Arjo Sara Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 39 documented symptoms for the Arjo Sara Plus, with 47 ranked causes, 45 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery not adequately charged
Battery display shows yellow or red range; insufficient charge for operation.
Ranked causes
- 1. Battery depleted below operational threshold (common)
- 2. Battery display in yellow range (common)
Check battery charge level on display before use
- Test point
- Battery display indicator
- Expected
- Green or adequate charge range, not yellow or red
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Steel Stopper Bar (SSB) distance adjustment
Adjust SSB composition and distance to specification by removing or unscrewing adjustment caps based on measured distance
Ranked causes
- 1. SSB composition B: distance 480–485 mm (unknown)
- 2. SSB composition C: distance 460–480 mm (unknown)
Measure SSB distance
- Test point
- SSB composition
- Expected
- 480–485 mm (18.912–19.109 inch) or 460–480 mm (18.124–18.912 inch), 480–485 or 460–480
- Tools
- Measuring device
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Steel Stopper Bar distance adjustment
Steel Stopper Bar (SSB) distance out of specification; requires adjustment of caps and screws based on measured distance
Ranked causes
- 1. SSB distance between 480 mm and 485 mm (unknown)
- 2. SSB distance between 460 mm and 480 mm (unknown)
Measure Steel Stopper Bar distance
- Test point
- SSB composition B
- Expected
- > 480mm and < 485mm, 480–485 mm
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Castor mounting pin loose or castor tread damaged patient safety
Castor mounting pin not tight on chassis or castor tread shows damage
Ranked causes
- 1. Loose castor mounting pin (occasional)
- 2. Damaged castor tread (occasional)
Check castor mounting pin tightness
- Test point
- castor mounting pin on chassis
- Expected
- Pin tight on chassis and chassis legs
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Castors do not rotate freely or rear brakes do not lock patient safety
Castors are stiff, do not roll freely, or rear brakes fail to lock position.
Ranked causes
- 1. Dirt, debris, or wear in castor bearing (common)
- 2. Rear brake mechanism failure or wear (occasional)
Manually push lift forward and observe castor rotation
- Test point
- All four castors
- Expected
- All castors rotate freely with minimal resistance
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift does not operate through full range patient safety
Dual control panel or cable remote control handset cannot raise or lower lift through complete motion range smoothly.
Ranked causes
- 1. Control panel or handset malfunction or cable damage (unknown)
- 2. Battery low or not fully charged (common)
Verify lift operates smoothly from dual control panel through full range
- Test point
- Control panel buttons (raise/lower)
- Expected
- Full smooth motion from lowest to highest position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rear castor mounting thread damaged or loose caution
Rear castors fitted to the machine become loose and cause damage to the thread in the chassis, making the chassis unserviceable.
Ranked causes
- 1. Machine pushed over large thresholds (steps) (common)
- 2. Shock loads during transport to different locations (occasional)
Visually examine castor and chassis underside for damage
- Test point
- Rear castor mounting points on chassis
- Expected
- No visible damage, threads intact
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steel Stopper Bar (SSB) installation and alignment caution
Installing SSB on Sara Plus patient lift requires proper positioning of set screws and actuator centralisation stop unit to ensure frame stability and correct leg angle.
Ranked causes
- 1. Set screws not properly aligned or tightened during SSB installation (common)
- 2. Actuator centralisation stop unit bolts loose or misaligned (occasional)
Verify actuator centralisation stop unit screw torque
- Test point
- 2x Soc Cap Head screws JX68400.0
- Expected
- Tight, no rotation with socket wrench
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arc-Rest handgrips insecure patient safety
Handgrips on Arc-Rest (Supportive armrests with handgrips) not secure or loose
Ranked causes
- 1. Loose Arc-Rest handgrip fasteners (occasional)
Check Arc-Rest handgrip security
- Test point
- Arc-Rest handgrips and fasteners
- Expected
- Handgrips firmly secured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arc-Rest handgrips not secure patient safety
Handgrips on Arc-Rest (Supportive armrests with handgrips) are loose, wobbly, or not fully tightened.
Ranked causes
- 1. Vibration loosening handgrip fasteners during use (common)
Test Arc-Rest handgrips for security and movement
- Test point
- All handgrips on Arc-Rest
- Expected
- Handgrips firm and secure; no wobbling or movement when hand pressure applied
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charger connector loose, wires cut, or casing damaged high voltage
Visual inspection reveals loose connectors, cut or exposed wires, or damage to battery charger housing.
Ranked causes
- 1. Impact damage, wear, or environmental exposure to charger (occasional)
Visually inspect battery charger for damage
- Test point
- Charger connectors, wires, and casing
- Expected
- No loose connectors, cut wires, or casing damage visible; all connections secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charger damaged or defective
Visual examination reveals loose connectors, cut wires, or damage to the casing
Ranked causes
- 1. Loose connectors, cut wires, or casing damage (unknown)
Visually examine battery charger for loose connectors, cut wires, and casing damage
- Test point
- connectors, wiring, casing
- Expected
- No loose connectors, cut wires, or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery in critical low state (red flashing)
Battery display shows red flashing light indicating critical discharge; audible warning may sound.
Ranked causes
- 1. Battery critically depleted during operation (occasional)
Monitor battery display during operation for red flashing light
- Test point
- Battery display indicator
- Expected
- No red flashing light present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery insufficiently charged for use
Battery display shows low charge state; audible warning device sounds when battery discharge indicator reaches red light range
Ranked causes
- 1. Battery at low charge state (yellow or red display) (common)
Check battery charge state display
- Test point
- battery display indicator
- Expected
- Green; not in yellow or red range
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not adequately charged for weekly service
Battery charge state insufficient at start of weekly maintenance
Ranked causes
- 1. Battery not recharged weekly (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Steel Stopper Bar (SSB) composition C distance adjustment
SSB distance measurement is between 460mm and 480mm; adjustment caps must be removed
Ranked causes
- 1. SSB composition C distance range requires cap removal (common)
Measure distance between SSB components
- Test point
- Steel Stopper Bar
- Expected
- > 460mm (18.124 inch) and < 480mm (18.912 inch), +10/-10 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Steel Stopper Bar caps removal
Adjustment caps require removal when SSB distance is below 480 mm specification
Ranked causes
- 1. SSB distance below 480 mm specification (unknown)
Verify Steel Stopper Bar distance specification
- Test point
- SSB composition C
- Expected
- > 460mm and < 480mm, 460–480 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Castor wear or degradation caution
Castors deteriorate due to age or use, requiring replacement.
Ranked causes
- 1. Normal wear and tear on castors (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chassis legs do not open and close smoothly
Chassis legs lack full travel or movement is not smooth
Ranked causes
- 1. Mechanical friction or binding in leg mechanism (occasional)
Operate chassis legs through full open and close cycle
- Test point
- chassis leg hinges and guides
- Expected
- Full travel with smooth movement in both directions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cord cleats M4 socket set screws loose patient safety
M4 socket set screws retaining cord cleats in the Arc-Rest are not fully tightened or show looseness.
Ranked causes
- 1. Vibration loosening M4 fasteners during equipment use (common)
Inspect M4 socket set screws retaining cord cleats in Arc-Rest
- Test point
- Both M4 socket set screws in cord cleats
- Expected
- Screws fully tightened; no movement or play in cord cleats when hand pressure applied
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cord cleats socket set screws loose caution
M4 socket set screws retaining cord cleats in Arc-Rest not tight
Ranked causes
- 1. Loose M4 socket set screws (occasional)
Check M4 socket set screw tightness at cord cleats
- Test point
- 2 M4 socket set screws in Arc-Rest cord cleats
- Expected
- Both screws tight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cracks or sharp edges on exposed parts patient safety
Visible cracks or sharp edges on parts with patient or user contact could cause injury or become unhygienic
Ranked causes
- 1. Material degradation or damage to exposed parts (occasional)
Visually inspect all exposed parts for cracks and sharp edges
- Test point
- all exposed surfaces, especially patient contact areas
- Expected
- No cracks or sharp edges present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cracks or sharp edges on exposed parts in contact with patient patient safety
Cracks, sharp edges, or corrosion found on exposed surfaces that contact patient skin; hygiene or injury risk.
Ranked causes
- 1. Material fatigue, impact, or corrosion from repeated use or decontamination (occasional)
Visually inspect all exposed surfaces for cracks, sharp edges, or corrosion
- Test point
- All exterior surfaces in patient contact area
- Expected
- No cracks, sharp edges, or corrosion visible; smooth finish
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External fittings loose or fasteners not tight patient safety
Screws, nuts, bolts, or other external fittings are loose or not securely fastened.
Ranked causes
- 1. Vibration loosening fasteners during normal operation (common)
Inspect all external fittings and fasteners for tightness
- Test point
- All visible screws, nuts, and bolts on equipment exterior
- Expected
- All fasteners tight; no visible movement or play when hand pressure applied
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External fittings or fasteners loose caution
External fittings not secure or screws and nuts not tight
Ranked causes
- 1. Loose fasteners or fittings (occasional)
Inspect all external fittings and fasteners
- Test point
- all external fittings, screws, and nuts
- Expected
- All fittings secure and fasteners tight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front clevis pin socket set screw loose or incorrectly torqued patient safety
Socket set screw retaining front clevis pin in upper lift arm is not tightened to specification (6.4 Nm / 56 lb in).
Ranked causes
- 1. Fastener loosened from vibration or incorrect torque applied during assembly (occasional)
Verify front clevis pin socket set screw torque
- Test point
- Socket set screw on upper lift arm retaining front clevis pin
- Expected
- 6.4 Nm (56 lb in) torque; fastener snug and secure, per specification
- Tools
- Torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front clevis pin socket set screw loose or undertorqued caution
Front clevis pin retaining socket set screw not tightened to specification
Ranked causes
- 1. Socket set screw not torqued to specification (occasional)
Verify front clevis pin socket set screw torque
- Test point
- socket set screw in upper lift arm retaining front clevis pin
- Expected
- 6.4 Nm (56 lb in), as specified
- Tools
- torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handset or cable damaged
Remote handset or its cable shows damage or deterioration
Ranked causes
- 1. Handset or cable physical damage (occasional)
Inspect remote handset and cable condition
- Test point
- handset body and entire cable length
- Expected
- No damage to handset or cable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handset or cable damaged or non-functional
Cable remote control handset or its connecting cable is cut, bent, pinched, or non-responsive to button input.
Ranked causes
- 1. Cable damage from pinching, crushing, or wear during storage or use (occasional)
Visually inspect handset and cable for cuts, kinks, or damage
- Test point
- Cable along full length and handset housing
- Expected
- No cuts, pinches, or visible damage; cable flexible and intact
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instruction labels loose, missing, or unreadable patient safety
Equipment instruction labels are not firmly attached, missing, or text is illegible.
Ranked causes
- 1. Label adhesive failure, wear, or decontamination damage (occasional)
Inspect all instruction labels for attachment and readability
- Test point
- All equipment instruction labels
- Expected
- Labels firmly attached and text completely legible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instruction labels missing, detached, or unreadable caution
Safety or instruction labels not firmly attached or not readable
Ranked causes
- 1. Label detached or damaged (occasional)
Inspect all instruction labels
- Test point
- all instruction labels
- Expected
- All labels firmly attached and fully readable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leg spacing out of tolerance after SSB assembly
Inner distance between legs not within 540–550 mm specification after Steel Stopper Bar installation
Ranked causes
- 1. Set screw B not properly adjusted (common)
Measure inner distance between both legs
- Test point
- Distance A between leg inner edges
- Expected
- 540–550 mm (21.67–21.828 inch), ±5 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Legs do not open or close smoothly patient safety
Leg movement is stiff, jerky, or produces disturbing noises during opening and closing cycles
Ranked causes
- 1. Inadequate lubrication at leg joints (common)
Open and close legs completely
- Test point
- Both leg hinges
- Expected
- Smooth movement without disturbing noises
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loop lock assemblies damaged or knot in cone knob defective patient safety
Loop lock assemblies show damage, wear, or the knot within the cone knob is loose, frayed, or non-functional.
Ranked causes
- 1. Wear or damage to knot or assembly from repeated patient transfers (occasional)
Visually inspect loop lock assemblies and cone knob for damage
- Test point
- All loop lock assemblies and knots within cone knobs
- Expected
- Knots intact and tight; assemblies show no fraying or damage; smooth operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loop lock assembly integrity compromised patient safety
Loop lock assembly shows damage or knot within cone knob is defective
Ranked causes
- 1. Loop lock assembly or cone knob knot damage (occasional)
Inspect loop lock assembly integrity and cone knob knot
- Test point
- loop lock assemblies and cone knob knot
- Expected
- Assembly intact and knot secure within cone knob
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lower Override knob not properly set
Lower Override knob not turned fully clockwise or not finger tight
Ranked causes
- 1. Lower Override knob not at correct position or tension (occasional)
Check Lower Override knob position and tightness
- Test point
- Lower Override knob
- Expected
- Knob turned fully clockwise and finger tight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling attachment cords or loop lock assembly frayed or damaged patient safety
Sling attachment cords or loop lock assembly show fraying, wear, or visible damage; safety risk during patient lift.
Ranked causes
- 1. Normal wear or impact damage to cords or assembly during patient handling (common)
Visually inspect sling attachment cords and loop lock assembly
- Test point
- All cords and loop lock assembly connection points
- Expected
- No fraying, tears, or visible damage; smooth surface and secure attachment
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling frayed or damaged patient safety
Sling straps or attachment clips are frayed, torn, or clips are damaged; sling is unsafe for patient use.
Ranked causes
- 1. Normal wear and tear or impact damage to sling (common)
Visually inspect sling for fraying, tears, or damage to straps
- Test point
- All sling straps and attachment clips
- Expected
- No fraying, tears, or damage visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Straight Line Steering Guide not locking rear castor inline patient safety
Straight Line Steering Guide (where installed) fails to lock rear castor in-line position.
Ranked causes
- 1. Steering guide mechanism jam or wear (occasional)
Operate Straight Line Steering Guide and verify rear castor locks inline
- Test point
- Steering guide mechanism and rear castor position
- Expected
- Castor locks in-line position when guide is engaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.