Baxter Colleague: troubleshooting and error codes

The ServiceForge knowledge base holds 108 documented symptoms for the Baxter Colleague, with 183 ranked causes, 141 diagnostic checks with test points and expected values, and 397 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Configuration transfer between pumps unsuccessful

Cannot transfer configuration data between two pumps

Ranked causes

  1. 1. Cable connections between pumps loose or disconnected (common)
  2. 2. Incompatible UIM Master software versions (common)
  3. 3. Configuration transfer cable continuity failed (occasional)
  4. 4. COMM port or internal connector fault (occasional)
  5. 5. COMM port dirt or contamination (occasional)
  6. 6. RS232 connector damaged (J12 on UIM PCB) (rare)
  7. 7. COMM port board faulty (rare)
First check

Verify cable connections

1 more check with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery icon lit when pump plugged into AC

Battery icon illuminated while pump connected to AC power; AC icon off indicating power issue

Ranked causes

  1. 1. AC power not available or external fuses blown (common)
  2. 2. Power cord defective (occasional)
  3. 3. Power Supply Module fuse blown (occasional)
  4. 4. Power Supply Module failure (occasional)
  5. 5. UIM PCB interconnection failure (occasional)
  6. 6. UIM PCB failure (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump cannot be turned off patient safety

ON/OFF CHARGE key press does not power down pump; may be related to UIM software or hardware issues

Ranked causes

  1. 1. UIM slave software version obsolete (occasional)
  2. 2. Front panel locked or manual tube release not reset (common)
  3. 3. Batteries just reconnected or first power-on after replacement (occasional)
  4. 4. Keypad interconnection failure (occasional)
  5. 5. Keypad failure (occasional)
  6. 6. UIM PCB failure (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

One or more keypad key presses are not accepted

Keypad buttons do not respond to operator input; specific keys not functioning

Ranked causes

  1. 1. Front panel locked (common)
  2. 2. Key function not available for current PERSONALITY feature set (common)
  3. 3. Keypad malfunction (occasional)
  4. 4. UIM PCB interconnection failure (occasional)
  5. 5. UIM PCB failure (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AIL PCB replacement indication

Conditions that require replacing the AIL PCB during calibration and verification procedures.

Ranked causes

  1. 1. Air In Line Sensor reading out of specification after calibration adjustment attempts (rare)
  2. 2. Empty tubing Air In Line Sensor value exceeds 128 at verification (occasional)
  3. 3. Fluid-filled tubing at 0 psig reads below 189 at verification (occasional)
  4. 4. Pressurized tubing Air In Line Sensor value exceeds 243 at verification (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Back light does not light when powered on

Display back light does not illuminate when pump powered on; display may be readable in bright light

Ranked causes

  1. 1. Back Light circuit disconnected or contacts loose (common)
  2. 2. CCFT assembly failure (occasional)
  3. 3. Inverter PCB failure (occasional)
  4. 4. UIM PCB failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Charge Level Indicator does not show increased charge after extended charging

Battery not charging despite extended charging period; AC icon displayed

Ranked causes

  1. 1. AC outlet not providing continuous power (common)
  2. 2. Battery fuse faulty (common)
  3. 3. Battery Charge Level Indicator calibration error (occasional)
  4. 4. Batteries defective (occasional)
First check

Check power outlet

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Contrast control has little or no effect

Rotating contrast control does not noticeably change display brightness or contrast; normal operation has slippage at ends

Ranked causes

  1. 1. Contrast control switch connection issue (common)
  2. 2. Display interconnection failure (occasional)
  3. 3. Display Module failure (occasional)
  4. 4. UIM PCB failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Downstream occlusion detected patient safety

Alarm 'Downstream Occlusion' displays; pump module shows 'DWN OCCL/Label'; red Alarm LED flashes; alarm tone sounds; infusion stops

Ranked causes

  1. 1. Distal clamp closed or stopcock occlusion (common)
  2. 2. Clogged filter or occluded tubing downstream (occasional)
  3. 3. Inappropriate occlusion pressure setting (occasional)
  4. 4. Pressure sensor malfunction (rare)
First check

Inspect administration set for closed clamps

Test point
Distal clamp, stopcock
Expected
All clamps and stopcocks open

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 715:xx Manual Tube Release Reset Time-out

Manual tube release reset operation times out; pump does not complete MTR cycle

Ranked causes

  1. 1. Overdrilled camshaft (rare)
  2. 2. MTR knob set screw binding plunger (occasional)
  3. 3. Broken encoder flag on tube loading motor (occasional)
  4. 4. LED/Sensor wires pinched on encoder PCB (occasional)
First check

Verify manual tube release tab is fully closed

Test point
Manual tube release tab
Expected
Tab fully closed after reset

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 803:07 Invalid slide clamp position sensor

Pump cannot determine slide clamp position; sensor reports invalid or inconsistent status

Ranked causes

  1. 1. Blue slide clamp inadvertently left in channel (common)
  2. 2. Flex cable J4 damaged or not seated (occasional)
  3. 3. Contamination or misalignment of sensor prism (occasional)
  4. 4. Pump mechanism sensor defect (rare)
First check

Verify no blue slide clamp in tubing channel

Test point
Tubing channel interior
Expected
No slide clamp visible

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PANEL LOCKOUT button presses are not accepted

PANEL LOCKOUT button does not respond when pressed to lock/unlock front panel

Ranked causes

  1. 1. PANEL LOCKOUT not available from current screen (common)
  2. 2. PANEL LOCKOUT button malfunction (occasional)
  3. 3. PANEL LOCKOUT switch harness discontinuity (occasional)
  4. 4. UIM PCB failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump cannot be turned on when plugged into AC outlet

AC power available but pump will not power on; AC icon lit

Ranked causes

  1. 1. Keypad contact issue or Manual Tube Release not fully closed (common)
  2. 2. Main batteries discharged or defective (common)
  3. 3. Keypad interconnection or malfunction (occasional)
  4. 4. UIM PCB interconnection or UIM PCB failure (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air bubble detected patient safety

Alarm 'Air Detected' displays with pop-up window; red Alarm LED flashes; alarm tone sounds; pump module shows 'AIR/Label'; infusion stops

Ranked causes

  1. 1. Air bubble in tubing larger than configured threshold (common)
  2. 2. AIL PCB miscalibration (occasional)
  3. 3. Pump mechanism failure (rare)
First check

Visually inspect tubing for air bubbles

Test point
Administration set tubing
Expected
No visible air bubbles or bubbles smaller than configured threshold

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible tone on constantly with no AC power

Continuous alarm beeping with pump operating on batteries only; AC icon off; critical alarm condition

Ranked causes

  1. 1. Failure code generated or abnormal battery condition (common)
  2. 2. Main batteries depleted or defective (common)
  3. 3. Pump will not power off normally (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery icon off while powered by main batteries

Battery icon does not illuminate when pump running on internal batteries; pump unplugged from AC

Ranked causes

  1. 1. UIM PCB to LED PCB interconnection failure (occasional)
  2. 2. LED PCB assembly failure (occasional)
  3. 3. UIM PCB failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 719:xx Manual tube release not fully closed

UIM detects manual tube release not fully closed at power-up; pump head unable to operate

Ranked causes

  1. 1. UIM to PHM ribbon cable not seated properly (common)
  2. 2. Manual tube release switch stuck or binding (occasional)
  3. 3. J12 flex cable damaged or unseated (occasional)
First check

Verify UIM to PHM ribbon cable seated at UIM PCB

Test point
UIM PCB ribbon cable connector
Expected
Cable fully seated, no gaps

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 720:xx UIM software communication timeout

UIM master software fails to receive response from PHM to failure message query; communication timeout

Ranked causes

  1. 1. Corrosion on UIM to PHM cable connector contacts (common)
  2. 2. PHM PCB failure or defect (occasional)
  3. 3. UIM PCB failure or defect (occasional)
First check

Inspect UIM to PHM cable connector contacts for corrosion

Test point
UIM to PHM cable connectors
Expected
Contacts clean, no discoloration

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

F3 fuse blown on UIM PCB caution

Fuse F3 protection circuit opened

Ranked causes

  1. 1. Power cable connector oriented incorrectly at P2 (common)
  2. 2. Loose metal debris causing short (occasional)
  3. 3. F3 fuse failed (common)
First check

Verify power connector orientation

Test point
Connector P2 on UIM PCB
Expected
Keyed connector properly oriented

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main display blank when powered on

Display screen appears all black or white with no graphics or text displayed after pump powered on

Ranked causes

  1. 1. Contrast Control set incorrectly (common)
  2. 2. Back Light off (common)
  3. 3. Display Assembly failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump cannot be turned on when powered by main batteries

Pump will not power on using internal batteries; AC Icon not lit indicating no AC power available

Ranked causes

  1. 1. Keypad contact issue or manual tube release not fully closed (common)
  2. 2. Main batteries discharged or defective (common)
  3. 3. Main battery fuse or harness failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

R118 resistor burnt on UIM PCB (underside below U66 and U4) caution

Resistor burnt due to improper fuse handling or battery connector grounding

Ranked causes

  1. 1. Fuse F103 not fully seated in holder (common)
  2. 2. Battery PCB connector contacted grounded surface (common)
  3. 3. R118 resistor damaged (common)
First check

Inspect fuse F103 seating

Test point
Fuse holder contacts
Expected
Both contacts fully within holder

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible tone on constantly with AC power patient safety

Continuous beeping or alarm tone heard while AC power connected; may or may not display failure code

Ranked causes

  1. 1. Failure code generated or manual tube release activated (common)
  2. 2. Unresolvable alarm condition (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery low caution

Main batteries discharged to 11.8 V; yellow Alert LED lights; alert tone sounds; status line displays 'Battery Low. Plug in Now!' and pump module shows 'BATT LOW'

Ranked causes

  1. 1. Main batteries discharged (common)
  2. 2. Fuses F1 or F2 open on battery harness (occasional)
First check

Measure main battery voltage

Test point
Battery terminals
Expected
11.8 V or lower

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad does not function

Pump keypad unresponsive to button presses

Ranked causes

  1. 1. Panel Lockout button inadvertently pressed (common)
  2. 2. Keypad malfunction or interconnection failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lithium backup battery low or disconnected

Alert 'Lithium Battery Low' displays; pump module shows 'BATT LOW'; yellow Alert LED lights

Ranked causes

  1. 1. Lithium backup battery charge depleted (occasional)
  2. 2. Lithium backup battery disconnected (occasional)
First check

Inspect lithium battery connection

Test point
Lithium backup battery connector
Expected
Battery connected and fully seated

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Manual tube release activated

RESET alarm when manual tube release (MTR) was activated; may indicate MTR snap action switch malfunction or slave microprocessor not responding

Ranked causes

  1. 1. MTR snap action switch not resetting or malfunctioning (occasional)
  2. 2. Slave microprocessor in UIM not responding (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse Call function does not work

Nurse Call system not responding

Ranked causes

  1. 1. Dirt or contamination in COMM port (common)
  2. 2. Nurse Call cable improperly wired (common)
First check

Perform Nurse Call Relay Test

Expected
Relay operates within tolerance; measured resistance values stabilize

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated failure codes 803:02, 803:03, or 803:09 caution

Pump module experiences repeated failure codes indicating possible tube loading motor drive gear or motor-to-gearbox fastening looseness

Ranked causes

  1. 1. Tube loading motor drive gear is loose (unknown)
  2. 2. Tube loading motor-to-gearbox fastening is loose (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tube loading in progress

Administration set not fully loaded in tubing channel when Start key pressed, or obstruction present

Ranked causes

  1. 1. Administration set not fully loaded (common)
  2. 2. Obstruction in tubing channel (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tube misloaded

Administration set improperly loaded or not fully removed from tubing channel, or hardware failure occurred

Ranked causes

  1. 1. Improperly loaded or incompletely removed administration set (common)
  2. 2. Hardware failure or obstruction (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tube not loaded

Administration set was not loaded prior to pressing START key

Ranked causes

  1. 1. Administration set not loaded before START key pressed (common)
  2. 2. Pump mechanism malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air in Line sensor value out of range patient safety

Air In Line Sensor reading does not meet specification during unpressurized, pressurized, or air-filled tubing test phases

Ranked causes

  1. 1. AIL PCB sensor drift or calibration error (occasional)
First check

Load fluid-filled unpressurized tubing

Test point
Pump channel with unused section
Expected
STOPPED message displayed; Air in Line Sensor value ≥ 190 counts, ≥190

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED malfunction

Alarm LED remains lit continuously during alarm or does not light when alarm occurs

Ranked causes

  1. 1. Pump head module keypad or flex cable issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm tone volume does not respond to volume control

Speaker test tone volume does not decrease or increase when volume control on pump handle is adjusted

Ranked causes

  1. 1. Volume control mechanical failure or loss of contact (occasional)
First check

Press Test Tone On soft key

Test point
Speaker output
Expected
Vaudio displays value; alarm tone sounds

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alert LED malfunction

Alert LED remains lit continuously or does not light when alert condition occurs

Ranked causes

  1. 1. Pump head module keypad or flex cable issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Back light toggle does not respond

Back Light key press does not turn light on or off as expected; back light always on or always off

Ranked causes

  1. 1. Keypad malfunction (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup beeper fails self-test

Backup beeper does not sound two distinct times during self-test

Ranked causes

  1. 1. Beeper or beeper circuit malfunction (unknown)
First check

Verify backup beeper during self-test

Test point
Beeper
Expected
Backup beeper sounds two distinct times

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup beeper tone constantly on with failure code

Continuous beeping, failure code displayed on Main Display, and FAILURE shown on pump head module displays

Ranked causes

  1. 1. Failure code generated requiring troubleshooting (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup beeper tones not heard during power-on

No beeping sound during power-on self-test sequence; Speaker Test soft key does not produce audible tone

Ranked causes

  1. 1. Speaker or backup beeper hardware failure or interconnection issue (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charging after UIM PCB replacement patient safety

After UIM PCB replacement, batteries must be charged and verified to reach full capacity with Amp Hours Left display at 3.800 before returning pump to service.

Ranked causes

  1. 1. New batteries require full charge cycle after installation (common)
First check

AC and Battery icons after plugging in

Test point
main display icons
Expected
AC (plug) icon lights; Battery icon extinguishes

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery completely depleted caution

No display on main screen; pump shows no display on module; 3-beep alarm tone from backup beeper; infusion stopped; alarm generated when Battery Depleted alarm persisted for longer than 3 minutes

Ranked causes

  1. 1. Main batteries fully discharged to 10.8 V or below (occasional)
First check

Power pump off and connect to AC power

Test point
AC power connection
Expected
Pump powered off and plugged into AC supply

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery depleted during minimum runtime test patient safety

Pump experiences Battery Depleted - Device Stopped alarm within 2.5 hour minimum test period (2 hours for COLLEAGUE 3) at 100 mL/hr rate with 9999 mL VTBI

Ranked causes

  1. 1. Defective or degraded main battery (common)
First check

Run battery discharge test at constant rate

Test point
Pump infusion system
Expected
No Battery Depleted - Device Stopped alarm within 2.5 hour period (2 hours COLLEAGUE 3) at 100 mL/hr, 9999 mL VTBI
Tools
Primed administration set, distilled water container, stopwatch

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery history lost during UIM PCB replacement

Disconnecting main batteries and backup battery during UIM PCB replacement results in loss of battery history data.

Ranked causes

  1. 1. Main batteries and battery harness must be replaced with UIM PCB (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery icon lights momentarily during self-test

Battery icon illuminates only during self-test sequence when pumped powered by main batteries; remains off during normal operation

Ranked causes

  1. 1. UIM PCB failure (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Low alert timing or duration incorrect patient safety

Battery Low. Plug in Now! alert does not occur, or pump fails to continue for 30 minutes after alert starts

Ranked causes

  1. 1. Defective or degraded backup battery or battery monitoring circuit (common)
First check

Continue infusion until Battery Low alert and verify continuation

Test point
Pump battery system and infusion output
Expected
Battery Low. Plug in Now! alert occurs after 2.5+ hours; pump continues infusion for 30 min after alert without Device Stopped alarm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

BCLI shows full but Amp Hours Left not 3.800

Battery Charge Level Indicator displays full but Amp Hours Left value does not reach 3.800, indicating possible BCLI calibration error.

Ranked causes

  1. 1. BCLI calibration drift or incorrect reference (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Calibration Step 14

Verify AIL PCB calibration value at Step 14 is 230±1 for all pump modules

Ranked causes

  1. 1. AIL PCB Step 14 calibration value out of range (unknown)
First check

Record AIL PCB Step 14 calibration value for pump modules A, B, C

Test point
AIL PCB calibration point Step 14
Expected
230±1 raw units, ±1

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Calibration Step 17

Verify AIL PCB calibration value at Step 17 is greater than 189 for all pump modules

Ranked causes

  1. 1. AIL PCB Step 17 calibration value below threshold (unknown)
First check

Record AIL PCB Step 17 calibration value for pump modules A, B, C

Test point
AIL PCB calibration point Step 17
Expected
>189 raw units, >189

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Calibration Step 18

Verify AIL PCB calibration value at Step 18 is less than 238 for all pump modules

Ranked causes

  1. 1. AIL PCB Step 18 calibration value exceeds maximum (unknown)
First check

Record AIL PCB Step 18 calibration value for pump modules A, B, C

Test point
AIL PCB calibration point Step 18
Expected
< 238 raw units, <238

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Calibration Step 21

Verify AIL PCB calibration value at Step 21 is less than 129 for all pump modules

Ranked causes

  1. 1. AIL PCB Step 21 calibration value exceeds maximum (unknown)
First check

Record AIL PCB Step 21 calibration value for pump modules A, B, C

Test point
AIL PCB calibration point Step 21
Expected
< 129 raw units, <129

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Step 18 minus Step 17 difference

Verify difference between Step 18 and Step 17 values is greater than 4

Ranked causes

  1. 1. AIL PCB Step 18 minus Step 17 difference not within specification (unknown)
First check

Calculate difference: Step 18 value minus Step 17 value

Test point
AIL PCB Steps 17 and 18
Expected
> 4

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Verification Step 2

Verify AIL PCB verification value at Step 2 is less than 128 for all pump modules

Ranked causes

  1. 1. AIL PCB verification Step 2 value exceeds limit (unknown)
First check

Record AIL PCB verification Step 2 value for pump modules A, B, C

Test point
AIL PCB verification point Step 2
Expected
< 128 raw units, <128

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Verification Step 5

Verify AIL PCB verification value at Step 5 is greater than 189 for all pump modules

Ranked causes

  1. 1. AIL PCB verification Step 5 value below threshold (unknown)
First check

Record AIL PCB verification Step 5 value for pump modules A, B, C

Test point
AIL PCB verification point Step 5
Expected
>189 raw units, >189

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AIL PCB Verification Step 7

Verify AIL PCB verification value at Step 7 is less than 243 for all pump modules

Ranked causes

  1. 1. AIL PCB verification Step 7 value exceeds maximum (unknown)
First check

Record AIL PCB verification Step 7 value for pump modules A, B, C

Test point
AIL PCB verification point Step 7
Expected
< 243 raw units, <243

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery Charge Level Indicator

UIM PCB battery charge level indicator requires calibration after replacement or when displaying negative current or values >25 mA with verified good batteries.

Ranked causes

  1. 1. UIM PCB replacement or BCLI loss of calibration (common)
First check

Verify pump powered off for minimum 2 hours

Test point
Pump electrical system
Expected
Minimum 2 hours elapsed

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery Charge Level Indicator (BCLI) Charge Current

Verify battery charge current diagnostics during BCLI calibration setup

Ranked causes

  1. 1. BCLI charge current out of calibration (unknown)
First check

Verify batteries are fully charged before BCLI calibration

Test point
Battery charging state
Expected
Batteries charged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: BCLI Closed Bezel High Value

Verify calibration high value is 501 to 522 during BCLI closed bezel calibration

Ranked causes

  1. 1. BCLI high calibration value out of range (unknown)
First check

Record calibration high value at Step 6

Test point
BCLI closed bezel high calibration point
Expected
501 to 522 raw units, 501–522

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: BCLI Closed Bezel Low Value

Verify stabilized calibration low value is 480 to 547 during BCLI closed bezel calibration

Ranked causes

  1. 1. BCLI low calibration value out of range (unknown)
First check

Record stabilized calibration low value at Step 4

Test point
BCLI closed bezel calibration point
Expected
480 to 547 raw units, 480–547

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: BCLI Value After Adjustment (Step 5)

Verify BCLI value after adjustment is 505 to 507

Ranked causes

  1. 1. BCLI adjusted value out of specification (unknown)
First check

Verify BCLI value after adjustment at Step 5

Test point
BCLI measurement point after adjustment
Expected
505 to 507 raw units, 505–507

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Calibration: BCLI Value After Adjustment (Step 8)
  • Calibration: Power Supply
  • Calibration: Power Supply output
  • Calibration: Power Supply Output Voltage
  • Channel Emergency Open not functioning
  • Charge Current Diagnostics - BCLI calibration verification
  • Damaged battery
  • Damaged Battery alarm after UIM PCB replacement
  • Device failure code 803:07, 865:01, or 865:19
  • Display fails self-test
  • External rattle or loose components detected
  • Flat ribbon cable ferrite ring stress damage
  • Flat ribbon cable J5 connector locked flange damage risk
  • Icon-encoded UIM PCB failures
  • Keypad not responding
  • LED or icon fails self-test
  • LED/icon PCB connector P15 orientation error
  • Loss of memory backup power
  • Manual tube release reset alarm on power-up
  • Motor encoder PCB wheel rotation check
  • Motor harness connector assembly
  • Nurse Call relay not responding
  • Panel Lockout button not responding
  • Piggyback/primary value out of range
  • Plug icon fails to light when AC power connected
  • Power-up beeps out of sequence
  • Pump does not complete power-on sequence
  • Pump module display fails self-test
  • Pump module PCB mounting
  • Running LED malfunction
  • Shuttle motor encoder collar alignment
  • Slide clamp snags during tubing removal
  • Slide clamp stuck in pump head module slot
  • Speaker audio test after UIM PCB replacement
  • Speaker test tone not heard
  • Temperature too high
  • Temperature too low
  • Three-beep tone with no visual displays
  • Tube loading motor encoder collar alignment
  • Tube misloaded alarm occurs frequently
  • Tube misloaded alarm with no tubing
  • Tube misloaded when no tubing is loaded
  • UIM PCB replacement loses configuration data
  • UIM PCB self-test verification after replacement
  • Unintended shutdown or power on
  • Unit does not power on or Plug icon not lit
  • Upstream occlusion
  • Voltage sensor readings out of range

48 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
100:0:0:0000Defective RAM on UIM PCBUIM software detects defective RAMerror
12:303:984:xxxxSoftware failure - Communication packet timeoutTransmission not completed within 1 seconderror
12:323:xxx:xxxxSoftware failure - Excessive keypress events or panel lockout violationToo many keypresses in short time, or multiple manual tube release operations during PANEL LOCKOUT, or single-channel UIM PCB on triple-channel pumperror
12:502:xxx:xxxxUIM software failure - Configuration transfer version mismatchTransfer from v4.XX.XX then attempted transfer to v5.XX.XX pumperror
12:520:xxx:xxxxUIM software failure - Configuration transfer version mismatchTransfer from v4.XX.XX then attempted transfer to v5.XX.XX pumperror
12:xxx:xxx:xxxxSoftware failure - Invalid value in RAMRAM corruption detectederror
13:xxx:xxx:xxxxSoftware failure - Out of memoryMemory allocation exceededcritical
15:xxx:xxx:xxxxSoftware failure - CRC variable check failedCyclic Redundancy Check on safe variable failedcritical
16:200:xxx:xxxxUIM software failure - Duplicate label in configuration transferConfiguration transfer between v5.04 or earlier and v5.09+ with duplicate labelserror
16:336:xxx:0000Obsolete UIM software - Configuration transfer incompleteOne pump returns to normal mode while transfer cable still connectedcritical
16:xxx:xxx:xxxxSoftware failure - Memory failureMemory insert to list operation failedcritical
18:111:xxxx:xxxxObsolete UIM software - Configuration received from obsolete softwareConfiguration data received from pump with obsolete UIM Master softwareerror
18:115:1083:xxxx5.XX.XX UIM softwareRate values less in active PERSONALITY feature set than in functional test modeerror
18:115:885:xxxxConfiguration settingsUser defined PERSONALITY feature set with rate limit below 100 mL or volume limit below 20.0 mL as power on default, then entered Functional test modeerror
18:115:xxxx:xxxxObsolete UIM softwareConversion of patient weight from lb to kg not performed properlyerror
18:200:xxx:xxxxUIM softwareRAM U2 and U3 cannot be accessed by UIM microprocessorerror
18:407:1143:0000UIM softwareAttempting to view Label Library setup of PERSONALITY with Library Setup enabled but power-on default has Library Setup disablederror
18:416:xxx:xxxxUIM softwareRAM U2 and U3 cannot be accessed by UIM microprocessorerror
18:xxx:xxx:xxxxObsolete UIM software - Invalid configuration transferConfiguration transfer attempted using drone pump with obsolete softwareerror
19:113:xxx:xxxxUIM softwareRAM U2 and U3 cannot be accessed by UIM microprocessorerror
197:xxx:xxx:xxxxVerified Memory Attribute CorruptMemory attribute corruption detected at power uperror
198:xxx:xxx:xxxxVerified Memory PROM Upgrade or software incompatibilityPROM upgrade in progress or detected software incompatibilitywarning
199:xxx:xxx:xxxxVerified Memory CRC Check FailedPower-on following total loss of AC and battery powererror
1:xxx:xxx:xxxxSoftware failure - Array Out Of RangeMessage to UIM failurecritical
200:xxx:xxx:xxxxHardware initialization failure80186 CPU failure during hardware initializationerror
201:xxx:xxx:xxxxInternal CPU Timer 080186 CPU Timer 0 failure during operationerror
202:xxx:xxx:xxxxInternal CPU Timer 180186 CPU Timer 1 failure during operationerror
203:xxx:xxx:xxxxInternal CPU Timer 280186 CPU Timer 2 failure during operationerror
204:xxx:xxx:xxxxInternal CPU Serial 080186 CPU Serial 0 failure during operationerror
205:xxx:xxx:xxxxInternal CPU Serial 1 (RS-232 Port)80186 CPU Serial 1 failure or RS-232 port malfunctionerror
206:xxx:xxx:xxxxMSM62X42B Clock: Init RTCReal Time Clock initialization failureerror
207:xxx:xxx:xxxx16550 DUART: Init SerialDUART serial device initialization failureerror
21:xxx:xxx:xxxxSafe Variable FailedSoftware failure during power-up or operationerror
30180186 internal Diagnostic CPU Intr Divide Errorunknown
30280186 internal Diagnostic CPU Intr Unexpectedunknown
30380186 internal Diagnostic CPU Port 1unknown
30480186 internal Diagnostic CPU Port 2unknown
30580186 internal Diagnostic CPU Serial 0unknown
30680186 internal Diagnostic CPU Serial 1unknown
30780186 internal Diagnostic CPU Timer 0unknown
30880186 internal Diagnostic CPU Timer 1unknown
30980186 internal Diagnostic CPU Timer 2unknown
310UPD71055 Parallel port Diagnostic CPU Port 0unknown
311UPD71055 Parallel port Diagnostic CPU Port 1unknown
312Ongoing ROM Check Diagnostic ROM CRCerror
313MSM62X42B Clock Diagnostic RTC 1 Hzerror
313:425MSM62X42B Clock - Poor continuityerror
314Software Diagnostic real time heap is corruptederror
314:432:201:0000Software Data heap corruptederror
315:423:514:1111Software Diagnostic Real Time Kernel Stack Overflowerror
316RAM Diagnostic Safe Var Corruptederror
31716550 DUART Diagnostic Serial Aerror
320MSM62X42B Clock Diagnostic RTCerror
321Obsolete software or manual tube release faultwarning
36:416:xxx:xxxxUIM softwareRAM U2 and U3 cannot be accessed by UIM microprocessorerror
38:xxx:xxx:xxxxInvalid StateEntering Configuration/Service Menu after pump module failure or during module malfunctionerror
39:xxx:xxx:xxxxOut of TimersTimer allocation exhausted during operationerror
400Slave Communication Failurecritical
401Slave Communication Init Failcritical
402Slave Communication Nak Failcritical
403Slave Communication Time-out Failcritical
40:xxx:xxx:xxxxInvalid allocation of memoryMemory allocation failure during operationerror
41:113:xxx:xxxxUIM softwareRAM U2 and U3 cannot be accessed by UIM microprocessorerror
49:121:xxxx:0003PHM SoftwareCOLLEAGUE 3 Pumps: PHM software versions not the same for all three pump head moduleserror
49:121:xxxx:0004Incorrect UIM Master software versionAttempt to use COLLEAGUE GUARDIAN Enhanced Feature software (version 6.XX.XX) in a monochrome pumperror
49:121:xxxx:xxxxObsolete/incompatible softwareIncompatibility among UIM master, UIM slave, or PHM software versionserror
4:xxx:xxx:xxxxSoftware failure - Selection Statement DefaultInvalid selection stateerror
500Stuck key detected (any key except ON/OFF CHARGE)warning
501Stuck key detected (ON/OFF CHARGE key)warning
502Invalid battery voltage or current datawarning
503Beep Enable lines are the same or improper power down sequenceerror
507General keypad read failurewarning
50:xxx:xxx:xxxxFloating point exception processing errorFloating point arithmetic or exception processing failureerror
511VKey power supply failurecritical
512Slave Main Power Supply or main batteries out of rangecritical
513Main Power Supply voltage too highcritical
514:xxx:xxx:xxxxA/D low true RMS current readings badBattery Charge Level Indicator A/D conversion failureerror
515:xxx:xxx:xxxxA/D high true RMS current readings badBattery Charge Level Indicator A/D conversion failureerror
516:xxx:xxx:xxxxA/D high true RMS voltage readings badBattery Charge Level Indicator A/D conversion failureerror
517:xxx:xxx:xxxxA/D low true RMS voltage readings badRepeated switching between AC and battery powererror
520:320:xxx:0000Slave memory corruptedCal Lo or Cal Hi keys pressed and held instead of releasedcritical
522:xxx:xxx:xxxxSpeaker and/or speaker circuit failureAudio circuit malfunction detectedwarning
526:xxx:xxx:xxxxMemory corrupted or channel count mismatchMemory corruption or configuration mismatch with power-up notificationerror
528:xxx:xxx:xxxxCould not obtain valid readingsPower supply or CPU A/D conversion failureerror
529:xxx:xxx:xxxxThree or more errors communicating with UIM MasterMaster/slave communication noise or failureerror
530:xxx:xxx:xxxxSpeaker and/or speaker circuit noisyAudio circuit noise detectedwarning
531:xxx:xxx:xxxxSlave memory corrupted from cascade failureDC power disconnected from UIM PCB while plugged into ACcritical
532:xxx:xxx:xxxxNo response from Master after message sentMaster/slave communication timeouterror
533:xxx:xxx:xxxxExcessive Nak's from MasterMaster/slave communication protocol failureerror
534:xxx:xxx:xxxxSlave timed out waiting for message from MasterMaster/slave communication timeouterror
535:xxx:xxx:xxxxNak received on status packetMaster/slave communication protocol failureerror
536:xxx:xxx:xxxxBad Image CRCROM firmware checksum failureerror
537:xxx:xxx:xxxxCoarse voltage reading badPower supply A/D conversion failureerror
538:xxx:xxx:xxxxSuccessive bad A/D readings from audio circuitAudio circuit sensor or A/D conversion failurewarning
539:xxx:xxx:xxxxBad readings from slave power supply A/D channelPower supply A/D conversion failureerror
540:xxx:xxx:xxxxA/D invalid VMem readingsPower supply A/D conversion failureerror
541:xxx:xxx:xxxxA/D VKey readings are badPower supply A/D conversion failureerror
542:xxx:xxx:xxxxCould not obtain valid readings from power supplyPower supply or CPU A/D conversion failureerror
543:xxx:xxx:xxxxTrue RMS and coarse voltage readings out of rangeBattery charge level indicator sensor failureerror
545:xxx:xxx:xxxxHigh true RMS voltages out of rangeBattery charge level indicator sensor failureerror
546:xxx:xxx:xxxxMaster/slave communication failure due to noiseCommunication circuit noise or interferenceerror
548:xxx:xxx:xxxxOn/off switch low at power downOn/off circuit malfunction at shutdownerror
549:xxx:xxx:xxxxSpeaker/audio circuit failure after battery deep dischargeBatteries discharged below 10.4 Verror
550:121:xxx:xxxxUser pressed No during power-up speaker testUser action during boot or possible speaker failurewarning
550:xxx:xxx:xxxxSpeaker voltage too low during audible periodSpeaker power supply failure or connection issuewarning
551:xxx:xxx:xxxxOn/off power switch monitor line high at wrong timeOn/off circuit malfunctionerror
552:xxx:xxx:xxxxStuck key or faulty on/off charge circuitKeypad or charge circuit malfunctionerror
559:xxx:xxx:xxxxOpen keys cannot be readKeypad circuit failureerror
567:xxx:xxx:xxxxFailure to complete power-up manual tube release testPower-up test sequence failureerror
568:xxx:xxx:xxxxSpeaker voltage low during alertSpeaker power supply failure or connection issuewarning
570Failure code when battery critically lowMain battery voltage drops below 10.0 VDCcritical
570:xxx:xxx:xxxxBatteries discharged to damaging levelLow battery voltage at power-uperror
589:xxx:xxx:xxxxFailed to shut down after battery depletion alarmPump unable to power off within three minutes of battery alarmcritical
599:320:xxxx:0000Transfer cable disconnected or abort during configuration transferConfiguration transfer interruptedwarning
599:xxx:xxx:xxxxMaster not provided valid service code from slave processorSlave processor communication failureerror
6:xxx:xxx:xxxxSoftware failure - Null Pointer AssignmentNULL pointer errorcritical
701:00Vmain DC power trace open circuit on UIM PCBConnector mismatch or disconnection at P2 or P8critical
701:xxPHM communication initiate failPHM communication circuit failureerror
702:xxx:xxx:xxxxPHM communication Nak failPHM communication protocol failureerror
703:xxx:xxx:xxxxPHM communication time-out failPHM communication timeouterror
704:xxxPHM Communications, PHM Comm Transmit Queue OverrunUIM to PHM communication queue overflow; too many messages queuederror
705:xxPHM Communications, PHM Rate MismatchPump rate data does not match between UIM and PHM moduleserror
706:00UIM software, PHM invalid switch overUIM software detects invalid state during PHM switchover operationerror
712:xxUIM/PHM software, No ack received from PHM after START pressedOperator presses START on pump but PHM fails to acknowledge commanderror
714:xxUIM master, PHM Software Communications, Jaws Open DeniedPump cannot open clamp mechanism when requestedwarning
715:xxUIM Master, PHM Software Communications, Manual Tube Release Reset Time-outManual tube release cycle times out; MTR does not complete reset within expected timeerror
719:xxManual tube release, UIM check indicates MTR not fully closed at power-upDuring power-up, UIM detects manual tube release not in fully closed position; may indicate PIU software mismatch or MTR activation during pump operationerror
720:xxSoftware, UIM master timeout to PHM failure message queryUIM sends failure message query to PHM but does not receive response within timeout perioderror
721:00Software, Real time clock problem detectedPump channel programmed for Delay Start; real time clock detected as faultyerror
7:xxx:xxx:xxxxSoftware failure - Too Many Enable LabelsLabel limit exceedederror
800:01System Error, CRC EEPROM check errorCyclic Redundancy Check fails on EEPROM data during self-testcritical
800:03System Error, Display Test ErrorDisplay self-test fails during startup or diagnosticcritical
801:04Pump Mechanical Error, Close FailedPump mechanism cannot close/compress tubing when commandedcritical
801:05Pump Mechanical Error, Power down state time outPump does not reach safe shutdown state within expected timecritical
801:06Pump Mechanical Error, Start attempted without tube or with keyed slide clamp in placeOperator attempts to start pump infusion without tubing loaded or with slide clamp engagedcritical
801:07Pump Mechanical Error, REGNO: Infusion register out of rangeInternal infusion register value exceeds valid rangeerror
801:08Pump Mechanical Error, VOLUMEZERO: Infusion register 0 VTBIVolume to be infused (VTBI) register shows zero when pump was operatingerror
801:09Pump Mechanical Error, Proximal or distal time outPump sensors do not detect expected tubing positions within time limitscritical
801:10Pump Mechanical Error, Pressure sensor nulling state errorPressure sensor zeroing/nulling operation fails during initializationerror
801:12Pump Mechanical Error, Reached infuse state without start commandPump enters infusion mode without operator pressing STARTcritical
801:13Pump Mechanical Error, Simultaneous Open and Start commandPump receives conflicting commands (open clamp and start infusion) at same timecritical
801:14Pump Mechanical Error, Manual tube release switch state error (Use-induced: Tab out at Open key press)MTR switch in unexpected state; typically operator pulls MTR tab during Open commandwarning
801:15Pump Mechanical Error, Attempted start with alarm stateOperator pressed START while pump was in alarm conditioncritical
801:16Pump Mechanical Error, Local index failed with tube in placePump cannot establish position reference (local index) while tubing is loadederror
801:17Pump Mechanical Error, Pump Module Open key stuckOpen keypad button appears to remain pressed or signals continuouslyerror
801:18Pump Mechanical Error, Too long in pre-infusion statePump remains in pre-infusion state longer than expected before transitioningerror
801:19Pump Mechanical Error, KVO setting out of rangeKeep Vein Open (KVO) rate parameter is outside valid rangeerror
801:20Pump Mechanical Error, Too long in stop infusion statePump remains in stop state longer than expected during shutdown sequenceerror
801:41Pump Mechanical Error, Pump Module state variable corruptedPump state variable fails CRC or parity checkcritical
801:42Pump Mechanical Error, Pump Module events 1 corruptedEvent history data block 1 fails CRC checkerror
801:43Pump Mechanical Error, Pump Module events 2 corruptedEvent history data block 2 fails CRC checkerror
801:44Pump Mechanical Error, Auto Tube Load disable corruptedAuto Tube Load disable flag fails integrity checkerror
801:45Pump Mechanical Error, UIM Start Register Number corruptedUIM start register number fails CRC checkerror
801:46Pump Mechanical Error, KVO Rate corruptedKVO rate value fails integrity checkerror
801:47Pump Mechanical Error, Pump module state variable corruptedPump state variable fails integrity checkcritical
801:48Pump Mechanical Error, New Pump module state variable out of rangeUpdated pump state variable value is outside valid rangeerror
802:04Shuttle Motor Mechanical Error, Shuttle failed to center after infusionShuttle mechanism cannot return to center position after infusion cycle completescritical
802:05Shuttle Motor Mechanical Error, Shuttle failed to center after tube loadShuttle mechanism cannot return to center position after tubing is loadedcritical
802:06Shuttle Motor Mechanical Error, Shuttle failed to move to tube load positionShuttle cannot position itself for tubing loading operationcritical
802:07Shuttle Motor Mechanical Error, REGNO: Start register number out of rangeShuttle start register number value exceeds valid rangeerror
802:08Shuttle Motor Mechanical Error, Attempt to start without Start commandShuttle motor moves without explicit start command from operator or control softwarecritical
802:09Shuttle Motor Mechanical Error, Shuttle failed to recover local indexShuttle mechanism cannot establish home position reference pointcritical
802:10Shuttle Motor Mechanical Error, Move to center from fill failedShuttle cannot move from fill position to center positioncritical
802:11Shuttle Motor Mechanical Error, Variable 1: State variable corruptedShuttle state variable fails integrity checkcritical
802:12Shuttle Motor Mechanical Error, Variable 2: Event flag variable corruptedShuttle event flag variable fails CRC checkcritical
802:13Shuttle Motor Mechanical Error, Variable 3: Start register number corruptedShuttle start register number fails integrity checkerror
802:14Shuttle Motor Mechanical Error, Shuttle move failure during tube recoveryShuttle cannot move properly when retracting tubingcritical
802:15Shuttle Motor Mechanical Error, Shuttle Joggle incorrect state errorShuttle joggle movement (fine adjustment) operation produces incorrect resulterror
802:16Shuttle Motor Mechanical Error, Shuttle Joggle movement failureShuttle joggle (fine adjustment) movement does not execute properlyerror
802:20Shuttle Motor Mechanical Error, No tube valve closed test failedInlet or outlet valve fails closure test when no tubing is loadederror
802:99Shuttle Motor Mechanical Error, State: State machine incorrect state errorShuttle state machine enters or detects invalid statecritical
803:01Tube Load Mechanical Error, Tubing channel not closedPump cannot fully close tubing channel when commandedcritical
803:02Tube Load Mechanical Error, Slide clamp has not moved to open position (Use-induced: manual tube release tab out while loading)Slide clamp remains in closed position when should be opening for tubing insertion; operator pulls MTR tab during tube loadingwarning
803:03Tube Load Mechanical Error, Slide Clamp not closed (Use-induced: Proximal tubing pulled while unloading)Slide clamp remains in open position when should be closing after tubing removal; operator pulls tubing before clamp fully closeswarning
803:04Tube Load Mechanical Error, Load Motor Overload (Use-induced: Proximal tubing pulled or manual tube release tab out during loading)Tube loading motor current exceeds safe operating limit; typically caused by mechanical binding from operator actionerror
803:05Tube Load Mechanical Error, Load Motor index time out (Use-induced: manual tube release tab out at power up)Tube loading motor fails to reach home position within expected time; typically after MTR tab pulled at startuperror
803:07Upstream under platen flex connector failure or sensor malfunctionFlex connector improperly seated in J4 or foiled end of ribbon cable damagederror
803:07Tube Load Mechanical Error, Invalid Slide Clamp Position/Sensor errorPump cannot determine slide clamp position; optical sensor reports invalid or inconsistent statuserror
803:08Tube Load Mechanical Error, Slide Clamp Sensor error at power up: No tubing channel closed testAt startup, tubing channel fails closure test and slide clamp sensor error detectederror
803:09Misload sensor value out of range (open)Manual tube release tab out during key press; PHM software version 3.01.00 or earlierwarning
803:11State variable corruptedPump mechanism state variable corruptionerror
803:12Event flag variable corruptedPump mechanism event flag corruptionerror
803:20Erroneous movement in tube loading mechanismUnauthorized motion in tube loaderwarning
803:21Overload impossibleTube loading motor overload detection failureerror
803:99State machine incorrect state errorState machine state variable corruption or faultcritical
804:21COMM ERR: 10 consecutive NAK'sCommunication NAK timeouterror
804:22COMM ERR: Packet without ACK or NAKPacket acknowledgment failureerror
804:23COMM ERR: Packet larger than limitOversized packet receivederror
804:24COMM ERR: Packet retry CRC errPacket CRC check failure on retryerror
804:25COMM ERR: Incorrect pump module numberPump module ID mismatcherror
804:26COMM ERR: Packet retry sequence errManual tube release used before speaker test completionwarning
804:27COMM ERR: Packet retry data size errPacket size mismatch on retryerror
804:28COMM ERR: Packet max data size exceededOversized packet dataerror
804:29COMM ERR: No ACK/NAK within 2 secondsAcknowledgment timeouterror
804:30COMM ERR: Received illegal commandInvalid command receivederror
804:31COMM ERR: No space to build bufferBuffer allocation failureerror
804:32COMM ERR: Transmit buffer out of spaceTransmit queue overflowerror
804:33COMM ERR: Power up UIM CRC incorrectUIM firmware checksum failure at power-uperror
804:34COMM ERR: Failed watchdog timer testWatchdog timer test failureerror
804:52COMM ERR: Serial communication frame errorSerial protocol frame errorwarning
804:54COMM ERR: Serial communication noise error flagSerial communication noise detectedwarning
805:01No bits left in nullingDistal pressure sensor nulling bit exhaustionerror
805:02Sensor did not settle in bandDistal pressure sensor unable to stabilizeerror
805:03Offset amplifier test failureDistal pressure offset amplifier self-test failureerror
805:04Call to null while renullingNull operation during renull sequenceerror
805:05Backup data corruption: base pressureBase pressure backup data checksum failureerror
805:06Backup data corruption: alarmAlarm backup data checksum failureerror
805:07Backup data corruption: pressurePressure backup data checksum failureerror
805:08Backup data corruption: threshold rangeThreshold range backup data checksum failureerror
805:09Pulse width expiredDistal pressure pulse generation timeouterror
805:10Gain slope garbageGain slope calibration data invaliderror
805:11Matrix CRC failureCalibration matrix checksum failureerror
805:12Failure during renullDistal pressure sensor renull operation failureerror
805:14Temperature compensation slope beyond limitsTemperature compensation calibration out of rangeerror
805:15Occlusion sensor failureOcclusion detection sensor malfunctionerror
805:16Threshold set out of rangeThreshold setting outside valid rangeerror
805:17Argument: Read distal matrix elementDistal pressure matrix read argument errorerror
806:01No bits left to adjust offset during nullingProximal pressure sensor nulling bit exhaustionerror
806:02Sensor did not settle in bandProximal pressure sensor unable to stabilizeerror
806:03Offset amplifier test failureProximal pressure offset amplifier self-test failureerror
806:04No bits left to adjust offset during trackingProximal pressure tracking bit exhaustionerror
806:05Backup data corruption: alarmAlarm backup data checksum failureerror
806:06Backup data corruption: threshold settingThreshold setting backup data checksum failureerror
806:07Backup data corruption: alarm thresholdAlarm threshold backup data checksum failureerror
806:08Backup data corruption: cumulative alarm thresholdCumulative threshold backup data checksum failureerror
806:10Proximal null failureProximal pressure sensor null operation failureerror
806:11Illegal threshold from UIMInvalid threshold value from user interface moduleerror
806:12More data corruptionProximal sensor data corruptionerror
807:00Trend monitor error inside 0 pump phasePosition sensor trend monitoring failurewarning
807:01Trend monitor error inside 1 pump phasePosition sensor trend monitoring failureerror
807:02Position monitor error at midpoint (Valve 2) side 0Gross position sensor error at valve midpointerror
807:03Position monitor error at midpoint (Valve 2) side 1Gross position sensor error at valve midpointerror
807:04Position monitor error at valve 1 side 0Gross position sensor error at valve 1error
807:05Position monitor error at valve 1 side 1Gross position sensor error at valve 1error
807:06Data corruptionGross position sensor data checksum failureerror
807:07Excessive movement while indexingPosition sensor movement beyond expected rangeerror
807:xxIntermittent/damaged flex-circuit connectionsGross position sensor flex-circuit damageerror
808:01Outlet valve not open, position 1Outlet valve failure to open at position 1error
808:02Inlet valve not open, position 1Inlet valve failure to open at position 1error
808:03Valve sensors: one valve not closedValve remains open when should be closederror
808:04Inlet valve faultyInlet valve malfunction detectederror
808:05Outlet valve faultyOutlet valve malfunction detectederror
808:06Outlet valve not open, position 2Outlet valve failure to open at position 2error
808:07Inlet valve not open, position 2Inlet valve failure to open at position 2error
808:08Inlet valve open low test faultInlet valve open voltage test failureerror
808:09Outlet valve open low test faultOutlet valve open voltage test failureerror
808:xxIntermittent/damaged flex-circuit connectionsValve sensor flex-circuit damageerror
809:00Temperature comparison failure: T1/T2Dual temperature sensor comparison mismatcherror
809:01Data corruption tube temperature bit (verify)Temperature backup data checksum failureerror
809:02Data corruption tube temperature (Temp 1)Temperature sensor 1 backup data checksum failureerror
809:03Data corruption tube temperature (Temp 2)Temperature sensor 2 backup data checksum failureerror
809:04Data corruption tube temperature bit (enable)Temperature enable flag backup data checksum failureerror
809:05Data corruption tube temperature bit (alarm)Temperature alarm flag backup data checksum failureerror
809:06Data corruption temperature bit (alarm switch)Temperature alarm switch flag backup data checksum failureerror
809:07Attempted to modify limit out of rangeTemperature limit modification outside valid rangeerror
809:08Array bounds check on calculated tube temperatureTemperature calculation array index out of boundserror
810:00Advance max override volume exceededAir-in-line sensor advance operation exceeds max volumeerror
810:01Prime max override volume exceededAir-in-line sensor prime operation exceeds max volumeerror
810:02Sensor no tube test failureAir-in-line sensor self-test failure (no tube condition)error
810:03Sensor -15 dB test failureAir-in-line sensor attenuation test failureerror
810:04Air sensor base line too highAir-in-line sensor baseline offset excessiveerror
810:05Backup data corruption: baselineSensor baseline backup data checksum failureerror
810:06Backup data corruption: threshold sumThreshold sum backup data checksum failureerror
810:07Backup data corruption: override activeOverride active flag backup data checksum failureerror
810:08Backup data corruption: alarm disableAlarm disable flag backup data checksum failureerror
810:09Backup data corruption: air detection thresholdAir detection threshold backup data checksum failureerror
810:10Argument passed to AIL override state setInvalid argument to air-in-line override functionerror
810:11Air sensor / circuits saturatedMoisture in tubing channelerror
810:12AIL check illegal state errorAIL circuit detects invalid stateerror
810:13Primary AIL failure 500 micro liters air detectedAir detected in AIL sensorerror
811:01Internal consistency failure: local motor indexMotor index mismatcherror
811:02Timing error w/slot (ADC's starting)ADC timing misalignment with motor sloterror
811:03Internal consistency failure: local motor indexMotor index mismatcherror
811:04Slot edge not found (clockwise)Shuttle motor slot detection failureerror
811:05Internal consistency failure: local motor indexMotor index mismatcherror
811:06Slot not found (after clockwise edge found)Shuttle motor slot detection incompleteerror
811:07Internal consistency failure: local motor indexMotor index mismatcherror
811:08Slot illegal widthShuttle motor slot width out of specificationerror
811:09In slot edge not found (counterclockwise)Shuttle motor counterclockwise slot detection failureerror
811:10Internal consistency failure: local motor indexMotor index mismatcherror
811:11Internal consistency failure: local motor indexMotor index mismatcherror
811:12No slot found over 360 degree searchShuttle motor slot detection complete failurecritical
811:13Found anomalous slot (second slot not found)Shuttle motor multi-slot detection failureerror
811:14Wrong slot(s) foundShuttle motor detected incorrect slotserror
811:15Time out going to center of slotShuttle motor timeout during slot centeringerror
811:16Internal consistency failure: local motor indexMotor index mismatcherror
811:17Illegal global index with tube presentMotor index error when tubing loadederror
811:18Argument to set motor position out of rangeMotor position command exceeds valid rangeerror
811:21Internal consistency failure: local motor indexMotor index mismatcherror
812:00Shuttle index error or motor index pulse timing errorMotor timing pulse detection failureerror
812:01Cycle time errorMotor cycle time exceeds specificationerror
812:02Motor excessive servo errorShuttle motor servo loop error exceeds limiterror
812:03State index monitor enable corruptedState monitoring data corruptionerror
812:04Shuttle index errorShuttle motor index detection failureerror
812:05Movement when stoppedUnwanted motor movement during stopped stateerror
813:00Argument to encoder monitor statusInvalid encoder status argumenterror
813:01Erroneous movementUnexpected motor movement detectederror
813:02Movement when stoppedUnwanted motor movement during stopped stateerror
813:03No movement detected while waiting to stopMotor does not reach stopped stateerror
814:00PHM power supply too highPower supply voltage exceeds specificationcritical
814:01PHM power supply too lowPower supply voltage below specification, battery depletedcritical
814:02Reference input outside toleranceReference voltage out of specificationerror
814:03Shuttle Motor temperature too highShuttle motor temperature exceeds limitunknown
814:04Time Base Error: 4.5 MHz outside toleranceSystem clock frequency out of tolerancecritical
814:05Time Base Error: Accumulator overflowSystem timer accumulator overflowerror
814:06Pump senses that keyed slide clamp is not open during infusionSlide clamp sensor detects clamp closed during infusionunknown
814:07ADC Failure: 5 volt calculationAnalog to digital converter voltage calculation errorerror
814:09Occlusion sensor failureOcclusion sensor not responding correctlyunknown
814:10Amplitude gain test: Distal high side failureDistal pressure sensor gain test failureerror
814:11Amplitude gain test: Proximal high side failureProximal pressure sensor gain test failureerror
814:12ADC Failure: Conversion not completedAnalog to digital converter timeouterror
814:13VAR1 Corrupted: Slide clamp open errorVariable memory corruptionerror
814:14VAR2 Corrupted: Gain test errorVariable memory corruptionerror
815:00VOL MON: Exceeded upper volume limitInfused volume exceeds programmed maximumunknown
815:01VOL MON: Did not reach lower volume limitInfused volume below programmed minimumunknown
815:02VOL MON: Ramp expungedVolume monitoring ramp data corruptederror
815:03VOL MON: Volume monitor action illegalVolume monitoring state errorerror
816:00Illegal argument with watchdog StatusWatchdog timer state errorerror
816:01Fast dog alarmWatchdog timer expiration too fasterror
816:02Slow dog alarmWatchdog timer expiration too slowerror
816:03No COP ServiceWatchdog service call not madeerror
817:00Infusion Register: mismatch, state corruptionInfusion register state mismatch detectederror
817:01Infusion Register: STATE illegal valueInfusion register contains invalid stateerror
817:02Infusion Register: STPTR illegal valueInfusion register pointer invaliderror
817:03Infusion Register: flow rate corruptedInfusion flow rate data corruptionerror
817:04Infusion Register: time out in prep stateInfusion preparation phase timeouterror
817:05Infusion Register: change while in prep stateInfusion settings changed during preperror
817:06Running while running bit not setInfusion state mismatcherror
817:07Last register mismatchInfusion register final state mismatcherror
818:01Pump status bit(s) have changed (Low 1)Status bit corruption or anomalous changeerror
818:02Pump status bit(s) have changed (High 2)Status bit corruption or anomalous changeerror
821:00Relocate error (relocate movement too far)Motor relocate command exceeded limitserror
821:01Illegal build state (building Cycle Table)Invalid state during cycle table buildingerror
861:xxSignal error: UIM Command "open channel"Channel open command not executederror
862:xxSignal error: load mechanism "fully open"Load mechanism not responding to open commanderror
863:xxSignal error: load mechanism "tube loaded"Load mechanism not responding to load commanderror
864:02Signal error: UIM Command 'close channel'Pump module failed to respond to channel close commanderror
865:01Upstream under platen flex connector or sensor failureFlex connector improperly seated in J4 or foiled end of ribbon cable damagederror
865:19Upstream under platen flex connector or sensor failureFlex connector improperly seated in J4 or foiled end of ribbon cable damagederror
865:xxSignal error: load mechanism 'tube misloaded'Tube position error during load cycleerror
866:xxSignal error: load mechanism 'fully closed'Pump failed to fully close loading mechanismerror
867:10Use-induced error when powering down during Delay StartUser presses ON/OFF CHARGE while pump powering down; Delay Start prepares to startwarning
867:xxSignal error: UIM Command 'start pumping'Pump failed to start infusion pump commanderror
868:xxSignal error: UIM Command 'resume pumping'Pump failed to resume infusionerror
869:xxSignal error: UIM Command 'stop pumping'Pump failed to stop infusionerror
870:xxSignal error: UIM Command 'power off'Pump failed to power off commanderror
871:xxSignal error: PHM volume = 0 'infusion complete'Pump reported infusion complete incorrectlyerror
873:xxSignal error: load mechanism 'close w/o tube'Attempted to close without tube loadederror
874:xxSignal error: load mechanism 'index done'Index operation completed unexpectedlyerror
875:xxSignal: shuttle 'local index failed'Shuttle motor failed to index properlyerror
876:xxSignal: Air 'AIL alarm'Air in line detectionerror
877:xxSignal: Occlusion 'Distal occlusion alarm'Distal occlusion detected in lineerror
878:xxSignal: Occlusion 'Proximal occlusion alarm'Proximal occlusion detected in lineerror
879:xxSignal: Drop Sensor 'no drops alarm'Drop sensor not detecting dropserror
880:xxSignal: Temperature 'tube temperature alarm'Tube temperature exceeds thresholderror
881:10Signal: UIM Command 'test enable mode' during Service FeaturesTest enable accessed from Service Features screenerror
881:xxSignal: UIM Command 'test enable mode'Test enable mode failederror
882:15Pump shuttle index failure without tubeShuttle failed to index when no tube presenterror
882:xxSignal: UIM Command 'power on with data'Power on with retained data failederror
883:xxSignal: UIM Command 'MTR hold clear'Motor hold clear command failederror
884:02Signal: PHM 'MTR alarm set'Motor alarm condition detectederror
884:14Signal: PHM 'MTR alarm set' (variant)Motor alarm condition detectederror
891:xxMotor Command: 'index to center' errorShuttle motor failed to index to centererror
892:xxMotor Command: 'start pumping' errorShuttle motor failed during start pump commanderror
893:xxMotor Command: 'resume pumping' errorShuttle motor failed during resume pump commanderror
894:xxMotor Command: 'tube load position' errorShuttle motor failed to reach tube load positionerror
896:01Motor Command: 'center position' error (with clamp/tubing)Motor failed to center when keyed clamp or tubing presenterror
896:xxMotor Command: 'center position' errorShuttle motor failed to reach center positionerror
897:xxMotor Command: 'stop & hold position' errorShuttle motor failed stop and holderror
898:xxMotor Command: 'Index not required' errorIndex not required command failederror
9:xxx:xxx:xxxxSoftware failure - Divide By ZeroDivision by zero attemptedcritical
Advance AirPump is in Advance Air mode to remove air bubbles from tubingUser presses soft key to enter Advance Air modeinfo
Battery DepletedBattery voltage critically low alarmMain battery voltage drops below 10.8 VDCcritical
Battery LowBattery voltage alertMain battery voltage drops below 11.8 VDCwarning
Changing Piggyback ProgramPump is displaying current infusion rate while user modifies piggyback infusion rate during active piggyback infusionUser modifies piggyback rate parameters while piggyback infusion is runninginfo
Changing Primary ProgramPump is displaying current infusion rate while user modifies primary infusion rate or dose during active primary infusionUser modifies primary rate or dose parameters while primary infusion is runninginfo
Channel StoppedPump is powered on but infusion is not running on affected channelInfusion stopped or never started after power-onwarning
CHK TUBETube misloadedImproperly loaded administration set or hardware failureerror
Close Regulating ClampAlarm indicating regulating clamp is open when it should be closed or tubing remains in slot after manual releaseOpen key pressed when administration set loaded and pump powered on, or manual tube release usederror
Confirm Power OffPump is requesting confirmation to power off; not a malfunctionUser presses ON/OFF CHARGE key once to initiate shutdowninfo
Damaged BatteryBattery failure alert requiring serviceCharge/discharge cycles >68 OR discharges below threshold OR installed date shows 01/01/90error
Damaged Battery! Service NowBattery system requires servicing or replacementBatteries are disconnected or depleted during front bezel assembly servicewarning
Dose out of Range LimitsProgrammed dose exceeds or falls below dose limits configured via COLLEAGUE GUARDIAN featureUser enters dose outside predefined limits in COLLEAGUE GUARDIAN-enabled systemwarning
FAILURE [with code]Pump has experienced an internal hardware or software failure; failure code displayedInternal system error detected during operation or self-testcritical
Incomplete Piggyback/Primary ProgramSTART key pressed before all required programming parameters were entered or confirmedUser presses START key with incomplete or unconfirmed programming dataerror
KVOVolume to be infused has reached zero; pump infusing at KVO (Keep Vein Open) rateProgrammed volume depleted during infusioninfo
Lithium Battery LowLithium backup battery charge is low or battery is disconnectedBackup battery voltage below threshold or connection lostwarning
LOADING/labelTube loading in progressAdministration set not fully loaded or obstruction detected during loadingwarning
NO TUBETube not loadedAdministration set not loaded before START key pressederror
Piggyback Callback at HH:MMPiggyback infusion completed; pump has switched to primary rate or KVO (whichever is lower); callback feature is enabledPiggyback infusion volume depleted while callback feature activeinfo
PrimingSystem displaying alert while Prime soft key is active; tubing is being primedUser presses and holds Prime soft keyinfo
Programming PiggybackPiggyback infusion programming is in progress while primary infusion is running; reminder alert to complete and start piggybackUser begins piggyback programming while primary infusion activeinfo
RESETManual tube release activatedManual tube release (MTR) snap action switch activatedwarning
STOPPEDPiggyback/Primary value out of rangePiggyback/primary value entered outside allowable range when Confirm Settings or START pressedwarning
TEMP HGH/labelTemperature too highOperating or administration set temperature above 38°Cwarning
TEMP LOW/labelTemperature too lowOperating or administration set temperature below 15°Cwarning
UPOCCL/labelUpstream occlusionTubing occluded between source container and pumpcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI