Baxter Flo-Gard 6301: troubleshooting and error codes
The ServiceForge knowledge base holds 160 documented symptoms for the Baxter Flo-Gard 6301, with 293 ranked causes, 233 diagnostic checks with test points and expected values, and 95 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
AIR alarm continuously on patient safety
AIR alarm message remains on display continuously despite troubleshooting
Ranked causes
- 1. Air trapped in tubing (common)
- 2. Incorrect or improperly loaded administration set (common)
- 3. Tubing scratched or deformed (occasional)
- 4. Air sensor out of calibration (occasional)
- 5. Air sensor dirty (occasional)
- 6. Air sensor failure (rare)
- 7. Occlusion sensor failure (rare)
Check for air in tubing
- Test point
- administration set tubing
- Expected
- no visible air bubbles
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Free flow occurs when door is closed patient safety
Solution flows past the pump mechanism when door is closed, indicating loss of occlusion
Ranked causes
- 1. Incorrect administration set type (occasional)
- 2. Tubing positioning or seating improper (occasional)
- 3. Temperature outside 60-100 F range (occasional)
- 4. Cracked door hinges (occasional)
- 5. Deformed coil springs on back plate (occasional)
- 6. Worn cams and pumping fingers (occasional)
- 7. Damaged or deformed door latch (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate infusion delivery patient safety
Actual delivered volume does not match programmed rate or volume to be infused
Ranked causes
- 1. Incorrect administration set type (occasional)
- 2. Tubing positioning or seating improper (occasional)
- 3. Temperature outside 60-100 F range (occasional)
- 4. Cracked door hinges (occasional)
- 5. Deformed coil springs on back plate (occasional)
- 6. Worn cams and pumping fingers (occasional)
- 7. Damaged or deformed door latch (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OCCLUSION alarm occurs with pump door closed caution
OCCLUSION or UPSTREAM OCCLUSION alarm activates when pump door is closed and START key is pressed
Ranked causes
- 1. Tubing positioning or seating improper (common)
- 2. Obstructions upstream or downstream of pump (common)
- 3. Incorrect administration set type (occasional)
- 4. Temperature below 60 degrees F (occasional)
- 5. Solution spills or residue on door or baseplate (occasional)
- 6. Damaged door latch or hinges (occasional)
- 7. Sensor calibration required (unknown)
Verify tubing positioning and seating in sensor
- Test point
- pump sensor and safety clamp
- Expected
- tubing properly seated and straight
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front panel keys not accepted caution
One or more front panel keys do not respond to operator input
Ranked causes
- 1. Front panel locked (common)
- 2. Key requires reset (occasional)
- 3. Front panel ribbon cable disconnected or faulty (occasional)
- 4. Front panel film defective (rare)
- 5. Sensor board failure (rare)
- 6. Display board failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Free flow occurs when door is opened patient safety
Solution flows past pump mechanism when door is opened, when flow should stop
Ranked causes
- 1. Incorrect administration set type (occasional)
- 2. Solution spills on safety clamp (occasional)
- 3. Tubing positioning or seating improper (occasional)
- 4. Temperature outside 60-100 F range (occasional)
- 5. Broken safety clamp spring or clamp (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INSERT SLIDECLAMP alert or alarm constantly on caution
INSERT SLIDECLAMP alert or alarm message remains on display continuously
Ranked causes
- 1. Slide clamp sensor dirty or contaminated (common)
- 2. Connection issue with sensor (occasional)
- 3. Sensor calibration required (occasional)
- 4. Sensor board failure (rare)
- 5. Safety/slide clamp assembly fault (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OCCLUSION or UPSTREAM OCCLUSION alarm with pump door closed patient safety
Alarm occurs when pump door closed and START key pressed
Ranked causes
- 1. Tubing mispositioned or pinched (common)
- 2. Obstruction in line upstream or downstream (common)
- 3. Wrong administration set type (occasional)
- 4. Ambient or solution temperature too low (occasional)
- 5. Spill or residue on door or base plate (occasional)
Verify tubing positioning and check for kinks
- Test point
- Tubing path through pump
- Expected
- Tubing properly seated, no pinches or kinks
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AIR alarm occurs with tubing in sensor patient safety
Air alarm triggered when tubing properly seated or pump door closed and START pressed
Ranked causes
- 1. Tubing not fully positioned into air sensor (common)
- 2. Tubing scratched or flattened (occasional)
- 3. Wrong administration set type (occasional)
- 4. Air sensor contaminated with solution spill (occasional)
Verify tubing positioning in air sensor
- Test point
- Air sensor
- Expected
- Tubing fully positioned
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detected in administration set patient safety
Ultrasonic sensor detects air bubble; pump stops infusing, red ALARM LED illuminates, AIR message displayed, audible alarm activated
Ranked causes
- 1. Air bubble at sensor (common)
- 2. Empty fluid container (common)
- 3. Improper set loading (common)
- 4. START key pressed with no set in pump (occasional)
Inspect administration set for air bubbles
- Test point
- Tubing at pump head
- Expected
- No visible air bubbles
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detected in infusion set patient safety
Ultrasonic sensor detects air bubble in administration set. Pump stops infusing, red ALARM LED illuminates, AIR displayed in message display, audible alarm activated.
Ranked causes
- 1. Air bubble at sensor (common)
- 2. Empty fluid container (common)
- 3. Improper set loading (occasional)
- 4. Start key pressed with no set in pump (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PANEL LOCK switch not accepted caution
PANEL LOCK switch on rear panel does not register or activate properly
Ranked causes
- 1. Switch operation malfunction (common)
- 2. CN701 or CNI/F2 connection fault (occasional)
- 3. PANEL LOCK switch hardware failure (occasional)
- 4. Sensor board fault (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power indicator not lit when plugged in
Plug icon not lit or battery icon lit when device plugged into AC outlet
Ranked causes
- 1. Power plug not firmly seated in outlet (common)
- 2. Rear power fuses failed (common)
- 3. AC outlet voltage too low (occasional)
- 4. Power cord continuity broken (occasional)
Verify power plug is tightly seated
- Test point
- AC outlet
- Expected
- Power plug fully inserted into outlet
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump door does not open or close smoothly caution
Door movement is difficult or restricted
Ranked causes
- 1. Tubing or slide clamp misaligned (common)
- 2. Wrong administration set type (occasional)
- 3. Solution spills or debris in door mechanism (occasional)
- 4. Door latch or hinge damage (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump door will not open or close smoothly caution
Door binding or difficult to operate
Ranked causes
- 1. Tubing and slide clamp mispositioned (common)
- 2. Wrong administration set type (occasional)
- 3. Door latch roller pin stiff or damaged (occasional)
- 4. Door hinge or latch damaged (occasional)
Check tubing and slide clamp positioning
- Test point
- Pump door interior guides and channels
- Expected
- Tubing seated in guides, slide clamp in slot, safety clamp open
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump operation louder than normal caution
Pump produces noise levels exceeding normal operating volume
Ranked causes
- 1. Incorrect administration set type (occasional)
- 2. Solution spills on pumping fingers or backplate (occasional)
- 3. Loose motor mounting bracket screws (occasional)
- 4. PC board or assembly vibrating internally (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety clamp does not latch open caution
Safety clamp fails to remain in open position or does not latch properly
Ranked causes
- 1. Improper positioning or seating of slide clamp (common)
- 2. Incompatible administration set (occasional)
- 3. Safety clamp arm cover not fully open (occasional)
- 4. Solution spills or residue (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety clamp will not latch open
Safety clamp cannot be opened or latches improperly
Ranked causes
- 1. Slide clamp not fully positioned in slot (common)
- 2. Wrong administration set type (occasional)
- 3. Safety clamp arm cover not in full open position (occasional)
- 4. Spill or residue on safety clamp (occasional)
Verify slide clamp positioned fully in slot
- Test point
- Slide clamp slot
- Expected
- Slide clamp fully seated
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slide clamp not inserted or improperly positioned patient safety
Safety slide clamp missing or not fully seated in pump slot; blocks infusion until corrected
Ranked causes
- 1. Slide clamp not inserted or misaligned (common)
- 2. Dirty slide clamp sensor (occasional)
- 3. Slide clamp sensor calibration out of range (occasional)
- 4. Slide clamp sensor or assembly defective (rare)
Slide clamp physical insertion
- Test point
- Slide clamp slot
- Expected
- Clamp fully inserted and seated
- Tools
- Visual inspection
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AIR alarm with no air present or door closed
Air detection alarm sounds without air in line or with door closed
Ranked causes
- 1. Tubing damaged or flattened (common)
- 2. Wrong administration set (occasional)
- 3. Air sensor contaminated (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup buzzer does not activate during self test
Backup alert buzzer fails to sound during power-on self test
Ranked causes
- 1. Buzzer wiring disconnected (common)
- 2. Battery fuse blown (occasional)
- 3. Backup buzzer defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Main battery voltage (MB)
Verification of main battery charging circuit voltage in Automatic Test Mode 1
Ranked causes
- 1. Main battery voltage out of specification (unknown)
- 2. DC power supply requires calibration (unknown)
- 3. Main battery requires replacement (unknown)
Check MB value in SEC RATE window
- Test point
- SEC RATE window on Main Display
- Expected
- 688 to 714 displayed counts, 688-714
- Tools
- Pump display
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication error with external device
Serial communication failure between pump and external computer; framing, overrun, or parity error detected
Ranked causes
- 1. Pump not in computer control mode (common)
- 2. Communication cable disconnected or damaged (common)
- 3. PC software error or mismatch (occasional)
Pump in computer control mode
- Test point
- Pump front panel display
- Expected
- Computer control mode active
- Tools
- Visual inspection of display
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device or pump unexpectedly stops with audible alarm
Pump stops running with alarm sound for no apparent cause
Ranked causes
- 1. ICP301 component failure (common)
- 2. Front panel connection failure (occasional)
- 3. Front panel film or Display board failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door not fully closed caution
Pump door not fully closed or latch magnet missing. Tubing not properly loaded. Pump will not operate until door fully closed with tubing properly seated.
Ranked causes
- 1. Door not fully closed (common)
- 2. Tubing not properly loaded (common)
- 3. Missing magnet in door latch (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DOOR OPEN alarm incorrect operation caution
DOOR OPEN alarm occurs when door is closed, or does not occur when door is opened
Ranked causes
- 1. Missing magnet on door latch (common)
- 2. CN851 or CN852 connection fault (occasional)
- 3. CPU board fault (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Downstream occlusion detected caution
Pressure buildup detected in line distal to pump; may indicate blockage or kink in delivery tubing
Ranked causes
- 1. Tubing damage, kink, or occlusion (common)
- 2. Temperature out of specification (occasional)
- 3. Dirty or contaminated occlusion sensor (occasional)
Inspect tubing for damage or kinks
- Test point
- Entire downstream tubing path
- Expected
- Tubing smooth, no kinks, no visible blockage
- Tools
- Visual inspection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INSERT SLIDE CLAMP alert or alarm
Alert or alarm when pump door is closed
Ranked causes
- 1. Slide clamp not in slide clamp slot (common)
- 2. Wrong administration set without slide clamp (occasional)
- 3. Spill or residue on slide clamp assembly (occasional)
Verify slide clamp is in slot
- Test point
- Slide clamp slot
- Expected
- Slide clamp fully inserted into slot
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INSERT SLIDECLAMP alert or alarm when door closed caution
INSERT SLIDECLAMP alert or alarm occurs when pump door is in closed position
Ranked causes
- 1. Slide clamp not inserted in slot (common)
- 2. Solution spills on slide clamp assembly (occasional)
- 3. Incompatible administration set (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery voltage detected caution
Battery voltage drops to 10.4 V or below; may indicate dead battery or charging system failure
Ranked causes
- 1. AC charger not connected or defective (common)
- 2. Power or battery fuse blown (occasional)
- 3. Battery internal failure or connection loose (occasional)
AC power connection and voltage
- Test point
- AC outlet, pump AC connector
- Expected
- AC power connected, voltage ≥105 VAC, +20/-15 VAC
- Tools
- Multimeter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or both pumps cannot be turned on
Pump fails to start when power button activated
Ranked causes
- 1. Low or dead backup battery (common)
- 2. Front panel cable short circuit (occasional)
- 3. CPU board failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not start or stops with failure code alarm patient safety
Pump fails to operate when START key pressed or immediately stops with alarm code displayed
Ranked causes
- 1. Administration set not primed or improperly loaded (common)
- 2. Door open or improperly closed (common)
- 3. Failure code alarm condition detected (occasional)
Verify administration set is primed and properly loaded
- Test point
- Tubing in pump guide channel and mechanism
- Expected
- Tubing clean, dry, seated properly with no slack, kinks or pinches; no resistance when door closes
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slide clamp not detected or INSERT SLIDECLAMP alert not clearing patient safety
After loading primed administration set with slide clamp into slide clamp slot, INSERT SLIDECLAMP message may appear on display or not clear after proper insertion.
Ranked causes
- 1. Slide clamp sensor contamination or misalignment (common)
- 2. Slide clamp sensor calibration error (occasional)
- 3. Slide clamp assembly or wiring failure (occasional)
Verify slide clamp feature configuration is enabled (option 2 or 3)
- Test point
- Configuration utility SLIDE CLAMP setting
- Expected
- Option 2 (ON or ALERT) or 3 (ALARM)
8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Upstream occlusion detected caution
Pressure drop or vacuum detected in proximal tubing; may indicate air leak, tubing damage, or blockage on supply side
Ranked causes
- 1. Tubing damage, leak, or blockage (common)
- 2. Dirty upstream occlusion sensor (occasional)
- 3. Temperature out of specification (occasional)
Upstream line pressure and connections
- Test point
- Supply tubing, connectors, catheter hub
- Expected
- Tight connections, no leaks, no blockage counts, +/-36 counts
- Tools
- Visual inspection, pump display reading
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air bubble detected in infusion line patient safety
Air detected in tubing during pump operation in Run mode; may trigger audible alarm
Ranked causes
- 1. Improper tubing load or tubing damage (common)
- 2. Dirty or contaminated air sensor (occasional)
Visual inspection of tubing
- Test point
- Entire infusion line path
- Expected
- No visible air bubbles, no flattening or scratches
- Tools
- Visual inspection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detection not functioning or incorrect sensitivity patient safety
Pump fails to detect air bubbles or air detection sensitivity is incorrect as determined by the Air Alarm Test.
Ranked causes
- 1. Air bubble size configuration setting incorrect (common)
- 2. Air sensor requires troubleshooting and repair (occasional)
Prime and load administration set with Y-site above pump
- Test point
- Administration set inlet
- Expected
- Set primed and loaded per Administration Set Preparation steps 1-6
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air sensor does not detect air bubbles patient safety
Pump does not alarm or display AIR message when air bubble reaches sensor
Ranked causes
- 1. Air bubble breaks up before reaching sensor (occasional)
- 2. Air sensor calibration incorrect (common)
Observe air sensor response to air bubble
- Test point
- Air sensor in pump
- Expected
- AIR displayed, red ALARM LED flashes, audible alarm activated, pump stops
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All front panel keys disabled except BACKLIGHT
Most front panel keys do not respond to presses
Ranked causes
- 1. Panel lock engaged (common)
- 2. Display board failure (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm always on caution
Alarm sounds continuously without stopping
Ranked causes
- 1. Battery disconnect/reconnect needed (common)
- 2. Component connection or short circuit (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound patient safety
Alarm fails to generate sound when triggered
Ranked causes
- 1. Software error or connection issue (common)
- 2. Speaker defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backlight is off when running on internal battery power
Display not lit while operating on battery
Ranked causes
- 1. Front panel locked (common)
- 2. Backlight not activated (common)
Verify front panel lock status
- Test point
- Main Display
- Expected
- No LOT indicator shown
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not detected after AC disconnection
BATTERY message not displayed on pump when disconnected from AC power, or BATTERY message does not disappear when plugged back in
Ranked causes
- 1. Battery absent or not functioning (common)
- 2. Battery charging voltage incorrect (occasional)
Verify BATTERY message displayed when AC disconnected
- Test point
- Main display
- Expected
- BATTERY message appears
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Beep does not sound when key pressed (non-BACKLIGHT/SILENCE)
No audible beep when pressing control buttons
Ranked causes
- 1. Connector disconnection or short circuit (common)
- 2. Speaker defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air sensor patient safety
Calibrate air detection sensor using fluid-filled and air-filled calibration tubing segments; required after pump head, CPU board, air sensor, door assembly, or latch replacement
Ranked causes
- 1. Air sensor replacement or related component replaced (common)
- 2. Air sensor values out of calibration range (occasional)
Check for pump head door modification
- Test point
- Pump head door opposite air sensor
- Expected
- White tubing guide shim or letter 'A' present
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication timeout with computer
No communication between device and controlling computer for specified time period; COMM TIMEOUT message displayed
Ranked causes
- 1. Disconnected communication cable (common)
- 2. Computer problem or malfunction (occasional)
Verify communication cable connection
- Test point
- Communications port connector
- Expected
- Cable fully inserted and secure at device and computer
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door open alarm does not activate when door is opened during operation patient safety
Pump continues to operate or alarm does not sound/display when pump door opened while pumping
Ranked causes
- 1. Door open sensor malfunction (rare)
- 2. Display or alarm circuits defective (occasional)
Program and start pump at specified settings
- Test point
- Pump select key; PRI RATE, PRI VTBI, PRI START keys
- Expected
- Primary rate 100 mL/hr, primary VTBI 10 mL, pump running
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety parameters out of specification high voltage
Leakage current exceeds 50 microamps or ground impedance exceeds specification; power cord shows wear, damage, or broken pins
Ranked causes
- 1. Leakage current or ground impedance out of specification (rare)
- 2. Power cord damaged or worn (occasional)
Measure leakage current per UL 544
- Test point
- Device under test per UL 544 standard
- Expected
- Not to exceed 50 microamps
- Tools
- Leakage current tester per UL 544
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External communication error from computer
EXT COMM ERROR message displayed; controlling computer sending multiple queries or commands without waiting for device replies
Ranked causes
- 1. Computer sending multiple queries without waiting for device responses (common)
- 2. Computer software malfunction (occasional)
Verify pump is responding to manual commands
- Test point
- Pump control buttons (SILENCE, PUMP 1, PUMP 2, OPTIONS)
- Expected
- Pump responds to button press and switches to manual mode
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Force Sensing Resistor (FSR) values unstable
FSR zero offset values jump more than 10 counts between power cycles
Ranked causes
- 1. FSR connector dirty or corroded (common)
- 2. FSR device defective (occasional)
Note left and right FSR values with pump on
- Test point
- FSR display readout
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INSERT SLIDECLAMP message appears when clamp not fully inserted
When the clamp is only inserted approximately one-third of the way into the mechanism and the pump door is closed, the INSERT SLIDECLAMP message incorrectly displays.
Ranked causes
- 1. Slide clamp sensor dirty or requires calibration (common)
- 2. Safety/slide clamp assembly defective (occasional)
Verify INSERT SLIDECLAMP message does not display with partial clamp insertion
- Test point
- Display
- Expected
- INSERT SLIDECLAMP does not appear on display
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INSERT SLIDECLAMP message does not appear on display
When the clamp is inserted approximately one-third of the way into the mechanism and the pump door is closed, the INSERT SLIDECLAMP message fails to display.
Ranked causes
- 1. Slide clamp sensor dirty or misaligned (common)
- 2. Safety/slide clamp assembly or wiring defective (occasional)
Verify INSERT SLIDECLAMP message displays after cleaning sensor
- Test point
- Display
- Expected
- INSERT SLIDECLAMP appears on display
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD blank, missing segments, or irrational display
LCD display is blank, shows missing characters, or displays incorrect information
Ranked causes
- 1. LCD connection or mounting issue (common)
- 2. Sensor board failure (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No communication with computer caution
No communication between pump and controlling computer for specified time period. Pump displays COM TIMEOUT alarm.
Ranked causes
- 1. Disconnected communication cable (common)
- 2. Computer communication problem (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more front panel keys not accepted
Individual keys fail to respond when pressed
Ranked causes
- 1. Key needs reboot to clear (common)
- 2. Front panel cable issue (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump delivers excessively high flow or fails drive defect test caution
Pump running at 1999 mL/hr with fluid-filled tubing produces downstream occlusion alarm or F-78/F-79 failure code within 15 seconds
Ranked causes
- 1. Drive mechanism defect (F-78 or F-79 alarm) (rare)
- 2. Downstream occlusion alarm triggered by normal high-speed operation (occasional)
Prepare and load fluid-filled administration set
- Test point
- Tubing loaded in pump head
- Expected
- Fluid-filled tubing properly seated, door closed, all clamps open
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not operate long on battery power
Battery runtime significantly shorter than expected
Ranked causes
- 1. Battery capacity degraded (common)
- 2. Battery harness continuity failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not run on AC but runs on battery
Unit operates only on internal battery; AC power not working
Ranked causes
- 1. AC power cord not firmly connected (common)
- 2. AC fuses blown (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump fails to detect air bubble in tubing patient safety
Air bubble present in administration set but no air alarm triggered; pump continues to run without stopping
Ranked causes
- 1. Air sensor calibration incorrect or sensor defective (occasional)
- 2. Air bubble too small to trigger alarm threshold (common)
Verify air sensor calibration values
- Test point
- Air sensor calibration per Section 5.3.2
- Expected
- Calibration values within specification for NORM or M/N setting
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RATE, VTBI, or START key not accepted by pump
Keypad input ignored or rejected
Ranked causes
- 1. Rate outside allowable configuration range (common)
- 2. VTBI outside allowable configuration range (common)
Verify rate within configuration limits
- Test point
- Configuration option
- Expected
- Rate within allowable range
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rear clamp fails to hold pump on IV pole patient safety
Rear clamp does not grip firmly enough to secure pump to intravenous pole
Ranked causes
- 1. Worn or missing friction pad (common)
- 2. Worn clamp threads or other defective parts (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Remote computer configuration cannot be set
- Slide clamp insert message not displayed or not clearing
- Tubing misloaded in pump mechanism
- Air detected in line
- ALARM and ALERT LEDs lit but pump does not operate
- ALARM LED always lit
- ALARM LED does not light
- ALARM LED(s) always lit
- ALARM LED(s) do not light
- Alarm volume not adjustable
- ALERT LED always lit
- ALERT LED does not light
- ALERT LED(s) always lit
- ALERT LED(s) do not light
- All front panel keys disabled except BACKLIGHT and TOT VOL/STATUS
- Audible alarm tone interval too long
- Audible alarm volume insufficient
- Audible alarm volume too low
- Backlight cannot be turned off
- Backlight does not light
- Backlight off while pump runs on internal battery
- Battery capacity degraded
- Battery does not provide backup power when AC unavailable
- Battery Icon lit during AC operation
- Battery icon lit while pump runs on AC
- Battery Icon off during battery operation
- Battery icon off while pump runs on internal battery
- Battery low
- Battery not charging
- Battery power exhausted
- Calibration: A/D Converter Reference Voltage
- Calibration: Air sensor recheck after value adjustment
- Calibration: Air sensor with no tubing
- Calibration: Air sensor with primed tubing
- Calibration: Air sensor with unprimed (air-filled) tubing
- Calibration: Backup battery voltage (BB)
- Calibration: Battery charging
- Calibration: CPU voltage (Vcpu)
- Calibration: DC Line Voltage
- Calibration: DC line voltages
- Calibration: Device temperature and initial state
- Calibration: Force Sensing Resistor (FSR) tube misloading sensor with door closed
- Calibration: Force Sensing Resistor (FSR) tube misloading sensor with door open
- Calibration: FSR sensor response to tubing misloading
- Calibration: Slide Shaft Sensor
- CHECK SET LOADING alarm when pump door closed
- CHECK SET/LOADING alarm when pump door closed
- Communication timeout with external device
- Computer communication error
- Device fails to run on internal battery
- Device runs on battery but not AC power
- Door Open Alarm does not trigger
- DOOR OPEN alarm incorrectly triggered or missing
- Door open alarm not activating
- Downstream occlusion not detected
- Downstream occlusion sensor not responding
- FLOW RATE or CHECK VTBI alarm when START pressed
- FLOW RATE or CHECK VTBI alarm with START key pressed
- Gravity flow not prevented by safety clamp
- Hi or Lo displayed during Volume-Time programming
- INSERT SLIDE CLAMP alert or alarm constantly on
- Keypad sealant defective or missing
- LCD displays blank, irrational, or missing segments
- Mounting clamp screw binding or clamp components damaged
- No beep on key press when pump stopped
- One or both pumps cannot be turned off
- One or both pumps do not power off
- One or both pumps do not power on
- Optional nurse call always off
- Optional nurse call always on
- Panel lock not functioning
- Plug icon not lit when plugged into AC
- Power cord insulation damaged or connectors compromised
- Pump delivers different volume than programmed VTBI
- Pump does not detect upstream occlusion within 3 minutes
- Pump does not stop or alarm when downstream tubing is pinched
- Pump exterior unclean or physically damaged
- Pump fails to run on internal battery (no LCD display)
- Pump flow rate accuracy out of specification
- Pump malfunction detected
- Pump stops with audible alarm unexpectedly
- Pump stops with BATTERY LOW alarm
- PUMPING LED always lit
- PUMPING LED does not light
- PUMPING LED(s) always lit
- PUMPING LED(s) do not light
- Rear clamp fails to hold device on IV pole
- Safety clamp arm fails to hold position or tubing removal not properly restricted
- Self-test does not complete properly
- Slide clamp alert or alarm does not trigger
- Slide clamp not loaded
- Slide clamp not loaded in slot
- START key causes BATTERY LOW alarm
- Tube misloading not detected
- Tubing not fully positioned in air sensor
- Tubing pinched at pump head door
- Tubing pinched in door
- Upstream occlusion alarm response delayed
- Upstream occlusion not detected
- Upstream occlusion sensor not responding
100 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0 | No alarms or failures | info | |
1 | Air bubble detected in tubing | Air detected during infusion in run mode | warning |
10 | Misloaded tubing detected | Infusion set tubing not properly positioned in pump mechanism | warning |
12 | Upstream occlusion (Revolution Counter) | Upstream occlusion detected by revolution counter interrupt | warning |
2 | Downstream occlusion detected | Sensor output increased by specified value | error |
20 | Door open detection circuit malfunction | Door open sensor circuit remains low when it should be high | critical |
21 | Occlusion detection frequency converter fault | Frequency converter for occlusion sensor P20 remains high or low | critical |
22 | Occlusion detection converter fault (duplicate) | Same as Code 21; frequency converter circuit failure | critical |
23 | Door Open detection circuit malfunction | P54 (Pump 1) or P55 (Pump 2) remains low | error |
23 | IRQ interrupt port stuck high | P50 IRQ port pin remains at +5 VDC when it should be low | critical |
24 | UP and DOWN occlusion values agree incorrectly | Upstream and downstream occlusion readings match when they should differ | critical |
25 | Generic occlusion circuit failure | Occlusion detection circuit malfunction | critical |
25 | Frequency converter malfunction for occlusion detection | P20 of Master CPU remains low | error |
27 | Slide clamp detection circuit malfunction | Frequency converter circuit failure | error |
28 | Four air sensors outputs cannot be selected | NORM and MIN outputs of both pumps cannot be selected | error |
29 | Two occlusion sensors cannot be selected | UP and DOWN sensors cannot be selected | error |
3 | Upstream occlusion detected | Pressure drop in proximal tubing detected | warning |
30 | A/D converter malfunction for air detection | A/D conversion did not complete in 50 microseconds | error |
31 | Air MIN detection circuit malfunction | A/D converter input remains high | error |
31 | Air MIN sensor A/D input stuck high | A/D converter input for MIN air transducer remains high with no oscillation | critical |
32 | Air NORM sensor A/D input stuck high | A/D converter input for NORM air transducer remains high | critical |
32 | Air NORM detection circuit malfunction | A/D converter input remains high | error |
33 | Air MIN sensor not oscillating but A/D not zero | MIN air transducer stopped oscillating but A/D input is not 0 V | critical |
34 | Air NORM sensor not oscillating but A/D not zero | NORM air transducer stopped oscillating but A/D input is not 0 V | critical |
35 | Front panel key stuck for 40+ seconds | Any front panel key held down for more than 40 seconds | unknown |
36 | Abnormal power ON/OFF signal | Master CPU received incorrect power control signal during startup/shutdown | critical |
37 | ON-OFF/CHARGE key pressed for more than 5 seconds when pump power was off | Key pressed for more than 5 seconds while powered off | warning |
38 | Malfunction in tube misloading detection circuitry | Tube misload detector hardware failure | critical |
4 | Upstream occlusion detected | Sensor output decreased by specified value | error |
41 | Abnormal operation of master CPU | Master CPU malfunction detected | critical |
42 | Malfunction in alarm control circuit | Alarm control circuitry failure | critical |
46 | PWRON1 signal is held low | PWRON1 signal stuck low | critical |
47 | PWRON1 signal is held high | PWRON1 signal stuck high | critical |
48 | RTC does not count time correctly | Real time clock malfunction | critical |
49 | Abnormal RTC data | RTC data corruption or defect | critical |
5 | Upstream occlusion (alternate) | Alternative detection method confirms upstream occlusion | warning |
50 | Master CPU received TRAP interrupt | TRAP interrupt signal to master CPU | critical |
51 | Master CPU received CMI interrupt | CMI interrupt signal detected | critical |
52 | Master CPU received SWI interrupt | SWI interrupt signal detected | critical |
53 | Master CPU received NMI interrupt | NMI interrupt signal detected | critical |
54 | Master CPU received OCI interrupt | OCI interrupt signal detected | critical |
55 | Master CPU received IRQ2 interrupt | IRQ2 interrupt signal detected | critical |
58 | TIMER 1 interrupt does not occur | TIMER 1 interrupt signal missing | critical |
59 | Program stuck in block 1 routine | Program execution halted in block 1 | critical |
6 | Low battery voltage detected | Battery voltage 10.4V or less | critical |
60 | Battery voltage detection input to A/D remains high | A/D converter input stuck high | critical |
61 | Battery voltage detection input to A/D remains low | A/D converter input stuck low | critical |
62 | Backup battery voltage detection input to A/D remains high | Backup battery voltage detection A/D stuck high | critical |
63 | Lithium backup battery voltage too low | Backup battery voltage insufficient | critical |
64 | A/D conversion channels 0-3 did not complete in 144 mS | A/D conversion timeout for channels 0-3 | critical |
65 | A/D conversion channels 4-7 did not complete in 144 μS | A/D conversion timeout for channels 4-7 | critical |
66 | CPU supply voltage is too high | CPU supply voltage exceeds acceptable range | critical |
67 | CPU supply voltage is too low | CPU supply voltage below acceptable range | critical |
68 | A/D conversion channels 8-9 did not complete in 72 μS | A/D conversion timeout for channels 8-9 | critical |
69 | Abnormal battery recharging voltage | Battery charging voltage out of specification | critical |
7 | Communication alarm detected | Framing, overrun or parity error during communication with external computer | error |
70 | Data in duplicate memory areas does not agree | Memory check detects mismatch in duplicate storage | critical |
71 | Slave CPU received no data from Master CPU | Master CPU fails to send data to Slave | critical |
72 | Slave CPU received undefined data 4 times from Master CPU | Multiple reception of undefined data | critical |
73 | Overcurrent to motor | Motor current exceeds safe limit | critical |
74 | Motor slipped at maximum current | Motor unable to maintain engagement at max current | critical |
77 | No revolution interrupt while motor was rotating | Motor rotation not detected by encoder | critical |
78 | Drive defect - overinfusion | Motor speed exceeds expected revolution count | critical |
79 | Drive defect - underinfusion | Motor speed below expected revolution count | critical |
8 | Communication timeout alarm detected | No communication between device and external PC for given period | error |
82 | No acknowledgment from Slave CPU for three trials | Slave CPU fails to acknowledge after three attempts | critical |
83 | Undefined message type received from Slave CPU | Invalid message received from Slave CPU | critical |
86 | LVDT reading of downstream occlusion sensor too low | Downstream occlusion sensor LVDT below threshold | warning |
87 | LVDT reading of downstream occlusion sensor too high | Downstream occlusion sensor LVDT above threshold | warning |
89 | Bank switching control circuit failure | Bank switching circuit malfunction detected | critical |
9 | Slide Clamp Alarm | Slide clamp not inserted or improperly inserted (software 1.09+) | warning |
90 | Checksum error of RATE, VTBI and VI data | Data checksum verification failed | critical |
91 | Checksum error of Programmed Delivery Profile stored data | PDP stored data checksum verification failed | critical |
92 | Checksum error of Programmed Delivery Profile working data | PDP working data checksum verification failed | critical |
93 | M5M5165 write data and read data do not agree | Memory write/read verification failed | critical |
94 | Configuration option settings checksum is not 0 | Configuration settings checksum verification failed | critical |
95 | Checksum error of alarm log data | Alarm log data checksum verification failed | critical |
96 | Checksum error of ROM code | ROM code checksum verification failed | critical |
99 | Undefined failure code received from Slave CPU | Invalid failure code from Slave CPU | critical |
AIR | Air bubble detected in infusion set | Ultrasonic sensor detects air bubble | unknown |
BATTERY LOW | Battery charge level critically low | Battery charge drops below minimum operating level | warning |
BATTERY LOW (critical) | Battery power exhausted | Battery depleted after initial 15-minute warning | critical |
C03 | NIBP kinked tubing | Internal tubing kink detected in NIBP system | warning |
C04 | NIBP disconnected tubing or module fault | NIBP tubing disconnected or module failure detected | warning |
CHECK SET/LOADING | Tubing pinched in pump door | Tubing detected between door and base plate | warning |
COMMON FAILURE with code number | Device-level microprocessor malfunction detected | Dual-pump system detects fault; code number identifies specific failure | critical |
COM TIMEOUT | No pump-computer communication | Communication timeout with computer | warning |
DOOR OPEN | Pump door not properly closed | Door sensor detects open position or improper latch | warning |
EXT COMM ERROR | Computer communication protocol error | Computer sends multiple commands without waiting for replies | warning |
FAILURE | Microprocessor detected pump malfunction | Hardware or software fault detected | error |
FAILURE with code number | Microprocessor detected pump malfunction | Device detects internal fault; code number identifies specific failure | critical |
INSERT SLIDECLAMP | Slide clamp not detected in slide clamp slot or slide clamp sensor unable to confirm insertion | Primed administration set loaded into pump mechanism without slide clamp inserted into slide clamp slot, or slide clamp sensor contaminated or not detecting clamp presence | warning |
INSERT SLIDECLAMP | Slide clamp not loaded | Slide clamp loading enabled but clamp not in slot | warning |
OCCLUSION | Downstream blockage detected | Increased back-pressure sensed between patient and pump | unknown |
UPSTREAM OCCLUSION | Upstream blockage detected | Closed clamp or blockage upstream of pump sensed | unknown |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.