Arjo Flowtron Universal AC600: troubleshooting and error codes
The ServiceForge knowledge base holds 17 documented symptoms for the Arjo Flowtron Universal AC600, with 29 ranked causes, 39 diagnostic checks with test points and expected values, and 3 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Low pressure alarm (LO) caution
Low pressure detected during cycle operation; device alarmed with audible signal, red lights flashing, and LCD message
Ranked causes
- 1. Garment hose disconnected at garment end (common)
- 2. Loose fitting garment (common)
- 3. Garment leak (common)
- 4. Kinked tube to transducer (occasional)
- 5. Damaged or disconnected internal tube (occasional)
- 6. Damaged hose set assembly (occasional)
- 7. Blocked air filters (occasional)
- 8. Motor and manifold assembly failed (rare)
- 9. Compressor assembly failure or low output (rare)
Hose connected at garment end?
7 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure alarm (HI) caution
High pressure detected during cycle; garment or hose kinked or obstructed
Ranked causes
- 1. Garment hose kinked (common)
- 2. Hose set assembly kinked (common)
Garment hose clear and unkinked?
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply PCB assembly handling high voltage
Safety precautions when handling power supply PCB assembly due to high voltage and electrostatic discharge risk
Ranked causes
- 1. Exposure to lethal voltages on power supply PCB assembly when mains power is connected (common)
- 2. Electrostatic discharge damage to power supply PCB assembly (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure transducer output incorrect
Output Pressure reading on LCD display does not match external pressure meter during calibration test.
Ranked causes
- 1. Zero calibration offset incorrect (common)
- 2. Gain calibration offset incorrect (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump fault alarm caution
Pump control circuit or power supply failure detected immediately; audible alarm sounds
Ranked causes
- 1. Control PCB faulty (occasional)
- 2. Power supply PCB faulty (occasional)
Control PCB functioning correctly?
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inlet filter requires replacement
Air inlet filter on compressor assembly needs periodic replacement for proper equipment operation
Ranked causes
- 1. Air inlet filter clogged or degraded (unknown)
Remove rear case from pump
- Test point
- Rear case
- Expected
- Rear case detached
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: pressure transducer accuracy
Verify pressure transducer output matches expected values on LCD display to confirm measurement accuracy
Ranked causes
- 1. Pressure transducer drift or inaccuracy (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hose set assembly damage caution
Screw cover hinges on hose set assemblies split, requiring full assembly replacement
Ranked causes
- 1. Repeated hose set removal (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hose set assembly integrity compromised caution
Elbow retaining cover screws may split if hose set assemblies are removed and reinstalled multiple times
Ranked causes
- 1. Repeated removal and reinstallation of hose set assemblies (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lo Alarm - Garment unplugged not triggered on disconnection patient safety
When garment is disconnected during operation, pump does not detect fault and display does not show Lo Alarm - Garment unplugged message within expected time.
Ranked causes
- 1. Pump pressure sensor not detecting garment disconnection (unknown)
Disconnect garment during pump operation
- Test point
- Garment connection to hose set assembly
- Expected
- Lo Alarm - Garment unplugged flashes on LCD within one complete inflation cycle (30-72 seconds)
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Motor and manifold assembly replacement caution
Procedure to install motor and manifold assembly onto control PCB and top case assemblies after removal or repair.
Ranked causes
- 1. Motor and manifold assembly removed for service (common)
O-ring correctly installed on each hose set elbow connector
- Test point
- Hose set elbow connectors
- Expected
- O-ring present and properly seated
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Fault - RFID error during calibration caution
LCD display shows Pump Fault - RFID error when attempting RFID calibration
Ranked causes
- 1. Hose set electrical connections not fitted correctly at pump end (common)
Inspect hose set electrical connections at pump end
- Test point
- pump connector interface on hose set assembly
- Expected
- connections fitted securely and correctly oriented
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump fault RFID alarm caution
Hose set electrical connections not properly fitted to pump; RFID tag not reading correctly
Ranked causes
- 1. Hose set electrical connections loose or disconnected (common)
Hose set electrical connections properly seated?
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump fault step alarm caution
Motor and manifold assembly malfunction detected; stepper motor or manifold failure
Ranked causes
- 1. Motor and manifold assembly failed (occasional)
Motor and manifold assembly functioning correctly?
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real time clock loses time and date settings
Control PCB battery disconnected longer than 20 seconds, clock requires reloading from PC-based service tool
Ranked causes
- 1. Battery disconnected or depleted (common)
Ensure battery swap completed within 20 seconds
- Test point
- 2-pin battery cable connector at PCB Con 5
- Expected
- Old battery disconnected and new battery connected with no gap exceeding 20 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace power supply PCB assembly high voltage
Procedure for installing power supply PCB onto front case with battery pack connector and sealing strips
Ranked causes
- 1. Power supply PCB assembly replacement or reinstallation (unknown)
Verify power supply PCB assembly is properly seated on front case
- Test point
- Fig 51, Item 10
- Expected
- Assembly flush with front case
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service maintenance due high voltage
Routine 12-month service inspection of pump components for damage, wear and operational function
Ranked causes
- 1. Scheduled 12-month maintenance interval (common)
Inspect outer casing
- Test point
- Outer casing
- Expected
- No damage; case secure
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Lo Alarm - Garment leak | Pressure loss detected in garment indicating possible leak | Pressure drop detected during garment inflation cycle | warning |
Lo Alarm - Garment unplugged | Garment has been disconnected from hose set assembly or connection lost | Garment disconnected during operation; no pressure detected within one inflation cycle | warning |
Pump Fault - RFID | RFID calibration or communication failure on pump hose set assemblies | occurs during RFID calibration process when electrical connections at pump are faulty | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.