Arjo Voyager 800: troubleshooting and error codes
The ServiceForge knowledge base holds 9 documented symptoms for the Arjo Voyager 800, with 16 ranked causes, 1 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Lift does not respond to hand control buttons patient safety
Pressing buttons on hand control produces no lift motion or response
Ranked causes
- 1. Lift positioned on charger (common)
- 2. Emergency stop activated (common)
- 3. Hand control improperly connected (common)
- 4. Hand control disconnection after troubleshooting (occasional)
- 5. Battery discharged (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charge indicator does not light on charger
Green-red charge indicator light not illuminating when lift positioned on charger
Ranked causes
- 1. Charger not plugged into power outlet (common)
- 2. Fuse is blown or defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger light remains red after overnight charge
Charge indicator stays red and does not turn green or white after full overnight charge
Ranked causes
- 1. Charger stopper unit defective (occasional)
- 2. Charger or battery assembly failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit starts and stops repetitively patient safety
Lift cycles on and off repeatedly during operation
Ranked causes
- 1. Load exceeds safe working load (common)
- 2. Battery low (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery dead after few transfers
Battery requires recharging after only 3 to 5 patient transfers
Ranked causes
- 1. Battery end-of-life (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger green but lift only works unloaded patient safety
Charge indicator green but lift stops when attempting transfer with patient load
Ranked causes
- 1. Battery failed or insufficient capacity (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger shows green but lift only does one or two transfers
Charge indicator green but lift power exhausts quickly after one or two uses
Ranked causes
- 1. Battery end-of-life (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift does not lower spreader bar at charger caution
Lift returns to charger when button pressed but fails to lower spreader bar; audible humming sound heard during operation.
Ranked causes
- 1. Charger has no power or is malfunctioning (common)
Measure AC voltage on charger contact points
- Test point
- Two contact points on charger (see diagram T-1)
- Expected
- 26-30 volts AC V AC, 26-30 V
- Tools
- Voltage meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift emits beep and red light flashes during operation patient safety
Unit beeps with red indicator during use; may stop lifting but lowering still possible
Ranked causes
- 1. Batteries are low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI