Agilent 6811B: troubleshooting and error codes
The ServiceForge knowledge base holds 45 documented symptoms for the Agilent 6811B, with 75 ranked causes, 139 diagnostic checks with test points and expected values, and 51 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power-on self-test fails with ERR annunciator
ERR annunciator illuminates on display during power-on self-test sequence
Ranked causes
- 1. Non-volatile RAM checksum error (common)
- 2. RAM selftest failure (occasional)
- 3. DAC selftest error (occasional)
- 4. Voltage selftest error (occasional)
- 5. Current selftest error (occasional)
- 6. Fan voltage failure (occasional)
- 7. Digital I/O selftest error (rare)
Query error code during self-test
- Test point
- *TST? command
- Expected
- Return value zero if all tests pass or error code of first failed test
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inverter assembly produces no output high voltage
A7 inverter assembly not generating proper AC output; no PWM signals or gate/drain waveforms
Ranked causes
- 1. INV_OFF signal low preventing inverter operation (common)
- 2. No PWM signal from control assembly (common)
- 3. Power FET gate or drain waveforms abnormal (common)
- 4. Shorted or broken FETs on inverter assembly (occasional)
Verify INV_OFF signal status
- Test point
- A10U1801-23
- Expected
- INV_OFF signal high
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error or fault
Calibration command rejected or calibration data incorrect
Ranked causes
- 1. CAL switch in incorrect position (common)
- 2. CAL password incorrect (occasional)
- 3. CAL not enabled or calibration constants incorrect (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External shunt line and load connections not secure or missing cable covers high voltage
Current shunt input/output cable connections are loose, incorrectly wired, or missing safety covers, exposing live line potential at barrier block terminals
Ranked causes
- 1. Cable covers not installed on shunt connectors after line connections (common)
- 2. Incorrect line/neutral/ground connection on shunt barrier block (occasional)
- 3. Cable connections to shunt barrier block screws not tight (common)
Verify cable covers installed on shunt connectors
- Test point
- External shunt input and output connectors, ac source input connector
- Expected
- All three hooded cable covers installed and fully covering barrier block screw terminals
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
GPIB/RS-232 communication failure
Unit not responding to GPIB or RS-232 controller commands
Ranked causes
- 1. Missing or low +5 Vdc supply to GPIB board (common)
- 2. Crystal oscillator not functioning (occasional)
- 3. GPIB/RS-232 transceiver fault (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement cable connections not secure or missing hoods high voltage
Cable connections at connector terminals are loose, improperly wired, or missing safety hoods, exposing live line potential
Ranked causes
- 1. Cable hoods not installed after wiring (common)
- 2. Wire connections not firmly tightened (common)
- 3. Incorrect wire gauge or routing in connector (occasional)
Verify cable hoods are installed on all connectors
- Test point
- All connector ends (ac source end, shunt end)
- Expected
- Cable hoods fully covering connector terminals, snapped together
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Analyzer input connector not operating properly high voltage
Connector plug connections not functioning correctly or frequency measurements inaccurate due to missing voltage sense inputs
Ranked causes
- 1. Voltage SENSE inputs not connected (common)
- 2. Incorrect cable routing or connection (occasional)
- 3. Frequency source incorrectly selected or unavailable (occasional)
Verify V SENSE HI and LO connections are made
- Test point
- V SENSE HI and LO terminals on ac source connector
- Expected
- Both HI and LO connections physically present and tight
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal output waveform at A9TP500
Oscilloscope shows incorrect waveform when probed at A9TP500; expected 500mV/div, 5.00us/div
Ranked causes
- 1. U502 and associated circuit components defective (occasional)
- 2. U500 relay 501 or voltage driver resistors defective (occasional)
Observe waveform at A9TP500 with oscilloscope
- Test point
- A9TP500
- Expected
- Normal waveform at 500mV/div, 5.00us/div
- Tools
- Oscilloscope
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication error between interfaces
Data communication failure between control assemblies during operation
Ranked causes
- 1. GPIB/DSP/Trigger assembly communication failure (common)
- 2. Front panel communication failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC Rail Assembly bias voltages incorrect or absent high voltage
Measured bias voltages at test points do not match expected values (+15V, -15V, +5V); backlight or control signals missing.
Ranked causes
- 1. A3 AC FET Assembly failure (common)
- 2. A5 DC Rail board component failure (common)
Disconnect cable from A3 FET Board
- Test point
- J1042 on A4 AC Input Board
- Expected
- Cable disconnected from connector
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FET board test and verification high voltage
Comprehensive troubleshooting of A3 FET board with power FETs Q201-204, Q211, Q222, Q233, Q244 using resistance measurements and signal testing
Ranked causes
- 1. Defective power FET or gate bias circuit (common)
- 2. Open circuit in FET gate input (occasional)
Measure resistance between +Rail and common
- Test point
- E502 to common (-Rail)
- Expected
- Greater than 20 MΩ, >
15 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
GPIB/RS-232 transceiver malfunction
Board fails to communicate via GPIB or RS-232 interface
Ranked causes
- 1. No +5 VDC supply to L100 (common)
- 2. Crystal oscillator Y100 not functioning (occasional)
Verify +5VDC supply voltage
- Test point
- L100
- Expected
- +5 Vdc
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Programming error or command violation
Invalid command sent or command incompatible with current settings
Ranked causes
- 1. List length mismatch or voltage/waveform incompatibility (common)
- 2. Command incompatible with current interface or configuration (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PWM driver output not toggling high voltage
PWM driver output at U491-8 to U491-9 shows no pulse activity or incorrect pulse train; FET switches not being driven.
Ranked causes
- 1. 300 Vdc rail not connected or defective (common)
- 2. PWM driver IC or gate drive circuit failure (occasional)
Verify 300 Vdc rail connection to A3 FET board
- Test point
- Cable from A3 FET to J1042 on A4 AC Input Board
- Expected
- Cable connected and secured
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PWM Enable signal absent or incorrect high voltage
PWM_EN signal at U436-6 not present or not at expected +5V level; prevents PWM control circuit operation.
Ranked causes
- 1. External synchronization signal missing or incorrect (common)
- 2. U423 Bias Detect Circuit failure (occasional)
Verify EXT_CLK_ISO signal present at U436-6
- Test point
- U436-6
- Expected
- WAVEFORM present (83 kHz external clock per Figure 4-5A)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote voltage sense at location other than external shunt not functioning high voltage
Voltage sensing at a point different from the external current shunt is not measuring correctly or connections are not secure
Ranked causes
- 1. Original voltage sense wires not completely removed from supplied cable (common)
- 2. Custom sense cable connections not made at remote location (occasional)
Verify voltage sense wires removed from supplied cable
- Test point
- Agilent supplied measurement cable connectors
- Expected
- No voltage sense wires present at either end of cable
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROM upgrade failure with OUT OF RANGE error
ROMUPD command fails showing OUT OF RANGE error, requiring re-initialization and recalibration. May occur after QA firmware or memory corruption.
Ranked causes
- 1. Previously initialized with QA firmware revision (occasional)
- 2. Non-volatile memory corruption (occasional)
Verify OUT OF RANGE error
- Test point
- Front panel display after ROMUPD command
- Expected
- OUT OF RANGE error message shown
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify DSP and Control Assembly operation high voltage
Check analog voltages and digital signals on A8 DSP and A10 Control boards after commands entered via front panel
Ranked causes
- 1. Incorrect J822 analog voltage levels (unknown)
- 2. Offset voltage not responding to offset command (unknown)
Measure J822 analog voltages with output disabled
- Test point
- J822 pins 1-33
- Expected
- J822-31: -3.8 Vdc, J822-32: -0.8 Vdc, J822-23: +1.0 Vac, J822-22: +5.0 Vdc, J822-15: -4.0 Vdc, J822-20: -15.0 Vdc, J822-13: +6.0 Vdc, J822-18: +15.0 Vdc, J822-9: -2.5 Vdc, J822-16: -1.5 Vdc, J822-5: +3.0 Vac, J822-12: -5.0 Vdc, J822-1: 0 Vac, J822-10: +6.0 Vdc, J822-8: +5.0 Vdc, J822-6: +6.0 Vdc
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both CV and CC annunciators illuminated caution
CV (constant voltage) and CC (constant current) indicator lights both on simultaneously
Ranked causes
- 1. Peak current limit set too low (common)
Check peak current limit setting
- Test point
- Peak current limit control
- Expected
- Limit set above normal operating current
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CAL DENIED message on display during calibration
Display shows CAL DENIED or error code 401 appears when calibrating over GPIB
Ranked causes
- 1. Internal INHIBIT CAL switch has been set (common)
Locate calibration switches on A8 DSP board
- Test point
- A8 DSP board Switches 1 and 2
- Expected
- Switches in Inhibit Cal position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current measurement using 4-terminal shunt high voltage
Connect voltmeter to current-monitoring terminals on 4-terminal current shunt to eliminate measurement error from load lead voltage drops
Ranked causes
- 1. Voltage drop error in load leads during current measurement (common)
Verify voltmeter connected to current-monitoring terminals on 4-terminal shunt
- Test point
- Current shunt terminals
- Expected
- Voltmeter connected to monitoring terminals, not across shunt element
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dual Power Analyzer Option 020/022 high voltage
Verification and calibration of ac source output voltage and current performance
Ranked causes
- 1. Initial verification before calibration required (common)
Assemble equipment: 5-pin connector plug, Digital Voltmeter (Agilent 3458A: 10 nV @ 1 V resolution, 8.5 digits, >20 ppm accuracy), Current Monitor (Guildline 7320/0.01: 0.01 Ω, ±200 ppm), Load Resistor (30 ohm 100 W minimum), Capacitor (1.0 µF 50 V, part 0160-3490), Function Generator (Agilent 33120A: 50 mV, 500 mV, 60 Hz sinewave), GPIB Controller (HP Series 200/300 or equivalent)
- Test point
- Equipment list per Table B-1
- Expected
- All required equipment present and functional
- Tools
- Specified equipment per table
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: External current measurement high voltage
Calibrate the Power Analyzer current sense terminals to measure external function generator signal accurately using an Agilent 3458A DVM as reference
Ranked causes
- 1. Current offset error in sense terminals (common)
Short current sense terminals
- Test point
- I SENSE
- Expected
- Shorted across terminals
10 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: External Current Shunt Setting
Verify and update the external current shunt resistance value in the unit to match the label on the physical shunt
Ranked causes
- 1. Shunt value setting mismatch (common)
Display current shunt value setting
- Test point
- SHUNT command in Calibration menu
- Expected
- Resistance value matches label on physical external current shunt
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: External voltage measurement high voltage
Calibrate the Power Analyzer voltage sense terminals to measure external ac voltage source output accurately using an Agilent 3458A DVM as reference
Ranked causes
- 1. Voltage offset error in sense terminals (common)
Reset unit to known state
- Test point
- Front panel
- Expected
- *RST display
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Initialization
Unit requires initialization to restore non-volatile memory after corruption or first setup. Erases all calibration, waveform, and list data.
Ranked causes
- 1. Non-volatile memory corruption or first-time setup (unknown)
Verify EEINIT command appears on display
- Test point
- Front panel display
- Expected
- EEINIT <model> shown
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Initialize unit
Non-volatile memory corruption or first-time initialization requiring front panel EEINIT command sequence
Ranked causes
- 1. Non-volatile memory corruption or first-time setup (unknown)
Turn unit on
- Test point
- Front panel power switch
- Expected
- Unit powers on
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Output Frequency Modulation
Configure modulating frequency, frequency deviation, and enable frequency modulation on the output
Ranked causes
- 1. Frequency modulation parameters not set (unknown)
Verify modulating frequency setting
- Test point
- FM:FREQ command
- Expected
- Modulating frequency value displayed Hz
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Post-repair adjustment
Supply requires calibration after replacing A8 DSP Assembly, A9 Output Filter/Relay Assembly, or A10 Control Assembly; A8 replacement requires prior initialization
Ranked causes
- 1. Component replacement triggers calibration requirement (common)
Verify initialization completed if A8 DSP Assembly was replaced
- Test point
- A8 DSP Assembly
- Expected
- Initialization completed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ROM upgrade
Update to newer ROM versions without destroying calibration data; identify and upgrade firmware revision
Ranked causes
- 1. Need to upgrade to newer ROM revision (unknown)
Identify firmware revision
- Test point
- Front panel display at power-on or GPIB *IDN? query
- Expected
- Firmware revision displayed for approximately 10 seconds or returned via GPIB
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV error amplifier output incorrect high voltage
CV (constant voltage) error amplifier not responding to RAIL_VMON stimulus; output at U447-7 or U463-3/4 not toggling as expected.
Ranked causes
- 1. CV error amplifier IC failure (occasional)
Apply 0V stimulus to RAIL_VMON
- Test point
- RAIL_VMON input
- Expected
- CV error amp responds; RAIL_OV_INH and RAIL_INH_ISO show expected pulses per test patterns
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Discrete Fault Indicator (DFI) not functioning
J113-1 does not respond to DFI commands or does not go low
Ranked causes
- 1. DFI signal path or control logic failure (unknown)
Connect +5 Vdc to DFI terminals
- Test point
- DFI terminals via 5K resistor
- Expected
- +5 Vdc applied
- Tools
- 5K resistor, power supply
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Initialization fails with OUT OF RANGE error
OUT OF RANGE error displayed during ROM upgrade, requiring re-initialization and re-calibration
Ranked causes
- 1. Instrument previously initialized with QA firmware revision or non-volatile memory corrupted (occasional)
Check for OUT OF RANGE error on front panel display
- Test point
- Front panel display
- Expected
- OUT OF RANGE error shown
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inverter assembly fails under loaded conditions high voltage
A7 passes static checks but fails when 500 VDC rail applied or under current load
Ranked causes
- 1. High voltage breakdown on FETs or capacitors (occasional)
Perform initial A7 inverter troubleshooting
- Test point
- Overall A7 assembly
- Expected
- All waveforms OK per Fig 4-7
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low rail detect circuit not functioning high voltage
Low rail warning or low rail final inhibit signals not toggling at expected thresholds when RAIL_VMON varied.
Ranked causes
- 1. Low rail detect comparator failure (occasional)
Measure RAIL_LOW_WARN threshold
- Test point
- U424-14 output
- Expected
- Toggles at approximately 10.6V RAIL_VMON
- Tools
- DMM or oscilloscope
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No communications pulses
Waveforms at communication test points never go high or low
Ranked causes
- 1. Communications circuit failure on A11 Trigger/BNC Board (unknown)
Check communications pulses
- Test point
- U150-1, U151-5, U153-3, U154-1
- Expected
- Similar waveforms that go high and low
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No trigger input recognition
Trigger output pulse not generated when Trig In BNC is shorted
Ranked causes
- 1. External trigger input path failure (unknown)
Set external trigger mode
- Test point
- Front panel
- Expected
- External trigger mode enabled
- Tools
- Power supply
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No trigger output pulse
Trigger Out BNC does not produce pulse when SHIFT TRIGGER command is entered
Ranked causes
- 1. Signal path failure from A8 DSP board to Trigger Out (unknown)
Set up trigger output
- Test point
- Trig Out BNC
- Expected
- Trigger pulse output
- Tools
- Oscilloscope
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Analyzer input measurement setup
Select Power Analyzer input as measurement source and configure frequency measurement source
Ranked causes
- 1. Measurement source not configured correctly (occasional)
Verify Power Analyzer input selection
- Test point
- Front panel display phase indicator
- Expected
- φ2 displayed to indicate Power Analyzer input active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rail overvoltage detect circuit not responding high voltage
RAIL_OV_INH (rail overvoltage inhibit) signal not toggling at expected threshold; overvoltage protection not triggering.
Ranked causes
- 1. Rail overvoltage comparator failure (occasional)
Measure RAIL_OV_INH response at U463-3
- Test point
- U463-3 output
- Expected
- Signal LOW (≈0V) when RAIL_VMON indicates overvoltage condition; HIGH when normal
- Tools
- DMM or oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ramp generator signal incorrect or missing high voltage
Ramp waveform at ramp generator test points not present or does not match expected sawtooth pattern.
Ranked causes
- 1. Ramp generator IC failure or missing input signal (common)
Measure ramp waveform at U475-7
- Test point
- U475-7 output
- Expected
- Sawtooth waveform at ~83 kHz per Figure 4-5D
- Tools
- Oscilloscope (2.00 V/div, 5.00 us/div)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote Inhibit (RI) not disabling output high voltage
Output does not go to 0 Vac when RI terminals are shorted
Ranked causes
- 1. RI control signal path failure (unknown)
Set 100 Vac at output
- Test point
- Output terminals
- Expected
- 100 Vac
- Tools
- Voltmeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit safety hazard from exposed screw terminals during normal operation high voltage
Connector and barrier block screw terminals remain exposed to contact with personnel during equipment operation, creating electrical shock and arc hazard
Ranked causes
- 1. Cable hoods or covers not installed or removed during operation (common)
Verify all cable hoods and covers are installed before powering unit
- Test point
- All Power Analyzer and external shunt connectors and barrier blocks
- Expected
- Cable hoods and covers fully installed, covering all exposed screw terminals
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify AC output voltage detection high voltage
Check ALC RMS voltage detection at J822 when cable to J627 connected and ac source can output
Ranked causes
- 1. ALC RMS detection not generating expected signals (unknown)
Measure J822-1 AC voltage during ALC RMS detection
- Test point
- J822-1
- Expected
- 3 Vac
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify current limit response with output short circuit high voltage
Check current limit programming response when output is shorted
Ranked causes
- 1. Current limit not responding correctly to short circuit (unknown)
Measure J822-3 with output shorted at 5V 4A limit
- Test point
- J822-3
- Expected
- -0.75 Vac
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Non-volatile RAM RD0 section checksum failed | Power-on self-test detects RAM error | error |
10 | RAM selftest failed | Power-on self-test detects RAM error | error |
11-31 | DAC selftest error, expected <n>, read <reading> | Power-on self-test detects DAC output mismatch | error |
2 | Non-volatile RAM CONFIG section checksum failed | Power-on self-test detects RAM error | error |
200 | Outgrd not responding | Runtime communication check fails | error |
201 | Front panel not responding | Runtime communication check fails | error |
210 | Ingrd receiver framing error | Runtime communication check fails | warning |
211 | Ingrd uart overrun status | Runtime communication check fails | warning |
212 | Ingrd received bad token | Runtime communication check fails | warning |
213 | Ingrd receiver buffer overrun | Runtime communication check fails | warning |
214 | Ingrd input buffer overrun | Runtime communication check fails | warning |
215 | Outgrd output buffer overrun | Runtime communication check fails | warning |
216 | RS-232 receiver framing error | Runtime RS-232 communication fails | warning |
217 | RS-232 receiver parity error | Runtime RS-232 communication fails | warning |
218 | RS-232 receiver overrun error | Runtime RS-232 communication fails | warning |
219 | Ingrd inbuf count sync error | Runtime communication check fails | warning |
220 | Front panel uart overrun | Runtime communication check fails | warning |
221 | Front panel uart framing | Runtime communication check fails | warning |
222 | Front panel uart parity | Runtime communication check fails | warning |
223 | Front panel buffer overrun | Runtime communication check fails | warning |
224 | Front panel timeout | Runtime communication check fails | warning |
3 | Non-volatile RAM CAL section checksum failed | Power-on self-test detects RAM error | error |
4 | Non-volatile RAM WAVEFORM section checksum failed | Power-on self-test detects RAM error | error |
40 | Voltage selftest error | Power-on self-test voltage output incorrect | error |
401 | Calibration command blocked due to INHIBIT CAL switch setting | Calibration command sent when INHIBIT CAL switch is active | error |
401 | CAL switch prevents calibration | Calibration command executed with CAL switch in wrong position | error |
402 | CAL password is incorrect | Calibration command executed with wrong password | error |
403 | CAL not enabled | Calibration command executed without enabling calibration mode | error |
404 | Computed readback cal constants are incorrect | Calibration process detects computed values out of range | error |
405 | Computed programming cal constants are incorrect | Calibration process detects computed values out of range | error |
406 | Incorrect sequence of calibration commands | Calibration commands executed out of order | error |
43 | Current selftest error | Power-on self-test current output incorrect | error |
5 | Non-volatile RAM STATE section checksum failed | Power-on self-test detects RAM error | error |
6 | Non-volatile RAM LIST section checksum failed | Power-on self-test detects RAM error | error |
600 | Systems in mode:list have different list lengths | Multiple channels configured with unequal list lengths | warning |
601 | Requested voltage and waveform exceeds peak voltage capability | Command requests voltage/waveform combination beyond rating | error |
602 | Requested voltage and waveform exceeds transformer volt-second rating | Command requests voltage/waveform combination exceeding transformer rating | error |
603 | Command only applies to RS-232 interface | Non-RS-232 interface executes RS-232-only command | warning |
604 | Trigger received before requested number of pre-trigger readings | Trigger arrives before pre-trigger buffer fills | warning |
605 | Requested RMS current too high for voltage range | Command requests current beyond voltage range capability | error |
606 | Waveform data not defined | Command references undefined waveform | error |
607 | VOLT, VOLT:SLEW, and FUNC:SHAP modes incompatible | Incompatible mode commands combined | error |
608 | Measurement overrange | Measurement exceeds input range of ADC | warning |
609 | Output buffer overrun | Output data rate exceeds buffer capacity | warning |
610 | Command cannot be given with present SYST:CONF setting | Command incompatible with current system configuration | error |
7 | Non-volatile RAM RST section checksum failed | Power-on self-test detects RAM error | error |
70 | Fan voltage failure | Power-on self-test detects fan voltage out of range | warning |
80 | Digital I/O selftest error | Power-on self-test detects digital I/O failure | error |
CAL DENIED | Calibration is locked and cannot be modified | Attempting to enter calibration mode when internal switch prevents calibration changes | error |
Error 613 | Frequency deviation exceeds maximum allowed value | Frequency deviation set to value greater than or equal to programmed output frequency, or within 0.001 Hz margin | error |
OUT OF RANGE | Entered calibration password is incorrect | Pressing Enter after entering wrong CAL ON password | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI