BD Infant Flow SiPAP: troubleshooting and error codes
The ServiceForge knowledge base holds 54 documented symptoms for the BD Infant Flow SiPAP, with 63 ranked causes, 90 diagnostic checks with test points and expected values, and 68 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Blender whistle caution
Audible alarm from blender unit. Verify inlet gas pressures, check watertraps and filters.
Ranked causes
- 1. Watertrap blocked, full or leaking (common)
- 2. Filters blocked (common)
- 3. Loss of wall pressure or imbalance in gas supply (occasional)
Inspect watertrap for blockage or fullness
- Test point
- Blender inlet watertrap
- Expected
- Clear, not full or leaking
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not start or fails self-test
Screen remains black and warning bar stays on after power switch is set to ON.
Ranked causes
- 1. Program memory fault (occasional)
- 2. Power supply not connected (common)
- 3. Emergency battery voltage low (occasional)
Verify mains power connection
- Test point
- AC power inlet
- Expected
- Mains power present and secure
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow meter fault caution
No flow indications or flow cannot be adjusted. Check flow meter knobs for damage or occlusion or kinks in flow meter tubes.
Ranked causes
- 1. Flow meter knobs damaged or occluded (common)
- 2. Kinks or blockage in flow meter tubes (common)
Visually inspect flow meter knobs
- Test point
- Flow meter control knobs
- Expected
- Intact, freely movable, no obstruction
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor calibration error caution
Error code displayed. Oxygen sensor incorrectly calibrated, damaged or depleted.
Ranked causes
- 1. Oxygen sensor incorrectly calibrated, damaged or depleted (common)
- 2. Oxygen sensor end of life or malfunction (occasional)
Perform oxygen sensor calibration
- Test point
- Oxygen sensor calibration port
- Expected
- Calibration completes successfully
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test fails
Screen remains darkened and warning bar remains on during power-up check
Ranked causes
- 1. Power Supply not connected (common)
- 2. Battery voltage low (common)
Verify power supply connection
- Test point
- Power cord to AC outlet
- Expected
- Power cord connected to AC power
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure reading low during nCPAP operation patient safety
Measured pressure on screen is below 4-5 cmH2O during nCPAP mode when prongs are occluded
Ranked causes
- 1. Circuit leak (common)
- 2. Circuit blockage (common)
Verify flow meter setting and measured pressure with prongs occluded
- Test point
- nCPAP/Pres Low flow meter and pressure display
- Expected
- 5 ±1 cmH2O at 8-9 lpm (depending on circuit type)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails Power-On Check patient safety
Screen remains darkened and warning bar remains on after power-up; unit does not display Power-up screen
Ranked causes
- 1. Power supply not connected (common)
- 2. Battery voltage low (occasional)
Verify AC power connection
- Test point
- Power cord and AC outlet
- Expected
- Power cord connected to AC outlet; outlet powered
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apnea detected patient safety
Breath rate zero for duration exceeding breath rate timeout. Intermittent high level audible alarm, Warning Bar flashes, breath rate digital display flashes, breath rate icon flashes.
Ranked causes
- 1. Patient breathing ceased or not detected by trigger module (common)
Monitor breath rate display and patient breathing
- Test point
- Breath rate digital display and patient
- Expected
- Breath rate > 0 and patient breathing detected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fault caution
Intermittent high level audible alarm, Warning Bar flashes intermittently, battery status indicator flashes, screen displays flashing error code.
Ranked causes
- 1. Battery disconnected or failing to take or hold charge (common)
Check battery connection and status indicator
- Test point
- Battery status indicator and physical battery connection
- Expected
- Steady indicator, secure connection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fault alarm
Intermittent high level audible alarm, warning bar flashes intermittently, battery status indicator flashes, screen displays flashing fault code (E number).
Ranked causes
- 1. Battery disconnected or failing to take or hold charge (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
BiPhasic/BiPhasic Tr mode fails to operate as set patient safety
Intermittent high level audible alarm, Warning Bar flashes intermittently, nCPAP Pres high on time display flashes, screen displays flashing error code. Cannot be silenced.
Ranked causes
- 1. BiPhasic or BiPhasic Tr mode software/hardware failure (rare)
Note error code on screen
- Test point
- Screen error code display
- Expected
- Specific error code per Chapter 9
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
BiPhasic/BiPhasic Tr mode fails to operate as set patient safety
Intermittent high level audible alarm, warning bar flashes intermittently, NCPAP Pres high on time display flashes, screen displays flashing fault code (E number).
Ranked causes
- 1. BiPhasic mode operation failure detected (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender whistle alarm caution
Audible alarm from blender.
Ranked causes
- 1. Water trap blocked, full or leaking; filters blocked; loss of wall pressure; imbalance of wall gas supply (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Alarms test setup and verification patient safety
Verify alarm system operation at specified unit settings (nCPAP mode, 8-9 lpm, 5 cmH2O, 30% O2)
Ranked causes
- 1. Alarm system calibration and reset required (unknown)
Unit mode configuration
- Test point
- Unit mode selector
- Expected
- nCPAP Mode
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blender oxygen delivery accuracy patient safety
Verify FiO2 output at blender control settings (21%, 30%, 60%, 90%, 100%) matches displayed values on SiPAP unit and external analyzer
Ranked causes
- 1. Blender oxygen concentration calibration required (unknown)
SiPAP FiO2 display at 21% blender setting
- Test point
- SiPAP display screen
- Expected
- 21% (±3%), ±3%
- Tools
- Calibrated oxygen analyzer
9 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen sensor patient safety
Initial O2 sensor calibration required before unit setup in patient mode. Performed during power-up startup sequence.
Ranked causes
- 1. Unit powered on without O2 sensor calibration (common)
Power switch set to ON position
- Test point
- Back panel power switch
- Expected
- Switch in ON position
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power-on check patient safety
Verify unit performs automatic self-checks on power-up including ROM, I/O, alarms, pressure sensor, and dump valve before clinical use
Ranked causes
- 1. Incomplete power-on self-test (unknown)
Flash ROM test
- Test point
- Internal memory
- Expected
- Test passes automatically during power-on
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Case/cover removal procedure caution
Step-by-step instructions for removing the plastic case/cover from the Infant Flow SiPAP unit to access internal components for service
Ranked causes
- 1. Case/cover must be removed for internal service (common)
Locate and remove eight screws on rear panel
- Test point
- Rear panel
- Expected
- All eight screws removed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPAP returns to wrong level after prong occlusion release patient safety
Pressure display shows 0-2 cmH2O when occlusion is removed, but does not return to 5 cmH2O when re-occluded or interface remains occluded
Ranked causes
- 1. Interface circuit occlusion not relieved (common)
Remove occlusion from prongs and measure pressure
- Test point
- Pressure display with nasal prongs open
- Expected
- 0-2 cmH2O
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Discharged battery
Battery voltage too low to power circuits; screen changes to Flat Battery screen
Ranked causes
- 1. Battery depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical fault high voltage
External power light does not match screen icon.
Ranked causes
- 1. External power indicator mismatch with display (rare)
Verify power indicator status
- Test point
- External power light and screen power icon
- Expected
- Both indicate same power status (on/off)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical fault caution
External power light does not match screen icon. Contact CareFusion SiPAP technical support.
Ranked causes
- 1. Power indicator light/software mismatch or internal electrical failure (rare)
Compare external power indicator light with screen display
- Test point
- Power indicator LED and screen power status icon
- Expected
- Light and icon state match (both on or both off)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External power disconnected caution
Intermittent high level audible alarm, Warning Bar flashes intermittently, battery status indicator and power indicator alternately flash.
Ranked causes
- 1. External power supply disconnected from unit (common)
Verify AC power connection
- Test point
- Power indicator and physical AC connector
- Expected
- Power indicator lit, AC power cord seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External power disconnected alarm
Intermittent high level audible alarm, warning bar flashes intermittently, battery status indicator and power indicator alternately flash.
Ranked causes
- 1. AC power supply disconnected from unit (common)
Verify power cord connection
- Test point
- AC power input connector
- Expected
- Power cord firmly connected and wall outlet powered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fault lockout patient safety
Software detects fault preventing correct operation; Fault Lockout screen displayed with fault code numbers
Ranked causes
- 1. Detectable software fault (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flowmeter fault caution
No flow indications or flow cannot be adjusted.
Ranked causes
- 1. Flowmeter malfunction (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High BiPhasic/BiPhasic Tr MAP pressure patient safety
High MAP pressure display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, digital high MAP pressure limit highlighted.
Ranked causes
- 1. Mean airway pressure exceeds 3 cmH2O above set point for 15 seconds (common)
Monitor high MAP pressure display
- Test point
- BiPhasic/BiPhasic Tr MAP pressure reading
- Expected
- ≤ 3 cmH2O above set point
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High BiPhasic/BiPhasic Tr pressure alarm patient safety
Intermittent high level audible alarm, warning bar flashes, high MAP pressure display flashes, digital high MAP pressure limit highlighted, alarm button flashes.
Ranked causes
- 1. Mean airway pressure exceeds 3 cmH2O above set point (unknown)
Verify MAP setting and actual pressure
- Test point
- High MAP pressure display
- Expected
- Within 3 cmH2O above set point
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High nCPAP pressure patient safety
Pressure display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, digital pressure high limit highlighted.
Ranked causes
- 1. Pressure exceeds 3 cmH2O above set point for 15 seconds (common)
Monitor pressure display
- Test point
- nCPAP pressure reading
- Expected
- ≤ 3 cmH2O above set point
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High oxygen concentration patient safety
FiO2 display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, digital O2 alarm high limit highlighted, Error Code appears.
Ranked causes
- 1. FiO2 exceeds 5% above set point for 15 seconds or exceeds 104% for 15 seconds (common)
Verify FiO2 level on display
- Test point
- FiO2 digital display
- Expected
- ≤ 5% above set point or ≤ 104%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak check: Gas connections caution
Test for leaks at all pipe connections, valves and fittings during pressurized gas flow
Ranked causes
- 1. Leaking tubing or connections (common)
Ensure whistle sounds during gas connection
- Test point
- Device audio output
- Expected
- Whistle audible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of Power
Loss of AC power and battery power together. Screen may go blank.
Ranked causes
- 1. Both AC power and battery are depleted or disconnected (rare)
Verify AC power connection and battery charge
- Test point
- AC power adapter and battery
- Expected
- AC power connected and/or battery above 11.10 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery charge caution
Battery status indicator changes from green to red. No audible alarm.
Ranked causes
- 1. Battery charge below 40% (common)
Check battery status indicator color
- Test point
- Battery status indicator on display
- Expected
- Green (charged above 40%)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery charge alarm
Battery status indicator changes from gray to red.
Ranked causes
- 1. Battery charge below 40% (common)
Observe battery status indicator color
- Test point
- Battery status indicator
- Expected
- Gray (normal) or red (low charge)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery voltage caution
Intermittent medium level audible alarm, Warning Bar flashes intermittently, battery status indicator flashes, unit starts controlled shut down.
Ranked causes
- 1. Battery voltage drops below 11.10 V for 5 seconds (common)
Verify battery voltage
- Test point
- Battery voltage measurement
- Expected
- ≥ 11.10 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery voltage alarm
Intermittent medium level audible alarm, warning bar flashes intermittently, battery status indicator flashes, unit starts controlled shutdown at set limit.
Ranked causes
- 1. Battery voltage drops below 10 V (common)
Monitor battery voltage
- Test point
- Battery status indicator and display
- Expected
- Above 10 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low breath rate alarm patient safety
Intermittent high level audible alarm, warning bar flashes, breath rate digital display flashes, breath rate icon flashes. In BiPhasic models with monitored NCPAP+Apnea, a single sigh is given. In BiPhasic Tr Mode backup sighs at set rate are given.
Ranked causes
- 1. Patient breath rate drops to zero for timeout period (occasional)
Observe breath rate display
- Test point
- Breath rate digital display and icon
- Expected
- Breath rate > 0 and above timeout threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low nCPAP pressure patient safety
Pressure display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, digital pressure low limit highlighted.
Ranked causes
- 1. Pressure drops 2 cmH2O below set point for 15 seconds or falls below 1 cmH2O (common)
Monitor pressure display
- Test point
- nCPAP pressure reading
- Expected
- ≥ 2 cmH2O below set point and ≥ 1 cmH2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen concentration patient safety
FiO2 display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, digital O2 alarm low limit highlighted.
Ranked causes
- 1. FiO2 drops 5% below set point for 15 seconds or falls below 20% for 15 seconds (common)
Verify FiO2 level on display
- Test point
- FiO2 digital display
- Expected
- ≥ 5% below set point and ≥ 20%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure alarm patient safety
Intermittent high level audible alarm, warning bar flashes, pressure display flashes, digital pressure low limit highlighted, alarm button flashes.
Ranked causes
- 1. Pressure below alarm threshold (unknown)
Verify actual patient pressure
- Test point
- Pressure display
- Expected
- Within 2 cmH2O above set point
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Minimum oxygen concentration too low patient safety
FiO2 display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, Error Code appears, digital O2 alarm low limit highlighted.
Ranked causes
- 1. FiO2 level below 18% (common)
Verify FiO2 level on display
- Test point
- FiO2 digital display
- Expected
- > 18%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 sensor calibration failure
O2 sensor fails to calibrate; red X displayed and unit alarms
Ranked causes
- 1. Internal oxygen sensor depleted or damaged (occasional)
Attempt O2 calibration procedure
- Test point
- Screen display
- Expected
- green Checkmark buttons at 21% and 100%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over pressure alarm patient safety
Intermittent high level audible alarm, warning bar flashes, O2 digital display flashes, digital O2 alarm low limit highlighted, alarm button flashes.
Ranked causes
- 1. Patient pressure exceeds 11 cmH2O in BiPhasic mode (unknown)
Measure patient pressure
- Test point
- Pressure display
- Expected
- Below 11 cmH2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over pressure condition patient safety
Pressure exceeds safe limits; software opens dump valve to release pressure
Ranked causes
- 1. Excessive system pressure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over pressure in BiPhasic or BiPhasic Tr mode patient safety
Pressure display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, digital O2 alarm low limit highlighted.
Ranked causes
- 1. Patient pressure exceeds 11 cmH2O in BiPhasic or BiPhasic Tr mode (common)
Monitor pressure display
- Test point
- Patient pressure reading
- Expected
- ≤ 11 cmH2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over pressure in nCPAP mode patient safety
Pressure display flashes, Warning Bar flashes, intermittent high level audible alarm, Alarm Button flashes, dump valve actuated for three seconds.
Ranked causes
- 1. Patient pressure exceeds 11 cmH2O in nCPAP mode (common)
Monitor pressure display
- Test point
- Patient pressure reading
- Expected
- ≤ 11 cmH2O
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen blender output not matching set value patient safety
Measured oxygen concentration differs from blender setting by more than 3%
Ranked causes
- 1. Oxygen sensor calibration drift (occasional)
Set oxygen blender and allow stabilization
- Test point
- %O2 control and measured FiO2 display
- Expected
- Measured value within 3% of set value, ±3%
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen cell calibration error patient safety
Difference between displayed and set oxygen value.
Ranked causes
- 1. Oxygen cell incorrectly calibrated, damaged or depleted (occasional)
Compare displayed O2 with set value
- Test point
- Oxygen display
- Expected
- Match between displayed and set oxygen percentage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord removal and installation high voltage
Procedure to safely disconnect and reconnect the power cord from the Infant Flow SiPAP unit
Ranked causes
- 1. Power cord requires replacement or maintenance (unknown)
Power cord physically disconnected from AC mains
- Test point
- AC power source
- Expected
- No power connection
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen displays in negative image during power-up
Screen shows inverted/negative image during power-up check sequence.
Ranked causes
- 1. Normal power-up check sequence (common)
Screen image is negative
- Test point
- Display screen
- Expected
- Negative image display during power-up checks
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software fault
Intermittent high level audible alarm, Warning Bar flashes intermittently, screen displays flashing error code.
Ranked causes
- 1. Software failure detected (rare)
Note error code displayed
- Test point
- Screen error code display
- Expected
- Specific error code per Chapter 9
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software fault alarm
Intermittent high level audible alarm, warning bar flashes intermittently, screen displays flashing fault code (E number).
Ranked causes
- 1. Software error detected during operation (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software not running with unit connected to power
Constant audible alarm, warning bar on, screen displays flashing fault code (E number).
Ranked causes
- 1. Software failed to start or load (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software not running with unit connected to power
Constant audible alarm, Warning Bar on, screen displays flashing error code.
Ranked causes
- 1. Software process failed to start on boot with power connected (rare)
Note error code on screen
- Test point
- Screen error code display
- Expected
- Specific error code per Chapter 9
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
- | Battery too discharged to operate LCD and valve circuits | Battery voltage below 6.5 V with no external power | critical |
- | Battery too discharged to operate LCD and valve circuits | Battery voltage drops below 6.5 V with no external power | critical |
- | Program memory checksum error | Software corruption during execution | critical |
- | Program memory checksum error | Software becomes corrupted during operation or storage | critical |
- | Battery voltage too low for valve operation but LCD operable | Battery voltage between 6.5 V and 10 V with no external power | critical |
Discharged battery | Battery voltage is too low to power the circuits | Battery charge drops below operating minimum | warning |
E10 | Non-volatile memory fault | Configuration or calibration data storage failure | error |
E10 | Non-volatile memory fault | Non-volatile memory failure | critical |
E11 | Calibration data lost | Loss of calibration data in memory | critical |
E11 | Calibration data lost | Sensor calibration data erased or unreadable from memory | error |
E12 | DIP settings mismatch or Patient Trigger board issue | DIP switch settings or PT PRESENT status differs from stored configuration | error |
E12 | Configuration DIP settings mismatch | DIP settings or PT PRESENT differ from non-volatile record | error |
E20 | Charged battery voltage too low | Battery voltage below 11 V under test load | warning |
E20 | Charged battery voltage too low | Battery voltage drops below 11 V under test load | warning |
E21 | External supply voltage too low to charge battery | AC supply voltage below 14 V when attempting to charge flat battery | error |
E21 | External supply voltage too low to charge battery | External supply voltage below 14 V | error |
E22 | Analog supply rails out of limits | Analog power supply voltage exceeds acceptable range | critical |
E22 | Analog supply rails out of limits | Power supply voltages for analog circuits exceed acceptable range | error |
E23 | Valve driver supply rails out of limits | Power supply voltages for valve driver circuits exceed acceptable range | error |
E23 | Valve driver supply rails out of limits | Valve driver power supply voltage exceeds acceptable range | critical |
E24 | Hardware safe-start watchdog disabled | Watchdog function failure preventing safe valve operation | critical |
E24 | Hardware safe-start watchdog disabled | Watchdog timer fails to enable valve control during startup | error |
E30 | Pressure sensor fault (ADC hits rail) | Pressure sensor analog-to-digital converter unable to convert signal | error |
E30 | Pressure sensor fault | Pressure sensor ADC hits rail indicating sensor failure | critical |
E31 | Zero valve not connected | Zero valve connection lost or not detected via sense line | critical |
E31 | Zero valve not connected (via sense) | Zero valve feedback signal not detected by pressure monitoring circuit | error |
E32 | Zero valve activation fault (via sense) | Zero valve feedback does not respond to activation command | error |
E32 | Zero valve activation fault | Zero valve fails to activate as detected via sense line | critical |
E33 | Unable to auto zero pressure sensor | Pressure sensor auto-zero function fails during startup | critical |
E33 | Unable to auto zero pressure sensor | Pressure sensor fails to reach zero calibration point | error |
E41 | Dump valve not connected (via sense) | Dump valve feedback signal not detected | warning |
E41 | Dump valve not connected | Dump valve connection lost or not detected via sense line | error |
E42 | Dump valve activation fault | Dump valve fails to activate as detected via sense line | error |
E42 | Dump valve activation fault (via sense) | Dump valve feedback does not respond to activation command | warning |
E50 | Oxygen sensor failure | Oxygen sensor malfunctions during normal operation | error |
E50 | Oxygen sensor fault | Oxygen sensor ADC hits rail indicating sensor failure | critical |
E51 | Oxygen sensor cannot be calibrated | User calibration fails with bad offset or high gain | critical |
E51 | Oxygen sensor failure variant | Oxygen sensor malfunction during operation | error |
E52 | Oxygen sensor failure variant | Oxygen sensor malfunction during operation | error |
E52 | Oxygen sensor calibrates but fuel cell worn out | O2 sensor calibration succeeds but shows low gain indicating fuel cell degradation | critical |
E53 | Oxygen sensor too noisy to calibrate | O2 sensor calibration timeout due to excessive noise | critical |
E53 | Oxygen sensor failure variant | Oxygen sensor malfunction during operation | error |
E54 | Oxygen calibration suspect (out of range) | O2 value measured outside normal range of 18-104% | error |
E54 | Oxygen sensor failure; high-elevation calibration issue | Oxygen sensor malfunction or high-altitude initialization (>6,600 feet or 2,000 meters) | error |
E54 | Oxygen calibration suspect | Oxygen reading measured outside valid range 18–104% | error |
E55 | Internal oxygen monitor disabled | User presses and holds alarm mute/reset button for 3 seconds, or oxygen sensor failure recovery | warning |
E61 | nCPAP Pressure high valve not connected | Pressure high valve feedback signal not detected | warning |
E61 | NCPAP Pres high valve not connected | NCPAP pressure high valve connection lost or not detected via sense | error |
E62 | NCPAP Pres high valve activation fault | NCPAP pressure high valve fails to activate as detected via sense | error |
E62 | nCPAP Pressure high valve activation fault | Pressure high valve feedback does not respond to activation | warning |
E70 | Patient Trigger module fault | PT module not responding or CAN bus communication failure | error |
E70 | PT module fault | Patient Trigger module failure (PTRDY or CAN bus failure) | warning |
E71 | No breath signal from PT module | No breath signal detected although CAN bus reports normal status | warning |
E71 | No breath signal from PT module | PT module not sending breath signal despite no apnea reported | error |
E72 | No trigger signal from PT module in BiPhasic Tr mode | PT module fails to send trigger signal in BiPhasic Trigger mode | error |
E72 | No trigger signal from PT module in BiPhasic Tr mode | PT module fails to provide trigger signal in BiPhasic Tr mode | warning |
E80 | Audible Alarm Module not functioning | Alarm output circuit fails to produce audible alert | warning |
E90 | Abnormal hardware, software or watchdog reset | Unplanned system reset detected during normal operation | warning |
E90 | Abnormal hardware, software or watchdog reset | Unexpected system reset during operation | critical |
E90* | Spurious software interrupt or hardware failure | Software interrupt, crystal failure, stack overflow/underflow, or CPU exception | critical |
E91 | Internal software error detected | Software integrity check fails | critical |
E91 | Internal software error detected | Software error detection routine identifies fault in firmware execution | error |
E99 | Unknown error detected | Error condition detected but error code generator unable to classify | error |
E99 | Unknown error detected | Unidentified error condition triggered | critical |
E (number) | Software fault, software not running, BiPhasic mode failure, or battery fault | Software error detected or fails to load, or BiPhasic mode cannot operate as set, or battery disconnected/failing | critical |
Fault Lockout | Software-detectable fault that prevents unit from operating correctly | Internal fault detected during operation | error |
Loss of Power | Complete loss of both AC power and battery power | AC power disconnected and battery depleted simultaneously | critical |
Oxygen cell calibration error | Oxygen cell is incorrectly calibrated, damaged, or depleted causing disparity between displayed and set O2 value | Automatic monitoring detects oxygen cell error | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI