Agilent 66311B: troubleshooting and error codes
The ServiceForge knowledge base holds 25 documented symptoms for the Agilent 66311B, with 45 ranked causes, 57 diagnostic checks with test points and expected values, and 23 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power-on self-test failure
ERR annunciator comes on during power-up self-test sequence; digital and DAC circuits tested
Ranked causes
- 1. Checksum error in Read-only Non-volatile ROM (unknown)
- 2. Checksum error in Config Non-volatile ROM (unknown)
- 3. Checksum error in Cal Non-volatile ROM (unknown)
- 4. Checksum error in State Non-volatile ROM (unknown)
- 5. Checksum error in RST Non-volatile ROM (unknown)
- 6. RAM test failed (unknown)
- 7. 12 bit DAC test failed at 0 input (unknown)
- 8. 12 bit DAC test failed at 4095/0 input (unknown)
- 9. 12 bit DAC test failed at 0/4095 input (unknown)
- 10. 12 bit DAC test failed at 4095 input (unknown)
- 11. 8 bit DAC test failed (unknown)
- 12. Digital I/O test failed (unknown)
- 13. RS-232 input buffer overrun (unknown)
- 14. RS-232 framing error (unknown)
- 15. RS-232 parity error (unknown)
- 16. RS-232 UART input overrun (unknown)
- 17. Front Panel comm UART input overrun (unknown)
- 18. Front Panel comm UART framing error (unknown)
- 19. Front Panel comm UART parity error (unknown)
- 20. Front Panel firmware input buffer overrun (unknown)
20 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Initialize dc source after A2 Interface board replacement
Procedure to reinitialize and calibrate supply when Interface board has been replaced
Ranked causes
- 1. A2 Interface board has been replaced (rare)
- 2. Calibration procedure required after initialization (common)
Enable Calibration mode
- Test point
- Front panel keys
- Expected
- Press Shift, then Cal, then Enter
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CAL DENIED appears on display during calibration
Display shows 'CAL DENIED' message when attempting to calibrate the unit
Ranked causes
- 1. Inhibit Calibration switch (S201) is set to ON (common)
Verify switch S201 position
- Test point
- S201 on A2 Interface board near GPIB connector
- Expected
- Switch 3 OFF, Switch 4 OFF for Normal mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration denied or error 401 on GPIB calibration
CAL DENIED message appears on display when calibration is attempted, or error code 401 occurs when calibrating over GPIB due to INHIBIT CAL switch being set
Ranked causes
- 1. INHIBIT CAL switch (S201) set to ON position (common)
Verify S201 switch position
- Test point
- S201 Switch 3 and Switch 4 on A2 Interface board
- Expected
- Switch 3 Off, Switch 4 Off for Normal operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Annual or component replacement
Calibration required annually and after replacing A1 power module or A2 interface board. If A2 replaced, must initialize first then calibrate using correct password.
Ranked causes
- 1. A1 or A2 component replaced (rare)
Verify component replacement
- Test point
- Physical inspection
- Expected
- A1 or A2 module replaced
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CC noise verification
Measure RMS current noise in constant current (CC) mode at full scale output to verify it does not exceed specification limits
Ranked causes
- 1. Excessive noise in CC mode output (unknown)
Verify test environment for external noise sources
- Test point
- Test setup
- Expected
- No external noise interference detected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: DC source reinitialization
After replacing A2 Interface board containing EEPROM with constants and references affecting alignment, supply must be reinitialized and calibrated
Ranked causes
- 1. A2 Interface board replaced (occasional)
Enable Calibration mode
- Test point
- Front panel controls
- Expected
- Calibration mode active
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Disable protection features caution
Disable overcurrent/thermal protection for troubleshooting. Default is protection enabled (NO-PROTECT OFF). Must be reset after power cycle. Overvoltage protection cannot be disabled.
Ranked causes
- 1. Protection feature needs to be disabled for troubleshooting (rare)
Access protection feature menu
- Test point
- Display
- Expected
- NO-PROTECT<OFF> shown
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Incorrect password
Wrong calibration password prevents calibration mode entry. Displays 'OUT OF RANGE' on front panel or error 402 on GPIB interface.
Ranked causes
- 1. Incorrect calibration password entered (occasional)
Enter calibration password
- Test point
- Display or GPIB
- Expected
- Calibration mode enabled
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Manual fan speed control
Reduce fan speed for acoustic testing or light load conditions. Default is Automatic; must be reset after power cycle.
Ranked causes
- 1. Fan speed adjustment needed for test conditions (common)
Select FAN:MODE menu
- Test point
- Display
- Expected
- FAN:MODE<AUTO.> shown
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Constant current sink mode not functioning
Output current does not decrease linearly as voltage increases in negative current mode
Ranked causes
- 1. Current sink circuit malfunction (unknown)
Verify negative current displayed at 0V output
- Test point
- Front panel display
- Expected
- Negative current value, approximately 2.8A
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Constant current source mode not activated
CC annunciator not illuminating when load is adjusted to full-scale voltage
Ranked causes
- 1. Load impedance too high in CV mode (common)
Verify CC annunciator is illuminated
- Test point
- Front panel CC indicator
- Expected
- CC annunciator on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 401 occurs during GPIB calibration
Error 401 displayed or reported when attempting calibration over GPIB interface
Ranked causes
- 1. Inhibit Calibration switch (S201) is set to ON (common)
Verify switch S201 position
- Test point
- S201 on A2 Interface board near GPIB connector
- Expected
- Switch 3 OFF, Switch 4 OFF for Normal mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low range current readback inaccurate
20mA range current readings do not match between DMM and front panel display
Ranked causes
- 1. Low range readback circuit out of specification (unknown)
Measure current at 0A with 20mA range selected
- Test point
- Between DMM and current monitoring resistor
- Expected
- Within specification at 0A per test record
- Tools
- DMM in current mode
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current changes with line voltage variation
Output current varies when AC line voltage changes from minimum to maximum
Ranked causes
- 1. Source regulation out of specification (unknown)
Measure output current at minimum rated line voltage
- Test point
- Across current monitoring resistor
- Expected
- Within specification for CC Source Effect
- Tools
- DVM, variable voltage transformer
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current changes with load impedance
Output current varies when load changes from full-scale voltage to short circuit
Ranked causes
- 1. Load regulation out of specification (unknown)
Measure output current at full-scale voltage
- Test point
- Across current monitoring resistor
- Expected
- Within specification for CC Load Effect test
- Tools
- DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current programming and readback inaccurate
Current values programmed and displayed do not match measured values across current monitoring resistor
Ranked causes
- 1. Out of specification current measurement (unknown)
Measure voltage drop across current monitoring resistor at 20mA programming
- Test point
- Across current monitoring resistor
- Expected
- Within specification limits for 0A range
- Tools
- DVM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output current ripple and noise excessive
RMS noise measured on output exceeds specification in 20Hz to 20MHz band
Ranked causes
- 1. PARD (Periodic and Random Deviations) out of specification (unknown)
Measure RMS voltage across current monitoring resistor
- Test point
- Across current monitoring resistor in 20Hz to 20MHz band
- Expected
- Within PARD specification limit
- Tools
- RMS voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output noise and ripple excessive
Peak-to-peak or RMS ac voltage superimposed on dc output exceeds specification limits
Ranked causes
- 1. EMI coupling, output filter degradation, or switching noise issue (unknown)
Verify oscilloscope configuration
- Test point
- Oscilloscope bandwidth and probe
- Expected
- 20 MHz bandwidth limit, RF tip probe on scope probe
- Tools
- Oscilloscope with RF probe
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage changes with load current
Output voltage is not stable when load current changes from full load to no load
Ranked causes
- 1. Output impedance or regulation circuit drift (unknown)
Verify CV regulation at full load
- Test point
- +S to -S terminals
- Expected
- CV annunciator on, front panel display at full-scale voltage
- Tools
- Digital voltmeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output voltage varies with ac line voltage
Output voltage is not stable when ac line input voltage changes between minimum and maximum rated values
Ranked causes
- 1. Source regulation circuit drift or transformer coupling issue (unknown)
Set line voltage to minimum rated value
- Test point
- Variable voltage transformer input
- Expected
- Lowest rated line voltage (e.g., 104 Vac for 115 Vac nominal)
- Tools
- Variable voltage transformer
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROM upgrade: U205 Interface ROM replacement caution
Procedure to upgrade Primary Interface ROM (U205) on A2 Interface board to update firmware revision
Ranked causes
- 1. Firmware update required (occasional)
Identify current firmware revision from front panel
- Test point
- Address key, Up/Down annunciator to ROM: <A.xx.xx>
- Expected
- ROM: <A.xx.xx> displayed
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient recovery time excessive
Output voltage does not recover to within specified value after load current change within required time
Ranked causes
- 1. Control loop response or compensation circuit issue (unknown)
Observe transient response on oscilloscope
- Test point
- (+) to (-) terminals
- Expected
- Initial negative dip settles within 35 microseconds, recovery within 20mV band of final voltage
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transient response time exceeds specification
Output voltage does not return to specified value within 35 microseconds after current transient
Ranked causes
- 1. Compensation setting incorrect or circuit performance degraded (common)
Verify compensation mode set to High capacitance
- Test point
- Front panel compensation setting
- Expected
- High capacitance mode selected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage programming and readback inaccurate
Front panel display and/or GPIB readback voltage values do not match expected values at 0V or full-scale settings
Ranked causes
- 1. Calibration or display offset error (unknown)
Measure output voltage at 0V setting
- Test point
- +S to -S terminals
- Expected
- Within limits specified in performance test record card for Voltage Programming and Readback @ 0 Volts
- Tools
- Digital voltmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
401 | Calibration denied | Attempted calibration when INHIBIT CAL switch S201 is set | error |
401 | Calibration denied due to Inhibit Calibration switch setting | Attempted calibration over GPIB when Inhibit Calibration switch (S201) is enabled | warning |
402 | Incorrect calibration password (GPIB interface) | Wrong password entered during GPIB calibration mode access attempt | error |
E1 | Checksum error in Read-only Non-volatile ROM | Power-on self-test detects ROM checksum failure | error |
E10 | RAM test failed | Power-on self-test detects RAM memory failure | error |
E11 | 12 bit DAC test failed at 0 input | DAC test: 0 written to U241A/B, ADC U242 does not read 133 +/- 7 counts | error |
E12 | 12 bit DAC test failed at 4095/0 input | DAC test: 4095 to U241A, 0 to U241B, ADC U242 does not read 71 +/- 7 counts | error |
E13 | 12 bit DAC test failed at 0/4095 input | DAC test: 0 to U241A, 4095 to U241B, ADC U242 does not read 71 +/- 7 counts | error |
E14 | 12 bit DAC test failed at 4095 input | DAC test: 4095 written to U241A/B, ADC U242 does not read 10 +/- 7 counts | error |
E15 | 8 bit DAC test failed | DAC test: 10 and 240 written to U244, ADC U242 does not read 10 and 240 +/- 7 counts | error |
E2 | Checksum error in Config Non-volatile ROM | Power-on self-test detects ROM checksum failure | error |
E213 | RS-232 input buffer overrun | Serial data received faster than processor can handle | warning |
E216 | RS-232 framing error | Invalid start or stop bit sequence detected on RS-232 input | warning |
E217 | RS-232 parity error | Parity bit does not match received data on RS-232 input | warning |
E218 | RS-232 UART input overrun | UART receive buffer filled before data retrieved | warning |
E220 | Front Panel comm UART input overrun | Front panel communication UART receive buffer filled before data retrieved | error |
E221 | Front Panel comm UART framing error | Invalid frame detected in front panel serial communication | error |
E222 | Front Panel comm UART parity error | Parity error detected in front panel serial communication | error |
E223 | Front Panel firmware input buffer overrun | Front panel firmware receive buffer filled before data processed | error |
E3 | Checksum error in Cal Non-volatile ROM | Power-on self-test detects ROM checksum failure | error |
E4 | Checksum error in State Non-volatile ROM | Power-on self-test detects ROM checksum failure | error |
E5 | Checksum error in RST Non-volatile ROM | Power-on self-test detects ROM checksum failure | error |
E80 | Digital I/O test failed | SEC_PCLR signal cannot be written low and high, or read-back through Xilinx fails | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI