Agilent E3631A: troubleshooting and error codes
The ServiceForge knowledge base holds 18 documented symptoms for the Agilent E3631A, with 31 ranked causes, 67 diagnostic checks with test points and expected values, and 47 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power on
Power supply is completely inoperative, no display or output.
Ranked causes
- 1. AC power cord not connected (common)
- 2. Front-panel power switch not pressed (common)
- 3. Power-line fuse missing or incorrect (common)
- 4. Incorrect power-line voltage setting (occasional)
AC power cord connection
- Test point
- Power supply inlet
- Expected
- Power cord firmly connected
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit is inoperative or does not power on
Power supply does not turn on; no output or indicator lights
Ranked causes
- 1. AC power cord not connected (common)
- 2. Front-panel power switch not depressed (common)
- 3. Power-line fuse blown or missing (common)
- 4. Wrong power-line voltage setting (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias supply voltage out of specification
Measured bias supply voltage outside specified limits
Ranked causes
- 1. Regulator input voltage insufficient (common)
- 2. Circuit failure or excessive load on supply (occasional)
- 3. AC oscillation on power supply output (rare)
Measure +5V Floating supply voltage
- Test point
- U1 pin 2 on top board
- Expected
- 4.75 V to 5.25 V, +/- 0.25 V
- Tools
- DMM
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails built-in self-test
Self-test does not complete or reports failures
Ranked causes
- 1. Wrong power-line voltage setting (common)
- 2. Terminal connections present during test (common)
- 3. DAC U12 integrated circuit failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Output voltage and current caution
Procedure to calibrate +6V, +25V, and -25V outputs. Unsecure power supply, disconnect loads, enter calibration mode by pressing Calibrate key at power-on for 5 seconds until beep.
Ranked causes
- 1. Output voltage requires recalibration (common)
- 2. Output current requires recalibration (common)
Unsecure the power supply
7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply fails self-test
Unit powers on but does not pass internal self-test; may display error or not initialize properly.
Ranked causes
- 1. Incorrect power-line voltage selected (common)
- 2. Load connected during self-test (common)
Power-line voltage setting correct
- Test point
- Voltage selector
- Expected
- Matches local AC mains voltage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit displays errors 740 through 748
Error codes 740-748 displayed on unit
Ranked causes
- 1. Power loss during calibration or state change (common)
- 2. Hardware failure (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration program configuration
Software setup parameters for GPIB addresses and current shunt value used in calibration.
Ranked causes
- 1. Current shunt value mismatch or incorrect GPIB addresses (common)
Verify current shunt value matches hardware
- Test point
- Current_shunt variable
- Expected
- 0.01 ohm or actual shunt value in use
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC Line Regulation caution
Measure change in output current from minimum to maximum AC line voltage (±10% nominal) in constant current mode
Ranked causes
- 1. CC line regulation out of specification (unknown)
Measure current at low line voltage (-10% nominal)
- Test point
- Across 0.1Ω current monitoring resistor
- Expected
- Record baseline current at low line voltage
- Tools
- Digital voltmeter, variable voltage transformer, electronic load, 0.1Ω resistor
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC Load Regulation
Measure change in output current from full-scale voltage to short circuit in constant current mode
Ranked causes
- 1. CC load regulation out of specification (unknown)
Measure current at full-scale output voltage (constant voltage mode)
- Test point
- Across 0.1Ω current monitoring resistor
- Expected
- Record baseline current with load in constant voltage mode
- Tools
- Digital voltmeter, electronic load, 0.1Ω resistor
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current Programming and Readback Accuracy
Verify that current programming over remote interface and front panel readback match specifications for 0A and full-scale current on +6V, +25V, and -25V outputs
Ranked causes
- 1. Current programming or readback out of specification (unknown)
Measure 0A output current at programming
- Test point
- Across 0.1Ω current monitoring resistor
- Expected
- +6V: 0A ±10mA; +25V: 0A ±4mA; -25V: 0A ±4mA, +/-10mA to ±4mA depending on output
- Tools
- Digital voltmeter, 0.1Ω resistor
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: E3631A voltage and current adjustment
Software calibration procedure using GPIB interface with Agilent 34401A DMM and 0.01 ohm current shunt to adjust output voltage and current accuracy for three power supply outputs.
Ranked causes
- 1. Voltage and current calibration required (unknown)
Disable calibration security
- Test point
- CAL:SEC:STAT
- Expected
- OFF
9 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Normal Mode Current Noise (CC Ripple and Noise)
Measure RMS output current noise in frequency range 20Hz to 20MHz during constant current operation
Ranked causes
- 1. CC ripple and noise exceeds specification (unknown)
Measure RMS current noise across resistive load
- Test point
- Across load resistor (1.2Ω for +6V, 25Ω for ±25V)
- Expected
- +6V: ≤2mA; +25V: ≤0.5mA; -25V: ≤0.5mA, ≤2mA for +6V, ≤0.5mA for ±25V
- Tools
- RMS voltmeter, resistive load (1.2Ω or 25Ω)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance verification
Verify power supply meets specifications after repairs to specific circuits
Ranked causes
- 1. Need to verify output specifications after circuit repair (common)
Perform power-on self-test
- Test point
- Front panel display
- Expected
- PASS displayed (or FAIL with ERROR annunciator if issue)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Rated voltage outputs high voltage
Check that power supply develops rated voltage outputs on all supplies (+6V, +25V, -25V) with no load and responds to front panel control
Ranked causes
- 1. Outputs not at rated voltage or unresponsive to control (unknown)
Power supply powers on and enters reset state
- Test point
- Front panel indicators
- Expected
- OFF annunciator on, +6V annunciator on
12 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Self-test verification
Verify power supply is operational using self-test function
Ranked causes
- 1. Initial verification of power supply operation required (common)
Observe automatic self-test result on power-up
- Test point
- Front panel display and ERROR annunciator
- Expected
- PASS displayed (indicates power supply operational)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Unsecure power supply with security code
Front panel procedure to enter security code and unsecure the E3631A for calibration operations
Ranked causes
- 1. Power supply secured with calibration security code (common)
Turn on calibration mode
- Test point
- Front panel Calibrate key
- Expected
- Beep heard after ~5 seconds, CAL MODE message displayed
- Tools
- None
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Unsecure power supply without security code caution
Internal jumper bypass procedure when security code is forgotten; requires opening instrument cover and shorting JP1
Ranked causes
- 1. Security code forgotten or unknown (occasional)
Disconnect power and load connections
- Test point
- Power cord and front terminals
- Expected
- All connections removed
- Tools
- None
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
-330 | Self-test failed | Self-test execution | critical |
-350 | Too many errors | Multiple errors detected during operation | critical |
501 | Isolator UART framing error | Communication between isolator and main processor | error |
502 | Isolator UART overrun error | Isolator UART buffer overflow during data transfer | error |
511 | RS-232 framing error | RS-232 serial communication | error |
512 | RS-232 overrun error | RS-232 buffer overflow during reception | error |
513 | RS-232 parity error | RS-232 data reception with parity mismatch | warning |
514 | Command allowed only with RS-232 | Attempt to execute RS-232-only command via different interface | warning |
521 | Input buffer overflow | Command input buffer capacity exceeded | error |
522 | Output buffer overflow | Output buffer capacity exceeded | error |
550 | Command not allowed in local | Attempt to execute remote command in local mode | warning |
601 | Front Panel Does not respond | Main controller U17 fails to establish serial communications with front panel controller U7 | error |
602 | RAM read/write failed | Checkerboard pattern test (55h and AAh) fails on RAM U15 | error |
603 | A/D sync stuck | A/D sync interrupt not recognized; sync pulse timeout occurs | error |
604 | A/D slope convergence failed | ADC integrator positive and negative slope decisions deviate beyond ±10% during 20 ms interval | error |
605 | Cannot calibrate rundown gain | Gain calibration procedure cannot complete due to hardware failure | error |
606 | Rundown gain out of range | Nominal gain between integrating ADC and U17 on-chip ADC exceeds ±10% tolerance | warning |
607 | Rundown too noisy | Gain noise exceeds ±64 lsb's of U17 on-chip ADC after eight repetitions of gain test | warning |
608 | Serial configuration readback failed | Serial configuration data (last 3 bytes) does not match when clocked back into U16 | error |
624 | Unable to sense line frequency | LSENCE logic input on U17 not toggling | warning |
625 | I/O processor did not respond | Communication not established between U16 (top board) and U10 (bottom board) via optically isolated serial link | error |
626 | I/O processor failed self-test | Earth referenced processor U10 internal RAM test fails | error |
630 | Fan test failed | Fan current not detected | error |
631 | System DAC test failed | DAC output digital conversion value outside valid range | error |
632 | P6V hardware test failed | Voltage and/or current error amplifiers for +6V circuit not operational | error |
633 | P25V hardware test failed | Voltage and/or current error amplifiers for +25V circuit not operational | error |
634 | N25V hardware test failed | Voltage and/or current error amplifiers for -25V circuit not operational | error |
701 | Cal security disabled by jumper | Calibration security jumper removed during operation | warning |
702 | Cal secured | Calibration mode locked after security code entry | info |
703 | Invalid secure code | Incorrect calibration security code entered | warning |
704 | Secure code too long | Calibration security code exceeds maximum length | warning |
708 | Cal output disabled | Calibration output selection command rejected | warning |
711 | Cal sequence interrupted | Calibration procedure stopped before completion | error |
712 | Bad DAC cal data | DAC calibration data read back differs from written | error |
713 | Bad readback cal data | Readback circuit calibration data verification failed | error |
740 | Cal checksum failed, secure state | Security state data checksum mismatch at startup | error |
741 | Cal checksum failed, string data | Calibration string data checksum mismatch | error |
742 | Cal checksum failed, store/recall data in location 1 | Store/recall memory location 1 checksum error | warning |
743 | Cal checksum failed, store/recall data in location 2 | Store/recall memory location 2 checksum error | warning |
744 | Cal checksum failed, store/recall data in location 3 | Store/recall memory location 3 checksum error | warning |
745 | Cal checksum failed, DAC cal constants | DAC calibration constants checksum mismatch | error |
746 | Cal checksum failed, readback cal constants | Readback calibration constants checksum mismatch | error |
747 | Cal checksum failed, GPIB address | GPIB address storage checksum error at startup | warning |
748 | Cal checksum failed, internal data | Internal calibration data checksum mismatch | error |
800 | P25V and N25V coupled by track system | Tracking coupling detected between positive and negative 25V supplies | error |
801 | P25V and N25V coupled by trigger subsystem | Trigger subsystem couples positive and negative 25V supplies | error |
ERROR | Error detected during output checkout | Output verification procedure encounters error condition | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI