Agilent 6812B: troubleshooting and error codes
The ServiceForge knowledge base holds 4 documented symptoms for the Agilent 6812B, with 4 ranked causes, 12 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Auxiliary Output Option 026 high voltage
Calibration procedure for the Auxiliary Output Option 026 of the Agilent 6812B AC power source. Verification procedures should be performed first to determine if recalibration is needed.
Ranked causes
- 1. Calibration required per maintenance schedule or after verification failure (unknown)
Verify auxiliary output voltage
- Test point
- Auxiliary output terminals
- Expected
- 0-26 V rms at resistive load
- Tools
- Voltmeter
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Auxiliary Output Voltage
Adjust auxiliary output voltage reading to match DVM measurement
Ranked causes
- 1. Auxiliary output voltage requires calibration (unknown)
Enable calibration mode by entering password
- Test point
- Display
- Expected
- CAL ON
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Output phase resolution high voltage
Verification and calibration of output phase resolution for the Auxiliary Output Option 026.
Ranked causes
- 1. Output phase resolution out of tolerance (unknown)
Measure output phase resolution
- Test point
- Auxiliary output terminals relative to main output
- Expected
- 0.1 degree steps
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Output voltage resolution high voltage
Verification and calibration of output voltage resolution for the Auxiliary Output Option 026.
Ranked causes
- 1. Output voltage resolution out of tolerance (unknown)
Measure output voltage resolution
- Test point
- Auxiliary output terminals
- Expected
- 10 mV steps
- Tools
- Voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
CAL DENIED | Calibration function access prevented | Attempt to enter calibration mode when internal switch set to prevent calibration changes | warning |
OUT OF RANGE | Measured or entered value exceeds acceptable range | Entered calibration data value outside permissible limits | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI