Agilent 669xA: troubleshooting and error codes
The ServiceForge knowledge base holds 17 documented symptoms for the Agilent 669xA, with 46 ranked causes, 60 diagnostic checks with test points and expected values, and 23 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
DAC defective
Digital-to-analog converter does not produce correct output steps or waveforms
Ranked causes
- 1. Bias and reference voltages out of specification (common)
- 2. One DAC output missing steps (occasional)
- 3. All DAC outputs missing steps (occasional)
- 4. CV or CC DAC defective (occasional)
- 5. Reference voltage out of specification (rare)
Verify bias and reference voltages
- Test point
- Test points 9 through 22 per table 6-3
- Expected
- Within specification
- Tools
- DMM
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Output held low
Output voltage cannot be raised above low level and PROT indicator is not illuminated
Ranked causes
- 1. Bias and reference voltages out of specification (common)
- 2. CC constant current annunciator is on (load impedance too low) (common)
- 3. UNR (underrange) annunciator is on (occasional)
- 4. A10 board defective (occasional)
- 5. A6 board defective (rare)
Verify bias and reference voltages
- Test point
- Test points 9 through 22
- Expected
- Within specification per table 6-3
- Tools
- DMM
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Restore factory calibration constants
Recover original factory calibration constants to operate power supply for troubleshooting or recalibration. Factory preset jumper position enables restoration.
Ranked causes
- 1. Factory calibration constants need restoration (unknown)
- 2. EEPROM requires initialization after replacement (unknown)
- 3. Transfer new calibration constants to EEPROM Factory Cal locations (unknown)
- 4. Pre-initialized A1 boards available (unknown)
Verify J3 jumper position on A1 Front Panel Board
- Test point
- J3 on A1 Front Panel Board
- Expected
- Jumper at FACTORY PRESET CAL position
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display
Unit powers on but display is blank or non-functional
Ranked causes
- 1. A1 front panel board defective (common)
- 2. LCD DSP1 or cable W2 defective (common)
- 3. A2 GPIB board defective (occasional)
- 4. A6 board defective (occasional)
Check display bias voltage
- Test point
- R1-6
- Expected
- +5V
- Tools
- DMM
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output held high high voltage
Output voltage is stuck at high level and cannot be reduced
Ranked causes
- 1. Bias and reference voltages out of specification (common)
- 2. CV OR gate diode A10D652 conducting (short circuit) (common)
- 3. Fast sense amplifier A10U608-7 defective (occasional)
- 4. A10 board defective (occasional)
Verify bias and reference voltages
- Test point
- Test points 9 through 22
- Expected
- Within specification per table 6-3
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial Down error
Display shows SERIAL DOWN error message indicating communication failure between front panel and control board
Ranked causes
- 1. GPIB board (A2) defective or not responding (common)
- 2. Cable W4 open or defective (occasional)
- 3. Secondary bias circuit defective (occasional)
- 4. Control board (A10) defective (occasional)
Check for serial data pulses at secondary interface
- Test point
- A2U110-8
- Expected
- SERIAL DATA PULSES present with respect to PRIMARY GROUND (test point 1)
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FET board (A3) defective high voltage
Power FET or bias circuit failure preventing proper output stage operation
Ranked causes
- 1. Power FET devices (Q201-204, Q301-304, Q211, Q311, Q222, Q322, Q233, Q333, Q244, Q344) shorted or defective (common)
- 2. Output capacitors (C201, C301) leaky or defective (occasional)
- 3. Bias supply or FET gate bias circuit defective (occasional)
Isolate A3 FET board from supply for testing
- Test point
- A3 board installation
- Expected
- Board removed from supply with heatsink attached
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OV at turn-on high voltage
Overvoltage circuit activates immediately when supply powers on instead of when threshold is exceeded
Ranked causes
- 1. Output voltage setting higher than OV setting in memory register (common)
- 2. Output momentarily exceeds voltage setting at turn-on (occasional)
- 3. A10 board or OV circuit defective (occasional)
Monitor output voltage at turn-on with scope
- Test point
- Output terminals
- Expected
- Output does not exceed programmed voltage
- Tools
- Oscilloscope, DC coupled
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OV will not fire
Overvoltage circuit does not respond when overvoltage condition is programmed or detected
Ranked causes
- 1. A6 board defective (common)
- 2. A10 board defective (common)
- 3. A10 or A9 downprogrammer board defective (occasional)
Check bias and reference voltages
- Test point
- Test points 15 through 22
- Expected
- Within specification
- Tools
- DMM
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Supply does not operate properly high voltage
Unit fails power-on selftest or operates abnormally after power-on
Ranked causes
- 1. GPIB board (A2) defective (common)
- 2. Front panel board (A1) defective (common)
- 3. Cable W5 defective (occasional)
Verify supply basic operation
- Test point
- Front panel
- Expected
- Unit powers on without error messages
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slow Down Programming error
Display shows SLOW DOWN PROGRAMMING error indicating slow response during programming of output voltage and current
Ranked causes
- 1. Control board (A10) defective or degraded performance (common)
- 2. Power supply board (A9) fuses open or defective (occasional)
Program output voltage and current to half scale
- Test point
- Front panel controls
- Expected
- Output voltage and current set to 1/2 full scale
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CAL DENIED displayed or error code 1 (GPIB)
Front panel calibration attempt shows CAL DENIED message or GPIB calibration returns error code 1 on Systems Supply, preventing access to calibration mode.
Ranked causes
- 1. INHIBIT CAL jumper installed (common)
Verify INHIBIT CAL jumper location
- Test point
- INHIBIT CAL jumper (Figure 3-15)
- Expected
- NORM position (not INHIBIT CAL position)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Current programming and readback accuracy high voltage
Verify programmed current matches actual output for each model
Ranked causes
- 1. Out-of-spec current output (unknown)
Measure output current
- Test point
- Output terminals
- Expected
- See Table 2-2 for model: 6690A 440A, 6691A 220A, 6692A 110A
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Post-repair adjustment
Annual calibration or after component replacement in CV/CC DACs, operational amplifiers, control circuits, readback DAC/comparators, front panel board, or EEPROM. Required to verify output accuracy after service.
Ranked causes
- 1. Component replacement in control or readback circuits (common)
Verify INHIBIT CAL jumper position
- Test point
- INHIBIT CAL jumper (Figure 3-15)
- Expected
- NORM position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Transfer factory preset calibration data
Transfer calibration constants from operating locations to EEPROM factory preset locations after calibration and alignment complete. Requires program execution.
Ranked causes
- 1. New calibration constants need storage in factory preset locations (unknown)
Input power supply model number to program
- Test point
- Program input prompt
- Expected
- 6690A, 6691A, or 6692A
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EEPROM initialization after component replacement
Reinitialize EEPROM AlU6 on A1 Front Panel Board after board or EEPROM replacement to load proper GPIB address, model number, and programming/calibration constants.
Ranked causes
- 1. Front Panel Board or EEPROM replaced (unknown)
Verify component replacement
- Test point
- A1 Front Panel Board and EEPROM AlU6
- Expected
- Board or EEPROM component replaced
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PASSWD ERROR displayed or error code 2 (GPIB)
Calibration mode entry failed with PASSWD ERROR message on front panel or error code 2 via GPIB, indicating incorrect or unknown calibration password.
Ranked causes
- 1. Incorrect calibration password entered (common)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Calibration inhibited or unable to enter calibration mode via GPIB on Systems Supply | INHIBIT CAL jumper in INHIBIT CAL position when GPIB calibration is attempted | error |
2 | Incorrect or unknown calibration password entered | User enters wrong password when attempting to enter calibration mode via front panel or GPIB | error |
E10 SEC 5V | Secondary interface 5 volt readback test failed | Power-on and *TST? | error |
E11 TEMP | Ambient temperature readback test failed | Power-on and *TST? | error |
E12 DACS | CV or CC DAC tests failed | Power-on selftest | error |
E1 FP RAM | Front panel RAM test failed | Power-on selftest | error |
E2 FP ROM | Front panel ROM test failed | Power-on and *TST? | error |
E3 EE CHKSM | Front panel EEPROM checksum test failed | Power-on and *TST? | error |
E4 PRI XRAM | Primary interface external RAM test failed | Power-on selftest | error |
E5 PRI IRAM | Primary interface internal RAM test failed | Power-on selftest | error |
E6 PRI ROM | Primary interface ROM test failed | Power-on and *TST? | error |
E7 GPIB | GPIB interface test failed | Power-on selftest | error |
E8 SEC RAM | Secondary interface RAM test failed | Power-on selftest | error |
E9 SEC ROM | Secondary interface ROM test failed | Power-on and *TST? | error |
EE WRITE ERR | EEPROM write failure | During operation or selftest | error |
SBUF FULL | Serial buffer full condition | During operation or selftest | error |
SBUF OVERRUN | Serial buffer failure | During operation or selftest | error |
SECONDARY DN | Serial data line failure on Main board | During operation or selftest | error |
SERIAL DOWN | Serial data line failure on A2 board | During operation or selftest | error |
SERIAL TIMOUT | Serial data line failure on A2 board | During operation or selftest | error |
UART FRAMING | UART failed framing check | During operation or selftest | error |
UART OVERRUN | UART buffer overrun | During operation or selftest | error |
UART PARITY | UART failed parity check | During operation or selftest | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI