B. Braun Infusomat Space: troubleshooting and error codes
The ServiceForge knowledge base holds 56 documented symptoms for the B. Braun Infusomat Space, with 74 ranked causes, 192 diagnostic checks with test points and expected values, and 92 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Caution sign at front does not flash yellow caution
Front caution indicator fails to illuminate yellow
Ranked causes
- 1. Downstream pressure sensor failed (common)
- 2. Processor PCB failed (occasional)
- 3. Safety clamp issue (rare)
Observe caution indicator at front of unit
- Test point
- Caution sign LED
- Expected
- The caution sign at the front of the unit flashes yellow
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door bolt does not secure operating unit caution
Door bolt fails to lock and tighten the operating unit
Ranked causes
- 1. Safety clamp failed (common)
- 2. Door bolt drive failed (occasional)
- 3. Processor PCB failed (occasional)
Verify door bolt tightens and secures unit
- Test point
- Door bolt and safety clamp
- Expected
- The operating unit is tightened by the door bolt and then fixed
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door bolt does not unlock operating unit caution
Door bolt mechanism fails to release and open unit
Ranked causes
- 1. Door bolt drive failed (common)
- 2. Processor PCB failed (occasional)
- 3. Operating unit hardware issue (occasional)
Attempt to unlock door bolt and open unit
- Test point
- Door bolt mechanism
- Expected
- The door bolt unlocks the operating unit which then opens
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service info not displayed on LCD caution
LCD fails to display Service info screen
Ranked causes
- 1. Air inline sensor failed (common)
- 2. Operating unit failed (occasional)
- 3. Processor PCB failed (occasional)
Power on and check if Service info appears on display
- Test point
- LC display
- Expected
- The Service info appears on the LC display
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inline sensor water value out of specification caution
Service info displays air inline sensor value outside TSC water specification
Ranked causes
- 1. Air inline sensor failed (common)
- 2. Processor PCB failed (occasional)
Display Service info and verify air inline sensor water value
- Test point
- air inline sensor reading on display
- Expected
- The value displayed (water value) is within the specifications of the TSC
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
At least one LED lights up but not all caution
Partial LED illumination on power on
Ranked causes
- 1. LCD display failure (common)
- 2. Message not displayed on LCD (occasional)
Observe LED illumination at power on
- Test point
- Individual LEDs
- Expected
- At least one LED lights up for a short moment
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery module discharges too fast
Battery depletes rapidly during use or storage
Ranked causes
- 1. Device not used for longer time (common)
- 2. Battery module not discharged and charged at regular intervals (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not functioning during power loss caution
Device does not operate on battery power when mains supply is disconnected
Ranked causes
- 1. Battery depleted or missing (unknown)
- 2. Battery alarm activation during cover opening (unknown)
Power down device
- Test point
- Power switch
- Expected
- Device powers off
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No beep sound heard at power on caution
Operating unit powered but no audio feedback; keyboard and display lit
Ranked causes
- 1. Loudspeaker or switch failed (common)
- 2. Processor PCB failed (occasional)
Power on unit and listen for audio feedback
- Test point
- Loudspeaker output
- Expected
- A short deep and then a short high beep sound
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Post-repair verification required patient safety
After disassembly and reassembly work (chapters 3.2-3.9), visual inspection and self-test must be performed before returning unit to service
Ranked causes
- 1. Completion of internal maintenance or repair (common)
- 2. Flow accuracy adjustment or pump work performed (occasional)
Cleanliness inspection
- Test point
- Entire device exterior and internal visible surfaces
- Expected
- Clean, free of debris, grease, ultrasonic gel or contamination
46 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure alarm does not trigger on line occlusion patient safety
Red LED on service connector does not light when supply line is occluded upstream or downstream; system fails to detect pressure threshold
Ranked causes
- 1. Electronic pressure sensor malfunction or threshold misconfiguration (occasional)
- 2. Service connector not properly connected to deactivate electronic cut-off during mechanical test (occasional)
Open operating unit and verify slide clamp is fully retracted
- Test point
- Operating unit; slide clamp position
- Expected
- Slide clamp completely pulled back on line
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not deliver or line is not vented caution
Pump delivery not audible or line venting fails
Ranked causes
- 1. Pump drive failed (common)
- 2. Processor PCB failed (occasional)
Listen for pump delivery and check line venting
- Test point
- Pump operation and line vent
- Expected
- You can hear the pump as it is delivering and the line is vented
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service info sensor values out of specification caution
Air inline sensor or temperature values exceed TSC specifications
Ranked causes
- 1. Operating unit failed (common)
- 2. Processor PCB failed (occasional)
Display Service info and verify air inline sensor value
- Test point
- air inline sensor reading on display
- Expected
- The value displayed is within the specifications of the TSC
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent line message not displayed caution
LCD does not display 'Vent line - 16 ml?' message
Ranked causes
- 1. Operating unit failed (common)
- 2. Processor PCB failed (occasional)
Check if vent line message is displayed
- Test point
- LC display
- Expected
- 'Vent line - 16 ml?' is displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air bubble not detected in supply line patient safety
Alarm not triggered when air bubble of approximately 0.4 ml (56 mm line length) is present in supply line during infusion
Ranked causes
- 1. Air inline sensor malfunction or miscalibration (unknown)
Generate air bubble in supply line
- Test point
- Supply line to pump
- Expected
- Air bubble approximately 0.4 ml (56 mm line length)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inline sensor does not detect or display air bubbles patient safety
Air inline sensor fails to trigger alarm or display air content value on LC display when water-filled line contains air
Ranked causes
- 1. Air sensor optical path blocked or electronic circuit malfunction; service connector not plugged in (occasional)
Verify service connector is plugged into connector P2
- Test point
- Service connector; connector P2
- Expected
- Service connector fully seated on P2
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inline sensor reading verification patient safety
Verification that air inline sensor displays correct value within technical specifications during device self-test with air-filled line
Ranked causes
- 1. Air inline sensor functioning correctly during device check (common)
Battery module not charging caution
Battery module fails to charge even with power supply connected
Ranked causes
- 1. Battery module defective (common)
Connect power supply and allow battery to charge for 16 hours
- Test point
- Battery module in device
- Expected
- Battery charges to full capacity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivery accuracy test fails patient safety
Measured volume delivered does not match expected volume within acceptable deviation; infusion rate not accurate
Ranked causes
- 1. Pump mechanism wear or calibration drift; improper test setup (occasional)
Verify test setup configuration matches Fig. 7-2 requirements
- Test point
- Test setup components; liquid level in container
- Expected
- Liquid level approx. 50 cm above middle of unit opening; all components properly connected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device alarm 2027 patient safety
Air inline sensor alarm triggered
Ranked causes
- 1. Air inline sensor defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device alarm 2119
LCD display manufacturing tolerance out of specification
Ranked causes
- 1. Manufacturing tolerance of LC-display too high (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device alarms 2021 to 2022 patient safety
Upstream pressure sensor fault alarms
Ranked causes
- 1. Upstream pressure sensor defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device alarms 2023 to 2024 patient safety
Downstream pressure sensor fault alarms
Ranked causes
- 1. Downstream sensor defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device check procedure validation patient safety
Systematic verification of Infusomat Space device functionality through a multi-step diagnostic sequence including power-up, LED indicators, audio signals, display functions, door mechanisms, line insertion, sensor readings, and pressure stage testing.
Ranked causes
- 1. Normal device operation during service check (common)
Device insertion or connection to SpaceStation/SpaceControl
- Test point
- Device connector
- Expected
- Device properly inserted or connected
45 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not operate after power-on caution
Connected Space system device fails to operate correctly after switching on with power supply connected
Ranked causes
- 1. Power supply not properly connected or locked (unknown)
Verify power supply connection and lock
- Test point
- Power supply connector to device
- Expected
- Connection secure with lock engaged
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door bolt drive installation and potentiometer engagement
Operating pin of linear potentiometer must engage in fork when spindle tappet is clipped in during door bolt drive installation
Ranked causes
- 1. Operating pin not engaging potentiometer fork during door bolt drive installation (occasional)
Install door bolt drive assembly
- Test point
- Door bolt drive position
- Expected
- Drive positioned for potentiometer engagement
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drop sensor does not alarm on air jet or line disconnect patient safety
Drop sensor fails to trigger alarm when drop chamber is squeezed to generate jet or when infusion line is disconnected from container
Ranked causes
- 1. Drop sensor not connected or malfunction; optical sensor dirty or misaligned (occasional)
Verify drop sensor is properly connected to device
- Test point
- Drop sensor connector
- Expected
- Sensor connected and enabled
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety compliance check patient safety
Verify that patient and housing leakage currents meet safety limits per standards IEC/EN 60601-1, VDE 0750, and VDE 0751
Ranked causes
- 1. Completion of electrical work or power supply replacement (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Housing bottom part and seal installation
Seal position relative to safety clamp must be verified, and hinge spring must be properly pressed into spring holder during bottom part assembly
Ranked causes
- 1. Improper seal position or hinge spring not fully seated in spring holder (occasional)
Verify seal position relative to safety clamp
- Test point
- Seal and safety clamp alignment
- Expected
- Seal correctly positioned in relation to safety clamp
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Housing front panel and inner frame assembly and alignment caution
Door bolt must be in open position before front panel mounting; membrane must be routed correctly; front frame lugs must engage inner frame openings; positioning hemispheres must align with bottom inner frame slots; final assembly must be parallel
Ranked causes
- 1. Improper door bolt position or membrane routing causing misalignment during front panel assembly (occasional)
Move door bolt to open position (full left)
- Test point
- Door bolt position
- Expected
- Door bolt at full left stop (open position)
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Housing upper part cable and component routing caution
Cables must not be pinched, connector seals must be correctly positioned, snap-in hook spring must be outside housing wall, and loudspeaker must be properly oriented during upper housing assembly
Ranked causes
- 1. Improper cable routing or component positioning causing pinching, seal misalignment, or speaker malfunction (occasional)
Inspect cable routing before upper housing installation
- Test point
- All cables and connectors in housing interior
- Expected
- Cables routed to avoid pinching; no cables in housing edge paths
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infusion or bolus does not deliver or buttons not responsive patient safety
Infusion fails to deliver fluid; intermediate bolus not functioning; button presses do not trigger expected functions
Ranked causes
- 1. Operating unit not properly closed or pump mechanism malfunction (common)
Verify infusion delivers fluid when operating unit is closed
- Test point
- Pump outlet; graduated cylinder
- Expected
- Fluid flows into graduated cylinder
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infusomat Space line does not seat or squeeze properly patient safety
Line does not slide into position or slide clamp does not disengage release lever; line is not squeezed by pump mechanism
Ranked causes
- 1. Line not properly inserted or slide clamp mechanism stuck (common)
Insert line and verify slide clamp mechanism engages
- Test point
- Infusomat Space line; slide clamp
- Expected
- Line inserts smoothly; slide clamp disengages release lever
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inner frame lubrication and assembly caution
Inner frame components must be properly lubricated to prevent stiff or jerky movements during installation
Ranked causes
- 1. Insufficient lubrication of inner frame components causing stiff or jerky movements (common)
Inspect inner frame components for lubrication adequacy
- Test point
- Plastic guides, spring bearings, toothed wheels, bearing guides
- Expected
- Sufficient grease film present; no stiff or jerky movement observed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard and LCD display not illuminated caution
No power indication at keyboard or display
Ranked causes
- 1. Processor PCB failed (common)
Power on unit and observe keyboard and display lighting
- Test point
- Keyboard and LCD display
- Expected
- The keyboard and the LC display are lighted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Membrane installation during assembly caution
Membrane must be properly installed over pump drive pushers using correct technique and lubricant
Ranked causes
- 1. Membrane requires careful installation over pump drive pushers (common)
Prepare new membrane and lubricant
- Test point
- Membrane and Meliseptol® availability
- Expected
- New membrane available; Meliseptol® lubricant prepared
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No message displayed on LCD display caution
LCD display is lit but no message appears
Ranked causes
- 1. Processor PCB failed (common)
Power on and observe LCD display and LEDs
- Test point
- LC display and all LEDs
- Expected
- All LEDs light up for a short moment and a message is displayed on the LC display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noise during opening or closing operating unit caution
Abnormal sounds heard when door opens or closes
Ranked causes
- 1. Toothed wheels in door bolt drive defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating unit hinge spring and connector assembly
Hinge spring must be correctly inserted in hinge cover and rear panel bore hole; keyboard PCB bonding must be observed; grooves must point to rear panel; connection cable and ribbon cable cover must be properly routed
Ranked causes
- 1. Improper hinge spring insertion or connector misalignment during operating unit assembly (occasional)
Insert hinge spring long side in hinge cover opening
- Test point
- Hinge cover opening
- Expected
- Spring long side fully inserted in opening
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating unit lock fails caution
Unit does not lock or unlock properly when stacked with another Infusomat Space device or when fitting battery compartment cover
Ranked causes
- 1. Lock mechanism worn or misaligned (occasional)
Verify lock engages when stacking unit on top of another device
- Test point
- Unit lock mechanism
- Expected
- Lock engages smoothly
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply mechanical lock malfunction caution
Power supply connector lock does not secure properly to Space system device
Ranked causes
- 1. Lock mechanism worn or misaligned (unknown)
Connect power supply to device
- Test point
- Power supply connector and device receptacle
- Expected
- Lock engages smoothly with audible or tactile click
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure adjustment unit calibration and cover cap removal patient safety
If pressure adjustment unit is serviced, cover cap must be removed from housing top before upper housing mounting and kept until calibration is complete
Ranked causes
- 1. Pressure adjustment unit serviced or replaced; calibration required (occasional)
Determine if pressure adjustment unit has been serviced, dismounted, or replaced
- Test point
- Work history and unit condition
- Expected
- Documentation confirms pressure adjustment unit service
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processor PCB positioning and lugs alignment
Adjusting lug of bottom inner frame must be positioned in processor PCB boring during assembly
Ranked causes
- 1. Adjusting lug misalignment with processor PCB boring (occasional)
Locate adjusting lug on bottom inner frame
- Test point
- Bottom inner frame lug area
- Expected
- Adjusting lug clearly visible and accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processor PCB replacement data restoration patient safety
After replacing processor PCB, previous pump data must be restored from saved backup; calibration data is not transferred and requires new calibration
Ranked causes
- 1. Processor PCB replaced; data backup available on PC (unknown)
Start Service Program and navigate to Modification data tab
- Test point
- Service Program screen
- Expected
- Modification data register tab displayed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump drive runs continuously
Pump motor operates without stopping
Ranked causes
- 1. Light barrier defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety clamp operating lever does not press or engage properly patient safety
Operating lever of safety clamp cannot be pressed down against spring tension; caution sign does not flash yellow; clamp does not secure line
Ranked causes
- 1. Safety clamp spring tension binding or wear; operating lever misaligned (occasional)
Measure residual pressure with operating unit unlocked but not opened
- Test point
- Pressure gauge; safety clamp
- Expected
- Residual pressure matches TSC specification after 10 seconds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Snap-in hook installation and housing assembly completion
Snap-in hook must be inserted before mounting housing bottom part; snap-in hooks must be engaged between housing top and front frame during final assembly
Ranked causes
- 1. Snap-in hook not inserted before bottom part mounting or not properly engaged during final assembly (occasional)
Insert snap-in hook before mounting housing bottom part
- Test point
- Snap-in hook position and housing bottom area
- Expected
- Hook inserted and positioned correctly before bottom part mounting
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software update failure or interruption caution
Device disconnected or powered off during software update; component corruption preventing PC-based recovery; device requires return to B. Braun for repair.
Ranked causes
- 1. Power loss or device disconnection during software update (rare)
Verify device power supply and data connection are stable
- Test point
- Power cable and PC connection
- Expected
- Stable connection maintained throughout update
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Special plastic screw installation caution
Special deformation-type screws for plastic housings must be installed with correct technique to establish proper thread and avoid stripping
Ranked causes
- 1. Special deformation screws require specific installation procedure to establish thread (common)
Position special screw on plastic housing thread
- Test point
- Screw and housing thread
- Expected
- Screw aligned and placed on thread
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature sensor reading deviation
Temperature sensor values on LC display differ from TSC specifications
Ranked causes
- 1. Temperature sensor calibration drift (unknown)
Read temperature sensor values
- Test point
- LC display
- Expected
- Values match TSC specifications, Per TSC specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature sensor reading verification patient safety
Verification that both temperature sensors (Temp1, Temp2) display values and that their difference falls within technical specifications
Ranked causes
- 1. Temperature sensors functioning correctly (common)
Unit does not function when battery module removed patient safety
Alarm should sound continuously for minimum 3 minutes when battery module is removed and unit is unpowered
Ranked causes
- 1. Battery module not properly installed or battery compartment cover magnetic function compromised (occasional)
Remove battery module from unit without external power supply
- Test point
- battery compartment
- Expected
- Continuous audible alarm for minimum 3 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not respond to power on caution
No LEDs light up when power is applied
Ranked causes
- 1. Power supply failed (common)
Power on unit and observe LEDs
- Test point
- All LEDs
- Expected
- All LEDs light up for a short moment
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit malfunction after service or repair
Device may fail or malfunction after a longer period of operation due to ESD damage to semiconductors during service work
Ranked causes
- 1. Electrostatic discharge (ESD) damage to semiconductors during service (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water inline sensor reading verification patient safety
Verification that water inline sensor displays correct value within technical specifications after line is filled with liquid during device check
Ranked causes
- 1. Water inline sensor functioning correctly with filled line (common)
Zero force insertion connector ribbon cable installation
Ribbon cable must be centered in plug guides and properly locked in zero force insertion connector before proceeding
Ranked causes
- 1. Ribbon cable misalignment or incomplete locking in zero force connector (occasional)
Position ribbon cable centered in zero force connector plug guides
- Test point
- Ribbon cable and plug guide alignment
- Expected
- Cable centered between left and right plug guides
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Battery nearly discharged | Device not connected to mains long enough or battery module defective/too old | warning |
2 | Battery discharged | Device not connected to mains long enough or battery module defective | error |
2001-2013 | Internal Error | Unknown | error |
2014 | Loudspeaker not off | Loudspeaker remains active when should be off | error |
2015 | Loudspeaker lost | Loudspeaker not detected | error |
2016 | Loudspeaker shorted | Loudspeaker short circuit detected | error |
2017 | KUP no work | KUP processor not functioning | error |
2018 | FUP-KUP OS Cycle Timeout | OS cycle communication timeout between processors | error |
2019 | FUP Flash Memory Error | Function processor flash memory error | error |
2020 | FUP different version KuP to FuP | Version mismatch between control and function processors | error |
2021 | FUP UPSTREAM TOO HI | Upstream pressure exceeds acceptable level | error |
2021 | Upstream pressure sensor fault | Upstream pressure sensor malfunction detected | error |
2022 | FUP UPSTREAM TOO LOW | Upstream pressure below acceptable level | error |
2022 | Upstream pressure sensor fault | Upstream pressure sensor malfunction detected | error |
2023 | FUP DOWNSTREAM TOO HI | Downstream pressure exceeds acceptable level | error |
2023 | Downstream pressure sensor fault | Downstream pressure sensor malfunction detected | error |
2024 | FUP DOWNSTREAM TOO LOW | Downstream pressure below acceptable level | error |
2024 | Downstream pressure sensor fault | Downstream pressure sensor malfunction detected | error |
2025 | FUP EA key closed too long | Emergency stop key held closed beyond timeout | error |
2026 | FUP Task inactive | Function processor task not responding | error |
2027 | Air inline sensor defective | Air inline sensor malfunction detected | error |
2027 | FUP Airsensor CALDATA CRC | Air inline sensor calibration data CRC error | error |
2028 | FUP Airsensor PROGDATA CRC | Air inline sensor program data CRC error | error |
2029 | FUP Airsensor Calibration Error | Air inline sensor calibration failed | error |
2030 | No Airtest Kup | Air test not completed on KUP | error |
2031 | FUP Temperature Sensors difference | Mismatch between temperature sensor readings | error |
2032 | FUP Temperature Raw Values difference | Mismatch between raw temperature values | error |
2033 | FUP Airsensor Signature Error | Air inline sensor signature verification failed | error |
2034 | FUP CAN paket scheduler list failure | CAN interface packet scheduling error | error |
2035 | Light barrier or motor defect/turn backwards | Light barrier malfunction or motor rotation direction incorrect | error |
2036 | FUP Language Flash Memory Error | Language file flash memory error | error |
2037 | FUP Language Version Error | Language file version mismatch | error |
2038 | FUP CRC Task inactive | CRC task not responding | error |
2039 | Two pumps with same serial number in SpaceStation | Duplicate serial numbers detected | error |
2100 | Timebase too fast | Processor clock running faster than specification | error |
2101 | Timebase too slow | Processor clock running slower than specification | error |
2102 | Timebase fail | Processor timebase not functioning | error |
2103 | Keyboard High | Keyboard voltage reading abnormally high | error |
2104 | EA_KEY defect 25sec | Emergency stop key malfunction lasting 25 seconds | error |
2105 | No keydecode | Keyboard input not being decoded | error |
2106 | ROM Romtest defect | ROM self-test failed | error |
2107 | ROM Program defect | ROM program area corrupted | error |
2109 | MPU_Test failed | Microprocessor unit self-test failed | error |
2110 | RAM_Test failed | RAM self-test failed | error |
2111 | KUP active reset | KUP processor underwent unexpected reset | error |
2112-2114 | Internal Error | Unknown | error |
2115 | Drive too fast | Motor drive speed exceeds specification | error |
2116 | Drive too slow | Motor drive speed below specification | error |
2117-2118 | Internal Error | Unknown | error |
2119 | lcd backlight on defect | LCD backlight cannot turn on | error |
2119 | LC-display manufacturing tolerance too high | Display tolerance manufacturing defect detected | error |
2120 | lcd backlight off defect | LCD backlight cannot turn off | error |
2121 | red led on defect | Red LED cannot turn on | error |
2122 | red led off defect | Red LED cannot turn off | error |
2123 | key pressed too long (without EA-Key) 60sec | Key held pressed for 60 seconds without emergency stop key | error |
2124-2127 | Internal Error | Unknown | error |
2128 | Drive motion rightless forward | Motor forward motion not detected | error |
2129 | Drive motion rightless backward | Motor backward motion not detected | error |
2130-2200 | Internal Error | Unknown | error |
2201 | different version FuP to KuP | Version mismatch between function and control processors | error |
2202 | stepmotor PH1A not 0 | Stepper motor phase A not at zero state | error |
2203 | stepmotor not middle current | Stepper motor mid-phase current not present | error |
2204 | stepmotor K_SM_CLK | Stepper motor clock signal issue | error |
2205 | stepmotor PH1A not 1 | Stepper motor phase A not at one state | error |
2206 | stepmotor not zero current | Stepper motor zero-state current not present | error |
2207 | stepmotor not middle current | Stepper motor mid-phase current not present | error |
2208 | stepmotor not high current | Stepper motor high-state current not present | error |
2209 | stepmotor PH1A,PH1B,PH2A,PH2B not 0 | Stepper motor all phases not at zero state | error |
2210 | K_PKM_UB not off | Door bolt drive signal not off | error |
2211 | K_PKM_UB not on | Door bolt drive signal not on | error |
2215 | no V_MOT | Motor supply voltage not present | error |
2216 | overvoltage test fail | Motor supply voltage exceeds specification | error |
2217 | no V_MOT | Motor supply voltage not present | error |
2218 | undervoltage test fail | Motor supply voltage below specification | error |
2220 | door poti defect | Door potentiometer not functioning | error |
2221 | tube change timeout | Tube change procedure timeout | error |
2222 | temperatur sensor difference | Temperature sensor reading mismatch | error |
2223 | airsensor defect | Air inline sensor not functioning | error |
2224 | rate compensation difference | Rate compensation calculation mismatch | error |
2225 | rate compensation max value | Rate compensation exceeds maximum | error |
2226 | rate compensation min value | Rate compensation below minimum | error |
2227 | received Airdata defect | Air sensor data corrupted or invalid | error |
2228 | defect temperature sensor switch | Temperature sensor switch malfunction | error |
2229 | FuP not stopped by KuP alarm | Function processor not halted by control processor alarm | error |
2230 | drive totphase is defect | Motor total phase control defective | error |
2231 | tube data set is defect | Tube dataset corrupted | error |
2237-2255 | Internal Error | Unknown | error |
3 | Battery cover open | Battery compartment cover not correctly closed or magnet missing or not recognized | error |
4 | Device alarm | Serious internal fault detected in system | critical |
Alarm / Check supply | Supply line check alarm triggered during upstream delivery test | Roller clamp closed or line squeezed between container and pump during upstream infusion | warning |
Alarm / Pressure too high | Maximum pressure limit exceeded for current pressure stage | Device delivers against closed downstream valve; maximum pressure threshold for selected pressure stage is reached | warning |
Devicealarm / 2111 | Battery removal alarm code indicating low or missing battery condition | Battery removed during powered device operation or device powered on without battery installed | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.