Alaris Pump Module 8100: troubleshooting and error codes
The ServiceForge knowledge base holds 3 documented symptoms for the Alaris Pump Module 8100, with 4 ranked causes, 3 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Pump Module keypad lifting or stuck keys patient safety
Keypad keys are lifting, stuck, or unresponsive during operation
Ranked causes
- 1. Supplier manufacturing process issue with keypad assembly (common)
- 2. Defective Pump Module Door Assembly Replacement Kit (common)
Inspect keypad for lifting, sticking, or unresponsive keys
- Test point
- Pump Module keypad
- Expected
- All keys move freely and respond to input
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad key stuck in pressed state patient safety
One or more keys become stuck in constantly pressed state due to fluid ingress. Alaris PC unit alarms and module exhibits Channel Error on scrolling marquee, resulting in infusion interruption.
Ranked causes
- 1. Keypad lifting allowing fluid ingress (common)
Observe for Channel Error alarm and marquee display
- Test point
- Alaris PC unit and pump module scrolling marquee
- Expected
- No Channel Error displayed; no PC alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad keys unresponsive patient safety
One or more keys on pump module keypad become unresponsive due to fluid ingress from lifting keypad. Module continues infusing as programmed. No alarm on PC unit; programming changes must be made on PC unit.
Ranked causes
- 1. Keypad lifting allowing fluid ingress (common)
Inspect keypad for signs of lifting
- Test point
- Pump module keypad surface
- Expected
- Keypad flush with module housing, no gaps or elevation
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Channel Error | Keypad key is stuck in pressed state on pump module | Stuck key on Alaris Pump Module keypad due to fluid ingress | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.