FR FieldReady Academy
Infusion pumps · Flo-Gard family

Baxter Flo-Gard 6301: troubleshooting and error codes

The ServiceForge knowledge base holds 160 documented symptoms for the Baxter Flo-Gard 6301, with 293 ranked causes, 233 diagnostic checks with test points and expected values, and 95 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

AIR alarm continuously on patient safety

AIR alarm message remains on display continuously despite troubleshooting

Ranked causes

  1. 1. Air trapped in tubing (common)
  2. 2. Incorrect or improperly loaded administration set (common)
  3. 3. Tubing scratched or deformed (occasional)
  4. 4. Air sensor out of calibration (occasional)
  5. 5. Air sensor dirty (occasional)
  6. 6. Air sensor failure (rare)
  7. 7. Occlusion sensor failure (rare)
First check

Check for air in tubing

Test point
administration set tubing
Expected
no visible air bubbles

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Free flow occurs when door is closed patient safety

Solution flows past the pump mechanism when door is closed, indicating loss of occlusion

Ranked causes

  1. 1. Incorrect administration set type (occasional)
  2. 2. Tubing positioning or seating improper (occasional)
  3. 3. Temperature outside 60-100 F range (occasional)
  4. 4. Cracked door hinges (occasional)
  5. 5. Deformed coil springs on back plate (occasional)
  6. 6. Worn cams and pumping fingers (occasional)
  7. 7. Damaged or deformed door latch (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate infusion delivery patient safety

Actual delivered volume does not match programmed rate or volume to be infused

Ranked causes

  1. 1. Incorrect administration set type (occasional)
  2. 2. Tubing positioning or seating improper (occasional)
  3. 3. Temperature outside 60-100 F range (occasional)
  4. 4. Cracked door hinges (occasional)
  5. 5. Deformed coil springs on back plate (occasional)
  6. 6. Worn cams and pumping fingers (occasional)
  7. 7. Damaged or deformed door latch (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

OCCLUSION alarm occurs with pump door closed caution

OCCLUSION or UPSTREAM OCCLUSION alarm activates when pump door is closed and START key is pressed

Ranked causes

  1. 1. Tubing positioning or seating improper (common)
  2. 2. Obstructions upstream or downstream of pump (common)
  3. 3. Incorrect administration set type (occasional)
  4. 4. Temperature below 60 degrees F (occasional)
  5. 5. Solution spills or residue on door or baseplate (occasional)
  6. 6. Damaged door latch or hinges (occasional)
  7. 7. Sensor calibration required (unknown)
First check

Verify tubing positioning and seating in sensor

Test point
pump sensor and safety clamp
Expected
tubing properly seated and straight

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Front panel keys not accepted caution

One or more front panel keys do not respond to operator input

Ranked causes

  1. 1. Front panel locked (common)
  2. 2. Key requires reset (occasional)
  3. 3. Front panel ribbon cable disconnected or faulty (occasional)
  4. 4. Front panel film defective (rare)
  5. 5. Sensor board failure (rare)
  6. 6. Display board failure (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Free flow occurs when door is opened patient safety

Solution flows past pump mechanism when door is opened, when flow should stop

Ranked causes

  1. 1. Incorrect administration set type (occasional)
  2. 2. Solution spills on safety clamp (occasional)
  3. 3. Tubing positioning or seating improper (occasional)
  4. 4. Temperature outside 60-100 F range (occasional)
  5. 5. Broken safety clamp spring or clamp (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INSERT SLIDECLAMP alert or alarm constantly on caution

INSERT SLIDECLAMP alert or alarm message remains on display continuously

Ranked causes

  1. 1. Slide clamp sensor dirty or contaminated (common)
  2. 2. Connection issue with sensor (occasional)
  3. 3. Sensor calibration required (occasional)
  4. 4. Sensor board failure (rare)
  5. 5. Safety/slide clamp assembly fault (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

OCCLUSION or UPSTREAM OCCLUSION alarm with pump door closed patient safety

Alarm occurs when pump door closed and START key pressed

Ranked causes

  1. 1. Tubing mispositioned or pinched (common)
  2. 2. Obstruction in line upstream or downstream (common)
  3. 3. Wrong administration set type (occasional)
  4. 4. Ambient or solution temperature too low (occasional)
  5. 5. Spill or residue on door or base plate (occasional)
First check

Verify tubing positioning and check for kinks

Test point
Tubing path through pump
Expected
Tubing properly seated, no pinches or kinks

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AIR alarm occurs with tubing in sensor patient safety

Air alarm triggered when tubing properly seated or pump door closed and START pressed

Ranked causes

  1. 1. Tubing not fully positioned into air sensor (common)
  2. 2. Tubing scratched or flattened (occasional)
  3. 3. Wrong administration set type (occasional)
  4. 4. Air sensor contaminated with solution spill (occasional)
First check

Verify tubing positioning in air sensor

Test point
Air sensor
Expected
Tubing fully positioned

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air detected in administration set patient safety

Ultrasonic sensor detects air bubble; pump stops infusing, red ALARM LED illuminates, AIR message displayed, audible alarm activated

Ranked causes

  1. 1. Air bubble at sensor (common)
  2. 2. Empty fluid container (common)
  3. 3. Improper set loading (common)
  4. 4. START key pressed with no set in pump (occasional)
First check

Inspect administration set for air bubbles

Test point
Tubing at pump head
Expected
No visible air bubbles

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air detected in infusion set patient safety

Ultrasonic sensor detects air bubble in administration set. Pump stops infusing, red ALARM LED illuminates, AIR displayed in message display, audible alarm activated.

Ranked causes

  1. 1. Air bubble at sensor (common)
  2. 2. Empty fluid container (common)
  3. 3. Improper set loading (occasional)
  4. 4. Start key pressed with no set in pump (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PANEL LOCK switch not accepted caution

PANEL LOCK switch on rear panel does not register or activate properly

Ranked causes

  1. 1. Switch operation malfunction (common)
  2. 2. CN701 or CNI/F2 connection fault (occasional)
  3. 3. PANEL LOCK switch hardware failure (occasional)
  4. 4. Sensor board fault (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power indicator not lit when plugged in

Plug icon not lit or battery icon lit when device plugged into AC outlet

Ranked causes

  1. 1. Power plug not firmly seated in outlet (common)
  2. 2. Rear power fuses failed (common)
  3. 3. AC outlet voltage too low (occasional)
  4. 4. Power cord continuity broken (occasional)
First check

Verify power plug is tightly seated

Test point
AC outlet
Expected
Power plug fully inserted into outlet

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump door does not open or close smoothly caution

Door movement is difficult or restricted

Ranked causes

  1. 1. Tubing or slide clamp misaligned (common)
  2. 2. Wrong administration set type (occasional)
  3. 3. Solution spills or debris in door mechanism (occasional)
  4. 4. Door latch or hinge damage (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump door will not open or close smoothly caution

Door binding or difficult to operate

Ranked causes

  1. 1. Tubing and slide clamp mispositioned (common)
  2. 2. Wrong administration set type (occasional)
  3. 3. Door latch roller pin stiff or damaged (occasional)
  4. 4. Door hinge or latch damaged (occasional)
First check

Check tubing and slide clamp positioning

Test point
Pump door interior guides and channels
Expected
Tubing seated in guides, slide clamp in slot, safety clamp open

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump operation louder than normal caution

Pump produces noise levels exceeding normal operating volume

Ranked causes

  1. 1. Incorrect administration set type (occasional)
  2. 2. Solution spills on pumping fingers or backplate (occasional)
  3. 3. Loose motor mounting bracket screws (occasional)
  4. 4. PC board or assembly vibrating internally (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety clamp does not latch open caution

Safety clamp fails to remain in open position or does not latch properly

Ranked causes

  1. 1. Improper positioning or seating of slide clamp (common)
  2. 2. Incompatible administration set (occasional)
  3. 3. Safety clamp arm cover not fully open (occasional)
  4. 4. Solution spills or residue (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety clamp will not latch open

Safety clamp cannot be opened or latches improperly

Ranked causes

  1. 1. Slide clamp not fully positioned in slot (common)
  2. 2. Wrong administration set type (occasional)
  3. 3. Safety clamp arm cover not in full open position (occasional)
  4. 4. Spill or residue on safety clamp (occasional)
First check

Verify slide clamp positioned fully in slot

Test point
Slide clamp slot
Expected
Slide clamp fully seated

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Slide clamp not inserted or improperly positioned patient safety

Safety slide clamp missing or not fully seated in pump slot; blocks infusion until corrected

Ranked causes

  1. 1. Slide clamp not inserted or misaligned (common)
  2. 2. Dirty slide clamp sensor (occasional)
  3. 3. Slide clamp sensor calibration out of range (occasional)
  4. 4. Slide clamp sensor or assembly defective (rare)
First check

Slide clamp physical insertion

Test point
Slide clamp slot
Expected
Clamp fully inserted and seated
Tools
Visual inspection

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AIR alarm with no air present or door closed

Air detection alarm sounds without air in line or with door closed

Ranked causes

  1. 1. Tubing damaged or flattened (common)
  2. 2. Wrong administration set (occasional)
  3. 3. Air sensor contaminated (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup buzzer does not activate during self test

Backup alert buzzer fails to sound during power-on self test

Ranked causes

  1. 1. Buzzer wiring disconnected (common)
  2. 2. Battery fuse blown (occasional)
  3. 3. Backup buzzer defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Main battery voltage (MB)

Verification of main battery charging circuit voltage in Automatic Test Mode 1

Ranked causes

  1. 1. Main battery voltage out of specification (unknown)
  2. 2. DC power supply requires calibration (unknown)
  3. 3. Main battery requires replacement (unknown)
First check

Check MB value in SEC RATE window

Test point
SEC RATE window on Main Display
Expected
688 to 714 displayed counts, 688-714
Tools
Pump display

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication error with external device

Serial communication failure between pump and external computer; framing, overrun, or parity error detected

Ranked causes

  1. 1. Pump not in computer control mode (common)
  2. 2. Communication cable disconnected or damaged (common)
  3. 3. PC software error or mismatch (occasional)
First check

Pump in computer control mode

Test point
Pump front panel display
Expected
Computer control mode active
Tools
Visual inspection of display

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device or pump unexpectedly stops with audible alarm

Pump stops running with alarm sound for no apparent cause

Ranked causes

  1. 1. ICP301 component failure (common)
  2. 2. Front panel connection failure (occasional)
  3. 3. Front panel film or Display board failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door not fully closed caution

Pump door not fully closed or latch magnet missing. Tubing not properly loaded. Pump will not operate until door fully closed with tubing properly seated.

Ranked causes

  1. 1. Door not fully closed (common)
  2. 2. Tubing not properly loaded (common)
  3. 3. Missing magnet in door latch (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DOOR OPEN alarm incorrect operation caution

DOOR OPEN alarm occurs when door is closed, or does not occur when door is opened

Ranked causes

  1. 1. Missing magnet on door latch (common)
  2. 2. CN851 or CN852 connection fault (occasional)
  3. 3. CPU board fault (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Downstream occlusion detected caution

Pressure buildup detected in line distal to pump; may indicate blockage or kink in delivery tubing

Ranked causes

  1. 1. Tubing damage, kink, or occlusion (common)
  2. 2. Temperature out of specification (occasional)
  3. 3. Dirty or contaminated occlusion sensor (occasional)
First check

Inspect tubing for damage or kinks

Test point
Entire downstream tubing path
Expected
Tubing smooth, no kinks, no visible blockage
Tools
Visual inspection

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

INSERT SLIDE CLAMP alert or alarm

Alert or alarm when pump door is closed

Ranked causes

  1. 1. Slide clamp not in slide clamp slot (common)
  2. 2. Wrong administration set without slide clamp (occasional)
  3. 3. Spill or residue on slide clamp assembly (occasional)
First check

Verify slide clamp is in slot

Test point
Slide clamp slot
Expected
Slide clamp fully inserted into slot

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

INSERT SLIDECLAMP alert or alarm when door closed caution

INSERT SLIDECLAMP alert or alarm occurs when pump door is in closed position

Ranked causes

  1. 1. Slide clamp not inserted in slot (common)
  2. 2. Solution spills on slide clamp assembly (occasional)
  3. 3. Incompatible administration set (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery voltage detected caution

Battery voltage drops to 10.4 V or below; may indicate dead battery or charging system failure

Ranked causes

  1. 1. AC charger not connected or defective (common)
  2. 2. Power or battery fuse blown (occasional)
  3. 3. Battery internal failure or connection loose (occasional)
First check

AC power connection and voltage

Test point
AC outlet, pump AC connector
Expected
AC power connected, voltage ≥105 VAC, +20/-15 VAC
Tools
Multimeter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

One or both pumps cannot be turned on

Pump fails to start when power button activated

Ranked causes

  1. 1. Low or dead backup battery (common)
  2. 2. Front panel cable short circuit (occasional)
  3. 3. CPU board failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump does not start or stops with failure code alarm patient safety

Pump fails to operate when START key pressed or immediately stops with alarm code displayed

Ranked causes

  1. 1. Administration set not primed or improperly loaded (common)
  2. 2. Door open or improperly closed (common)
  3. 3. Failure code alarm condition detected (occasional)
First check

Verify administration set is primed and properly loaded

Test point
Tubing in pump guide channel and mechanism
Expected
Tubing clean, dry, seated properly with no slack, kinks or pinches; no resistance when door closes

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Slide clamp not detected or INSERT SLIDECLAMP alert not clearing patient safety

After loading primed administration set with slide clamp into slide clamp slot, INSERT SLIDECLAMP message may appear on display or not clear after proper insertion.

Ranked causes

  1. 1. Slide clamp sensor contamination or misalignment (common)
  2. 2. Slide clamp sensor calibration error (occasional)
  3. 3. Slide clamp assembly or wiring failure (occasional)
First check

Verify slide clamp feature configuration is enabled (option 2 or 3)

Test point
Configuration utility SLIDE CLAMP setting
Expected
Option 2 (ON or ALERT) or 3 (ALARM)

8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Upstream occlusion detected caution

Pressure drop or vacuum detected in proximal tubing; may indicate air leak, tubing damage, or blockage on supply side

Ranked causes

  1. 1. Tubing damage, leak, or blockage (common)
  2. 2. Dirty upstream occlusion sensor (occasional)
  3. 3. Temperature out of specification (occasional)
First check

Upstream line pressure and connections

Test point
Supply tubing, connectors, catheter hub
Expected
Tight connections, no leaks, no blockage counts, +/-36 counts
Tools
Visual inspection, pump display reading

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air bubble detected in infusion line patient safety

Air detected in tubing during pump operation in Run mode; may trigger audible alarm

Ranked causes

  1. 1. Improper tubing load or tubing damage (common)
  2. 2. Dirty or contaminated air sensor (occasional)
First check

Visual inspection of tubing

Test point
Entire infusion line path
Expected
No visible air bubbles, no flattening or scratches
Tools
Visual inspection

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air detection not functioning or incorrect sensitivity patient safety

Pump fails to detect air bubbles or air detection sensitivity is incorrect as determined by the Air Alarm Test.

Ranked causes

  1. 1. Air bubble size configuration setting incorrect (common)
  2. 2. Air sensor requires troubleshooting and repair (occasional)
First check

Prime and load administration set with Y-site above pump

Test point
Administration set inlet
Expected
Set primed and loaded per Administration Set Preparation steps 1-6

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air sensor does not detect air bubbles patient safety

Pump does not alarm or display AIR message when air bubble reaches sensor

Ranked causes

  1. 1. Air bubble breaks up before reaching sensor (occasional)
  2. 2. Air sensor calibration incorrect (common)
First check

Observe air sensor response to air bubble

Test point
Air sensor in pump
Expected
AIR displayed, red ALARM LED flashes, audible alarm activated, pump stops

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

All front panel keys disabled except BACKLIGHT

Most front panel keys do not respond to presses

Ranked causes

  1. 1. Panel lock engaged (common)
  2. 2. Display board failure (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm always on caution

Alarm sounds continuously without stopping

Ranked causes

  1. 1. Battery disconnect/reconnect needed (common)
  2. 2. Component connection or short circuit (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound patient safety

Alarm fails to generate sound when triggered

Ranked causes

  1. 1. Software error or connection issue (common)
  2. 2. Speaker defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backlight is off when running on internal battery power

Display not lit while operating on battery

Ranked causes

  1. 1. Front panel locked (common)
  2. 2. Backlight not activated (common)
First check

Verify front panel lock status

Test point
Main Display
Expected
No LOT indicator shown

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not detected after AC disconnection

BATTERY message not displayed on pump when disconnected from AC power, or BATTERY message does not disappear when plugged back in

Ranked causes

  1. 1. Battery absent or not functioning (common)
  2. 2. Battery charging voltage incorrect (occasional)
First check

Verify BATTERY message displayed when AC disconnected

Test point
Main display
Expected
BATTERY message appears

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Beep does not sound when key pressed (non-BACKLIGHT/SILENCE)

No audible beep when pressing control buttons

Ranked causes

  1. 1. Connector disconnection or short circuit (common)
  2. 2. Speaker defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Air sensor patient safety

Calibrate air detection sensor using fluid-filled and air-filled calibration tubing segments; required after pump head, CPU board, air sensor, door assembly, or latch replacement

Ranked causes

  1. 1. Air sensor replacement or related component replaced (common)
  2. 2. Air sensor values out of calibration range (occasional)
First check

Check for pump head door modification

Test point
Pump head door opposite air sensor
Expected
White tubing guide shim or letter 'A' present

10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication timeout with computer

No communication between device and controlling computer for specified time period; COMM TIMEOUT message displayed

Ranked causes

  1. 1. Disconnected communication cable (common)
  2. 2. Computer problem or malfunction (occasional)
First check

Verify communication cable connection

Test point
Communications port connector
Expected
Cable fully inserted and secure at device and computer

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door open alarm does not activate when door is opened during operation patient safety

Pump continues to operate or alarm does not sound/display when pump door opened while pumping

Ranked causes

  1. 1. Door open sensor malfunction (rare)
  2. 2. Display or alarm circuits defective (occasional)
First check

Program and start pump at specified settings

Test point
Pump select key; PRI RATE, PRI VTBI, PRI START keys
Expected
Primary rate 100 mL/hr, primary VTBI 10 mL, pump running

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrical safety parameters out of specification high voltage

Leakage current exceeds 50 microamps or ground impedance exceeds specification; power cord shows wear, damage, or broken pins

Ranked causes

  1. 1. Leakage current or ground impedance out of specification (rare)
  2. 2. Power cord damaged or worn (occasional)
First check

Measure leakage current per UL 544

Test point
Device under test per UL 544 standard
Expected
Not to exceed 50 microamps
Tools
Leakage current tester per UL 544

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

External communication error from computer

EXT COMM ERROR message displayed; controlling computer sending multiple queries or commands without waiting for device replies

Ranked causes

  1. 1. Computer sending multiple queries without waiting for device responses (common)
  2. 2. Computer software malfunction (occasional)
First check

Verify pump is responding to manual commands

Test point
Pump control buttons (SILENCE, PUMP 1, PUMP 2, OPTIONS)
Expected
Pump responds to button press and switches to manual mode

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Force Sensing Resistor (FSR) values unstable

FSR zero offset values jump more than 10 counts between power cycles

Ranked causes

  1. 1. FSR connector dirty or corroded (common)
  2. 2. FSR device defective (occasional)
First check

Note left and right FSR values with pump on

Test point
FSR display readout

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

INSERT SLIDECLAMP message appears when clamp not fully inserted

When the clamp is only inserted approximately one-third of the way into the mechanism and the pump door is closed, the INSERT SLIDECLAMP message incorrectly displays.

Ranked causes

  1. 1. Slide clamp sensor dirty or requires calibration (common)
  2. 2. Safety/slide clamp assembly defective (occasional)
First check

Verify INSERT SLIDECLAMP message does not display with partial clamp insertion

Test point
Display
Expected
INSERT SLIDECLAMP does not appear on display

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INSERT SLIDECLAMP message does not appear on display

When the clamp is inserted approximately one-third of the way into the mechanism and the pump door is closed, the INSERT SLIDECLAMP message fails to display.

Ranked causes

  1. 1. Slide clamp sensor dirty or misaligned (common)
  2. 2. Safety/slide clamp assembly or wiring defective (occasional)
First check

Verify INSERT SLIDECLAMP message displays after cleaning sensor

Test point
Display
Expected
INSERT SLIDECLAMP appears on display

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD blank, missing segments, or irrational display

LCD display is blank, shows missing characters, or displays incorrect information

Ranked causes

  1. 1. LCD connection or mounting issue (common)
  2. 2. Sensor board failure (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No communication with computer caution

No communication between pump and controlling computer for specified time period. Pump displays COM TIMEOUT alarm.

Ranked causes

  1. 1. Disconnected communication cable (common)
  2. 2. Computer communication problem (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

One or more front panel keys not accepted

Individual keys fail to respond when pressed

Ranked causes

  1. 1. Key needs reboot to clear (common)
  2. 2. Front panel cable issue (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump delivers excessively high flow or fails drive defect test caution

Pump running at 1999 mL/hr with fluid-filled tubing produces downstream occlusion alarm or F-78/F-79 failure code within 15 seconds

Ranked causes

  1. 1. Drive mechanism defect (F-78 or F-79 alarm) (rare)
  2. 2. Downstream occlusion alarm triggered by normal high-speed operation (occasional)
First check

Prepare and load fluid-filled administration set

Test point
Tubing loaded in pump head
Expected
Fluid-filled tubing properly seated, door closed, all clamps open

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump does not operate long on battery power

Battery runtime significantly shorter than expected

Ranked causes

  1. 1. Battery capacity degraded (common)
  2. 2. Battery harness continuity failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump does not run on AC but runs on battery

Unit operates only on internal battery; AC power not working

Ranked causes

  1. 1. AC power cord not firmly connected (common)
  2. 2. AC fuses blown (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump fails to detect air bubble in tubing patient safety

Air bubble present in administration set but no air alarm triggered; pump continues to run without stopping

Ranked causes

  1. 1. Air sensor calibration incorrect or sensor defective (occasional)
  2. 2. Air bubble too small to trigger alarm threshold (common)
First check

Verify air sensor calibration values

Test point
Air sensor calibration per Section 5.3.2
Expected
Calibration values within specification for NORM or M/N setting

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

RATE, VTBI, or START key not accepted by pump

Keypad input ignored or rejected

Ranked causes

  1. 1. Rate outside allowable configuration range (common)
  2. 2. VTBI outside allowable configuration range (common)
First check

Verify rate within configuration limits

Test point
Configuration option
Expected
Rate within allowable range

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rear clamp fails to hold pump on IV pole patient safety

Rear clamp does not grip firmly enough to secure pump to intravenous pole

Ranked causes

  1. 1. Worn or missing friction pad (common)
  2. 2. Worn clamp threads or other defective parts (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Remote computer configuration cannot be set
  • Slide clamp insert message not displayed or not clearing
  • Tubing misloaded in pump mechanism
  • Air detected in line
  • ALARM and ALERT LEDs lit but pump does not operate
  • ALARM LED always lit
  • ALARM LED does not light
  • ALARM LED(s) always lit
  • ALARM LED(s) do not light
  • Alarm volume not adjustable
  • ALERT LED always lit
  • ALERT LED does not light
  • ALERT LED(s) always lit
  • ALERT LED(s) do not light
  • All front panel keys disabled except BACKLIGHT and TOT VOL/STATUS
  • Audible alarm tone interval too long
  • Audible alarm volume insufficient
  • Audible alarm volume too low
  • Backlight cannot be turned off
  • Backlight does not light
  • Backlight off while pump runs on internal battery
  • Battery capacity degraded
  • Battery does not provide backup power when AC unavailable
  • Battery Icon lit during AC operation
  • Battery icon lit while pump runs on AC
  • Battery Icon off during battery operation
  • Battery icon off while pump runs on internal battery
  • Battery low
  • Battery not charging
  • Battery power exhausted
  • Calibration: A/D Converter Reference Voltage
  • Calibration: Air sensor recheck after value adjustment
  • Calibration: Air sensor with no tubing
  • Calibration: Air sensor with primed tubing
  • Calibration: Air sensor with unprimed (air-filled) tubing
  • Calibration: Backup battery voltage (BB)
  • Calibration: Battery charging
  • Calibration: CPU voltage (Vcpu)
  • Calibration: DC Line Voltage
  • Calibration: DC line voltages
  • Calibration: Device temperature and initial state
  • Calibration: Force Sensing Resistor (FSR) tube misloading sensor with door closed
  • Calibration: Force Sensing Resistor (FSR) tube misloading sensor with door open
  • Calibration: FSR sensor response to tubing misloading
  • Calibration: Slide Shaft Sensor
  • CHECK SET LOADING alarm when pump door closed
  • CHECK SET/LOADING alarm when pump door closed
  • Communication timeout with external device
  • Computer communication error
  • Device fails to run on internal battery
  • Device runs on battery but not AC power
  • Door Open Alarm does not trigger
  • DOOR OPEN alarm incorrectly triggered or missing
  • Door open alarm not activating
  • Downstream occlusion not detected
  • Downstream occlusion sensor not responding
  • FLOW RATE or CHECK VTBI alarm when START pressed
  • FLOW RATE or CHECK VTBI alarm with START key pressed
  • Gravity flow not prevented by safety clamp
  • Hi or Lo displayed during Volume-Time programming
  • INSERT SLIDE CLAMP alert or alarm constantly on
  • Keypad sealant defective or missing
  • LCD displays blank, irrational, or missing segments
  • Mounting clamp screw binding or clamp components damaged
  • No beep on key press when pump stopped
  • One or both pumps cannot be turned off
  • One or both pumps do not power off
  • One or both pumps do not power on
  • Optional nurse call always off
  • Optional nurse call always on
  • Panel lock not functioning
  • Plug icon not lit when plugged into AC
  • Power cord insulation damaged or connectors compromised
  • Pump delivers different volume than programmed VTBI
  • Pump does not detect upstream occlusion within 3 minutes
  • Pump does not stop or alarm when downstream tubing is pinched
  • Pump exterior unclean or physically damaged
  • Pump fails to run on internal battery (no LCD display)
  • Pump flow rate accuracy out of specification
  • Pump malfunction detected
  • Pump stops with audible alarm unexpectedly
  • Pump stops with BATTERY LOW alarm
  • PUMPING LED always lit
  • PUMPING LED does not light
  • PUMPING LED(s) always lit
  • PUMPING LED(s) do not light
  • Rear clamp fails to hold device on IV pole
  • Safety clamp arm fails to hold position or tubing removal not properly restricted
  • Self-test does not complete properly
  • Slide clamp alert or alarm does not trigger
  • Slide clamp not loaded
  • Slide clamp not loaded in slot
  • START key causes BATTERY LOW alarm
  • Tube misloading not detected
  • Tubing not fully positioned in air sensor
  • Tubing pinched at pump head door
  • Tubing pinched in door
  • Upstream occlusion alarm response delayed
  • Upstream occlusion not detected
  • Upstream occlusion sensor not responding

100 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0No alarms or failuresinfo
1Air bubble detected in tubingAir detected during infusion in run modewarning
10Misloaded tubing detectedInfusion set tubing not properly positioned in pump mechanismwarning
12Upstream occlusion (Revolution Counter)Upstream occlusion detected by revolution counter interruptwarning
2Downstream occlusion detectedSensor output increased by specified valueerror
20Door open detection circuit malfunctionDoor open sensor circuit remains low when it should be highcritical
21Occlusion detection frequency converter faultFrequency converter for occlusion sensor P20 remains high or lowcritical
22Occlusion detection converter fault (duplicate)Same as Code 21; frequency converter circuit failurecritical
23Door Open detection circuit malfunctionP54 (Pump 1) or P55 (Pump 2) remains lowerror
23IRQ interrupt port stuck highP50 IRQ port pin remains at +5 VDC when it should be lowcritical
24UP and DOWN occlusion values agree incorrectlyUpstream and downstream occlusion readings match when they should differcritical
25Generic occlusion circuit failureOcclusion detection circuit malfunctioncritical
25Frequency converter malfunction for occlusion detectionP20 of Master CPU remains lowerror
27Slide clamp detection circuit malfunctionFrequency converter circuit failureerror
28Four air sensors outputs cannot be selectedNORM and MIN outputs of both pumps cannot be selectederror
29Two occlusion sensors cannot be selectedUP and DOWN sensors cannot be selectederror
3Upstream occlusion detectedPressure drop in proximal tubing detectedwarning
30A/D converter malfunction for air detectionA/D conversion did not complete in 50 microsecondserror
31Air MIN detection circuit malfunctionA/D converter input remains higherror
31Air MIN sensor A/D input stuck highA/D converter input for MIN air transducer remains high with no oscillationcritical
32Air NORM sensor A/D input stuck highA/D converter input for NORM air transducer remains highcritical
32Air NORM detection circuit malfunctionA/D converter input remains higherror
33Air MIN sensor not oscillating but A/D not zeroMIN air transducer stopped oscillating but A/D input is not 0 Vcritical
34Air NORM sensor not oscillating but A/D not zeroNORM air transducer stopped oscillating but A/D input is not 0 Vcritical
35Front panel key stuck for 40+ secondsAny front panel key held down for more than 40 secondsunknown
36Abnormal power ON/OFF signalMaster CPU received incorrect power control signal during startup/shutdowncritical
37ON-OFF/CHARGE key pressed for more than 5 seconds when pump power was offKey pressed for more than 5 seconds while powered offwarning
38Malfunction in tube misloading detection circuitryTube misload detector hardware failurecritical
4Upstream occlusion detectedSensor output decreased by specified valueerror
41Abnormal operation of master CPUMaster CPU malfunction detectedcritical
42Malfunction in alarm control circuitAlarm control circuitry failurecritical
46PWRON1 signal is held lowPWRON1 signal stuck lowcritical
47PWRON1 signal is held highPWRON1 signal stuck highcritical
48RTC does not count time correctlyReal time clock malfunctioncritical
49Abnormal RTC dataRTC data corruption or defectcritical
5Upstream occlusion (alternate)Alternative detection method confirms upstream occlusionwarning
50Master CPU received TRAP interruptTRAP interrupt signal to master CPUcritical
51Master CPU received CMI interruptCMI interrupt signal detectedcritical
52Master CPU received SWI interruptSWI interrupt signal detectedcritical
53Master CPU received NMI interruptNMI interrupt signal detectedcritical
54Master CPU received OCI interruptOCI interrupt signal detectedcritical
55Master CPU received IRQ2 interruptIRQ2 interrupt signal detectedcritical
58TIMER 1 interrupt does not occurTIMER 1 interrupt signal missingcritical
59Program stuck in block 1 routineProgram execution halted in block 1critical
6Low battery voltage detectedBattery voltage 10.4V or lesscritical
60Battery voltage detection input to A/D remains highA/D converter input stuck highcritical
61Battery voltage detection input to A/D remains lowA/D converter input stuck lowcritical
62Backup battery voltage detection input to A/D remains highBackup battery voltage detection A/D stuck highcritical
63Lithium backup battery voltage too lowBackup battery voltage insufficientcritical
64A/D conversion channels 0-3 did not complete in 144 mSA/D conversion timeout for channels 0-3critical
65A/D conversion channels 4-7 did not complete in 144 μSA/D conversion timeout for channels 4-7critical
66CPU supply voltage is too highCPU supply voltage exceeds acceptable rangecritical
67CPU supply voltage is too lowCPU supply voltage below acceptable rangecritical
68A/D conversion channels 8-9 did not complete in 72 μSA/D conversion timeout for channels 8-9critical
69Abnormal battery recharging voltageBattery charging voltage out of specificationcritical
7Communication alarm detectedFraming, overrun or parity error during communication with external computererror
70Data in duplicate memory areas does not agreeMemory check detects mismatch in duplicate storagecritical
71Slave CPU received no data from Master CPUMaster CPU fails to send data to Slavecritical
72Slave CPU received undefined data 4 times from Master CPUMultiple reception of undefined datacritical
73Overcurrent to motorMotor current exceeds safe limitcritical
74Motor slipped at maximum currentMotor unable to maintain engagement at max currentcritical
77No revolution interrupt while motor was rotatingMotor rotation not detected by encodercritical
78Drive defect - overinfusionMotor speed exceeds expected revolution countcritical
79Drive defect - underinfusionMotor speed below expected revolution countcritical
8Communication timeout alarm detectedNo communication between device and external PC for given perioderror
82No acknowledgment from Slave CPU for three trialsSlave CPU fails to acknowledge after three attemptscritical
83Undefined message type received from Slave CPUInvalid message received from Slave CPUcritical
86LVDT reading of downstream occlusion sensor too lowDownstream occlusion sensor LVDT below thresholdwarning
87LVDT reading of downstream occlusion sensor too highDownstream occlusion sensor LVDT above thresholdwarning
89Bank switching control circuit failureBank switching circuit malfunction detectedcritical
9Slide Clamp AlarmSlide clamp not inserted or improperly inserted (software 1.09+)warning
90Checksum error of RATE, VTBI and VI dataData checksum verification failedcritical
91Checksum error of Programmed Delivery Profile stored dataPDP stored data checksum verification failedcritical
92Checksum error of Programmed Delivery Profile working dataPDP working data checksum verification failedcritical
93M5M5165 write data and read data do not agreeMemory write/read verification failedcritical
94Configuration option settings checksum is not 0Configuration settings checksum verification failedcritical
95Checksum error of alarm log dataAlarm log data checksum verification failedcritical
96Checksum error of ROM codeROM code checksum verification failedcritical
99Undefined failure code received from Slave CPUInvalid failure code from Slave CPUcritical
AIRAir bubble detected in infusion setUltrasonic sensor detects air bubbleunknown
BATTERY LOWBattery charge level critically lowBattery charge drops below minimum operating levelwarning
BATTERY LOW (critical)Battery power exhaustedBattery depleted after initial 15-minute warningcritical
C03NIBP kinked tubingInternal tubing kink detected in NIBP systemwarning
C04NIBP disconnected tubing or module faultNIBP tubing disconnected or module failure detectedwarning
CHECK SET/LOADINGTubing pinched in pump doorTubing detected between door and base platewarning
COMMON FAILURE with code numberDevice-level microprocessor malfunction detectedDual-pump system detects fault; code number identifies specific failurecritical
COM TIMEOUTNo pump-computer communicationCommunication timeout with computerwarning
DOOR OPENPump door not properly closedDoor sensor detects open position or improper latchwarning
EXT COMM ERRORComputer communication protocol errorComputer sends multiple commands without waiting for replieswarning
FAILUREMicroprocessor detected pump malfunctionHardware or software fault detectederror
FAILURE with code numberMicroprocessor detected pump malfunctionDevice detects internal fault; code number identifies specific failurecritical
INSERT SLIDECLAMPSlide clamp not detected in slide clamp slot or slide clamp sensor unable to confirm insertionPrimed administration set loaded into pump mechanism without slide clamp inserted into slide clamp slot, or slide clamp sensor contaminated or not detecting clamp presencewarning
INSERT SLIDECLAMPSlide clamp not loadedSlide clamp loading enabled but clamp not in slotwarning
OCCLUSIONDownstream blockage detectedIncreased back-pressure sensed between patient and pumpunknown
UPSTREAM OCCLUSIONUpstream blockage detectedClosed clamp or blockage upstream of pump sensedunknown

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI