Agilent 6812B: troubleshooting and error codes

The ServiceForge knowledge base holds 4 documented symptoms for the Agilent 6812B, with 4 ranked causes, 12 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Auxiliary Output Option 026 high voltage

Calibration procedure for the Auxiliary Output Option 026 of the Agilent 6812B AC power source. Verification procedures should be performed first to determine if recalibration is needed.

Ranked causes

  1. 1. Calibration required per maintenance schedule or after verification failure (unknown)
First check

Verify auxiliary output voltage

Test point
Auxiliary output terminals
Expected
0-26 V rms at resistive load
Tools
Voltmeter

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Auxiliary Output Voltage

Adjust auxiliary output voltage reading to match DVM measurement

Ranked causes

  1. 1. Auxiliary output voltage requires calibration (unknown)
First check

Enable calibration mode by entering password

Test point
Display
Expected
CAL ON

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Output phase resolution high voltage

Verification and calibration of output phase resolution for the Auxiliary Output Option 026.

Ranked causes

  1. 1. Output phase resolution out of tolerance (unknown)
First check

Measure output phase resolution

Test point
Auxiliary output terminals relative to main output
Expected
0.1 degree steps
Tools
Oscilloscope

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Output voltage resolution high voltage

Verification and calibration of output voltage resolution for the Auxiliary Output Option 026.

Ranked causes

  1. 1. Output voltage resolution out of tolerance (unknown)
First check

Measure output voltage resolution

Test point
Auxiliary output terminals
Expected
10 mV steps
Tools
Voltmeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
CAL DENIEDCalibration function access preventedAttempt to enter calibration mode when internal switch set to prevent calibration changeswarning
OUT OF RANGEMeasured or entered value exceeds acceptable rangeEntered calibration data value outside permissible limitserror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI