Berchtold Chromophare D 650: troubleshooting and error codes
The ServiceForge knowledge base holds 56 documented symptoms for the Berchtold Chromophare D 650, with 66 ranked causes, 67 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Red LED indication on reverse polarity protection module high voltage
LED on reverse polarity protection module shows red instead of green, or no indication at all
Ranked causes
- 1. Wrong polarity (reversed positive and negative connections) (common)
- 2. Alternating current (A.C.) supplied instead of D.C. (occasional)
- 3. No voltage connected to input terminals (occasional)
Verify input cable polarity
- Test point
- terminals 2 and 3
- Expected
- black cable on terminal 2 (negative), red cable on terminal 3 (positive)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
A.C. voltage supplied to light electronics high voltage
Direct A.C. voltage applied to light intensity power circuit board (part no. 65604) destroys electronics
Ranked causes
- 1. A.C. voltage directly connected without filtering board (common)
- 2. Filtering board omitted when only D.C. available (rare)
Verify input voltage type before connection
- Test point
- Main supply input
- Expected
- D.C. voltage 23–26 V or A.C. voltage with filtering board installed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Glare or tissue damage from thermal radiation patient safety
Operator vision impaired or tissue heating/drying at surgical site; cover glass or filter system compromised
Ranked causes
- 1. Cover glass or filter system damaged or destroyed (common)
- 2. Improper distance from patient surface (occasional)
Inspect cover glass and filter system for damage
- Test point
- Light emission surface
- Expected
- Cover glass intact, filter system undamaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect ceiling installation or mounting patient safety
Light mounting to operating room ceiling does not follow BERCHTOLD procedures; risk of ceiling damage and light falling.
Ranked causes
- 1. Installation not performed per BERCHTOLD mounting instructions or by unauthorized personnel (unknown)
- 2. Electrical installation not per IEC 60364-7-10 or missing fuse protection and mains switch (unknown)
Verify light secured to ceiling using BERCHTOLD anchor plate and spacer block per mounting instructions
- Test point
- Ceiling mounting bolts and anchor plate
- Expected
- Secure mounting with no movement or play when gently pushed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red indication on reverse polarity protection LED caution
The LED (V3) on the reverse polarity protection module illuminates red
Ranked causes
- 1. Reverse polarity (input connections reversed) (common)
- 2. AC voltage applied instead of DC voltage (occasional)
Measure voltage polarity and type at input terminals
- Test point
- Terminal 1 to Terminal 2 (input)
- Expected
- 24–27 V D.C. with correct polarity (red positive, black negative)
- Tools
- DMM with polarity indication
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse polarity protection LED not illuminated high voltage
No indication on LED V3 of the reverse polarity protection module, suggesting no voltage is connected or module is non-functional.
Ranked causes
- 1. No voltage connected to input terminal (common)
- 2. Reverse polarity protection module malfunction (occasional)
Measure input voltage at terminal 1
- Test point
- Terminal 1 input
- Expected
- 24 – 27 V D.C., 24-27 V
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse polarity protection LED showing red indication high voltage
Red LED on reverse polarity protection module indicates wrong polarity or AC voltage applied instead of DC.
Ranked causes
- 1. Reversed input polarity (common)
- 2. AC voltage supplied instead of DC (occasional)
Check voltage polarity at input terminals
- Test point
- Terminal 2 (negative) and Terminal 3 (positive)
- Expected
- negative (black cable) at terminal 2, positive (red cable) at terminal 3
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse polarity protection LED shows no indication
LED on reverse polarity protection module is not lit, indicating no voltage or module failure
Ranked causes
- 1. No input voltage (common)
- 2. Reverse polarity protection module fault (occasional)
Measure input voltage at terminals 1 and 2
- Test point
- terminal 1 (positive), terminal 2 (negative)
- Expected
- 24–27 V D.C., +3 V
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse polarity protection LED shows red indication high voltage
Red LED on reverse polarity protection module indicates wrong polarity or AC voltage applied
Ranked causes
- 1. Input polarity reversed (common)
- 2. AC voltage applied instead of DC (occasional)
Verify input cable polarity at reverse polarity protection module
- Test point
- terminal 2 (negative), terminal 3 (positive)
- Expected
- Red cable at terminal 3, black cable at terminal 2
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 24 V DC power supply voltage high voltage
Verify terminal block and sliding contact voltages under load and no-load conditions
Ranked causes
- 1. Power supply out of specification (unknown)
Measure DC voltage at terminal block under load
- Test point
- Terminal block
- Expected
- 24 V, +1.5 / -0.3 V
- Tools
- DMM
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: DC alignment of halogen lamp voltage high voltage
Terminal voltage under load on ceiling tube exceeds 24.5 V DC, causing halogen lamp tension to exceed safe operating limit (22.0 V +/- 0.2 V), affecting lamp life and illumination.
Ranked causes
- 1. Terminal voltage exceeds 24.3 V DC due to power supply or circuit characteristics (common)
Switch on emergency supply and light
- Test point
- Ceiling terminal
- Expected
- Light activates with emergency supply
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EndoLite bulb voltage high voltage
Set maximum EndoLite bulb output voltage via adjustable resistor R21 on touch control board back side
Ranked causes
- 1. Initial calibration or voltage drift (unknown)
Measure EndoLite bulb voltage at R21 adjustment test point
- Test point
- TP on touch control board back side (Part no. 68823/68824/68825/68826)
- Expected
- max. 11.0 V A.C. und D.C. RMS, -1 V
- Tools
- DMM (AC/DC RMS capable)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: halogen bulb voltage high voltage
Set maximum halogen bulb output voltage via adjustable resistor R20 on touch control board back side
Ranked causes
- 1. Initial calibration or voltage drift (unknown)
Measure halogen bulb voltage at R20 adjustment test point
- Test point
- TP on touch control board back side (Part no. 68823/68824/68825/68826)
- Expected
- D 650: 22.0 V A.C. und D.C. RMS; D 650plus/D 530plus/D 530: 22.8 V A.C. und D.C. RMS, +0.3 V
- Tools
- DMM (AC/DC RMS capable)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Halogen lamp voltage alignment high voltage
Terminal voltage under load on ceiling tube exceeds 24.5 V D.C., causing halogen lamp tension to exceed safe 22.0 V setpoint, reducing lamp life and illumination stability
Ranked causes
- 1. Excessive terminal voltage on ceiling supply (common)
Measure terminal voltage on ceiling terminal under load with emergency supply on
- Test point
- Ceiling tube terminal
- Expected
- 24.0 V D.C., +0.3 V
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Main bulb voltage adjustment high voltage
Set maximum bulb voltage via adjustable resistor R 20
Ranked causes
- 1. Bulb voltage out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Spring arm height adjustment
Adjust spring arm to horizontal position suitable for low ceiling installation
Ranked causes
- 1. Spring arm requires positioning adjustment (occasional)
Remove protective cap from spring arm
- Test point
- spring arm cap
- Expected
- cap removed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Spring arm height position
Adjust the spring arm to a horizontal position suitable for low ceiling installations
Ranked causes
- 1. Spring arm requires height position adjustment for ceiling clearance (common)
Remove the cap from the spring arm assembly
- Test point
- Spring arm cap
- Expected
- Cap removed without damage
- Tools
- Standard hand tools
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Transformer secondary tap adjustment high voltage
Set transformer secondary tap to match emergency supply terminal voltage with reference value measured behind filtering board
Ranked causes
- 1. Transformer secondary tap requires adjustment to correct voltage output (common)
Measure terminal voltage on ceiling terminal
- Test point
- ceiling terminal
- Expected
- matches reference value (D.C.)
- Tools
- voltmeter, DMM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Weight compensation on spring arm caution
Light head drifts upward or moves downward on the spring-balanced support arm; requires adjustment of circular nut on acrobat arm.
Ranked causes
- 1. Spring arm tension misaligned (common)
Observe light head position during arm movement
- Test point
- Spring-balanced support arm
- Expected
- Light head remains level without drifting
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chromophare D 530plus/D 500plus/D 530/D 500 spring arm adjustment (classic cardanic) caution
Weight compensation adjustment procedure for D 530plus, D 500plus, D 530, and D 500 models with classic cardanic spring arm support.
Ranked causes
- 1. Spring arm requires weight compensation calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chromophare D 650plus/D 650 spring arm adjustment (classic cardanic) caution
Weight compensation adjustment procedure for D 650plus and D 650 models with classic cardanic spring arm support.
Ranked causes
- 1. Spring arm requires weight compensation calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collision damage to supporting arms or lamp housings patient safety
Supporting arms or lamp housings struck, dented, or parts knocked off; risk of debris falling into surgical area.
Ranked causes
- 1. Contact between arms and housings during installation or movement (unknown)
Visually inspect supporting arms and lamp housings for contact marks, dents or missing parts
- Test point
- All supporting arms and lamp housing exteriors
- Expected
- No dents, cracks, contact marks or loose parts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control unit malfunction or fault indicator illuminated caution
Red fault indicator on control unit illuminates, indicating fault in main circuit of control electronics; light operates at maximum intensity and cannot be controlled
Ranked causes
- 1. Main circuit control electronics fault (rare)
Observe red fault indicator on control unit and test control key functions
- Test point
- Control unit indicator and keys
- Expected
- No red fault indicator; all keys function normally and adjust light intensity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control unit not responding to key inputs
Individual function keys do not control light intensity or movement; light operates at maximum luminous intensity only.
Ranked causes
- 1. Control electronics main circuit fault or keypad malfunction (unknown)
Activate individual function keys and observe light response
- Test point
- Control unit keypad and main circuit
- Expected
- Each key controls corresponding function; light intensity variable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical system missing fuse protection or mains switch high voltage
On-site electrical installation lacks required fuse protection or all-pole mains disconnect switch.
Ranked causes
- 1. Electrical installation not completed per code or requirements omitted (unknown)
Verify fuse protection installed in electrical circuit
- Test point
- Circuit fuse or breaker location
- Expected
- Fuse installed with rating appropriate to light specification
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Halogen lamp replacement high voltage
Surgical light halogen lamp has failed or needs replacement; reserve lamp management procedure
Ranked causes
- 1. Halogen lamp failure or routine replacement (common)
Switch off lighting system
- Test point
- Main power switch
- Expected
- System powered off and cooled
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High terminal voltage with cable voltage drop
Terminal voltage under load exceeds 2.0 V DC above set-point, preventing alignment to maximum allowed lamp tension when light intensity power circuit board is in separate wall box.
Ranked causes
- 1. Excessive voltage drop in supply cable to separate wall-mounted power circuit board (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper light positioning
Inadequate or uneven illumination of surgical field due to incorrect light distance from patient
Ranked causes
- 1. Light positioned too close to patient surface (occasional)
Measure distance from light emission surface to patient surface
- Test point
- Light head to surgical field
- Expected
- Minimum 60 cm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inadequate or incorrect illumination of surgical field patient safety
Surgical area may not be properly illuminated for recognition and treatment of patient condition
Ranked causes
- 1. Light positioned too close to patient surface (occasional)
Measure distance from light emission surface to patient surface
- Test point
- Light emission surface to surgical field
- Expected
- Not less than 60 cm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lamp failure with automatic switchover to reserve caution
Red indicator lights up on lamp housing when main lamp fails and system switches to reserve lamp
Ranked causes
- 1. Main halogen bulb end of life or failure (common)
Observe red indicator on lamp housing
- Test point
- Lamp housing indicator
- Expected
- Red indicator lit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light assembly instability or falling objects in surgical area patient safety
Placing or hanging objects on lamp housing or arms may compromise fixing stability and cause objects to fall into surgical field
Ranked causes
- 1. Objects placed on or hung from lamp housing or arms (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light assembly overheating during operation caution
Covering lamp housing during operation prevents heat exchange and risks overheating the surgical light
Ranked causes
- 1. Lamp housing covered during operation (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light does not emit after power-on patient safety
Main lamp fails to produce light output when unit is switched on; in combination lights, additional lamp also does not emit.
Ranked causes
- 1. Main lamp failure or incorrect installation (unknown)
Verify light emission from main lamp after power-on
- Test point
- Main lamp housing
- Expected
- Light emitted from main lamp
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light emission lens damaged or broken patient safety
Lens of lamp housing is damaged or broken, risk of glass splinters falling into surgical area
Ranked causes
- 1. Impact or collision damage to lamp housing lens (occasional)
Inspect light emission lens for damage or cracks
- Test point
- Lens surface of lamp housing
- Expected
- Lens intact, no cracks or chips visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light failure during operation patient safety
Reserve lamp auto-switch or APU switchover activated; single light units have incomplete component redundancy
Ranked causes
- 1. Lamp failure or electronic monitoring system fault (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light failure during surgery patient safety
Surgical light stops functioning during patient treatment, loss of illumination in operating field
Ranked causes
- 1. Component failure (not all components duplicated in single light) (occasional)
Verify light configuration in operating room
- Test point
- Installed light units
- Expected
- Two or more light heads in combination for critical surgical operations
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light glare or eye injury patient safety
Operator or patient eye damage from high illumination intensity or reflected light beams
Ranked causes
- 1. Direct viewing of switched-on light or reflective objects in beam path (occasional)
Check for reflective objects in light beam
- Test point
- Operating field and light path
- Expected
- No reflective objects present in beam
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light head drifts upward
The light head moves upward unintentionally during use
Ranked causes
- 1. Cardan arm requires tension adjustment (common)
Locate the circular nut in the cardan arm adjustment
- Test point
- One of two windows on the cardan arm
- Expected
- Circular nut becomes visible
- Tools
- Phillips screwdriver, steel pin diameter 4 mm
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light head moves downward caution
Spring arm support does not maintain light head position; head drifts downward
Ranked causes
- 1. Weight compensation spring arm tension insufficient (common)
Observe light head position after manual repositioning
- Test point
- Light head
- Expected
- Head maintains position without drifting downward
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light housing stability compromised or objects fall into surgical area patient safety
Heavy objects placed on or hung from lamp housing; mechanism damaged; risk of contamination or patient injury
Ranked causes
- 1. Objects placed on lamp housing or arms (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light overheating caution
Excessive heat buildup in lamp housing, potential thermal shutdown or damage
Ranked causes
- 1. Lamp housing covered during operation (occasional)
Verify lamp housing is uncovered
- Test point
- Lamp housing exterior
- Expected
- No covering or obstruction present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main lamp failed caution
Red indicator lights up on lamp housing; system automatically switches to reserve lamp
Ranked causes
- 1. Halogen bulb end of life (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mechanical instability or falling objects patient safety
Light fixture becomes unstable or objects fall from arms/housing into surgical field, risk of contamination or injury
Ranked causes
- 1. Objects placed or hung on lamp housing or arms (occasional)
Inspect lamp housing and arms for objects
- Test point
- Entire light fixture
- Expected
- No objects on housing or hanging from arms
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Movement mechanism not operating smoothly caution
Mechanical movement of supporting arms or rotation mechanism restricted, stiff, or not functioning
Ranked causes
- 1. Mechanical friction, misalignment, or wear in movement mechanism (occasional)
Manually operate all movement axes (swivel and rotate) for smooth operation
- Test point
- Supporting arms and lamp housing pivots
- Expected
- Smooth, unrestricted movement in all directions without grinding or binding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No indication on reverse polarity protection LED
The LED (V3) on the reverse polarity protection module does not light up (neither green nor red)
Ranked causes
- 1. No voltage connected to input terminals 1-3 (common)
Measure voltage at input terminals
- Test point
- Terminal 1 to Terminal 2 (input)
- Expected
- 24–27 V D.C., 24–27 V
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage exceeds specification high voltage
Light body output voltage above maximum rated value
Ranked causes
- 1. Power circuit board malfunction or misconfiguration (unknown)
Measure output voltage at light connector
- Test point
- Pins 3 and 1 (or 4 and 2)
- Expected
- max. 22.6 V A.C./D.C., not to exceed 22.6 V
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red indicator light displayed caution
Control unit electronics fault indicated by red indicator light
Ranked causes
- 1. Fault in control unit electronics (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reserve lamp warning indicator not functioning caution
Red warning indicator does not illuminate when reserve lamp is activated, or reserve lamp fails to switch on when main lamp fails
Ranked causes
- 1. Main lamp failure with defective reserve lamp mechanism (rare)
Simulate main lamp failure and observe reserve lamp activation and indicator
- Test point
- Red warning indicator on lamp housing
- Expected
- Red indicator illuminates and reserve lamp swings into position when main lamp fails
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reserve lamp warning indicator not illuminating
Red warning indicator on lamp housing does not light when reserve lamp activates after main lamp failure.
Ranked causes
- 1. Reserve lamp indicator malfunction or electrical fault (unknown)
Observe reserve lamp warning indicator when main lamp fails or is removed
- Test point
- Red warning indicator on lamp housing
- Expected
- Red warning indicator illuminates
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse polarity protection LED showing green indication
Green LED is illuminated on the reverse polarity protection module, indicating correct polarity and functioning module.
Ranked causes
- 1. Normal operation (common)
Confirm LED status
- Test point
- Indication LED V3
- Expected
- green
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse polarity protection LED shows green indication
Green LED on reverse polarity protection module indicates correct polarity and module functioning
Ranked causes
- 1. Normal operation (common)
Reverse polarity protection module not functioning high voltage
Module does not indicate green LED despite correct polarity and D.C. voltage supplied
Ranked causes
- 1. Reverse polarity protection module failure (rare)
Verify all input and output cable connections are secure
- Test point
- terminals 1–6
- Expected
- all cables firmly seated, correct polarity confirmed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Risk of eye injury from bright light patient safety
Operator or staff may suffer glare or eye damage from looking directly into switched-on light or from reflective objects in the beam path
Ranked causes
- 1. Direct viewing of operating light or reflective objects in beam path (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Terminal voltage exceeds maximum safe level high voltage
Terminal voltage under load on ceiling tube higher than 2.0 V D.C. prevents halogen lamp voltage alignment to maximum allowed setpoint
Ranked causes
- 1. Cable voltage drop exceeds 2.0 V D.C. (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tissue damage from thermal radiation patient safety
Heat injury, tissue necrosis, drying of surgical wound tissue due to exposed infrared radiation
Ranked causes
- 1. Damaged or destroyed cover glass or filter system (occasional)
Inspect cover glass and filter system for damage
- Test point
- Cover glass and filter assembly
- Expected
- No cracks, discoloration, or deterioration visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tissue damage from thermal radiation exposure patient safety
Surgical area may be heated, dried out, or suffer necrosis if light cover glass or filter is damaged
Ranked causes
- 1. Damaged or destroyed cover glass or filter system (occasional)
Inspect cover glass and filter system for damage
- Test point
- Cover glass and filter system
- Expected
- No damage or destruction visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Fault | Control unit has a fault | LED indicator on control unit | error |
Fault LED (red) | Fault in the control unit electronics | Indicator light shows red | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.