Alaris Medley Syringe Module 8110: troubleshooting and error codes

The ServiceForge knowledge base holds 19 documented symptoms for the Alaris Medley Syringe Module 8110, with 38 ranked causes, 10 diagnostic checks with test points and expected values, and 19 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Check Syringe patient safety

Syringe insertion or engagement issue detected; drivearm obstruction or sensor engagement problem

Ranked causes

  1. 1. Syringe and/or drivearm obstruction (common)
  2. 2. Plunger head not engaged (common)
  3. 3. Barrel Clamp not closed (common)
  4. 4. Flange detect sensor not engaged (occasional)
  5. 5. Binary switches not working properly or sensors defective (occasional)
  6. 6. Force sensor harness disconnected and/or defective (occasional)
  7. 7. Syringe Size Sensor Assembly harness disconnected and/or defective (occasional)
  8. 8. Linear Sensor harness disconnected and/or defective (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive Not Engaged patient safety

Plunger drive is not engaging with syringe; force sensor or carriage assembly defective or cam lock issue

Ranked causes

  1. 1. Force Sensor Assembly defective (occasional)
  2. 2. Plunger head rotates due to dislodged or broken dowel pin and/or tab in Carriage Assembly (occasional)
  3. 3. Cam Lock stripped or not engaged (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion patient safety

Back pressure detected during infusion; tubing obstructed, pressure sensor defective, or excessive downward force applied

Ranked causes

  1. 1. Increased back pressure sensed while infusing (common)
  2. 2. Pressure Sensor Board and/or Pressure Sensor Harness defective (occasional)
  3. 3. Excessive downward pressure applied on drive head (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe Not Recognized patient safety

Syringe not recognized; wrong syringe brand/size, syringe size sensor defective, or logic board software problem

Ranked causes

  1. 1. Odd size or wrong brand syringe installed (common)
  2. 2. Syringe Size Sensor defective (occasional)
  3. 3. Logic Board software problem (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel Disconnected

Module disconnected from PCU during operation or communication problem with Inter Unit Interface detected

Ranked causes

  1. 1. Channel disconnected during operation (common)
  2. 2. Channel communication problem with Inter Unit Interface (IUI) (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure Disc Installed patient safety

Pressure sensing disc installed during an infusion or sensor not sensing disc properly

Ranked causes

  1. 1. Pressure sensing disc installed during an infusion (occasional)
  2. 2. Sensor not sensing disc (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure Disc Removed patient safety

Pressure sensing disc removed during an infusion or sensor sensing disc when it should not be present

Ranked causes

  1. 1. Pressure sensing disc removed during an infusion (occasional)
  2. 2. Sensor is sensing disc (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe Calibration

System detected Force Sensor Board, pressure disc, or Barrel Clamp requires calibration or part fails calibration

Ranked causes

  1. 1. System detected Force Sensor Board, pressure disc, or Barrel Clamp requires calibration (common)
  2. 2. Force Sensor Board, pressure disc, Syringe Size Sensor Assembly, or Logic Board fails calibration (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe Drive-Head Error patient safety

Syringe plunger travel impeded or Cam Lock stripped/not engaged during infusion

Ranked causes

  1. 1. Syringe plunger travel impeded (occasional)
  2. 2. Cam Lock stripped or not engaged (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe Empty patient safety

System reports syringe empty when it is not, or plunger travel is impeded

Ranked causes

  1. 1. Force Sensor Board Assembly defective (occasional)
  2. 2. Syringe plunger travel impeded (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Syringe Module 8110 patient safety

Regular and preventive maintenance calibration using BD Alaris System Maintenance Software to ensure proper device operation.

Ranked causes

  1. 1. Routine calibration requirement (common)
First check

Perform regular inspection before each use

Test point
Syringe Module 8110
Expected
Device in good operating condition

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel Errors

Module error detected indicating a problem with channel operation

Ranked causes

  1. 1. Module error detected (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device dropped or severely jarred patient safety

Physical impact from dropping or jarring the instrument; potential internal damage from trauma

Ranked causes

  1. 1. Physical trauma or impact damage (common)
First check

Visual inspection for physical damage

Test point
Exterior housing, connectors, and display
Expected
No cracks, dents, bent connectors, or visible damage

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Instrument continues to alarm during operation patient safety

Unit alarms and displays error message after power-up self-test passed, indicating detection of internal malfunction during runtime

Ranked causes

  1. 1. Internal malfunction detected during continuous self-test (unknown)
First check

Observe displayed error message

Test point
Display screen
Expected
Error message indicating type of internal malfunction

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

PCA Handset Stuck

PCA Module handset is stuck or incorrectly connected

Ranked causes

  1. 1. Handset is stuck or incorrectly connected (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit continues to alarm during operation patient safety

Instrument repeatedly triggers alarms and displays messages indicating internal malfunction during normal operation after startup

Ranked causes

  1. 1. Internal malfunction detected by continuous self-test (unknown)
First check

Monitor for alarms and error messages during operation

Test point
Display screen and audio alert
Expected
No alarms or error messages; normal operation without self-test failures

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not beep or display LEDs do not flash on startup patient safety

Instrument powers on but fails to produce beep sound or show LED segment flash during startup sequence, indicating self-test failure

Ranked causes

  1. 1. Internal malfunction detected by self-test (common)
First check

Verify instrument beeps on power-up

Test point
Audio output
Expected
Audible beep

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not beep or LEDs do not flash on power-up patient safety

When powering up the instrument, expected beep and LED flash do not occur, indicating self-test failure or malfunction

Ranked causes

  1. 1. Internal malfunction detected during self-test (unknown)
First check

Verify beep and LED flash sequence on power-up

Test point
Display LEDs and audio output
Expected
Beep heard and all display LED segments flash

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit has been dropped or severely jarred patient safety

Instrument has sustained physical impact from being dropped or violently shaken

Ranked causes

  1. 1. Potential internal damage from physical impact (unknown)
First check

Visual inspection for physical damage

Test point
Entire device housing and connectors
Expected
No cracks, dents, or loose components

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
350.6600Watchdog circuit failed power-on self testPower-on self testcritical
350.6630Pressure not tracked by force sensorForce sensor monitoring during operationerror
350.6680Timer bases; clock frequency out of syncPower-on self test or during operationcritical
351.6610Unexpected countdown interruptMotor watchdog monitoringerror
351.6660Movement error; predicted change in linear position sensor inconsistent with expected positionDuring plunger movementerror
351.6720Motor steps accumulating too fast or too slowMotor step monitoringerror
351.6730Too many steps in motor step sampleMotor step monitoringerror
351.6740Nut not engaged during infusion; lever not openedDuring infusion startuperror
351.6750Attempt to engage nut timed out before engagingDuring drive engagementerror
351.6760Calibration flag not set. Module not calibrated.Power-on or start of infusionwarning
351.6770Pressure sensing disc circuit test failed during power-on self testPower-on self test (Syringe Module only)error
352.6640Force sensor output is zero or is outside acceptable rangeForce sensor monitoringerror
352.6700Force sensor circuit check failurePower-on self test or circuit monitoringerror
353.6650Linear position sensor output is zero or is outside acceptable rangeLinear position sensor monitoringerror
354.6670Pressure sensor output is outside acceptable rangePressure sensor monitoring (Syringe Module only)error
354.6690Pressure Sensor current monitor: current is outside acceptable rangePressure sensor current monitoringerror
355.6620Syringe sizer output is outside acceptable rangeSyringe sizer monitoringerror
356.6710Digital sensor is outside acceptable range during no-power testPower-on self testerror
356.6780System detected PCA Handset Key stuck at power-upPower-up of PCA Moduleerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI