Agilent 654xA: troubleshooting and error codes
The ServiceForge knowledge base holds 18 documented symptoms for the Agilent 654xA, with 25 ranked causes, 46 diagnostic checks with test points and expected values, and 59 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power on
Power switch turned on but no display response or fan activation
Ranked causes
- 1. LINE switch not in Off position before startup (common)
- 2. Power cord not connected (common)
- 3. Incorrect line voltage fuse installed (occasional)
Verify LINE switch position
- Test point
- Front panel LINE switch
- Expected
- Off(0)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EEPROM initialization after component replacement
EEPROM chip A3U6 or front panel board A3 replaced; supply requires reinitialization with proper constants before calibration
Ranked causes
- 1. EEPROM chip A3U6 replaced (occasional)
- 2. Front panel board A3 replaced (occasional)
Run EEPROM initialization program (Figure 3-17)
- Test point
- Program prompt
- Expected
- Program pauses to select Initialization (I) or Factory Preset Replacement (F)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Output current adjustment high voltage
Procedure to set and verify output current control and overcurrent protection limits
Ranked causes
- 1. Current limit requires programming via keypad (common)
- 2. Overcurrent protection must be tested and set (common)
Connect short circuit to output terminals
- Test point
- Output terminals
- Expected
- Short circuit applied per test diagram (Fig 9)
- Tools
- Test jumper or wire
15 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Output voltage adjustment high voltage
Procedure to set and verify output voltage control and overvoltage protection limits
Ranked causes
- 1. Requires factory-specified voltage programming (common)
- 2. Overvoltage protection must be tested and set (common)
Set output voltage to 4 volts
- Test point
- Voltage control and display
- Expected
- VOLT 4, then 4.000 displayed in meter mode
- Tools
- Keypad
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output protection circuit tripped patient safety
Output voltage or current protection activated; Prot LED illuminated after short circuit or overvoltage condition
Ranked causes
- 1. Overvoltage protection (OVP) threshold exceeded (common)
- 2. Overcurrent protection (OCP) activated by output short (common)
Identify protection circuit that tripped
- Test point
- LED indicators and display
- Expected
- Prot LED on; display shows OV or OC message
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Troubleshoot power supply fault high voltage
Power supply does not deliver correct output voltage or current, or exhibits instability
Ranked causes
- 1. Failed or damaged circuit board (unknown)
- 2. Defective component on circuit board (unknown)
Verify problem is with power supply and not associated circuit
- Test point
- Input power and output connections
- Expected
- Problem isolated to power supply
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration denied or error code 1
CAL DENIED message on display or error code 1 occurs when attempting calibration via GPIB (models 664xA & 665xA)
Ranked causes
- 1. INHIBIT CAL jumper installed (common)
Verify INHIBIT CAL jumper position
- Test point
- INHIBIT CAL jumper location shown in Figure 3-3
- Expected
- Jumper removed from INHIBIT CAL position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EEPROM initialization and factory preset restoration
After replacing front panel board A3 or EEPROM chip A3U6, the supply must be reinitialized and calibrated. Program Figure 3-17 must be run twice: once with selection 'I' to initialize, then after calibration with selection 'F' to transfer constants to factory preset locations.
Ranked causes
- 1. Front panel board A3 or EEPROM chip A3U6 replaced (common)
Run initialization program with selection I
- Test point
- Power supply at GPIB address 705
- Expected
- Program pauses for Initialization (I) or Factory Preset Replacement (F) selection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Front panel board replacement
Recalibration required after replacing A3 Front Panel Board or EPROM chip A3U6
Ranked causes
- 1. A3 Front Panel Board or EPROM A3U6 replacement (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Post-repair output adjustment
Recalibration required after replacing CV/CC DACs, amplifiers, readback circuits, or reference components on A1 Main Board
Ranked causes
- 1. Component replacement on A1 Main Board (CV/CC DACs, amplifiers, readback circuits) (common)
Verify bias and reference voltages
- Test point
- See Table 6-3 in Chapter 6 for A2 GPIB board, A2 Isolator Board, and A1 Board test points
- Expected
- Within specification per service manual
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Transfer calibration constants to Factory Preset locations
Transfer newly obtained calibration constants to EEPROM Factory Cal locations to enable recovery using FACTORY PRESET CAL jumper
Ranked causes
- 1. Need to back up calibration constants after supply calibration (common)
Run EEPROM program for Factory Preset transfer
- Test point
- Program prompt
- Expected
- Program pauses to select Initialization (I) or Factory Preset Replacement (F)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Fuses indicator blinking high voltage
Check Fuses LED flashes; one or two of three-phase line fuses are open. If all three are open, the indicator does not blink and the unit requires service.
Ranked causes
- 1. One or two three-phase line fuses open (common)
Observe Check Fuses indicator status
- Test point
- Front panel Check Fuses LED
- Expected
- Blinking or steady, indicates fuse condition
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Fuses indicator not blinking with three open fuses caution
Check Fuses indicator does not blink when all three phase fuses are open; unit requires service
Ranked causes
- 1. All three phase fuses open (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows error message during self-test
LCD displays error code or message instead of normal asterisk pattern or meter mode
Ranked causes
- 1. Self-test failure detected (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan not operating caution
Power switch on but no fan noise or airflow detected
Ranked causes
- 1. Fan not running during normal self-test (occasional)
Listen and feel for fan operation
- Test point
- Rear panel grille
- Expected
- Audible fan noise and/or airflow detected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Moisture detected inside unit caution
Dew LED illuminated indicating internal humidity levels exceed safe operating range
Ranked causes
- 1. High humidity environment or condensation (occasional)
Observe Dew LED status at power-on
- Test point
- Dew LED indicator (668xA series)
- Expected
- Dew LED off during normal operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on in high humidity
Power supply fails to turn on when dew forms due to excessive moisture approaching 100%
Ranked causes
- 1. Excessive moisture (approaching 100% humidity) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VOLT or AMPS display shows +OL or –OL
Display indicates out-of-range reading when output voltage or current exceeds meter circuit range
Ranked causes
- 1. Output voltage or current exceeds meter range (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Addr | Power supply is addressed for Listen or Talk mode | GPIB system addressing | info |
Cal | Power supply is in calibration mode | Calibration procedure initiated | info |
CC | Power supply output in constant current mode | Output shorted or current limit reached | info |
Check Fuses | Fuse open (668xA series only) | Internal fuse failure detected | error |
Check Fuses | Power supply fuse is open | Excessive current or internal fault | error |
CV | Output in constant voltage mode | Output open or load draws minimal current | info |
Dew | High moisture inside power supply (668xA series only) | Humidity sensor detects excessive moisture | warning |
Dis | Power supply output is disabled | Output disable command or protection activation | warning |
Dis | Output disabled or non-functional | Output control failure or protection activation | error |
E1 | FP RAM error | Front panel RAM checksum failure during self-test | error |
E1 | Front panel RAM | Self-test failure | error |
E10 | Secondary 5V ADC read error | 5V ADC reading failure during self-test | error |
E10 SEC 5 V | Secondary interface 5 volt read back test failed | Power-on self-test; also *TST? for 664xA & 665xA models | error |
E11 | Secondary thermistor read error | Thermistor reading failure during self-test | error |
E11 TEMP | Ambient temperature read back test failed | Power-on self-test; also *TST? for 664xA & 665xA models | error |
E12 | Secondary VDAC/IDAC error | VDAC or IDAC functionality failure during self-test | error |
E12 DACS | CV or CC DAC tests failed | Power-on self-test | error |
E1 FP RAM | Front panel RAM test failed | Power-on self-test detection | error |
E2 | FP ROM checksum error | Front panel ROM checksum failure during self-test | error |
E2 FP ROM | Front panel ROM test failed | Power-on self-test; also *TST? for 664xA & 665xA models | error |
E3 | EEPROM error | EEPROM checksum failure during self-test | error |
E3 EE CHKSM | Front panel EEPROM checksum test failed | Power-on self-test; also *TST? for 664xA & 665xA models | error |
E4 | Primary external RAM error | Primary XRAM failure during self-test | error |
E4 PRI XRAM | Primary interface external RAM test failed | Power-on self-test; 664xA & 665xA models only | error |
E5 | Primary internal RAM error | Primary IRAM failure during self-test | error |
E5 PRI IRAM | Primary interface internal RAM test failed | Power-on self-test; 664xA & 665xA models only | error |
E6 | Primary ROM checksum error | Primary ROM checksum failure during self-test | error |
E6 PRI ROM | Primary interface ROM test failed | Power-on self-test; also *TST? for 664xA & 665xA models | error |
E7 | GPIB communication error | Serial port communication failure during self-test | error |
E7 | GPIB serial port | Self-test failure | error |
E7 GPIB | GPIB interface test failed | Power-on self-test | error |
E8 | Secondary RAM error | Secondary RAM failure during self-test | error |
E8 SEC RAM | Secondary interface RAM test failed | Power-on self-test | error |
E9 | Secondary ROM checksum error | Secondary ROM checksum failure during self-test | error |
E9 SEC ROM | Secondary interface ROM test failed | Power-on self-test; also *TST? for 664xA & 665xA models | error |
EE WRITE ERR | EEPROM write failure | Attempt to write to EEPROM during calibration or configuration | error |
Err | Error message generated from remote operation | Remote control command error | error |
OC | Overcurrent protection active | Output current exceeds protection threshold | warning |
OCP | Overcurrent protection circuit is operational | OCP circuit self-test or monitoring | info |
OV | Overvoltage protection active | Output voltage exceeds limit | error |
Prot | Protection circuit has shut down the power supply | Overcurrent or overvoltage condition | error |
Prot | Power supply shutdown due to protection circuit activation | Protection limit exceeded | critical |
Rmt | Power supply in remote control mode | Remote control enabled or GPIB addressed | info |
SBUF FULL | Serial buffer failure due to full condition | Serial buffer remains full; data not being processed | warning |
SBUF OVERRUN | Serial buffer failure due to overrun | Serial buffer exceeded capacity during reception | warning |
SBUV FULL | Serial buffer overflow | Message too long for buffer capacity | warning |
SECONDARY DN | Serial data line failure on main board or isolator board | Secondary interface communications loss | error |
SERIAL DOWN | Serial data line failure on GPIB or isolator board | Communications failure during operation | error |
SERIAL DOWN | GPIB/front panel communication failure | GPIB unable to communicate with front panel | error |
SERIAL TIMEOUT | Serial data line failure on GPIB or isolator board | Communications timeout during operation or self-test | error |
Shift | Shift key has been pressed | Shift key (blue unlabeled key below Recall) pressed | info |
Shift | Shift key pressed (blue unlabeled key below Recall) | Shift key depression detected | info |
SRQ | Power supply requesting service from controller | Service request condition detected | warning |
STK OVERFLOW | Front panel stack overflow | Stack buffer exceeded during front panel operation | error |
UART FRAMING | UART byte framing error | UART received byte with framing error | warning |
UART OVERRUN | UART failed due to input buffer overrun | Serial data arrival faster than processor can handle | warning |
UART PARITY | UART failed due to parity error | Serial communications parity check failure | error |
UART PARTY | UART byte parity error | Serial communication error | error |
Unr | Output voltage/current not regulated to setpoint | Output not in CV or CC mode | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI