Arjo Voyager: troubleshooting and error codes

The ServiceForge knowledge base holds 2 documented symptoms for the Arjo Voyager, with 2 ranked causes, 2 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Ceiling lift moves without caregiver intervention patient safety

V4 and V4i lifts activate and move without operator input; potential electromagnetic interference (EMI) from facility RF sources

Ranked causes

  1. 1. Electromagnetic interference from facility RF signals (occasional)
First check

Verify IFU revision level on installed units

Test point
Device documentation
Expected
IFU #001.14000 rev 8 or #001.14150 rev 7

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit moves without caregiver intervention patient safety

Ceiling lift activates and moves without operator control, potentially due to electromagnetic interference from WiFi, radio transmitters, nurse call systems, pagers, or cellular phones in healthcare facilities

Ranked causes

  1. 1. Electromagnetic interference (EMI) from RF sources (occasional)
First check

Verify device serial number to determine if revised IFU is required

Test point
Device serial number plate
Expected
V4-4209-58957 or earlier requires updated IFU; V4-4209-58958 and later have current IFU

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.