Arthrex AR-6480 DualWave: troubleshooting and error codes

The ServiceForge knowledge base holds 51 documented symptoms for the Arthrex AR-6480 DualWave, with 60 ranked causes, 98 diagnostic checks with test points and expected values, and 19 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Warning messages appear on operator display patient safety

One or more critical warning messages (*Critical Failure*, *Power Failure*, *OVP Detect Fail*, *Sensor Failure*, or *Pressure Fault*) appear on the pump's operator display during startup or operation.

Ranked causes

  1. 1. Critical Failure message present (rare)
  2. 2. Power Failure message present (rare)
  3. 3. OVP Detect Fail message present (rare)
  4. 4. Sensor Failure message present (rare)
  5. 5. Pressure Fault message present (rare)
First check

Verify no *Critical Failure* message on first display line

Test point
Operator display first line
Expected
No *Critical Failure* message

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure relief setting caution

Calibrate overpressure and pressure fault thresholds during preventive maintenance

Ranked causes

  1. 1. Overpressure threshold calibration required (unknown)
  2. 2. Pressure Fault threshold calibration required (unknown)
First check

Set pump pressure to specified level

Test point
Pump pressure display
Expected
120 mmHg

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Over Pressure message does not appear at 300 mmHg patient safety

Over Pressure message should appear when sensed pressure exceeds 300 mmHg software limit

Ranked causes

  1. 1. Pressure sensor failure (rare)
  2. 2. Software limit not active (rare)
First check

Set pressure to 120 mmHg on pump display

Test point
pump screen pressure setting
Expected
120 mmHg

3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Pressure Fault message does not appear at 40 mmHg with clamped tubing

Pressure Fault message should appear after waiting with tubing clamped at 40 mmHg set pressure

Ranked causes

  1. 1. Pressure sensor malfunction (unknown)
  2. 2. Tubing not properly clamped (common)
First check

Unclamp tubing from previous test

Test point
tubing clamp
Expected
Tubing flows freely

3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Tubing Out or Door Open messages persist after corrective action

The *Tubing Out* or *Door Open* messages continue to display on the operator display even after tubing is loaded and the door is closed, indicating the sensors are not clearing properly.

Ranked causes

  1. 1. Tubing not fully seated in sensor coupler (common)
  2. 2. Door not fully closed or door latch defective (common)
First check

Verify *Tubing Out* and *Door Open* messages clear after loading tubing and closing door

Test point
Operator display
Expected
*Tubing Out* and *Door Open* messages are not shown

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tubing Out or Door Open messages persist after door closure

Tubing Out and Door Open messages should not appear after inflow tubing is loaded and door is closed

Ranked causes

  1. 1. Tubing not fully loaded (common)
  2. 2. Door not fully closed (common)
First check

Load inflow tubing into pump

Test point
tubing path
Expected
Tubing fully seated

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Boost button not responding in inflow-only mode

Cannot select between Boost options or Boost button not functioning when pump in inflow-only mode

Ranked causes

  1. 1. Boost button not accessible or responding (unknown)
First check

Verify Boost button accessible in inflow-only mode

Test point
Touchscreen Boost button
Expected
Boost button appears and allows selection between different options

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Boost mode options not selectable in inflow-only mode

Boost button appears but user cannot select between different Boost options when pump is in inflow-only mode

Ranked causes

  1. 1. Touchscreen button or Boost mode selection malfunction (unknown)
First check

Boost button functionality and option selection in inflow-only mode

Test point
Touchscreen Boost button and menu
Expected
Able to select between different Boost options

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Overpressure alarm response patient safety

Verify pump raises overpressure message when pressure exceeds software threshold

Ranked causes

  1. 1. Overpressure detection calibration required (unknown)
First check

Verify overpressure alarm message displays when pressure exceeds threshold

Test point
pump display
Expected
"Over Pressure" message appears when pressure exceeds 300 mmHg

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure fault alarm response patient safety

Verify pump raises pressure fault message when clamped tubing maintains low pressure condition

Ranked causes

  1. 1. Pressure fault detection calibration required (unknown)
First check

Verify pressure fault alarm message displays when tubing remains clamped at 40 mmHg

Test point
pump display
Expected
"Pressure Fault" message appears when tubing clamped at 40 mmHg setting

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure fault detection at 40 mmHg patient safety

Procedure to verify pressure fault alarm triggers when tubing is clamped at 40 mmHg setting

Ranked causes

  1. 1. Pressure fault alarm verification (unknown)
First check

Set pressure to 40 mmHg and start pump

Test point
Pump pressure control and display
Expected
Pump running at 40 mmHg

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump pressure adjustment and overpressure alarm verification patient safety

Preventive maintenance procedure to verify pump pressure setting to 120 mmHg and test overpressure alarm function at 300 mmHg threshold

Ranked causes

  1. 1. Maintenance verification step (unknown)
First check

Set pump pressure using + button

Test point
Pump pressure display
Expected
120 mmHg

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump pressure setting caution

Preventive maintenance procedure to verify and set pump pressure output at 120 mmHg and verify overpressure alarm

Ranked causes

  1. 1. Pressure setting drift or calibration needed (common)
First check

Set pressure to 120 mmHg

Test point
pump display pressure setting
Expected
120 mmHg

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump pressure setting and overpressure detection patient safety

Procedure to verify and adjust pump pressure to 120 mmHg and confirm overpressure alarm triggers at 300 mmHg software limit

Ranked causes

  1. 1. Calibration verification (unknown)
First check

Verify 'Tubing Out' and 'Door Open' messages are not displayed

Test point
Pump display
Expected
Messages not shown

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannula button outflow rate selection not working

Cannot select between different outflow rates when Cannula button is pressed

Ranked causes

  1. 1. Cannula button not responding or rate options not accessible (unknown)
First check

Verify Cannula button rate selection

Test point
Touchscreen Cannula button
Expected
Able to select between different outflow rates

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannula mode outflow rates not selectable

Pressing Cannula button but user cannot select between different outflow rate options; system should provide rate selection

Ranked causes

  1. 1. Cannula mode button or outflow rate selection malfunction (unknown)
First check

Cannula button functionality and rate selection

Test point
Touchscreen Cannula button and outflow rate menu
Expected
Able to select between different outflow rates

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Critical Failure message displayed patient safety

Critical Failure message appears on first line of display

Ranked causes

  1. 1. Unknown critical system fault (unknown)
First check

Verify Critical Failure message does not appear

Test point
Display first line
Expected
No Critical Failure message

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door Open message appears incorrectly

Message 'Door Open' (door/flap open) displays when roller housing door is properly closed

Ranked causes

  1. 1. Door/flap sensor malfunction or misalignment (unknown)
First check

Confirm Door Open message appears only when door open

Test point
Roller housing door position
Expected
Message appears when door open; disappears when door closed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door Open message does not appear

Door Open message should appear when roller housing door is open

Ranked causes

  1. 1. Door sensor detection failure (unknown)
First check

Open roller housing door

Test point
Roller housing door
Expected
Door Open message appears

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door Open message does not appear when expected

Door Open message should appear when roller housing door is open

Ranked causes

  1. 1. Door sensor malfunction (unknown)
First check

Open the roller housing door

Test point
roller housing door
Expected
Door Open message appears

Door Open message does not appear when roller housing door is open

The *Door Open* message should display on the pump's operator display when the roller housing door is open, but it does not appear.

Ranked causes

  1. 1. Door sensor malfunction or disconnection (unknown)
First check

Verify *Door Open* message displays when door is open

Test point
Roller housing door sensor
Expected
*Door Open* message appears on operator display

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot pedal not detected on display

Foot pedal icon does not appear on pump display after connection to foot pedal port

Ranked causes

  1. 1. Foot pedal not properly connected or faulty (unknown)
First check

Verify foot pedal connection and icon display

Test point
foot pedal port and display
Expected
foot pedal icon appears on display

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lavage button pressure not responding

Pressure does not increase according to user-defined specifications when Lavage button is pressed

Ranked causes

  1. 1. Lavage mode not activating pressure increase (unknown)
First check

Verify Lavage button increases pressure

Test point
Touchscreen Lavage button
Expected
Pressure increases according to user-defined specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lavage mode pressure does not match user-defined specifications

Pressing Lavage button and setting pressure does not achieve user-configured pressure levels; pressure should increase according to settings

Ranked causes

  1. 1. Pressure control or pump calibration error (unknown)
First check

Lavage pressure matches user-defined specifications

Test point
Pump display pressure readout
Expected
Pressure increases according to configured user specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Logo does not appear on display at startup

'Arthrex DualWave' message does not display when power switch is turned on

Ranked causes

  1. 1. Power supply failure (common)
First check

Verify mains power present at pump inlet

Test point
Power input connector
Expected
Mains voltage present
Tools
DMM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Menu button not functioning or settings not accessible

Menu button does not respond, digital buttons not working, or settings not accessible from menu

Ranked causes

  1. 1. Menu button not responding or menu system malfunction (unknown)
First check

Verify Menu button opens menu

Test point
Touchscreen Menu button
Expected
Menu opens when pressed and all options are selectable

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Menu button not responding or settings not accessible

Pressing Menu button when pump is stopped does not open menu; digital buttons within menu do not respond or settings are not accessible

Ranked causes

  1. 1. Menu button malfunction or touchscreen unresponsive (unknown)
First check

Menu button opens menu interface

Test point
Touchscreen menu screen
Expected
Menu interface appears when Menu button pressed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Over Pressure alarm does not appear patient safety

Over Pressure message fails to display when tubing is clamped and pressure exceeds 300 mmHg software limit

Ranked causes

  1. 1. Over Pressure alarm function not working (common)
First check

Verify pump pressure adjustment

Test point
Pump display pressure control
Expected
120 mmHg
Tools
Pump controls

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OVP Detect Fail message displayed high voltage

Overvoltage Protection Detect Fail message appears on pump display

Ranked causes

  1. 1. Overvoltage protection circuit failure (unknown)
First check

Verify OVP Detect Fail message does not appear

Test point
Display
Expected
No OVP Detect Fail message

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Failure message displayed patient safety

Power Failure (Stromversorgung) message appears on pump display

Ranked causes

  1. 1. Power supply fault (unknown)
First check

Verify Power Failure message does not appear

Test point
Display
Expected
No Power Failure message

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure Fault alarm does not appear patient safety

Pressure Fault message fails to display when tubing is clamped and pump is running at 40 mmHg until fault condition triggers

Ranked causes

  1. 1. Pressure Fault alarm function not working (common)
First check

Set pump pressure to 40 mmHg

Test point
Pump display pressure control
Expected
40 mmHg
Tools
Pump controls

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure Fault message displayed patient safety

Pressure Fault (Druckstörung) message appears on pump display

Ranked causes

  1. 1. Pressure system fault (unknown)
First check

Verify Pressure Fault message does not appear

Test point
Display
Expected
No Pressure Fault message

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump does not recognize outflow tubing

Pump fails to automatically recognize AR-6430 outflow tubing or does not switch to inflow/outflow mode

Ranked causes

  1. 1. Tubing detection sensor malfunction (unknown)
First check

Verify tubing is properly loaded

Test point
Outflow tubing port
Expected
AR-6430 tubing fully seated and recognized

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump does not start when Run button pressed

Run button appears but pump motor does not start when activated; pump should begin operation

Ranked causes

  1. 1. Run button malfunction or motor control failure (unknown)
First check

Pump motor starts when Run button pressed

Test point
Pump motor
Expected
Motor activates and runs

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump does not transition to inflow/outflow mode when tubing loaded

After loading outflow tubing (AR-6430) and pressing Stop, pump does not automatically recognize tubing or switch to inflow/outflow mode; features like Rinse, Cannula, Shaver should enable

Ranked causes

  1. 1. Tubing sensor malfunction or tubing not properly seated (unknown)
First check

Tubing detection and automatic mode switching

Test point
Pump mode indicator on display
Expected
Pump automatically recognizes tubing and transitions to inflow/outflow mode

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump motor does not stop when Stop button pressed patient safety

Stop button appears on display but pump motor does not halt when button activated during operation

Ranked causes

  1. 1. Stop button malfunction or motor control failure (unknown)
First check

Pump motor stops when Stop button pressed

Test point
Pump motor
Expected
Motor stops when Stop button pressed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Remote control not detected on display

Remote control icon does not appear on pump display after connection to remote control port

Ranked causes

  1. 1. Remote control not properly connected or faulty (unknown)
First check

Verify remote control connection and icon display

Test point
remote control port and display
Expected
remote control icon appears on display

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rinse button outflow not responding

Outflow does not increase according to user-defined specifications when Rinse button is pressed

Ranked causes

  1. 1. Rinse mode not activating outflow increase (unknown)
First check

Verify Rinse button increases outflow

Test point
Touchscreen Rinse button
Expected
Outflow increases according to user-defined specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rinse mode outflow does not increase per specifications

Pressing Rinse button does not increase outflow according to user-defined specifications; pump should adjust outflow pressure/flow

Ranked causes

  1. 1. Outflow pressure control malfunction or calibration error (unknown)
First check

Rinse mode outflow performance

Test point
Pump display and outflow sensor
Expected
Pump increases outflow according to user-defined specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Run button not starting pump

Pump does not start when Run button is pressed on touchscreen

Ranked causes

  1. 1. Run button not responding (unknown)
First check

Verify Run button starts pump

Test point
Touchscreen Run button
Expected
Pump motor starts when button is pressed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor Failure message displayed patient safety

Sensor Failure (Sensorausfall) message appears on pump display

Ranked causes

  1. 1. Sensor malfunction or disconnection (unknown)
First check

Verify Sensor Failure message does not appear

Test point
Display
Expected
No Sensor Failure message

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Shaver button suction and pressure options not responding

Cannot select between different suction levels or pressure increases when Shaver button is pressed

Ranked causes

  1. 1. Shaver button not responding or options not accessible (unknown)
First check

Verify Shaver button suction and pressure selection

Test point
Touchscreen Shaver button
Expected
Able to select between different suction levels and pressure increases

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Shaver mode suction and pressure options not selectable

Pressing Shaver button but user cannot select between different suction levels or pressure options; system should provide option selection

Ranked causes

  1. 1. Shaver mode button or suction/pressure selection malfunction (unknown)
First check

Shaver button functionality and suction/pressure selection

Test point
Touchscreen Shaver button and settings menu
Expected
Able to select between different suctions and pressure increases

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stop button not stopping pump motor

Pump motor does not stop when Stop button is pressed on touchscreen

Ranked causes

  1. 1. Stop button not responding (unknown)
First check

Verify Stop button halts motor

Test point
Touchscreen Stop button
Expected
Pump motor stops when button is pressed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen pressure adjustment not working

Pressure display does not increase or decrease when pressing + or - buttons on touchscreen

Ranked causes

  1. 1. Touchscreen buttons not functioning (unknown)
First check

Verify + button increases pressure

Test point
Touchscreen + button
Expected
Pressure increases in increments of 5

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen pressure control not responding to increment buttons

Pressing '+' or '-' buttons does not change pressure display; pressure should increase/decrease in 5-unit increments

Ranked causes

  1. 1. Touchscreen malfunction or unresponsive buttons (unknown)
First check

Pressure increment when '+' button pressed

Test point
Touchscreen display
Expected
Pressure increases in 5-unit increments

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tubing Out message appears incorrectly

Message 'Tubing Out' (tube not connected) displays when tubing is properly connected to the tubing sensor coupling

Ranked causes

  1. 1. Tubing sensor malfunction or false detection (unknown)
First check

Verify standard pressure preset is set

Test point
Pressure preset setting
Expected
Standard default pressure setting

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tubing Out message does not appear

Tubing Out message should appear when inlet tube is disconnected from sensor coupling

Ranked causes

  1. 1. Tubing sensor detection failure (unknown)
First check

Disconnect inlet tubing from sensor coupling

Test point
Tubing sensor coupling
Expected
Tubing Out message appears

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tubing Out message does not appear when expected

Tubing Out message should appear when tubing is not connected to the tubing sensor coupler

Ranked causes

  1. 1. Tubing sensor coupler malfunction (unknown)
First check

Disconnect tubing from sensor coupler

Test point
tubing sensor coupler
Expected
Tubing Out message appears

Tubing Out message does not appear when tubing disconnected

The *Tubing Out* message should display on the pump's operator display when inflow tubing is not connected to the tubing sensor coupler, but it does not appear.

Ranked causes

  1. 1. Tubing sensor malfunction or disconnection (unknown)
First check

Verify *Tubing Out* message displays when tubing is disconnected

Test point
Tubing sensor coupler
Expected
*Tubing Out* message appears on operator display

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Warning messages displayed on startup caution

One or more warning messages appear when pump is powered on

Ranked causes

  1. 1. System fault or sensor error detected during initialization (unknown)
First check

Verify no warning messages on startup

Test point
Display screen after power on
Expected
No warning messages present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Critical FailurePump has experienced a critical system failureDetected on pump startup or operationcritical
Door OpenPump door not fully closed or latchedDoor closure or latch switch not engaged when pump powered onwarning
Door OpenDoor or cover openDoor or access panel not fully closedwarning
Door OpenRoller housing door is not properly closedRaised when door sensor detects roller housing door is openwarning
Over PressurePump pressure has exceeded maximum safe thresholdOccurs when measured pressure exceeds 300 mmHgerror
Over PressurePump pressure has exceeded safe operating limitMeasured pressure exceeds software overpressure threshold of 300 mmHgcritical
Over PressureOverpressure condition detectedMeasured pressure exceeds 300 mmHg software limiterror
OVP Detect FailOverpressure protection system has failedDetected during pump operation when overpressure protection cannot engagecritical
Power FailurePower supply error detectedPower supply circuit malfunction or power input problemerror
Power FailurePump power supply has failed or is unstableDetected during pump operation or initializationcritical
Pressure FaultPump pressure system is not functioning correctlyDetected when pressure does not reach expected levels or fluctuates abnormallyerror
Pressure FaultPressure drop or fault detected during operationPressure falls or fails to build when expected during clamped tubing conditionerror
Pressure FaultPump pressure below minimum threshold for normal operationMeasured pressure drops below fault threshold at set pressureerror
Pressure FaultPressure fault condition detectedPressure cannot be maintained when tubing is clamped at setpointerror
Pressure FaultPump pressure monitoring circuit fault or pressure loss detectedPressure cannot be maintained at set level; transducer detects abnormal readingerror
Sensor FailureSensor malfunction detectedPressure sensor, door sensor, or tubing sensor circuit failureerror
Sensor FailureOne or more system sensors have failedDetected during pump initialization or operationerror
Tubing OutTubing not connected or tubing sensor failureTubing disconnected from sensor coupling or tubing sensor malfunctionwarning
Tubing OutTubing connector not properly seated or disconnectedTubing not connected or seated when pump powered onwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI