Agilent E3620A: troubleshooting and error codes
The ServiceForge knowledge base holds 32 documented symptoms for the Agilent E3620A, with 65 ranked causes, 36 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
V2 preregulator control circuit fault (at 10V)
V2 output voltage control circuit not functioning correctly when set to 10V ±1V
Ranked causes
- 1. U1 or Q3 defective (high voltage at U4 pin 1) (occasional)
- 2. U4 defective (occasional)
- 3. U6 defective (occasional)
- 4. U7 defective (occasional)
- 5. U5 defective (occasional)
Measure voltage at pin 1 of U4
- Test point
- U4 pin 1
- Expected
- Low voltage (0 V) indicates normal; high voltage (+0.7 V) indicates U1 or Q3 defective
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
V2 preregulator control circuit fault (at 19V)
V2 output voltage control circuit not functioning correctly when set to 19V ±1V
Ranked causes
- 1. U2 or Q4 defective (high voltage at U4 pin 7) (occasional)
- 2. U4 defective (occasional)
- 3. U6 defective (occasional)
- 4. U7 defective (occasional)
- 5. U5 defective (occasional)
Measure voltage at pin 7 of U4
- Test point
- U4 pin 7
- Expected
- Low voltage (0 V) indicates normal; high voltage (+0.7 V) indicates U2 or Q4 defective
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low output voltage (V1 supply)
V1 output voltage below the rated minimum value
Ranked causes
- 1. Q1 open (common)
- 2. Q6 shorted (occasional)
- 3. U10A defective (occasional)
- 4. U10B defective or shorted CR14, CR15 or R48, C9 (occasional)
Attempt to turn on Q1 by disconnecting emitter of Q10
- Test point
- Q10 emitter
- Expected
- Output voltage should increase if Q1 can be turned on
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low output voltage (V2 supply)
V2 output voltage below the rated minimum value
Ranked causes
- 1. Q1 open (common)
- 2. Q6 shorted (occasional)
- 3. U10A defective (occasional)
- 4. U10B defective or CR14, CR15, R48, C9 shorted or leaky (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High output voltage (V1 supply)
V1 output voltage exceeds the rated maximum value
Ranked causes
- 1. Q1 shorted (common)
- 2. Open CR8 or R48 or defective U10B (occasional)
- 3. Q6 defective (occasional)
Attempt to turn off Q2 by shorting emitter-to-collector of Q10
- Test point
- Q2 emitter-to-collector
- Expected
- Output voltage should decrease if Q2 can be turned off
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High output voltage (V2 supply)
V2 output voltage exceeds the rated maximum value
Ranked causes
- 1. Q1 shorted (common)
- 2. Open CR8 or R48 or defective U10B (occasional)
- 3. Q6 defective (occasional)
Attempt to turn off Q1 by shorting emitter-to-collector of Q6
- Test point
- Q6 emitter-to-collector
- Expected
- Output voltage should decrease
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High ripple on output
Excessive voltage ripple observed on supply output
Ranked causes
- 1. Ground loop in operating setup (common)
- 2. Main rectifiers open (occasional)
- 3. Supply operating in current limit mode (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oscillation or poor transient response
Output exhibits ringing, oscillation, or slow recovery following load changes
Ranked causes
- 1. Open C13 or C35 (high frequency oscillation) (occasional)
- 2. Defective output capacitor (occasional)
- 3. Open C34 or C12 (current limit mode oscillation) (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
V1 preregulator control circuit fault
V1 output voltage control circuit not functioning correctly at 16V setting
Ranked causes
- 1. U18B defective (when pin 7 voltage near +11.3V) (occasional)
- 2. U3 or Q5 defective (occasional)
- 3. Open Q9 or R6 (occasional)
Measure voltage at pin 7 of U18
- Test point
- U18 pin 7
- Expected
- -12V
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
-12V bias voltage out of spec
The -12V bias supply voltage is not within normal range (±5%)
Ranked causes
- 1. U14 or U20 open (for high voltage) (occasional)
- 2. U14 or U20 shorted (for low voltage) (occasional)
Measure -12V bias voltage
- Test point
- -12V supply
- Expected
- -12V ± 5%, ±5%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
-5V reference voltage out of spec
The -5V reference supply voltage is not within normal range (±2%)
Ranked causes
- 1. U12 or U21 open (for high voltage) (occasional)
- 2. U12 or U21 shorted (for low voltage) (occasional)
Measure -5V reference voltage
- Test point
- -5V reference
- Expected
- -5V ± 2%, ±2%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
+12V bias voltage out of spec
The +12V bias supply voltage is not within normal range (±5%)
Ranked causes
- 1. U13 or U23 shorted (for high voltage) (occasional)
- 2. U13 or U23 open (for low voltage) (occasional)
Measure +12V bias voltage
- Test point
- +12V supply
- Expected
- +12V ± 5%, ±5%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
+5V reference voltage out of spec
The +5V reference supply voltage is not within normal range (±2%)
Ranked causes
- 1. U11 or U22 open (for high voltage) (occasional)
- 2. U11 or U22 shorted (for low voltage) (occasional)
Measure +5V reference voltage
- Test point
- +5V reference
- Expected
- +5V ± 2%, ±2%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current limit circuit verification required caution
Verify current limit functions and supplies minimum rated current when output is shorted
Ranked causes
- 1. Current limit circuit malfunction or drift (unknown)
- 2. Short circuit protection circuit failure (unknown)
Measure current with shorted output at maximum voltage
- Test point
- Across current sampling resistor RS
- Expected
- Minimum 1.05 A (1 A + 5%), +5%
- Tools
- DVM, current sampling resistor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive heat caution
Power supply generates abnormally high temperatures during operation
Ranked causes
- 1. Preregulator control circuit fault (common)
- 2. Q3, Q4, or Q5 shorted (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overload condition detected during load operation caution
OVERLOAD LED illuminates during V1 or V2 operation when load current exceeds preset current limit (minimum 1 A + 5%)
Ranked causes
- 1. Load current exceeds preset current limit setting (common)
- 2. Short circuit or near-short load impedance (occasional)
Monitor OVERLOAD LED during load operation
- Test point
- OVERLOAD LED indicator
- Expected
- OVERLOAD LED OFF during normal operation with load current below 1.0 A + 5% limit
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor load and line regulation
Output voltage varies excessively with changes in load or input line voltage
Ranked causes
- 1. Bias or reference voltage out of spec (common)
- 2. Main rectifiers or filters open (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zero output voltage on V1 or V2
Supply output is at zero volts and does not respond to adjustment
Ranked causes
- 1. AC line fuse blown (common)
- 2. Bias or reference voltage out of spec (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ammeter accuracy check required
Front panel ammeter reading must match calculated current from sampling resistor voltage drop within tolerance
Ranked causes
- 1. Front panel ammeter calibration drift (unknown)
Measure voltage drop across current sampling resistor RS
- Test point
- Across RS in series with load
- Expected
- Current 0.993 to 1.007 A (±0.5% + 2 counts), ±(0.5% + 2 counts)
- Tools
- DVM, current sampling resistor (100 mΩ per Table A-1)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Common mode current measurement
Measure AC common mode current between output and chassis ground at V1 supply
Ranked causes
- 1. Common mode current circuit evaluation (not a failure symptom) (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive noise or regulation degradation with pulse load
Peak current during pulse-type loading exceeds current limit, causing supply crossover from constant-voltage to current-limit mode and performance degradation
Ranked causes
- 1. Peak current in pulse load exceeds preset current limit (occasional)
Measure peak output current during pulse load operation
- Test point
- Output terminals with pulse load applied
- Expected
- Peak current remains below 1.0 A + 5% during pulse duration; average current below limit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line regulation check required high voltage
Verify output voltage stability when AC input varies from -10% to +10% of nominal line voltage
Ranked causes
- 1. Line regulation circuit drift or transformer coupling degradation (unknown)
Set low line input (-10% nominal) and record output voltage
- Test point
- V1 output terminals
- Expected
- 25.0 V at low line
- Tools
- Variable autotransformer, electronic load, DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Load delivers reverse current to supply during operation caution
Active load connected to supply pumps current back into supply during portion of its operating cycle, risking loss of voltage regulation and damage to output capacitor
Ranked causes
- 1. Active load circuit topology returns current to supply (occasional)
Monitor output voltage during active load operation
- Test point
- V1 or V2 output voltage
- Expected
- Output voltage stable and regulated at set value throughout active load cycle; no voltage dips or transients
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Load regulation check required
Verify output voltage stability when load changes from maximum current to no load (open circuit)
Ranked causes
- 1. Load regulation circuit drift or component degradation (unknown)
Record output voltage at full load (1.000 A)
- Test point
- V1 output terminals
- Expected
- 25.0 V
- Tools
- Electronic load (constant current mode), DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Open fuse high voltage
AC line fuse F1 is open or blown
Ranked causes
- 1. Transient or fatigue-induced failure (common)
Visually inspect fuse F1
- Test point
- F1 in power inlet
- Expected
- Filament intact and not discolored
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage clamped
Output voltage does not respond to adjustment or remains fixed at a clamped level
Ranked causes
- 1. Preregulator control circuit fault (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage higher or lower than rating
Supply output voltage deviates from nominal rating
Ranked causes
- 1. Bias or reference voltage out of spec (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Peak-to-peak ripple and noise measurement
Verify peak-to-peak noise on output is within specification using oscilloscope in AC mode
Ranked causes
- 1. Output filter degradation or high-frequency noise injection in supply (unknown)
Measure peak-to-peak noise on V1 output at 25 V with oscilloscope AC mode, 20 MHz BW
- Test point
- V1 +/- terminals
- Expected
- Less than 1.5 mV peak-to-peak
- Tools
- Oscilloscope (20 MHz bandwidth), AC coupling
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse voltage applied across output terminals high voltage
External voltage source or load connected with reverse polarity, threatening output capacitors and series regulator transistors
Ranked causes
- 1. Load or external source connected with reversed output polarity (rare)
Inspect output terminal connections for correct polarity
- Test point
- V1 +/- terminals and V2 +/- terminals
- Expected
- Load connected +V to +V terminal, -V to -V terminal; no reverse connections
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RMS ripple and noise measurement
Verify RMS noise voltage on output is within specification using true RMS voltmeter
Ranked causes
- 1. Output filter degradation or noise injection in supply circuit (unknown)
Measure RMS noise voltage on V1 output at 25 V
- Test point
- V1 +/- terminals
- Expected
- Less than 0.35 mV RMS
- Tools
- True RMS voltmeter, bandwidth 20 Hz to 20 MHz
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltmeter accuracy check required
Front panel voltmeter reading must match external DVM measurement within specified tolerance
Ranked causes
- 1. Front panel voltmeter calibration drift (unknown)
Measure output voltage with external DVM at 17 V setting
- Test point
- V1 +/- terminals
- Expected
- 16.90 to 17.10 V, ±(0.5% + 2 counts)
- Tools
- DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Will not current limit
Supply does not enter current limiting mode when output current exceeds limit
Ranked causes
- 1. Open OR-gate diodes or defective current limit amplifiers (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.