B. Braun Perfusor Space: troubleshooting and error codes

The ServiceForge knowledge base holds 35 documented symptoms for the B. Braun Perfusor Space, with 73 ranked causes, 153 diagnostic checks with test points and expected values, and 77 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit fails post-repair checks patient safety

Device does not pass one or more required checks after repair or calibration

Ranked causes

  1. 1. Incorrect or incomplete repair execution (common)
  2. 2. Visual defects in housing or components (occasional)
  3. 3. Mechanical wear or misalignment (occasional)
  4. 4. Functional test failures (occasional)
  5. 5. Syringe recognition malfunction (occasional)
  6. 6. Excessive syringe leakage at stopping (occasional)
  7. 7. Syringe alarm malfunction (occasional)
  8. 8. Pressure stage measurements out of tolerance (occasional)
  9. 9. Motor power limitation test fails (occasional)
  10. 10. Housing version mismatch or incompatibility (rare)
First check

Visual inspection of operating unit

Test point
Front cover, LC-display area, keyboard foil, magnets on operating unit, metal pins on upper housing, rear panel
Expected
No deep scratches, readable display, clean interior, tight magnet/pin fit, no damage to rear panel

32 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blue LEDs flashing periodically in SpaceStation or connected device

Blue LEDs flash periodically when pump is placed in SpaceStation or connected to device communicating with pump, indicating connectivity loss

Ranked causes

  1. 1. Contaminants on connector P2 (common)
  2. 2. Damaged contact surface on P2 connector (occasional)
  3. 3. Cold or broken soldering joints on P2 connector (occasional)
  4. 4. Damaged PCB or broken CAN-Components (rare)
First check

Inspect connector P2 contact surfaces for contamination

Test point
Connector P2
Expected
Clean, no visible contamination

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery cover open patient safety

Operating alarm triggered; infusion stops, red LED flashes, staff call triggered

Ranked causes

  1. 1. Battery compartment cover not correctly closed (common)
  2. 2. Magnet in battery cover missing (occasional)
  3. 3. Battery cover not recognized by battery module (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery discharged

Operating alarm; infusion stops, red LED flashes, audible signal released, staff call triggered

Ranked causes

  1. 1. Device not connected to mains long enough (common)
  2. 2. Connector P2 corroded or defective (occasional)
  3. 3. Battery module defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery nearly discharged

Pre-alarm triggered when battery charge is low; beep sounds and message displays

Ranked causes

  1. 1. Device not connected to mains long enough (common)
  2. 2. Battery connector P2 corroded or defective (occasional)
  3. 3. Battery module defective or too old (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blue LED flashing on pump

Continuous or intermittent blue LED flashing on Operating Unit indicates CAN bus communication error

Ranked causes

  1. 1. CAN lists for 2 seconds after pushing into a Station (common)
  2. 2. Duplicate address assignment (occasional)
  3. 3. CAN dead - Duplicate serial number (occasional)
First check

Power cycle SpaceStation

Test point
Mains supply
Expected
Fault remedied after disconnecting and reconnecting

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device alarm 1010/2010 occurs

Flash memory takes too long to store data when pump is turned off, or alarm occurs after software update

Ranked causes

  1. 1. Flash memory takes too long to store data from RAM on older motherboards (common)
  2. 2. New algorithm in software N calculates reduced allowed time frame for storing to flash memory (occasional)
  3. 3. Memory module defective due to high age or extended operation (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive blocked patient safety

Operating alarm; infusion stops, red LED flashes, audible signal and staff call triggered

Ranked causes

  1. 1. Drive manually blocked (common)
  2. 2. Driving force too low (occasional)
  3. 3. Drive physically damaged (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Functional test: Device not operating after startup patient safety

Device fails to complete self-test sequence, LED indicators not flashing, audible tones not present, or display not showing expected messages

Ranked causes

  1. 1. Device power insufficient or battery not charged (common)
  2. 2. Mechanical damage or incomplete assembly (occasional)
  3. 3. Syringe not properly inserted or piston brake malfunction (occasional)
First check

Verify battery charged during power-on self-test

Test point
Device display
Expected
Display shows "Self test active"
Tools
Power Supply SP

12 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Operating unit does not contact both magnets caution

Gap visible between operating unit and upper housing magnet after assembly; affects magnetic alignment

Ranked causes

  1. 1. Incorrect hinge plate positioning (common)
  2. 2. Operating unit not aligned with magnets on right side (common)
  3. 3. Operating unit not aligned with magnets on left side (occasional)
First check

Close operating unit and verify magnet contact

Test point
Operating unit to upper housing magnets
Expected
Operating unit held by both magnets with no visible gap

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Push-button has no contact patient safety

Operating alarm; device cannot detect syringe push-button engagement

Ranked causes

  1. 1. Negative pressure in syringe system (occasional)
  2. 2. Syringe removed without opening syringe holder (occasional)
  3. 3. Push-button sensor defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery age determination and check

Read batch number from battery label to determine age; replace if older than 4 years or perform battery check.

Ranked causes

  1. 1. Battery older than 4 years (occasional)
  2. 2. Battery fails capacity check (occasional)
First check

Open battery compartment and read batch number from label

Test point
Battery type plate
Expected
Batch code visible and readable

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: After software update

Device requires reconfiguration and calibration following a firmware update to ensure proper operation and apply customer settings

Ranked causes

  1. 1. Software update installed on device (common)
  2. 2. Incompatible HiBaSeD service program version (occasional)
First check

Verify HiBaSeD service program version matches unit software

Test point
Service Program Help > Info
Expected
HiBaSeD version compatible with installed unit software per Table 1-2
Tools
HiBaSeD service program

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Housing lower part replacement and shielding plate rework caution

Procedure for replacing lower housing (Valox to Romiloy material transfer) and reworking shielding plate cutout for compatibility with new housing web

Ranked causes

  1. 1. Lower housing material change from Valox to Romiloy adds web requiring shielding plate rework (common)
  2. 2. Old shielding plate must be reworked if reused with new lower housing Romiloy version (common)
First check

Heat rating plate on old lower housing evenly with hot air

Test point
Rating plate surface on old lower housing
Expected
Plate heated uniformly without scorching; approx. 8 s heating duration

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device alarm 1013/2013 occurs

Update of motherboard to higher software version fails during process, with red LED blinking or all three LEDs lit

Ranked causes

  1. 1. Flash memory incorrectly programmed during failed update (common)
  2. 2. Damaged motherboard with incorrectly programmed flash memory (occasional)
First check

Observe LED status on device front

Test point
Red LED on front
Expected
Red LED blinking followed by device alarm 1013/2013 (State 1), or alarm without LED blinking (State 2), or alarm with all three LEDs (State 3)

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device update failure or incomplete update caution

Software update does not complete when using Space Station or standalone update procedure, or pumps do not power down after update

Ranked causes

  1. 1. Excessive number of devices updated simultaneously (common)
  2. 2. CAN cable disconnected before pumps power down (occasional)
First check

Count number of devices and verify within serial number limits

Test point
Device serial number plate
Expected
Total devices to update does not exceed Table 1-1 limits for device serial numbers
Tools
Table 1-1 reference

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Malfunction of claws patient safety

Operating alarm; syringe piston not recognized or claws/drive damaged; infusion stops

Ranked causes

  1. 1. Syringe piston not recognized (common)
  2. 2. Claws or claw drive damaged (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor power limitation failure patient safety

Drive does not stop with 'Drive blocked' alarm when testing motor power at designated pressure stages

Ranked causes

  1. 1. Motor power test adapter not properly installed (occasional)
  2. 2. Motor drive circuit or power limiting circuit malfunction (rare)
First check

Prepare motor power test adapter

Test point
Syringe gauge and test adapter
Expected
Push-button plate removed; threaded end inserted into adapter; #LehreOPS50 syringe type selected

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe recognition not working patient safety

Device fails to recognize syringe gauge after insertion; FuP value out of tolerance range (29.6-30.4)

Ranked causes

  1. 1. Syringe holder miscalibration (common)
  2. 2. Syringe holder connection to processor PCB defective (rare)
First check

Insert syringe gauge and access Service Info

Test point
Service Info display
Expected
FuP value displayed in right corner

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery status check failure caution

Battery does not charge properly or does not reach 1/3 charge threshold before full discharge detection

Ranked causes

  1. 1. Battery charging circuit defect or degraded battery (common)
First check

Check battery initial charge state

Test point
Battery indicator display
Expected
Battery at least 1/3 charged for test to proceed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery fixation retrofit

Improved battery fixation for devices up to serial number 91061 to prevent wear of battery contacting from vibrations

Ranked causes

  1. 1. Wear of battery contacting from vibration in older pumps (occasional)
First check

Remove battery compartment cover and battery module

Test point
Battery compartment
Expected
Battery module removed

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Piston brake motor assembly installation caution

Assembly procedure for piston brake motor into housing bottom part with syringe holder spring and guide rail, with torque-dependent tightening steps

Ranked causes

  1. 1. Piston brake motor requires proper assembly torque based on housing material version (common)
First check

Identify housing material version (Valox old or Romiloy new)

Test point
Lower housing labeling side
Expected
Romiloy material has imprint '2' starting Q4 2021 for identification

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure Stages Test high voltage

Measure and verify pressure thresholds at pressure stages 1, 3, 6, and 8 using strain gauge and motor power adapter.

Ranked causes

  1. 1. Pressure stage calibration out of tolerance (unknown)
First check

Insert syringe gauge with push-button plate and select pressure stage 1

Test point
Syringe gauge and device menu
Expected
Syringe type "#LehreOPS50" selected, rate 200 mL/h, pressure stage 1 active

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Processor PCB and syringe wing sensor installation caution

Installation of processor PCB, shielding plate, and syringe wing sensor into housing bottom with torque-dependent fastening steps

Ranked causes

  1. 1. Processor PCB and EMC shielding require correct torque based on housing material (common)
First check

Verify shielding plate rework if reusing old plate with new housing

Test point
Old shielding plate dimensions
Expected
Cutout extended at marked position to clear new housing web and sit flush

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Stopping Distance Test

Verify device stops infusion within permissible distance; maximum two drops allowed when drive head engages plunger.

Ranked causes

  1. 1. Drive head not stopping within tolerance (unknown)
First check

Fill and prepare 3 mL syringe with water

Test point
Syringe outlet
Expected
Air completely expelled, water surface flush with outlet

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Syringe Alarm Test patient safety

Verify pre-alarm and end-alarm functions trigger correctly; pre-alarm at ≥5 mL remaining, end-alarm when syringe empty.

Ranked causes

  1. 1. Pre-alarm threshold incorrectly configured (occasional)
First check

Fill 20 mL syringe with 5.5 mL air

Test point
Syringe volume
Expected
5.5 mL air, select syringe type in menu

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Syringe holder and piston brake assembly caution

Installation of syringe holder housing, circlip, carrier, bearing cover, piston brake PCB, blade, and cover with torque-dependent fastening

Ranked causes

  1. 1. Piston brake assembly requires precise torque settings and attention to sharp blade (common)
First check

Insert syringe holder housing into housing bottom

Test point
Housing bottom cavity for syringe holder
Expected
Holder housing fully seated and aligned

12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Syringe Recognition Test

Verify device correctly recognizes syringe diameter using test gauges; FuP value must be within tolerance.

Ranked causes

  1. 1. Syringe holder out of calibration (common)
First check

Check 9 mm diameter gauge FuP value

Test point
Service Info display, right corner
Expected
8.6 to 9.4 (9±0.4) mm, ±0.4

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CAN communication failure in SpaceStation

All pumps in SpaceStation flash blue LED, indicating no CAN communication with SpaceCover

Ranked causes

  1. 1. Station or pump defect after isolation testing (unknown)
First check

Perform single pump in affected slot test

Test point
Affected SpaceStation slot
Expected
Communication restored with single pump

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CAN-Dead state elimination

Pump flashing when removed from SpaceStation due to CAN-Dead state or duplicate serial number

Ranked causes

  1. 1. Duplicate serial number (INR) in system (occasional)
First check

Connect HiBaSeD device

Test point
HiBaSeD interface
Expected
HiBaSeD connected and ready for calibration

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device alarm - internal fault detected patient safety

Operating alarm; serious internal fault detected in system; red LED flashes, staff call triggered

Ranked causes

  1. 1. Serious internal fault in system (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Mount connector gasket P2 (self-adhesive)

Install protective gasket onto P2 connector enclosure to protect from contaminants

Ranked causes

  1. 1. Need for P2 connector protection during service or assembly (common)
First check

Clean P2 adapter with isopropanol

Test point
P2 connector adapter surface
Expected
Clean, no residue, air dried

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure regulation failure - pressure stages test patient safety

Device pressure response does not match expected values for pressure stages 1, 3, or 8 during strain gauge measurement

Ranked causes

  1. 1. Incorrect pressure calibration for stage (common)
First check

Prepare strain gauge pressure measurement

Test point
Syringe gauge with push-button plate
Expected
Syringe gauge inserted; syringe type #LehreOPS50 selected

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rework: Battery contact strip replacement

Replace battery contact strip and circuit board assembly due to wear or damage

Ranked causes

  1. 1. Worn or damaged battery contact strip (occasional)
First check

Remove battery module

Test point
Battery compartment
Expected
Battery module removed

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit handed over to field after completion patient safety

When all checks pass, unit is ready for redeployment

Ranked causes

  1. 1. Successful completion of all required checks (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1001-1009Internal ErrorDevice operationcritical
1010Device alarm on Perfusor® Space due to flash memory data storage timeoutFlash memory takes too long to store data from RAM when pump is turned off, or after software update to version N or highererror
1010Internal ErrorDevice operation during boot or runtimecritical
1011-1012Internal ErrorDevice operationcritical
1013Internal ErrorDevice operation, software update, or impact damagecritical
1013Device alarm on Perfusor® Space due to failed motherboard software updateUpdate of motherboard to higher software version fails during update processerror
1014-1016Loudspeaker errorDevice attempting to produce audible alarms or signalswarning
1017KuP switchoff path defect (K_SM_CLK)Device attempting to power offcritical
1018ADC pressure out of rangeDevice operation during pressure measurementerror
1019Internal ErrorDevice operationcritical
1020FUP Flash Memory ErrorDevice operation during software managementcritical
1021FUP different version KuP to FuPDevice operation with mismatched firmware versionswarning
1022FUP pressure zero test failDevice calibration or operationerror
1023FUP pressure offset test failDevice calibration or operationerror
1024FUP EA key closed too long (20 seconds)User holds emergency/alarm key depressed for extended durationwarning
1025-1041Internal ErrorDevice operationcritical
1042-1051FUP Syncdata KuPDevice operation during synchronizationerror
1052-1053Internal ErrorDevice operationcritical
1060-1087TCI ErrorDevice operation with TCI therapy and special settings enablederror
1088-1090Internal ErrorDevice operationcritical
1100Timebase too fastDevice operation during timing verificationcritical
1101Timebase too slowDevice operation during timing verificationcritical
1102Timebase failDevice operation during timing verificationcritical
1103Keyboard HighDevice operation during keyboard scanningerror
1104EA_KEY defect (≥25 seconds)Emergency/alarm key held depressed for 25 seconds or longerwarning
1105No keydecodeDevice attempting to process keyboard inputerror
1106ROM Romtest defect SoftwareDevice operation during ROM verificationcritical
1107ROM Program defectDevice operation during ROM verificationcritical
1108Internal ErrorDevice operationcritical
1109MPU_Test failedDevice operation during microprocessor verificationcritical
1110RAM_Test failedDevice operation during RAM verificationcritical
1111KUP active resetVoltage supply interruption during device operationwarning
1112-1114Internal ErrorDevice operationcritical
1115Drive too fastDevice operation during motor controlcritical
1116Drive too slowDevice operation during motor controlcritical
1117-1118Internal ErrorDevice operationcritical
1119LCD backlight on defectiveDevice attempting to illuminate LCD backlightwarning
1120LCD backlight off defectiveDevice attempting to dim LCD backlightwarning
1121Red LED on defectiveDevice attempting to illuminate red alarm LEDwarning
1122Red LED off defectiveDevice attempting to turn off red alarm LEDwarning
1123Key pressed too long (without EA-Key) (≥60 seconds)User holds a key depressed for 60 seconds or longer without using emergency/alarm keywarning
1124-1127Internal ErrorUnknown internal processor error conditionerror
1128Drive motion rightless forwardMotor drive recognition failure in forward directionerror
1129Drive motion rightless backwardMotor drive recognition failure in backward directionerror
1130-1159Internal ErrorUnknown internal processor error conditionerror
1160-1162PCA ErrorError in relation to PCA therapy together with special settingserror
1170-1173DLOCK ErrorError in relation to Datalock together with special settingserror
1180-1185DRUG Error / DIFF ErrorError in relation to dose calculation together with special settingserror
1199-1200Internal ErrorUnknown internal processor error conditionerror
1201different version FuP to KuPSoftware version mismatch between FuP and KuP componentserror
1202-1209E_ERROR_STEPMOTORMotor drive or direction of rotation recognition failureerror
1210-1218E_ERROR_DCMOTORPiston brake drive faulterror
1220Syringe holder defectSyringe holder or potentiometer defect detectederror
1221Syringe change timeoutSyringe change operation exceeded timeout periodwarning
1222Plunger plate sensor defectiveMissing protection cap at drive head or light barrier of plunger plate sensor defectiveerror
1236-1238Internal ErrorUnknown internal processor error conditionerror
1239plunger plate sensor defectiveLight barrier of plunger plate sensor defectiveerror
1240-1255Internal ErrorUnknown internal processor error conditionerror
1260-1289TCI ErrorError in relation to TCI therapy together with special settingserror
1290-1301Internal ErrorUnknown internal processor error conditionerror
1320-1328RTM ErrorError in relation to Ramp and Tapper therapy together with special settingserror
1340-1344INTBOL ErrorError in relation to Interval Bolus therapy together with special settingserror
1360-1363PGM ErrorError in relation to Program Mode therapy together with special settingserror
1380-1421Internal ErrorUnknown internal processor error conditionerror
2010Device alarm on Infusomat® Space due to flash memory data storage timeoutFlash memory takes too long to store data from RAM when pump is turned off, or after software update to version N or highererror
2013Device alarm on Infusomat® Space due to failed motherboard software updateUpdate of motherboard to higher software version fails during update processerror
8000-8049Invalid Message / Invalid TCI InputError in relation to TCI therapy together with special settingswarning
8100-8102TOM ErrorError in relation to Take over Mode together with special settingserror
9000-9999Internal ErrorUnknown internal processor error conditionerror
Alarm / drive blockedMotor power limitation or drive mechanism obstruction detectedMotor reaches maximum power limit or drive head cannot move under loaderror
Alarm / pressure highDownstream occlusion detected or pressure limit exceededPressure measurement exceeds programmed stage limit during infusionwarning
Calibration errorCalibration did not terminate successfullyError occurs during claw or syringe holder calibration stepserror
Calibration interruptedCalibration process did not complete successfullyUser interrupts calibration or system error occurs during procedurewarning
CAN Error Code 1Duplicate serial number in systemCAN bus reports duplicate serial number (INR)warning
CAN Error Code 2Duplicate address assigned in systemCAN bus address assignment conflictwarning
Drive blockedMotor drive stalled or blocked by excessive loadMotor unable to advance syringe due to resistance exceeding motor power limit at given pressure stagewarning
Pressure highPressure detected exceeds stage-specific thresholdStrain gauge pressure exceeds programmed limit during infusion at selected pressure stagewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI