Alaris Medley SpO2 Module 8220 Masimo: troubleshooting and error codes
The ServiceForge knowledge base holds 7 documented symptoms for the Alaris Medley SpO2 Module 8220 Masimo, with 7 ranked causes, 5 diagnostic checks with test points and expected values, and 13 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Bad Sensor patient safety
Broken, unknown or nonsystem sensor or patient cable attached.
Ranked causes
- 1. Incorrect or incompatible sensor attached (common)
Verify sensor type matches module requirements
- Test point
- Sensor connection
- Expected
- Correct system sensor attached
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad sensor detected patient safety
Alarm indicates broken, unknown or nonsystem sensor or patient cable attached; visual display and symbol shown when audio alarm is silenced
Ranked causes
- 1. Incorrect or incompatible sensor attached (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Sensor - Electrical or Optical Interference
External interference detected on sensor signal.
Ranked causes
- 1. External electromagnetic or optical interference (occasional)
Inspect sensor for damage or contamination
- Test point
- Sensor condition
- Expected
- Clean, intact sensor with proper connection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Sensor - Light
Light interference detected on sensor.
Ranked causes
- 1. Ambient or external light exposure on sensor (common)
Assess lighting conditions at sensor site
- Test point
- Sensor location
- Expected
- Minimal direct or ambient light exposure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check sensor - light interference
Light interference on sensor detected; visual alarm indication displayed
Ranked causes
- 1. Light interference on sensor (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Sensor - Low Perfusion patient safety
Patient's low perfusion has inhibited monitoring.
Ranked causes
- 1. Poor blood perfusion at sensor site (occasional)
Evaluate perfusion at current sensor site
- Test point
- Sensor location
- Expected
- Adequate pulsatile blood flow visible or palpable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Sensor - Low Signal I.Q. patient safety
Low signal quality being measured by the sensor.
Ranked causes
- 1. Incorrect sensor placement or poor perfusion (common)
Verify sensor placement per product instructions
- Test point
- Sensor application site
- Expected
- Sensor positioned correctly on well-perfused patient site
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Bad Sensor | Broken, unknown or nonsystem sensor or patient cable attached | Detection of incompatible or defective sensor or patient cable | warning |
Check Sensor - Electrical or Optical Interference | External interference on sensor | Detection of electrical or optical interference affecting sensor signal | warning |
Check Sensor - Light | Light interference on sensor | Detection of excessive ambient light affecting sensor readings | warning |
Check Sensor - Low Perfusion | Patient's low perfusion has inhibited monitoring | Detection of insufficient blood perfusion at sensor site | warning |
Check Sensor - Low Signal I.Q. | Low signal quality being measured | Low quality signal detected during pulse oximetry measurement | warning |
High Pulse Rate Alarm | High pulse rate alarm limit has been exceeded. | Pulse rate exceeds configured alarm limit | warning |
High SpO2 Alarm | High SpO2 alarm limit has been exceeded. | SpO2 value exceeds configured alarm limit | warning |
Low Pulse Rate Alarm | Low pulse rate alarm limit has been exceeded. | Pulse rate falls below configured alarm limit | warning |
Low SpO2 Alarm | Low SpO2 alarm limit has been exceeded. | SpO2 value falls below configured alarm limit | warning |
No Sensor | Sensor not properly attached to patient cable or patient cable not properly attached to SpO2 Module. | Sensor or cable connection lost or not detected | error |
No Signal | Failure to find a patient signal after 30 seconds of searching. | No detectable pulse signal within 30 seconds of sensor application | error |
Remove Module (Max=1) | More than 1 SpO2 Module attached. | Multiple SpO2 modules connected simultaneously | error |
Sensor Off | Sensor not properly attached to patient. | Sensor connection to patient lost or not detected | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.