Bio-Med Devices Air/Oxygen Blender: troubleshooting and error codes
The ServiceForge knowledge base holds 6 documented symptoms for the Bio-Med Devices Air/Oxygen Blender, with 14 ranked causes, 11 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Blender is alarming patient safety
Alarm condition activated on the blender
Ranked causes
- 1. Air and oxygen source pressures differ by more than 20 PSI (common)
- 2. Alarm system out of calibration (occasional)
- 3. Dirty gas contaminating alarm system (occasional)
- 4. Pressure balance chamber not working properly (rare)
Verify source pressure differential
- Test point
- Air and oxygen source inlet pressures
- Expected
- Within 20 PSI of each other
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen analyzer reading does not agree with blender setting
Analyzer output does not match the oxygen concentration set on the blender
Ranked causes
- 1. Oxygen analyzer out of calibration (common)
- 2. Blender out of calibration (common)
- 3. Bleed orifice on bottom of blender restricted (occasional)
- 4. Air introduced after blender diluting concentration (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender does not alarm on supply loss patient safety
Alarm circuit fails to activate when air or oxygen supply pressure is lost
Ranked causes
- 1. Air supply pressure loss (common)
- 2. Oxygen supply pressure loss (common)
Verify air supply pressure present
- Test point
- Air inlet
- Expected
- Supply pressure available
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender output concentration incorrect patient safety
Oxygen concentration at blender outlet does not match set percentage within tolerance
Ranked causes
- 1. Blender needs calibration adjustment (common)
- 2. Analyzer not stabilized (occasional)
Allow analyzer to stabilize after setting blender concentration
- Test point
- Analyzer display
- Expected
- Steady reading maintained
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air/Oxygen Blender patient safety
Periodic preventive maintenance to ensure continued proper operation including performance testing, cleaning, and calibration
Ranked causes
- 1. Scheduled maintenance interval due (common)
Perform performance test prior to each use
- Test point
- Blender outlet
- Expected
- Correct gas mixture delivery
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas inlet filters clogged
Air or oxygen supply inlet filters become restricted or clogged, affecting gas flow
Ranked causes
- 1. Contaminated compressed gas source or environmental debris (occasional)
Inspect inlet filter elements
- Test point
- Air and oxygen supply inlet fittings
- Expected
- Clear, unobstructed filter elements
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI