Agilent 66312A: troubleshooting and error codes
The ServiceForge knowledge base holds 34 documented symptoms for the Agilent 66312A, with 65 ranked causes, 101 diagnostic checks with test points and expected values, and 22 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Output current less than rating high voltage
Power supply output current below specified maximum rating; control voltage and feedback circuits may be at fault
Ranked causes
- 1. Interface board current control failure (unknown)
- 2. Current monitoring voltage incorrect (unknown)
- 3. U324A/B current feedback amplifier failure (unknown)
- 4. Control or downprogrammer circuit fault (unknown)
- 5. U321A or R360 circuit fault (unknown)
- 6. U321B or voltage sense circuit failure (unknown)
- 7. Output transistor stage not conducting (unknown)
Measure CC_Prog voltage at R336
- Test point
- R336 (CC_Prog)
- Expected
- <= -4.8V
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test failure
Display shows ERR annunciator; output is not disabled; query unit to determine specific error codes
Ranked causes
- 1. ROM checksum error (common)
- 2. RAM test failure (common)
- 3. DAC test failure (common)
- 4. RS-232 communication error (occasional)
- 5. Front Panel communication error (occasional)
Query unit to determine specific error code
- Test point
- Unit interface
- Expected
- Error code E1–E5 (ROM checksums), E10–E15 (DAC tests), E80 (I/O test), E213–E223 (communication errors)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Protection activated high voltage
Unit is in protection mode (FS, OT, OC, OV, RI); output disabled
Ranked causes
- 1. Overvoltage protection triggered (common)
- 2. Overcurrent protection triggered (common)
- 3. Over-temperature protection (occasional)
- 4. Remote inhibit active (occasional)
Observe protection annunciator state
- Test point
- front panel protection indicator
- Expected
- annunciator off
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias and reference voltage out of range
Check bias and reference voltages before troubleshooting circuits; voltages out of spec indicate power supply issue
Ranked causes
- 1. Primary +5V out of specification (common)
- 2. Secondary voltage supply failure (common)
- 3. Reference voltage out of tolerance (occasional)
Measure primary +5V supply
- Test point
- A1 J206-1
- Expected
- +5V, +/- 0.15V
- Tools
- DMM
8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Post-repair procedures
Annual calibration or recalibration required after component replacement in voltage/current monitor or interface circuits
Ranked causes
- 1. Component replacement in A1 Control Board voltage/current monitor circuits (common)
- 2. Component replacement in A1 Control Board current monitor resistors R425/R431 (common)
- 3. A2 Interface Board replacement (common)
Verify INHIBIT CAL switch S201 configuration
- Test point
- S201 on A2 Interface board, switches 3 and 4
- Expected
- Switch 3 Off, Switch 4 Off for Normal operation
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV/CC annunciator troubleshooting with no output
CV or CC annunciator on indicates problem in associated control circuit; UNR indicates neither circuit in control
Ranked causes
- 1. Voltage Amplifier circuit failure (common)
- 2. Current Amplifier circuit failure (common)
- 3. Output stage failure (occasional)
Measure voltage drop across CV gating diode
- Test point
- A1 D317
- Expected
- Forward drop present if CV mode active; low signal at U328C output (CV_DETECT*)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display on unit high voltage
Front panel display is not powered on when unit is turned on
Ranked causes
- 1. AC input failure (common)
- 2. Line voltage selection incorrect (occasional)
- 3. Bias supply failure (occasional)
Verify display is powered after power-on
- Test point
- front panel display
- Expected
- display illuminated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slow down programmer output voltage incorrect high voltage
When external supply applies 5V at 2A across output, meter indicates incorrect voltage or current
Ranked causes
- 1. U321A output amplifier error (unknown)
- 2. U308A output stage failure (unknown)
- 3. Q306 or Q309 control stage issue (unknown)
Measure voltage at meter output with 5V 2A applied
- Test point
- Meter output
- Expected
- 5V -1.5A
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration password incorrect or forgotten
Incorrect password entered during calibration attempt; OUT OF RANGE or error 402 displayed
Ranked causes
- 1. Incorrect calibration password entered (common)
- 2. Calibration password forgotten (occasional)
Verify password entry
- Test point
- Front panel display or GPIB calibration mode
- Expected
- Correct password accepted; calibration mode enabled
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current output out of specification high voltage
Programmed current limit does not match actual measured output current
Ranked causes
- 1. Current calibration needed (common)
- 2. Interface board failure (occasional)
Check CC_PROG voltage at R336
- Test point
- R336
- Expected
- 0 to -4.8 V for 0 to full scale programmed current, 0 to -4.8 V
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan not operating caution
Cooling fan does not run even when unit is powered on
Ranked causes
- 1. Low or missing 15V bias supply (common)
- 2. Fan driver circuit failure (occasional)
Observe fan operation during power-on
- Test point
- cooling fan
- Expected
- fan rotates smoothly
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
J307 voltage measurement out of range
J307 connects A1 Main Board to A2 Interface Assembly; out-of-range voltages indicate circuit problem on A1 or A2
Ranked causes
- 1. CV Mode voltage outside specification (common)
- 2. CC Mode voltage outside specification (common)
Measure PM_INHIBIT at J307 pin 1
- Test point
- A1 J307 pin 1
- Expected
- 0V when enabled
- Tools
- DMM
9 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output voltage high voltage
Output terminals show zero voltage or less than minimum programmed value
Ranked causes
- 1. Gated bias supply failure (common)
- 2. Series regulator failure (occasional)
Measure gated +/-15V at fuse holders F302 and F304
- Test point
- F302, F304
- Expected
- ±15 V DC, ±15 V
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output held high high voltage
Output remains at high voltage level and will not decrease when attempting to reduce programmed voltage
Ranked causes
- 1. Series pass element stuck (occasional)
- 2. Feedback circuit failure (occasional)
Disable OV protection via DIAG:SCR 0 command
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage out of specification high voltage
Programmed voltage does not match actual measured output voltage
Ranked causes
- 1. Voltage calibration needed (common)
- 2. Interface board failure (occasional)
Program full scale voltage and current; enable output
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage at turn-on high voltage
Output voltage spikes high when unit is powered on; OV protection activates
Ranked causes
- 1. OV detection circuit failure (occasional)
- 2. Incorrect OV protection setting (common)
Connect scope DC coupled across output during power-on
- Test point
- output terminals
- Expected
- output stays below OV limit
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
20 mA range current readback accuracy out of specification at 0 A
Current readback in 20 mA range differs from DMM measurement at zero amps beyond specified limits
Ranked causes
- 1. Low current range readback circuit calibration error (unknown)
Measure 20 mA range current readback at 0 A
- Test point
- DMM in current mode
- Expected
- difference within specified limits per performance test record chart 20mA RANGE CURRENT READBACK ACCURACY @ 0A
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
20 mA range current readback accuracy out of specification at 20 mA
Current readback in 20 mA range differs from DMM measurement at 20 mA beyond specified limits
Ranked causes
- 1. Low current range readback circuit error at 20 mA (unknown)
Measure 20 mA range current readback at 20 mA
- Test point
- DMM in current mode
- Expected
- difference within specified limits per performance test record chart 20mA RANGE CURRENT READBACK ACCURACY @ 20mA
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration switch CAL DENIED or error 401
Calibration mode denied due to INHIBIT CAL switch setting
Ranked causes
- 1. INHIBIT CAL switch S201 set to inhibit calibration (common)
Check S201 switch 3 and 4 configuration
- Test point
- S201 on A2 Interface board
- Expected
- Switch 3 Off, Switch 4 Off for Normal; Switch 3 On, Switch 4 Off for Inhibit Calibration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power supply initialization
Re-initialize EEPROM on A2 Interface board after replacement; required before calibration
Ranked causes
- 1. Interface board replaced (unknown)
Enable Calibration mode
- Test point
- Front panel
- Expected
- Calibration mode active
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ROM upgrade procedure
Upgrade Interface board ROM without affecting existing calibration data in EEPROM
Ranked causes
- 1. ROM upgrade required (unknown)
Enable Calibration mode and press '0' and '1' keys simultaneously
- Test point
- Front panel
- Expected
- EEINIT <model> displayed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CC load effect out of specification
Change in output current from full-scale voltage to short circuit exceeds specified limit
Ranked causes
- 1. Output stage or current feedback regulation inadequate (unknown)
Verify CC mode operation at full-scale voltage
- Test point
- UUT CC annunciator
- Expected
- annunciator illuminated
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CC source effect out of specification high voltage
Change in output current from minimum to maximum line voltage exceeds specified limit
Ranked causes
- 1. Line voltage regulation or supply regulation inadequate (unknown)
Set line voltage to nominal value
- Test point
- Variable voltage transformer
- Expected
- nominal line voltage
- Tools
- Variable voltage transformer
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current programming and readback accuracy out of specification at 0 A
Current readback does not match programmed value within specified limits at zero amps
Ranked causes
- 1. Current monitoring circuit calibration drift (unknown)
Measure current readback at 0 A programming point
- Test point
- Current monitoring resistor
- Expected
- within specified limits per performance test record card CC PROGRAMMING @ 0 AMPS
- Tools
- DVM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current programming and readback accuracy out of specification at full-scale
Current readback does not match programmed value within specified limits at full-scale current
Ranked causes
- 1. Current monitoring circuit calibration drift at full-scale (unknown)
Measure current readback at full-scale programming
- Test point
- Current monitoring resistor
- Expected
- within specified limits per performance test record card CC PROGRAMMING @ FULL-SCALE
- Tools
- DVM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current sink operation or readback out of specification
Negative current (sink) operation does not function correctly or readback differs from expected value
Ranked causes
- 1. Current sink circuitry malfunction or readback error (unknown)
Verify current sink operation produces ~-1.4 A display reading
- Test point
- UUT display
- Expected
- approximately −1.4 A
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV load effect exceeds specification high voltage
Output voltage change when load current changes from full load to no load exceeds maximum allowable value
Ranked causes
- 1. Output impedance too high or regulation circuit degraded (unknown)
Record DVM voltage with full-scale current load
- Test point
- +S and -S terminals
- Expected
- Stable reading at programmed full-scale voltage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV noise (PARD) exceeds specification
Peak-to-peak or RMS ripple and noise voltage on output exceeds maximum allowable limits
Ranked causes
- 1. Output filter capacitors degraded or switching noise coupling to output (unknown)
Measure peak-to-peak noise on oscilloscope
- Test point
- (+) and (-) output terminals
- Expected
- Peak-to-peak voltage not to exceed CV NOISE (PARD) limit per performance test record chart
- Tools
- Oscilloscope with 20 MHz bandwidth limit, RF tip probe
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CV source effect exceeds specification high voltage
Output voltage change when input ac line voltage varies from minimum to maximum rated value exceeds limit
Ranked causes
- 1. Line voltage regulation circuit degraded or ac filter insufficient (unknown)
Set transformer to nominal line voltage and program supply
- Test point
- AC power input
- Expected
- Supply stable at nominal voltage
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disable protection features for troubleshooting caution
Protection features can be disabled except for overvoltage protection; resets to default at power-up
Ranked causes
- 1. Protection features interfering with troubleshooting (rare)
Press 0 and 1 keys simultaneously
- Test point
- Front panel keys
- Expected
- EEINIT <model> displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual fan speed control caution
Reduce fan speed manually for testing where fan noise interferes; resets to Automatic at power-up
Ranked causes
- 1. Fan noise interfering with test (rare)
Press 0 and 1 keys simultaneously
- Test point
- Front panel keys
- Expected
- EEINIT <model> displayed
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slow down-programming high voltage
Output voltage decreases slowly or not at all when programmed to lower value
Ranked causes
- 1. Down-programmer circuit fault (occasional)
Program voltage up and then down
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient recovery time out of specification
Output voltage does not return to specified value within specified time following 50% load current change
Ranked causes
- 1. Transient response characteristics degraded (unknown)
Verify output voltage recovery time on loading transient
- Test point
- +S and -S terminals
- Expected
- within specified voltage in less than specified time
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage programming and readback accuracy out of spec high voltage
DVM and front panel display readings do not match specification limits when programming to zero volts or full-scale voltage
Ranked causes
- 1. Voltage programming circuit malfunction (unknown)
Measure DVM voltage at +S to -S at zero volts programming
- Test point
- +S and -S terminals
- Expected
- Within CV PROGRAMMING @ 0 VOLTS limits per performance test record chart
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
401 | Calibration inhibited by INHIBIT CAL switch setting | Attempt to calibrate over GPIB when S201 INHIBIT CAL switch is set to On | warning |
402 | Incorrect calibration password entered over GPIB | GPIB calibration command with wrong password | warning |
E1 | Checksum in Read-only Non-volatile ROM | ROM checksum validation failure during self-test | critical |
E10 | RAM test failed | RAM memory validation failure during self-test | critical |
E11 | 12 bit DAC test failed at 0 input | 0 written to DAC U241A and B; ADC U242 not reading 133 +/- 7 counts | critical |
E12 | 12 bit DAC test failed at 4095/0 input | 4095 written to DAC U241A and 0 to B; ADC U242 not reading 71 +/- 7 counts | critical |
E13 | 12 bit DAC test failed at 0/4095 input | 0 written to DAC U241A and 4095 to B; ADC U242 not reading 71 +/- 7 counts | critical |
E14 | 12 bit DAC test failed at 4095 input | 4095 written to DAC U241A and B; ADC U242 not reading 10 +/- 7 counts | critical |
E15 | 8 bit DAC test failed | 10 and 240 written to DAC U244; ADC U242 not reading 10 and 240 +/- 7 counts | critical |
E2 | Checksum in Config Non-volatile ROM | Configuration ROM checksum validation failure during self-test | critical |
E213 | RS-232 input buffer overrun | Serial communication data received faster than processed | error |
E216 | RS-232 framing error | RS-232 data frame format violation detected | error |
E217 | RS-232 parity error | RS-232 parity check validation failure | error |
E218 | RS-232 UART input overrun | RS-232 UART receives data while buffer still processing previous data | error |
E220 | Front Panel communication UART input overrun | Front Panel UART receives data while buffer still processing | error |
E221 | Front Panel communication UART framing error | Front Panel UART data frame format violation detected | error |
E222 | Front Panel communication UART parity error | Front Panel UART parity check validation failure | error |
E223 | Front Panel firmware input buffer overrun | Front Panel firmware receives data faster than processed | error |
E3 | Checksum in Cal Non-volatile ROM | Calibration ROM checksum validation failure during self-test | critical |
E4 | Checksum in State Non-volatile ROM | State ROM checksum validation failure during self-test | critical |
E5 | Checksum in RST Non-volatile ROM | Reset ROM checksum validation failure during self-test | critical |
E80 | Digital I/O test failed | SEC_PCLR written low and high; not read back correctly through Xilinx | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI