FR FieldReady Academy
Pulse oximetry modules · Medley family

Alaris Medley SpO2 Module 8220 Masimo: troubleshooting and error codes

The ServiceForge knowledge base holds 7 documented symptoms for the Alaris Medley SpO2 Module 8220 Masimo, with 7 ranked causes, 5 diagnostic checks with test points and expected values, and 13 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Bad Sensor patient safety

Broken, unknown or nonsystem sensor or patient cable attached.

Ranked causes

  1. 1. Incorrect or incompatible sensor attached (common)
First check

Verify sensor type matches module requirements

Test point
Sensor connection
Expected
Correct system sensor attached

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bad sensor detected patient safety

Alarm indicates broken, unknown or nonsystem sensor or patient cable attached; visual display and symbol shown when audio alarm is silenced

Ranked causes

  1. 1. Incorrect or incompatible sensor attached (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Sensor - Electrical or Optical Interference

External interference detected on sensor signal.

Ranked causes

  1. 1. External electromagnetic or optical interference (occasional)
First check

Inspect sensor for damage or contamination

Test point
Sensor condition
Expected
Clean, intact sensor with proper connection

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Sensor - Light

Light interference detected on sensor.

Ranked causes

  1. 1. Ambient or external light exposure on sensor (common)
First check

Assess lighting conditions at sensor site

Test point
Sensor location
Expected
Minimal direct or ambient light exposure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check sensor - light interference

Light interference on sensor detected; visual alarm indication displayed

Ranked causes

  1. 1. Light interference on sensor (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Sensor - Low Perfusion patient safety

Patient's low perfusion has inhibited monitoring.

Ranked causes

  1. 1. Poor blood perfusion at sensor site (occasional)
First check

Evaluate perfusion at current sensor site

Test point
Sensor location
Expected
Adequate pulsatile blood flow visible or palpable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Sensor - Low Signal I.Q. patient safety

Low signal quality being measured by the sensor.

Ranked causes

  1. 1. Incorrect sensor placement or poor perfusion (common)
First check

Verify sensor placement per product instructions

Test point
Sensor application site
Expected
Sensor positioned correctly on well-perfused patient site

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Bad SensorBroken, unknown or nonsystem sensor or patient cable attachedDetection of incompatible or defective sensor or patient cablewarning
Check Sensor - Electrical or Optical InterferenceExternal interference on sensorDetection of electrical or optical interference affecting sensor signalwarning
Check Sensor - LightLight interference on sensorDetection of excessive ambient light affecting sensor readingswarning
Check Sensor - Low PerfusionPatient's low perfusion has inhibited monitoringDetection of insufficient blood perfusion at sensor sitewarning
Check Sensor - Low Signal I.Q.Low signal quality being measuredLow quality signal detected during pulse oximetry measurementwarning
High Pulse Rate AlarmHigh pulse rate alarm limit has been exceeded.Pulse rate exceeds configured alarm limitwarning
High SpO2 AlarmHigh SpO2 alarm limit has been exceeded.SpO2 value exceeds configured alarm limitwarning
Low Pulse Rate AlarmLow pulse rate alarm limit has been exceeded.Pulse rate falls below configured alarm limitwarning
Low SpO2 AlarmLow SpO2 alarm limit has been exceeded.SpO2 value falls below configured alarm limitwarning
No SensorSensor not properly attached to patient cable or patient cable not properly attached to SpO2 Module.Sensor or cable connection lost or not detectederror
No SignalFailure to find a patient signal after 30 seconds of searching.No detectable pulse signal within 30 seconds of sensor applicationerror
Remove Module (Max=1)More than 1 SpO2 Module attached.Multiple SpO2 modules connected simultaneouslyerror
Sensor OffSensor not properly attached to patient.Sensor connection to patient lost or not detectederror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.