Alaris PC 8000: troubleshooting and error codes

The ServiceForge knowledge base holds 39 documented symptoms for the Alaris PC 8000, with 105 ranked causes, 1 diagnostic checks with test points and expected values, and 476 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Check syringe message displayed patient safety

Syringe module reports issue with syringe installation, positioning, or sensor engagement

Ranked causes

  1. 1. Syringe not properly positioned or installed (common)
  2. 2. Plunger head not properly engaged (common)
  3. 3. Barrel clamp not closed (common)
  4. 4. Flange detect sensor not engaged (occasional)
  5. 5. Binary switches miscalibrated (occasional)
  6. 6. Force sensor harness disconnected (occasional)
  7. 7. Syringe size sensor harness disconnected (occasional)
  8. 8. Linear sensor harness disconnected (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air in line detected patient safety

AIL (Air-In-Line) sensor detects air bubbles in administration set or fluid line

Ranked causes

  1. 1. Air present in administration set (common)
  2. 2. AIL sensor threshold set too sensitive (occasional)
  3. 3. AIL transmitter or receiver contaminated or misaligned (occasional)
  4. 4. AIL connector loose or damaged (occasional)
  5. 5. AIL assembly defective (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Instrument will not turn on

Unit does not power up when power button is pressed or when plugged into AC

Ranked causes

  1. 1. No AC power supplied (common)
  2. 2. Battery missing or defective (common)
  3. 3. Blown fuse (occasional)
  4. 4. Off-line switcher failure (occasional)
  5. 5. Power supply board failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check IV set message displayed patient safety

Alarm or message indicating administration set installation issue or sensor problem

Ranked causes

  1. 1. Administration set not correctly installed (common)
  2. 2. Door ajar or FLO-Stop sensor malfunction (occasional)
  3. 3. Patient side occlusion sensor miscalibrated (occasional)
  4. 4. Fluid side occlusion sensor miscalibrated or defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door open alarm triggered

Pump module door is detected as open; infusion cannot proceed until door is closed and latched

Ranked causes

  1. 1. Administration set not correctly installed (common)
  2. 2. Door latch failure or damage (occasional)
  3. 3. Door sear worn or broken (occasional)
  4. 4. AIL assembly sensor failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion detected in syringe module patient safety

Syringe module detects pressure increase indicating blockage on fluid or patient side

Ranked causes

  1. 1. Administration set installed incorrectly or tubing obstructed (common)
  2. 2. Binary switches or pressure sensors miscalibrated (occasional)
  3. 3. Pressure sensor miscalibrated or defective (occasional)
  4. 4. Force sensor miscalibrated or defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Accuracy verification fails patient safety

Pump module fails accuracy verification test for infusion rate delivery

Ranked causes

  1. 1. Mechanical damage or improper administration set installation (common)
  2. 2. Calibration drift or incorrect calibration (occasional)
  3. 3. Pump mechanism defective or worn (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery depleted or low

Battery charge insufficient for operation, may trigger low battery alarm

Ranked causes

  1. 1. Battery requires reconditioning (common)
  2. 2. Battery defective or worn out (occasional)
  3. 3. Logic board failure (rare)
First check

Plug instrument into AC power

Test point
AC inlet
Expected
Unit receives AC power and begins charging

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display problem (SpO2 Module) caution

Display screen not functioning or showing incorrect information

Ranked causes

  1. 1. Loose or disconnected display cable (common)
  2. 2. Defective display module (occasional)
  3. 3. Defective logic board (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display problem or malfunction

LCD display not functioning correctly, text not visible, contrast issues, or display unresponsive

Ranked causes

  1. 1. Display cable connection loose or damaged (common)
  2. 2. Display module defective (occasional)
  3. 3. Logic board failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive not engaged patient safety

Syringe module drive mechanism is not engaging the syringe plunger; infusion cannot proceed

Ranked causes

  1. 1. Binary switches miscalibrated or force sensor issue (common)
  2. 2. Lower housing or carriage assembly mechanical failure (occasional)
  3. 3. Actuator knob or cam lock binding or worn (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent operation or dropout patient safety

Unit operates normally at times but occasionally stops responding or cuts out unexpectedly

Ranked causes

  1. 1. Loose connection to logic board (common)
  2. 2. PC unit logic board failure (occasional)
  3. 3. Module logic board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Key stuck alarm

Alarm triggered indicating a keypad key appears to be pressed or stuck, preventing normal operation

Ranked causes

  1. 1. Transient software glitch (common)
  2. 2. Keypad assembly defective (occasional)
  3. 3. Logic board failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD contrast is dark or light

Display text is difficult to read due to contrast being too dark or too light

Ranked causes

  1. 1. Contrast setting not optimized (common)
  2. 2. Main LCD module failure (occasional)
  3. 3. Logic board failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump chamber blocked patient safety

Tubing inside pump chamber is blocked, clogged, or obstructed, preventing fluid movement

Ranked causes

  1. 1. Tubing blockage in pump chamber (common)
  2. 2. Patient side pressure sensor miscalibrated (occasional)
  3. 3. Fluid side pressure sensor miscalibrated or defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe not recognized patient safety

System cannot identify syringe model or type

Ranked causes

  1. 1. Incompatible or wrong syringe type (common)
  2. 2. Syringe Size Sensor Assembly malfunction (occasional)
  3. 3. Logic Board firmware mismatch (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Will not turn on (SpO2 Module) caution

Module cannot be powered on or powered on but displays no function

Ranked causes

  1. 1. Blown fuse (common)
  2. 2. Defective off-line switcher (occasional)
  3. 3. Defective power supply board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bad sensor (SpO2) patient safety

Pulse oximetry sensor or patient cable not functioning correctly

Ranked causes

  1. 1. Incorrect sensor or patient cable type (common)
  2. 2. Patient cable malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel disconnected patient safety

PCA or Syringe module is not connected to PC Unit or connection is lost; module not recognized

Ranked causes

  1. 1. Module not physically connected to PC Unit (common)
  2. 2. IUI connector damaged or worn (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel errors detected patient safety

PCA or Syringe module reports communication or operational errors on its channel

Ranked causes

  1. 1. Specific error code indicates root cause (unknown)
  2. 2. Module requires calibration or verification (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check sensor - electrical or optical interference patient safety

Pulse oximetry sensor reading corrupted by external electrical or optical signals

Ranked causes

  1. 1. External electrical or optical interference (occasional)
  2. 2. Patient cable defect (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check sensor - excessive ambient light (Model 8210) patient safety

Pulse oximetry sensor overwhelmed by environmental light

Ranked causes

  1. 1. Excessive ambient light exposure (common)
  2. 2. Patient cable malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check sensor - high pulse amplitude (Model 8210) patient safety

Pulse oximetry reading corrupted by patient motion or artifact interference

Ranked causes

  1. 1. Motion artifact from excessive pulse amplitude at current site (common)
  2. 2. Patient cable malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Close door alarm or message

Door is not fully closed or latch is not engaged; system prevents operation until resolved

Ranked causes

  1. 1. Door not fully closed or latched (common)
  2. 2. Door sensor misalignment or miscalibration (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Downstream occlusion alarm patient safety

Pump detects pressure increase on the patient/downstream side of the tubing, indicating blockage

Ranked causes

  1. 1. Tubing kinked, clogged filter, or other obstruction (common)
  2. 2. Pressure verification needed or sensor miscalibrated (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

FLO-Stop open or door not closed

FLO-Stop mechanism is open or door closure is incomplete; system prevents infusion

Ranked causes

  1. 1. Roller clamp open or door not fully closed (common)
  2. 2. FLO-Stop or door sensor malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Instrument malfunction patient safety

Unit exhibits abnormal behavior, fault detected, or unexpected error conditions

Ranked causes

  1. 1. Transient software error or power anomaly (common)
  2. 2. Specific fault indicated in alarm history (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent operation (SpO2 Module) patient safety

Device operates inconsistently with intermittent failures or disconnections

Ranked causes

  1. 1. Loose or intermittent connection to Logic Board (common)
  2. 2. Defective logic board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Key stuck alarm (SpO2 Module) caution

Keypad stuck or unresponsive, triggering system alarm

Ranked causes

  1. 1. Stuck keypad contact (common)
  2. 2. Defective logic board (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No sensor (SpO2) patient safety

Pulse oximetry sensor not detected or connected

Ranked causes

  1. 1. Sensor not attached to patient cable (common)
  2. 2. Patient cable malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No signal (SpO2) patient safety

Pulse oximetry sensor connected but receiving no signal from patient

Ranked causes

  1. 1. Incorrect sensor placement (common)
  2. 2. Patient cable malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occluded fluid side or empty container patient safety

Fluid side of pump is obstructed or medication container is empty; no fluid can be drawn

Ranked causes

  1. 1. Fluid side obstruction or empty drip chamber (common)
  2. 2. Pressure sensor miscalibrated or defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occluded patient side patient safety

Patient/downstream side of pump is blocked, preventing fluid delivery to patient

Ranked causes

  1. 1. Patient side obstruction or kinked tubing (common)
  2. 2. Patient side pressure sensor miscalibrated or defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PCA handset stuck patient safety

PCA module patient control handset is not responsive, stuck, or cannot be disconnected

Ranked causes

  1. 1. Handset not correctly connected or handset defective (common)
  2. 2. Handset hardware failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor off (SpO2) patient safety

Pulse oximetry sensor became detached or powered off

Ranked causes

  1. 1. Sensor detached from patient (common)
  2. 2. Patient cable malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe drive-head error patient safety

Drivetrain blockage or plunger head cannot achieve full travel with cam lock engagement/disengagement

Ranked causes

  1. 1. Drivetrain blockage (common)
  2. 2. Plunger head travel restriction or cam lock issue (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe empty patient safety

Syringe depleted or sensor not detecting syringe level

Ranked causes

  1. 1. Force Sensor or Linear Sensor malfunction (occasional)
  2. 2. Plunger head travel restriction or cam lock issue (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Disposable disconnected (EtCO2) patient safety

Microstream disposable not connected to EtCO2 module

Ranked causes

  1. 1. Disposable not attached to instrument (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient not detected (EtCO2) patient safety

EtCO2 module cannot detect patient connection or breathing

Ranked causes

  1. 1. Disposable malfunction or improper connection (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1000General Failureunknown
100.1010Error detected during CPU testerror
100.1070Hardware watchdog test failureerror
100.1080Data written to I/O-protected RAM not verifying correctlyerror
100.1090Error detected during DRAM testerror
100.1100Error detected during SRAM testerror
100.1110Flash file device CRC test failureerror
100.1120Boot flash device CRC test failureerror
100.1130Time read from battery-backed RTC not changingerror
100.1140Time read from RTC not changingerror
100.1180RTC differs from OS clock by more than 1 minuteerror
100.1210Option board ID conflict detectederror
100.1220Processor exception or corrupted codeerror
100.1230Power supply or auxiliary processor unexpectedly reseterror
100.1240Power supply lost powererror
100.1250Unknown board type detected in option board sloterror
100.2070Power losterror
100.5010Boot block software version not compatibleerror
100.50208051 processor not reporting software version immediatelyerror
100.5070Software access key invalid in system config modeerror
100.6010Failed attempt to write to battery-backed RAMerror
100.6020Critical value data integrity failureerror
100.6110Time/date exceeds maximum RTC dateerror
100.6130Configuration data invalid in normal modeerror
100.6160Unexpected hardware interrupt in processorerror
100.6190Current facility Data Set damagederror
100.6400General CBIT failureerror
100.8000Operating system errorerror
1010CPU Failureunknown
1020ROM Failureunknown
1030RAM Failureunknown
1040Flash Failureunknown
1050Timebase Failureunknown
1060Watchdog Failureunknown
1070Watchdog Test Failureunknown
1080Protected RAM Failureunknown
1090DRAM Failureunknown
1100SRAM Failureunknown
110.1010CPU failure detected during POSTerror
110.1020ROM failure detected during POSTerror
110.1160Internal RAM failure detected during POSTerror
110.2000Communication errorerror
110.2040Unknown message typeerror
110.3000Key or tamper switch stuck closederror
110.50108051 software version not compatible with flashed PCerror
110.5020Processor not reporting version after System Onerror
110.6000One or more POST failures detectederror
110.6010RAM failure detected during POSTerror
110.6020Keypad failure detected during POSTerror
110.6110Projected life span exceededSystem time rolled over past maximum dateerror
1110Flash File Failureunknown
111.2000Keypad Processor communication errorKeypad Processor quit respondingerror
1120Flash Boot Failureunknown
1130Timer Stall Failureunknown
1140Timer Range Failureunknown
1150Time Base Failureunknown
1160Internal RAM Failureunknown
1170External RAM Failureunknown
1180Timer Sync Failureunknown
1200UNINIT Interrupt Failureunknown
120.1010CPU failure during POSTCPU failure detected during power-on self-testcritical
120.1020ROM failure during POSTROM failure detected during power-on self-testcritical
120.1160RAM failure during POSTInternal RAM failure detected during power-on self-testcritical
120.431024V supply voltage too lowOff-Line Switcher defective; low voltage on 24V supplycritical
120.432024V supply voltage too highOff-Line Switcher defective; high voltage on 24V supplycritical
120.4330Nonswitched 5V supply voltage too lowLow voltage on nonswitched 5V supply linecritical
120.4340Nonswitched 5V supply voltage too highHigh voltage on nonswitched 5V supply linecritical
120.43505V supply voltage too lowLow voltage on 5V supply linecritical
120.43605V supply voltage too highHigh voltage on 5V supply linecritical
120.43708V supply voltage too lowLow voltage on 8V supply linecritical
120.43808V supply voltage too highHigh voltage on 8V supply linecritical
120.43903.3V supply voltage too lowLow voltage on 3.3V supply linecritical
120.44003.3V supply voltage too highHigh voltage on 3.3V supply linecritical
120.4410Backup alarm capacitor voltage too lowLow backup voltage at capacitor used to power backup alarmerror
120.4420Backup alarm capacitor voltage too highHigh backup voltage at capacitor used to power backup alarmerror
120.4430Battery backup capacitor voltage too lowLow battery voltage at capacitor used to power backup alarmerror
120.4440Battery backup capacitor voltage too highHigh battery voltage at capacitor used to power backup alarmerror
120.4450Watchdog 3.3V supply voltage too lowLow watchdog 3.3 supply voltagecritical
120.4460Watchdog 3.3V supply voltage too highHigh watchdog 3.3 supply voltagecritical
120.4470Processor charging parameter update failedError updating processor charging parametererror
120.4490Processor could not set charge currentProcessor could not set charge currenterror
120.4500Processor charge current too highProcessor charge current too high for present charge level settingerror
120.4510Battery missing or disconnected during useBattery missing or disconnected during operationcritical
120.4520Battery precharge current response abnormalBattery did not respond appropriately to precharge currenterror
120.4530Battery voltage exceeded safe limitsBattery voltage exceeded safe limitscritical
120.4540Charger chip watchdog failed to shut downCharger chip could not be turned off by its watchdogerror
120.4550ADC low positive reference voltage self-test failedADC failed. Incorrect low positive reference voltageerror
120.4560ADC high positive reference voltage self-test failedADC failed. Incorrect high positive reference voltageerror
120.4570ADC low negative reference voltage self-test failedADC failed. Incorrect low negative reference voltageerror
120.4580ADC high negative reference voltage self-test failedADC failed. Incorrect high negative reference voltageerror
120.4590ADC low average reference voltage self-test failedADC failed. Incorrect low average reference voltageerror
120.4600ADC high average reference voltage self-test failedADC failed. Incorrect high average reference voltageerror
120.4610Processor charge current too lowProcessor charge current too low for present charge levelerror
120.4620Battery voltage too low when chargingBattery voltage too low when chargingerror
120.4630Battery voltage too highBattery voltage too highcritical
120.50108051 software version incompatible8051 software version not compatible with PC Unit softwareerror
120.5020Processor not reporting software versionProcessor not reporting software version data after System Onerror
120.6000Multiple failures detected during POSTOne or more failures detected during power-on self-testcritical
120.6010RAM failure during POSTRAM failure detected during power-on self-testcritical
120.6170Battery charging disabled; high compartment temperatureBattery not being charged and compartment temperature too highunknown
120.6180Battery charging disabled; high temperatureBattery not being charged and temperature too highunknown
1210Option Board Multiple IDunknown
121.2000Power Supply Processor not respondingPower Supply Processor quit respondingerror
1220Processor Exceptionunknown
1230Processor Resetunknown
1240Processor Power On Resetunknown
1250Option Board Unknown Typeunknown
133.6080Backup speaker failedBackup speaker component failureerror
133.6090Main speaker failedMain speaker component failureerror
133.6100Backup speaker power supply failedBackup speaker power supply component failureerror
2000Comm Failureunknown
200.1010Error detected during CPU testerror
200.103064K to flash file DRAM test failederror
200.1040Unable to write into flash memory chip used for loggingerror
200.1060Watchdog failureerror
200.1070Hardware watchdog test failureerror
200.1110Flash file device containing application failed CRC testerror
200.1120Boot flash file device containing boot code failed CRC testerror
200.1200Uninitialized interrupt occurred - default handler callederror
200.4160Motor shutoff safety switch test failederror
200.5010Boot block software version not compatible with Pump Moduleerror
200.50208051 processor not reporting software version after System Onerror
200.5030Serial number reported by Pump Module not a recognized typeerror
200.5040Pump Module software version not compatible with flashed PC Uniterror
200.6120Critical value data integrity failureerror
2010Comm Watchdog Failureunknown
2020Comm Incorrect Response Failureunknown
2030Comm Corrupted Response Failureunknown
2040Comm Unexpected Response Failureunknown
2050Comm Bad Commandunknown
2060Comm Heartbeat Failureunknown
2070Comm No Response Failureunknown
2080Comm Poll Sync Failureunknown
2090Comm Too Many Units Failureunknown
2100Comm Transmit Failureunknown
2100Lost interrupt check response (log only)Interrupt check response lost during operationinfo
210.1010CPU failure detected during POSTerror
210.1020ROM failure detected during POSTerror
210.1160Internal RAM failure detected during POSTerror
210.2040Response message received when not expectederror
210.4040Clear Display BSYNC failure detected during POSTerror
210.50108051 software version not compatible with Pump Moduleerror
210.5020Processor not reporting software version after System Onerror
210.6000One or more failures detected during POSTerror
210.6010RAM failure detected during POSTerror
210.6020Keypad failure detected during POSTerror
210.6030Alarm LED failure detected during POSTerror
210.6040Display failure detected during POSTerror
211.2000Unspecified communication errorerror
2200Serial device error (log only)2892 and 16x50 serial device errorinfo
2200Comm Break Interruptunknown
220.1000Safety Processor internal errorerror
220.1010CPU failure detected during POSTerror
220.1020ROM failure detected during POSTerror
220.1160Internal RAM failure detected during POSTerror
220.50108051 software version not compatible with Pump Moduleerror
220.5020Processor not reporting software version after System Onerror
220.6000One or more failures detected during POSTerror
220.6010RAM failure detected during POSTerror
2210Comm Timer Interruptunknown
221.2000Unspecified communication errorerror
221.2010Timely response to watchdog message not receivederror
221.2020Incorrect response message receivederror
221.2030Corrupted response message receivederror
221.2040Response message received when not expectederror
221.2050Received command that cannot be interpretederror
221.2060Not receiving heartbeat messages in timely mannererror
222.1000Measured illegal or inaccurate action by main processorerror
222.2030Corrupted response message receivederror
222.2040Response message received when not expectederror
222.4010High infusion rate detectederror
222.4020Low infusion rate detectederror
222.4030Motor stall detectederror
222.4040Detected encoder out of sync with motor pulseserror
222.4050Excessive air-in-line detectedunknown
222.4060Over-infusion state detected (infusion past VTBI limit)error
222.4160Motor shutoff safety switch test failederror
222.4170Infusion parameter out-of-range detectederror
222.4180Improper door and/or flow stop state during infusionunknown
222.4190Received command to open switch when already open, or close when closederror
222.4200Motor activity detected when motor should be idleerror
2230Comm Modem Interruptunknown
2300Model 800 cleanup task interrupt lost (log only)Model 800 cleanup task lost interrupt acknowledgeinfo
2300Comm Lost End of Interruptunknown
240.4140Bottle-side pressure sensor voltage too lowerror
240.4150Bottle-side pressure sensor voltage too higherror
241.4140Patient-side pressure sensor voltage too lowerror
241.4150Patient-side pressure sensor voltage too higherror
242.4000Encoder feature not recognized; unable to determine rotation speederror
242.4010Fluid delivery mechanism turning too quickly (high rate)Motor speed too high during infusionerror
242.4020Fluid delivery mechanism turning too quickly (low rate)Motor speed too high during low-rate infusionerror
242.4030Motor stall detectedMotor unable to rotate or stopped unexpectedlyerror
243.4070Air-in-Line assembly self-test failedAIL component failure during POSTerror
244.3020Safety clamp reads safe when unpoweredSafety clamp read CLOSED when should read OPEN when unpoweredcritical
245.3020Door sensor reads safe when unpoweredDoor sensor read CLOSED when should read OPEN when unpoweredcritical
260.40803.3V AD voltage too low3.3V AD voltage at least 8.4% lower than specificationcritical
260.40903.3V AD voltage too high3.3V AD voltage at least 8.4% higher than specificationcritical
260.41005V AD voltage too low5V AD voltage at least 8.4% lower than specificationcritical
260.41105V AD voltage too high5V AD voltage at least 8.4% higher than specificationcritical
260.4120Pressure Sensor voltage too lowPressure Sensor voltage at least 8.4% lower than specificationerror
260.4130Pressure Sensor voltage too highPressure Sensor voltage at least 8.4% higher than specificationerror
2XX.2080Module poll not in odd/even sequenceCommunication protocol violation in module pollingerror
2XX.2090Seventh or greater module attachedMore than 6 modules detected on systemwarning
2XX.3000Keypad key stuck closedKey held down continuously during self-testerror
2XX.4700ADC calibration timeoutADC failed to calibrate within specified timeerror
2XX.4710ADC conversion timeoutADC failed to complete conversion within specified timeerror
2XX.5080Module version data not validatedModule version not validated by PC Unit within specified timewarning
2XX.6400General CBIT failureGeneral built-in self-test failure not tied to specific malfunctionerror
3000Stuck Closed Failureunknown
3010Stuck Open Failureunknown
3010Nurse Call switch open (log only)Nurse Call switch open when expected to be closedinfo
3020Bad Unpowered State Failureunknown
3030Power Supply Failureunknown
350.6600Watchdog circuit POST failureWatchdog circuit failed power-on self-testcritical
350.6630Pressure not tracked by force sensorForce sensor output does not track pressure changeserror
350.6680Clock frequency out of syncTimer base clock frequency is out of synccritical
351.6610Unexpected motor watchdog countdown interruptMotor watchdog timer triggered unexpectedlyerror
351.6660Linear Sensor position mismatchPredicted change in Linear Sensor inconsistent with expected positionerror
351.6720Motor steps accumulating incorrectlyMotor steps too fast or too slow compared to expected rateerror
351.6730Excessive motor steps in sampleToo many motor steps detected in motor step sampleerror
351.6740Engagement nut not engaged during infusionNut not engaged when lever opened during operationcritical
351.6750Nut engagement attempt timeoutAttempt to engage nut timed out before full engagementerror
351.6760Module calibration flag not setModule not calibrated before operationwarning
351.6770Pressure sensing disc circuit test failedPressure sensing disc circuit test failed during POSTerror
352.6640Force sensor output out of rangeForce sensor output is zero or outside acceptable rangeerror
352.6700Force sensor circuit check failureForce sensor circuit check failed during self-testerror
353.6650Linear Sensor output out of rangeLinear Sensor output is zero or outside acceptable rangeerror
354.6670Pressure sensor output out of rangePressure sensor output is outside acceptable rangeerror
354.6690Pressure Sensor current out of rangePressure Sensor current outside acceptable rangeerror
356.6710Digital sensor out of range during no-power testDigital sensor outside acceptable range during no-power testerror
356.6780PCA Handset key stuck at power-upSystem detected PCA Handset key stuck when powering uperror
3XX.2080Module poll not in odd/even sequenceCommunication protocol violation in module pollingerror
3XX.2090Seventh or greater module attachedMore than 6 modules detected on systemwarning
3XX.3000Keypad key stuck closedKey held down continuously during self-testerror
3XX.4700ADC calibration timeoutADC failed to calibrate within specified timeerror
3XX.4710ADC conversion timeoutADC failed to complete conversion within specified timeerror
3XX.5080Module version data not validatedModule version not validated by PC Unit within specified timewarning
3XX.6400General CBIT failureGeneral built-in self-test failure not tied to specific malfunctionerror
3XX.7200Plunger position intermittent sensor contactLinear Sensor readings indicate potentiometer wiper making intermittent contactwarning
3XX.7210Linear Sensor contamination detectedPlunger position sampler detected possible sensor contamination continuouslyunknown
3XX.7220Plunger approaching lower mechanical stopMotor running and plunger about to contact lower mechanical stopunknown
4000General Rate Failureunknown
4010High Rate Failureunknown
4020Low Rate Failureunknown
4030Motor Stall Failureunknown
4040Sync Failureunknown
4050AIL Check Failureunknown
4060Over Infused Failureunknown
4070Sensor Test Failureunknown
4080ADC 3 Volt Low Voltage Failureunknown
4090ADC 3 Volt High Voltage Failureunknown
4100ADC 5 Volt Low Voltage Failureunknown
4110ADC 5 Volt High Voltage Failureunknown
4120Sensor Supply Low Voltage Failureunknown
4130Sensor Supply High Voltage Failureunknown
4140Sensor Broken Low Failureunknown
4150Sensor Broken High Failureunknown
4160Safety Switch Test Failureunknown
4170Parameter Range Failureunknown
4180Door or Safety Clamp State Failureunknown
4190Safety Switch State Failureunknown
4200Illegal Motion Failureunknown
4310Low 24V Failureunknown
4320High 24V Failureunknown
4330Low NSW 5V Failureunknown
4340High NSW 5V Failureunknown
4350Low 5V Failureunknown
4360High 5V Failureunknown
4370Low 8V Failureunknown
4380High 8V Failureunknown
4390Low 3P3V Failureunknown
4400High 3P3V Failureunknown
4410Low Backup Voltage Failureunknown
4420High Backup Voltage Failureunknown
4430Low Battery Voltage Failureunknown
4440High Battery Voltage Failureunknown
4450Low WD3P3V Failureunknown
4460High WD3P3V Failureunknown
4470Power Supply Processor Parameter Failureunknown
4490Power Supply Processor Set Charge Level Failureunknown
4500Power Supply Processor Charge Level High Failureunknown
4510Power Supply Processor Missing Battery Failureunknown
4520Power Supply Processor Precharge Failureunknown
4530Power Supply Processor Charge Voltage High Failureunknown
4540Power Supply Processor Watchdog Failureunknown
4550Low VREF Plus Failureunknown
4560High VREF Plus Failureunknown
4570Low VREF Minus Failureunknown
4580High VREF Minus Failureunknown
4590Low VREF Average Failureunknown
4600High VREF Average Failureunknown
4610Power Supply Processor Charge Level Low Failureunknown
4620Power Supply Processor Charge Voltage Low Failureunknown
4630Power Supply Processor Charger FET Errorunknown
4700ADC Calibration Timeout Failureunknown
4710ADC Conversion Timeout Failureunknown
4XX.2080Module poll not in odd/even sequenceCommunication protocol violation in SpO2 module pollingerror
4XX.2090Seventh or greater module attachedMore than 6 modules detected on systemwarning
4XX.3000Keypad key stuck closedSpO2 module keypad key held down continuouslyerror
4XX.5080Module version data not validatedSpO2 module version not validated within specified timewarning
4XX.6200SpO2 continuous built-in tests failedSpO2 Board failed continuous built-in testserror
4XX.6210SpO2 communications errorSpO2 Board communication error detectederror
4XX.6220SpO2 configuration errorSpO2 Board configuration errorerror
4XX.6230SpO2 received value out of rangeSpO2 Board received value out of acceptable rangeunknown
4XX.6240SpO2 missing sensor statusSpO2 Board missing sensor status signalunknown
4XX.63007-segment display shorted segmentShorted segment in 7-segment displayerror
4XX.63107-segment display open segmentOpen segment in 7-segment displayerror
4XX.6320Alarm LED shortedShorted alarm LED in display circuiterror
4XX.6330Alarm LED openOpen alarm LED in display circuiterror
4XX.6400General CBIT failureGeneral built-in self-test failure in SpO2 moduleerror
4XX.6500Serious SpO2 module errorSerious module error in SpO2 Boardcritical
4XX.6510SpO2 error requires power cycleModule error requires power to be cyclederror
4XX.6520SpO2 error requires module resetModule error requires reseterror
4XX.6530SpO2 transient errorTransient error in SpO2 modulewarning
4XX.6560SpO2 power-on test errorPower-on test error in SpO2 Boarderror
4XX.6580SpO2 unknown errorUnknown error in SpO2 Boarderror
5000Hardware-Software Compatibility Failureunknown
5010Software Version Compatibility Failureunknown
5020Peripheral Version Response Failureunknown
5030Unknown Device Typeunknown
5040Module Software Version Compatibility Failureunknown
5050Duplicate Serial Number Failureunknown
5060Unsupported Module Type Failureunknown
5070Invalid Software Access Key Failureunknown
5080Module Software Version Timeout Failureunknown
5XX.2080Module poll not in odd/even sequenceCommunication protocol violation in EtCO2 module pollingerror
5XX.2090Seventh or greater module attachedMore than 6 modules detected on systemwarning
5XX.3000Keypad key stuck closedEtCO2 module keypad key held down continuouslyerror
5XX.5080Module version data not validatedEtCO2 module version not validated within specified timewarning
5XX.6200EtCO2 continuous built-in tests failedEtCO2 Board failed continuous built-in testserror
5XX.6210EtCO2 communications errorEtCO2 Board communication error detectederror
5XX.6220EtCO2 configuration errorEtCO2 Board configuration errorerror
5XX.6230EtCO2 received value out of rangeEtCO2 Board received value out of acceptable rangeunknown
5XX.6240EtCO2 missing sensor statusEtCO2 Board missing sensor status signalunknown
5XX.6400General CBIT failureGeneral built-in self-test failure in EtCO2 moduleerror
5XX.6500EtCO2 Board action timeoutEtCO2 Board took too long to perform an actionerror
6000Post General Failureunknown
600.6800CI Board POST test failedCI Board failed power-on self-testerror
600.6820CI Board initialization failureCI Board initialization failed during startuperror
600.6830CI Board configuration failureCI Board configuration errorerror
600.6840CI Board server connection failureCI Board unable to connect to servererror
600.6850CI Board not respondingCI Board communication failureerror
600.6860CI Board server disconnect failureCI Board failed to disconnect from server properlyerror
600.6870CI Board communications errorCI Board communication error with moduleserror
6010Post RAM Failureunknown
6020Keypad Failureunknown
6030Alarm LED Failureunknown
6040Display Failureunknown
6050Encoder Failureunknown
6060ADC Failureunknown
6070PC Unit RTC battery failure (log only)Real-time clock battery failurewarning
6080Backup Speaker Failureunknown
6090Main Speaker Failureunknown
6100Backup Speaker Power Supply Failureunknown
6110Compliance Timeframe Failureunknown
6120Data Integrity Failureunknown
6130System Configuration Data Integrity Failureunknown
6140Unexpected interrupt or exception (log only)Unexpected interrupt or exception detectederror
6150Operating system error (log only)Operating system error detectederror
6160Unexpected Hardware Interrupt Failureunknown
6170Charger Battery Temp Too High Failureunknown
6180Charger Battery Box Temp Too High Failureunknown
6190Hospital Profile Data Integrity Failureunknown
6200CBIT Failureunknown
6210Comm Down Failureunknown
6220Config Settings Failureunknown
6230Received Value Out of Rangeunknown
6240Missing Sensor Statusunknown
6300Shorted Segment in 7-Segment Displayunknown
6310Open Segment in 7-Segment Displayunknown
6320Shorted Alarm LEDunknown
6330Open Alarm LEDunknown
63407-Segment Failureunknown
63407-segment display status report (log only)Display status change loggedinfo
63507-segment display status report (log only)Display status change loggedinfo
6400CBIT General Failureunknown
6500Serious Module Errorunknown
6510Module Error Requires Power Cycleunknown
6520Module Error Requires Resetunknown
6530Transient Errorunknown
6540SpO2 informational reports (log only)SpO2 module status loggedinfo
6550Digital Errorunknown
6560Module-Specific Errorunknown
6570SpO2 warmer error (log only)SpO2 warmer component failureunknown
6580Unknown Errorunknown
6600Syringe Module Motor Watchdog Test Failureunknown
6610Syringe Module Watchdog Failureunknown
6630Syringe Plunger Force Not Tracking Pressureunknown
6640Syringe Plunger Force Sensor Failureunknown
6650Syringe Plunger Position Sensor Failureunknown
6660Syringe Plunger Position Failureunknown
6670Syringe Pressure Disc Sensor Failureunknown
6680Syringe Timebases Out of Syncunknown
6690Syringe Pressure Sensor Current Monitor Failureunknown
6700Syringe Force Sensor Circuit Test Failureunknown
6710Syringe Digital Sensors Illegal Stateunknown
6720Syringe Module Motor Step Rate Failureunknown
6730Syringe Module Motor Step Sample Too Many Stepsunknown
6740Syringe Nut Disengagedunknown
6750Syringe Nut Engage Failureunknown
6760Syringe Module Not Calibratedunknown
6770Syringe Module Pressure Sensor Circuit Test Failureunknown
6780PCA Handset Stuckunknown
6800Wireless Network POST Failureunknown
6810Wireless Network Not Configuredunknown
6820Wireless Network Initialization Failureunknown
6830Wireless Network Configuration Failureunknown
6840Wireless Network Connect Failureunknown
6850Wireless Network Not Respondingunknown
6860Wireless Network Disconnect Failureunknown
6870Wireless Network Communication Errorunknown
6880Wireless Network Hardware Failureunknown
6900Wireless Network Data Link Went Downunknown
6910Wireless Network Communication Timed Outunknown
6920Wireless Network Rebootedunknown
7000Auto-ID Module Message Corruptedunknown
7010Auto-ID Comm Errorunknown
7020Auto-ID Interrupt Errorunknown
7030Auto-ID General Errorunknown
7040Auto-ID Command Buffer Errorunknown
7050Auto-ID Internal Command Errorunknown
7060Auto-ID External Command Errorunknown
7070Auto-ID Internal Not Respondingunknown
7080Auto-ID External Not Respondingunknown
7090Auto-ID POST General Errorunknown
7100Auto-ID Keypad Failureunknown
7110Auto-ID Software Version Compatibility Failureunknown
7120Auto-ID Internal Version Compatibility Failureunknown
7130Auto-ID External Version Compatibility Failureunknown
7200Syringe Plunger Position Sensor Contaminatedunknown
7210Syringe Plunger Position Reset Calibrationunknown
7220Syringe Plunger Position Sensor Illegal Motionunknown
7500System time correction failure (log only)System time cannot be corrected by skew adjustmenterror
7500Invariant Time Failureunknown
8000Model 8015 General Operating System Failureunknown
800.8000Software fault detectedSoftware fault in operating systemerror
810.9000POLO software fault or OS errorSoftware fault or OS error while running POLOerror
810.9010POLO boot failed; missing application ELF filePOLO not able to boot - nonexistent application ELF fileerror
810.9020POLO boot failed; invalid application ELF filePOLO not able to boot - invalid application ELF fileerror
810.9030POLO boot failed; application ELF checksum errorPOLO not able to boot - application ELF file has invalid checksumerror
810.9050POLO parameter file update failedError occurred while updating POLO parameter fileerror
810.9060POLO configuration mismatchPOLO running on Logic Board not configured forerror
810.9070Internal compact flash card not detectedPOLO did not detect presence of internal compact flash card in slot Berror
810.9080Internal compact flash card reset/access failedPOLO could not reset and access internal compact flash card in slot Berror
810.9090Compact flash card slot B not a disk typeCompact flash card in slot B not a disk type carderror
810.9100Compact flash card not supported by POLOCompact flash card in slot B not supported by POLOerror
810.9105Wireless card not supported by POLOWireless card not supported by POLOerror
810.9110ATA driver initialization failed for flash cardPOLO could not initialize ATA driver for disk type carderror
810.9120Flash card volume mount failedPOLO could not mount volume on disk type carderror
810.9130Configuration tuple read error from flash card slot BConfiguration tuple read from compact flash card slot B incorrecterror
810.9135Configuration tuple read error from wireless cardConfiguration tuple read from wireless card slot incorrecterror
810.9140Function ID tuple error from flash card slot BFunction ID tuple read from compact flash card slot B incorrecterror
810.9145Function ID tuple error from wireless cardFunction ID tuple read from wireless card incorrecterror
810.9150POLO directory creation failedPOLO not able to create POLO directory in internal FFX file systemerror
810.9160Watchdog timer test failed during POLO bootWatchdog timer test failed while POLO was booting applicationcritical
810.9170POLO unable to read attribute tuples from compact flash carderror
810.9175POLO unable to read attribute tuples from wireless carderror
810.9340POLO parameter file open failed during updatePOLO parameter file could not be opened while processing updateerror
9000POLO General Failureunknown
9010POLO ELF File Not Foundunknown
9020POLO ELF File Invalidunknown
9030POLO ELF File CRC Errorunknown
9040POLO Parameter File Not Foundunknown
9050POLO Parameter File Update Errorunknown
9060POLO Incompatible Boardunknown
9070POLO Compact Flash Card Missingunknown
9080POLO Compact Flash Card Not Readyunknown
9090POLO Compact Flash Card Invalidunknown
9100POLO Compact Flash Card Not Supportedunknown
9110POLO Compact Flash Card Initialization Failureunknown
9120POLO Compact Flash Card Mount Failureunknown
9130POLO Compact Flash Card Configuration Errorunknown
9140POLO Compact Flash Card Functional ID Errorunknown
9150POLO Make Directory Failureunknown
9160POLO Watchdog Test Failureunknown
9170POLO Read Tuple Failureunknown

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI