Agilent 6811B: troubleshooting and error codes

The ServiceForge knowledge base holds 45 documented symptoms for the Agilent 6811B, with 75 ranked causes, 139 diagnostic checks with test points and expected values, and 51 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Power-on self-test fails with ERR annunciator

ERR annunciator illuminates on display during power-on self-test sequence

Ranked causes

  1. 1. Non-volatile RAM checksum error (common)
  2. 2. RAM selftest failure (occasional)
  3. 3. DAC selftest error (occasional)
  4. 4. Voltage selftest error (occasional)
  5. 5. Current selftest error (occasional)
  6. 6. Fan voltage failure (occasional)
  7. 7. Digital I/O selftest error (rare)
First check

Query error code during self-test

Test point
*TST? command
Expected
Return value zero if all tests pass or error code of first failed test

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inverter assembly produces no output high voltage

A7 inverter assembly not generating proper AC output; no PWM signals or gate/drain waveforms

Ranked causes

  1. 1. INV_OFF signal low preventing inverter operation (common)
  2. 2. No PWM signal from control assembly (common)
  3. 3. Power FET gate or drain waveforms abnormal (common)
  4. 4. Shorted or broken FETs on inverter assembly (occasional)
First check

Verify INV_OFF signal status

Test point
A10U1801-23
Expected
INV_OFF signal high

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error or fault

Calibration command rejected or calibration data incorrect

Ranked causes

  1. 1. CAL switch in incorrect position (common)
  2. 2. CAL password incorrect (occasional)
  3. 3. CAL not enabled or calibration constants incorrect (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

External shunt line and load connections not secure or missing cable covers high voltage

Current shunt input/output cable connections are loose, incorrectly wired, or missing safety covers, exposing live line potential at barrier block terminals

Ranked causes

  1. 1. Cable covers not installed on shunt connectors after line connections (common)
  2. 2. Incorrect line/neutral/ground connection on shunt barrier block (occasional)
  3. 3. Cable connections to shunt barrier block screws not tight (common)
First check

Verify cable covers installed on shunt connectors

Test point
External shunt input and output connectors, ac source input connector
Expected
All three hooded cable covers installed and fully covering barrier block screw terminals

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

GPIB/RS-232 communication failure

Unit not responding to GPIB or RS-232 controller commands

Ranked causes

  1. 1. Missing or low +5 Vdc supply to GPIB board (common)
  2. 2. Crystal oscillator not functioning (occasional)
  3. 3. GPIB/RS-232 transceiver fault (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement cable connections not secure or missing hoods high voltage

Cable connections at connector terminals are loose, improperly wired, or missing safety hoods, exposing live line potential

Ranked causes

  1. 1. Cable hoods not installed after wiring (common)
  2. 2. Wire connections not firmly tightened (common)
  3. 3. Incorrect wire gauge or routing in connector (occasional)
First check

Verify cable hoods are installed on all connectors

Test point
All connector ends (ac source end, shunt end)
Expected
Cable hoods fully covering connector terminals, snapped together

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Analyzer input connector not operating properly high voltage

Connector plug connections not functioning correctly or frequency measurements inaccurate due to missing voltage sense inputs

Ranked causes

  1. 1. Voltage SENSE inputs not connected (common)
  2. 2. Incorrect cable routing or connection (occasional)
  3. 3. Frequency source incorrectly selected or unavailable (occasional)
First check

Verify V SENSE HI and LO connections are made

Test point
V SENSE HI and LO terminals on ac source connector
Expected
Both HI and LO connections physically present and tight

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormal output waveform at A9TP500

Oscilloscope shows incorrect waveform when probed at A9TP500; expected 500mV/div, 5.00us/div

Ranked causes

  1. 1. U502 and associated circuit components defective (occasional)
  2. 2. U500 relay 501 or voltage driver resistors defective (occasional)
First check

Observe waveform at A9TP500 with oscilloscope

Test point
A9TP500
Expected
Normal waveform at 500mV/div, 5.00us/div
Tools
Oscilloscope

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication error between interfaces

Data communication failure between control assemblies during operation

Ranked causes

  1. 1. GPIB/DSP/Trigger assembly communication failure (common)
  2. 2. Front panel communication failure (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DC Rail Assembly bias voltages incorrect or absent high voltage

Measured bias voltages at test points do not match expected values (+15V, -15V, +5V); backlight or control signals missing.

Ranked causes

  1. 1. A3 AC FET Assembly failure (common)
  2. 2. A5 DC Rail board component failure (common)
First check

Disconnect cable from A3 FET Board

Test point
J1042 on A4 AC Input Board
Expected
Cable disconnected from connector

7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

FET board test and verification high voltage

Comprehensive troubleshooting of A3 FET board with power FETs Q201-204, Q211, Q222, Q233, Q244 using resistance measurements and signal testing

Ranked causes

  1. 1. Defective power FET or gate bias circuit (common)
  2. 2. Open circuit in FET gate input (occasional)
First check

Measure resistance between +Rail and common

Test point
E502 to common (-Rail)
Expected
Greater than 20 MΩ, >

15 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

GPIB/RS-232 transceiver malfunction

Board fails to communicate via GPIB or RS-232 interface

Ranked causes

  1. 1. No +5 VDC supply to L100 (common)
  2. 2. Crystal oscillator Y100 not functioning (occasional)
First check

Verify +5VDC supply voltage

Test point
L100
Expected
+5 Vdc
Tools
DMM

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Programming error or command violation

Invalid command sent or command incompatible with current settings

Ranked causes

  1. 1. List length mismatch or voltage/waveform incompatibility (common)
  2. 2. Command incompatible with current interface or configuration (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PWM driver output not toggling high voltage

PWM driver output at U491-8 to U491-9 shows no pulse activity or incorrect pulse train; FET switches not being driven.

Ranked causes

  1. 1. 300 Vdc rail not connected or defective (common)
  2. 2. PWM driver IC or gate drive circuit failure (occasional)
First check

Verify 300 Vdc rail connection to A3 FET board

Test point
Cable from A3 FET to J1042 on A4 AC Input Board
Expected
Cable connected and secured

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PWM Enable signal absent or incorrect high voltage

PWM_EN signal at U436-6 not present or not at expected +5V level; prevents PWM control circuit operation.

Ranked causes

  1. 1. External synchronization signal missing or incorrect (common)
  2. 2. U423 Bias Detect Circuit failure (occasional)
First check

Verify EXT_CLK_ISO signal present at U436-6

Test point
U436-6
Expected
WAVEFORM present (83 kHz external clock per Figure 4-5A)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Remote voltage sense at location other than external shunt not functioning high voltage

Voltage sensing at a point different from the external current shunt is not measuring correctly or connections are not secure

Ranked causes

  1. 1. Original voltage sense wires not completely removed from supplied cable (common)
  2. 2. Custom sense cable connections not made at remote location (occasional)
First check

Verify voltage sense wires removed from supplied cable

Test point
Agilent supplied measurement cable connectors
Expected
No voltage sense wires present at either end of cable

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ROM upgrade failure with OUT OF RANGE error

ROMUPD command fails showing OUT OF RANGE error, requiring re-initialization and recalibration. May occur after QA firmware or memory corruption.

Ranked causes

  1. 1. Previously initialized with QA firmware revision (occasional)
  2. 2. Non-volatile memory corruption (occasional)
First check

Verify OUT OF RANGE error

Test point
Front panel display after ROMUPD command
Expected
OUT OF RANGE error message shown

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verify DSP and Control Assembly operation high voltage

Check analog voltages and digital signals on A8 DSP and A10 Control boards after commands entered via front panel

Ranked causes

  1. 1. Incorrect J822 analog voltage levels (unknown)
  2. 2. Offset voltage not responding to offset command (unknown)
First check

Measure J822 analog voltages with output disabled

Test point
J822 pins 1-33
Expected
J822-31: -3.8 Vdc, J822-32: -0.8 Vdc, J822-23: +1.0 Vac, J822-22: +5.0 Vdc, J822-15: -4.0 Vdc, J822-20: -15.0 Vdc, J822-13: +6.0 Vdc, J822-18: +15.0 Vdc, J822-9: -2.5 Vdc, J822-16: -1.5 Vdc, J822-5: +3.0 Vac, J822-12: -5.0 Vdc, J822-1: 0 Vac, J822-10: +6.0 Vdc, J822-8: +5.0 Vdc, J822-6: +6.0 Vdc

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Both CV and CC annunciators illuminated caution

CV (constant voltage) and CC (constant current) indicator lights both on simultaneously

Ranked causes

  1. 1. Peak current limit set too low (common)
First check

Check peak current limit setting

Test point
Peak current limit control
Expected
Limit set above normal operating current

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CAL DENIED message on display during calibration

Display shows CAL DENIED or error code 401 appears when calibrating over GPIB

Ranked causes

  1. 1. Internal INHIBIT CAL switch has been set (common)
First check

Locate calibration switches on A8 DSP board

Test point
A8 DSP board Switches 1 and 2
Expected
Switches in Inhibit Cal position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Current measurement using 4-terminal shunt high voltage

Connect voltmeter to current-monitoring terminals on 4-terminal current shunt to eliminate measurement error from load lead voltage drops

Ranked causes

  1. 1. Voltage drop error in load leads during current measurement (common)
First check

Verify voltmeter connected to current-monitoring terminals on 4-terminal shunt

Test point
Current shunt terminals
Expected
Voltmeter connected to monitoring terminals, not across shunt element

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Dual Power Analyzer Option 020/022 high voltage

Verification and calibration of ac source output voltage and current performance

Ranked causes

  1. 1. Initial verification before calibration required (common)
First check

Assemble equipment: 5-pin connector plug, Digital Voltmeter (Agilent 3458A: 10 nV @ 1 V resolution, 8.5 digits, >20 ppm accuracy), Current Monitor (Guildline 7320/0.01: 0.01 Ω, ±200 ppm), Load Resistor (30 ohm 100 W minimum), Capacitor (1.0 µF 50 V, part 0160-3490), Function Generator (Agilent 33120A: 50 mV, 500 mV, 60 Hz sinewave), GPIB Controller (HP Series 200/300 or equivalent)

Test point
Equipment list per Table B-1
Expected
All required equipment present and functional
Tools
Specified equipment per table

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: External current measurement high voltage

Calibrate the Power Analyzer current sense terminals to measure external function generator signal accurately using an Agilent 3458A DVM as reference

Ranked causes

  1. 1. Current offset error in sense terminals (common)
First check

Short current sense terminals

Test point
I SENSE
Expected
Shorted across terminals

10 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: External Current Shunt Setting

Verify and update the external current shunt resistance value in the unit to match the label on the physical shunt

Ranked causes

  1. 1. Shunt value setting mismatch (common)
First check

Display current shunt value setting

Test point
SHUNT command in Calibration menu
Expected
Resistance value matches label on physical external current shunt

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: External voltage measurement high voltage

Calibrate the Power Analyzer voltage sense terminals to measure external ac voltage source output accurately using an Agilent 3458A DVM as reference

Ranked causes

  1. 1. Voltage offset error in sense terminals (common)
First check

Reset unit to known state

Test point
Front panel
Expected
*RST display

8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Initialization

Unit requires initialization to restore non-volatile memory after corruption or first setup. Erases all calibration, waveform, and list data.

Ranked causes

  1. 1. Non-volatile memory corruption or first-time setup (unknown)
First check

Verify EEINIT command appears on display

Test point
Front panel display
Expected
EEINIT <model> shown

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Initialize unit

Non-volatile memory corruption or first-time initialization requiring front panel EEINIT command sequence

Ranked causes

  1. 1. Non-volatile memory corruption or first-time setup (unknown)
First check

Turn unit on

Test point
Front panel power switch
Expected
Unit powers on

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Output Frequency Modulation

Configure modulating frequency, frequency deviation, and enable frequency modulation on the output

Ranked causes

  1. 1. Frequency modulation parameters not set (unknown)
First check

Verify modulating frequency setting

Test point
FM:FREQ command
Expected
Modulating frequency value displayed Hz

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Post-repair adjustment

Supply requires calibration after replacing A8 DSP Assembly, A9 Output Filter/Relay Assembly, or A10 Control Assembly; A8 replacement requires prior initialization

Ranked causes

  1. 1. Component replacement triggers calibration requirement (common)
First check

Verify initialization completed if A8 DSP Assembly was replaced

Test point
A8 DSP Assembly
Expected
Initialization completed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ROM upgrade

Update to newer ROM versions without destroying calibration data; identify and upgrade firmware revision

Ranked causes

  1. 1. Need to upgrade to newer ROM revision (unknown)
First check

Identify firmware revision

Test point
Front panel display at power-on or GPIB *IDN? query
Expected
Firmware revision displayed for approximately 10 seconds or returned via GPIB

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CV error amplifier output incorrect high voltage

CV (constant voltage) error amplifier not responding to RAIL_VMON stimulus; output at U447-7 or U463-3/4 not toggling as expected.

Ranked causes

  1. 1. CV error amplifier IC failure (occasional)
First check

Apply 0V stimulus to RAIL_VMON

Test point
RAIL_VMON input
Expected
CV error amp responds; RAIL_OV_INH and RAIL_INH_ISO show expected pulses per test patterns

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Discrete Fault Indicator (DFI) not functioning

J113-1 does not respond to DFI commands or does not go low

Ranked causes

  1. 1. DFI signal path or control logic failure (unknown)
First check

Connect +5 Vdc to DFI terminals

Test point
DFI terminals via 5K resistor
Expected
+5 Vdc applied
Tools
5K resistor, power supply

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Initialization fails with OUT OF RANGE error

OUT OF RANGE error displayed during ROM upgrade, requiring re-initialization and re-calibration

Ranked causes

  1. 1. Instrument previously initialized with QA firmware revision or non-volatile memory corrupted (occasional)
First check

Check for OUT OF RANGE error on front panel display

Test point
Front panel display
Expected
OUT OF RANGE error shown

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inverter assembly fails under loaded conditions high voltage

A7 passes static checks but fails when 500 VDC rail applied or under current load

Ranked causes

  1. 1. High voltage breakdown on FETs or capacitors (occasional)
First check

Perform initial A7 inverter troubleshooting

Test point
Overall A7 assembly
Expected
All waveforms OK per Fig 4-7

7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low rail detect circuit not functioning high voltage

Low rail warning or low rail final inhibit signals not toggling at expected thresholds when RAIL_VMON varied.

Ranked causes

  1. 1. Low rail detect comparator failure (occasional)
First check

Measure RAIL_LOW_WARN threshold

Test point
U424-14 output
Expected
Toggles at approximately 10.6V RAIL_VMON
Tools
DMM or oscilloscope

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No communications pulses

Waveforms at communication test points never go high or low

Ranked causes

  1. 1. Communications circuit failure on A11 Trigger/BNC Board (unknown)
First check

Check communications pulses

Test point
U150-1, U151-5, U153-3, U154-1
Expected
Similar waveforms that go high and low
Tools
Oscilloscope

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No trigger input recognition

Trigger output pulse not generated when Trig In BNC is shorted

Ranked causes

  1. 1. External trigger input path failure (unknown)
First check

Set external trigger mode

Test point
Front panel
Expected
External trigger mode enabled
Tools
Power supply

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No trigger output pulse

Trigger Out BNC does not produce pulse when SHIFT TRIGGER command is entered

Ranked causes

  1. 1. Signal path failure from A8 DSP board to Trigger Out (unknown)
First check

Set up trigger output

Test point
Trig Out BNC
Expected
Trigger pulse output
Tools
Oscilloscope

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Analyzer input measurement setup

Select Power Analyzer input as measurement source and configure frequency measurement source

Ranked causes

  1. 1. Measurement source not configured correctly (occasional)
First check

Verify Power Analyzer input selection

Test point
Front panel display phase indicator
Expected
φ2 displayed to indicate Power Analyzer input active

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rail overvoltage detect circuit not responding high voltage

RAIL_OV_INH (rail overvoltage inhibit) signal not toggling at expected threshold; overvoltage protection not triggering.

Ranked causes

  1. 1. Rail overvoltage comparator failure (occasional)
First check

Measure RAIL_OV_INH response at U463-3

Test point
U463-3 output
Expected
Signal LOW (≈0V) when RAIL_VMON indicates overvoltage condition; HIGH when normal
Tools
DMM or oscilloscope

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ramp generator signal incorrect or missing high voltage

Ramp waveform at ramp generator test points not present or does not match expected sawtooth pattern.

Ranked causes

  1. 1. Ramp generator IC failure or missing input signal (common)
First check

Measure ramp waveform at U475-7

Test point
U475-7 output
Expected
Sawtooth waveform at ~83 kHz per Figure 4-5D
Tools
Oscilloscope (2.00 V/div, 5.00 us/div)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Remote Inhibit (RI) not disabling output high voltage

Output does not go to 0 Vac when RI terminals are shorted

Ranked causes

  1. 1. RI control signal path failure (unknown)
First check

Set 100 Vac at output

Test point
Output terminals
Expected
100 Vac
Tools
Voltmeter

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit safety hazard from exposed screw terminals during normal operation high voltage

Connector and barrier block screw terminals remain exposed to contact with personnel during equipment operation, creating electrical shock and arc hazard

Ranked causes

  1. 1. Cable hoods or covers not installed or removed during operation (common)
First check

Verify all cable hoods and covers are installed before powering unit

Test point
All Power Analyzer and external shunt connectors and barrier blocks
Expected
Cable hoods and covers fully installed, covering all exposed screw terminals

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verify AC output voltage detection high voltage

Check ALC RMS voltage detection at J822 when cable to J627 connected and ac source can output

Ranked causes

  1. 1. ALC RMS detection not generating expected signals (unknown)
First check

Measure J822-1 AC voltage during ALC RMS detection

Test point
J822-1
Expected
3 Vac

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verify current limit response with output short circuit high voltage

Check current limit programming response when output is shorted

Ranked causes

  1. 1. Current limit not responding correctly to short circuit (unknown)
First check

Measure J822-3 with output shorted at 5V 4A limit

Test point
J822-3
Expected
-0.75 Vac

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1Non-volatile RAM RD0 section checksum failedPower-on self-test detects RAM errorerror
10RAM selftest failedPower-on self-test detects RAM errorerror
11-31DAC selftest error, expected <n>, read <reading>Power-on self-test detects DAC output mismatcherror
2Non-volatile RAM CONFIG section checksum failedPower-on self-test detects RAM errorerror
200Outgrd not respondingRuntime communication check failserror
201Front panel not respondingRuntime communication check failserror
210Ingrd receiver framing errorRuntime communication check failswarning
211Ingrd uart overrun statusRuntime communication check failswarning
212Ingrd received bad tokenRuntime communication check failswarning
213Ingrd receiver buffer overrunRuntime communication check failswarning
214Ingrd input buffer overrunRuntime communication check failswarning
215Outgrd output buffer overrunRuntime communication check failswarning
216RS-232 receiver framing errorRuntime RS-232 communication failswarning
217RS-232 receiver parity errorRuntime RS-232 communication failswarning
218RS-232 receiver overrun errorRuntime RS-232 communication failswarning
219Ingrd inbuf count sync errorRuntime communication check failswarning
220Front panel uart overrunRuntime communication check failswarning
221Front panel uart framingRuntime communication check failswarning
222Front panel uart parityRuntime communication check failswarning
223Front panel buffer overrunRuntime communication check failswarning
224Front panel timeoutRuntime communication check failswarning
3Non-volatile RAM CAL section checksum failedPower-on self-test detects RAM errorerror
4Non-volatile RAM WAVEFORM section checksum failedPower-on self-test detects RAM errorerror
40Voltage selftest errorPower-on self-test voltage output incorrecterror
401Calibration command blocked due to INHIBIT CAL switch settingCalibration command sent when INHIBIT CAL switch is activeerror
401CAL switch prevents calibrationCalibration command executed with CAL switch in wrong positionerror
402CAL password is incorrectCalibration command executed with wrong passworderror
403CAL not enabledCalibration command executed without enabling calibration modeerror
404Computed readback cal constants are incorrectCalibration process detects computed values out of rangeerror
405Computed programming cal constants are incorrectCalibration process detects computed values out of rangeerror
406Incorrect sequence of calibration commandsCalibration commands executed out of ordererror
43Current selftest errorPower-on self-test current output incorrecterror
5Non-volatile RAM STATE section checksum failedPower-on self-test detects RAM errorerror
6Non-volatile RAM LIST section checksum failedPower-on self-test detects RAM errorerror
600Systems in mode:list have different list lengthsMultiple channels configured with unequal list lengthswarning
601Requested voltage and waveform exceeds peak voltage capabilityCommand requests voltage/waveform combination beyond ratingerror
602Requested voltage and waveform exceeds transformer volt-second ratingCommand requests voltage/waveform combination exceeding transformer ratingerror
603Command only applies to RS-232 interfaceNon-RS-232 interface executes RS-232-only commandwarning
604Trigger received before requested number of pre-trigger readingsTrigger arrives before pre-trigger buffer fillswarning
605Requested RMS current too high for voltage rangeCommand requests current beyond voltage range capabilityerror
606Waveform data not definedCommand references undefined waveformerror
607VOLT, VOLT:SLEW, and FUNC:SHAP modes incompatibleIncompatible mode commands combinederror
608Measurement overrangeMeasurement exceeds input range of ADCwarning
609Output buffer overrunOutput data rate exceeds buffer capacitywarning
610Command cannot be given with present SYST:CONF settingCommand incompatible with current system configurationerror
7Non-volatile RAM RST section checksum failedPower-on self-test detects RAM errorerror
70Fan voltage failurePower-on self-test detects fan voltage out of rangewarning
80Digital I/O selftest errorPower-on self-test detects digital I/O failureerror
CAL DENIEDCalibration is locked and cannot be modifiedAttempting to enter calibration mode when internal switch prevents calibration changeserror
Error 613Frequency deviation exceeds maximum allowed valueFrequency deviation set to value greater than or equal to programmed output frequency, or within 0.001 Hz marginerror
OUT OF RANGEEntered calibration password is incorrectPressing Enter after entering wrong CAL ON passworderror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.