Agilent E3614A: troubleshooting and error codes
The ServiceForge knowledge base holds 41 documented symptoms for the Agilent E3614A, with 78 ranked causes, 79 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Low output voltage
Output voltage is lower than expected or cannot reach set value.
Ranked causes
- 1. Q1 or Q4 open (occasional)
- 2. Q9 shorted (occasional)
- 3. Constant current comparator circuit defective (occasional)
- 4. U9A defective (occasional)
- 5. U10 and U9D defective or R85 open (occasional)
- 6. U9B defective (occasional)
- 7. U9C defective (occasional)
Disconnect Q9 emitter; check output voltage
- Test point
- Q9 emitter
- Expected
- Output voltage increases if Q1/Q4 are functional
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preregulator control circuit troubleshooting
Systematic troubleshooting of preregulator and control circuit for E3614A, E3615A, E3616A, E3617A models.
Ranked causes
- 1. CR18, CR15, Q7, Q8 defective (E3614A/E3615A) (occasional)
- 2. CR15, CR18, U2, U21 defective (E3616A) (occasional)
- 3. U3 defective (occasional)
- 4. U4 defective (occasional)
- 5. U5 defective (occasional)
- 6. U6 defective (occasional)
- 7. CR10, CR12, Q5, Q6 defective (occasional)
Set output voltage to model-specific value
- Test point
- Output terminals
- Expected
- E3614A: 4.5 V ±0.5 V; E3615A: 10 V ±1 V; E3616A: 15 V ±1 V; E3617A: 26 V ±5 V
10 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High ripple on output
Output AC ripple exceeds specification or is visibly noisy
Ranked causes
- 1. Ground loop in test setup (common)
- 2. Output not properly grounded (occasional)
- 3. Supply operating in constant current mode (occasional)
- 4. Low voltage across filter capacitor or regulator transistor (occasional)
- 5. Excessive ripple on reference voltages (occasional)
Inspect test setup for ground loops
- Test point
- Test setup connections
- Expected
- No ground loops present
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or no output voltage
Output voltage is at or near zero, or lower than commanded
Ranked causes
- 1. Line fuse open (common)
- 2. Series regulator feedback loop failure (common)
- 3. Preregulator feedback loop failure (occasional)
- 4. CR20 shorted (occasional)
Verify input power available and check line fuse
- Test point
- Rear panel line fuse
- Expected
- Fuse intact, proper rating for line voltage
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor stability in constant current mode
Output current is unstable or drifts in constant current operation mode.
Ranked causes
- 1. +10 V reference voltage unstable (common)
- 2. CR24, CR25, CR29, CR30 leaky (occasional)
- 3. U9 and U10 defective (occasional)
- 4. Noisy programming resistor R85 (rare)
Measure +10 V reference voltage
- Test point
- Reference voltage
- Expected
- +10 V stable
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor stability in constant voltage mode
Output voltage is unstable or drifts in constant voltage operation mode.
Ranked causes
- 1. +10 V reference voltage unstable (common)
- 2. CR27, CR28, CR23, CR26 leaky (occasional)
- 3. U9 defective (occasional)
- 4. Noisy programming resistor R83 (rare)
Measure +10 V reference voltage
- Test point
- Reference voltage
- Expected
- +10 V stable
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High output voltage
Output voltage remains high or does not respond to control adjustments.
Ranked causes
- 1. Q1 or Q4 shorted (occasional)
- 2. Q9 open (occasional)
- 3. U9B defective (occasional)
Short Q9 emitter to collector; check output voltage
- Test point
- Q9 emitter and collector
- Expected
- Output voltage decreases if Q9 is functional V
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor load regulation in constant current mode
Output current does not maintain stability under varying load conditions in constant current operation.
Ranked causes
- 1. +10 V reference voltage low or unstable (common)
- 2. CR1, CR19, CR20, C2, C31 leaky (occasional)
- 3. Supply crossing over to constant voltage operation (common)
Measure +10 V reference voltage
- Test point
- Reference voltage point
- Expected
- +10 V
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ammeter and CC set high voltage
Front panel AMPS display reading does not match DVM value divided by Rs, or CC set does not respond correctly
Ranked causes
- 1. Ammeter display out of calibration (common)
- 2. CC set circuit out of calibration (common)
Verify AMPS display reads correct value
- Test point
- Front panel AMPS display
- Expected
- DVM value divided by Rs
- Tools
- DVM, test setup Figure A-9
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Voltmeter and OVP set high voltage
Front panel VOLTS display does not match DVM value, or OVP set does not trip at correct voltage
Ranked causes
- 1. Voltmeter display out of calibration (common)
- 2. OVP set circuit out of calibration (common)
Measure output voltage on DVM
- Test point
- Output terminals
- Expected
- DVM value
- Tools
- DVM
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive heat generation caution
Power supply generates excessive heat during normal operation.
Ranked causes
- 1. Preregulator control circuit malfunction (common)
- 2. CR10, CR12, CR15, CR18 shorted (occasional)
Test CR10, CR12, CR15, CR18 for short circuits
- Test point
- Diode leads
- Expected
- No short
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVP (Overvoltage Protection) circuit shutdown
Overvoltage protection circuit activates, shutting down the supply or limiting output.
Ranked causes
- 1. OVP Adjust screw not fully clockwise (common)
- 2. Overvoltage protection circuit defective (occasional)
Verify OVP Adjust screw position
- Test point
- Front panel OVP Adjust control
- Expected
- Fully clockwise
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVP shutdown continues after reset
OVP LED remains lit or shutdown recurs repeatedly after power cycle
Ranked causes
- 1. Load or sense terminal connections loose or incorrect (common)
- 2. OVP limit setting incorrect for load (common)
Inspect load and sense terminal connections
- Test point
- Load and sense terminals at rear panel
- Expected
- All connections secure, correct polarity, no damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor line regulation
Output voltage changes more than specified when input line voltage changes
Ranked causes
- 1. Reference voltage out of specification (common)
- 2. U9 regulator circuit failure (common)
Measure +10V reference voltage
- Test point
- Reference voltage output (per Table A-2)
- Expected
- +10.5 +/- 0.2 Vdc, ripple 2 mV
- Tools
- DVM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor load regulation in constant voltage mode
Output voltage does not maintain stability under varying load conditions in constant voltage operation.
Ranked causes
- 1. +10 V reference voltage low or unstable (common)
- 2. Supply entering current limit operation (common)
Measure +10 V reference voltage
- Test point
- Reference voltage point
- Expected
- +10 V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor load voltage regulation when remote sensing used
Output voltage drops more than specified when load resistance or current changes in remote sensing mode
Ranked causes
- 1. Sense lead resistance exceeds 0.5 ohms per lead (common)
- 2. Noise picked up on sense leads (occasional)
Measure sense lead resistance
- Test point
- +S and -S terminals to load
- Expected
- 0.5 ohms per lead or less
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Added external capacitance reduces current limiting protection caution
Externally added output capacitance improves pulse current capability but decreases safety from current limiting circuit, allowing high-current pulses to damage load components before current limit operates.
Ranked causes
- 1. External capacitance added without considering impact on current limiting (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
False OVP shutdowns occurring
OVP LED lights unexpectedly, output voltage and current drop to zero during normal operation
Ranked causes
- 1. OVP shutdown voltage set too close to operating voltage (common)
Verify OVP shutdown voltage setting relative to operating voltage
- Test point
- OVP Adjust control
- Expected
- 4% of output voltage +2.0 V or more above operating voltage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect line voltage configuration high voltage
Rear panel label and fuse setting do not match the local input line voltage
Ranked causes
- 1. Supply set for wrong line voltage (common)
Verify rear-panel label matches input line voltage
- Test point
- Rear panel label
- Expected
- Label indicates correct line voltage for your location
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current line regulation out of spec caution
CC line regulation exceeds 0.01% plus 250 µA when AC input voltage varies ±10% from nominal
Ranked causes
- 1. AC input voltage not varied through full range (common)
Connect DVM across current monitoring resistor Rs
- Test point
- across Rs
- Expected
- DVM across Rs for current measurement
- Tools
- Digital voltmeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current load regulation out of spec
CC load regulation exceeds 0.01% plus 250 µA when load changes from short circuit to full load
Ranked causes
- 1. Electronic load not set to full rated voltage (common)
Connect DVM across current monitoring resistor Rs
- Test point
- across Rs
- Expected
- DVM across Rs for current measurement
- Tools
- Digital voltmeter
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current setting does not change with CURRENT control high voltage
Rotating the CURRENT control fully counter-clockwise and clockwise does not adjust the current control set value
Ranked causes
- 1. CURRENT control not functioning (unknown)
While pressing OVP/CC SET switch, rotate CURRENT control fully counter-clockwise then clockwise
- Test point
- AMPS display
- Expected
- Display changes with control rotation
Output impedance and voltage recovery affected by capacitance
Internal and external output capacitance affects output impedance frequency response, voltage recovery time during load changes, and surge current during rapid load resistance reduction.
Ranked causes
- 1. Output capacitance design during constant current operation (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output instability or oscillations in remote sensing mode
Output voltage oscillates or exhibits poor transient response when supply configured for remote sensing
Ranked causes
- 1. Load lead inductance and capacitance forming filter in feedback loop (occasional)
Verify load lead length and routing
- Test point
- Load leads from +/- output terminals to load
- Expected
- Leads as short as possible, twisted together
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output oscillates in constant current mode
Output current exhibits oscillation or instability when operating in constant current mode.
Ranked causes
- 1. C31 and C33 defective in constant current circuit (common)
Test C31 and C33 for defects
- Test point
- C31 and C33 capacitors
- Expected
- Normal operation, no oscillation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output oscillates in constant voltage mode
Output voltage exhibits oscillation or instability when operating in constant voltage mode.
Ranked causes
- 1. C29 and C36 defective in constant voltage circuit (common)
Test C29 and C36 for defects
- Test point
- C29 and C36 capacitors
- Expected
- Normal operation, no oscillation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage does not change with VOLTAGE control high voltage
Rotating the VOLTAGE control fully counter-clockwise and clockwise does not decrease or increase output voltage to 0 Vdc and maximum respectively
Ranked causes
- 1. VOLTAGE control not functioning (unknown)
Rotate VOLTAGE control fully counter-clockwise
- Test point
- VOLTS display
- Expected
- 0 Vdc
Output voltage drift exceeds 0.1% plus 5 mV over 8 hours
CV drift measurement shows output voltage change exceeding 0.1% plus 5 mV during 8-hour test after 30-minute warmup
Ranked causes
- 1. Insufficient warmup period or unstable ambient conditions (occasional)
Connect DVM across current monitoring resistor Rs
- Test point
- across Rs in test setup
- Expected
- DVM across Rs as in Figure A-4
- Tools
- Digital voltmeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage line regulation out of spec caution
CV line regulation exceeds 0.01% plus 2 mV when AC input voltage varies ±10% from nominal
Ranked causes
- 1. AC input voltage not varied through full range (common)
Set AC line to minimum voltage via autotransformer
- Test point
- AC mains input
- Expected
- 104 Vac (115V nominal), 90 Vac (100V nominal), or 207 Vac (230V nominal)
- Tools
- Variable autotransformer, AC voltmeter
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage load regulation out of spec
CV load regulation exceeds 0.01% plus 2 mV when load changes from open circuit to full load
Ranked causes
- 1. Electronic load not set to full rated current (common)
Verify output voltage with load at full current
- Test point
- across load resistor
- Expected
- 8 V for E3614A, 20 V for E3615A, 35 V for E3616A, 60 V for E3617A
- Tools
- Digital voltmeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage peak-to-peak ripple exceeds 1 mV
CV PARD peak-to-peak measurement of output noise exceeds 1 mV p-p over 20 Hz to 20 MHz bandwidth
Ranked causes
- 1. Supply not at full rated output voltage in CV mode (common)
Connect oscilloscope in AC mode to output
- Test point
- across output terminals
- Expected
- Oscilloscope AC coupled, bandwidth 20 MHz
- Tools
- Oscilloscope, AC probe
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage RMS ripple and noise exceeds 200 µV
CV PARD RMS measurement of output voltage noise exceeds 200 µV rms over 20 Hz to 20 MHz bandwidth
Ranked causes
- 1. Supply not at full rated output voltage in CV mode (common)
Connect true RMS voltmeter across output
- Test point
- across output terminals
- Expected
- RMS voltmeter connected with BW 20 Hz-20 MHz
- Tools
- True RMS voltmeter, AC coupling
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage transient response exceeds 50 µsec
CV output voltage takes longer than 50 µsec to settle within 15 mV of baseline after load change
Ranked causes
- 1. Electronic load not in transient operation mode (common)
Replace oscilloscope for DVM in test setup
- Test point
- across output
- Expected
- AC coupled oscilloscope connected to output
- Tools
- Oscilloscope, AC coupling probe
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage shutdown activated caution
OVP LED indicator is illuminated, indicating output has been shut down by overvoltage condition
Ranked causes
- 1. Overvoltage condition occurred (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVP crowbar SCR damage during battery charging caution
When OVP is inadvertently triggered during battery charging, the crowbar SCR shorts the output and continuously sinks large current from the battery, damaging the supply.
Ranked causes
- 1. OVP crowbar SCR triggered without series output diode protection during battery charging (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVP shutdown setting is below minimum threshold caution
OVP shutdown voltage setting is not above the minimum required value for the model
Ranked causes
- 1. OVP shutdown incorrectly set (common)
While pressing OVP/CC SET switch, verify OVP shutdown setting
- Test point
- VOLTS display (showing OVP setting)
- Expected
- Above 8.0 Vdc (E3614A), 20.0 Vdc (E3615A), 35.0 Vdc (E3616A), or 60.0 Vdc (E3617A)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parallel operation damage from unequal supply turn-on timing caution
In parallel or auto-parallel operation, if one supply is turned on before the other, reverse voltage across the series regulator transistors can cause damage.
Ranked causes
- 1. Missing series regulator protection diode in parallel operation without simultaneous turn-on (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse loading causes unwanted current limit crossover
High peak currents during pulse loading exceed preset current limit, causing automatic crossover from constant-voltage to constant-current operation when not desired.
Ranked causes
- 1. Preset current limit set to average instead of peak requirement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse current from active load causes regulation loss and capacitor damage caution
Active load delivers reverse current to supply during portion of operating cycle, causing loss of output regulation and possible damage to output capacitor.
Ranked causes
- 1. Active load delivers reverse current without dummy load preconditioning (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse voltage from series operation damages supply without protection diode caution
When AC is removed from one supply in series operation, reverse polarity voltage can damage output electrolytic capacitors and series regulator transistors in the unenergized supply.
Ranked causes
- 1. Missing reverse polarity protection diode across output terminals (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage regulation error when remote sensing sense leads open patient safety
Output voltage uncontrolled or excessive when sense lead connection is accidentally disconnected during remote sensing operation
Ranked causes
- 1. Sense lead connection accidentally opened or disconnected (occasional)
Inspect sense lead connections for security
- Test point
- +S and -S rear panel terminals
- Expected
- Connections tight, no movement, no damage to conductors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.