Biodex AtomLab 500: troubleshooting and error codes
The ServiceForge knowledge base holds 86 documented symptoms for the Biodex AtomLab 500, with 124 ranked causes, 238 diagnostic checks with test points and expected values, and 16 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Isotope list management radiation
Configure isotope list, routine isotopes, button assignments, dial values, and sealed sources for dose calibrator measurement operations.
Ranked causes
- 1. Need to select current isotope from full list (common)
- 2. Need to configure routine isotopes list (common)
- 3. Need to assign isotopes to measurement screen buttons (common)
- 4. Need to add custom isotope to isotope list (occasional)
- 5. Need to remove custom isotope from list (occasional)
- 6. Need to change isotope dial values (common)
- 7. Need to set QA test acceptance variance (occasional)
- 8. Need to configure sealed sources (common)
Navigate to Select Current Isotope screen
- Test point
- Home > Utilities > Dose Calibrator Configuration > Isotope List
- Expected
- Isotope List screen displayed after password entry
7 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Y-90 dial setting determination
Establishing correct AtomLab 500 dial setting for Y-90 activity measurement in plastic syringes, starting from radiopharmacy reference dose and adjusting within acceptable tolerance range.
Ranked causes
- 1. Initial dial setting selection and constancy verification (common)
- 2. Dial value within acceptable range (common)
- 3. Dial value outside acceptable range (occasional)
- 4. Acceptability of calculated error sum (common)
Obtain decay-corrected Y-90 activity from radiopharmacy
- Test point
- Y-90 dose vial or syringe label
- Expected
- Activity value in mCi or other stated unit
10 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System will not power up
Dose calibrator does not boot; no display response when power switch activated.
Ranked causes
- 1. ON/OFF switch not turned ON (common)
- 2. Power cord not plugged in (common)
- 3. Power supply issue or internal reset needed (occasional)
- 4. Incorrect outlet voltage (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power up
Atomlab dose calibrator fails to start; display does not boot.
Ranked causes
- 1. On/Off switch not activated (common)
- 2. Power cord not connected to display (common)
- 3. Power cord not seated in power pack receptacle (common)
- 4. Incorrect outlet voltage (occasional)
Verify on/off switch position
- Test point
- Display on/off switch
- Expected
- Switch in ON position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Activity measurement fluctuations radiation
Display shows unstable or varying activity values after isotope insertion and calibration key selection
Ranked causes
- 1. Normal electrometer averaging at low activity levels (common)
- 2. High calibration value causing electrometer gain increase and noise amplification (common)
- 3. Mo-99 normal high initial fluctuations during averaging (common)
Observe activity display after isotope insertion and calibration selection
- Test point
- Display screen
- Expected
- Display stabilizes after short time with expected fluctuations per isotope
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dial value adjustment for isotope
Determining modified dial value for isotope using known activity sample; accounts for configuration, container, and material variations
Ranked causes
- 1. Need to adjust dial value for different sample configurations (common)
- 2. Variation in wall thickness and material affects Beta sample activity (common)
- 3. F-18 dial setting modification for Ge-68/F-18 source (occasional)
Access dial value adjustment screen
- Test point
- User interface
- Expected
- Dial Value Adjustment screen displayed (Figure 5.3)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose calibrator setup and environmental conditions radiation
Installation planning and environmental requirements for optimal measurement accuracy and radiation safety
Ranked causes
- 1. Improper work surface or location affecting measurement accuracy (common)
- 2. Inadequate detector well setup or liner missing (common)
- 3. Improper sample handling increasing radiation exposure (occasional)
Verify work surface stability and load capacity
- Test point
- Tabletop
- Expected
- Free of vibration, does not wobble, supports 300 pounds
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Wide window QA verification radiation
Verify wide window count rate, efficiency and ROI settings for Cs-137 isotope in dose calibrator
Ranked causes
- 1. Wide window calibration drift or misconfiguration (occasional)
- 2. ROI setting out of specification (occasional)
- 3. Wide window efficiency drift (occasional)
Measure wide window net count
- Test point
- Wide window counter
- Expected
- 40 DPM
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Environmental operating conditions compliance
Temperature and humidity within specified ranges for optimal dose calibrator performance
Ranked causes
- 1. Operating environment outside specified temperature range (occasional)
- 2. Operating environment outside specified humidity range (occasional)
- 3. High ambient radiation level (occasional)
Verify ambient temperature
- Test point
- Operating environment
- Expected
- 10 to 30°C
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement accuracy affected by container type or wall thickness
Isotope measurement accuracy depends on sample container material, volume, and wall thickness, particularly for low-energy photon emitters.
Ranked causes
- 1. Container material or thickness differs from NIST calibration standard (occasional)
- 2. Standardized container not used consistently (occasional)
- 3. Low-energy photon isotopes (I-125, I-123, Xe-133) affected by wall thickness (common)
Verify sample container matches NIST calibration standard
- Test point
- Container physical specification
- Expected
- 5 ml liquid in sealed borosilicate glass ampoule (0.6 mm wall thickness) or plastic syringe as per NIST reference
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Activity display fluctuations
Reading varies or oscillates at low activity levels after isotope insertion
Ranked causes
- 1. Normal stabilization period (common)
- 2. High calibration value noise amplification (occasional)
Monitor display for stabilization
- Test point
- Activity display reading
- Expected
- Reading stabilizes within a few seconds
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Atomlab 500 dose calibrator measurement verification radiation
Verify measurement accuracy and response time for radioactive isotope measurements across specified activity ranges
Ranked causes
- 1. Factory calibration verification for isotope measurement accuracy (common)
- 2. Response time verification across activity ranges (common)
Measure known Tc-99m source activity
- Test point
- Well detector
- Expected
- Within 3% or 0.3 µCi of actual activity, whichever is greater, +/-3%
- Tools
- Tc-99m calibration source, reference activity
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constancy radiation
Daily reproducibility check of dose calibrator measurements using long-lived reference source with decay correction
Ranked causes
- 1. Cs-137 reference source activity drift or contamination (common)
- 2. Dose calibrator measurement error (occasional)
Press Zero Background button to zero background radiation
- Test point
- Dose calibrator display
- Expected
- Display zeroes and background is automatically adjusted
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constancy test radiation
Daily reproducibility measurement of same source with decay correction using Cs-137 reference source
Ranked causes
- 1. Constancy measurement exceeds 5% of predicted activity (unknown)
- 2. Constancy result exceeds 10% of predicted activity (rare)
Press Zero Background button
- Test point
- Zero Background (Bkgnd) button
- Expected
- Display zeroes
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Determine dial value adjustment for isotope patient safety
Measure activity of known isotope sample to establish or modify dial value setting, especially for Beta isotopes in different configurations or Ge-68/F-18 sources
Ranked causes
- 1. Beta isotope samples vary by configuration or pharmacy source (common)
- 2. F-18 dial setting requires adjustment for NIST traceable Ge-68/F-18 source (occasional)
Select correct isotope
- Test point
- Isotope key selection
- Expected
- System recognizes selected isotope and applies corresponding dial value
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Geometry Test radiation
Verify that the dose calibrator correctly measures activity at different volumes as a dose is diluted, ensuring geometry independence (indicated activity does not change with volume or configuration).
Ranked causes
- 1. Need to verify volume-dependent measurement accuracy (common)
- 2. Need to establish baseline normalized volume (common)
Select isotope for test
- Test point
- Geometry Test screen
- Expected
- Defaults to first isotope on button list, usually Tc-99m
9 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Linearity radiation
Test proportionality of dose calibrator measurement results across intended activity range using Tc-99m decay
Ranked causes
- 1. Mo-99 contamination in Tc-99m sample affects low-activity measurements (occasional)
- 2. Dose calibrator nonlinearity error exceeds ±5 percent (rare)
Zero background radiation using Zero Background button
- Test point
- Dose calibrator display
- Expected
- Display zeroes
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Lineator Test radiation
Comparison of initial linearity factors with current factors to verify dose calibrator accuracy within ±5% tolerance. Performed at installation and quarterly thereafter.
Ranked causes
- 1. Linearity drift or degradation (unknown)
- 2. Linearity drift or degradation (>55 mCi range) (unknown)
Access Linearity Test function from QA menu
- Test point
- QA Dose Calibrator menu
- Expected
- Linearity Testing screen displayed
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communications error caution
Error code 7 displayed; detector offline, repeated communication attempts failed
Ranked causes
- 1. Bad detector or disconnected/faulty cable (common)
- 2. Defective detector (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Linearity test shows correction factor <0.90 or >1.10 radiation
Calibrator may be non-linear or contaminated; requires diagnostic testing and vendor contact
Ranked causes
- 1. Mo-99 contamination in Tc-99m sample (occasional)
- 2. Dose calibrator non-linearity or malfunction (occasional)
Review cf_m values for declining trend at low activity measurements after >48h decay period
- Test point
- Linearity form data at lowest M values
- Expected
- Stable or increasing cf_m; declining cf_m suggests Mo-99 contamination
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Accuracy test radiation
Determine dose calibrator's absolute error using NIST-traceable reference sources with different principal photon energies
Ranked causes
- 1. Deviation in dose calibrator measurement accuracy from NIST standard (common)
Zero dose calibrator background
- Test point
- Dose calibrator display
- Expected
- Display reads zero after pressing Zero Background button
- Tools
- Dose calibrator
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Attenuation Tube (Sleeve) Method for Linearity radiation
Alternative linearity test using lead sleeves or attenuators of various thicknesses to simulate activity decay
Ranked causes
- 1. Mo-99 contamination causes failure at low activity with sleeve method (occasional)
Perform Decay Method linearity test first to establish baseline
- Test point
- Tc-99m sample activity over decay period
- Expected
- cfM values 0.95-1.05 or acceptable per criteria
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Calicheck Linearity Test radiation
Periodic verification of dose calibrator measurement accuracy using Calicheck test vials with black and red tubes
Ranked causes
- 1. Calibration verification required at installation and quarterly intervals (common)
Access Calicheck Test function
- Test point
- QA Dose Calibrator menu
- Expected
- Linearity Testing screen displayed
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Co-60 reference dial value setting radiation
Establish baseline dial value for Co-60 isotope as reference standard
Ranked causes
- 1. Co-60 selected as reference isotope for detector calibration (common)
Confirm Co-60 reference dial value
- Test point
- Dial valueCo-60
- Expected
- 5.0
- Tools
- Calibration manual
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constancy verification radiation
Daily constancy check using long-lived reference source (Cs-137) with decay correction to verify reproducibility of dose calibrator measurements over time.
Ranked causes
- 1. Need to verify measurement reproducibility (common)
Touch Zero Background button
- Test point
- Display
- Expected
- Display zeroes and background is automatically adjusted
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Container correction factor determination radiation
Correction factor or corrected dial value may be needed when container wall thickness or material differs significantly from NIST standards used in dose calibrator calibration.
Ranked causes
- 1. Container configuration differs from NIST standards (occasional)
Assess container specification
- Test point
- Container material and wall thickness
- Expected
- Matches NIST standard configuration or documented difference recorded
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Container Dial Value (CDV) for unsealed sources radiation
Determine accurate CDV for isotope measurements in non-standard source containers (vial, syringe) to ensure displayed activity reflects true measurement
Ranked causes
- 1. Source container composition differs from calibration reference (plastic syringe with 1mm wall) (common)
Assay isotope source in standard plastic syringe
- Test point
- Supplied source dipper syringe support
- Expected
- Activity reading for standard container
- Tools
- AtomLab 500 dose calibrator, standard plastic syringe
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Container Dial Value determination radiation
Procedure to determine appropriate Dial Value (DV) for accurate measurement of unsealed sources when container type changes from standard plastic syringe.
Ranked causes
- 1. Container type differs from standard plastic syringe (occasional)
Assay isotope source in standard plastic syringe
- Test point
- Dose calibrator display
- Expected
- Activity reading recorded as baseline (Plastic Syringe Activity)
- Tools
- Dose calibrator, plastic syringe nominal 1mm wall thickness
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Container Dial Value for container with pre-loaded source radiation
Determine CDV when isotope source is already in the target container at start of calibration procedure
Ranked causes
- 1. CDV required for container type with isotope already present (occasional)
Assay container with isotope source material present
- Test point
- Supplied source dipper or appropriate holder
- Expected
- Activity reading for loaded container
- Tools
- AtomLab 500 dose calibrator, set DV to published value
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Decay activity calculation using calculator method radiation
Procedure to calculate decayed activity of a radioisotope between calibration and measurement times using a scientific calculator with Y^x function
Ranked causes
- 1. Need to determine decayed activity for radioisotope measurement (common)
Measure current activity of radioisotope
- Test point
- A₀
- Expected
- Current measured activity value
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector well liner installation radiation
Proper placement and verification of detector well liner prior to radioisotope measurements to ensure accurate dose calibration.
Ranked causes
- 1. Missing or improperly installed detector well liner (common)
Verify detector well liner is present
- Test point
- Detector well
- Expected
- Clear plastic liner installed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector well liner placement radiation
Proper placement and use of the detector well liner to protect the ionization detector from contamination during radioactive source measurement
Ranked causes
- 1. Detector contamination risk without liner (common)
Verify well liner is installed in detector
- Test point
- Detector well opening
- Expected
- Plastic liner present and seated in well
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector well source positioning radiation
Radioactive sources must be positioned correctly within the detector well for accurate activity measurement. Sources can be in vials (using bottom cup of sample holder) or syringes (using syringe holder ring). Vertical orientation required.
Ranked causes
- 1. Improper source positioning affects measurement accuracy (common)
Verify source holder type matches sample (vial cup or syringe ring)
- Test point
- Sample holder assembly
- Expected
- Vial in bottom cup or syringe in syringe holder ring
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Determine dial value for custom isotope radiation
User needs to establish a dial value (detector sensitivity factor) for an isotope not in Biodex's published table, using a sample with known activity.
Ranked causes
- 1. Custom isotope requires user-defined dial value (common)
Obtain isotope sample with known activity value
- Test point
- Sample documentation
- Expected
- Activity value provided by isotope producer, pharmacy, or independent measurement method with accuracy estimate
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Determine dial values for radioisotopes radiation
Procedure to establish calibration dial values for radioisotope detection based on detector sensitivity curves relative to Co-60 reference standard.
Ranked causes
- 1. Dial values must be calculated from detector sensitivity to photon energy and radioisotope emission intensity (common)
Verify detector sensitivity curve has been determined using NIST calibrated isotopes
- Test point
- Atomlab detector calibration
- Expected
- Sensitivity curve plotted against photon energy with measurements from Ga-67, Tc-99m, In-111, I-125, I-131, Tl-201, F-18, Co-57, Co-60, Ba-133, Cs-137
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Determine personal dial value for unmeasured isotope radiation
User needs to establish a dial value for an isotope without a published reference value by measuring a sample of known activity in the dose calibrator.
Ranked causes
- 1. No published dial value available for isotope (unknown)
Obtain sample with known activity value
- Test point
- Source
- Expected
- Known activity documented from isotope producer, pharmacy, or independent measurement
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dial value determination for radioisotope selection radiation
Procedure to calculate and set dial values for different radioisotopes in the dose calibrator detector, with Co-60 as reference
Ranked causes
- 1. Different radioisotopes require different detector sensitivity calibration values (common)
Obtain detector sensitivity curve by plotting sensitivity versus photon energy
- Test point
- Detector sensitivity Se
- Expected
- Sensitivity curve in nA/mCi as function of photon energy
- Tools
- Reference tabulated photon sensitivity data and radioisotope emission spectra
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dial Value Settings and Container Configuration radiation
Establish correct Dial Value for isotope measurement in sealed and unsealed source containers (syringes, vials)
Ranked causes
- 1. Container composition affects measurement accuracy for beta and low-energy gamma sources (common)
Assay isotope quantity in plastic syringe with nominal 1mm wall thickness
- Test point
- Dose calibrator chamber
- Expected
- Activity reading recorded as Plastic Syringe Activity₁
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Geometry independence test radiation
Verify that indicated activity does not change with volume or configuration of source material using syringe and vial tests with Tc-99m
Ranked causes
- 1. Volume-dependent measurement errors in dose calibrator (common)
Mix Tc-99m solution
- Test point
- Small vial
- Expected
- 2.0 ml solution with activity concentration between 1 and 10 mCi/ml
- Tools
- Syringe, vials
9 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Geometry test verification radiation
Repeat geometry test if significant volume correction results to verify measurements are within acceptable tolerance.
Ranked causes
- 1. Significant volume correction detected (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Isotope selection and response verification
Selecting isotope for measurement, verifying detector response time and activity display accuracy
Ranked causes
- 1. Isotope selection required before measurement (common)
Select isotope and place source in detector
- Test point
- Detector well
- Expected
- Activity level displayed immediately
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Lineator Initial Test (setup phase) radiation
Initial capture of linearity reference factors using Tc-99m vial and multiple tubes prior to performing comparison tests.
Ranked causes
- 1. Baseline linearity factors required for future testing (unknown)
Record reading for Tube 1 & 2
- Test point
- Lineator Tube 1 & 2 in detector
- Expected
- Live reading displayed and stable
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Passcode change
Change user access passcode in System Configuration
Ranked causes
- 1. User requires passcode update (common)
Highlight Passcode field
- Test point
- Passcode field
- Expected
- Field highlighted and keypad displayed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Source holder positioning radiation
Proper placement of radioactive sources in the detector well for accurate measurement within the 4-pi geometry
Ranked causes
- 1. Source positioning affects measurement accuracy (common)
Measure source position from bottom of detector well without liner
- Test point
- Source location in detector well
- Expected
- 5 to 15 cm from bottom, 5 to 15 cm
- Tools
- Measuring tape or ruler
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Source positioning for measurement radiation
Radioactive sources must be positioned correctly in the detector well to ensure accurate activity measurement. Sources are placed in sample holders (vials in bottom cup, syringes in holder ring) and oriented vertically 5 to 15 cm from detector well bottom.
Ranked causes
- 1. Improper source positioning affects measurement accuracy (common)
Measure source position distance from detector well bottom
- Test point
- Detector well opening
- Expected
- 5 to 15 cm from bottom of detector well (measured without well liner), +/- 0.5%
- Tools
- Ruler or measuring tape
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tone Volume adjustment
Set the volume of the audible tone output from the Dose Calibrator
Ranked causes
- 1. Tone volume requires adjustment per user preference (common)
Verify tone volume setting range
- Test point
- Tone Volume field
- Expected
- 0 – 10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Work surface setup radiation
Establishing proper physical location and environmental conditions for dose calibrator operation to ensure accurate measurements and safe handling of radioactive sources.
Ranked causes
- 1. Inadequate work surface specifications (common)
Verify work surface dimensions and capacity
- Test point
- Tabletop area
- Expected
- 30 in. deep, 48 in. wide, 36 in. high minimum; support 300 pounds
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Y-90 dial setting for vials radiation
Determining dial setting for Y-90 measurements using vial containers instead of plastic syringes
Ranked causes
- 1. Y-90 dial setting needs adjustment for vial geometry (occasional)
Prepare vial with known Y-90 activity
- Test point
- vial source
- Expected
- vial contains documented Y-90 activity
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Change selected isotope
Switch isotope selection during measurement; new isotope activity displays within one second
Ranked causes
- 1. User needs to measure different isotope (common)
Select new isotope from list
- Test point
- Isotope selection menu
- Expected
- New isotope highlighted and selectable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Changing selected isotope during operation
Switching isotope selection after measurement started; new activity value displays within one second
Ranked causes
- 1. Need to change isotope for different measurement (common)
Change isotope selection during operation
- Test point
- Isotope selection interface
- Expected
- New isotope activity value displayed within one second of selection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration: Display units
Select between Curies or Becquerels unit display
Ranked causes
- 1. User preference for unit display (common)
Touch unit selection radio box
- Test point
- Units selection
- Expected
- Curies or Becquerels selected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration: Screen brightness
Adjust display brightness from 0 to 10
Ranked causes
- 1. User preference for screen brightness (common)
Select brightness level
- Test point
- Brightness control
- Expected
- Value from 0 to 10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration: Screen saver timeout
Set time in minutes before screen saver activates
Ranked causes
- 1. User preference for screen saver (common)
Enter screen saver timeout value
- Test point
- Screen Saver field
- Expected
- Time in minutes entered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration: Tone volume
Adjust audio tone volume from 0 to 10
Ranked causes
- 1. User preference for tone volume (common)
Select tone volume level
- Test point
- Tone Volume control
- Expected
- Value from 0 to 10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector bias voltage anomaly patient safety
Error code 8 displayed; detector offline, bias voltage out of specification
Ranked causes
- 1. Bias voltage outside 325V–375V range (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector error status displayed
Error code and description shown on detector status screen with WARNING or FAIL indicator
Ranked causes
- 1. Detector fault condition detected (unknown)
Verify detector power and communication status
- Test point
- Display screen and detector connection
- Expected
- Error code with brief description displayed; WARNING if usable, FAIL if dropped
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector error status reported during operation
Display shows popup screen with error code and audio tone alert when detector reports error condition; code marked as WARNING (detector usable) or FAIL (detector dropped from selection)
Ranked causes
- 1. Detector communication error or severe fault state (unknown)
Verify detector power and connection status
- Test point
- Detector connection to display unit
- Expected
- Detector powered and connected, no communication error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector gas leak or gas low warning caution
Gas leak or gas low warning appears on detector status screen with 24-hour reminder implemented
Ranked causes
- 1. Detector gas leak or gas pressure below threshold (unknown)
Monitor 24-hour reminder reminder schedule
- Test point
- Display screen timestamp
- Expected
- Reminder displays at 8:00 am if warning is in effect and 4+ hours elapsed since first detection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector gas low, correctable patient safety
Warning code 5 displayed; detector remains usable but no linearity correction applied
Ranked causes
- 1. Argon gas loss exceeds 5% of factory value (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector gas too low to use patient safety
Error code 6 displayed; detector automatically taken offline, no measurements possible
Ranked causes
- 1. Argon gas loss exceeds 10% of factory value (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Detector gas warning triggered
- Detector leaking gas or gas low warning
- Detector status error during operation
- Detector well liner installation and maintenance
- EEPROM data corrupt
- Electrometer monitor voltage anomaly
- Electrometer needs recalibration
- Firmware checksum error
- Fluctuations in activity measurements
- Improve measurement accuracy for low-activity sources
- Initial Lineator factors replacement needed
- Lead shield backscatter affects sensitivity at 88–200 keV photon energy
- Maintenance: Dial value checks
- Maintenance: Dipper and well liner inspection
- Measurement range verification for Tc-99m
- Mo-99 display fluctuations
- No available detector error
- No available detector selected
- Overrange detector reading
- Power supply configuration and site preparation
- Request QA test record count
- Restore factory dial values for all isotopes
- Test completion and documentation
- Unexpected activity reading
- Unexpected activity readings
- Y-90 dial setting for vial measurements
26 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Between 0 and -5% (single source) | Measurement within acceptable negative tolerance | One reference source reads 0 to -5% error | info |
Between 0 and +5% (single source) | Measurement within acceptable positive tolerance | One reference source reads 0 to +5% error | info |
Both high within tolerance | Both sources measure high but within acceptable limits | Both sources exceed true activity within tolerance band | info |
Both high with tolerance | Both sources read high but within specified tolerance | Both sources measure above true activity but within acceptable range | warning |
Both skewed high | Both reference sources measure significantly above true activity | Both sources read high (positive error) | error |
Both skewed low | Both reference sources measure significantly below true activity | Both sources read low (negative error) | error |
Both sources between ±5% | Both reference sources measure within ±5% error | Average percent errors for both sources within ±5% | info |
Communication error | Loss of communication between detector and display unit | Detector not communicating or in severe fault state | critical |
Detector Gas low | Detector gas supply level depleted or below operating threshold | Gas level monitored by detector self-test | warning |
Detector leaking Gas | Detector gas container has a leak | Gas leak detected by detector self-test | warning |
Large difference, skewed high | Both sources measure high with large discrepancy between them | Both sources read high but with significant variation between readings | error |
Large difference, skewed low (one source > 5%, < 10%) | One source measures significantly low, other differs markedly | One source error exceeds 5% to 10%, other differs substantially | warning |
Large transition during initialization | Unusually large change in activity detected during background initialization or Zero Background procedure | Large activity change (nn.n pa) detected when initializing background or performing Zero Background | warning |
Measure Screen Not Active | The Measure screen must be on view for this command to work | Command issued when Measure screen is not displayed | error |
No Available Chamber | The dose calibrator detector cannot be found or is not available | Detector connection lost or not detected | error |
Power Supply Configuration | AtomLab 500 uses ICC Nexergy model ME60A1503-F02 modified power supply | System power requirements specification | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.