Agilent E3633A: troubleshooting and error codes
The ServiceForge knowledge base holds 64 documented symptoms for the Agilent E3633A, with 91 ranked causes, 103 diagnostic checks with test points and expected values, and 64 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power-on self-test fails
Unit performs automatic self-tests (checks 601-604, 624-632) upon power-on. Test failures prevent normal operation.
Ranked causes
- 1. Front panel controller communication failure (test 601) (unknown)
- 2. RAM read/write error (test 602) (unknown)
- 3. A/D sync pulse not recognized (test 603) (unknown)
- 4. A/D slope convergence out of range (test 604) (unknown)
- 5. Rundown gain calibration failure (test 605) (unknown)
Front panel controller communication
- Test point
- U19 and U1 serial interface
- Expected
- Bidirectional communication established
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias supplies problems caution
Bias supply voltages out of specification or unstable
Ranked causes
- 1. Regulator input voltage insufficient (occasional)
- 2. Circuit overload pulling down bias supplies (occasional)
- 3. Local bias supply inactive (occasional)
- 4. AC oscillations on bias supplies (rare)
+5V Floating bias supply voltage
- Test point
- U10 pin 2
- Expected
- +4.75 V to +5.25 V, +/-0.25 V
- Tools
- Digital multimeter
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
Power supply is completely inoperative, no display or output
Ranked causes
- 1. AC power cord not connected (common)
- 2. Front-panel power switch not depressed (common)
- 3. Power-line fuse missing or incorrect type (common)
- 4. Incorrect power-line voltage setting (occasional)
AC power cord connection
- Test point
- Power inlet connector
- Expected
- Power cord fully inserted and seated
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit is inoperative caution
Power supply does not turn on or respond to front-panel power switch
Ranked causes
- 1. AC power cord not connected (common)
- 2. Front-panel power switch not depressed (common)
- 3. Power-line fuse missing or incorrect (common)
- 4. Incorrect power-line voltage setting (occasional)
Verify AC power cord connection
- Test point
- power cord to power supply
- Expected
- AC power cord connected
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output ripple and noise exceeds limit
Peak-to-peak or rms voltage ripple and noise outside specification in frequency range 20 Hz to 20 MHz
Ranked causes
- 1. Output filter capacitors degraded or missing (common)
- 2. Common mode noise coupled into measurement (occasional)
- 3. Switching supply feedback loop oscillation (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply fails self-test
Unit powers on but fails internal self-test diagnostics
Ranked causes
- 1. Incorrect power-line voltage selected (common)
- 2. Load connected during self-test (occasional)
- 3. Terminal connections not removed during self-test (occasional)
Power-line voltage setting verification
- Test point
- Voltage selector
- Expected
- Set to correct input voltage (100, 115, or 230 Vac)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails self-test
Unit fails internal self-test upon startup or manual execution
Ranked causes
- 1. Incorrect power-line voltage setting (common)
- 2. Terminal connections present during self-test (common)
- 3. DAC U22 failure on PC board (occasional)
Verify correct power-line voltage setting
- Test point
- voltage selector switch
- Expected
- Matches input voltage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant Current (CC) Programming and Readback Accuracy
Verify that current programming and GPIB or RS-232 readback functions are within specifications; readback values over remote interface should match front panel display.
Ranked causes
- 1. Current monitoring resistor accuracy or measurement setup (common)
- 2. Full-range current accuracy (common)
Setup monitoring resistor and verify zero current condition
- Test point
- Current monitoring resistor (RM)
- Expected
- 0 A ± 10 mA, ±10 mA
- Tools
- Digital voltmeter, 0.01 Ω resistor
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Full voltage output reading incorrect
Digital voltmeter reading at full rated voltage setting is outside tolerance, or remote readback value differs from expected
Ranked causes
- 1. Voltage regulation circuit fault (common)
- 2. Remote readback accuracy degraded (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line regulation exceeds limit caution
Output voltage change when AC input varies from low to high line voltage exceeds specified tolerance
Ranked causes
- 1. Input filter or holdup capacitor degraded (occasional)
- 2. Voltage regulator feedback network drift (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Load regulation exceeds limit
Output voltage change from full load to no load exceeds specified tolerance
Ranked causes
- 1. Voltage regulation loop gain insufficient (common)
- 2. Load sense line disconnected or high impedance (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overcurrent condition triggered (OCP TRIPPED)
OCP annunciator blinks and OCP TRIPPED message appears on display when output current exceeds trip level
Ranked causes
- 1. Output current exceeded OCP trip point (common)
- 2. External voltage source (e.g., battery) connected to output (occasional)
Verify OCP annunciator blinking and OCP TRIPPED message
- Test point
- Display screen and OCP annunciator LED
- Expected
- OCP annunciator flashing, OCP TRIPPED message displayed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient response time exceeds limit
Recovery time after load transient exceeds 50 µsec to settle within 15 mV of nominal voltage
Ranked causes
- 1. Transient response compensation network incorrect (common)
- 2. Regulation loop bandwidth too low (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit reports errors 740 to 750
Unit displays error codes 740 to 750 during operation or startup
Ranked causes
- 1. Power interrupted during calibration (occasional)
- 2. Hardware failure (occasional)
Power cycle and recalibrate
- Test point
- unit state
- Expected
- Unit recalibrated and error cleared
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zero voltage output reading incorrect
Digital voltmeter reading at zero volt setting is outside tolerance, or remote readback value differs from front panel display
Ranked causes
- 1. Voltage programming error (common)
- 2. Remote interface calibration offset (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D converter slope convergence failed
ADC integrator slope decisions do not balance, indicating measurement accuracy degradation
Ranked causes
- 1. ADC integrator not producing balanced positive and negative slope decisions (unknown)
ADC integrator slope balance during 20 ms interval
- Test point
- ADC integrator input amplifier
- Expected
- Equal positive and negative slope decisions within ±10%, ±10%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D converter sync signal not recognized
A/D converter sync pulse fails to latch value in slope counters, causing measurement system failure
Ranked causes
- 1. A/D sync pulse not latching or timeout occurring (unknown)
A/D sync pulse recognition and latch operation
- Test point
- U19, U20 A/D converters
- Expected
- Sync interrupt recognized, slope counters latch successfully before timeout
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D slope convergence failed (Test 604)
ADC integrator does not produce balanced positive and negative slope decisions within tolerance
Ranked causes
- 1. Input amplifier or ADC integrator circuit drift or component failure (unknown)
Measure ADC slope decision balance
- Test point
- ADC integrator output, 10 V range, MZ state
- Expected
- Positive and negative slope decisions balanced within ±10% over 20 ms interval percent tolerance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D sync stuck (Test 603)
A/D converter sync interrupt not recognized or timeout occurs
Ranked causes
- 1. A/D sync pulse circuit failure in U19 or U20 (unknown)
Verify A/D sync pulse issued and interrupt recognized
- Test point
- U19, U20 sync line
- Expected
- Sync interrupt recognized without timeout within 20 ms window
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC Load Effect
Current load regulation: maximum current change under load variation must be within limits
Ranked causes
- 1. Excessive current load regulation variation (unknown)
Vary load resistance under constant current
- Test point
- Output terminals
- Expected
- E3633A: <2.25 mA change; E3634A: <0.95 mA change, E3633A <2.25 mA; E3634A <0.95 mA
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC PARD Common Mode
Constant current ripple and noise in common mode must be within specification
Ranked causes
- 1. Excessive current ripple in common mode (unknown)
Measure common mode current ripple
- Test point
- Output terminals (differential measurement)
- Expected
- <1.5 µA rms, <1.5 µA rms
- Tools
- Differential oscilloscope probe
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC PARD Normal Mode
Constant current ripple and noise in normal mode must be within specification
Ranked causes
- 1. Excessive current ripple in normal mode (unknown)
Measure output current ripple (normal mode)
- Test point
- Output terminals or current sense resistor
- Expected
- <2 mA rms, <2 mA rms
- Tools
- Oscilloscope 20 MHz bandwidth
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC Programming Accuracy
Current output programming accuracy at 0A and full scale must meet specification limits
Ranked causes
- 1. Out-of-specification current calibration (unknown)
Measure output current at 0A
- Test point
- Output terminals via ammeter or shunt
- Expected
- +0.0100 A to -0.0100 A, +/-0.0100 A
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC Readback Accuracy
Current readback accuracy verification at 0A and full scale
Ranked causes
- 1. Out-of-specification current readback (unknown)
Measure current readback at 0A
- Test point
- Output terminals; compare to front panel display
- Expected
- IO +0.0050 A to IO -0.0050 A, +/-0.0050 A from measured IO
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC Source Effect
Current line regulation: maximum current change with input line variation must be within limits
Ranked causes
- 1. Excessive current line regulation variation (unknown)
Vary AC input line +/-10%
- Test point
- Output current measurement
- Expected
- E3633A: <2.25 mA change; E3634A: <0.95 mA change, E3633A <2.25 mA; E3634A <0.95 mA
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Common Mode Current Noise
Measure ac current component between output lines and chassis ground to verify it does not exceed specification limits.
Ranked causes
- 1. Out-of-specification common mode current noise (unknown)
Set up measurement circuit with 100 kΩ resistor and 2200 pF capacitor in parallel between (-) terminal and chassis ground
- Test point
- (-) terminal to chassis ground
- Expected
- Proper circuit connection
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant Current Load Effect (Load Regulation)
Measure change in output current when load varies from full rated output voltage to short circuit in constant current mode.
Ranked causes
- 1. Load regulation at full-rated to short-circuit transition (common)
Measure output current at full-rated voltage
- Test point
- Current monitoring resistor (RM)
- Expected
- Full rated value
- Tools
- Digital voltmeter, electronic load, 0.01 Ω resistor
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant Current PARD (Ripple and Noise)
Measure periodic and random deviations (ripple and noise) in output current in frequency range 20 Hz to 20 MHz with supply in constant current operation.
Ranked causes
- 1. Ripple and noise measurement setup and verification (common)
Setup load and measurement instruments for PARD test
- Test point
- Current monitoring resistor (RM)
- Expected
- RMS ripple and noise per specification
- Tools
- RMS voltmeter, load resistor 0.4 Ω or 3.5 Ω
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Constant Current Source Effect (Line Regulation) caution
Measure change in output current resulting from AC line voltage variation from 10% below to 10% above nominal input voltage in constant current mode.
Ranked causes
- 1. Line regulation at minimum and maximum AC input voltage (common)
Setup transformer and measure current at low line voltage
- Test point
- Current monitoring resistor (RM)
- Expected
- Full rated output current at minimum AC input
- Tools
- Digital voltmeter, variable transformer, electronic load
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current and OCP
Software-based current output and overcurrent protection calibration using GPIB interface, DMM measuring across 0.001 ohm shunt
Ranked causes
- 1. Current calibration required at minimum, middle, and maximum levels (common)
Verify DAC DNL error correction completed
- Test point
- Syst:Err?
- Expected
- 0
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CV Load Effect
Load regulation verification: maximum voltage change under load variation must be within limits
Ranked causes
- 1. Excessive load regulation variation (unknown)
Vary load from zero to full rated current
- Test point
- Output terminals
- Expected
- E3633A: <4 mV change; E3634A: <7 mV change, E3633A <4 mV; E3634A <7 mV
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CV PARD
Power supply ripple and noise (PARD) verification in normal and common modes
Ranked causes
- 1. Excessive output ripple or noise (unknown)
Measure output ripple and noise
- Test point
- Output terminals
- Expected
- Normal mode: <3 mV p-p; E3633A: <350 µV rms; E3634A: <500 µV rms, <3 mV p-p; E3633A <350 µV rms; E3634A <500 µV rms
- Tools
- Oscilloscope 20 MHz bandwidth
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CV Programming Accuracy
Voltage output programming accuracy at 0V and full scale must meet specification limits for model verification
Ranked causes
- 1. Out-of-specification voltage calibration (unknown)
Measure DVM reading at 0V output
- Test point
- Output terminals
- Expected
- +0.0100 V to -0.0100 V, +/-0.0100 V
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CV Readback Accuracy
Voltage readback accuracy measurement verification at 0V and full scale
Ranked causes
- 1. Out-of-specification readback accuracy (unknown)
Measure readback accuracy at 0V
- Test point
- Output terminals; compare to front panel display
- Expected
- DVM +0.0050 V to DVM -0.0050 V, +/-0.0050 V from DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CV Source Effect
Line regulation verification: maximum voltage change with input line variation must be within limits
Ranked causes
- 1. Excessive line regulation variation (unknown)
Vary AC input line +/-10%
- Test point
- Output terminals
- Expected
- E3633A: <4 mV change; E3634A: <7 mV change, E3633A <4 mV; E3634A <7 mV
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electronic correction factors
Procedure to calibrate output voltage and current using closed-case electronic calibration with input reference values
Ranked causes
- 1. Calibration due per maintenance interval (unknown)
Determine calibration interval requirement
- Test point
- Application accuracy specification
- Expected
- 1-year interval adequate for most applications; do not extend beyond 1 year
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Load Transient Response Time
Verification that output settles within specified time when load changes
Ranked causes
- 1. Slow transient response (unknown)
Apply step load change
- Test point
- Output terminals
- Expected
- Settling time <50 µsec, <50 µsec
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance Verification
Verify power supply operates within performance specifications listed in chapter 1. Two levels: Self-Test (internal verification) and Performance Verification Tests (specification compliance).
Ranked causes
- 1. Performance drift or component degradation after repair (unknown)
Power-on self-test result
- Test point
- Front panel display
- Expected
- PASS displayed
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Unsecure power supply with security code
Power supply is [withheld] (E3633A) or [withheld] (E3634A). Technician must enter correct code to access calibration mode from front panel.
Ranked causes
- 1. Power supply in secured state at factory (common)
Verify calibration mode activated
- Test point
- Front panel display
- Expected
- CAL MODE message displayed after long beep
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Unsecure power supply without security code caution
When security code is forgotten, jumper method allows unsecuring by temporarily bridging JP5 pads inside unit, then entering any code.
Ranked causes
- 1. Security code forgotten or lost (rare)
Disconnect power and open enclosure
- Test point
- Power supply exterior
- Expected
- Power cord disconnected, all loads removed, instrument cover removed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Voltage and OVP
Software-based voltage output and overvoltage protection calibration using GPIB interface with DMM and 0.001 ohm current shunt
Ranked causes
- 1. Voltage calibration required at minimum, middle, and maximum levels (common)
Unsecure power supply for calibration
- Test point
- Cal:Sec:Stat?
- Expected
- Response should not contain '1'
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot calibrate rundown gain (Test 605)
Procedure fails to complete due to hardware failure; integrating ADC to U19 on-chip ADC gain mismatch
Ranked causes
- 1. Integrating ADC or U19 on-chip ADC component failure or gain circuit drift (unknown)
Verify ADC gain calibration can complete
- Test point
- Integrating ADC to U19 ADC interface
- Expected
- Gain calibration procedure completes without timeout or hardware error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Complete self-test initiated
User holds any front panel key except Error key for more than 5 seconds while powering on; unit beeps when test starts
Ranked causes
- 1. Operator initiates extended diagnostics (common)
Confirm unit beeps when test starts
- Test point
- Audible output
- Expected
- Single beep at start of complete self-test
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display returns to OCP TRIPPED message after timeout
After exiting OCP menu, display reverts to OCP TRIPPED message after several seconds instead of remaining in normal mode
Ranked causes
- 1. Overcurrent condition not fully cleared (common)
Ensure OCP clear mode is reached and completed
- Test point
- Display screen
- Expected
- DONE message displayed, OCP annunciator stops blinking
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan cooling circuit failure caution
Fan does not operate or fan current not detected, risking overtemperature condition
Ranked causes
- 1. Fan current not detected, fan stopped or circuit open (common)
Fan current detection at fan drive circuit
- Test point
- Fan drive output
- Expected
- Fan current flowing, fan spinning
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
I/O processor communication failure
Communications cannot be established between main controller and front panel processor through optical link
Ranked causes
- 1. Optically isolated serial link (U2, U5) failure (unknown)
Bidirectional serial communication through optical isolators
- Test point
- U2, U5 optoisolators, serial data signals
- Expected
- Communication established successfully both directions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
I/O processor internal RAM test failed
Front panel processor U1 internal RAM test does not pass, indicates processor memory or logic failure
Ranked causes
- 1. U1 internal RAM failure or processor logic error (unknown)
U1 internal RAM and processor self-test execution
- Test point
- U1 front panel processor
- Expected
- Internal RAM test passes without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line frequency detection failed
Power supply cannot sense AC line frequency, system defaults to 50 Hz operation
Ranked causes
- 1. LSENCE logic input signal not toggling or not detected (unknown)
LSENCE logic input signal presence and toggling
- Test point
- U19 LSENCE input
- Expected
- Logic input toggling at line frequency
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OCP annunciator not enabled after setting trip level
OCP annunciator fails to turn on after completing OCP menu configuration steps
Ranked causes
- 1. OCP circuit not properly enabled in menu (common)
Verify OCP menu entry and trip level setting
- Test point
- Display screen
- Expected
- LEVEL message displayed with trip level value
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage protection trip cleared by trip level adjustment
OVP condition resolved by raising OVP trip level above current output voltage
Ranked causes
- 1. Output voltage at or above OVP trip level (common)
Verify OVP trip level is raised above output voltage
- Test point
- OVP trip level display
- Expected
- OVP trip level > current output voltage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage protection trip cleared by voltage adjustment
OVP condition resolved by lowering output voltage below trip point
Ranked causes
- 1. Output voltage exceeds OVP trip point (common)
Verify output voltage is below OVP trip point
- Test point
- Output display
- Expected
- Output voltage < OVP trip level setting
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage protection trip occurs
Output voltage reaches or exceeds OVP trip level; power supply output drops to near zero, OVP annunciator blinks, CC annunciator turns on, 'OVP TRIPPED' message displayed
Ranked causes
- 1. External voltage source connected to output (e.g., battery) (occasional)
Verify external voltage source presence
- Test point
- Output terminals
- Expected
- No external voltage source connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply does not respond to front panel input
Front panel keys do not function or are unresponsive to user commands
Ranked causes
- 1. Front panel controller U1 communication failure (unknown)
Verify front panel controller U1 responds and display segments activate
- Test point
- U1 front panel controller
- Expected
- All display segments illuminated, serial communication established both directions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RAM read/write failed (Test 602)
Unit fails memory test on RAM U14; error readable only from remote interface
Ranked causes
- 1. RAM U14 defective or communication error (unknown)
Write and read test patterns to RAM U14
- Test point
- RAM U14 address range
- Expected
- 55h and AAh patterns read back correctly from all addresses
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RAM read/write failure
Memory test fails during self-test, indicates RAM data corruption or hardware failure
Ranked causes
- 1. RAM U14 defective or data path corrupted (unknown)
Read/write test to RAM U14 with checker board pattern
- Test point
- U14 RAM
- Expected
- 55h and AAh pattern written and read back successfully at all addresses
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rated output voltage not achievable
Front-panel voltmeter does not respond to knob control or voltage cannot be adjusted from zero to maximum rated value (8.0 V for E3633A or 25.0 V for E3634A)
Ranked causes
- 1. Knob control not responding or voltage adjustment mechanism failure (unknown)
Verify power supply is turned on and outputs enabled
- Test point
- Front panel OFF annunciator and 8V/25V annunciator
- Expected
- OFF annunciator off; 8V or 25V, OVP, OCP, and CV annunciators lit
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rated output voltage not achievable (E3634A)
Front-panel voltmeter does not respond to knob control or voltage cannot be adjusted from zero to 25.0 volts maximum rated value
Ranked causes
- 1. Knob control not responding or voltage adjustment mechanism failure (unknown)
Verify power supply is turned on and outputs enabled
- Test point
- Front panel OFF annunciator and 25V annunciator
- Expected
- OFF annunciator off; 25V, OVP, OCP, and CV annunciators lit
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rundown gain calibration cannot complete
Rundown gain calibration procedure cannot finish, indicates hardware failure in measurement circuit
Ranked causes
- 1. Hardware failure prevents rundown gain calibration completion (unknown)
Rundown gain calibration execution
- Test point
- Integrating ADC, U19 on-chip ADC
- Expected
- Calibration completes without timeout or error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rundown gain out of specification
Measured gain between integrating ADC and U19 on-chip ADC exceeds tolerance limits
Ranked causes
- 1. Gain measured outside ±10% tolerance (unknown)
Rundown gain measurement
- Test point
- Integrating ADC to U19 ADC signal path
- Expected
- Gain within ±10% of nominal value, ±10%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rundown gain repeatability excessive
Gain measurement noise exceeds specification, causing unstable measurement accuracy
Ranked causes
- 1. Gain noise exceeds ±64 LSB specification (unknown)
Rundown gain noise measurement (8 repeated tests)
- Test point
- U19 on-chip ADC output
- Expected
- Gain noise less than ±64 LSB, ±64 LSB
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Self-Test: Complete verification
- Serial configuration readback failed
- System DAC hardware not functional
- Voltage or current error amplifier not operational
4 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
-330 | Self-test failed | Self-test routine execution | error |
-350 | Too many errors | Error threshold exceeded during operation | error |
501 | Isolator UART framing error | Serial communication with isolator circuit | error |
502 | Isolator UART overrun error | Isolator serial input buffer overflow | warning |
511 | RS-232 framing error | Serial port RS-232 communication | error |
512 | RS-232 overrun error | RS-232 input buffer overflow | warning |
513 | RS-232 parity error | RS-232 serial data reception | error |
514 | Command allowed only with RS-232 | Non-RS-232 interface command attempt | error |
521 | Input buffer overflow | Input buffer capacity exceeded | error |
522 | Output buffer overflow | Output buffer capacity exceeded | error |
550 | Command not allowed in local | Remote command in local mode | error |
601 | Front panel controller communication failure | Main controller U19 cannot establish serial communications with U1 on front panel board | error |
601 | Front Panel Does not respond | Main controller U19 cannot establish serial communication with front panel controller U1 | error |
602 | RAM read/write failed | Checkerboard pattern test (55h and AAh) fails on RAM U14 | error |
603 | A/D sync signal stuck | A/D sync interrupt not recognized or timeout occurs before sync completes | error |
604 | A/D slope convergence failed | ADC slope decisions not balanced within ±10% in 10 V range | error |
605 | Cannot calibrate rundown gain | Rundown gain calibration procedure cannot complete due to hardware failure | error |
606 | Rundown gain out of range | Measured gain between integrating ADC and U19 on-chip ADC outside ±10% tolerance | warning |
607 | Rundown gain too noisy | Gain noise from repeated tests exceeds ±64 LSB specification on U19 ADC output | warning |
608 | Serial configuration readback failed | 3-byte serial configuration data does not match after transmission through U23 | error |
624 | Unable to sense line frequency | LSENCE logic input to U19 not toggling or not detected | warning |
625 | I/O processor did not respond | Communication not established between U19 and U1 through optically isolated link in either direction | error |
626 | I/O processor failed self-test | U1 internal RAM test fails or processor logic error detected during execution | error |
630 | Fan test failed | Fan current not detected at power-on self-test, fan not running | critical |
631 | System DAC test failed | DAC output readback after reference voltage test outside valid range | error |
632 | Hardware test failed (error amplifier failure) | Voltage and/or current error amplifier not operational or both amplifiers not responding | critical |
701 | Cal security disabled by jumper | Calibration startup check | info |
702 | Cal secured | Calibration access attempt when secured | warning |
703 | Invalid secure code | Calibration security code entry | error |
704 | Secure code too long | Security code entry during calibration | error |
705 | Cal aborted | Calibration routine interrupted | warning |
708 | Cal output disabled | Calibration output verification | error |
712 | Bad DAC cal data | DAC calibration data readback | error |
713 | Bad readback cal data | Calibration readback verification | error |
714 | Bad OVP cal data | Overvoltage Protection calibration data check | error |
715 | Bad OCP cal data | Overcurrent Protection calibration data verification | error |
716 | Bad DAC DNL error correction data | DAC Differential Nonlinearity correction data check | error |
717 | Cal OVP or OCP status enabled | Calibration verification check | warning |
718 | Gain out of range for gain error correction | Gain calibration or correction routine | error |
740 | Cal checksum failed, secure state | Calibration security checksum verification | critical |
741 | Cal checksum failed, string data | Calibration string data checksum verification | critical |
742 | Cal checksum failed, store/recall data in location 0 | Calibration storage location 0 checksum check | critical |
743 | Cal checksum failed, store/recall data in location 1 | Calibration storage location 1 checksum verification | critical |
744 | Cal checksum failed, store/recall data in location 2 | Calibration storage location 2 checksum check | critical |
745 | Cal checksum failed, store/recall data in location 3 | Calibration storage location 3 checksum verification | critical |
746 | Cal checksum failed, DAC cal constants | DAC calibration constants checksum check | critical |
747 | Cal checksum failed, readback cal constants | Readback calibration constants checksum verification | critical |
748 | Cal checksum failed, GPIB address | GPIB address calibration checksum check | critical |
749 | Cal checksum failed, internal data | Internal calibration data checksum verification | critical |
750 | Cal checksum failed, DAC DNL error correction data | DAC DNL correction data checksum check | critical |
CHANGED | OVP settings have been successfully updated | User modifies OVP trip level and exits OVP menu | info |
CHANGED | OCP settings have been modified and new trip level is now active | User changes OCP trip level setting and exits OCP menu | info |
DONE | Overcurrent condition has been successfully cleared and power supply has returned to normal operation | User completes OCP clear procedure and presses confirmation key | info |
ERROR | Error detected during output checkout procedure | Detected during steps 1-6 of rated current output checkout | warning |
Error | Incorrect security code entered or jumper left connected during unsecure attempt | Invalid security code entry from front panel; JP5 jumper not removed after unsecure operation | error |
NO CHANGE | OCP settings were not modified during menu session | User exits OCP menu without changing trip level value | info |
OCP TRIPPED | Overcurrent protection circuit has activated due to output current exceeding the programmed trip level | Output current exceeds OCP trip point setting | warning |
OVP TRIPPED | Overvoltage protection circuit has activated due to output voltage reaching or exceeding the OVP trip level | Output voltage rises to or above the programmed OVP trip level | warning |
SECURED | Power supply is in secured state and calibration access is blocked | Power supply powered on while security code protection is active | warning |
Test 601 | Front panel communication failed | Serial link between U19 and U1 does not establish bidirectional communication | error |
Test 602 | RAM read/write failed | Checker board pattern (55h, AAh) not read back correctly from U14 | error |
Test 603 | A/D sync stuck | Sync interrupt not recognized or timeout during A/D sync pulse latch | error |
Test 604 | A/D slope convergence failed | Positive and negative slope decisions not balanced within ±10% in 20 ms interval | error |
Test 605 | Cannot calibrate rundown gain | Gain calibration procedure fails to complete due to hardware error | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI