AirSep NewLife: troubleshooting and error codes
The ServiceForge knowledge base holds 31 documented symptoms for the AirSep NewLife, with 90 ranked causes, 45 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Limited or low flow output
Oxygen output flow at regulator is less than expected or set value; insufficient output.
Ranked causes
- 1. Restriction in humidifier or tubing (common)
- 2. Product regulator set too low (common)
- 3. Leak in system (occasional)
- 4. Weak compressor (occasional)
- 5. Defective solenoid valve (occasional)
- 6. Reduced air intake (suction) (occasional)
- 7. Defective circuit board (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not sound
Audio alarm fails to activate when condition triggers or unit powers on; no buzzer sound.
Ranked causes
- 1. Dead battery (common)
- 2. Incorrectly installed battery (common)
- 3. Faulty electrical connection (occasional)
- 4. Defective ON/OFF switch (occasional)
- 5. Defective buzzer (occasional)
- 6. Defective pressure switch (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flowmeter fluctuates
Flowmeter display shows unstable or rapidly changing flow reading; not steady.
Ranked causes
- 1. Improperly set or faulty product regulator (common)
- 2. Leak in system (occasional)
- 3. Worn compressor (occasional)
- 4. Defective circuit board (occasional)
- 5. Defective flowmeter (occasional)
- 6. Defective solenoid valve (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit temperature is too high caution
Cabinet or compressor temperature rises to unsafe levels during normal operation.
Ranked causes
- 1. Blocked air intake or defective cabinet fan (common)
- 2. Contaminated sieve beds (occasional)
- 3. Defective solenoid valve (occasional)
- 4. Restriction in exhaust muffler (occasional)
- 5. Restriction in suction resonator (occasional)
- 6. Restriction in EcoCheck valve (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor does not run
Unit powered on but compressor motor does not start; constant audio alarm heard with ON/OFF switch in ON position.
Ranked causes
- 1. No power to unit (common)
- 2. Unit circuit breaker tripped or faulty (common)
- 3. Faulty electrical connections (occasional)
- 4. Defective circuit board (occasional)
- 5. Defective ON/OFF power switch (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor relief valve releases with popping sound caution
Loud popping sound heard from relief valve during normal operation; valve venting excess pressure.
Ranked causes
- 1. Defective circuit board (occasional)
- 2. Defective solenoid valve (occasional)
- 3. Contaminated sieve beds (occasional)
- 4. Defective relief valve (occasional)
- 5. Faulty electrical connection at waste valve (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor thermally cut off; excessive heat detected caution
Compressor shuts down due to overheating; will not restart until cooled down; cabinet fan operates.
Ranked causes
- 1. Blocked air intake or defective cabinet fan (common)
- 2. Defective capacitor (occasional)
- 3. Defective circuit board (occasional)
- 4. Faulty electrical connection for compressor (occasional)
- 5. Defective high temperature switch (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Constant alarm with circuit breaker repeatedly trips caution
Alarm sounds continuously with ON/OFF switch in ON position; unit circuit breaker trips repeatedly when reset.
Ranked causes
- 1. Defective circuit breaker (common)
- 2. Defective capacitor (occasional)
- 3. Defective compressor (occasional)
- 4. Defective circuit board (occasional)
- 5. Faulty electrical connection (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low operating pressure
Operating pressure lower than normal specifications.
Ranked causes
- 1. Restriction in suction resonator or optional air intake filter (common)
- 2. Improperly operating circuit board or solenoid valve (occasional)
- 3. Leak in the unit allowing system pressure to escape (occasional)
- 4. Compressor with reduced output (occasional)
- 5. Obstructed or vertically positioned EcoCheck valve (rare)
Verify all filters are clean
- Test point
- Air intake filter, sieve bed inlet
- Expected
- Clean, unobstructed
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen concentration output
Oxygen purity at outlet is below specification; may indicate degraded oxygen separation.
Ranked causes
- 1. Occluded internal felt filter (common)
- 2. Leak in system (occasional)
- 3. Weak compressor (occasional)
- 4. Contaminated sieve beds (occasional)
- 5. Defective solenoid valve (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio alarm sounds continuously
Battery-powered alarm system activates with continuous loud alarm sound
Ranked causes
- 1. Power failure (common)
- 2. High pressure indicator activated (occasional)
- 3. Low pressure indicator activated (occasional)
- 4. Oxygen concentration below therapeutic level (optional monitor) (occasional)
Verify power connection
- Test point
- power outlet
- Expected
- Unit plugged in and powered on
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor shuts down intermittently
Compressor starts normally but then stops unexpectedly during operation; cycle repeats.
Ranked causes
- 1. Restricted air flow through unit (common)
- 2. Unit overheating due to improper location (occasional)
- 3. Defective cabinet fan (occasional)
- 4. Defective high temperature switch (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio alarm sounds intermittently patient safety
Alarm system activates with intermittent alarm sound pattern
Ranked causes
- 1. High pressure indicator activated (occasional)
- 2. Low pressure indicator activated (occasional)
- 3. Oxygen concentration below therapeutic level (optional monitor) (occasional)
Check high pressure indicator
- Test point
- high pressure indicator
- Expected
- Indicator not activated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor runs with intermittent high pressure alarm; low oxygen concentration output
Compressor runs but high pressure alarm sounds intermittently; oxygen concentration below specification.
Ranked causes
- 1. Defective sieve beds (common)
- 2. Faulty solenoid valve (occasional)
- 3. Restriction in exhaust muffler (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet fan does not turn
Cabinet cooling fan is not rotating; no audible fan noise during operation.
Ranked causes
- 1. Defective cabinet fan (common)
- 2. Defective electrical connections (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor runs with intermittent low pressure alarm; low oxygen concentration output
Compressor operates but low pressure alarm sounds intermittently; oxygen concentration output below specification.
Ranked causes
- 1. Leak in system (common)
- 2. Defective circuit board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Solenoid valve circuit malfunction indicated by LED caution
One of five green indicator lights on circuit board matrix fails to illuminate during operation; indicates disconnected or faulty solenoid coil or electrical malfunction in that valve circuit
Ranked causes
- 1. Faulty solenoid coil (common)
- 2. Electrical malfunction in valve circuit (occasional)
Observe LED matrix during normal operation
- Test point
- Five green indicator lights on circuit board
- Expected
- Two green lights always illuminated and rotating/cycling
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Solenoid valve noisy
Audible noise from one of the five solenoid valves during operation.
Ranked causes
- 1. Solenoid valve internal parts worn or damaged (common)
- 2. Solenoid valve coil non-functional (occasional)
Observe which green circuit board light illuminates when noise occurs
- Test point
- Green circuit board lights
- Expected
- One light illuminates corresponding to noisy valve location
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery alarm test absent on startup caution
Five-second audible alarm does not sound when unit is turned on, indicating potential battery problem
Ranked causes
- 1. Battery in poor condition or dead (common)
Turn unit ON and listen for five-second alarm
- Test point
- audio output
- Expected
- Loud audible alarm for approximately five seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Back Pressure Correction at 5 LPM
Adjusting oxygen flow to 5 lpm when long tubing and humidifier bottle cause back pressure reduction
Ranked causes
- 1. Back pressure from extended tubing reduces oxygen flow below 5 lpm (occasional)
Verify unit is on, pressurized with full patient setup
- Test point
- ON/OFF switch with humidifier, tubing and cannula connected
- Expected
- ON position, 3+ minutes runtime
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Product Regulator Cleaning or Rebuilding
Cleaning or rebuilding regulator diaphragm, seat and spring when flowmeter ball fluctuates or lockout fails
Ranked causes
- 1. Flowmeter ball fluctuates more than 1/4 lpm or regulator cannot be adjusted for lockout (occasional)
Verify unit is powered down
- Test point
- ON/OFF switch and power cord
- Expected
- OFF position, unplugged
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Product Regulator Lockout patient safety
Limiting maximum oxygen flow output using the regulator lockout feature for patient safety
Ranked causes
- 1. Maximum flow limit needs to be set for patient requirements (occasional)
Verify unit is on and pressurized with patient connection
- Test point
- ON/OFF switch with humidifier and cannula connected
- Expected
- ON position, 3+ minutes runtime, accessories connected
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Product Regulator Normal Operation
Setting the product regulator to ensure flowmeter ball centers on 5.5 lpm line for proper oxygen output
Ranked causes
- 1. Regulator requires adjustment during setup or maintenance (common)
Verify unit is on and pressurized
- Test point
- ON/OFF switch
- Expected
- ON position, 3+ minutes runtime
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor cannot start
Unit fails to power on or compressor does not engage when switched on.
Ranked causes
- 1. Capacitor defective (common)
Verify compressor does not start when unit is turned on
- Test point
- Unit operation
- Expected
- Compressor engages and runs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor does not start with cold ambient
Compressor motor will not start when unit is at very cold temperature; ON/OFF switch in ON position, intermittent alarm audible, cabinet fan turns on.
Ranked causes
- 1. Extreme cold ambient temperature (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor runs with intermittent alarm; oxygen concentration normal at 5 LPM
Compressor motor runs but alarm sounds intermittently; output oxygen concentration meets specification at 5 liters per minute.
Ranked causes
- 1. Defective high or low pressure switch (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maximum flow does not register 5.5 lpm
Flowmeter ball does not appear to split the middle of the 5.5 lpm line when flowmeter adjustment knob is turned wide open with no outlet connections
Ranked causes
- 1. Product regulator requires adjustment (common)
Turn flowmeter adjustment knob counterclockwise until it stops
- Test point
- flowmeter adjustment knob
- Expected
- Knob at wide open position (counterclockwise stop)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen Monitor amber light remains on patient safety
Oxygen Monitor's amber light stays on after startup, does not turn off when oxygen concentration reaches therapeutic level
Ranked causes
- 1. Oxygen concentration below 85% ± 3% (common)
Measure oxygen concentration at outlet
- Test point
- oxygen outlet
- Expected
- 85% ± 3%, ±3%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power failure alarm does not activate caution
Unit does not produce audio alarm when unplugged and turned ON, indicating alarm system failure
Ranked causes
- 1. Alarm system malfunction (rare)
Unplug unit from power outlet
- Test point
- power outlet
- Expected
- Power disconnected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure alarm system indication abnormal caution
Red pressure indicator lights (high and low pressure) on circuit board display abnormally
Ranked causes
- 1. Pressure anomaly or sensor malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Solenoid valve coil does not illuminate
Green circuit board light for a solenoid valve does not light up during operation.
Ranked causes
- 1. Solenoid valve coil defective or not energized (common)
Observe if green circuit board light illuminates for the solenoid valve
- Test point
- Green circuit board light
- Expected
- Light illuminates when valve is energized
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI