AngioDynamics RITA 1500X: troubleshooting and error codes
The ServiceForge knowledge base holds 46 documented symptoms for the AngioDynamics RITA 1500X, with 61 ranked causes, 38 diagnostic checks with test points and expected values, and 35 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unintended burns or thermal injury to patient patient safety
Patient sustains burns at non-target sites, skin damage from Dispersive electrode adhesive, or thermal injury from RF energy.
Ranked causes
- 1. Cables contacting patient or other leads (common)
- 2. Improper Dispersive Electrode application or placement (common)
- 3. Skin-to-skin contact on patient (occasional)
- 4. Patient contact with grounded metal parts or high capacitance objects (occasional)
- 5. Incompatible or faulty accessory cable (occasional)
- 6. RF power on simultaneously with infusion (rare)
Position electrosurgical cables away from patient and other leads
- Test point
- Active and return electrosurgical leads
- Expected
- Leads positioned and secured not in contact with patient or other electrical equipment
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low RF power output or generator fails to deliver RF energy high voltage
RF Generator does not deliver expected power at normal settings; electrosurgical equipment fails to function correctly.
Ranked causes
- 1. Faulty application of Dispersive Electrode (common)
- 2. Failure of electrical lead (occasional)
- 3. Dispersive Electrode repositioning needed after generator shutdown (occasional)
Verify Dispersive Electrode contact with patient
- Test point
- Dispersive Electrode pad surface
- Expected
- Entire surface area in contact with patient skin
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pump detected on RS-232 Port B
LCD displays 'NO PUMP: (RF) TO START, (MODE) TO CANCEL'.
Ranked causes
- 1. Pump cable not connected or not properly seated (common)
- 2. Defective pump cable (occasional)
- 3. Pump disconnected or powered off (common)
Verify pump cable connected to RS-232 Port B
- Test point
- Port B on generator, RS-232 cable
- Expected
- Cable fully seated at both ends
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor tissue conduction with RF shutdown patient safety
LCD displays 'POOR TISSUE CONDUCTION – RF SHUT OFF' during procedure.
Ranked causes
- 1. Device not optimally positioned in tissue (common)
- 2. Poor connection of device, main cable, or dispersive pad (occasional)
- 3. Defective device or main cable (occasional)
Reposition device in tissue
- Test point
- Device electrode positioning
- Expected
- Device in optimized tissue contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incomplete ablation or lesion only partially destroyed patient safety
Ablation procedure completed but imaging shows lesion was not fully destroyed or only partially treated.
Ranked causes
- 1. High power settings caused tissue desiccation impeding ablation (occasional)
- 2. Device not fully deployed in target tissue (occasional)
Set appropriate RF power level
- Test point
- SET POWER (W) display
- Expected
- Lowest power setting suitable for intended ablation purpose
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infusion pump not responding patient safety
LCD displays 'PUMPS NOT RESPONDING'.
Ranked causes
- 1. Pump manually stopped or turned off within first 15 seconds or during procedure (occasional)
- 2. Loss of communication with pump or pump failure (occasional)
Verify pump is running continuously
- Test point
- Infusion pump display and flow indicator
- Expected
- Pump running; flow active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No device present
LCD displays 'NO DEVICE PRESENT' during operation when RF is activated.
Ranked causes
- 1. Device not properly connected to Device port (common)
- 2. Defective device cable or device (occasional)
Verify device connection to Device port
- Test point
- Device port connector
- Expected
- Device firmly seated and connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit displays error or fails self-test on power-up
Generator self-test fails after power on, error message or no display
Ranked causes
- 1. Software module not installed or defective (common)
- 2. Internal hardware malfunction (rare)
Verify software module is installed
- Test point
- Software Module Access panel on rear
- Expected
- Module visible and button not protruding
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature out of range caution
LCD displays 'AMBIENT TEMP OUT OF RANGE' at startup.
Ranked causes
- 1. Ambient temperature outside operating range during startup (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad device or connection with thermocouple error codes
LCD displays 'BAD DEVICE OR CONNECTION / TC' followed by position indicators (e.g., TC1----, TC-2---, etc.) or 'BAD DEVICE OR CONNECTION / NO ID, NO TCs'.
Ranked causes
- 1. Device or connection cable fault with specific thermocouple open or short (common)
Reconnect device and main cable
- Test point
- Device port, Main Cable, Dispersive Pad, Pad Adapter Cable
- Expected
- All connections seated firmly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device in contact with patient during purge patient safety
LCD displays 'REMOVE DEVICE FROM PATIENT, PRESS (MODE)' during purge operation.
Ranked causes
- 1. Device in contact with patient tissue or return pads during purge (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device not recognized
LCD displays 'DEVICE NOT RECOGNIZED BAD DEVICE OR CONNECTION / TC-----'.
Ranked causes
- 1. Device not properly connected or unrecognized device type (common)
Verify device connection and orientation
- Test point
- Device port
- Expected
- Device fully seated and aligned
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse blown or generator will not power on high voltage
No power to RF Generator, controls unresponsive
Ranked causes
- 1. Fuse open or blown (common)
Inspect fuses visually
- Test point
- Two fuse modules in Line Input module
- Expected
- Fuse element intact (no break in filament visible)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas embolism during laparoscopic ablation patient safety
Air or gas enters blood vessels during laparoscopic procedure, potentially causing vascular obstruction.
Ranked causes
- 1. Device activation without tissue contact during laparoscopic procedure (rare)
Verify device contact with target tissue during laparoscopic ablation
- Test point
- Device position relative to target tissue
- Expected
- Device in direct contact with target lesion; efficiency reading 6-10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injury to adjacent non-target structures or critical nerves patient safety
RF ablation damages structures near the target site including nerves, causing incontinence or other serious complications.
Ranked causes
- 1. Inadequate pre-procedure evaluation for proximity of target to critical structures (occasional)
Verify minimum distance from device to non-target structures
- Test point
- Device placement relative to critical structures
- Expected
- At least 1 cm clearance from non-target structures, particularly nerve structures
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal load impedance too high
LCD displays 'SYSTEM FAILURE 5: INTERNAL LOAD HIGH TURN POWER OFF'.
Ranked causes
- 1. Internal impedance measurement anomaly or circuit fault (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal load impedance too low
LCD displays 'SYSTEM FAILURE 4: INTERNAL LOAD LOW TURN POWER OFF'.
Ranked causes
- 1. Internal impedance measurement anomaly or circuit fault (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid spill or water damage to RF Generator high voltage
Unit may be damaged by moisture intrusion, connections corroded, shorts present
Ranked causes
- 1. Liquid allowed to enter unit enclosure (rare)
Inspect external moisture
- Test point
- RF Generator exterior and vents
- Expected
- Dry, no visible moisture
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loop overrun failure
LCD displays 'SYSTEM FAILURE 8: LOOP OVERRUN TURN POWER OFF' during operation.
Ranked causes
- 1. Real-time control loop execution timeout (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of pain relief durability after bone metastasis ablation
Pain relief from RF ablation of painful bone metastases diminishes after initial treatment period.
Ranked causes
- 1. Limited evidence of pain relief durability beyond 4 weeks (unknown)
No connection or poor connection with device
LCD displays 'NO CONNECTION OR POOR CONNECTION'.
Ranked causes
- 1. Needle device not properly connected to generator (common)
Verify device and main cable connections
- Test point
- Device port, Main Cable connectors
- Expected
- All connections fully seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pathologic fracture after bone metastasis ablation patient safety
Ablated bone fractures during or after treatment, particularly in weight-bearing bones.
Ranked causes
- 1. Impending fracture present in weight-bearing bone (occasional)
Evaluate for impending fracture before ablation
- Test point
- Weight-bearing bone lesion imaging
- Expected
- No evidence of impending fracture on imaging assessment
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor connection patient safety
LCD displays 'POOR CONNECTION' during ablation procedure.
Ranked causes
- 1. Needle tissue contact out of allowable range (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply failure high voltage
LCD displays 'SYSTEM FAILURE 3: POWER SUPPLY TURN POWER OFF'.
Ranked causes
- 1. Internal power supply circuit failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump connected during active RF procedure patient safety
LCD displays 'PUMP CONNECTED DURING PROCEDURE- RESTART'.
Ranked causes
- 1. Infusion pump connected to generator while RF active without proper initialization (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RAM memory check failure
LCD displays 'SYSTEM FAILURE 1: RAM CHECK TURN POWER OFF' at self-test.
Ranked causes
- 1. Internal RAM memory failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Re-heating to target temperature patient safety
LCD displays 'RE-HEATING TO TARGET TEMP' during procedure.
Ranked causes
- 1. Temperature decreased below target during ablation (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF Generator shuts down or does not activate high voltage
System fails to turn on, does not respond to RF ON/OFF button, or automatically shuts down during operation.
Ranked causes
- 1. Target temperature not reached within 30 minutes (occasional)
Monitor timer and temperature during ablation
- Test point
- TIMER (min) and SET TEMP (°C) displays
- Expected
- Target temperature reached and maintained; timer counts down from set time
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF off current detection failure high voltage
LCD displays 'SYSTEM FAILURE 14: RF OFF: CURRENT DET TURN POWER OFF'.
Ranked causes
- 1. Current detection circuit malfunction when RF off (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF off voltage detection failure high voltage
LCD displays 'SYSTEM FAILURE 13: RF OFF: VOLTAGE DET TURN POWER OFF'.
Ranked causes
- 1. Voltage detection circuit malfunction when RF off (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF pulse high voltage failure high voltage
LCD displays 'SYSTEM FAILURE 15: RF PULSE: HI VOLTAGE TURN POWER OFF'.
Ranked causes
- 1. RF output voltage exceeds high limit during pulse (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF pulse low voltage failure high voltage
LCD displays 'SYSTEM FAILURE 16: RF PULSE: LO VOLTAGE TURN POWER OFF'.
Ranked causes
- 1. RF output voltage below low limit during pulse (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF short patient safety
LCD displays 'RF SHORT' during ablation procedure.
Ranked causes
- 1. Needle tissue contact out of allowable range or system setup error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RF was turned off
LCD displays 'RF WAS TURNED OFF' during operation.
Ranked causes
- 1. User manually turned off RF energy via RF ON/OFF button or foot pedal (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROM memory check failure
LCD displays 'SYSTEM FAILURE 2: ROM CHECK TURN POWER OFF / CRC = XXXX'.
Ranked causes
- 1. Internal ROM memory or firmware corruption (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Shorted panel key(s)
LCD displays 'SYSTEM FAILURE 12: SHORTED PANEL KEY(S) POWER OFF, PRESS KEYS ONCE AND POWER ON'.
Ranked causes
- 1. One or more front panel keys stuck or shorted (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software module installation failure
LCD does not display information after power-on following software module installation
Ranked causes
- 1. Software module not properly inserted or oriented (common)
Verify software module orientation
- Test point
- Software module arrow and label
- Expected
- Arrow facing insertion slot, wording right side up
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software module not connected
LCD displays 'SOFTWARE MODULE NOT CONNECTED TURN POWER OFF' at self-test startup.
Ranked causes
- 1. Software module not fully inserted or incorrect orientation (common)
Verify software module seating and orientation
- Test point
- Software module slot
- Expected
- Module fully inserted with arrow symbol aligned and cover closed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spurious interrupt failure
LCD displays 'SYSTEM ERROR 11: SPURIOUS INTERRUPT TURN POWER OFF' during operation.
Ranked causes
- 1. Unexpected interrupt signal in control electronics (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System on too long caution
LCD displays 'SYSTEM FAILURE 10: SYSTEM ON TOO LONG TURN POWER OFF'.
Ranked causes
- 1. Unit left on continuously without performing self-test on startup (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System temperature too high caution
LCD displays 'SYSTEM TEMP TOO HIGH' during self-test or operation.
Ranked causes
- 1. RF Generator vents blocked restricting airflow (common)
Inspect RF Generator vents for blockage
- Test point
- Back and bottom vents
- Expected
- Vents clear and unobstructed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Target temperature not reached patient safety
LCD displays 'TARGET TEMP NOT REACHED' during procedure.
Ranked causes
- 1. Device unable to reach target temperature during ablation (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature reference failure
LCD displays 'SYSTEM FAILURE 6: TEMP REF BAD TURN POWER OFF'.
Ranked causes
- 1. Temperature sensor or reference circuit malfunction (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Time error failure
LCD displays 'SYSTEM FAILURE 9: TIME ERROR TURN POWER OFF' during operation.
Ranked causes
- 1. Real-time clock or timing reference malfunction (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unsupported pump connected
LCD displays 'PUMP NOT SUPPORTED, SEE USER'S MANUAL'.
Ranked causes
- 1. Non-IntelliFlow infusion pump connected to generator (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Watchdog timer failure
LCD displays 'SYSTEM FAILURE 7: WATCHDOG TURN POWER OFF' during operation.
Ranked causes
- 1. Watchdog timer circuit or software timeout (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AMBIENT TEMP OUT OF RANGE | Ambient operating temperature outside acceptable range at startup | Power-on self-test detects ambient temperature outside specifications | warning |
BAD DEVICE OR CONNECTION | Device communication error with specific thermocouple open/short indicator | Device port read fails or thermocouple sensor open/short detected during runtime | error |
DEVICE NOT RECOGNIZED | Connected device type not recognized or not supported by generator software | Device port detects device with unknown or mismatched ID code | error |
IMPEDE-OUT RECOVERY | Loss of tissue impedance contact detected; automatic recovery sequence initiated | Runtime impedance monitoring detects sudden impedance rise above threshold during ablation | warning |
NO CONNECTION OR POOR CONNECTION | Needle device not properly connected to generator or connection intermittent | Device port detects communication timeout or signal loss during startup or operation | error |
NO DEVICE PRESENT | No RF ablation device detected on Device port at startup or during operation | Device port communication failure during initialization or runtime checks | error |
NO PUMP | No infusion pump detected on RS-232 Port B at startup or during operation | Device port scan at startup or runtime detects no pump on Port B | warning |
POOR CONNECTION | Needle tissue contact impedance outside acceptable range | Runtime impedance monitoring detects contact resistance beyond limits | warning |
POOR TISSUE CONDUCTION | Tissue conductivity too low to deliver RF energy safely; RF automatically shut off | Runtime conduction testing shows impedance rise above safe maximum during active ablation | error |
PUMP CONNECTED DURING PROCEDURE | Infusion pump connected to generator while RF active without proper system initialization | Pump detected on RS-232 port after RF already activated without pump present | error |
PUMP NOT SUPPORTED | Connected infusion pump model not supported by generator software | Pump ID code detected is not in supported device list | error |
PUMPS NOT RESPONDING | Infusion pump not responding to generator commands or communication lost | RS-232 command timeout or no pump heartbeat detected after procedure start | error |
RE-HEATING TO TARGET TEMP | Temperature dropped below target during procedure and re-heating is in progress | Runtime temperature monitoring detects drop below setpoint during active ablation | info |
REMOVE DEVICE FROM PATIENT | Ablation device in contact with patient tissue or return pads during purge operation | Purge cycle detects device impedance/contact signal while purge active | critical |
RF SHORT | Detected RF short circuit or impedance anomaly preventing energy delivery | Safety monitoring detects short-circuit-like condition or impedance collapse | error |
RF WAS TURNED OFF | RF energy was manually disabled by operator via RF ON/OFF button or foot pedal | User presses RF ON/OFF button or foot pedal during active ablation | info |
SOFTWARE MODULE NOT CONNECTED | Software module not fully seated or incorrectly oriented at startup self-test | Power-on self-test failure when module slot is empty or improperly inserted | error |
SYSTEM ERROR 11 | Spurious or unexpected interrupt signal detected in control electronics | Interrupt handler receives interrupt from non-enabled source or at unexpected time | error |
SYSTEM FAILURE 1 | RAM memory self-test failure | Power-on self-test detects RAM read/write error | critical |
SYSTEM FAILURE 10 | Generator left on continuously without performing startup self-test | System detects extended continuous operation without power-on reset | warning |
SYSTEM FAILURE 12 | One or more front panel keys detected as stuck or shorted | Self-test detects stuck key contact during power-on | error |
SYSTEM FAILURE 13 | Voltage detection circuit failure during RF off state | Self-test or runtime monitoring detects voltage sensor malfunction when RF disabled | critical |
SYSTEM FAILURE 14 | Current detection circuit failure during RF off state | Self-test or runtime monitoring detects current sensor malfunction when RF disabled | critical |
SYSTEM FAILURE 15 | RF pulse output voltage exceeds high limit during energy transmission | Output voltage monitoring detects peak voltage above maximum allowed during RF pulse | critical |
SYSTEM FAILURE 16 | RF pulse output voltage below low limit during energy transmission | Output voltage monitoring detects peak voltage below minimum during RF pulse attempt | critical |
SYSTEM FAILURE 2 | ROM or firmware self-test failure with CRC value | Power-on self-test detects ROM read error or firmware checksum mismatch | critical |
SYSTEM FAILURE 3 | Power supply circuit self-test failure | Power-on or runtime self-test detects power supply regulation fault | critical |
SYSTEM FAILURE 4 | Internal load impedance measurement below lower limit | Self-test detection of low impedance in internal RF load circuit | error |
SYSTEM FAILURE 5 | Internal load impedance measurement above upper limit | Self-test detection of high impedance in internal RF load circuit | error |
SYSTEM FAILURE 6 | Temperature sensor reference circuit failure | Self-test detection of temperature reference sensor malfunction | error |
SYSTEM FAILURE 7 | Watchdog timer failure detected during operation | Watchdog timer interrupt not cleared by control software within timeout period | error |
SYSTEM FAILURE 8 | Real-time control loop execution overrun during operation | Control loop task fails to complete within allocated time cycle | error |
SYSTEM FAILURE 9 | Real-time clock or timing reference failure during operation | Timing reference query returns invalid or out-of-range value | error |
SYSTEM TEMP TOO HIGH | RF Generator internal temperature exceeds thermal limit during self-test or operation | Temperature monitoring detects sensor reading above acceptable limit | warning |
TARGET TEMP NOT REACHED | Ablation device failed to reach target temperature during procedure | Procedure timer expires without reaching target temperature setpoint | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI