Agilent E4350B: troubleshooting and error codes
The ServiceForge knowledge base holds 11 documented symptoms for the Agilent E4350B, with 31 ranked causes, 22 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Output held low high voltage
Supply output is low and PROT annunciator is not on
Ranked causes
- 1. Secondary bias or reference voltage out of spec (common)
- 2. Rail voltages incorrect (common)
- 3. Primary circuit failure (power transformer, rectifiers, filter capacitors) (occasional)
- 4. CV/CC OR gate diodes conducting (occasional)
- 5. A1D411 diode conducting (occasional)
- 6. VMon amplifier A1U401 defective (rare)
- 7. CV comparator A1U408 defective (rare)
Measure secondary bias and reference voltages
- Test point
- TP 9 through 12
- Expected
- Per Table 6-3
- Tools
- DMM
5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not turn on
No display output visible; fan may or may not be running
Ranked causes
- 1. AC input fault or line fuse open (common)
- 2. A3 front panel board defective (occasional)
- 3. A3 microprocessor or ROM defective (rare)
- 4. LCD or cable W19 defective (occasional)
Is display on?
1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan does not run caution
Supply is powered but cooling fan does not operate
Ranked causes
- 1. AC bias voltage fault or fan cable disconnected (common)
- 2. Fan speed control circuit defective (occasional)
- 3. Fuses F700, F701 open or capacitor/resistor fault (occasional)
Measure voltage at A1Q770 heatsink (TP 12) vs common (TP 9)
- Expected
- +24V
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output held high high voltage
Supply output is high and cannot be reduced to commanded voltage; output regulation failure
Ranked causes
- 1. Secondary bias or reference voltage incorrect (common)
- 2. CV feedback control path faulty (common)
- 3. Tunnel board defective (occasional)
Turn supply off and disable overvoltage circuit
- Test point
- A1U410 pins 1&2, A1R447
- Expected
- Overvoltage circuit disabled for testing
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage condition triggered at turn-on caution
OV protection activates immediately when power supply is turned on, preventing normal operation
Ranked causes
- 1. Output voltage exceeds OV setting at start-up (common)
- 2. OV control circuit latch transistor defective (occasional)
- 3. Output momentarily exceeds voltage setting (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output voltage or current despite enabling
Output remains disabled or at zero volts; supply does not respond to programming commands
Ranked causes
- 1. GPIB interface fault (common)
- 2. Primary interface or front panel control circuit defective (occasional)
Connect GPIB controller; program supply to full scale voltage and current; verify it accepts commands
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current not within specification caution
Programmed output current does not meet expected value; may read approximately correct but outside tolerance
Ranked causes
- 1. Current readback circuit defective (common)
- 2. Cable from A1 to A5 DSP board disconnected or defective (occasional)
Connect short across output; program full scale voltage and current; measure with external ammeter
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage held at zero or low value
Output voltage cannot be programmed above zero or remains at reduced value
Ranked causes
- 1. Output shorted or current-limited (common)
- 2. Voltage monitor circuit defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage not within specification caution
Programmed output voltage does not meet expected value; may read approximately correct but outside tolerance
Ranked causes
- 1. Voltage readback circuit defective (common)
- 2. Cable from A1 to A5 DSP board disconnected or defective (occasional)
Program full scale voltage and current; check output voltage with DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage protection will not trigger caution
OV protection does not function; supply does not enter OV condition when programmed
Ranked causes
- 1. OV circuit comparator or switch defective (common)
- 2. OV feedback network resistors or capacitor defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current held at zero or low value
Output current cannot be programmed above zero or remains at reduced value despite voltage enabled
Ranked causes
- 1. Current monitor circuit defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E1 | Front panel RAM test failed | Microprocessor detects RAM error during self-test | error |
E10 | Secondary interface 5V readback test failed | Voltage measurement error on 5V supply during self-test | error |
E11 | Ambient temperature readback test failed | Temperature sensor or measurement circuit error during self-test | error |
E12 | CV or CC DAC test failed | Digital-to-analog converter error during self-test | error |
E2 | Front panel ROM test failed | Microprocessor detects ROM error during self-test | error |
E3 | Front panel EEPROM checksum failed | Checksum error detected during self-test | error |
E30 | Secondary interface DAC test failed | Digital-to-analog converter error on A5 board | error |
E31 | Secondary interface DAC test failed | Digital-to-analog converter error on A5 board | error |
E32 | Secondary interface DAC test failed | Digital-to-analog converter error on A5 board | error |
E4 | Primary interface external RAM test failed | RAM error detected on A2 GPIB board during self-test | error |
E45 | Secondary interface I/O test failed | Input/output circuit error on A5 board | error |
E46 | Secondary interface nonvolatile RAM test failed | NVRAM error on A5 board during self-test | error |
E5 | Primary interface internal RAM test failed | RAM error detected on A2 GPIB board | error |
E6 | Primary interface ROM test failed | ROM error detected on A2 GPIB board | error |
E7 | GPIB interface test failed | GPIB communication error during self-test | error |
E8 | Secondary interface RAM test failed | RAM error detected on A5 DSP board | error |
E9 | Secondary interface ROM test failed | ROM error detected on A5 DSP board | error |
EE WRITE ERR | EEPROM write failure on front panel board | EEPROM programming failed | error |
SBUF FULL | Serial buffer full on GPIB board | Serial buffer capacity exceeded | warning |
SBUF OVERRUN | Serial buffer overrun on GPIB board | Serial buffer received more data than capacity | warning |
SERIAL DOWN | Serial data line failure on DSP board | Serial communication failure on secondary interface | error |
SERIAL TIMEOUT | Serial data line failure on GPIB board | Serial communication timeout during self-test | error |
STK OVERFLOW | Stack overflow failure on DSP board | Microprocessor stack exceeded during execution | error |
UART FRAMING | UART framing error on GPIB board | Invalid framing detected in serial communication | error |
UART OVERRUN | UART overrun error on GPIB board | Serial data received faster than processed | error |
UART PARITY | UART parity error on GPIB board | Parity bit mismatch during serial communication | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.