Monitor display PEEP reading out of range
- Monitor display pressure reading out of range
- Monitor display tidal volume (VT) out of range
- Over Pressure Relief Valve does not function
- Oxygen concentration reading out of range at 100%
- Oxygen concentration reading out of range at 21%
- Oxygen concentration reading out of range at 30%
- Oxygen concentration reading out of range at 60%
- Oxygen concentration reading out of range at 90%
- Oxygen inlet pressure alarm not triggering at correct threshold
- Patient demand trigger does not activate at correct sensitivity
- PEEP/CPAP plus Pressure Support reading incorrect
- Reset button does not cancel visual alarm
- Safety Valve dump time exceeds specification
- Tidal volume out of range
- Vent hop alarm activates on power loss
- Verification: Exhalation Valve Drive Current
- Verification: Fiber Optic Link Integrity
- Verification: Open Potentiometer Detection Circuit
- Verification: Volume Monitor Event Code Status
- Verification: Volume Monitor Fiber Optic Loopback Test
- Verification: Volume Monitor Purge Function
- Verification: Volume Monitor Solenoid Leak Test
- Volume Monitor Solenoid Test fails (UVT)
26 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Circuit Fault | Circuit fault condition detected on ventilator | Exhalation valve outlet blockage or proximal pressure line disconnection | error |
CPU A executed unused memory space | CPU A executed code in unused memory area | CPU A instruction pointer jumped to unused memory | critical |
CPU B executed unused memory space | CPU B executed code in unused memory area | CPU B instruction pointer jumped to unused memory | critical |
CPU B interrupted CPU A with VENT INOP | CPU B interrupted CPU A causing ventilation inoperative condition | CPU B generated VENT INOP interrupt to CPU A | critical |
CPU C executed unused memory space | CPU C executed code in unused memory area | CPU C instruction pointer jumped to unused memory | critical |
Flow control valve did not home (multiple) | Flow control valve failed to return to home position for specified breath type | Stepper motor control failure during breath cycle initialization | error |
High Breath Rate | Measured breath rate exceeds High Breath Rate alarm threshold | Actual respiratory rate falls below High Breath Rate control setting | warning |
High Tidal Volume | Delivered tidal volume exceeds High Tidal Volume alarm threshold | Actual tidal volume falls below High Tidal Volume control setting | warning |
Low Inlet Gas | Inlet gas (air or oxygen) pressure falls below alarm threshold | Supply pressure reduced below 30 PSIG or air/oxygen supply removed | error |
Low Minute Volume | Delivered minute volume falls below Low Minute Volume alarm threshold | Actual minute volume exceeds Low Minute Volume control setting | warning |
Mode switch failed | Ventilator mode selector switch malfunction | Switch reading does not correspond to physical switch position | warning |
PROM configuration bytes do not match | PROM configuration data inconsistency detected | Configuration bytes read from PROM do not match expected values | error |
Vent hop | Ventilator hop-up alarm; power loss detected or brownout voltage condition | Electrical power removed or voltage falls below brownout threshold | critical |
Watchdog timer failed | Hardware watchdog timer circuit malfunction | Watchdog logic does not time out as expected during self-test | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.