Agilent E3631A: troubleshooting and error codes

The ServiceForge knowledge base holds 18 documented symptoms for the Agilent E3631A, with 31 ranked causes, 67 diagnostic checks with test points and expected values, and 47 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit does not power on

Power supply is completely inoperative, no display or output.

Ranked causes

  1. 1. AC power cord not connected (common)
  2. 2. Front-panel power switch not pressed (common)
  3. 3. Power-line fuse missing or incorrect (common)
  4. 4. Incorrect power-line voltage setting (occasional)
First check

AC power cord connection

Test point
Power supply inlet
Expected
Power cord firmly connected

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit is inoperative or does not power on

Power supply does not turn on; no output or indicator lights

Ranked causes

  1. 1. AC power cord not connected (common)
  2. 2. Front-panel power switch not depressed (common)
  3. 3. Power-line fuse blown or missing (common)
  4. 4. Wrong power-line voltage setting (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bias supply voltage out of specification

Measured bias supply voltage outside specified limits

Ranked causes

  1. 1. Regulator input voltage insufficient (common)
  2. 2. Circuit failure or excessive load on supply (occasional)
  3. 3. AC oscillation on power supply output (rare)
First check

Measure +5V Floating supply voltage

Test point
U1 pin 2 on top board
Expected
4.75 V to 5.25 V, +/- 0.25 V
Tools
DMM

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit fails built-in self-test

Self-test does not complete or reports failures

Ranked causes

  1. 1. Wrong power-line voltage setting (common)
  2. 2. Terminal connections present during test (common)
  3. 3. DAC U12 integrated circuit failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Output voltage and current caution

Procedure to calibrate +6V, +25V, and -25V outputs. Unsecure power supply, disconnect loads, enter calibration mode by pressing Calibrate key at power-on for 5 seconds until beep.

Ranked causes

  1. 1. Output voltage requires recalibration (common)
  2. 2. Output current requires recalibration (common)
First check

Unsecure the power supply

7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply fails self-test

Unit powers on but does not pass internal self-test; may display error or not initialize properly.

Ranked causes

  1. 1. Incorrect power-line voltage selected (common)
  2. 2. Load connected during self-test (common)
First check

Power-line voltage setting correct

Test point
Voltage selector
Expected
Matches local AC mains voltage

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit displays errors 740 through 748

Error codes 740-748 displayed on unit

Ranked causes

  1. 1. Power loss during calibration or state change (common)
  2. 2. Hardware failure (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration program configuration

Software setup parameters for GPIB addresses and current shunt value used in calibration.

Ranked causes

  1. 1. Current shunt value mismatch or incorrect GPIB addresses (common)
First check

Verify current shunt value matches hardware

Test point
Current_shunt variable
Expected
0.01 ohm or actual shunt value in use

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CC Line Regulation caution

Measure change in output current from minimum to maximum AC line voltage (±10% nominal) in constant current mode

Ranked causes

  1. 1. CC line regulation out of specification (unknown)
First check

Measure current at low line voltage (-10% nominal)

Test point
Across 0.1Ω current monitoring resistor
Expected
Record baseline current at low line voltage
Tools
Digital voltmeter, variable voltage transformer, electronic load, 0.1Ω resistor

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CC Load Regulation

Measure change in output current from full-scale voltage to short circuit in constant current mode

Ranked causes

  1. 1. CC load regulation out of specification (unknown)
First check

Measure current at full-scale output voltage (constant voltage mode)

Test point
Across 0.1Ω current monitoring resistor
Expected
Record baseline current with load in constant voltage mode
Tools
Digital voltmeter, electronic load, 0.1Ω resistor

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Current Programming and Readback Accuracy

Verify that current programming over remote interface and front panel readback match specifications for 0A and full-scale current on +6V, +25V, and -25V outputs

Ranked causes

  1. 1. Current programming or readback out of specification (unknown)
First check

Measure 0A output current at programming

Test point
Across 0.1Ω current monitoring resistor
Expected
+6V: 0A ±10mA; +25V: 0A ±4mA; -25V: 0A ±4mA, +/-10mA to ±4mA depending on output
Tools
Digital voltmeter, 0.1Ω resistor

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: E3631A voltage and current adjustment

Software calibration procedure using GPIB interface with Agilent 34401A DMM and 0.01 ohm current shunt to adjust output voltage and current accuracy for three power supply outputs.

Ranked causes

  1. 1. Voltage and current calibration required (unknown)
First check

Disable calibration security

Test point
CAL:SEC:STAT
Expected
OFF

9 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Normal Mode Current Noise (CC Ripple and Noise)

Measure RMS output current noise in frequency range 20Hz to 20MHz during constant current operation

Ranked causes

  1. 1. CC ripple and noise exceeds specification (unknown)
First check

Measure RMS current noise across resistive load

Test point
Across load resistor (1.2Ω for +6V, 25Ω for ±25V)
Expected
+6V: ≤2mA; +25V: ≤0.5mA; -25V: ≤0.5mA, ≤2mA for +6V, ≤0.5mA for ±25V
Tools
RMS voltmeter, resistive load (1.2Ω or 25Ω)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Performance verification

Verify power supply meets specifications after repairs to specific circuits

Ranked causes

  1. 1. Need to verify output specifications after circuit repair (common)
First check

Perform power-on self-test

Test point
Front panel display
Expected
PASS displayed (or FAIL with ERROR annunciator if issue)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Rated voltage outputs high voltage

Check that power supply develops rated voltage outputs on all supplies (+6V, +25V, -25V) with no load and responds to front panel control

Ranked causes

  1. 1. Outputs not at rated voltage or unresponsive to control (unknown)
First check

Power supply powers on and enters reset state

Test point
Front panel indicators
Expected
OFF annunciator on, +6V annunciator on

12 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Self-test verification

Verify power supply is operational using self-test function

Ranked causes

  1. 1. Initial verification of power supply operation required (common)
First check

Observe automatic self-test result on power-up

Test point
Front panel display and ERROR annunciator
Expected
PASS displayed (indicates power supply operational)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Unsecure power supply with security code

Front panel procedure to enter security code and unsecure the E3631A for calibration operations

Ranked causes

  1. 1. Power supply secured with calibration security code (common)
First check

Turn on calibration mode

Test point
Front panel Calibrate key
Expected
Beep heard after ~5 seconds, CAL MODE message displayed
Tools
None

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Unsecure power supply without security code caution

Internal jumper bypass procedure when security code is forgotten; requires opening instrument cover and shorting JP1

Ranked causes

  1. 1. Security code forgotten or unknown (occasional)
First check

Disconnect power and load connections

Test point
Power cord and front terminals
Expected
All connections removed
Tools
None

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
-330Self-test failedSelf-test executioncritical
-350Too many errorsMultiple errors detected during operationcritical
501Isolator UART framing errorCommunication between isolator and main processorerror
502Isolator UART overrun errorIsolator UART buffer overflow during data transfererror
511RS-232 framing errorRS-232 serial communicationerror
512RS-232 overrun errorRS-232 buffer overflow during receptionerror
513RS-232 parity errorRS-232 data reception with parity mismatchwarning
514Command allowed only with RS-232Attempt to execute RS-232-only command via different interfacewarning
521Input buffer overflowCommand input buffer capacity exceedederror
522Output buffer overflowOutput buffer capacity exceedederror
550Command not allowed in localAttempt to execute remote command in local modewarning
601Front Panel Does not respondMain controller U17 fails to establish serial communications with front panel controller U7error
602RAM read/write failedCheckerboard pattern test (55h and AAh) fails on RAM U15error
603A/D sync stuckA/D sync interrupt not recognized; sync pulse timeout occurserror
604A/D slope convergence failedADC integrator positive and negative slope decisions deviate beyond ±10% during 20 ms intervalerror
605Cannot calibrate rundown gainGain calibration procedure cannot complete due to hardware failureerror
606Rundown gain out of rangeNominal gain between integrating ADC and U17 on-chip ADC exceeds ±10% tolerancewarning
607Rundown too noisyGain noise exceeds ±64 lsb's of U17 on-chip ADC after eight repetitions of gain testwarning
608Serial configuration readback failedSerial configuration data (last 3 bytes) does not match when clocked back into U16error
624Unable to sense line frequencyLSENCE logic input on U17 not togglingwarning
625I/O processor did not respondCommunication not established between U16 (top board) and U10 (bottom board) via optically isolated serial linkerror
626I/O processor failed self-testEarth referenced processor U10 internal RAM test failserror
630Fan test failedFan current not detectederror
631System DAC test failedDAC output digital conversion value outside valid rangeerror
632P6V hardware test failedVoltage and/or current error amplifiers for +6V circuit not operationalerror
633P25V hardware test failedVoltage and/or current error amplifiers for +25V circuit not operationalerror
634N25V hardware test failedVoltage and/or current error amplifiers for -25V circuit not operationalerror
701Cal security disabled by jumperCalibration security jumper removed during operationwarning
702Cal securedCalibration mode locked after security code entryinfo
703Invalid secure codeIncorrect calibration security code enteredwarning
704Secure code too longCalibration security code exceeds maximum lengthwarning
708Cal output disabledCalibration output selection command rejectedwarning
711Cal sequence interruptedCalibration procedure stopped before completionerror
712Bad DAC cal dataDAC calibration data read back differs from writtenerror
713Bad readback cal dataReadback circuit calibration data verification failederror
740Cal checksum failed, secure stateSecurity state data checksum mismatch at startuperror
741Cal checksum failed, string dataCalibration string data checksum mismatcherror
742Cal checksum failed, store/recall data in location 1Store/recall memory location 1 checksum errorwarning
743Cal checksum failed, store/recall data in location 2Store/recall memory location 2 checksum errorwarning
744Cal checksum failed, store/recall data in location 3Store/recall memory location 3 checksum errorwarning
745Cal checksum failed, DAC cal constantsDAC calibration constants checksum mismatcherror
746Cal checksum failed, readback cal constantsReadback calibration constants checksum mismatcherror
747Cal checksum failed, GPIB addressGPIB address storage checksum error at startupwarning
748Cal checksum failed, internal dataInternal calibration data checksum mismatcherror
800P25V and N25V coupled by track systemTracking coupling detected between positive and negative 25V supplieserror
801P25V and N25V coupled by trigger subsystemTrigger subsystem couples positive and negative 25V supplieserror
ERRORError detected during output checkoutOutput verification procedure encounters error conditionerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI