Alaris Medley Syringe Module 8110: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Alaris Medley Syringe Module 8110, with 38 ranked causes, 10 diagnostic checks with test points and expected values, and 19 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Check Syringe patient safety
Syringe insertion or engagement issue detected; drivearm obstruction or sensor engagement problem
Ranked causes
- 1. Syringe and/or drivearm obstruction (common)
- 2. Plunger head not engaged (common)
- 3. Barrel Clamp not closed (common)
- 4. Flange detect sensor not engaged (occasional)
- 5. Binary switches not working properly or sensors defective (occasional)
- 6. Force sensor harness disconnected and/or defective (occasional)
- 7. Syringe Size Sensor Assembly harness disconnected and/or defective (occasional)
- 8. Linear Sensor harness disconnected and/or defective (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive Not Engaged patient safety
Plunger drive is not engaging with syringe; force sensor or carriage assembly defective or cam lock issue
Ranked causes
- 1. Force Sensor Assembly defective (occasional)
- 2. Plunger head rotates due to dislodged or broken dowel pin and/or tab in Carriage Assembly (occasional)
- 3. Cam Lock stripped or not engaged (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion patient safety
Back pressure detected during infusion; tubing obstructed, pressure sensor defective, or excessive downward force applied
Ranked causes
- 1. Increased back pressure sensed while infusing (common)
- 2. Pressure Sensor Board and/or Pressure Sensor Harness defective (occasional)
- 3. Excessive downward pressure applied on drive head (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe Not Recognized patient safety
Syringe not recognized; wrong syringe brand/size, syringe size sensor defective, or logic board software problem
Ranked causes
- 1. Odd size or wrong brand syringe installed (common)
- 2. Syringe Size Sensor defective (occasional)
- 3. Logic Board software problem (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel Disconnected
Module disconnected from PCU during operation or communication problem with Inter Unit Interface detected
Ranked causes
- 1. Channel disconnected during operation (common)
- 2. Channel communication problem with Inter Unit Interface (IUI) (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure Disc Installed patient safety
Pressure sensing disc installed during an infusion or sensor not sensing disc properly
Ranked causes
- 1. Pressure sensing disc installed during an infusion (occasional)
- 2. Sensor not sensing disc (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure Disc Removed patient safety
Pressure sensing disc removed during an infusion or sensor sensing disc when it should not be present
Ranked causes
- 1. Pressure sensing disc removed during an infusion (occasional)
- 2. Sensor is sensing disc (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe Calibration
System detected Force Sensor Board, pressure disc, or Barrel Clamp requires calibration or part fails calibration
Ranked causes
- 1. System detected Force Sensor Board, pressure disc, or Barrel Clamp requires calibration (common)
- 2. Force Sensor Board, pressure disc, Syringe Size Sensor Assembly, or Logic Board fails calibration (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe Drive-Head Error patient safety
Syringe plunger travel impeded or Cam Lock stripped/not engaged during infusion
Ranked causes
- 1. Syringe plunger travel impeded (occasional)
- 2. Cam Lock stripped or not engaged (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe Empty patient safety
System reports syringe empty when it is not, or plunger travel is impeded
Ranked causes
- 1. Force Sensor Board Assembly defective (occasional)
- 2. Syringe plunger travel impeded (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Syringe Module 8110 patient safety
Regular and preventive maintenance calibration using BD Alaris System Maintenance Software to ensure proper device operation.
Ranked causes
- 1. Routine calibration requirement (common)
Perform regular inspection before each use
- Test point
- Syringe Module 8110
- Expected
- Device in good operating condition
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel Errors
Module error detected indicating a problem with channel operation
Ranked causes
- 1. Module error detected (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device dropped or severely jarred patient safety
Physical impact from dropping or jarring the instrument; potential internal damage from trauma
Ranked causes
- 1. Physical trauma or impact damage (common)
Visual inspection for physical damage
- Test point
- Exterior housing, connectors, and display
- Expected
- No cracks, dents, bent connectors, or visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instrument continues to alarm during operation patient safety
Unit alarms and displays error message after power-up self-test passed, indicating detection of internal malfunction during runtime
Ranked causes
- 1. Internal malfunction detected during continuous self-test (unknown)
Observe displayed error message
- Test point
- Display screen
- Expected
- Error message indicating type of internal malfunction
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PCA Handset Stuck
PCA Module handset is stuck or incorrectly connected
Ranked causes
- 1. Handset is stuck or incorrectly connected (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit continues to alarm during operation patient safety
Instrument repeatedly triggers alarms and displays messages indicating internal malfunction during normal operation after startup
Ranked causes
- 1. Internal malfunction detected by continuous self-test (unknown)
Monitor for alarms and error messages during operation
- Test point
- Display screen and audio alert
- Expected
- No alarms or error messages; normal operation without self-test failures
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not beep or display LEDs do not flash on startup patient safety
Instrument powers on but fails to produce beep sound or show LED segment flash during startup sequence, indicating self-test failure
Ranked causes
- 1. Internal malfunction detected by self-test (common)
Verify instrument beeps on power-up
- Test point
- Audio output
- Expected
- Audible beep
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not beep or LEDs do not flash on power-up patient safety
When powering up the instrument, expected beep and LED flash do not occur, indicating self-test failure or malfunction
Ranked causes
- 1. Internal malfunction detected during self-test (unknown)
Verify beep and LED flash sequence on power-up
- Test point
- Display LEDs and audio output
- Expected
- Beep heard and all display LED segments flash
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit has been dropped or severely jarred patient safety
Instrument has sustained physical impact from being dropped or violently shaken
Ranked causes
- 1. Potential internal damage from physical impact (unknown)
Visual inspection for physical damage
- Test point
- Entire device housing and connectors
- Expected
- No cracks, dents, or loose components
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
350.6600 | Watchdog circuit failed power-on self test | Power-on self test | critical |
350.6630 | Pressure not tracked by force sensor | Force sensor monitoring during operation | error |
350.6680 | Timer bases; clock frequency out of sync | Power-on self test or during operation | critical |
351.6610 | Unexpected countdown interrupt | Motor watchdog monitoring | error |
351.6660 | Movement error; predicted change in linear position sensor inconsistent with expected position | During plunger movement | error |
351.6720 | Motor steps accumulating too fast or too slow | Motor step monitoring | error |
351.6730 | Too many steps in motor step sample | Motor step monitoring | error |
351.6740 | Nut not engaged during infusion; lever not opened | During infusion startup | error |
351.6750 | Attempt to engage nut timed out before engaging | During drive engagement | error |
351.6760 | Calibration flag not set. Module not calibrated. | Power-on or start of infusion | warning |
351.6770 | Pressure sensing disc circuit test failed during power-on self test | Power-on self test (Syringe Module only) | error |
352.6640 | Force sensor output is zero or is outside acceptable range | Force sensor monitoring | error |
352.6700 | Force sensor circuit check failure | Power-on self test or circuit monitoring | error |
353.6650 | Linear position sensor output is zero or is outside acceptable range | Linear position sensor monitoring | error |
354.6670 | Pressure sensor output is outside acceptable range | Pressure sensor monitoring (Syringe Module only) | error |
354.6690 | Pressure Sensor current monitor: current is outside acceptable range | Pressure sensor current monitoring | error |
355.6620 | Syringe sizer output is outside acceptable range | Syringe sizer monitoring | error |
356.6710 | Digital sensor is outside acceptable range during no-power test | Power-on self test | error |
356.6780 | System detected PCA Handset Key stuck at power-up | Power-up of PCA Module | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.