Arjo Tenor: troubleshooting and error codes
The ServiceForge knowledge base holds 39 documented symptoms for the Arjo Tenor, with 43 ranked causes, 51 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery low charge warning indicated
Red light on unit; battery charge insufficient for safe operation
Ranked causes
- 1. Battery charge depleted during use (common)
- 2. Regular charging intervals not followed (occasional)
Verify charger light status during battery charging
- Test point
- Charger indicator light
- Expected
- Constant light during charging; flashing light when fully charged
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging
Charger connected but light does not indicate charging status; battery remains discharged
Ranked causes
- 1. Battery not fully inserted or clicked into charger (common)
- 2. Charger not connected to grounded mains outlet (common)
Verify charger is plugged into grounded AC outlet
- Test point
- Charger main plug and wall socket
- Expected
- Charger main plug fully inserted into grounded outlet
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
M12 bolt worn or seized caution
Spreader bar does not move freely or excessive friction at pivot
Ranked causes
- 1. M12 bolt worn or deteriorated (occasional)
- 2. M12 bolt lacks lubrication (common)
Visually inspect M12 bolt condition
- Test point
- M12 bolt shaft and threads
- Expected
- No visible wear, corrosion or deterioration
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spreader bar loose or rattling patient safety
Movement or noise at spreader bar connection to centre section during operation
Ranked causes
- 1. M8 nuts loose on spreader bar attachment (common)
- 2. Grub screws loose (common)
Measure torque of eight M8 nuts on spreader bar arms
- Test point
- M8 nuts retaining spreader bar arms to centre section
- Expected
- 40Nm (30lbf ft)
- Tools
- 13mm Socket and Spanner, torque wrench
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal noise or rough actuator operation patient safety
Actuator produces unusual noise or operates roughly during raise/lower cycles
Ranked causes
- 1. Actuator wear or internal damage (common)
Operate handset or mast control switches to raise and lower actuator through full working range
- Test point
- Actuator during operation
- Expected
- Smooth operation with no abnormal noise
- Tools
- Handset or mast control switches
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Actuator bolts or nuts loose patient safety
Actuator moves relative to support or produces noise during operation
Ranked causes
- 1. Actuator top and bottom bolts or nyloc nuts loose (common)
Check security of actuator top and bottom bolts and nyloc nuts
- Test point
- Actuator top and bottom bolts and nyloc nuts
- Expected
- Tight, no movement
- Tools
- Appropriate socket and spanner
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Actuator does not hold position when depressed button released patient safety
Jib moves from restrained position after handset button is released and hand restraint removed
Ranked causes
- 1. Actuator brake or load-holding mechanism failure (common)
Position jib at approximately mid lifting height; lower with handset button depressed while restraining with hand
- Test point
- Handset button and jib position
- Expected
- Actuator motor stops; jib held by hand restraint
- Tools
- Handset
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Actuator making contact with hoist parts patient safety
Actuator contacts hoist structure other than upper and lower installation points during operation
Ranked causes
- 1. Misalignment or bending of actuator or hoist frame (occasional)
Visually check actuator for contact with hoist parts during raise/lower operation
- Test point
- Actuator and surrounding hoist structure
- Expected
- No contact except at upper and lower installation points
- Tools
- Visual inspection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Axle bearing worn or stiff caution
Spreader bar moves with difficulty or grinding noise at bearing
Ranked causes
- 1. Axle bearing lacks lubrication or is worn (common)
Visually inspect axle and bearing
- Test point
- Axle and bearing assembly
- Expected
- No visible wear, corrosion or scoring
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braked castors do not hold effectively patient safety
Braked castors do not prevent movement or do not engage properly
Ranked causes
- 1. Brake mechanism wear or damage (occasional)
Test effectiveness of braked castors by attempting to move hoist with brakes engaged
- Test point
- Braked castor locking mechanism
- Expected
- Hoist remains stationary; brakes hold under load
- Tools
- Manual load test
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Castors damaged or not operating freely patient safety
Castors show signs of damage, deterioration, or do not roll freely
Ranked causes
- 1. Castor wheel wear, bearing damage, or debris (common)
Visually inspect all castors and manually rotate wheels
- Test point
- Castor wheels and bearings
- Expected
- No visible damage or deterioration; wheels rotate freely
- Tools
- Visual and manual inspection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger cannot be mounted securely to wall patient safety
Charger installation fails; charger not firmly fastened or wobbles
Ranked causes
- 1. Screws are insufficient for charger weight or wall type (common)
Confirm wall and screw type are suitable for charger weight
- Test point
- Wall material and screw specifications
- Expected
- Screws rated for charger weight; wall type compatible with fasteners
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Emergency stop button does not stop all power functions patient safety
Pressing emergency stop button does not render all power functions inoperative
Ranked causes
- 1. Emergency stop switch malfunction or electrical fault (rare)
Push in emergency stop button
- Test point
- Emergency stop button
- Expected
- All power functions inoperative
- Tools
- Control panel inspection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handset or cable damaged patient safety
Handset cracked, cable frayed, or connection hook damaged
Ranked causes
- 1. Handset, securing hook or cable physically damaged (occasional)
Visually inspect handset, hook and cable
- Test point
- Handset unit, securing hook, and cable assembly
- Expected
- No cracks, breaks, fraying or visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hoist operates with noise or stiffness patient safety
Excessive noise or jerking motion when raising or lowering, or smooth operation lost
Ranked causes
- 1. Actuator internal wear or failure (occasional)
Operate hoist fully and observe for noise and smoothness
- Test point
- Hoist raising and lowering cycle
- Expected
- Smooth operation with no excessive noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect charger being used for batteries patient safety
Customer is not using the appropriate charger for the hoist batteries
Ranked causes
- 1. Wrong charger selected or charger substitution (occasional)
Verify charger model and battery type
- Test point
- Battery specification plate and charger label
- Expected
- Charger model matches battery type and capacity requirements
- Tools
- Service documentation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leg spread/close or overload device malfunction patient safety
Legs do not open and close smoothly or electrical overload device does not activate at extremes
Ranked causes
- 1. Overload device or leg mechanism wear or electrical fault (occasional)
Operate handset to open and close legs through full range
- Test point
- Legs and overload device
- Expected
- Legs open and close smoothly; overload device activates at extremes of travel
- Tools
- Handset
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose front castor patient safety
Front castor has insufficient torque, may affect mobility or stability
Ranked causes
- 1. Front castor bolts loose or insufficient torque (common)
Check torque on front castor
- Test point
- Front castor
- Expected
- 40Nm (30 lbf ft)
- Tools
- ST287 (19mm Crowfoot Spanner), torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose front castor bracket screws patient safety
Screws retaining front castor bracket to chassis leg are loose, may affect castor alignment or stability
Ranked causes
- 1. Front castor bracket screws loose or insufficient torque (common)
Check torque of front castor bracket screws
- Test point
- Front castor bracket to chassis leg
- Expected
- 60Nm (44 lbf ft)
- Tools
- ST55 (6mm Socket Allen Key), torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose leg pivot bolts patient safety
Leg pivot bolts have insufficient torque, may cause instability or leg movement during operation
Ranked causes
- 1. Leg pivot bolts loose or insufficient torque (common)
Check torque of leg pivot bolts
- Test point
- Leg pivot bolts
- Expected
- 60Nm (44 lbf ft)
- Tools
- Torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose M8 button head cap screw (rear castor bracket) patient safety
M8 button head cap screw retaining rear castor bracket is loose, may affect structural integrity
Ranked causes
- 1. M8 button head cap screw loose (common)
Check security of M8 button head cap screw at rear castor bracket
- Test point
- M8 button head cap screw
- Expected
- Secure
- Tools
- ST274 (5mm Socket Allen Key)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose M8 button head screw (front castor bracket) patient safety
M8 button head screw retaining front castor bracket is loose, may affect structural integrity
Ranked causes
- 1. M8 button head screw loose at front castor bracket (common)
Check security of M8 button head screw at front castor bracket
- Test point
- M8 button head screw
- Expected
- Secure
- Tools
- ST247 (5mm Socket Allen Key)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose rear castor patient safety
Rear castor has insufficient torque, may affect mobility or stability
Ranked causes
- 1. Rear castor bolts loose or insufficient torque (common)
Check torque on rear castor using extension spanner
- Test point
- Rear castor
- Expected
- 35Nm (26 lbf ft) actual applied is 40Nm (29 lbf ft) due to extension
- Tools
- ST302 (24mm Extension Spanner), torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose rear castor bracket screws patient safety
Screws retaining rear castor bracket to chassis leg are loose, may affect castor alignment or stability
Ranked causes
- 1. Rear castor bracket screws loose or insufficient torque (common)
Check torque of rear castor bracket screws
- Test point
- Rear castor bracket to chassis leg
- Expected
- 60Nm (44 lbf ft)
- Tools
- ST274 (5mm Socket Allen Key), ST55 (6mm Socket Allen Key), torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lower actuator support loose patient safety
Actuator support rattles or moves at mast attachment
Ranked causes
- 1. Lower actuator support M10 button head screws loose (common)
Measure torque of lower actuator support M10 button head screws
- Test point
- Two M10 button head screws retaining actuator support to mast
- Expected
- 60Nm (44 lbf ft)
- Tools
- ST55 (6mm Socket Allen Key), torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual override does not lower jib patient safety
Twisting the red manual override switch on lift actuator does not cause jib to lower
Ranked causes
- 1. Manual override mechanism jammed or not engaged (occasional)
Position jib at approximately mid lifting height; operate manual system override twist switch (red switch on actuator)
- Test point
- Red twist switch on lift actuator
- Expected
- Hoist lowers
- Tools
- Manual override switch
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mast support screws loose patient safety
Mast support rattles or moves relative to mast structure
Ranked causes
- 1. Mast to mast support screws loose (common)
Measure torque of mast to mast support screws
- Test point
- Two mast to mast support screws
- Expected
- 40Nm (30 lbf ft)
- Tools
- ST274 (5mm Socket Allen Key), torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mast support to chassis loose patient safety
Mast rattles or moves excessively at chassis mounting point
Ranked causes
- 1. Mast support to chassis screws loose (common)
Check security of mast support to chassis screws
- Test point
- Two mast support to chassis screws
- Expected
- Tight, no movement
- Tools
- ST274 (5mm Socket Allen Key)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mast support to positioning block loose patient safety
Mast support moves or rattles at positioning block connection
Ranked causes
- 1. Mast support to positioning block screws loose (common)
Measure torque of upper mast support to positioning block screw
- Test point
- Upper mast support to positioning block screw
- Expected
- 40Nm (30 lbf ft)
- Tools
- ST274 (5mm Socket Allen Key), torque wrench
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mast to jib pivot loose or stiff patient safety
Mast does not pivot smoothly or rattles at connection point
Ranked causes
- 1. Mast to jib pivot bolt or nyloc nut loose (common)
Check security of mast to jib pivot bolt and nyloc nut
- Test point
- Mast to jib pivot bolt and nyloc nut
- Expected
- Tight, no movement
- Tools
- Appropriate socket and spanner
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Plastic plugs missing from steering / push handle ends caution
Plastic plugs not installed or missing from both ends of handle
Ranked causes
- 1. Plastic plugs not installed or lost during use (occasional)
Visually verify plastic plugs present at both ends of handle
- Test point
- Each end of steering / push handle
- Expected
- Plastic plug installed at each end
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recoil clip loose or damaged caution
Spreader bar recoil clip rattles or does not hold position
Ranked causes
- 1. Recoil clip button head screws loose (common)
Measure torque of recoil clip button head screws
- Test point
- Two button head screws retaining recoil clip
- Expected
- 25Nm (18 lbf ft)
- Tools
- 3mm Allen Key, torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling damage or deterioration patient safety
Stitching, seams, fabric position compromised; plastic support clips cracked or deformed; SWL label illegible
Ranked causes
- 1. Incorrect laundry processes or normal wear (common)
Visually inspect stitching and seams
- Test point
- All sling seams and stitched areas
- Expected
- Stitching intact, seams in good position
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling identification missing or unclear patient safety
Sling lacks proper identification or dating per service procedure
Ranked causes
- 1. Sling not properly identified after inspection (occasional)
Confirm sling identification method
- Test point
- Sling serial number label or corner marking area
- Expected
- Sling has either serial number label or blank corner space for marking
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steering / push handle bolts or nuts loose caution
M10 bolts or nyloc nuts securing handle are not firmly tightened
Ranked causes
- 1. M10 bolts or nyloc nuts loose due to vibration or incomplete assembly (common)
Remove four plastic caps covering bolt heads and nyloc nuts
- Test point
- Steering / push handle attachment points
- Expected
- Four plastic caps removed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steering / push handle damaged caution
Physical damage visible on steering/push handle structure
Ranked causes
- 1. Impact or wear damage to handle (unknown)
Inspect steering / push handle for cracks, dents or deformation
- Test point
- Entire handle assembly
- Expected
- No visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steering / push handle weld integrity compromised caution
Welds on handle structure show signs of cracking or separation
Ranked causes
- 1. Weld crack or separation (unknown)
Visually inspect all welds on steering / push handle
- Test point
- Weld joints on handle
- Expected
- No cracks, gaps or separation visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Upper actuator support loose patient safety
Actuator support rattles or moves at jib attachment
Ranked causes
- 1. Upper actuator support M10 button head screws loose (common)
Measure torque of upper actuator support M10 button head screws
- Test point
- Two M10 button head screws retaining actuator support to jib
- Expected
- 60Nm (44 lbf ft)
- Tools
- ST55 (6mm Socket Allen Key), torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wear or damage to actuator installation points patient safety
Upper or lower actuator installation bolts or holes show signs of wear, deformation, or ovality
Ranked causes
- 1. Bolt hole oval or worn installation points (occasional)
Visually inspect upper and lower actuator installation points
- Test point
- Installation bolts and bolt holes
- Expected
- No wear, deformation, or oval bolt holes
- Tools
- Visual inspection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.